This manual supplements the Concise Product manual(s) supplied with each instrument at
the time of shipment. Information in this installation, wiring and operation manual is subject to
change without notice.
Copies of this manual are available in electronic format on the West Instruments web site
(www.westinstruments.com) Printed versions are available from West or its agents at the
price published on the front cover.
Note:
It is strongly recommended that applications incorporate a high or low limit protective
device, which will shut down the equipment at a preset process condition in order to
prevent possible damage to property or products.
WARNING:
THE INTERNATIONAL HAZARD SYMBOL IS INSCRIBED ADJACENT TO THE REAR
CONNECTION TERMINALS. IT IS IMPORTANT TO READ THIS MANUAL BEFORE
INSTALLING OR COMMISSIONING THE UNIT.
WARNING:
THIS SYMBOL MEANS THE EQUIPMENT IS PROTECTED THROUGHOUT BY DOUBLE
INSULATION.
Products covered by this manual are suitable for Indoor use, Installation Category II,
Pollution category 2 environments.
This user guide covers all versions of the West ProVU4 controller.
These products are sold by West Instruments under the warranties set forth in the following
paragraphs. Such warranties are extended only with respect to a purchase of these products,
as new merchandise, directly from West Instruments or from a West Instruments distributor,
representative or reseller and are extended only to the first buyer thereof who purchases
them other than for the purpose of resale.
Warranty
These products are warranted to be free from functional defects in material and workmanship
at the time the products leave West Instruments factory and to conform at that time to the
specifications set forth in the relevant West instruction manuals sheet or sheets, for such
products for a period of three years.
THERE ARE NO EXPRESSED OR IMPLIED WARRANTIES, WHICH EXTEND BEYOND THE
WARRANTIES HEREIN AND ABOVE SET FORTH. NO WARRANTY IS MADE OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO THE
PRODUCTS.
Limitations
West shall not be liable for any incidental damages, consequential damages, special
damages, or any other damages, costs or expenses excepting only the cost or expense of
repair or replacement as described above. Products must be installed and maintained in
accordance with West Instruments instructions. There is no warranty against damage to the
product resulting from corrosion. Users are responsible for the suitability of the products to
their application.
For a valid warranty claim, the product must be returned carriage paid to the supplier within
the warranty period. The product must be properly packaged to avoid damage from
Electrostatic Discharge or other forms of harm during transit.
This manual is structured to give easy access to the information required for all aspects of
the installation and use and of the Graphical Controller. The main sections are shown here,
followed by a full table of contents.
Section 1:Introduction - A brief description of the product and it’s features.
Section 2:Installation - Unpacking, installing and panel mounting instructions.
Section 3:Field Upgrade Options - Installation of the plug-in option modules.
Section 4:Wiring Instructions - Guidance on good wiring practice, noise avoidance,
wiring diagrams and input/output connections.
Section 5:Powering Up - Powering up procedure and descriptions of displays & switches.
Section 7:Configuration & Use - Describes operating and configuration modes available.
These include Operation Mode; the Main and Configuration menus; the Easy Setup Wizard;
Supervisor Mode; Automatic tuning; Product and Service Information. Also available on
some models are menus to setup and use the USB, Data Recorder and Profiler features.
Section 8:The USB Interface Option – Describes uploading or download of instrument
settings, profiles or recorder log to a USB memory stick.
Section 9:The Data Recorder Option – Describes the Data recorder feature. This allows
process data to be stored in to memory for later download and analysis.
Section 10:The Profiler Option – Describe the Profiler feature. A profile controls the value
of the setpoint over time; increasing, decreasing or holding its value as required.
Section 11:Manually Tuning Controllers - Advice on manually adjusting the controller to
the Process characteristics.
Sections 12: Serial Communications - Details the physical layer and message formats
used for the RS485 and Ethernet communications options.
Sections 13: Modbus Parameters - Details the parameter addresses and data formats used
for the Modbus RTU and TCP communications protocols.
Section 14:Calibration - Step-by-step instructions to calibrate the instrument. This section
is intended for use by suitably qualified personnel.
Appendix 1: Glossary - Explanations of the terms used and product features.
Appendix 2: PC Software – Using the software suite.
Appendix 3: Specifications - Technical specifications for all products in the range.
USB Data Transfer Problems...............................................................................................................40
Getting Help ........................................................................................................................................40
7Configuration and Use .........................................................................................................41
Single Control Tuning (PID with Primary Output only) ..........................................................................73
Dual Control Tuning (PID with Primary and Secondary Outputs) ..........................................................74
PI Tuning (Valve, Damper& Speed Controller) .....................................................................................74
Fine Tuning. ........................................................................................................................................76
CPU ..................................................................................................................................................139
Deviation Alarm Value .......................................................................................................................140
Digital Input .......................................................................................................................................140
Direct Acting Control..........................................................................................................................140
Linear Input .......................................................................................................................................142
Linear Output.....................................................................................................................................142
PI Control ..........................................................................................................................................149
Digital Inputs......................................................................................................................................174
This product is a 1/4 DIN size (96 x 96mm front) microprocessor based graphical process
controller, featuring a 160 x 80 pixel, monochrome LCD with a dual colour (red/green)
backlight. It can control process variables from a variety of sources such as temperature,
pressure, flow and level.
The operating voltage is either 100-240V at 50/60 Hz or 24V-48V AC/DC depending on the
model purchased. Optional features include a USB interface, RS485 or Ethernet
communications, profile controlling and data recording. Non-volatile memory protects against
data or configuration loss during power outages. If the unit is left un-powered, a lithium
battery powers the data recorder’s real-time clock for a minimum of one year.
The USB Interface option allows uploading or downloading instrument configuration settings
to/from a USB memory stick, for easy configuration of multiple instruments or transfer to/from
the PC configuration software. If the Data Recorder of Profiler options are fitted, recordings
and profile information can also be transferred via the memory stick.
The Data Recorder option allows the user to make recordings over time. Recordings can be
transferred to a memory stick using the USB Port or downloaded using one of the
communications options.
The Profiler option allows the user to predefine up 255 segments, shared amongst up to 64
Setpoint Profiles. These control the setpoint level over time, increasing, decreasing or
holding its value as required. When combined with the real-time clock of the Data Recorder
option, the profiling capabilities are expanded to allow automatic program start at a defined
time and day.
Inputs are user configurable for thermocouple and RTD probes, as well as linear process
signal types such as mVDC, VDC or mADC. Multipoint scaling can compensate for nonlinear signals. Output options include relays, SSR drivers, triacs or linear mV/voltage
modules. These can be used for process control, alarms or retransmission of the process
variable or setpoint to external devices. A Transmitter Power Supply option module can
provide an unregulated 24V DC (22mA) auxiliary output voltage for external signal
transmitters.
Alarm indication is standard on all instruments; up to five alarms can be defined. Alarms may
be set as process high or low, deviation (active above or below controller setpoint), band
(active both above and below setpoint), rate of input change, control loop or signal break
types. Alarm status can be indicated by lighting an LED’s, changing the display backlight
colour or viewing the alarm status screen. These alarms can be linked to any suitable output.
The controller can be programmed for on-off, time proportioning, or current proportioning
control implementations, depending on the output modules fitted, and feature manual or
automatic tuning of the PID parameters. A secondary control output is available when
additional output modules are fitted. Optional analogue Remote Setpoint inputs can be
included. Configuration of the major settings is made easy by a Setup Wizard that runs
automatically at first ever power-up or whenever option modules have been changed. Access
to the full range of parameters is via a simple menu driven front panel interface, or the PC
based configuration software.
Page 14Table Of Content59407-6, Issue 1 – March 2008
1. Remove the product from its packing. Retain the packing for future use, in case it is
necessary to transport the instrument to a different site or to return it to the supplier for
repair/testing.
2. The instrument is supplied with a panel gasket and push fit fixing strap. A single sheet
concise manual is also supplied in one or more languages. Examine the delivered items
for damage or defects. If any are found, contact your supplier immediately.
Installation
CAUTION:
Installation should be only performed by technically competent personnel. It is the
responsibility of the installing engineer to ensure that the configuration is safe.
Local Regulations regarding electrical installation & safety must be observed (e.g.
US National Electrical Code (NEC) or Canadian Electrical Code).
Figure 1. Main dimensions
Panel-Mounting
The mounting panel must be rigid and may be up to 6.0mm (0.25 inches) thick. The cut-out
size is:
Instruments may be mounted side-by-side in a multiple installation, but instrument to panel
moisture and dust sealing will be compromised. Allow a 20mm gap above, below and behind
the instrument for ventilation. The cut-out width (for n instruments) is:
(96n - 4) mm or (3.78n - 0.16) inches
If panel sealing must be maintained, mount each instrument into an individual cut-out with
6mm or more clearance between the edges of the holes.
Note:
The mounting clamp tongues may engage the ratchets either on the sides or the
top/bottom faces of the Instrument housing. When installing several Instruments side-byside in one cut-out, use the ratchets on the top/bottom faces.
CAUTION:
Ensure the inside of the panel remains within the instrument operating temperature
and that there is adequate airflow to prevent overheating.
Gasket
Slide mounting clamp over
the instrument housing
Mounting Panel
towards rear face of
Clamp
Ratchets
Housing
mounting panel until the
tongues engage in ratchets
position.
Hold instrument firmly in
position (apply pressure to
bezel only)
Figure 2. Panel-Mounting the instrument
CAUTION:
Do not remove the panel gasket, as this may result in inadequate clamping and
sealing of the instrument to the panel.
Once the instrument is installed in its mounting panel, it may be subsequently removed from
it’s housing, if necessary, as described in the Fitting and Removing Option Modules section.
Cleaning
Clean the front panel by washing with warm soapy water and dry immediately If the USB
option is fitted, close the USB port cover before cleaning.
The available plug-in modules, options and accessories are shown in below:
Table 1.Options & Accessories
PART
NUMBER
OPTION SLOT 1
PO1-R10
PO1-S20
PO1-T80
PO1-C21
OPTION SLOT 2 or 3
PO2-R10
PO2-W09
PO2-S20
PO2-S22
PO2-T80
PO2-C21
PO2-W08
OPTION SLOT A
PA1-W03
PA1-W04
PA1-W06
PA1-ETH
OPTION SLOT B
PB1-W0R
OPTION SLOT 4
PO4-R10
ACCESSORIES
PS1-PRF
PS1-PRW
DESCRIPTION
Single Relay Output for option slot 1
Single SSR Driver Output for option slot 1
Triac Output for option slot 1
Linear mA / Voltage Output module for option slot 1
Single Relay Output for option slot 2 or 3
Dual Relay Output for option slot 2 or 3
Single SSR Driver Output for option slot 2 or 3
Dual SSR Driver Output for option slot 2 or 3
Triac module Output for slot 2 or 3
Linear mA / Voltage Output for option slot 2 or 3
24VDC Transmitter Power Supply for option slot 2 or 3
Digital Input for option slot A
Basic Auxiliary Input for option slot A
RS485 Serial Communications for option slot A
Ethernet Communications for option slot A
Full Auxiliary Input (inc digital input B) for option slot B
4-Relay Output for option slot 4
Profiler Enable Key-code
PC Configuration Software & Lead
BOARD
IDENTIFICATION
NUMBER
716/01
716/02
716/03
639/01
717/01
644/01
717/02
644/02
647/01
640/01
642/01
641/02
653/01
680/01
707/01
641/01
703/01
Note:
Modules can be either pre-installed at the time of manufacture, or retrofitted in the field.
CAUTION:
Plastic pegs prevent fitting of older non-reinforced single relay modules (Board
Identification Numbers 637/01 and 638/01). Fitting the older relay modules reduces
the isolation rating to Basic 240V Isolation and is therefore not recommended.
Remove this peg when fitting Dual Relay Modules.
Note:
All dual relay modules have reinforced isolation.
59407, Issue 1 – March 2008Field Upgrade OptionsPage 17
Before removing the instrument from it’s housing, ensure that all power has been
removed from the rear terminals. Modules / boards should only be replaced by a
trained technician.
1. Remove the instrument from its housing by gripping the edges of the front panel (there is a
finger grip on each edge) and pull the instrument forwards. This will release the instrument
from the rear connectors in the housing and will give access to the boards.
2. Take note of the orientation of the instrument for subsequent replacement into the
housing. The positions of the boards in the instrument are shown above.
Page 18Field Upgrade Options59407, Issue 1 – March 2008
1. To remove or replace modules in Option Slots 1, 2, 3, A or B, it is necessary to detach the
Power Supply and Input boards from the front panel by lifting first the upper and then
lower mounting struts.
2. Remove or fit the modules to the connectors on the Power Supply and Input boards. The
location of the connectors is shown below. Plastic pegs prevent fitting of older nonreinforced single relay modules – Remove the peg to fit dual relay modules
3. Assemble the Power Supply and Input boards together. Tongues on each option module
locate into a slots cut into the main boards, opposite each of the connectors. Hold the
Power and Input boards together and relocate them back on their mounting struts.
4. Remove or replace the Slot C and 4 modules as required.
5. Push the boards forward to ensure correct connection to the front Display/CPU board.
CAUTION:
Check for correct orientation of the modules and that all pins are located correctly.
59407, Issue 1 – March 2008Field Upgrade OptionsPage 19
Before replacing the instrument in it’s housing, ensure that all power has been
removed from the rear terminals.
With the required option modules correctly located into their respective positions the
instrument can be replaced into it’s housing as follows:
1. Hold the Power Supply and Input boards together.
2. Align the boards with the guides in the housing.
3. Slowly and firmly, push the instrument in position.
CAUTION:
Ensure that the instrument is correctly orientated. A mechanical stop will operate if
an attempt is made to insert the instrument in the wrong orientation, this stop MUST
NOT be over-ridden.
Auto Detection of Option Modules
The instrument automatically detects which option modules have been fitted into each slot.
The menus and screens change to reflect the options compatible with the hardware fitted.
The modules fitted can be viewed in the products information menu, as detailed in the
Product Information Mode section of this manual.
Replacement of Power Supply or Input Boards
It is recommend that users change these boards only if unavoidable.
1. Remove the instrument from it’s housing as detailed above.
2. Remove all option modules.
3. Replace the Power Supply or Input board as required. Carefully observe the transformer
colour and the case labelling to check the supply voltage when replacing the power
supply board.
4. Reassemble the unit in it’s case.
5. If the input board has to be replaced, a full recalibration must be carried out before the
instrument is used. Refer to the calibration section of this manual for instructions.
CAUTION:
Replacement of boards must be carried out by a trained technician.
If the Power Supply board does not match the labelling, users may apply incorrect
voltage resulting in irreparable damage.
Page 20Field Upgrade Options59407, Issue 1 – March 2008
If installed, the Data Recorder memory and Real Time Clock (RTC) components are located
on a plug-in daughter board attached to the front Display/CPU board.
CAUTION:
Servicing of the Data Recorder/RTC circuit and replacement of the lithium battery
should only be carried out by a trained technician.
Profiler Enabling
If you purchased a controller with the Profiler option installed, these features will be enabled
during manufacture.
Controllers supplied without the Profiler option installed can be upgraded in the field by
purchasing a licence code number from your supplier. A unique code must be purchased to
enable profiling on each controller that requires it.
Entering A Profiler Enable Code
Hold down the and keys during the power-up “splash screen”.
Using the or keys, enter the 16-character licence code in the displayed screen.
Press to move on to the next character. Press to move back to the previous
character.
Press after entering the final character.
To confirm if profiling is installed in your instrument, refer to the Controller Feature
Information in Product Information mode.
59407, Issue 1 – March 2008Field Upgrade OptionsPage 21
Installation should be only performed by technically competent personnel. It is the
responsibility of the installing engineer to ensure that the configuration is safe.
Local Regulations regarding electrical installation & safety must be observed (e.g.
US National Electrical Code (NEC) or Canadian Electrical Code).
Installation Considerations
Ignition transformers, arc welders, motor drives, mechanical contact relays and solenoids are
examples of devices that generate electrical noise in typical industrial environments. The
following guidelines MUST be followed to minimise their effects.
1. If the instrument is being installed in existing equipment, the wiring in the area should be
checked to ensure that good wiring practices have been followed.
2. Noise-generating devices such as those listed should be mounted in a separate
enclosure. If this is not possible, separate them from the instrument, by the largest
distance possible.
3. If possible, eliminate mechanical contact relays and replace with solid-state relays. If a
mechanical relay being powered by an output of this instrument cannot be replaced, a
solid-state relay can be used to isolate the instrument.
4. A separate isolation transformer to feed only the instrumentation should be considered.
The transformer can isolate the instrument from noise found on the AC power input.
AC Power Wiring - Neutral (for 100 to 240V AC versions)
It is good practice to ensure that the AC neutral is at or near ground (earth) potential. A
proper neutral will help ensure maximum performance from the instrument.
Wire Isolation
Four voltage levels of input and output wiring may be used with the unit:
1. Analogue input or output (for example thermocouple, RTD, VDC, mVDC or mADC)
2. Relays & Triac outputs
3. SSR Driver outputs
4. AC power
CAUTION:
The only wires that should run together are those of the same category.
If any wires need to run parallel with any other lines, maintain a minimum space of 150mm
between them.
If wires MUST cross each other, ensure they do so at 90 degrees to minimise interference.
Page 22Wiring Instructions59407, Issue 1 – March 2008
All analogue signals must use shielded cable. This will help eliminate electrical noise
induction on the wires. Connection lead length must be kept as short as possible keeping the
wires protected by the shielding. The shield should be grounded at one end only. The
preferred grounding location is at the sensor, transmitter or transducer.
Noise Suppression at Source
Usually when good wiring practices are followed, no further noise protection is necessary.
Sometimes in severe electrical environments, the amount of noise is so great that it has to be
suppressed at source. Many manufacturers of relays, contactors etc supply 'surge
suppressors' which mount on the noise source. For those devices that do not have surge
suppressors supplied, Resistance-Capacitance (RC) networks and/or Metal Oxide Varistors
(MOV) may be added.
Inductive coils:- MOVs are recommended for transient suppression in inductive coils,
connected in parallel and as close as possible to the coil. Additional protection may be
provided by adding an RC network across the MOV.
Figure 5. Transient suppression with inductive coils
Contacts:- Arcing may occur across contacts when they open and close. This results in
electrical noise as well as damage to the contacts. Connecting a properly sized RC network
can eliminate this arc.
For circuits up to 3 amps, a combination of a 47 ohm resistor and 0.1 microfarad capacitor
(1000 volts) is recommended. For circuits from 3 to 5 amps, connect two of these in parallel.
Figure 6. Contact noise suppression
59407, Issue 1 – March 2008Wiring InstructionsPage 23
If the temperature probe is to be subjected to corrosive or abrasive conditions, it must be
protected by an appropriate thermowell. The probe must be positioned to reflect true process
temperature:
1. In a liquid media - the most agitated area
2. In air - the best circulated area
CAUTION:
The placement of probes into pipe work some distance from the heating vessel
leads to transport delay, which results in poor control.
For a two wire RTD a wire link should be used in place of the third wire. Two wire RTDs must
only be used with lead lengths less than 3 metres. Use of three wire RTDs is strongly
recommended.
Thermocouple Wire Identification Chart
The different thermocouple types are identified by their wires colour, and where possible, the
outer insulation as well. There are several standards in use throughout the world.
The table below shows the wire and sheath colours used for most
common thermocouple types. The format used in this table is:
+ Wire
- Wire
Sheath
Table 2.Thermocouple Extension Wire Colours
TypeInternational
IEC584-3
BlackWhiteYellowYellowRed
J
White
-
BrownBlueWhiteYellowRed
+
T
+*
White
-
GreenYellowBrownYellowRed
K
N
+
White
-*
PinkOrangeOrange
+
White
-
GreyGreyRed
+
B
White
-
Black
Brown
Green
Pink
Grey
USA ANSI
MC 96.1
Black
Red
Blue
Red
Yellow
Red
Orange
Red
Grey
Red
British
BS1843
Blue
Blue
Blue
Blue
Black
Blue
Red
Orange
French
NFC 42-324
Black
Black
Blue
Blue
Yellow
Purple
German
DIN 43710
Blue
Blue
Brown
Brown
Green
Green
Grey
Grey
R & S
C (W5)
+
White
-
+
-
Orange
Red
White
Red
Green
White
Blue
Green
Green
Green
White
White
OrangeBlackWhiteYellowRed
Note:
* = Wire is magnetic
Page 24Wiring Instructions59407, Issue 1 – March 2008
This symbol means the equipment is protected throughout by double insulation.
CAUTION:
All external circuits connected must provide double insulation. Failure to comply
with the installation instructions may impact the protection provided by the unit.
WARNING:
TO AVOID ELECTRICAL SHOCK, AC POWER WIRING MUST NOT BE CONNECTED TO THE
SOURCE DISTRIBUTION PANEL UNTIL ALL WIRING PROCEDURES ARE COMPLETED.
CHECK THE INFORMATION LABEL ON THE CASE TO DETERMINE THE CORRECT
VOLTAGE BEFORE CONNECTING TO A LIVE SUPPLY.
Note:
The wiring diagram below shows all possible combinations to the main connections
(numbered 1 to 24) in the centre of the case rear. The actual connections required depend
upon the features available on the model and the modules and options fitted.
Figure 7. Main Rear terminals
Note:
59407, Issue 1 – March 2008Wiring InstructionsPage 25
The wiring diagram below shows the additional connections (numbered 25 to 42) at the
sides of the case rear. These are required for Options Slots 4 and C if fitted.
Figure 8. Additional Option terminals
Note:
Use single strand (1.2mm / AWG18 max size) copper wire throughout, except for the
thermocouple input, where the correct thermocouple or compensating cable and
connectors must be used.
Power Connections
Power Connections - Mains Powered Instruments
Mains powered instruments operate from a 100 to 240V (±10%) 50/60Hz supply. Power
consumption is 20VA. Connect the line voltage (live and neutral) as illustrated via a two-pole
IEC60947-1 & IEC60947-3 compliant isolation switch / circuit breaker and a UL listed fuse
type: 250V AC 1Amp anti-surge. If the instrument has relay outputs with contacts carrying
mains voltage, it is recommended that the relay contacts supply should be switched and
fused in a similar manner, but should be separate from the instruments mains supply.
Figure 9. Mains Power Connections
WARNING:
CHECK THE INFORMATION LABEL ON THE CASE TO DETERMINE THE CORRECT
VOLTAGE BEFORE CONNECTING TO A LIVE SUPPLY.
CAUTION:
This equipment is designed for installation in an enclosure that provides adequate
protection against electric shock. The isolation switch should be located in close
proximity to the unit, in easy reach of the operator and appropriately marked.
Page 26Wiring Instructions59407, Issue 1 – March 2008
Power Connections - 24/48V AC/DC Powered Instruments
24/48V AD/DC powered instruments will operate from a 20 to 48V AC or 22 to 55V DC
supply. AC power consumption is 15VA max, DC power consumption is 12 watts max.
Connection should be via a two-pole IEC60947-1 & IEC60947-3 compliant isolation switch /
circuit breaker and a UL listed fuse type: 65v dc 1Aamp anti-surge.
Figure 10.24/48V AC/DC Power Connections
WARNING:
CHECK THE INFORMATION LABEL ON THE CASE TO DETERMINE THE CORRECT
VOLTAGE BEFORE CONNECTING TO A LIVE SUPPLY.
CAUTION:
This equipment is designed for installation in an enclosure that provides adequate
protection against electric shock. The isolation switch should be located in close
proximity to the unit, in easy reach of the operator and appropriately marked.
Universal Input Connections
Universal Input Connections - Thermocouple (T/C)
Use only the correct thermocouple wire or compensating cable from the probe to the
instrument terminals avoiding joints in the cable if possible. Failure to use the correct wire
type will lead to inaccurate readings. Ensure correct polarity of the wires by crossreferencing the colours with a thermocouple reference table.
Figure 11.Thermocouple Input Connections
59407, Issue 1 – March 2008Wiring InstructionsPage 27
For three wire RTDs, connect the resistive leg and the common legs of the RTD as
illustrated. For a two wire RTD a wire link should be used in place of the third wire (shown by
dotted line). Two wire RTDs should only be used when the leads are less than 3 metres long.
Avoid cable joints.
Figure 12.RTD Input Connections
Four wire RTDs can be used, provided that the fourth wire is left unconnected. This wire
should be cut short or tied back so that it cannot contact any of the terminals on the rear of
the instrument.
Universal Input Connections - Linear Volt, mV or mA input
Linear DC voltage, millivolt or milliamp input connections are made as illustrated. Carefully
observe the polarity of the connections.
mV/V
mA
Figure 13.DC Volt, mV & mA Input Connections
Page 28Wiring Instructions59407, Issue 1 – March 2008
If option slot 1 is fitted with a single relay output module, make connections as illustrated.
The relay contacts are SPDT and rated at 2 amps resistive, 240 VAC.
Figure 14.Option Slot 1 – Single Relay Module
Option Slot 1 – Single SSR Driver Output Module
If option slot 1 is fitted with a single SSR driver output module, make connections as
illustrated. The solid-state relay driver is a 0-10V DC signal, load impedance must be no less
than 500 ohms. SSR driver outputs are not isolated from the signal input or other SSR driver
outputs.
Figure 15.Option Slot 1 – Single SSR Driver Module
Option Slot 1 - Triac Output Module
If option slot 1 is fitted with a Triac output module, make connections as shown. This output
is rated at 0.01 to 1 amp @ 280V AC 50/60Hz. A snubber should be fitted across inductive
loads to ensure reliable switch off the Triac.
Figure 16.Option Slot 1 - Triac Module
59407, Issue 1 – March 2008Wiring InstructionsPage 29
Loading...
+ 154 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.