Hand-held thermometer, precision version
Model CTH6500
Model CTH65I0, Ex version
for further approvals
see page 2 - 3
Applications
■
Calibration of thermometers
■
Temperature measurement for the purposes of quality
assurance
■
Measurements in service and maintenance applications
■
Long-term monitoring and online documentation
Special features
Calibration technology
WIKA data sheet CT 55.10
■
High accuracy of 0.03 K with Pt100
■
One- and two-channel versions
■
Connection possibilities for various probe types
■
Intrinsically safe version Ex ib IIB T4 Gb
Description
The all-purpose model CTH6500 hand-held thermometer, for
superior mobile temperature measurement, is notable for its
precision, exibility and ease of handling.
In addition to Pt100 resistance thermometers, it can
also process signals from typical thermocouples. Thus
temperatures from -200 ... +1,500 °C (-328 ... +2,732 °F)
can be measured. The explosion proof version is only for
measuring Pt100 resistance thermometers available.
Through its high accuracy of 0.03 K in ranges from
-100 ... +150 °C (-148 ... +302 °F), this instrument can
also be used as a reference instrument in biotechnology,
pharmaceutical and food industries. The CTH6500 is thus
also ideal for all service and maintenance tasks.
Hand-held thermometer model CTH6500
In this way it is possible to reduce measuring errors to a
minimum and ensure a high display accuracy.
Additional elds of application
The instrument has been primarily designed for temperature
measurement, though it can, with the appropriate probes,
also be used for:
■
Humidity measurement with a combined temperaturehumidity probe
■
Flow measurement from 0.1 ... 40 m/s with a vane sensor
The calibration and adjustment possibilities above are also
applicable to these measurement parameters.
Low-drift measuring ampliers ensure small measurement
errors, while easy-to-use adjustment features considerably
simplify adjustments and calibrations:
■
Calibration by code for fast setting of standard probes via
identication numbers
■
Physical calibration of probe and display at one, two or
three dierent temperatures
∙ 09/2017
Data sheets showing similar products:
Hand-held thermometer; models CTH6300, CTH63I0; see data sheet CT 51.05
Page 1 of 6WIKA data sheet CT 55.10
Specications
Hand-held thermometer Model CTH6500 Model CTH65I0
Probe types Pt100, thermocouples, humidity, ow Pt100
Measuring inputs 1 or 2 1 or 2
Measuring ranges
Pt100 -200 ... +600 °C (-328 ... +1,112 °F) -200 ... +600 °C (-328 ... +1,112 °F)
Thermocouples -200 ... +1,500 °C (-328 ... + 2,732 °F) -
Humidity 0 ... 100 % r. h. -
Flow 0 ... 40 m/s -
Accuracies
Resistance thermometer
type Pt100
Thermocouple types K, J, L, N
and T
Thermocouple types R and S 1 K + 0.1 % of reading -
Humidity 1.5 % r. h. -
Flow 0.5 % of full-scale value -
Digital indicator
Display
Screen Large 4 1/2-digit 2-line LC display with backlighting
Resolution 0.01 K up to 200 °C, then 0.1 K
Functions
Measuring rate 4/s (“fast”); 1/s (“slow”)
Memory Min./Max.
Functions via key press Min./Max. memory, Hold, Tare, Zero-point adjustment
Real-time clock integrated clock with date
Voltage supply
Power supply DC 9 V battery or rechargeable battery
Battery life approx. 20 hours of operation with battery
Permissible ambient conditions
Operating temperature 0 ... 40 °C (32 ... 104 °F)
Storage temperature -10 ... +50 °C (14 ... 122 °F)
Communication
Interface USB via interface cable
Case
Material impact-resistant ABS plastic, transparent screen
Dimensions (L x W x H) 200 x 93 x 44 mm (7.87 x 3.66 x 1.73 in)
Weight 350 g (0.77 lbs.)
0.03 K for -50 ... +199.99 °C (-58 ... +394.98 °F)
0.05 K for -200 ... -50.01 °C (-328 ... -58.02 °F)
otherwise 0.05 % of reading
0.2 K for 0 ... 200 °C (32 ... 392 °F)
0.5 K for 200 ... 1,000 °C (392 ... 1,832 °F)
1 K above 1,000 °C (1,832 °F)
0.03 K for -50 ... +199.99 °C (-58 ... +394.98 °F)
0.05 K for -200 ... -50.01 °C (-328 ... -58.02 °F)
otherwise 0.05 % of reading
-
Approvals
Logo Description Land
EU declaration of conformity for CTH6500
■
EMC directive
EN 61326 emission (group 1, class B) and interference immunity (industrial application)
■
RoHS directive
Page 2 of 6 WIKA data sheet CT 55.10 ∙ 09/2017
European Union