WIKA CTH7000 Specifications

Hand-held thermometer Model CTH7000
Applications
Precision thermometer for very accurate temperature measurements in a range of -200 ... +962 °C
Self-contained, complete system also suitable for on-site measurements/calibrations
Calibration technology
WIKA data sheet CT 55.50
Special features
Temperature range: -200 ... +962 °C
Stability: < 0.005 °C per year
Dual inputs measuring platinum resistance thermometers
(PRTs) and thermistors single or dierential measurement
High long-term stability
Description
The model CTH7000 hand-held thermometer is a high performance 2-channel thermometer for Pt100 and thermistor probes. The hand-held thermometer CTH7000 is a step up in measurement accuracy with a battery power.
The CTH7000 can handle all needs, with accuracies and resolutions normally associated with a bench top thermometer. Two inputs give direct temperatures from Pt100 or thermistor probes or can display the temperature
dierence between them. Measurements can be logged
directly to memory or, use the USB port to control and data log with the PC.
The CTH7000 will display a statistical analysis of measurements, with average value, min and max values and standard deviation. Because the CTH7000 will display in ohms as well as temperature, it can even be used to calibrate
Hand-held thermometer model CTH7000
one probe against another. Most instruments require to enter
calibrated probe data into memory to optimise measurement accuracy, which can mean a lot of data entering if you change probes regularly.
With ASL's SMART connector on the probes, you only need
to store the data once - in the connector! The calibration data stays with the probe - permanently. The probe can even be used on another CTH7000 without any further action.
The SMART connector saves time and reduces error. If you
have existing calibrated or uncalibrated probes, no problem,
CTH7000 automatically registers if a probe is SMART or
normal.
Data sheets showing similar products: Hand-held thermometer; models CTH6500, CTH65I0; see data sheet CT 55.10 Reference thermometer; model CTP5000; see data sheet CT 61.20
Page 1 of 6WIKA data sheet CT 55.50 ∙ 04/2014
Specications
Model CTH7000
Hand-held thermometer
Probe types Pt100 and Pt25 resistance thermometers, NTC thermistors
Measuring inputs 2
Data entry format ITS 90 and CVD for calibrated probes; or EN 60751 for uncalibrated probes
Measuring ranges
Probe current DC 1 mA (Pt100); Auto-select 1 mA, 10 µA and 3 µA for thermistors
Temperature range -200 ... +962 °C, depending on thermometer probe
Accuracy
Resistance thermometers
NTC thermistors 0 ... 400 Ω
1) The accuracy in K denes the deviation between the measured value and the reference value. (Only valid for indicating instruments.)
1)
Digital indicator
Display
Screen Large 4 1/2-digit two-line LCD screen with backlighting
Resolution 0.001 °C
Functions
Memory approx. 8,000 values
Functions via key press Statistical analysis: Min/Max, average and standard deviation, Hold, Zero, one-shot measure-
Real-time clock integrated clock with date and year
Voltage supply
Power supply Nickel-Metal Hydride (Ni-MH) rechargeable battery
Battery life approx. 20 hours of operation
Permissible ambient conditions
Operating temperature 15 ... 25 °C
Storage temperature -20 ... +50 °C
Communication
Interface USB
Case
Dimensions 232 x 97 x 53 mm (L x D x W)
Weight 500 g
0.015 K
400 Ω ... 50 kΩ 50 ... 400 kΩ
ment, measurement log
Charging socket built-in low battery indicator
±0.006 Ω
±0.01 % of reading ±0.02 % of reading
CE conformity, certicates
CE conformity
EMC directive 2004/108/EC, EN 61326 emission (group 1, class B) and interference immunity (portable test
and measuring equipment)
Certicates
Calibration Standard: 3.1 calibration certicate per DIN EN 10204
Option: DKD/DAkkS calibration certicate
Recommended calibration interval 1 year (depending on the conditions of usage)
Approvals and certicates, see website
Page 2 of 6 WIKA data sheet CT 55.50 ∙ 04/2014
Temperature probes
Resistance thermometer
Spring strain relief
2 metres
Engraved line
300 mm maximum insertion depth
350 or 450
Resistance thermometer (immersion probe) Temperature range
Pt100, d = 6 mm, l = 350 mm -50 ... +250 °C
Pt100, d = 6 mm, l = 350 mm -70 ... +450 °C
Pt100, d = 6 mm, l = 450 mm -70 ... +650 °C
ø6 or ø7
nominal
Centreline of sensor
Resistance thermometer
Brazed weld
3 metres
Probe type (immersion probe) Temperature range
Pt100, d = 3 mm, l = 30 mm -50 ... +200 °C
NTC thermistor
Nylotron handle
Probe type (immersion probe) Temperature range
NTC thermistor R25 = 10 KΩ, d = 4 mm, l = 300 mm -50 ... +125 °C
Page 3 of 6WIKA data sheet CT 55.50 ∙ 04/2014
Resistance thermometer connection, 4-wire (5-pin DIN connector)
View towards front panel connector
Shield
Options
With DIN plug or SMART plug
With ASL's SMART connector on the probes, storing the data
is needed only once - in the connector! The calibration data stays with the probe - permanently. It can even be used on another instrument without any further action.
Viewed from top panel
Shield
The SMART connector saves time and reduces error.
If there are existing calibrated or uncalibrated probes, no problem, CTH7000 automatically registers if a probe is
SMART or normal
Features of the hand-held thermometer
Simple handling
Large display with dual temperature display
Min/Max value for monitoring of temperature limits
Mean value function for statistical evaluation
Selectable channel can be switched o to improve the
clarity of the display data
Recording and visualisation of temperature cycles with the help of the ULog software
Data logger
The large graphic LCD screen is your direct link to the
instrument, presenting you with the measurement results and information or menus to set and control the instrument.
The LCD screen is designed for viewing without any
reections under normal lighting. A backlight is provided in case of darker ambient conditions.
1
4
3
2
1 2 Connection ports for temperature probe inputs
2 USB connection port for PC
3 Keypad
4 Large LCD display
Page 4 of 6 WIKA data sheet CT 55.50 ∙ 04/2014
Operation
The keys are used to select the various menu options.
Generally, no more than two menu levels are used. A few
(infrequently used options) require three menu levels. Operation is simple once you are familiar with instrument.
Key Function
☼ ▲ ▼ ◀ ▶
A/B 0 II
Connection ports for inputs
There are two input channels; two 5-pin DIN input-sockets
are located at the rear of the instrument. These are designed
to take either platinum resistance thermometer (PRT) or
thermistor probes. Channel A is colour-coded red. Channel B is colour-coded blue.
Menu display
Backlight
Up arrow
Down arrow
Left arrow
Right arrow
Clear (view statistics)
Enter
A/B channel select
Zero (oset) reading
Hold reading
ON/OFF
Keypad of CTH7000
2 Connection port for temperature probe inputs
Page 5 of 6WIKA data sheet CT 55.50 ∙ 04/2014
Scope of delivery
Model CTH7000 hand-held thermometer incl. battery
charger and USB cable
3.1 calibration certicate per DIN EN 10204
Choice of temperature probes
Option
DKD/DAkkS calibration certicate
UKAS calibration certicate
Accessories
Temperature probes
Immersion probe
Customer-specic probes are available on request
Voltage supply
Replacement battery charger with multiple power plugs
Spare battery pack to clip onto rear of CTH7000
Test case
Transport case, robust
Software
ULog
Ordering information
Model / Probe at input 1 / Connection input 1 / Probe at input 2 / Connection input 2 / Calibration / Software / Carrying case /
Additional order information
© 2013 WIKA Alexander Wiegand SE & Co. KG, all rights reserved.
The specications given in this document represent the state of engineering at the time of publishing. We reserve the right to make modications to the specications and materials.
Page 6 of 6 WIKA data sheet CT 55.50 ∙ 04/2014
WIKA Alexander Wiegand SE & Co. KG
Alexander-Wiegand-Straße 30 63911 Klingenberg/Germany Tel. +49 9372 132-0 Fax +49 9372 132-406 info@wika.de www.wika.de
04/2014 GB