WIKA CTM9100-150 Specifications

Temperature multi-function calibrator Model CTM9100-150
for further approvals see page 3
Applications
Testing and calibration of all types of temperature measuring instruments
Reference instrument for works laboratories for the calibration of thermometers
Also suitable for on-site calibration
Special features
Multi-function instrument with four controller parameter sets
Calibration with an external reference
Low weight and compact design
Simple operation
Temperature multi-function calibrator model CTM9100-150
Calibration
WIKA data sheet CT 41.40
Description
Versatile in application
Nowadays, fast and simple testing of thermometers is a “must” when it comes to the operational safety of machines and plants. The portable calibrators of the CTx9100 family are particularly suited for on-site calibrations and extremely user-friendly. Due to their compact design and their low weight, the instruments can be taken and used almost anywhere.
Using a dry-well calibrator or a micro calibration bath to calibrate either surface thermometers or non-contact
thermometers does not reect the application and can result
in false values. In these cases, the model CTM9100-150 temperature multi-function calibrator should be used.
With this temperature multi-function calibrator, in the temperature range from -20 ... +150 °C [-4 ... +302 °F], you can cover not only the common functions, but also, with special inserts, this can be used as a surface temperature calibrator and an infrared black body. Like having four instruments in one!
WIKA data sheet CT 41.40
08/2018
Easy to use
The temperature multi-function calibrator combines four application possibilities in one instrument. The simple instrument design enables fast and simple switching, back and forth, between individual applications. Changing between the four possibilities can be made very easily using one of the four keys and the toggle switch on the front.
It is possible to calibrate nearly every thermometer with
the dierent inserts, even contact (TC, Pt, surface) or non-contact (infrared) thermometers.
The calibration temperature, adjusted simply using two keys on the controller, can be very quickly controlled. The actual and set temperature can be displayed simultaneously on a large 4-digit, high-contrast LC display. Thus reading errors are virtually eliminated.
Page 1 of 7
Specications Model CTM9100-150
Display
Temperature range -20 ... +150 °C [-4 ... 302 °F]
-35 ... +165 °C [-31 ... 329 °F] Use as a micro calibration bath
Accuracy
Stability
1)
2)
Resolution 0.01 to 100 °C, then 0.1
Temperature control
Heating time depending upon usage and area of application
Cooling time depending upon usage and area of application
Stabilisation time
3)
Insert
Immersion depth 150 mm [5.91 in]
Insert dimensions Ø 60 x 170 mm [Ø 2.36 x 6.69 in]
Insert material Aluminium
Voltage supply
Power supply AC 100 ... 240 V, 50/60 Hz
Power consumption 400 VA
Fuse 6.3 A slow blow fuse
Power cord for Europe, 230 V
Communication
Interface RS-485
Case
Dimensions in [W x D x H] 215 x 305 x 425 mm [8.46 x 12.0 x 16.73 in]
Weight 12 kg [26.5 lbs]
1) Is dened as the measuring deviation between the measured value and the reference value.
2) Maximum temperature dierence at a stable temperature over 30 minutes.
3) Time before reaching a stable value.
±0.2 K ±0.3 K ±1 K ±1 K
±0.05 K ±0.05 K ±0.2 K ±0.2 K
depending upon usage and area of application
Use as a micro calibration bath Use as a dry-well calibrator Use as an infrared black body Use as a surface temperature calibrator
Use as a micro calibration bath Use as a dry-well calibrator Use as an infrared black body Use as a surface temperature calibrator
The measurement uncertainty is dened as the total measurement uncertainty (k = 2), which contains the following shares:
accuracy, measurement uncertainty of reference, stability and homogeneity.
WIKA data sheet CT 41.40 ∙ 08/2018 Page 2 of 7
Approvals
Logo Description Country
EU declaration of conformity
EMC directive
EN 61326, emission (group 1, class B) and interference immunity (industrial application)
Low voltage directive EN 61010, safety requirements for electrical equipment for measurement, control and laboratory use
RoHS directive
EAC
EMC directive
Low voltage directive
GOST
Metrology, measurement technology
KazInMetr
Metrology, measurement technology
- MTSCHS Permission for commissioning
BelGIM
Metrology, measurement technology
European Union
Eurasian Economic Community
Russia
Kazakhstan
Kazakhstan
Belarus
Certicates
Certicate
Calibration Standard: 3.1 calibration certicate per DIN EN 10204
Option: DKD/DAkkS calibration certicate
Recommended recalibration interval 1 year (dependent on conditions of use)
Approvals and certicates, see website
WIKA data sheet CT 41.40 ∙ 08/2018 Page 3 of 7