WIKA 130.15 User Manual

Operating Instructions
GB
Pressure Gauges
Examples:
Model 131.11.063
Model 131.11.050
Notes in accordance with Pressure Equipment Directive 97/23/EC
The pressure gauges are "pressure
accessories" in accordance with article 1, paragraph 2.1.4
The volume of the pressure bearing
housings of WIKA pressure gauges is < 0.1 L
The pressure gauges carry the
CE marking for fluid group 1G in accordance with annex 2, table 1 when their permissible working pressure exceeds 200 bar
Pressure gauges that do not carry
the CE marking are manufactured in accordance with article 3, paragraph 3 "sound engineering practice".
Applied standards
EN 837-1 Bourdon tube pressure
EN 837-2 Selection and installation
EN 837-3 Diaphragm and capsule
Specifications: see data sheet at www.wika.de
Subject to technical modifications.
© WIKA Alexander Wiegand SE & Co. KG 2009
WIKA Alexander Wiegand SE & Co. KG
Alexander-Wiegand-Straße 30 63911 Klingenberg • Germany Tel. (+49) 93 72/132-0 Fax (+49) 93 72/132-406 E-Mail info@wika.de www.wika.de
gauges, dimensions, metrology, requirements and testing
recommendations for pressure gauges
pressure gauges, dimen­sions, metrology, require­ments and testing
1. Safety instructions
WARNING!
Before installation, commissioning and
operation, ensure that the appropriate pressure gauge has been selected in terms of measuring range, design and suitable wetted material (corrosion) for the specific measuring conditions. In order to guarantee the measur­ing accuracy and long-term stability specified, the corresponding load limits are to be observed.
Only qualified persons authorised by the plant manager are permitted to install, maintain and service the pressure gauges.
For hazardous media such as oxygen, acetylene, flammable or toxic gases or liquids, and refrigeration plants, compressors, etc., in addition to all standard regulations, the appropriate existing codes or regulations must also be followed.
After an external fire pressure media can leak out particularly at soft solder joints. All gauges have to be checked and, if necessary, replaced before recommissioning the plant.
Non-observance of the respective regulations can result in serious injury and/or damage to equipment.
2. Mechanical connection
In accordance with the general techni­cal regulations for pressure gauges (e.g. EN 837-2). When screw-fitting the gauges the force required for this must not be applied through the case or terminal box but just through the spanner flats (with suitable tool) provided for this purpose.
Mounting with
spanner
Correct sealing of pressure gauge connections with parallel thread shall be by means of a suitable sealing ring, sealing washer or WIKA profile seals. The sealing of tapered threads (e.g. NPT threads) is made by providing the thread with additional sealing material like, for example, PTFE tape (EN 837-2).
Spanner flats
Sealing face
The torque depends on the seal used. Gauge connection by means of a clamp socket or a union nut is recom­mended to simplify correct orientation of the gauge. When a blow-out device is fitted to a pressure gauge, it shall be protected against blocking by debris and dirt.
Sealing in the thread
With safety pressure gauges (see dial symbol k) it has to be ensured that the free space behind the blow-out back will be at least 15 mm.
After mounting, set the compensating valve (if available) from CLOSE to OPEN.
Requirements on the installation point
If the line to the gauge is not adequately stable for a vibration-free mounting, a measuring instrument support (possibly via a flexible capil­lary) should be used for fastening. If vibrations cannot be avoided by a suitable installation, liquid-filled gauges should be used. The instru­ments should be protected against coarse dirt and wide fluctuations in ambient temperature.
3. Admissible ambient and operating temperatures
When mounting the pressure gauge it has to be ensured that, taking the influence of convection and heat radiation into consideration, no upper or lower deviation from the permissi­ble ambient and medium temperatures can occur. The influence of tempera­ture on the indication accuracy is to be observed.
4. Storage
The pressure gauge should remain in its original packing until installation. The gauge should be protected from external damage during storage. Storage temperature range:
-40 ... +70 °C.
Protect the gauges from humidity and dust.
5. Maintenance / Repairs
The instruments are maintenance­free. Tests should be carried out on a regular basis to guarantee the measur­ing accuracy of the pressure gauge. The tests or recalibrations have to be carried out by qualified skilled person­nel with the appropriate equipment. When dismounting, close the compen­sating valve (if available).
WARNING!
Residual media in dismounted pressure gauges can result in a risk to persons, the environ­ment and equipment. Take sufficient precautionary measures.