Welch Allyn assumes no responsibility for injury or for any illegal or improper use of the product that may result from failure
to use this product in accordance with the instructions, cautions, warnings, or indications for use published in this manual.
Welch Allyn, Acuity, Micropaq, and FlexNet
Nellcor
is a registered trademark of Nellcor Puritan Bennett.
are registered trademarks of Welch Allyn.
Masimo and SET are registered trademarks of Masimo Corporation.
Software in this product is copyright by Welch Allyn or its vendors. All rights are reserved. The software is protected by
United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the
licensee is entitled to use the copy of the software incorporated within this instrument as intended in the operation of the
product in which it is embedded. The software may not be copied, decompiled, reverse-engineered, disassembled or
otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title
and ownership of the software remain with Welch Allyn or its vendors.
For information about any Welch Allyn product, please call Welch Allyn Technical Support:
USA+ 1 315 685 4560
Canada 800 561 8797China+ 86 216 327 9631
European Call
The Micropaq® monitor is intended to be used by clinicians for single or multiparameter
vital signs monitoring of ambulatory and nonambulatory pediatric and adult patients in
health care facilities. The monitor is able to withstand light rain exposure over short
periods of time (uniform distribution of approximately 1 mm of water per minute for 10
minutes or less).
The Micropaq monitor is intended to operate with an Acuity
through wireless communication over the Welch Allyn
connects multiple devices to the Acuity Central Monitoring System through hardwired
Ethernet networks and Wireless Local Area Networks (WLANs). If the Micropaq monitor
is moved out of range or loses communication with the FlexNet network, it continues to
monitor the patient, display patient data, and generate local patient alarms or alert
messages.
•The ECG channel is intended primarily for five-lead ECG monitoring, although
•The Pulse Oximetry channel is intended for continuous noninvasive monitoring of
The most likely locations for patients monitored by this device are step-down units,
telemetry departments, general medical/surgical floors, emergency departments, and inhospital transport.
®
®
three-lead ECG monitoring is supported.
functional oxygen saturation of arterial hemoglobin (SpO
(measured by an SpO
sensor).
2
Central Monitoring System
FlexNet® network. FlexNet
) and pulse rate
2
This guide was written for clinicians. Although this guide may describe some monitoring
techniques, Welch Allyn expects that the operator is a trained clinician who knows how to
take and interpret a patient’s vital signs.
Federal USA law restricts sale of the device identified in this manual to, or on the order of,
a licensed medical practitioner.
2Chapter 1 General informationWelch Allyn Micropaq Monitor
IPX1
C
US
Li++
Li++
Symbols
Warning Warning statements in this manual identify conditions or practices that could result in personal
injury.
Caution Caution statements in this manual identify conditions or practices that could result in damage to the
equipment or other property.
Caution On the product, means “Consult the accompanying documentation.”
The following symbols appear on the monitor or accessories.
SymbolDefinitionSymbolDefinition
Direct currentEnclosure Protection Drip proof: Classification
IPX1 per EN60529: 1991
Alternating current (battery charger)Fuse
The CE Mark and Notified Body Registration
Number signify the device has met all essential
requirements of European Medical Device
Directive 93/42/EEC
Restrictions for use of wireless device in
Europe. European Communities Class 2 radio
equipment
This device has been tested and certified by
the Canadian Standards Association
International to comply with applicable U.S.
and Canadian medical safety standards.
Signifies the device has met all essential
requirements of European Medical Device
Directive 93/42/EEC for a Class 1 product
(battery charger)
Protective earth ground (battery charger)Separate batteries from other disposables for
recycling
Lithium Ion batteryFor indoor use only (battery charger)
Caution: Refer to Directions For Use and
Keep away from rain
accompanying documentation
See the accompanying manualRecycle the monitor and battery
separately from other waste. Refer to
www.welchallyn.com/weee for collection
point and additional information.
Alarm(s) offPatient connections are Type CF, isolated for
direct cardiac application, and protected
against defibrillation
This way upStacking limit (by number)
Humidity limitAltitude limit
FragileIATA/ICAO Hazard Class 9 Package
(International Air Transport Association/
9
International Civil Aviation Organization)
Temperature limitsSingle use only
Directions for UseChapter 1 General information3
This device complies with FCC and
Industry Canada requirements for
international radiators (802.11 wireless)
FCC ID:
PGUWA11AO7
IC:
4168a-WA11A07
Symbol DefinitionSymbolDefinition
Non-ionizing electromagnetic radiation. This
device contains an approved RLAN module
frequency 2402 to 2480 MHz
The monitor is connected to AcuityThe monitor is not connected to Acuity
(Flashing) The monitor is searching for a
connection to Acuity
Not magnetic resonance safe.
Monitor Front Panel Keys
Select Key and Silence Patient Alarm/
Equipment Alert Key- Selects the choice
highlighted on the menu. During patient alarms,
silences the tone at the monitor and at Acuity (if
connected) for 90 seconds. During equipment
alerts, silences or acknowledges (dismisses) the
alert.
Snapshot Key - When connected to Acuity,
pressing this key sends Acuity a snapshot print
to the Acuity central station printer. A total of
21 seconds of patient numeric and waveform
data (14 seconds of history, 7 seconds after the
key is pressed) will be sent to the printer. See
”Snapshot key” on page 10 for more
information.
Non-ionizing electromagnetic radiation. This
device contains an approved RLAN module
frequency 5150 to 5825 MHz
Scroll Up Key and Reset Alarm Tone KeyScrolls up menus on the display. During patient
alarms, resets the tone at the monitor and at
Acuity (if connected).
Scroll Down Key and Main Menu Key- Pressing
this key scrolls down menus on the display, or
causes the Main Menu to appear if no menu is
displayed.
4Chapter 1 General informationWelch Allyn Micropaq Monitor
Battery charger labels and LEDs
Eight-bay battery charger
(008-0651-XX)
Green LED on continuouslyBattery is fully charged.
Green LED flashing
Green LED flashing very slowlyBattery detected and waiting to be charged.
Yellow LED on continuouslySomething is wrong with the battery or the
kj
General warnings and cautions
Familiarize yourself with all warnings and cautions before using the monitor.
WARNING When considering a treatment protocol that involves wireless
communication of patient data, be sure to recognize some limitations inherent
in wireless communications. When the monitor is not connected to the
network:
•There are no patient alarms or alerts at the Acuity Central Station.
•Acuity does not perform arrhythmia and ST analysis on the patient data
and does not generate related alarms.
•Patient data is not saved.
WARNING Do not try to monitor neonatal patients with the monitor. The
monitor is intended for adult or pediatric patients. It is not intended for use with
pediatric patients (or infants) weighing less than 22 lbs (10 kg).
Battery is charging.
charger. (See ”Battery Status and Possible
Response” on page 44.)
WARNING Always check the patient mode at Acuity when monitoring a new
patient. The patient mode determines default alarm limits and internal algorithm
settings.
WARNING The monitor may not meet its performance specifications if stored
or used outside the specified temperature and humidity ranges.
WARNING Do not connect more than one patient to a monitor. Do not connect
more than one monitor to a patient.
WARNING During defibrillation, keep the discharge paddles away from ECG
and other electrodes, as well as other conductive parts in contact with the
patient.
WARNING Do not operate this product in the presence of flammable
anesthetics or other flammable substances in combination with air, oxygenenriched environments, or nitrous oxide; explosion can result.
WARNING Do not use the monitor in a Magnetic Resonance Imaging (MRI)
suite or a hyperbaric chamber. Such use can cause fire or explosion resulting in
patient injury and monitor damage.
Directions for UseChapter 1 General information5
WARNING Magnetic Resonance (MR) Safety. This monitor is designated Not
MR Safe, as defined in the ACR Guidance for Safe MR Practices, 2007. Do not
place this monitor in Zones 3 or 4, in MR environments for which Zones have
been established and marked in accordance with the Guidance. In MR
environments for which Zones have not been designated, do not place this
monitor into a magnetic field with field strength larger than the field strength
that your facility permits to exist in areas that are accessible by personnel,
patients, and guests without MR safety precautions. Failure to follow these
precautions creates a risk that the magnetic field will damage the monitor,
causing it to overheat and potentially burn patents and cause a fire. The damage
to the monitor caused by exposure to magnetic fields may be immediate or
latent. If the damage is latent, it can cause the monitor to overheat at some time
subsequent to its exposure to the magnetic field.
WARNING Electronic equipment that emits strong electromagnetic or radio
frequency signals can cause electrical interference with monitor operation. This
interference may distort the ECG signal, thereby preventing accurate rhythm
analysis. Avoid operating this device near equipment of this type.
WARNING Exposure to Radio Frequency (RF) radiation. To comply with Federal
Communications Commission (FCC) RF exposure requirements, this device
shall be used in accordance with the operating conditions and instructions
provided in this manual, including the section ”Install the carrying pouch” on
page 28.
WARNING Pacemaker signals can differ from one pacemaker to the next. The
Association for Advancement of Medical Instrumentation (AAMI) cautions that
“in some devices, rate meters may continue to count the pacemaker rate during
occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon rate
meter alarms. All pacemaker patients should be kept under close or constant
observation.” See ”ECG specifications” on page 53 for disclosure of the
pacemaker pulse rejection capability of this instrument.
WARNING This wireless medical device was tested and, when used with a
metal-free accessory between the monitor and the patient, complies with FCC
RF Exposure (SAR) guidelines. The use of accessories containing metal may not
ensure compliance with FCC RF exposure guidelines. Specific Absorption Rate
(SAR) is a measurement of radio frequency energy. The FCC permits a
maximum SAR value of 1.6 mW/g. The highest SAR value for this patient
monitor, when worn by a patient in accordance with the directions for use, is
0.560 mW/g.
WARNING Military radars are allocated as primary users in the bandwidths
between 5.25 - 5.35 GHz and 5.47 to 5.725 GHz. In the event a radar signature
is detected, the Access Point moves to a new channel, which can temporarily
interrupt patient monitoring. If the device is operated near a military radar, the
radar could cause damage to the device.
WARNING Changes or modifications not expressly approved by Welch Allyn
could void the purchaser’s authority to operate the equipment. This product
does not contain any user serviceable components. Any unauthorized product
changes or modifications will invalidate Welch Allyn’s warranty and all applicable
regulatory certifications and approvals.
WARNING Motion artifact can affect the accuracy of patient vital sign
measurements. Minimize patient motion whenever possible.
6Chapter 1 General informationWelch Allyn Micropaq Monitor
WARNING For patients with a pacemaker, position the monitor to maintain a
minimum 6-inch distance between the monitor and pacemaker. Immediately
turn the monitor off and provide appropriate patient care if you have any reason
to suspect that the monitor is interfering with the pacemaker. The Health
Industry Manufacturers Association recommends this minimum 6-inch distance
between a hand-held wireless radio and a pacemaker, which is consistent with
the independent research by, and recommendations of, Wireless Technology
Research.
WARNING Make frequent electrical and visual checks on cables, sensors, and
electrode wires. All cables, sensors, and electrode wires must be inspected and
properly maintained and in proper working order to allow the equipment to
function properly and protect patient safety.
WARNING Avoid electrosurgery burns at monitoring sites by ensuring proper
connection of the electrosurgery return circuit so that the return paths cannot be
made through monitoring electrodes and probes.
WARNING Use of ECG and SpO
cables not specified by Welch Allyn may
2
negate defibrillator protection and risk patient injury.
WARNING Use of Masimo LNOP
®
sensors/cables will not provide protection
in accordance with IEC defibrillation standards when used with this device.
WARNING To ensure patient safety, the conductive parts of the ECG
electrodes (including associated connectors) and other patient-applied parts
should not contact other conductive parts, including earth ground, at any time.
WARNING Use only accessories supplied by Welch Allyn or recommended in
the Welch Allyn Products and Accessories booklet (810-0409-XX). The monitor
will only meet the listed specifications when using accessories listed by Welch
Allyn. Use accessories according to your facility’s standards and the
manufacturer’s recommendations. Always refer to the manufacturer’s Directions
for Use.
WARNING As with all medical equipment, carefully route the patient cabling to
reduce the possibility of patient entanglement or strangulation. Use the supplied
garment clips to secure the cable properly.
WARNING When positioning the monitor pouch on the patient, make sure the
straps do not entangle the patient’s neck or cause choking. Make sure the straps
do not restrict the movement of the patient’s limbs or create a hazard when
walking or moving.
WARNING If a product has been dropped or severely abused, send it to a
qualified service person to confirm proper operation.
WARNING Do not use the pulse oximeter as a replacement or substitute for
ECG-based arrhythmia analysis.
Caution Do not autoclave the monitor. Autoclave accessories only if the
manufacturer’s instructions clearly approve it. Many accessories can be severely
damaged by autoclaving.
It is possible for the monitor to detect a problem that prevents the monitor from operating
properly. If this occurs, the monitor displays an error message and error number. Report
such errors to Welch Allyn. The monitor should be serviced only by a Welch Allyn service
technician while under warranty. Contact Welch Allyn for information about post-warranty
period service.
Directions for UseChapter 1 General information7
Introducing the monitor
The monitor is a patient-worn vital signs monitor for use by adult or pediatric ambulatory
patients.
•One or two ECG channels displayed
•Up to 2 ECG leads displayed at the monitor:
I, II, III, V, aV
•Up to 7 ECG leads displayed at Acuity:
I, II, III, V, aV
•One ECG lead displayed at the monitor and
at Acuity: Fixed lead II with 3-lead cable, or
5-lead cable with only RA, LA and LL
electrodes attached.
•Pulse oximetry (SpO
408 only)
•Two-way wireless communication within
Welch Allyn’s FlexNet network
•LCD for display of ECG waveforms, SpO
and heart rate/pulse rate data, and
messages from Acuity
•Standalone operation with patient alarms
when out of range of the network
•Patient alarm limits that can be set at the
monitor or at Acuity
•Configurable formats for single- or dual-waveform ECG display
•Internal antenna
•Snapshot key
•Lightweight (less than two pounds with battery)
•Rugged and tolerant of brief water exposure
•Rechargeable battery
•Sleep mode to extend battery life
•Your model may be shipped with an attached identification number on the front of the
monitor.
, aVL, or aVF with 5-lead cable
R
, aVL, or aVF with 5-lead cable
R
) monitoring (Model
2
2
Model 406
Model 408
ECG monitoring
ECG monitoring and either one of two pulse oximetry (SpO2) monitoring options:
•SpO
•SpO
with Masimo® SET® technology, indicated by:
2
with NELLCOR® OxiMax® technology, indicated by:
2
8Chapter 1 General informationWelch Allyn Micropaq Monitor
HR
SpO2
80
97
FlexNet Network
Micropaq monitors
Access Points
To Other
Access
Points
To Other Acuity
Central Stations
Acuity Central Station
HR
SpO2
80
97
Understanding the monitor and the FlexNet Network
The monitor is intended to operate with an Acuity® Central Station as part of Welch Allyn’s
FlexNet network. FlexNet allows multiple devices to communicate through hardwired
Ethernet networks and Wireless Local Area Networks (WLANs). The Acuity Central
Station provides the primary display and entry of patient data for a patient connected to
the monitor.
HR
SpO2
80
97
HR
SpO2
80
97
Each patient-worn monitor supports two-way communication with an Acuity Central
Station through an access point in the FlexNet network. The access point is a digital radio
transceiver that connects to the FlexNet network. During monitoring, the monitor sends
the patient data to Acuity. Acuity and the monitor continuously analyze the data. Acuity
provides appropriate alarm or alert messages at the Central Station and other network
devices such as a hallway message panel and the monitor itself. Acuity also stores the
patient data for viewing or report printing.
If the monitor is moved out of range or loses communication with the FlexNet network
and Acuity, it continues to monitor the patient and display patient data. While not
communicating with Acuity, the monitor continues to generate local patient alarms or
alert messages. Patient data is not stored and Acuity does not perform waveform analysis
or generate arrhythmia messages while the monitor is not communicating with Acuity.
When the monitor is returned to within range of the FlexNet network, it automatically
reconnects to Acuity.
Directions for UseChapter 1 General information9
SpO2 Connector
ECG Connector
Visual Alarm Indicator
Liquid Crystal Display
Snapshot key
Scroll Up Key and Reset Patient Alarm Tone Key
Scroll Down Key and Main Menu Key
Select Key and
Silence Patient Alarm/
Equipment Alert Key
Back
Battery pack latch
Battery
SpO
2
connector clip
(model 408) or
cover (model
406)
Monitor features
Controls and connectors
10Chapter 1 General informationWelch Allyn Micropaq Monitor
Note
Note
Visual alarm indicator
GreenFlashes slowly during normal operation.
RedFlashes during patient alarm, remains on continuously when alarms are
silenced or suspended.
YellowFlashes during an equipment alert or while not connected to the network.
Remains on continuously if the operator suspends an alert at Acuity for 90
seconds or acknowledges (dismisses) a low battery alert from the monitor
or Acuity.
The flashing green LED indicates that the monitor is connected to the network
but not necessarily connected to a patient. If the monitor is actively monitoring a
patient, the green LED indicates no alarms or alerts are detected.
Audible alarm indicator
Beeps to indicate a patient alarm, and beeps faster for life-threatening arrhythmia
alarms (see ”Patient alarm and equipment alert specifications” on page 58).
Snapshot key
Beeps to indicate when the equipment needs attention. This beep tone is slower than
patient alarm tones (see ”Patient alarm and equipment alert specifications” on
page 58).
Volume can be configured as high, low, or off (configured at Acuity).
Volume can be configured differently for network connection or stand-alone operation
(configured at Acuity).
When connected to Acuity, pressing this key sends a snapshot of the patient’s
numeric and waveform data to the Acuity Central Monitoring System. Depending on
how Acuity is configured, this will cause Acuity to print a 21-second snapshot (14
seconds of history, 7 seconds of data after the button is pressed) to the Acuity central
station printer.
Snapshot is the default selection of the monitor. However, the connected
monitor will inherit the configuration previously defined by Acuity. For example,
if Acuity has defined the Snapshot key to respond with a Nurse Call function and
a new monitor is introduced to the system, the Snapshot key definition will
remain as Nurse Call.
For more information about using the Acuity Central Monitoring System, refer to
Acuity Directions For Use.
Scroll Up key and Reset Patient Alarm Tone key
Scrolls up menus on the display.
Resets a silenced patient alarm tone.
Directions for UseChapter 1 General information11
Note
Scroll Down key and Main Menu key
Scrolls down menus on the display.
Displays the Main Menu.
Select key and Silence Patient Alarm/Equipment Alert key
Selects the choice highlighted on the menu.
During patient alarms, silences the tone at the monitor and Acuity (if connected) for
90 seconds. During equipment alerts, silences or acknowledges (dismisses) the alert
at the monitor and Acuity.
Battery
Insert the battery to turn on power. Remove the battery to turn off power. (While the
battery is removed, the monitor does not perform patient monitoring.)
If you do not use END TELE to disconnect from the network as described
above, the Acuity Central Station generates a DROPOUT equipment alert at
Acuity.
If you want to monitor this same patient at a later time, you will need to reselect
the patient name from the monitor or confirm the patient ID at Acuity.
Recharge the battery while it is removed from the monitor. (See ”Recharge a battery”
on page 43.)
To order a new battery, see ”Battery Status and Possible Response” on page 44.
12Chapter 1 General informationWelch Allyn Micropaq Monitor
Patient name as entered at
Acuity.
If the patient name has not
been entered, the monitor
displays the last four digits of
the monitor serial number, such
as:
ID:6472
SpO2 numeric data is a
percentage value.
HR indicates the heart
rate is from ECG.
PR (pulse rate) is
displayed if SpO
2
is active
and ECG is not (pulse rate
is derived from SpO2).
If the monitor detects
a vital sign outside
the measurable range,
it displays:
- - - (below the range)
+ + + (above the range).
ECG lead is selectable
Low battery icon flashes to
indicate monitor will shut off in
30 minutes or less.
Indicates the monitor
is connected to
Acuity.
Waveform scale is
selectable
Indicates one or
more patient alarms
are disabled (off).
ECG waveform is displayed
when active.
Dashed lines indicate the
monitor detects a pacemaker
signal (display of pacer
detection can be enabled or
disabled at Acuity)
This symbol indicates the monitor is not
communicating with Acuity:
Flashing indicates the monitor is associated
with an access point, but not communicating
with Acuity.
Continuous on indicates the monitor is not
communicating with an access point or Acuity.
Symbol is displayed at the
monitor whenever the
Snapshot key is pressed.
SpO2 pulse amplitude indicator
(not proportional to pulse
volume)
Display
Although the Acuity Central Station is the primary location for viewing patient data, the
monitor provides information to support patient care.
Display sleep mode
In order to extend battery life, the display becomes blank after two minutes if no keys are
pressed. The display becomes active again if an alarm or alert occurs, a key is pressed,
the initial Acuity connection occurs, a cable is inserted, or an electrode is attached.
The display will not become blank if a patient alarm is occurring, an Acuity message is
displayed, or the monitor is in Demo mode or Service mode.
Directions for UseChapter 1 General information13
Press to select or change the highlighted choice.
Press to move through the menu.
Main Menu
When you first press , the Main Menu appears:
EXITExit the Main Menu (the menu disappears).
ACUITY...Access the Acuity Menu with network options. The Acuity Menu
is only accessible while connected to Acuity.
EXITExit all menus and return to the monitoring
screen.
END TELEDiscontinue monitoring a patient.
NEW ROOMReassign a patient to a new room in the same
unit.
TRANSFERTransfer a patient to a new room in a new unit.
NEW PATIENTAssign the monitor to a new patient.
PATIENT INFODisplay patient information such as ID, name,
unit and room.
Whenever the monitor is connected to Acuity and you select
ACUITY... from the Main Menu, the monitor displays the message
ACUITY CONTACTED to confirm that Acuity has been contacted.
The monitor continues to display this message until Acuity
responds, or you press to acknowledge the message and clear
the screen. If the monitor detects an alarm or alert, it clears the
screen to display the appropriate alarm or alert message. The
length of time required for Acuity to respond to your selection at
the monitor can vary widely depending on the amount of network
traffic and other conditions.
ECG LEAD...Access a menu to change the ECG 1 or ECG 2 lead selection (I, II,
III, aV
, aVL, aVF, or V). Available vectors depend on the connected
R
electrodes.
14Chapter 1 General informationWelch Allyn Micropaq Monitor
Note
ECG SCALE...Change the scale of the ECG waveform. If two waveforms are
displayed, both have the same scale.
1 WAVEFORMThere are four possible ECG waveform display selections:
1 WAVEFORMthe default selection
2 WAVEFORMS
5 SECONDS
FULL SCREEN
Pressing changes to the next selection. This change does not
take effect until after you exit the Main Menu. See ”Display” on
page 12 for descriptions.
LIMITS...Enter the Alarm Limits Menu (”Customize patient alarm limits at
the monitor” on page 37) and change alarm limits.
SYSTEM
INFORMATION
Display information about the network connection and SpO
module.
SERVICE MENU Enter Service Mode for a demonstration mode (Demo, see
”Demonstration mode” on page 17) or service functions for
technicians. Service Mode is not available if any cables are
plugged in.
To restrict access to the Main Menu, a Menu Lock option can be configured for
the monitor at the Acuity Central Station. When the Menu Lock is enabled, the
operator must press and hold down and for two seconds to gain access
to the Main Menu. The Menu Lock is disabled if the monitor loses
communication with Acuity.
2
Directions for UseChapter 1 General information15
1 Waveform
The single ECG 1 (lead II)
waveform is displayed.
Full Screen
The single ECG 1 (lead II) waveform is
allowed to occupy most of the screen.
2 Waveforms
ECG 1 (lead II) and ECG 2
(lead V) are both displayed.
5 Seconds
ECG 1 (lead II) cascades from
one line to the other.
Waveform options
There are four ECG waveform options as shown:
To change the waveform selection during operation:
1.Press to display the Main Menu.
2. Press as needed to highlight the current waveform selection. Then press as
needed to select the desired display.
16Chapter 1 General informationWelch Allyn Micropaq Monitor
Messages from Acuity
The monitor displays messages sent from Acuity as needed, including patient alarms and
equipment alerts. When Acuity messages are displayed, they temporarily override
information displayed on the lower half of the monitor screen.
Accessories
Battery charger (8-battery)Micropaq Directions For Use
BatteryECG electrodes
3-lead ECG cable (optional)5-lead ECG cable
ECG extension cable (optional)Carrying pouch
sensors (Masimo or Nellcor)SpO2 cable (Masimo or Nellcor)
SpO
2
WARNING Use only accessories supplied by Welch Allyn or recommended in
the Welch Allyn Products and Accessories booklet (810-0409-XX). The monitor
will only meet the listed specifications when using accessories listed by Welch
Allyn. Use accessories according to your facility’s standards and the
manufacturer’s recommendations. Always refer to the manufacturer’s Directions
for Use.
Operating settings
The following monitor operating settings can be set at the monitor or at the Acuity Central
Station:
•Patient alarm limit settings (ECG and SpO
•ECG lead and scale selection
•ECG display format
Many other monitor operating settings (such as patient mode and alarms volume) can
only be set at the Acuity Central Station. See ”Operating settings” on page 51 for a list of
all settings and where they are set.
Default settings
When the monitor connects to Acuity for a new patient, the Acuity Central Station
downloads the appropriate default settings stored at Acuity. While the monitor is
connected to Acuity, settings can be changed either at the monitor or at the Acuity Central
Station.
If the monitor is temporarily disconnected from Acuity and the operator changes settings
at the monitor, those settings are transmitted to and stored at Acuity when the monitor
reconnects.
).
2
Directions for UseChapter 1 General information17
Power-Up Screen
Main Menu
Demonstration mode
You can practice using the monitor with the Demo mode of operation, including
connection to Acuity.
The Demo mode cannot be activated while you are monitoring a patient or if any cables
have been plugged into the monitor. During the Demo mode, the monitor and Acuity
display the message SIMULATION.
To practice with the monitor in Demo mode:
1.Disconnect all patient cables connected to the monitor.
2. Remove the battery (if installed).
3. Insert the battery and watch for the Power-Up screen.
4. After the Power-Up screen disappears, press to display the Main Menu.
18Chapter 1 General informationWelch Allyn Micropaq Monitor
Service Menu
A simulated waveform and
numerics are displayed.
Demonstration Mode
SIMULATION indicates the
Demo mode is active.
5. Press to highlight SERVICE MENU, then press to display the Service Menu.
6. Press to highlight DEMO MENU, then press to display the Demo Menu.
7.Press to highlight DEMO 1 or DEMO 2, then press to start.
Demo Mode Display Values and Alarm Limits
DisplayDemo 1Demo 2
ECG WaveformNormal sinus rhythm,
normal ST
ECG Heart Rate
SpO2 Pulse Rate
Saturation%9788Lower 90
SpO
2
a. Demo 2 will cause patient alarms.
80125Lower 50
a
Normal sinus rhythm,
normal ST
Alarm Limits (On)
(not applicable)
Upper 120
Upper 100
8. While in Demo mode you can practice changing settings such as ECG lead selection
and alarm limit adjustment. (These changes only affect the Demo mode and are
erased when you exit the Demo mode.)
9. To change to the other Demo selection, press to display the menu, then scroll
down to highlight TOGGLE DEMO MODE and press .
10. To exit the Demo mode, either insert a patient cable or remove and insert the battery.
The monitor restarts and enters the normal monitoring mode.
19
If any patient cables are
connected, there will be
some patient data displayed.
The format depends on the
monitor default settings.
Example of Initial Monitor Screen
The monitor is
searching for a
network
connection.
Possible Acuity Unit
selections.
Press to view more.
Example of Acuity Unit Selection
2
Monitoring
Connect a new patient
Connect to the network
1.Insert a battery into the monitor to turn it on. After a few seconds the monitor PowerUp Screen is replaced by an initial monitoring screen.
2. After the network connection is established, the monitor may prompt you to select an
Acuity Unit (if your facility has more than one Acuity unit):
428-02-2392, Hopkins, Bill J
520-29-0319, Phillips, Mary L
532-94-8372, Smith, Frank R
▼
Select Patient at Central
3. Press or to highlight the desired Acuity unit, then press .
When you press or to highlight the desired Acuity unit and then press ,
your selection will begin to flash between normal and reverse video to confirm that
the monitor is communicating your selection to Acuity. You cannot scroll to another
selection during this time. The selection continues to flash until Acuity responds back
to the monitor. Then the monitor displays the next appropriate screen (such as a list of
possible patients). The length of time required for Acuity to respond to your selection
at the monitor can vary widely depending on the amount of network traffic and other
conditions.
Be sure to select an Acuity unit. Even though the monitor is
connected to the network (as indicated by the green LED and
network connection symbol), the Acuity Central Station may not
display any indication of this monitor until after you have selected an
Acuity unit.
4. The monitor displays a list of possible patients.
If your patient has been pre-admitted to the selected Acuity unit, they will be included
in the list.
5. Scroll through the patient list to look for your patient’s name.
•If your patient is not in the list, highlight Select Patient at Central and press .
The patient name will need to be entered later at the Acuity Central Station.
WARNING If you do not select the patient name at the monitor at this time, do
not adjust any alarm limits until after the patient name and ID are confirmed at
Acuity. When the patient name and ID are confirmed at Acuity, Acuity
downloads the default settings and patient alarm limits for that Acuity unit to the
monitor, thereby overriding any previous settings and alarm limits.
At power-up, the monitor retains the most recent patient mode. The patient
mode can only be changed at Acuity. If the patient is being monitored when the
patient mode is changed, there is a brief interruption in the display and recording
of ECG and SpO
patient data.
2
Directions for UseChapter 2 Monitoring21
Possible
rooms to
select.
Example of Room List
SELECT
ROOM
1104A
1104B
1105A
Select Room at Central
▼
•If your patient is in the list, highlight the name and press . Within a few
seconds the monitor displays a list of unassigned rooms.
•If you want to assign the patient to a room, highlight the room and press .
•If you do not want to assign a room at this time, highlight Select Room at
Central and press . The patient room will need to be entered later at the
monitor (see “Reassign a monitored patient to a new room in the same unit” on
page 32) or at Acuity (see “Monitor patient at Acuity” on page 41).
6. If you need to customize alarm limits for your patient, see “Customize patient alarm
limits at the monitor” on page 37.
Perform ECG monitoring
WARNING Motion artifact can cause incorrect heart rate readings. Minimize
patient motion whenever possible.
WARNING If a disconnected lead is in too close proximity to other electrical
devices, it may cause false heart rate readings.
WARNING The monitor does not provide internal arrhythmia analysis.
Therefore, arrhythmias may cause the monitor to display inaccurate heart rates.
WARNING The monitor will show + + + for HR numerics between 301 and
350 beats per minute. Above 350 beats per minute, it may display incorrectly
low heart rates, due to intermittent picking of R-waves.
WARNING Do not use the monitor in a Magnetic Resonance Imaging (MRI)
suite or a hyperbaric chamber. Such use can cause fire or explosion resulting in
patient injury and monitor damage.
WARNING Pacemaker signals can differ from one pacemaker to the next. The
Association for Advancement of Medical Instrumentation (AAMI) cautions that
“in some devices, rate meters may continue to count the pacemaker rate during
occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon rate
meter alarms. All pacemaker patients should be kept under close or constant
observation.” See “ECG specifications” on page 53 for disclosure of the
pacemaker pulse rejection capability of this instrument.
Loading...
+ 57 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.