West Control Solutions KS98-1 User Manual

PMA Prozeß- und Maschinen-Automation GmbH
KS 98-1 Multi-function unit
Engineering manual
Valid from: 1.02.2011 Order number : 9499-040-82711
û
A publication of:
PMA Prozeß- und Maschinen-Automation GmbH P.O.Box 310 229 • D-34058 Kassel • Germany
All rights reserved. No part of this document may be reproduced or published in any form or by any means without prior written permission from the copyright owner.
Symbols used on the instrument
Liability and warranty
Any information and notes in these operating instructions were composed under consideration of the applicable regulations, the present state of the art and our extensive know-how and experien ce.
With special versions, additional ordering options or due to the la test technical modifications, the actual scope of delivery may vary from the descriptions and drawings in this manual.
For questions, please, contact the manufacturer.
a
Before starting to work with the instrument and be fore commissioning, in particular, these operating instructions must be read carefully! The manufacturer cannot be held responsible for damage and trouble resulting from failure to com ply with the information given in this manual.
-
-
-
-
à
a
a
!
l
g
+
EU conformity marking
Caution, Follow the operating instructions!
Symbols in the text
Danger of injury
Danger for the instrument, or of faulty function
Danger of destroying electronic components due to electrostatic discharge (ESD)
Additional information or reference to further sources of information.
Important hint for avoiding frequent operator faults.
This product may be subject to change due to im provements of the product features in the course of further development.
Copyright
This operating manual should be considered as confidential infor­mation, intended only for persons who work with the instrument.
Contraventions are subject to payment of damages. Further claims reserved.
-
I-3
Content
I. Operating instruction .............................9
I-1 Description .......................................9
I-2 Safety notes .....................................10
I-3 Technical data ....................................12
I-4 Versions ........................................18
I-4.1 I/O-Modules ..................................19
I-4.2 Ex-factory setting ...............................19
I-4.3 Auxiliary equipment ..............................19
I-5 Mounting .......................................20
I-5.1 Function of wire-hook switches........................21
I-5.2 Retro-fitting and modific. of I/O-ext. (*watch connecting diagram).....22
I-5.3 I/O extension with CANopen .........................22
I-6 Electrical connections - Safety hints .......................23
I-6.1 Electromagnetic compatibility ........................23
I-6.2 Measurement earth connection........................23
I-6.3 RC protective circuitry ...........................24
I-6.4 Galvanic isolations ..............................24
I-6.5 General connecting diagram .........................24
I-6.6 Analog inputs .................................26
I-6.7 Digital in- and outputs.............................27
I-6.8 Connecting diagram i/o-modules .......................28
I-7 Commissioning ....................................29
I-8 Operation .......................................30
I-8.1 Front view ...................................30
I-8.2 Menu structure ................................31
I-8.3 Navigation, page selection ..........................32
I-8.4 Adjusting values................................33
I-9 Instrument settings in the main menu .......................34
I-9.1 CAN-Status ..................................34
I-9.2 Profibus-Status ................................34
I-9.3 ModC-Status .................................34
I-9.4 Calibration ...................................35
I-9.5 Online/Offline .................................35
I-10 Operating pages ...................................36
I-10.1 List display...................................36
I-10.2 Bargraph display................................36
I-10.3 Alarm display .................................37
I-10.4 Graphic trend curve ..............................37
I-10.5 Programmer ..................................38
I-10.6 Controller ...................................41
I-10.7 Cascade controller...............................46
I-11 Maintenance, test, trouble shooting........................48
I-11.1 Cleaning ....................................48
I-11.2 Behaviour in case of trouble .........................48
I-11.3 Shut-down ...................................48
I-11.4 Test engineering as basic equipment.....................48
I-11.5 I/O-Test ....................................50
I-3
II. Engineering-Tool ...............................51
II-1 Survey .........................................51
II-1.1 Scope of delivery ...............................51
II-2 Installation ......................................52
II-2.1 Hardware and software prerequisites ....................52
II-2.2 Software installation .............................52
II-2.3 Licencing....................................53
II-2.4 Software start .................................53
II-3 Menu reference to the engineering tool .....................54
II-3.1 Menu ‘File’...................................54
II-3.2 Menu ‘Edit’ ..................................60
II-3.3 Menu ‘Functions’ ...............................63
II-3.4 Menu ‘Fixed funct.’ ..............................63
II-3.5 Menu ‘Device’ .................................64
II-3.6 Menu ‘Options’ ................................65
II-3.7 Menu ‘Window’ ................................67
II-3.8 Menu ‘Help’ ..................................67
II-4 Engineering tool operation .............................68
II-4.1 Fundamentals of the engineering tool operation ..............68
II-4.2 Function block placement ...........................68
II-4.3 Function block shifting ............................68
II-4.4 Creating connections .............................69
II-4.5 Online operation................................71
II-4.6 Trend function .................................72
II-5 Building an engineering...............................77
II-6 Tips and tricks ....................................81
II-6.1 Function keys .................................81
II-6.2 Mouse key functions .............................82
II-6.3 Tips and tricks .................................83
III. Function blocks: ...............................89
III-1 Scaling and calculating functions.........................91
III-1.1 ABSV (absolute value (No. 01)) ........................91
III-1.2 ADSU ( addition/subtraction (No. 03)) ....................91
III-1.3 MUDI ( Multiplication / division (No. 05))...................92
III-1.4 SQRT ( square root function (No. 08)) .....................92
III-1.5 SCAL ( scaling (No. 09)) ............................93
III-1.6 10EXP (10s exponent (No. 10)) ........................93
III-1.7 EEXP (e-function (No. 11)) ...........................94
III-1.8 LN (natural logarithm (No. 12)) ........................94
III-1.9 LG10 (10s logarithm (No. 13)) .........................95
III-2 Non-linear functions.................................96
III-2.1 LINEAR (linearization function (No. 07)) ...................96
III-2.2 GAP (dead band (No. 20)) ...........................98
III-2.3 CHAR (function generator (No. 21)) ......................99
III-3 Trigonometric functions ..............................100
III-3.1 SIN (sinus function (No. 80)) .........................100
III-3.2 COS (cosinus function (No. 81)) .......................100
III-3.3 TAN (tangent function (No. 82)) .......................101
I-4
III-3.4 COT (cotangent function (No. 83)) ......................102
III-3.5 ARCSIN (arcus sinus function (No. 84)) ...................103
III-3.6 ARCCOS (arcus cosinus function (No. 85))..................104
III-3.7 ARCTAN (arcus tangent function (No. 86)) .................105
III-3.8 ARCCOT (arcus cotangent function (No. 87)) ................105
III-4 Logic functions ...................................106
III-4.1 AND (AND gate (No. 60)) ..........................106
III-4.2 NOT (inverter (No. 61)) ............................106
III-4.3 OR (OR gate (No. 62)) ............................107
III-4.4 BOUNCE (debouncer (No. 63)) ........................108
III-4.5 EXOR (exclusive OR gate (No. 64))......................108
III-4.6 FLIP (D flipflop (No. 65)) ...........................109
III-4.7 MONO (monoflop (No. 66))..........................110
III-4.8 STEP (step function for sequencing (No. 68)) ................111
III-4.9 TIME1 (timer (No. 69)) ............................112
III-5 Signal converters ..................................114
III-5.1 AOCTET (data type conversion (No. 02))...................114
III-5.2 ABIN (analog i binary conversion (No. 71)) ................115
III-5.3 TRUNC (integer portion (No. 72)) ......................117
III-5.4 PULS (analog pulse conversion (No. 73)) ..................118
III-5.5 COUN (up/down counter (No. 74)) ......................120
III-5.6 MEAN (mean value formation (No. 75)) ...................122
III-6 Time functions....................................124
III-6.1 LEAD ( differentiator (No. 50)) ........................124
III-6.2 INTE (integrator (No. 51)) ..........................126
III-6.3 LAG1 ( filter (No. 52)).............................128
III-6.4 DELA1 ( delay time (No. 53)) .........................129
III-6.5 DELA2 ( delay time (No. 54) ) ........................130
III-6.6 FILT ( filter with tolerance band (No. 55) ) ..................131
III-6.7 TIMER ( timer (No. 67) ) ...........................132
III-6.8 TIME2 ( timer (No. 70) ) ...........................133
III-7 Selecting and storage ...............................134
III-7.1 EXTR ( extreme value selection (No. 30)) ..................134
III-7.2 PEAK ( peak value memory (No. 31)).....................135
III-7.3 TRST ( hold amplifier (No. 32))........................136
III-7.4 SELC ( Constant selection (No. 33)) .....................136
III-7.5 SELD (selection of digital variables - function no. 06) ...........137
III-7.6 SELP ( parameter selection (No. 34)) ....................138
III-7.7 SELV1 ( variable selection (No. 35)) .....................139
III-7.8 SOUT ( Selection of output (No. 36)).....................140
III-7.9 REZEPT ( recipe management (No. 37)) ...................141
III-7.10 2OF3 ( 2-out-of-3 selection with mean value formation (No. 38)) .....143
III-7.11 SELV2 ( cascadable selection of variables (No. 39)) ............145
III-8 Limit value signalling and limiting ........................146
III-8.1 ALLP ( alarm and limiting with fixed limits(No. 40)).............146
III-8.2 ALLV ( alarm and limiting with variable limits (No. 41))...........148
III-8.3 EQUAL ( comparison (No. 42) )........................150
III-8.4 VELO ( rate-of-change limiting (No. 43) ) ..................151
III-8.5 LIMIT ( multiple alarm (No. 44)) .......................152
III-8.6 ALARM ( alarm processing (No. 45) )
....................153
I-5
III-9 Visualization .....................................154
III-9.1 TEXT (text container with language-dependent selection (No. 79)) ....154
III-9.2 VWERT ( display / definition of process values (No. 96) )..........156
III-9.3 VBAR ( bargraph display (No. 97) )......................161
III-9.4 VPARA ( parameter operation (No. 98) ) ...................164
III-9.5 VTREND ( trend display(No. 99) ) ......................166
III-10 Communication ...................................169
III-10.1 L1READ ( read level1 data(No. 100) ) ....................169
III-10.2 L1WRIT ( write level1 data (No. 101)) ....................170
III-10.3 DPREAD ( read level1 data via PROFIBUS (No. 102) ) ............171
III-10.4 DPWRIT ( write level1 data via PROFIBUS (No. 103)) ............172
III-10.5 MBDATA (read and write parameter data via MODBUS - no. 104) ....173
III-11 I/O extensions with CANopen...........................175
III-11.1 RM 211, RM212 and RM213 basic modules ................175
III-11.2 C_RM2x (CANopen fieldbuscoupler RM 201 (No. 14)) ...........176
III-11.3 RM_DI (RM 200 - (digital input module (No. 15)) ..............177
III-11.4 RM_DO (RM 200 - digital output module (No. 16)) .............177
III-11.5 RM_AI (RM200 - analog input module (No. 17)) ..............178
III-11.6 RM_AO (RM200 - analog output module (No. 18)) .............180
III-11.7 RM_DMS (strain gauge module (No. 22)) ..................181
III-12 KS 98-1- KS 98-1 cross communication (CANopen) ..............183
III-12.1 CRCV (receive mod. block no's 22,24,26,28 (No.56)) ............183
III-12.2 CSEND (Send mod. blockno.'s 21, 23, 25, 27 - (No. 57)) ..........184
III-13 Connection of KS 800 and KS 816 .........................185
III-13.1 C_KS8x (KS 800 and KS 816 node function - (No. 58)) ...........186
III-13.2 KS8x (KS 800/ KS 816 controller function - (No. 59)) ............187
III-14 Description of KS 98-1 CAN bus extension ...................189
III-14.1 CPREAD (CAN-PDO read function (No. 88)) .................193
III-14.2 CPWRIT (CAN-PDO write function (No. 89)) ................194
III-14.3 CSDO (CAN-SDO function (No. 92)) ....................195
III-15 Programmer .....................................201
III-15.1 APROG ( analog programmer (No. 24)) / APROGD ( APROG data (No. 25)) . 201 III-15.2 DPROG ( digital programmer (No. 27)) / DPROGD ( DPROG data(No. 28)) . 219
III-16 Controller.......................................223
III-16.1 CONTR (Controllerfunction with one parameterset (No. 90)) ........223
III-16.2 CONTR+ (Controllerfunction with six parametersets (No. 91)) .......224
III-16.3 Parameter and Konfiguration for CONTR, CONTR+ .............226
III-16.4 Control behaviour ..............................228
III-16.5 Optimizing the controller...........................240
III-16.6 Empirical optimization ............................241
III-16.7 Self-tuning r controller adaptation to the process ............242
III-16.8 PIDMA (Control function with particular self-tuning behaviour (No. 93) .....247
III-16.9 Parameter and configuration for PIDMA ..................250
III-16.10 Controller characteristics and self-tuning with PIDMA ...........252
III-16.11 Controller applications: ...........................256
III-16.12 Setpoint functions ..............................260
III-16.13 Process value calculation ..........................265
III-16.14 Small controller ABC ...........................271
III-17 Inputs .........................................274
I-6
III-17.1 AINP1 ( analog input 1 (No. 08)) .......................274
III-17.2 AINP3...AINP5 ( analog inputs 3...5 (No. 112...114) ) ............281
III-17.3 AINP6 ( analog input 6 (No. 115)) ......................282
III-17.4 DINPUT ( digital inputs (No. 121)) ......................285
III-18 Outputs ........................................286
III-18.1 OUT1 and OUT2 ( process outputs 1 and 2 (No. 116, 117)).........286
III-18.2 OUT3 ( process output 3 (No. 118)).....................287
III-18.3 OUT4 and OUT5 ( process outputs 4 and 5 (No. 119...120))........288
III-18.4 DIGOUT ( digital outputs (No. 122)).....................289
III-19 Additional functions ................................290
III-19.1 LED (LED display (No. 123)).........................290
III-19.2 CONST ( constant function (No. 126) )...................291
III-19.3 INFO ( information function (No. 124) )...................292
III-19.4 STATUS ( status function (No. 125)).....................293
III-19.5 CALLPG (Function for calling up an operating page (no. 127)) .......296
III-19.6 SAFE ( safety function (No. 94)).......................297
III-19.7 VALARM (display of all alarms on alarm operating pages (function no. 109)298
III-20 Modular I/O - extension-modules ........................300
III-20.1 TC_INP (analog input card TC, mV, mA) ...................300
III-20.2 F_Inp (frequency/counter input) .......................302
III-20.3 R_Inp (analog input card ) ..........................303
III-20.4 U_INP (analog input card -50...1500mV, 0...10V) ..............305
III-20.5 I_OUT (analog output card 0/4...20mA, +/-20mA) .............307
III-20.6 U_OUT (analog output card 0/2...10V, +/-10V) ...............308
III-20.7 DIDO (digital input/output card) .......................309
III-21 Function management ...............................310
III-21.1 Memory requirement and calculation time .................310
III-21.2 Sampling intervals ..............................311
III-21.3 EEPROM data ................................311
III-22 Examples.......................................312
III-22.1 Useful small engineerings ..........................312
III-22.2 Controller applications ............................313
III-23 Index .........................................314
I-7
Preface
This manual consists of three parts.
I. Operating instructions
II. Engineeringtool description
III. Function block description
Section I holds the required information for identification, mounting, connection and electrical commissioning of the unit under consideration of safety notes of the application and environmental conditions.
The basic principles of operation are explained: Controls and indicators, menu structure and navigation with the cursor, selection of sub-menus and properties as well as adjustment of e.g. s and parameters.
Section II comprises the handling of the engineering-tool, the building of a simple engineering and transmission to the KS 98-1.
9499-040-82711
+
+
Section III presents the particular fnction blocks in detail.
For functional commissioning, additional descriptions are required; please, order them separately or load them from the PMA homepage: www.pma-online.de
As the functions provided in KS 98-1 are composed individually for each application using an Engineering Tool ET/KS 98, entire comprehension of the operating functions requires the relevant Project description with the Enginee ­ring !
Supplementary documentation:
PROFIBUS-protocol (GB) 9499-040-82811 ISO 1745-protocol (GB) 9499-040-82911
I-8
9499-040-82711 Description
I Operating instruction
I-1 Description
1
2
KS 98-1 advanced
4
3
Fig. 1 Frontview
The instrument is a compact automation unit the function of which can be configured freely by means of function blocks. Each unit contains a comprehensive function library for selection, configuration, parameter setting and connec ­tion of max. 450 function blocks by means of an engineering tool.
I.e. complex mathematical calculations, multi-channel control structures and sequence controllers can be realized in a single unit.
Various operating pages are indicated via the front-panel LCD matrix display, e.g.
Numeric input and output of analog and digital signals, values and parameters as well as
w
full-graphics display of bargraphs, controllers, programmers and trends.
w
Display colour red / green and direct / inverse display can be switched over dependent on event, or by operation
w
dependent on the engineering.
Dependent on version, the basic unit is equipped with analog and digital inputs, outputs and relays.
Additional inputs and outputs are available either with option C or “modular option C”, which contains four sockets for various I/O modules.
Optionally, the unit may be equipped with 2 additional communication interfaces:
Option B: serial TTL/RS422 interface or Profibus-DP
w
Option CAN: CANopen-compliant interface port for the I/O extension with modular I/O system RM200
w
I-9
Safety notes 9499-040-82711
The user is responsible that the instrument is operated only by
I-2 Safety notes
This section provides a survey of all important safety aspects: op timum protection of personnel and safe, trouble-free operation of the instrument.
Additionally, the individual chapters include specific safety notes for prevention of immediate hazards, which are marked with sym bols. Moreover, the hints and warnings given on labels and in scriptions on the instruments must be followed and kept in readable condition continuously.
General
Software and hardware are programmed or developed in compli ance with the state of the art applicable at the time of develop ment, and considered as safe.
Before starting to work, any person in charge of work at the prod uct must have read and understood the operating instructions.
trained and authorized persons. Maintenance and repair may be done only by trained and specialized persons who are familiar
-
with the related hazards.
Operation and maintenance of the instrument are limited to reli able persons. Any acts susceptible to impair the safety of persons or of the environment have to be omitted. Any persons who are under effect of drugs, alcohol or medication affecting reaction are
­precluded from operation of the instrument.
-
Instrument Safety
This instrument was built and tested according to VDE 0411 / EN61010-1 and was shipped in safe condition. The unit was tested before delivery and has passed the tests required in the
­test plan.
-
In order to maintain this condition and to ensure safe operation, the user must follow the hints and warnings given in these safety
­notes and operating instructions.
-
The plant owner is recommended to request evidence for knowl edge of the operating instructions by the personnel.
-
Correct use for intended application
The operating safety is only ensured when using the products cor­rectly for the intended application. The instrument can be used as a multiple function controller for open and closed control loops in industrial areas within the limits of the specified technical data and environmental conditions.
Any application beyond these limits is prohibited and considered as non-compliant.
Claims of any kind against the manufacturer and /or his lawful agents, for damage resulting from non-compliant use of the instru ment are precluded, liability is limited to the user.
User responsibility
The user is responsible:
for keeping the operating manual in the immediate vicinity
w
of the instrument and always accessible for the operating personnel.
for using the instrument only in technically perfect and safe
w
condition.
Apart from the work safety notes given in these operating instruc tions, compliance with the generally applicable regulations for safety, accident prevention and environment protection is compul sory.
The user and the personnel authorized by the user are responsible for perfect functioning of the instrument and for clear definition of competences related to instrument operation and maintenance. The information in the operating manual must be followed com pletely and without restrictions!
-
The unit is intended exclusively for use as a measuring and control instrument in technical installations.
The insulation meets standard EN 61010-1 with the values for overvoltage category, degree of contamination, operating voltage range and protection class specified in the operating instructions / data sheet.
The instrument may be operated within the specified environmen­tal conditions (see data sheet) without impairing its safety.
The instrument is intended for mounting in an enclosure. Its con­tact safety is ensured by installation in a housing or switch cabi­net.
Unpacking the instrument
­Remove instrument and accessories from the packing. Enclosed
standard accessories:
operating notes or operating instructions
fixing elements.
Check, if the shipment is correct and complete and if the instru ment was damaged by improper handling during transport and storage.
!
-
We recommend to keep the original packing for shipment in case
­of maintenance or repair.
Warning! If the instrument is so heavily damaged that safe operati on seems impossible, the instrument must not be taken into operation.
-
-
I-10
9499-040-82711 Safety notes
Mounting
Mounting is done in dustfree and dry rooms. The ambient temperature at the place of installation must not ex
ceed the permissible limits for specified accuracy given in the technical data. When mounting several units with high packing density, sufficient heat dissipation to ensure perfect operation is required.
For installation of the unit, use the fixing clamps delivered with the unit.
The sealing devices (e.g. sealing ring) required for the relevant protection type must also be fitted.
Electrical connections
All electrical wiring must conform to local standards (e.g. VDE 0100 in Germany). The input leads must be kept separate from signal and mains leads.
The protective earth must be connected to the relevant terminal (in the instrument carrier). The cable screening must be connected to the terminal for grounded measurement. In order to prevent stray electric interference, we recommend using twisted and screened input leads.
The electrical connections must be made according to the relevant connecting diagrams.
See electrical safety, page 23
Electrical safety
The insulation of the instrument meets standard EN 61 010-1 (VDE 0411-1) with contamination degree 2, overvoltage category III, working voltage 300 V r.m.s. and protection class I.
Galvanically isolated connection groups are marked by lines in the connecting diagram.
Commissioning
Before instrument switch- on, ensure that the rules given below were followed:
Ensure that the supply voltage corresponds to the specifica
w
tion on the type label.
All covers required for contact safety must be fitted.
w
Before instrument switch- on, check if other equipment and
w
/ or facilities connected in the same signal loop is / are not affected. If necessary, appropriate measures must be taken.
On instruments with protection class I, the protective earth
w
must be connected conductingly with the relevant terminal in the instrument carrier.
Operation
Switch on the supply voltage. The instrument is now ready for op eration. If necessary, a warm- up time of approx. 1.5 min. should
­be taken into account.
a
Warning ! Any interruption of the protective earth in the in strument carrier can impair the instrument safety. Purposeful interruption is not permissible. If the instrument is damaged to an extent that safe operation seems impossible, shut it down and protect it against accidental operation.
Shut- down
For permanent shut- down, disconnect the instrument from all voltage sources and protect it against accidental operation.
Before instrument switch- off, check that other equipment and / or facilities connected in the same signal loop is / are not affected. If necessary, appropriate measures must be taken.
Maintenance and modification
The instrument needs no particular maintenance. Modifications, maintenance and repair may be carried out only by
trained, authorized persons. For this, the user is invited to contact the service.
For correct adjustment of wire-hook switches (page 21 ) and for in­stallation of modular option C, the unit must be withdrawn from the housing.
a
Before carrying out such work, the instrument must be discon nected from all voltage sources.
After completing such work, re- shut the instrument and re-fit all
­covers and components. Check, if the specifications on the type la bel are still correct, and change them, if necessary.
Warning! When opening the instruments, or when remo ving covers or components, live parts or termi nals can be exposed.
Explosion protection
Non-intrinsically safe instruments must not be operated in explo sion-hazarded areas. Moreover, the output and input circuits of the instrument / instrument carrier must not lead into explo sion-hazarded areas.
-
-
-
-
-
-
-
-
The instrument must be operated only when mounted in its
w
enclosure.
I-11
Technical data 9499-040-82711
I-3 Technical data
General
Housing
Plug-in module, inserted from front. Material: Makrolon 9415, flame-retardant, self-extinguishing Flammability class: UL 94 VO
Front-panel / display
160 x 80-dot LCD matrix display 4 LEDs, 4 foil-covered keys
Front interface port (standard)
Front-panel socket for PC adapter (see „Accessories“ page 19).
Protection mode
(to EN 60 529, DINVDE 0470) Front: IP 65 Housing: IP 20 Terminals: IP 00
Safety tests
Complies with EN 61 010-1
Overvoltage category: III
w
Contamination class: 2
w
Working voltage range: 300 VAC
w
Protection class: I
w
Electrical connections
Screw terminals for conductor cross-section from 0,5 - 2,5 mm
Mounting method
Panel mounting with 4 fixing clamps at top/bottom
Mounting position
Not critical
Weight
Approx. 0.75 kg with all options
Environmental conditions
Permissible temperatures
For operation: 0...55°C For specified accuracy: 0...60°C, optionC For UL-Devices: Operation and protected operation 0...50°C Storage and transport: –20...60 °C
Temperature effect: £ 0.15% / 10 K
Climatic category
KUF to DIN 40 040 Relative humidity: £ 75% yearly average, no condensation
Shock and vibration
Vibration test Fc: To DIN 68-2-6 (10...150 Hz) Unit in operation: 1g or 0.075 mm, Unit not in operation: 2g or 0.15 mm Shock test Ea:To DIN IEC 68-2-27 (15g, 11 ms)
2
DIN EN 14597
The device may be used as temperature control and limiting equip ment according to DIN EN 14597.
CE marking
Meets the EuropeanDirectives regarding „Electromagnetic Compa tibility“ and „Low-voltage equipment“(r page 23)
The unit meets the following European guidelines:
q
Electromagnetic compatibility (EMC): 89/336/EEC (as amended by 93/97/EEC)
q
Electrical apparatus (low-voltage guideline): 73/23/EEC (as amended by 93/68/EEC).
Evidence of conformity is provided by compliance with EN 61326-1 and EN 61010-1
cULus certification
cULus-certification
(Type 1, indoor use) File: E 208286
a
RC protective circuitry must be fitted with inducti ve load !
-
-
-
I-12
9499-040-82711 Technical data
Connections
Depending on version and selected options, the following inputs and outputs are available:
DI DO AI AO
4 relays
di1*
or
2 relays
+
2AO
OPTION B
OPTION C*
or
modular
OPTION C*
Not available with option CAN!
*
di2*
di1* di2*
di3 di4
di5 di6 di7
di8
di9
di10 di11 di12
OUT1 OUT2 OUT4
OUT5
OUT4 OUT5
do1 do2 do3
do4
do5 do6
INP1 INP5 INP6
INP1 INP5 INP6
––
INP3 INP4
Depending on module type
OUT1 OUT2
OUT3
Inputs
Resistance thermometer
Pt 100 to DIN IEC 751, and temperature difference2xPt100
Range Error Resolution
–200.0...250.0 °C £ 0.5 K 0.024 K
–200.0...850.0 °C £ 1.0 K 0.05 K 2 x –200.0...250.0 °C £ 0.5 K 0.024 K 2 x –200.0...250.0 °C £ 0.1 K 0.05 K
Linearization in °C or ° F Two wire connection with compensation or three-wire connection
without. Two-wire connection with lead resistance adjustment. Lead resistance: £ 30 W per lead Sensor current: £ 1mA Input circuit monitoring for sensor/lead break, and lead short circuit.
Potentiometric transducer
R
incl. 2 x R
total
L
0...500 0.1 % £ 0.02 W
Resistance-linear Sensor current: £ 1mA Matching/scaling with sensor connected. Input circuit monitoring for sensor/lead break, and lead short circuit.
Error Resolution
Universal input INP1
Limiting frequency: fg=1Hz,Measurementcycle: 200 ms
Thermocouples
according to DIN IEC 584
Type Range Error Resolution L –200...900°C £ 2K 0,05 K J –200...900°C £ 2 K 0,05 K K –200...1350°C £ 2 K 0,072 K N –200...1300°C £ 2 K 0,08 K S –50...1760°C £ 3 K 0,275 K R –50...1760°C £ 3 K 0,244 K
1)
B
(25)400...1820°C £ 3 K 0,132 K
T –200.. .400°C £ 2 K 0,056 K
2)
W(C) E –200... 900°C £ 2 K 0,038 K
With linearization (temperature-linear in °C or °F) Input resistance: ³ 1MW Cold-junction compensation (CJC): built in
Input circuit monitor:
Current through sensor: £ 1mA Reverse-polarity monitor is triggered at 10 °C below span start.
Additional error of internal CJC
£ 0.5 K per 10 K terminal temperature External temperature selectable: 0...60 °C or 32...140 °F
0...2300°C £ 2 K 0,18 K
*
1 ) from 400 °C
*
2 ) W5Re/W26Re
Resistance input
Range Error Resolution
0...250 0.25 W < 0.01 W
0...500 0.5 W < 0.02 W
Direct current 0/4...20mA
Range Error Resolution 0/4...20 mA £ 0.1 % £ 0.8 mA
Input resistance: 50 W Input circuit monitor with 4...20mA: triggered when I £ 2mA
Direct voltage
Range Error Resolution 0/2...10 V £ 0.1 % £ 0.4 mV
Input resistance ³ 100 kW
I-13
Technical data 9499-040-82711
Signal input INP5 Differential amplifier input
Up to 6 controllers can be cascaded, if there is no other galvanic connection between them. If there is, only 2 inputs can be cascaded.
Direct current and voltage
Technical data as for INP1, except for: Limiting frequency: = 0.25 Hz, Measurement cycle: 800 ms Input resistance (voltage): ³ 500 kW
Signal input INP6
Limiting frequency: = 0.5 Hz, Measurement cycle: 400 ms
Potentiometric transducer
Technical data as for INP1, except for:
R
inkl. 2 x RL Error Resolution
total
0...1000 0.1 % £ 0.04 W
Direct current 0/4...20 mA
Technical data as for INP1
Signal inputs INP3, INP4 (option C)
Galvanically-isolated differential amplifier inputs. Limiting frequency: fg = 1 Hz, Measurement cycle: 100 ms
Direct current
Technical data as for INP1, except Ri = 43 W
Outputs
Relay outputs (OUT4, OUT5)
Relays have potential-free change-over contacts. Max. contact rating: 500 VA, 250 V, 2 A with 48...62 Hz, cos0,9 Minimum rating: 12 V, 10 mA AC/DC Number of switching cycles electrical: for I = 1A/2A ³ 800,000 /
500,000 (at ~ 250V / (resistive load))
When connecting a contactor to a relay output, RC
!
protective circuitry to specification of the contactor manufacturer is required. Varistor protection is not recommended!
Outputs OUT1, OUT2
Relay or current/logic signal, depending on version:
OUT1, OUT2 as current outputs
Galvanically isolated from the inputs 0/4...20 mA, selectable Signal range: 0...22 mA Resolution: £ 6mA (12 bits) Error: £ 0.5% Load: £ 600 W Load effect: < 0.1% Limiting frequency: approx. 1 Hz Output cycle: 100ms
Control inputs di1...di12
di1, di2: standard di3...d7: Option B di8...di12: Option C
Opto-coupler
Nominal voltage: 24 VDC, external Current sink (IEC 1131 Type 1) Logic „0“ (Low): –3...5 V, Logic „1“ (High): 15...30 V Current demand: approx. 6 mA (for galvanic connections and isola
tion see page 25).
Built-in transmitter supply (optional)
Can be used to energize a two-wire transmitter or up to 4 opto-coupler inputs.Galvanically isolated.
Output: ³ 17.5 VDC, max. 22 mA Short-circuit proof.
Factory setting
The transmitter supply is available at terminals A12 and A14
+
See page 25, Configuration see page 21
OUT1, OUT2 as logic signal
0/³ 20 mA with a load ³ 600 W 0 / > 12 V with a load ³ 600 W
Output OUT3 (Option C)
Technical data as for OUT1, OUT2, as current output
Control outputs do1..do6
Galvanic isolated Opto-coupler outputs, galvanic isolation see
­page 25 and text.
Grounded load: common positive control voltage
Switch rating: 18...32 VDC; Imax £ 70 mA Internal voltage drop: £ 0.7V with I
Protective circuit: thermal, switches off with overload. Supply 24 V DC external Residual ripple £ 5%
ss
max
I-14
9499-040-82711 Technical data
Modular Option C
Each module has two channels which can be configured independently.
A/D-Converter
Resolution: 20,000 (50Hz) or 16,667 (60Hz) steps for the selected measuring range.
Conversion time: 20ms (50Hz) or 16.7ms (60Hz).
D/A-Converter
Resolution:12Bit Refresh-Rate: 100 ms
Cut-off frequency
Analog: fg=10Hz Digital: fg=2Hz Measurement cycle: 100 ms per module
R_INP Resistive input module
(9407-998-0x201) Connection method: 2, 3 or 4-wire circuit (with 3 und 4-wire con
nection, only one channel can be used).
Sensor current:
£ 0,25mA
Resistance thermometer
Type Range °C Overall error Resolution K/Digit Pt100 -200...850°C £ 2 K 0,071 Pt100 -200...100°C £ 2 K 0,022 Pt1000 -200...850°C £ 2 K 0,071 Pt1000 -200...100°C £ 2 K 0,022 Ni100 -60...180°C £ 2 K 0,039 Ni1000 -60...180°C £ 2 K 0,039
Linearization: in °C oder °F
Lead resistance
Pt (-200...850°C): £ 30 Pt (-200...100°C), Ni: £ 10
Lead resistance compensation
3 and 4-wire connection: not necessary. 2-wire connection: compensation via the front with short-circuited
sensor. The calibration values are stored in a non-volatile memory.
Lead resistance effect
3 and 4-wire connection: negligible
Sensor monitoring
Break: sensor or lead. Short-circuit: triggers at 20 K below mea suring range.
Resistance / Potentiometer
Range R
0...160 W
0...450 W
0...1600 W
0...4500 W
ges
/ W
W per lead
W per lead
Overall error
£ 1% 0,012 £ 1% 0,025 £ 1% 0,089 £ 1% 0,025
-
Resolution W/Digit
Characteristic: resistance-linear Lead resistance or 0%/100% compensation:
via the front with short-circuited sensor. The calibration values are stored in a non-volatile memory.
Variable resistance (only 2-wire connection): compensation for 0%
w
Potentiometer: compensation for 0% and 100%
w
Lead resistance effect
3 and 4-wire connection: negligible
Sensor monitoring:
break: sensor or lead
TC_INP Thermocouple, mV, mA Module
(9407-998-0x211)
Thermocouples
To DIN IEC 60584 (except L, W/C, D)
Type Range
L
­J
K
N S R
(1)
B
T
W(C)
D
E
*
(1) Values apply from 400°C
-200...900°C
-200...900°C
-200...1350°C
-200...1300°C
-50...1760°C
-50...1760°C
(25) 400...1820°C
-200...400°C
0...2300°C
0...2300°C
-200...900°C
Overall
error
£ 2 K 0,080 £ 2 K 0,082 £ 2 K 0,114 £ 2 K 0,129 £ 3 K 0,132 £ 3 K 0,117 £ 3 K 0,184 £ 2 K 0,031 £ 2 K 0,277 £ 2 K 0,260 £ 2 K 0,063
Linearization:in°Cor°F Linearity error: negligible
Input resistance:
³1MW
Temp. compensation (CJC): built in
£ 0.5K/10K
Error: External CJC possible: 0...60 °C or. 32...140 °F Source resistance effect: 1mV/kW
Sensor monitoring:
Sensor current:
£ 1mA
Reversed polarity monitor: triggers at 10K below measuring range
mV input
Range
0...30 mV
0...100 mV
0...300 mV
Input resistance: ³1MW Break monitoring: built in. Sensor current: £1mA
Overall error Resolution
£ 45 mV 1,7 mV £ 150 mV 5,6 mV £ 450 mV 17 mV
K/Digit
I-15
Technical data 9499-040-82711
mA input
Range 0/4...20 mA
Input resistance:10W Break monitor:<2mA (only for 4...20 mA) Over range monitor: >22mA
Overall error
£ 40 mA 2 mA
Resolution
U_INP High-impedance voltage input
(9407-998-0x221)
Range
-50...1500 mV
0...10 V
Characteristic: voltage-linear Input resistance: >1G Source resistance effect: £ 0,25 mV/MW Sensor monitoring: none
Overall error Resolution mV/Digit
£ 1,5 mV 0,09
£ 10 mV 0,56
W
U_OUT Voltage output module
(9407-998-0x301)
Signal ranges: 0/2...10V, -10...10V configurable Resolution: approx.. 5.4 mV/digit Load: ³ 2kW
Load effect : £ 0.1%
I_OUT Current output module
(9407-998-0x311)
Signal ranges: 0/4...20mA, -20...20mA configurable by channel Resolution: ca. 11mA/Digit Load: £ 400 W
Load effect: £ 0.1%/100W
F_INP Frequency/counter Module
(9407-998-0x411)
Current sink: to IEC 1131 Type 1
Logic „0“: -3...5V Logic „1“: 15...30V
Galvanic isolation: via Opto-couplers Nominal voltage: 24 VDC external Input resistance:12k Selectable functions:
Control input (2 channels)
w
Pulse counter (2 channels)
w
Frequency counter (2 channels)
w
Up/down counter (1 channel)
w
Quadrature counter (1 channel)
w
Frequency range: Signal shape: any (square 1:1 with 20kHz) Gate time: 0.1...20s adjustable (only relevant with frequency mea
surement)
W
£ 20 kHz
-
Influencing factors
Temperature effect: £ 0,1%/10K Supply voltage: negligible Common mode interference: negligible up to 50 Vrms Series mode interference: negligible up to 300 mVrms (TC),
30 mVrms (RT), 10 Vrms (U), 5 Vrms (F)
CAN I/O extension
The unit offers a port conforming to CANopen for connection of the RM 200 system and of KS 800 or additional KS 98-1 units with max. five CAN nodes.
DIDO Digital I/O Module
(9407-998-0x401)
Number of channels: 2 (configurable as input or output by channel)
Input
Current sink: to IEC 1131 Type 1) Logic „0“: -3...5V, Logic „1“: 15...30V
Measurement cycle: 100 ms Galvanic isolation : via opto-couplers Nominal voltage: 24 VDC external
Input resistance:5k
W
Output
Grounded load (common positive control voltage)
Switch rating: 18...32 VDC; Internal voltage drop: Refresh-Rate: 100 ms Galvanic isolation : via opto-couplers Protective circuit: thermal, switches off with overload.
I-16
£70 mA
£ 0.7 V
+ +
Control inputs di1 and di2 are not available !
Modular option C is not available
Power supply
Dependent on version:
Alternating current
90...250 VAC Frequency: 48...62 Hz Power consumption: approx. 17,1 VA; 9,7 W (max. configuration).
Universal current 24 V UC
24 VAC, 48...62 Hz/24 VDC Tolerance: +10...–15% AC 18...31.2 VDC Power consumption AC: approx. 14.1 VA; 9.5 W DC: 9,1 W (max. Configuration).
9499-040-82711 Technical data
Behaviour after power failure
Storage in non-volatile EEPROM for structure, configuration, pa rameter setting and adjusted setpoints.
Storage in a capacitor-buffered RAM (typ. > 15 minutes) for data of time functions (programmers, integrators, counters, ...)
-
Real-time clock(Option B, RS 422)
Buffer capacitor provides back-up for at least 2 days (typical).
Bus interface (Option B)
TTL and RS422 / 485-interface
Galvanically isolated, choice of TTL or RS 422/485 operation.
Number of multi function unitsper bus:
With RS 422/485: 99 With TTL signals: 32 interface modules (9404 429 980x1). Address
range ((00...99)See Documentation 9499-040-82911) .
PROFIBUS-DP interface
According to EN 50 170, Vol. 2. Reading and writing of all process data, parameters, and configu-
ration data.
Transmission speeds and cable lengths
automatic transm. speed detection, 9,6 kbit/s ...12 Mbit/s
Addresses
0...126 (factory setting: 126).Remote addressing is possible.
Other functions
Sync and Freeze
Terminating resistors
Internally selectable with wire-hook switches.
Cable
According to EN 50 170, Vol. 2 (DIN 19 245T3).
Required accessories
Engineering Set PROFIBUS-DP, consisting of:
GSD file, Type file
w
PROFIBUS manual (9499-040-82911)
w
Function block(s) for Simatic S5/S7
w
Electromagnetic compatibility
Complies with EN 50 081-2 and EN 50 082-2.
Electrostatic discharge
Test to IEC 801-2, 8 kV air discharge 4 kV contact discharge
High-frequency interference
Test to ENV 50 140 (IEC 801-3)
80...1,000 MHz, 10 V/m Effect: £ 1%
HF interference on leads
Test to ENV 50141 (IEC 801-6)
0.15...80 MHz, 10 V Effect: £ 1%
Fast pulse trains (burst)
Test to IEC 801-4 2 kV applied to leads for supply voltage, and signal leads.
Effect: £ 5%resp. restart
High-energy single pulses (surge)
Test to IEC 801-5 1 kV symmetric or 2 kV asymmetric on leads for supply voltage.
0.5 kV symmetric or 1 kV asymmetric on signal leads.
I-17
Versions 9499-040-82711
I-4 Versions
The instrument version results from the combination of various variants from the following table.
Fig. 2
Please mindfootnotes! KS98-1 with screw terminals only!
KS 98 Standard
KS
1
89
-
-
000
-
0
BASIC UNIT
POWER SUPPLY AND
CONTROL OUTPUTS
OPTION B
INTERFACE
OPTION
(standard)
OPTION
(modular)
SETTING
KS 98 with transmitter power supply KS 98 with CANopen I/O
1)
90...250V, AC 4 relays 24V UC, 4 relays 1
90...250V AC 2 relays+ 2 mA/logic, 4 24V UC, 2 + 2 mA/logicrelays 5 no interface 0 TTL interface + di/do
RS422 + di/do + clock 2 PROFIBUS DP + di/do 3
no extensions
C
INP3, INP4, OUT3, di/do Motherboard without modules
C
2)
Motherboard ordered modules inserted
Standard configuration Customer-specific configuration
Operating instruction
3)
Standard (CE certification)
1 2
0
1
0 1 3
2)
4
0 9
0
0
I-18
APPROVALS
cULus approval
DIN EN 14597 certified
1)
Not possible with Modular Option C! RM 200 not included in cULus approval !
2)
Not possible with CANopen ! I/O modules must be ordered separately! Mind possible combinations and power limitations; Text!®
3)
Detailled system manual can be ordered separatelyor downloaded (www.pma-online.de)
U D
9499-040-82711 Versions
)
I-4.1 I/O-Modules
Can be installed on instruments with modular option C basic card.
Fig. 3 Table of I/O-module Versions
9407 998 0 1
I-4.2
Singular order(separate delivery) 0
SOCKETS
Module group 1
In KS 98-1 pinned to socket 1 In KS 98-1 2pinned to socket In KS 98-1 3pinned to socket
2Module group
In KS 98-1 4pinned to socket
3)
3)
3)
3)
1 2 3 4
R_INP: Pt100/1000, Ni100/1000, resistor 20
ANALOG INPUTS
TC_INP: Thermocouple, mV, 0/4...20mA 21 U_INP: -50...1500mV (z.B. Lambda-probe), 0...10V 22 U_OUT: Voltage output 30
ANALOG OUTPUTS
I_OUT: Current output
4)
31
DIDO: digitale in-/outputs 40
DIGITAL SIGNALS
3) Please note on your order: "mounted in KS in orderposition X" 4
Max. 1 current output module
F_INP: Frequency-/counter inputs 41
98-1
Ex-factory setting
All delivered units permit operation, parameter setting and configuration via the front-panel keys.
Instruments with default setting are delivered with a test engineering which permits testing of the basic instrument in ­puts and outputs (no I/O extension) without auxiliary means.
+
This engineering is not suitable for controlling a system. For this purpose, a customer-specific engineering is required (see versions, section: Setting)
Instruments with "setting to specification" are delivered complete with an engineering. Code no. KS 98-1-1xx-xx09x-xxx is specified on the type label.
Accessories delivered with the instrument:
Operating manual, 4 fixing clamps
I-4.3 Auxiliary equipment
Engineering Tool ET/KS 98
Simulation SIM/KS 98-1
PC-Adapter
Adapter cable for connecting a PC (Engineering Tool) to the front-panel interface socket of the KS 98-1.
Updates and Demos on the PMA- Homepage (www.pma-online.de)
I/O-Modules I-19
Mounting 9499-040-82711
I-5 Mounting
96 96
?24
+0,8
92
1
160
2
3
1
KS 98-1advanced
4
1
2
3
92
4
+0,8
2
3
4
1...16
KS 98-1advanced
l
96
KS 98-1 advanced
96
max.
60°C
min. 0°C
max.
95% rel.
Fig. 4 Mounting
The instrument must be installed as described below. Required dimensions of the control cabinet cut-out and minimum clearances for installation of further units are shown in the drawing.
For mounting, insert the unit into the control cabinet or cabinet door cut-out. Push the instrument module fully home and mount it firmly by means of the locking screw. Four fixing clamps are delivered with the unit.
!
+
Fig. 5 Inserting the locking screws
Ü
These clamps have to be fitted in the instrument from the control cabinet inside: 2 at top and bot
-
tom.
Ü
*
Now, the threaded pins of the fixing clamps can be screwed against the control cabinet from insi
-
de.
¡
A rubber seal is fitted on the rear of the instrument front panel (in mounting direction). This rubber seal must be in perfect condition, flush and cover the cut-out edges completely to ensure tightness!
cULus certification
For compliance with cUus certificate, the technical data at the beginning must be taken into account (see technical data, page 12 .
I-20 Auxiliary equipment
9499-040-82711 Mounting
I-5.1 Function of wire-hook switches
For closing the wire-hook switches, release the locking screw, withdraw the instrument module from the housing and close the wire-hook switch. Re-insert the instrument and lock it.
Ex-factory setting
S open DP open - terminating resistor not active CAN open - terminating resistor not active TPS A 14/12
l
The unit contains electrostatically sensitive components. Comply with rules for protection against ESD during mounting.
Wire-hook switch S:
The switching status is signalled by function STATUS and can be used in the engineering for blocking e.g. operating pages and other settings.
Wire-hook switch for PROFIBUS DP (Option B):
The bus terminating resistor can be connected by 2 wire-hook switches (DP) in KS98-1.Both wire-hook switches must always be open or closed .
Wire-hook switch for CAN bus: (only Option CANbus):
Both ends of the CAN bus must be terminated (wire-hook switch closed).
Wire-hook switch for transmitter supply
Versions (KS98-11x-xxxxx) with transmitter supply are provided with a potential-free supply voltage for energization of a 2-wire transmitter, or of max. 4 control inputs.
The output connections can be made to terminals A4(+) - A1(-) by means of 3 wire-hook switches.
If A14/A12 is used for di1/di2, A12 must be linked with A1.
Connectors Ü
14 (+) 12 (-) T open closed
4 (+) 1 (-) D closed open
Remarks
Only available with INP1 configured for current or thermocouple The voltage input of INP5 is then not available
Fig. 6 Position of wire-hook switches
View from below
TPS 3
View from above
CAN
*
Ü
TPS 1
TPS 2
TPS-card
*
Ü
Ö
Function of wire-hook switches I-21
Mounting 9499-040-82711
p
00201
R_Inp
8368
Energizing digital inputs (e.g. di1...di4)
_
+
1 (12)
2 3
A
*
di 1 di 2
4 (14)
B
di 3 di 4
1
3 4
Connection 2-wire-transducer (e.g. INP1)
+
tion)
(O
_
1 (12)
4 (14)
+
13
_
15
A
I-5.2 Retro-fitting and modific. of I/O-ext. (*watch connecting diagram)
Fig. 7 Mounting of i/o modules
*)
socket
-
43 2 1
l
Only for instruments with modular option C!
The instrument contains electrostatically sensitive components. Original packing protects against elec trostatic discharge (ESD), transport only in original packing.
When mounting, rules for ESD protection must be
taken into account.
+
INP1
_
Connection:
KS 98-1 engineering must be taken into account, be­cause it determines pin allocation and signification of connections.
Moreover, the rules for the performance limits must be followed (see manual r 9499-040-82711).
Mounting
After releasing the locking screw, withdraw the KS 98-1 module from the housing.
Insert the module into the required socket with the
a
printed label pointing down-wards into the green connector and click it in position in the small, white contact b
at the top.
00201 R_Inp
The various modules can be identified by the printed label. The last 5 digits of the order number are given in the upper line.
8368
I-5.3 I/O extension with CANopen
The unit offers a CANopen-compliant interface port for connection of the RM 200 system, KS 800 or additional KS 98-1 units with max. five CAN nodes. See installation notes in the CANopen system manual (9499-040-62411).
I-22 Retro-fitting and modific. of I/O-ext. (*watch connecting diagram)
9499-040-82711 Electrical connections - Safety hints
I-6 Electrical connections - Safety hints
a
a
a
Following the safety notes starting on page 6 is indispensable!
The installation must be provided with a switch or power circuit breaker which must be marked accordingly. The switch must be located near the instrument and readily accessible for the operator.
When the instrument module is withdrawn from the housing, protection against dropping of conducting parts into the open housing must be fitted.
The protective earth terminal (P3) must be connected to the cabinet ground. This applies also to 24V supply.
I-6.1 Electromagnetic compatibility
European guideline 89/336/EEC. The following European standards are met: EN 61326-1.
The unit can be installed in industrial areas (risk of radio interference in residential areas). A considerable increase of the electromagnetic compatibility is possible by the following measures:
Installation of the unit in an earthed metal control cabinet.
w
Keeping power supply cables separate from signal and measurement cables.
w
Using twisted and screened measurement and signal cables (connect the screening to measurement earth).
w
Providing connected motor actuators with protective circuitry to manufacturer specifications. This measure pre-
w
vents high voltage peaks which may cause trouble for the instrument.
I-6.2 Measurement earth connection
+
Measurement earth connection is required for protection against interference. External interference voltages (incl. ra ­dio frequencies) may affect the perfect function of the device.
To ensure protection against interference, the measurement earth must be connected to cabinet ground.
Terminals A11(measurement earth) and P3 (protective earth) must be connected to cabinet ground via a short cable (approx. 20 cm)! The protective earth conductor of the power supply cable must also be connected to this earth poten ­tial (cabinet ground). r See also diagram on page 24
Electromagnetic compatibility I-23
Electrical connections - Safety hints 9499-040-82711
I-6.3 RC protective circuitry
Load current free connections between the ground po tentials must be realized so that they are suitable both for the low-frequency range (safety of persons, etc.) and the high-frequency range (good EMC values). The connections must be made with low impedance.
All metal grounds of the components installed in the cabinet Ü or in the cabinet door * must be scre wed directly to the sheet-metal grounding plate to en sure good and durable contact.
In particular, this applies to earthing rails ä, protecti ve earth rail #, mountingplates for switching units
> and door earthing strips <.Controllers KS40/50/90 y and KS98-1 x are shown as an example for eart
hing.
The max. length of connections is 20 cm (see relevant operating instructions).
Generally, the yellow/green protective earth is too long to provide a high-quality ground connection for high-frequency interferences.
Braided copper cables Ö provide a high frequency conducting, low-resistance ground connection, especi­ally for connecting cabinet Ü and cabinet door *.
-
-
Fig. 8 Protective circuitry
ä Ö
-
-
-
>
Ö
Ü
#
P3
A11
ä
*
x
6
y
<
Because of the skin effect, the surface rather than the cross section is decisive for low impedance. All connections must have large surfaces and good contact. Any lacquer on the connecting surfaces must be removed.
!
Due to better HF properties, zinc-plated mounting plates and compartment walls are more suitable for large-surface grounding than chromated mounting plates.
I-6.4 Galvanic isolations
Galvanically isolated connection groups are marked by lines in the connecting diagram.
Measuring and signal circuits: functional isolation up to a working voltage £ 33 VAC / 70 VDC against earth (to
w
DIN 61010-1; dashed lines).
Mains supply circuits 90...250 VAC, 24 VUC: safety isolation between circuits and against earth up to a working
w
voltage £ 300 Vr.m.s. (to EN 61010-1; full lines).
Instruments with I/O extension modules (KS98-1xx-x3xxx and KS98-1xx-x4xxx): sockets 1-2 and 3-4 are galvani
w
cally isolated from each other and from other signal inputs/outputs (functional isolation).
I-6.5 General connecting diagram
a a
a
The max. permissible working voltage on input and signal circuits is 33 VAC / 70 VDC against earth ! Ot herwise, the circuits must be isolated and marked with warning label for “contact hazard”.
The max. permissible working voltage on mains supply circuits may be 250 VAC against earth and against each other ! Only on versions with transmitter power supply (factory setting: connection across terminals A12-A14)
Additionally, the units must be protected by fuses for a max. power consumption of 12,3VA/7,1W per in strument individually or in common (standard fuse ratings, min. 1A)!
-
-
-
I-24 RC protective circuitry
9499-040-82711 Electrical connections - Safety hints
Fig. 9 connecting diagram
OUT2
OUT1
INP4
INP3
OUT3
+
1
P
2 3 4
OUT5
5 6
7
OUT4
ßßß500VA, 250V, 2A
8 9
+
0/4...20mA
_
+
0/4...20mA
_
ßßß500VA, 250V, 2A
-
24 V
.
+
di 8 ( )
+
di 9 ( )
+
di 10 ( )
+
di 11 ( )
+
di 12 ( )
+
do 5 do 6
GND
+
0/4...20mA
_
+
0/4...20mA
_
+
0/4...20mA
_
Nur bei Geräten mit Transmitter-Speisung
1 1
For instruments with built-in transmitter power supply
1
Seulementpour les appareils avec alimentation transmetteur
For instruments with modular option C r see connecting diagram on page 28
10 11 12
13 14 15
1 2 3 4 5 6 7 8
9 10 11 12 13 14 15
16
C
(Option)
A
oder
or/ou
10 11 12 13 14
1
15
16
B
(Option)
10 11 12 13 14 15 16
1 2 3 4 5 6
di (-) di1 (+) di2 (+) +Volt +mA
_
7 8 9
1
2 3 4 5 6 7 8 9
ABCP
1
1
INP5
Volt / mA
100%
_
0%
0/4...20mA
+
}}2
1
abcdef
-
24 V
+
+
di 3 ( )
+
di 4 ( )
+
di 5 ( )
+
di 6 ( )
+
di 7 ( ) do 1 do 2 do 3 do 4
RXD-B
GND
RXD-A
TXD-B
TXD-A
RS422 RS485 TTL PROFIBUS -DP
RGND
100 [
DATA B
DATA A
0%
+5V
GND
TRE
TXD
RXD
16
INP6
+
mA
_
100%
RxD/TxD-N
RxD/TxD-P
16
+
Volt
_
VP
GND
AB
OUT
IN
INTERBUS
INP1
a a
General connecting diagram I-25
The protective earth must be connected also with 24 V DC / AC supply (see safety notes on page 23 ). The polarity is uncritical.
Only on versions with transmitter power supply (factory setting: connection to terminals A12-A14). Connection of the transmitter power supply is determined by the wire-hook switch r page 21.
Electrical connections - Safety hints 9499-040-82711
I-6.6 Analog inputs
Thermocouples
see general connecting diagram on page 25. No lead resistance adjustment.
Internal temperature compensation:
compensating lead up to the instrument terminals. With AINP1, STK = int.CJC must be configured.
External temperature compensation:
Use separate cold junction reference with fixed reference temperature.
Compensating lead is used up to the cold junction reference. Copper lead between reference and instrument. With AINP1, STK = ext.CJC and TKref = reference temperature must be configured.
Resistance thermometer
Pt 100 in 3-wire connection. Lead resistance adjustment is not necessary, if RL1 is equal
RL2.
Resistance thermometer
Pt 100 in 2-wire connection. Lead resistance adjustment is necessary: Ra must be equal to RL1 + RL2.
Two Resistance thermometer
Pt100 for difference. Lead resistance compensation: proceed as described in chapter calibration page 27 - I-9.4.
Resistance transducer
Measurement calibration r proceed as described in chapter calibration page 27 - I-9.4.
16
15
RL1
}
16
RL1
Ra = RL1+RL2
16 15
RL1
15
}1 }2
RL1 = RL2
RL2
Ra
RL2
}
xeff = 1 - 2}}
RL2
14
14
14
Standard current signals 0/4...20 mA
Input resistance: 50 [, configure scaling and digits behind the decimal point.
Standard voltage signals 0/2...10V
Input resistance: ? 100 k[ (Voltage input module U_INP: >1 G[), configure scaling and digits behind the decimal point.
+
a
I-26 Analog inputs
INP5 is a difference input, the reference potential of which is connected to terminal A9. With voltage input, A6 must always be connected to A9.
The inputs INP1 / INP6 are interconnected (common reference potential). This must be taken into account if both inputs must be used for standard current signal. If necessary, galvanic isolation should be applied.
9499-040-82711 Electrical connections - Safety hints
I-6.7 Digital in- and outputs
The digital inputs and outputs must be energized from one or several external 24 V DC sources. Power consumption is 5 mA per input. The max. load is 70 m A per output.
Examples:
Digital inputs (connector A)
A
()()-
24V (ext.)
+
Imax. 5 mA Imax. 5 mA
Digital inputs and outputs at one voltage source (e.g. connector B) 70mA!
di 1 di 2
1 2 3
B
()()-
1
24V (ext.1)
+
Imax. 5 mA Imax. 5 mA Imax. 5 mA Imax. 5 mA Imax. 5 mA
di 3 di 4 di 5 di 6 di 7
10 11
2 3 4 5 6
Imax. 0,1 A
Imax. 0,1 ARLImax. 0,1 A
Imax. 0,1 A
7
do 1
8
do 2
9
do 3 do 4
Digital inputs and outputs at two voltage sources (e.g. connector B)
B
()()-
24V
(ext.1)
+
Imax. 6 mA Imax. 6 mA Imax. 6 mA Imax. 6 mA Imax. 6 mA
di 3 di 4 di 5 di 6 di 7
10 11
1
()()-
24V 2 3
(ext.2)
+
4 5 6 7 8 9
do 1 do 2 do 3 do 4
Imax. 70 m A
Imax. 70 m ARLImax. 70 m A
Imax. 70 m A
Digital in- and outputs I-27
Electrical connections - Safety hints 9499-040-82711
I-6.8 Connecting diagram i/o-modules
(Modular Option C)
+
CAN and modular option C are mutually precluding.
The inputs and outputs of multi-function unit KS98-1 can be adapted to the individual application by means of “modu lar optionC". The supporting card is firmly installed in the unit.
The card contains four sockets for various I/O modules which can be combined, whereby the positions of the various connection types are dependent on engineering.
The KS98-1 programming personnel must provide a connecting diagram corresponding to the block diagram ( r page ) for installation of the device.
4-wire 3-wire 2-wire Poti
R1
R1
R1
Resistive input (9407-998-0x201)
TC, mV, mA / V inputs
(9407-998-0x211) /
(9407-998-0x221)
R1
R2
+ _
I1
I2
U1
+ _
U2
R2
+
_
+
_
+
}1
+
}2
-
I1
I2
di 1 (+) di 2 (+)
di 1/2 (-)
di 1 (+) di 2 (+)
di 1/2 (-)
+ _
+ _
di 1 (+) di 2 (+)
di 1/2 (-)
di 1 (+) di 2 (+)
di 1/2 (-)
do1/2(+)
di 1 (+) di 2 (+)
di 1/2 (-)
U1
U2
do 1 do 2
do 1/2 (-)
+
_
+
_
Fig. 10 Quadruple counter up/down counter 2 x counter and2xfrequency
Voltage / current outputs
V
(9407-998-0x301) /
V
(9407-998-0x311)
Combined digital inputs/ outputs
(9407-998-0x401)
Frequency/ Counter inputs
(9407-998-0x241)
I-28 Connecting diagram i/o-modules
9499-040-82711 Commissioning
I-7 Commissioning
Before switching on the instrument, ensure that the following points are taken into account:
The supply voltage must correspond to the specification on the type label!
w
All covers required for contact protection must be fitted.
w
Before operation start, check that other equipment in the same signal loop is not affected. If necessary, appropri
w
ate measures must be taken.
-
a
a
The unit is freely configurable. For this reason, the input and output behaviour is determined by the loaded engi
w
neering. Before commissioning, make sure that the right instructions for system and instrument commissioning are available.
Unless an application-specific engineering was loaded, the unit is equipped with the IO test engineering described on page 49.
Before instrument switch-on, the plant-specific input and output signal types must be adjusted on the in strument, in order to avoid damage to the plant and to the device.
On instruments without default setting, partial I/O signal testing is possible.
The effect on connected equipment must be taken into account.
After supply voltage switch-on, start-up logo and Main menu wait! are displayed, followed by display of the main menu during several seconds .
Unless a selection is made during this time, the first operating page (e.g. a controller) is displayed automatically, wit ­hout marking a line or a field.
-
-
Connecting diagram i/o-modules I-29
Operation 9499-040-82711
Ü¡ ¢£¤
I-8 Operation
The operation of the device is menu-guided. The menu is divided into several levels which can all be influenced via the engineering, i.e. the final scope of the menu is dependent on the engineering.
This manual describes the operating functions which are independent of the engineering.
I-8.1 Front view
LEDs (Ü¡¢£):
for indication of conditions determined by the engineering, e.g. alarms or switching states.
HDIM keys (|–©¨):
Operation is by four keys for selection of pages and input on pages.
DI
The two functions of keys up/down are:
Navigation in menus and on pages
Changing input values (e.g. )
M
The two significations of the selector key are dependent on the selected field:
– Pressing the selector key (confirmation / Enter): startspage changing, – starts value alteration via the up/down keys and confirms the adjustment subsequently ( r page 32).
H
The functions of the auto/manual key are dependent on operating page, i.e. this key is sometimes called function key.
– Controller: auto/manual switchover –
Programmer: programmer control
Adjustment of digital values.
Fig. 11
¥|
©
y
x
#
Locking screw:
for locking the instrument module in the housing
<
PC interface:
PC interface for engineering tool ET/KS98 and BlueControl. The tools permit structuring/soft-wiring/configurati on/parameter setting/operation.
y
Display/operating page:
LCD point matrix (160 x 80 dots),
changeable green/red back lighting, direct/inverse display mode.
Display is dependent on configured functions.
I-30 Front view
-
+ 290 hidden pages