General
•
Available in two configurations:
- Standalone: model 802LTON
- Networked via wireless radio: model
802LTRN
•
Adult, paediatric and neonatal
patient modes
•
Electrosurgery interference suppression
(ESIS): all channels except impedance
respiration
•
All channels electronically isolated
ECG
•
Automatic 3- and 5-lead monitoring and
cable type detection
•
Selectable leads: I, II, III, aVR, VL, VF, V
•
Lead fault detection and autoconfigure
•
Bandwidth:
- Adult: 0.5 to 80 Hz monitor mode;
0.05 to 80 Hz extended mode
- Paediatric/neonate: 0.5 to 80 Hz monitor
mode; 0.05 to 80 Hz extended mode
•
Hypertronics connector
•
Heart rate measurement range:
25 to 300bpm (display)
•
Heart rate accuracy: ± 3bpm or 3%
•
Pacer detection and display
•
Electrosurgery and defibrillator protected
•
Sweep speed: 25 mm/sec; (6.25 mm/sec
respiration only)
•
User-selectable size (mV/cm):
8, 4, 2, 1, 0.5, 0.2
•
QRS tone: med/low/off
•
3.0 sec waveform duration at 25 mm/sec for
single waveform; 6.0 sec waveform duration
at 25 mm/sec for cascading waveform
•
QRS detector range:
- Adult or paediatric mode: 0.22 to 5.0 mV
(RTI)
- Neonatal mode: 0.1 to 5.0 mV
- Neonatal width: 40 to 120 msec
- Paediatric width: 40 to 120 msec
- Adult width: 70 to 120 msec
NIBP
•
Oscillometric method
•
Automatic (intervals between 1 to 60min)
and manual modes
•
Quick-action NIBP start/stop button
•
TurboCuf: 5min of repeated NIBP readings
•
Systolic, diastolic and mean display
•
On-screen manometer
•
Store and view up to 150 readings
•
Standard cuff sizes:
Neonates #1 to #5, infant, child, small adult,
adult, large adult, thigh
•
Cuff overpressure protection
•
Typical measurement time: 30 to 45 sec
•
Smartcuf®motion-tolerant NIBP technology
•
Blood pressure range:
- Adult:
Systolic: 30 to 260 mmHg
Diastolic: 20 to 235 mmHg
Mean: 20 to 255 mmHg
- Paediatric:
Systolic: 30 to 160 mmHg
Diastolic: 15 to 130 mmHg
Mean: 15 to 140 mmHg
- Neonate:
Systolic: 25 to 120 mmHg
Diastolic: 10 to 105 mmHg
Mean: 10 to 110 mmHg
•
Blood pressure target inflation
- Adult: 160 mmHg
- Paediatric: 120 mmHg
- Neonate: 90 mmHg
•
Pulse rate accuracy: ± 3 bpm or 3%
•
Pulse rate range: 30 to 220 bpm
SpO
2
(Masimo)
•
Masimo SET®technology
•
Perfusion Index and Signal IQ display
•
Sensitivity modes: Normal, MAX, APOD
•
Masimo SET®sensors
•
Waveform and pulse amplitude meter
•
Pulse tone pitch indicator
•
Saturation range: 1 to 100%
•
Probe accuracy: 25 to 41 °C
No motion:
Adult/Paediatric: 70 to 100% ±2 digits
0 to 69% unspecified
Neonatal: 70 to 100% ±3 digits
0 to 69% unspecified
During motion:
Adult/Paediatric/Neonatal:
70 to 100% ±3 digits
0 to 69% unspecified
•
Pulse rate measurement range:
(Adult/Paediatric/Neonatal):
No motion: 26 to 239 bpm
During motion: 26 to 239 bpm
•
Pulse rate accuracy:
No motion: ±3 digits
During motion: ±5 digits
SpO2(Nellcor)
•
Nellcor
®
OxiMax
®
sensors
•
Waveform and pulse amplitude meter
•
Normal, fast and slow response modes
•
Saturation range: 1 to 100%
•
Accuracy: (Probe: 70 to 100%, 28 to 42 °C)
- Adult/Paediatric digit: ± 2 counts
- Neonate digit: ± 3 counts
•
Pulse rate accuracy: no motion: ± 3 bpm
•
Pulse rate range: 25 to 300 bpm
Respiration
•
Breath rate source: impedance pneumography
•
Two user-selectable leads: RA-LA, RA-LL
•
Automatic cardiovascular artifact (CVA)
rejection
•
Respiration rate range:
- Adult/Paediatric modes: 2 to 150 br/m
- Neonate mode: 3 to 150 br/m
•
Respiration detection threshold: 175
milliohms or 2X CVA, whichever is greater
•
Respiration rate accuracy: ± 2 br/m or ± 2%
•
Sweep speed: 6.25 mm/sec
Physical
•
Monitor
- Height: 137.1 mm
- Width: 190.5 mm
- Depth: 53.3 mm
- Weight: 904.3 g with FlexNet radio;
765 g without radio
•
Charging station
- Height: 144.7 mm
- Width: 193 mm
- Depth: 152.4 mm
- Weight: 680.3 g
•
Colour display
•
3.5 in colour transflective LCD
•
Viewable in direct sunlight
Environmental
•
Operating temperature: 0 to 40 °C
•
Operating relative humidity: 5 to 95%,
noncondensing
•
Monitor operating altitude: -610 to 4,572 m
•
Cradle operating altitude: 152 to 4,572 m
•
Shipping/storage temperature: -20 to 60 °C
•
Shock: 75 g
•
Vibration, Random and Sinusoidal:
designed to meet RTCA/DO-160D, categories
section 8 modified
•
Electromagnetic compatibility (EMC): per IEC
601-1-2
•
Drip-proof per IEC 529, level IPX1
Power
•
Rechargeable lithium ion internal battery pack
•
External recharger
•
Input voltage: 12 to 28 VDC
•
AC power adapter: 110 and 220 V, DC
vehicle conversion
•
Operating time on battery, typical*: ECG/Resp
only: 24 hours; ECG/Resp, SpO
2
and NIBP:
8 hours
•
Battery recharge time: < 3 hours
Performance may vary based on monitor
configuration and battery condition
* Fully charged, new battery at 25°C, NIBP every 15
minutes, display and backlight operating in power
save mode.
Communications
•
PC Utility
- USB interface via charging station
- Supports Microsoft Windows XP/2000
and later
- Autoprint utility allows printing of tabular
trends and snapshots to connected laser
printer
•
Acuity®Central
Station Wireless FlexNet™Network
- 802.11a, 5 GHz
- Able to operate on shared 802.11a/b/g
network
Alarms
•
All parameters: upper/lower limits specific to
patient mode
•
Factory default or programmable settings for
all patient modes
•
Alarm indicator: red
•
Alarm(s) off indicator: yellow
•
Audible alarm tone: high/med/low
•
Alarm silence/reset: 90 sec
•
One-button “Param-set” for adjusting
specific parameter alarm limits,
programmable
•
Equipment alert: yellow
Trends
•
Tabular numeric and graphical formats
•
All parameters trended/viewable
•
Resolution: non-NIBP trends entered
every 1 min
•
NIBP trends entered after each reading
•
Duration: 24 hr non-NIBP trends; up to 150
NIBP readings
•
Page up/down trend view
•
Up to 20 data snapshots (7 sec behind and
14 sec ahead of snapshot event) of displayed
data captured
Certifications/Standards
•
Complies with all relevant AAMI, IEC, EN,
CSA, and UL standards. CE marked according
to the European Medical Device Directive.
CSA certificate of compliance for use in the
U.S. and Canada
Warranty
•
Monitor and recharger: two years
•
Accessories: 90 days
•
Extended warranties available
All specifications subject to
change without notice.
All Propaq vital signs monitors come standard with
Smartcuf technology, a new standard in NIBP accuracy.
This patented software offers industry-leading
performance and accuracy in the presence of
motion artifact.
®
UK Customer Services,
Welch Allyn UK Ltd,
Cublington Road,
Aston Abbotts,
Buckinghamshire HP22 4ND,
England
Tel: +44 (0) 207 365 6780
Fax: +44 (0) 207 365 9694
South Africa Customer Services,
Welch Allyn SA,
394 Surrey Avenue,
Ferndale 2194,
P O Box 809,
Fontainebleau 2032,
South Africa
Tel: +27 11 777 7555
Fax: +27 11 777 7556
Welch Allyn
European Customer Services,
Navan Business Park,
Dublin Road,
Navan,
Co. Meath,
Republic of Ireland
Tel: +353 46 90 67790
Fax: +353 46 90 67754
© 2007 Welch Allyn
All rights reserved. Acuity, Propaq and Smartcuf are registered
trademarks and FlexNet is a trademark of Welch Allyn. Other
product and company names mentioned herein are for
identification purposes only and may be the trademarks or
registered trademarks of their respective companies.
SM2833EU Rev.E
Propaq LT Specifications
Only Nellcor
®
OxiMax
®
pulse oximetry sensors,
should be used with the Nellcor pulse oximetry
option. Nellcor and
OxiMax
®
are registered
trademarks of Nellcor Puritan Bennett. Inc.
Only Masimo SET SpO2sensors, including
LNCS series of cables and NR value disposable
sensors, should be used with Masimo pulse
oximetry. Masimo, Masimo SET, LNCS and NR
are registered trademarks of Masimo Corporation.
This monitor is approved for use during all
phases of flight aboard U.S. Army aircraft.
MIL STD 461E
Army: CE101, CE102, CS101, CS114, CS115, CS116, RE102, RS103