Wet Labs ac-9 Plus Owner's Manual

DH-4 Data Handler
User’s Guide
The user’s guide is an evolving document. If you find sections that are unclear, or missing information, please let us know. Check our website periodically for updates.
WET Labs, Inc. PO Box 518 Philomath, OR 97370 (541) 929-5650 wetlabs@wetlabs.com
DH-4 Revision E5 19 June 2006
Attention!
Return Policy for Instruments with Anti-fouling Treatment
WET Labs cannot accept instruments for servicing or repair that are treated with anti-fouling compound(s). This includes but is not limited to tri-butyl tin (TBT), marine anti-fouling paint, ablative coatings, etc.
Please ensure any anti-fouling treatment has been removed prior to returning instruments to WET Labs for service or repair.
DH-4 Revision E5 19 June 2006
DH-4 Warranty
This unit is guaranteed against defects in materials and workmanship for one year from the original date of purchase. Warranty is void if the factory determines the unit was subjected to abuse or neglect beyond the normal wear and tear of field deployment, or in the event the pressure housing has been opened by the customer.
To return the instrument, contact WET Labs for a Return Merchandise Authorization (RMA) and ship in the original container. WET Labs is not responsible for damage to instruments during the return shipment to the factory. WET Labs will supply all replacement parts and labor and pay for return via 3rd day air shipping in honoring this warranty.
Shipping Requirements for Warranty and Out-of-warranty Instruments
1. Please retain the original shipping material. We design the shipping container to meet
stringent shipping and insurance requirements, and to keep your meter functional.
2. To avoid additional repackaging charges, use the original box (or WET Labs-approved
container) with its custom-cut packing foam and anti-static bag to return the instrument.
If using alternative container, use at least 2 in. of foam (NOT bubble wrap or
Styrofoam “peanuts”) to fully surround the instrument.
Minimum repacking charge for ac meters: $50.00.
3. Clearly mark the RMA number on the outside of your shipping container and on all
packing lists.
4. Return instruments using 3rd day air shipping or better: do not ship via ground.
DH-4 Revision E5 19 June 2006
Table of Contents
1. Getting Started....................................................................................................... 1
1.1
Install Host Program ....................................................................................... 1
1.2
Verify Electrical Functionality .......................................................................... 1
1.3
Verify System Operation ................................................................................. 2
2. Setting Operational Modes ................................................................................... 7
2.1
Profiling Mode ................................................................................................. 8
2.2
Moored Mode ................................................................................................ 13
2.3
Real-time Mode ............................................................................................ 19
3. Reference: Status Displays................................................................................. 25
3.1
Setup Status ................................................................................................. 25
3.2
Host Status ................................................................................................... 25
3.3
Logger Port Setup ......................................................................................... 25
3.4
Logger Status ............................................................................................... 26
4. Reference: Pull-down Menus .............................................................................. 29
4.1
Files .............................................................................................................. 29
4.2
Modes ........................................................................................................... 29
4.3
Commands ................................................................................................... 29
4.4
Logger_Port .................................................................................................. 32
5. Reference: Window Tabs .................................................................................... 35
5.1
Port Setup ..................................................................................................... 35
5.2
Logger Control .............................................................................................. 41
5.3
Data Status ................................................................................................... 47
5.4
Upload Logger Data ...................................................................................... 48
5.5
Real Time Data ............................................................................................. 50
5.6
Recording: OFF/ON ...................................................................................... 51
5.7
PC-Logger .................................................................................................... 52
6. Hardware .............................................................................................................. 55
6.1
Specifications ................................................................................................ 55
6.2
Standard DH-4 Connector Configuration ....................................................... 56
6.3
Standard Data Format .................................................................................. 58
6.4
Maintenance ................................................................................................. 58
Appendix A: Configuration Worksheets ............................................................... 59
Sample Worksheet .................................................................................................... 59
Blank Custom Configuration Worksheet .................................................................... 62
Appendix B: Real-Time Data Format ................................................................... 64
Appendix C: File Handling .................................................................................... 65
DH-4 Revision E5 19 June 2006 i
1. Getting Started
The Data Handler (DH-4) and its associated software, WET Labs Host, controls four, six or eight input-output ports for a variety of serial or analog instruments, providing for the simultaneous collection, time stamping, storage and merging of data streams from all of the ports.
Depending on the capabilities of the host computer, output from one to all of the instruments can be viewed in real time through the host software.
The DH-4 (data logger) is delivered with the following components:
Test cable
Dummy plugs with lock collar for each bulkhead connector
This user’s guide
Custom Configuration Sheet
WET Labs Host software on CD
WET Labs File Archive Processing (WAP) software on CD
WAP User’s Guide
1.1 Install Host Program
Insert the WET Labs Host software CD into the host computer and copy the contents to a desired location on the host computer. The WAP software may also be installed at this time. Refer to the WAP User’s Guide for details on using this data processing software.
1.2 Verify Electrical Functionality
We recommend following the steps below to verify the system’s electrical functionality when you first receive your DH-4. You will need the following items:
A clean, solid lab table or work bench.
A multimeter.
The DH-4.
Test cable (includes terminations for an RS-232 connection into a host computer,
power leads that connect to a power supply, and a connector to the DH-4).
Any meter(s) that will be used with the DH-4.
A 12–15 volt power supply.
A computer with WL Host installed.
1.2.1 Check Test Cable Voltage:
1. Connect the power leads to a 12–15 volt power supply (safe operating range).
The black lead is the V+ lead. Use a multimeter to check the input power before connecting the cable to the instrument.
DH-4 Revision E5 19 June 2006 1
2. With the power supply turned on, touch the multimeter ground probe to socket 1
on the test cable connector. Connect the “hot” probe to pin 4 (the pin directly opposite from pin 1). You should measure 12–15 volts across these two pins.
3. Turn the power supply off.
1.3 Verify System Operation
1. If necessary, attach the test cable power leads to a stable power source that supplies
12–15 VDC. Make sure the polarity is correct before switching on the power supply.
2. Connect the RS-232 connector to the desired serial port of the host PC.
Note that if your instrument is sending data in the RS-485 format, an RS-485 to RS­232 converter is required to allow proper operation.
WARNING!
Power input on this unit is diode-protected from reverse polarity power-up, but this is not
100 percent insurance against damaging the meter, nor will it protect it from over-voltage.
3. Plug the submerged (wet) end of the test cable into the DH-4.
Applying electrical grade silicone spray or an equivalent to the rubber base of the instrument bulkhead connector makes the plug insertion easier and provides greater assurance of a good seal. Do NOT use stopcock greases or products like WD-40. These will cause de-lamination of the rubber from the metal surface of the bulkhead connector. Use a connector lock ring if one is available.
Tips
Refer to the Custom Configuration Sheet that ships with your DH-4 for a summary of factory configuration settings.
The Configuration Sheet in Appendix A refers to the example setup used in this manual.
4. Ensure the DH-4 is connected to the host computer and power supply as above.
5. Connect any additional meter(s) to the DH-4. It may help to refer to the Custom
Configuration Sheet that ships with each DH-4: when possible, configurations settings specific to your system are set at the factory and documented on the Custom Configuration Sheet.
6. Start WET Labs Host program.
7. Turn on power to the DH-4.
2 DH-4 Revision E5 19 June 2006
8. The first time WET Labs host is started, the
program screen will be mostly empty, with only a Configure Host Port window appearing.
9. Input the COM port and the baud rate (default 9600). The red backgrounds will turn white. Close the window. Both time indicators on the left of the window (Host Status and Logger Status) will increment, indicating the DH-4 and PC are communicating.
10. Go to the Modes pull-down menu and select Profile Logging.
11. Go to the now-visible
Port Setup
tab and press
Get Setup
to load the logger’s settings
into the Host program.
12. Review the Serial Port Type, associated Port Setup baud rate, Analog Ports, and Pump Ports settings. These settings can also be found on the Custom Configuration Sheet that ships with your data logger. See Reference section 5.1 for details about Port Setup options.
Note
For purposes of illustration in this user’s guide, WET Labs Host software runs the following:
Host Port: DH-4
Port 1: ac-s
Port 2: ECO VSFS, Volume Scattering Function Meter with Bio-wiper
Port 3: CTD
Port 4: Pump
Port 5: ECO FLNTUS, fluorometer-turbidity sensor with Bio-wiper
DH-4 Revision E5 19 June 2006 3
Port Setup tab settings
13. Select the Logger
Setup
tab to view the factory-set logger control settings. Refer
to section 5.2 for details on Logger Control configuration options.
Logger Control tab settings
14. To save these settings on the host computer: Select the Save Setup Files button and in the resulting window, name the setup and click OK. This will allow you to retrieve the factory settings in the future.
4 DH-4 Revision E5 19 June 2006
15. There are two options to verify that the meters connected to DH-4 are functional and communicating. Option 1 allows checking multiple meters. Option 2 tests one meter at a time, but also provides baud rate information.
Option 1: test multiple meters
In the Logger Control tab select Begin Sample. Sampling begins with the program cycling through the preset sampling parameters. For example, there will be 5 seconds for Delay Before Start and Pre-
Warm Up Flush, and 10 for Meter Warm­Up before the system begins active sampling
for 5 minutes.
Select the Data Status tab. The status of each port is displayed. Depending on the operation the data logger is performing, either the Bytes/Second and/or the File Size will increment. While the logger is performing a pre-sampling warm­up or flush operation, only the bytes/second will increase. When sampling begins, the file size will increment as well.
Return to the Logger Control tab and select Stop.
Option 2: test a single meter
Select Commands/Test One Logger Port
DH-4 Revision E5 19 June 2006 5
a. Select the data port (and thus instrument) you
wish to test.
b. Select the instrument’s baud rate. c. Select the host computer’s baud rate. d. Select Start Test. The Real_Time Data window
appears, showing scrolling data.
e. Select Stop Test when communication is
verified.
16. To upload and save the test
run from above, select the Upload Logger Data tab, then Get DH-4 Directory. Select a Run # to upload, then select Upload Logger Data. Choose a location to store the data. The upload status bar indicates file size and status of upload progress.
17. Select Clear DH-4 Memory to erase the data from the DH-4.
18. Exit the program by selecting Exit under the File menu.
Tip:
6 DH-4 Revision E5 19 June 2006
We highly recommend you experiment with changing and retrieving settings to get a feel for the flexibility of the host software before deploying your system.
2. Setting Operational Modes
WET Labs Host can operate in several modes. Which you choose, of course, depends on your application. This section describes setup for the three most frequently used modes of operation:
1. Profile Logging Mode—typically used to assess a column of water, with the logger and
other instruments moving vertically through the column.
2. Moored Logging Mode—typically used for stationary deployments. WET Labs host
provides settings options for low power, or “sleep” intervals, as well as flushing options. Data is stored at the logger.
3. Real Time Output—similar to profiling, but no data is stored to the logger; it is sent
directly to a host PC. Selecting the appropriate settings to save data in WET Labs Host are critical in this mode.
Less frequently used, and requiring an advanced understanding of WET Labs Host software:
Analog Output—Transmits ASCII data from analog meters in real time.
Advanced Logging/Output—Used only for custom instrument configurations, or
when a combination of modes is required.
DH-4 Revision E5 19 June 2006 7
2.1 Profiling Mode
This subsection details the steps and selections to execute a profile for the example system listed below:
Host Port: DH-4
Port 1: ac-s
Port 2: ECO VSFS, Volume Scattering Function Meter with Bio-wiper
Port 3: CTD
Port 4: Pump
Port 5: ECO FLNTUS, fluorometer-turbidity sensor with Bio-wiper
When possible, configuration settings specific to your system are set at the factory. The Custom Configuration Sheet that ships with your DH-4 represents a hard-copy summary of the DH-4’s flash settings as they appear in the host program’s setup windows.
2.1.1 Changing or Verifying Configuration Settings
Supply power to the DH-4. Start the host program. Select COM port and baud rate (default: 9600). Select Profile Logging from the Modes menu.
Port Setup tab:
1. Load configuration settings using one of the options below:
• To retrieve the DH-4’s current configuration, select Get Setup.
To load configuration settings saved to the host PC, select Open Setup
File, select the desired file, then select Send Setup to send those settings to the DH-4.
To load factory default settings, go to the Commands pull-down menu, then
Access Factory Settings > Load Default Flash Setup. These are also
available on the Custom Configuration Sheet that ships with your DH-4.
2. If necessary, change the settings under the Serial Port Type, Port Setup, and Analog Port options. Note that Setup Not Current will appear on a yellow banner on the status portion of the host program window if any settings are changed.
8 DH-4 Revision E5 19 June 2006
Tip
Sample Port Setup tab settings for profiling mode
Logger Control tab:
1. Check the Low Voltage Cut Off setting. Set to the highest value of the minimum voltage requirement for all the meters being used. For example, if you’re deploying an ac-9 (min. requirement 10.0 V), and an ECO fluorometer (min. requirement 7.5 V), set the Low Voltage Cut Off to 10.0 V.
2. Check Analog/Data Rate setting. Set to the fastest meter’s data rate. For example, if you’re deploying an ac-s (8 Hz), ac-9 (6 Hz), and an ECO fluorometer (1 Hz), set the Sample Rate to 8 Hz.
If the logger is being used with meters whose total output is more than 1024 bytes of data per sample, set the Analog/Data Rate higher than the meter data rate to prevent data loss. For example, if the total output is 2500 bytes/second, set the logger to at least 3 Hz.
Set the rate of data acquisition to the fastest meter’s data rate.
ac-s meters: ac-9 meters:
ECO meters:
Analog meters:
4 or 8 Hz 6 Hz 1 Hz 1 Hz
DH-4 Revision E5 19 June 2006 9
If your system uses analog meters:
Each analog bulkhead on the DH-4 can optionally support two analog channels. Therefore, each analog bulkhead has two analog definitions in the DH-4 host program. (Analog 1 and Analog 2 correlate with analog bulkhead 1, Analog 3 and Analog 4 correlates to analog bulkhead 2, etc.)
To use both channels on a given bulkhead, there must be a Y cable that originates from the DH-4 bulkhead and terminates at each of the supported instruments. If a Y cable is not available, each analog port will only support one analog device at the odd numbered Analog ID (Analog Bulkhead 1 = Analog 1, Analog Bulkhead 2 = Analog 3, etc.)
3. When the desired settings are input: Select Send Setup to send the configuration settings to the DH-4’s flash memory. To save the configuration settings to the host computer, select Save Setup File.
4. To begin sampling, select the Begin Sample button.
Sample Logger Control tab settings for profiling mode
10 DH-4 Revision E5 19 June 2006
2.1.2 Tracking Incoming Data
Select the Data Status tab to display the current data status of all meters. Refer to Reference section 5.3 for details of what each status line represents.
Viewing Data Status tab, profiling mode
To stop the sample, press Stop at the Logger Control tab.
2.1.3 Uploading Data
1. Go to the Port Setup tab and press Get Setup to ensure the most recent flash settings are loaded.
2. Under the Logger Port Setup combo box, choose the fastest data rate that the host cable and COM port can support. Three-meter cables can run at 115,200 baud.
3. Press Send Setup to send the baud rate change to the data logger. The logger will change to the meter’s baud rate and the host program will follow suit. There will be a brief pause as the logger and PC change their baud rates.
Occasionally, the logger and PC will lose track of each other while the baud rate change is occurring. This will be indicated by a red background for baud rate. Press Send Setup again. It may be necessary to toggle power to the data logger if re­sending the command doesn’t synchronize baud rates.
DH-4 Revision E5 19 June 2006 11
Upload Logger Data in profiling mode
4. At the Upload Logger Data tab, select Get DH-4 Directory to retrieve all the DH-4 file information.
5. Select one or more of the runs that appear in the Data Logger Run Directory. In the example above run 001 will be uploaded.
6. Select WL Archive from the combo box. This will create a data file that combines the data from the selected files.
7. Select the runs to be uploaded, then select Upload Data to upload them. The status section will update while uploading is occurring.
8. Selecting Cancel Upload will stop data upload. If multiple runs have been selected, you will be prompted to cancel all uploads or individual run uploads.
2.1.4 Clearing DH-4 Memory
Caution!
Once “Clear DH-4 Memory” is pressed, the data is not retrievable.
WARNING!
Turning off the data logger while memory clearing is in progress
corrupted data logger file system.
To clear the DH-4 memory, select the Clear DH-4 Memory button in the Upload Logger Data tab (above).
12 DH-4 Revision E5 19 June 2006
WILL
result in a
2.2 Moored Mode
This subsection details the steps and selections to execute a moored operation for the example system listed below:
Host Port: DH-4
Port 1: ac-s
Port 2: ECO VSFS, Volume Scattering Function Meter with Bio-wiper
Port 3: CTD
Port 4: Pump
Port 5: ECO FLNTUS, fluorometer-turbidity sensor with Bio-wiper
When possible, configuration settings specific to your system are set at the factory. The Custom Configuration Sheet that ships with your DH-4 represents a hard-copy summary of the DH-4’s flash settings as they appear in the host program setup windows.
2.2.1 Moored Logging Configuration Settings
Supply power to the DH-4; start the WET Labs Host program. Select COM port and baud rate (default: 9600). Select Moored Logging from the Modes menu.
Port Setup tab:
1. Load configuration settings using one of the options below:
• To retrieve the DH-4’s current configuration, select Get Setup.
To load configuration settings saved to the host PC, select Open Setup
File, select the desired file, then select Send Setup to send those settings to the DH-4.
To load factory default settings, go to the Commands pull-down menu,
then Access Factory Settings > Load Default Flash Setup. Note that these are also available on the Custom Configuration Sheet that ships with your DH-4.
2. If necessary, change the settings under the Serial Port Type, Port Setup, Analog Port options. Note that Setup Not Current will appear on a yellow banner on the status portion of the host program window if any settings are changed.
3. Check the Once Per Hour checkbox to allow instruments with a high current draw to run only during the first sample period after the beginning (“top”) of the hour. This allows you to obtain several samples per hour from instruments with lower current draws without the power penalty of instruments with a high current draw.
DH-4 Revision E5 19 June 2006 13
Sample Port Setup settings for moored mode
Logger Control tab:
1. Check the Low Voltage Cut Off setting. Set to the highest value of the minimum voltage requirement for all the meters being used. For example, if you’re deploying an ac-9 (min. requirement 10.0 V), and an ECO fluorometer (min. requirement 7.5 V), set the Low Voltage Cut Off to 10.0 V.
2. Check Analog/Data Rate setting. Set to the fastest meter’s data rate. For example, if you’re deploying an ac-s (4 or 8 Hz), ac-9 (6 Hz), and an ECO fluorometer (1 Hz), set the Sample Rate to 8 Hz.
If the logger is being used with meters whose total output is more than 1024 bytes of data per sample, set the Analog/Data Rate higher than the meter data rate to prevent data loss. For example, if the total output is 2500 bytes/second, set the logger to at least 3 Hz.
Tip
Set the rate of data acquisition to the fastest meter’s data rate.
ac-s meters: ac-9 meters:
ECO meters:
Analog meters:
If your system uses analog meters:
Each analog bulkhead on the DH-4 can optionally support two analog channels. Therefore, each analog bulkhead has two analog definitions in the DH-4 host program. (Analog 1 and Analog 2 correlate with analog bulkhead 1, Analog 3 and Analog 4 correlates to analog bulkhead 2, etc.)
To use both channels on a given bulkhead, there must be a Y cable that originates from the DH-4 bulkhead and terminates at each of the supported instruments. If a Y cable is not available, each analog port will only support one analog device at the odd numbered Analog ID (Analog Bulkhead 1 = Analog 1, Analog Bulkhead 2 = Analog 3, etc.)
4 or 8 Hz 6 Hz 1 Hz 1 Hz
14 DH-4 Revision E5 19 June 2006
3. Commands drop down menu:
Synchronize the DH-4 and host PC clocks. Select
Synchronize Logger Time
then
Synchronize Time Now
,
.
Note
Turning on “Automatic Time Sync will prevent you from seeing the DH-4 clock drift and should
not be used for moored applications.
4. When the desired settings are input: Select Send Setup to send the configuration settings to the DH-4’s flash memory. To save the configuration settings to the host computer, select Save Setup File.
5. To begin sampling, select the Begin Sample button.
Sample Logger Control settings for moored mode
DH-4 Revision E5 19 June 2006 15
Loading...
+ 51 hidden pages