Welch Allyn Medical Diagnostic Equipment LXi User Manual

Welch Allyn Spot Vital Signs®LXi
Specifications

Features

> SureBP™with pulse rate and MAP > SureTemp®Plus or Braun ThermoScan®
PRO 4000 thermometry
> Masimo®or Nellcor®pulse oximetry > Manual weight, height, respiration
rate, and pain level entry
> Compatible with select digital scales > Calculates Body Mass Index > 50-reading memory > Connectivity-ready > Fast, accurate, flexible > Easy to use
*Use a letter suffix to designate language followed by a number to designate voltage (e.g., 45MT0-E1).
Languages: E=English, C=Chinese, DN=Danish, DT=Dutch, F=French, FN=Finnish, G=German, I=Italian, N=Norwegian, P=Portuguese, S=Spanish, SW=Swedish
Voltages: 1=North America, 2=Europe, 4=UK, 6=Australia
For more information, contact your Distributor Representative or Welch Allyn directly at 800.535.6663.

Ordering Information

450T0* 450E0* Spot LXi with SureBP and PRO 4000 Thermometer 45MT0* Spot LXi with SureBP, Masimo SpO 45ME0* Spot LXi with SureBP, Masimo SpO 45NT0* Spot LXi with SureBP, Nellcor SpO2, and SureTemp Plus Thermometer 45NE0* Spot LXi with SureBP, Nellcor SpO 4500-60 Mobile stand with basket 4500-62 Wall mount with basket 4500-500 External printer 4500-910 Barcode scanner 4500-920 Wireless radio 4500-925 Connectivity accessory kit—USB cable 4500-150E Inservice CD
Spot LXi with SureBP and Sur
eTemp Plus Thermometer
, and SureTemp Plus Thermometer
2
, and PRO 4000 Thermometer
2
, and PRO 4000 Thermometer
2
Welch Allyn Spot Vital Signs®LXi

Specifications

Noninvasive Blood Pressure (NIBP)
uff Pressure Range 0 to 300 mmHg (0 to 40.0 kPa)
C
ystolic Range 60 to 250 mmHg (8.0 to 33.3 kPa)
S
iastolic Range 30 to 160 mmHg (4.0 to 21.3 kPa)
D
ean Arterial 40 to 190 mmHg (5.3 to 25.3 kPa)
M Pressure Range Blood Pressure Accuracy Meets or exceeds AAMI SP10:2002
Overpressure Cutoff 315 mmHg ±15 mmHg (42.0 kPa ± 1.9 kPa)
Blood Pressure Typical: 15 seconds
etermination Time
D
Pulse Rate Range 35 to 199 bpm
Pulse Rate Accuracy
±5.0%
Temperature
emperature Range SureTemp
T
Calibration Accuracy SureTemp Plus: ±0.2° F (0.1° C) (Monitor Mode)
Display Resolution 0.1° F or ° C
SureTemp Plus Predict Time Oral Approx. 4 to 6 seconds
Probe and Probe Well Oral/Axillary Blue
SpO
2
Performance Measurement 1 to 100%
Range 25 to 240 bpm
SpO2Accuracy 70% to 100%
Pulse Rate Accuracy 25 to 240 bpm ±
®
lus: 80° F to 110° F (26.7° C to 43.3° C)
P
Braun ThermoScan® PRO 4000: 68° F to 108° F (20° C to 42.2° C)
Braun ThermoScan PRO 4000 for displayed temperature ranges: ± 0.4° F (±0.2° C) 95.9° F to 107.6° F (35.5° C to 42° C), ±0.5° F (±0.25° C) (outside this temperature range)
Adult Axillary Approx.12 to 15 seconds (age 18 years and older) Pediatric Axillary Approx. 10 to 13 seconds (age 17 years and younger) Rectal Approx. 10 to 13 seconds
Rectal Red
3 digits
Mechanical
Dimensions Height 10.63 “ (27 cm)
Length/PRO 4000 Length/SureTemp Plus 7.5” (19 cm) Depth 5.25” (13.34 cm)
Weight 7.5 lbs (3.4 kg)
8” (20.32 cm)
Electrical
• AC power supply: 8V DC at 0.75 A
• Rechargeable sealed lead acid internal battery
• Battery recharge time: 90% to 100% after 6 hours; 1 additional hour for PRO 4000; 4 additional hours for external printer
• Simultaneously operates device and charges battery when AC power adapter is plugged in
Wireless Radio
Wireless Network Interface IEEE 802.11b DSSS, WiFi compliant
Frequency 2.4 to 2.4835 GHz
Environmental
Operating Temperature 50° F to 104° F (+10° C to +40° C) (Device)
Operating Relative 15% to 95% (noncondensing) Humidity
Operating Altitude -557 ft to +16,000 ft (-170 m to +4,877 m)
Storage Temperature Device with SureTemp Plus: -13° F to 131°F (-25° C to 55°C)
Storage Humidity 15% to 95% (noncondensing)
EMC Standard IEC/EN 60601-1-2:2001
Device with PRO 4000: -4° F to 122°F (-20° C to 50°C)
Warranty
Two-year warranty on device parts and labor
• Service contracts available
Only Masimo SET SpO2sensors, including LNOP durable adhesive and NR value disposable sensors should be used with Masimo pulse oximetry. Masimo, Masimo SET, LNOP, and NR are registered trademarks of Masimo Corporation.
Only Nellcor with the Nellcor pulse oximetry option. Nellcor and OxiMax are registered trademarks of Nellcor Puritan Bennett, Inc.
OxiMax®pulse oximetry sensors should be used
4341 State Street Road, PO Box 220, Skaneateles Falls, NY 13153-0220 USA (p) 800.535.6663 (f) 315.685.2174
.welchallyn.com
www
SureTemp and Spot Vital Signs are registered trademarks of Welch Allyn, Inc. © 2005 Welch Allyn. All rights reserved. SM2871 Rev B
Braun ThermoScan® is a registered trademark of the Braun Company.