Welch Allyn and Flexible Monitoring are registered trademarks of Welch Allyn. POEM and FlexNet are trademarks of Welch
Allyn.
Nellcor and Oximax are registered trademarks of Nellcor Puritan Bennett.
SET, LNCS, and Masimo are registered trademarks of Masimo Corporation. Possession or purchase of a Masimo SpO
-
2
equipped monitor does not convey any express or implied license to use the device with unauthorized sensors or cables
which would, alone or in combination with this device, fall within the scope of one or more of the patents relating to this
device.
Software in this product is copyright by Welch Allyn or its vendors. All rights are reserved. The software is protected by
United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the
licensee is entitled to use the copy of the software incorporated with this instrument as intended in the operation of the
product in which it is embedded. The software may not be copied, decompiled, reverse-engineered, disassembled or
otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title
and ownership of the software remain with Welch Allyn or its vendors.
For information about any Welch Allyn product, call Welch Allyn Technical Support:
This device complies with Part 15 of the FCC rules and with the rules of the Canadian ICES-003. Operation is subject to the
following two conditions: (1) This device may not cause harmful interference and (2) this device must accept any interference
received, including interference that may cause undesired operation.
Caution! Changes or modifications not expressly approved by Welch Allyn could void the purchaser’s authority to operate
the equipment.
810-2811-02 (CD)
Manual Part Number 810-1651-04 Rev A, 09/2011
Welch Allyn, Inc.
8500 SW Creekside Plac
Beaverton, OR 97008-7107 USA
Regulatory Affairs Representative
Welch Allyn Limited
Navan Business Park
Dublin Road, Navan
County Meath, Republic of Ireland
Remove and disconnect the battery........................................58
Separate the front and rear chassis ........................................59
Disassemble the front chassis assembly ....................................60
Remove the LCD display from the main board ...............................62
Disassemble the rear chassis assembly.....................................63
Remove the main board .................................................64
Disassemble and remove the NIBP assembly ................................67
Remove and replace the speaker ..........................................69
Remove and disassemble the printer assembly...............................71
Disassemble the temperature module ......................................79
Disassemble the Nellcor SpO2assembly....................................82
Disassemble the Masimo SpO2assembly ...................................85
Replacement parts ..............................................87
Updates and service bulletins......................................99
DC power cable clamp ..................................................99
ivWelch Allyn Vital Signs Monitor 300 Series
1
1
Safety
All personnel must read and understand all safety information presented in this manual
before using or repairing the monitor.
Caution Use only the Welch Allyn Monitor Service Utility (810-1784-XX) with
this monitor.
Software service tools, such as the “Custom Repair Software” (part number
130S29E), distributed for use with previous models of the Vital Signs Monitor,
must not be used with the Model 300 Series. Use of any tool other than the
Welch Allyn Monitor Service Utility (810-1784-XX) may set the monitor in an
undefined and unrecoverable state.
United States federal law restricts this device to sale, distribution, or use by or on the
order of a licensed medical practitioner.
General safety considerations
Always consider the following safety points when using the monitor:
•Place the monitor and accessories in locations where they cannot harm the patient
should they fall from a shelf or mount.
•Do not connect more than one patient to a monitor.
•Do not connect more than one monitor to a patient.
•Do not use the monitor in an MRI suite or hyperbaric chamber.
•Do not autoclave the monitor.
•Accessories can be autoclaved only if the manufacturer’s instructions clearly approve
it. Many accessories can be severely damaged by autoclaving.
•Inspect the power adapter cord periodically for fraying or other damage. Replace the
adapter as needed. Do not operate the monitor from mains power if the adapter, the
adapter cord, or the cord plug is damaged.
•Frequently check all cables, both electrically and visually.
•To avoid explosion, do not operate the monitor in the presence of flammable
anesthetics.
WARNING Use only accessories approved by Welch Allyn. Visit
www.welchallyn.com. The use of any other accessories can result in inaccurate
patient data, can damage the equipment, and can void your product warranty.
WARNING Always use accessories according to the standards of your facility
and according to the manufacturer's directions for use.
2SafetyWelch Allyn Vital Signs Monitor 300 Series
•A monitor that has been dropped or otherwise damaged or abused must not be used
until it has been tested and verified by qualified service personnel for proper
operation.
•If the monitor detects an unrecoverable problem, an error code and a brief message
appear in the message display. Report all such errors to Welch Allyn.
•While under warranty, the monitor must be serviced only by a Welch Allyn service
technician.
Electrostatic discharge (ESD)
CAUTION
SENSITIVE ELECTRONIC DEVICES
DO NOT SHIP OR STORE NEAR STRONG
ELECTROSTATIC, ELECTROMAGNETIC,
MAGNETIC OR RADIOACTIVE FIELDS.
ATTENTION
OBSERVE PRECAUTIONS
FOR HANDLING
ELECTROSTATIC
SENSITIVE DEVICES
WARNING Electrostatic discharge (ESD) can damage or destroy electronic
components. Handle static-sensitive components only at static-safe workstation.
WARNING Consider all electrical and electronic components of the monitor as
static-sensitive.
Electrostatic discharge is a sudden current flowing from a charged object to another
object or to ground. Electrostatic charges can accumulate on common items such as
foam drinking cups, cellophane tape, synthetic clothing, untreated foam packaging
material, untreated plastic bags, and work folders, to name only a few.
Electronic components and assemblies, if not properly protected against ESD, can be
permanently damaged or destroyed when near or in contact with electrostatically charged
objects. When you handle components or assemblies that are not in protective bags and
you are not sure whether they are static-sensitive, assume that they are static-sensitive
and handle them accordingly.
•Perform all service procedures in a static-protected environment. Always use
techniques and equipment designed to protect personnel and equipment from
electrostatic discharge.
•Remove static-sensitive components and assemblies from their static-shielding bags
only at static-safe workstations—a properly grounded table and grounded floor mat—
and only when you are wearing a grounded wrist strap (with a resistor of at least 1
megohm in series) or other grounding device.
•Use only grounded tools when inserting, adjusting, or removing static-sensitive
components and assemblies.
•Remove or insert static-sensitive components and assemblies only with monitor
power turned off.
•Insert and seal static-sensitive components and assemblies into their original staticshielding bags before removing them from static-protected areas.
•Always test your ground strap, bench mat, conductive work surface, and ground cord
before removing components and assemblies from their protective bags and before
beginning any disassembly or assembly procedures.
Service manualSafety3
Symbols
The symbols illustrated on the following pages appear on the monitor or in this document.
Documentation symbols
WARNING Indicates conditions that could lead to illness, injury, or death.
Caution In this manual, indicates conditions that could damage equipment or other property.
Caution On the product, means “Consult accompanying documentation.”
Certification and operation labels
This device has been tested and certified by the
Canadian Standards Association International to
comply withapplicable U.S. and Canadian medical
safety standards.
Patient connections are Type BF, and protected
against defibrillation.
Australian Registered ImporterSealed lead-acid battery, 6V 4 Ah
Shipping, storing, and environment labels
Keep this end of the package or shipping crate up.Protect the monitor from exposure to rain.
Do not expose the monitor to relative humidity
above this limit.
Fragile contents—handle with care.Limit stacking to this number of units.
Do not expose the monitor to temperatures
outside these limits.
Urgent alarm notification (output to Nurse Call
system)
Recycle used batteries properly and in accordance
with local regulations.
Do not dispose of batteries in refuse containers.
Do not subject the monitor to altitudes outside
these limits.
Monitor connector labels
Temperature Probe Cable ConnectorSpO
RS232 Cable ConnectorAC Power Adapter Cable Connector
Nurse Call Cable ConnectorNIBP Hose Connector
SpO2Sensor Cable Connector
2
4SafetyWelch Allyn Vital Signs Monitor 300 Series
ºC
ºF
M
Printer door label
Press to open the printer doorLoad paper this direction
The monitor front panel controls are described in more detail throughout this document.
Front panel controls
Set alarm limitsPower on/off
Silence alarmsPrint patient data
SYS
DIA
SpO
2
message
window
ºC
Scroll up/down, Scroll forward/back,
Increase/decrease value
(The scroll icon appears as these two
arrows in the documentation.)
This service manual is a reference for periodic preventive maintenance and corrective
service procedures for the Vital Signs Monitor 300 Series.
Corrective service is supported to the level of field-replaceable units. These include some
circuit-board assemblies and some subassemblies, case parts, and other parts. (See
“Replacement parts” on page 87 for a complete list of user-replaceable service parts.)
Note
Repair and replacement of the main board is not supported. All service work on
the main board must be performed by certified and qualified service personnel at
an authorized Welch Allyn service center.
Caution No component-level repair of circuit boards and subassemblies is
supported. Use only the repair procedures described in this manual.
WARNING When performing a service procedure, follow the instructions
exactly as presented in this manual. Failure to do so could damage the monitor,
invalidate the product warranty, and lead to serious personal injury.
This guide provides troubleshooting information, assembly procedures, and instructions
for functional testing and performance verification. It is intended for use only by
technically qualified service personnel.
This guide applies only to the Vital Signs Model 300 Series. For servicing the previous
(52000-series) version of the Vital Signs Monitor, refer to Welch Allyn service manual
95P445E, which is available on the TechView CD (900298-1).
Technical support services
Welch Allyn offers the following technical support services:
Telephone support
Loaner equipment
Service agreements
Service training
Replacement service parts
Factory Service
For information on any of these services, contact Welch Allyn at the Technical Support:
numbers listed on page ii.
6Service overviewWelch Allyn VSM 300 Series
Returning products
To return a product for service, contact Welch Allyn Technical Support and request a
Return Material Authorization (RMA) number.
Note
Welch Allyn does not accept returned products without an RMA.
When requesting an RMA, please have the following information available:
•Product name, model number, and serial number
•A complete return shipping address, including a contact name and phone
number; include any special shipping instructions
•A purchase-order number or credit-card number if the product is not covered by
warranty or service agreement
•A full description of the problem or service request
To ship the monitor, please observe these packing guidelines:
•Remove from the package all hoses, connectors, cables, sensors, power cords,
and other ancillary products and equipment, except those items that might be
associated with the problem.
•Use the original shipping carton and packing materials, or as close an
approximation as possible.
•Include a packing list.
•Write the Welch Allyn RMA number with the Welch Allyn address on the outside
of the shipping carton.
United States federal regulations require that any unit received by Factory Service must
be free from blood-borne pathogens before processing. All incoming products are cleaned
as well as possible, but products that cannot be effectively cleaned cannot be accepted
for repair. Please thoroughly clean all organic residues from the product before shipment.
This will ensure safe receipt, processing and repair, and will help expedite the return of
your monitor.
Product configurations
Model numbers for the configurations are as follows:
Model FeaturesModel Features
53000 Standard (NIBP, Pulse Rate, and MAP)530T0 Standard + Temperature
5300P Standard + Printer530TP Standard + Temperature + Printer
®
53N00 Standard + Nellcor
53NT0 Standard + Nellcor SpO2+ Temperature53ST0 Standard + Masimo SpO2+ Temperature
53N0P Standard + Nellcor SpO
53NTP Standard + Nellcor SpO
SpO
2
+ Printer53S0P Standard + Masimo SpO2+ Printer
2
+ Temperature + Printer 53STP Standard + Masimo SpO2+ Temperature + Printer
2
53S00 Standard + Masimo®SpO
2
Service manualService overview7
Recommended service intervals
Interval or ConditionAction RecommendedProcedurePage
Every6-24months
(per hospital protocols)
Battery does not hold a charge Check battery capacity
All repairs on products under warranty must be performed or approved by Welch Allyn.
Refer all warranty service to Welch Allyn Factory Service or another authorized Welch
Allyn Service Center. Obtain an RMA number for all returns to Welch Allyn Factory
Service – see “Returning products” on page 6.
Caution Unauthorized repairs will void the product warranty.
Complete functional testFunctional verification9
Replace battery
Complete functional testFunctional verification9
Troubleshooting and repair followed by
functional test; if necessary, return to
authorized service center
Functional verification
Disassembly procedure
Troubleshooting and repair
Disassembly procedure
Functional verification
9
55
33
55
9
Non-warranty service
Welch Allyn Factory Service and authorized Service Centers support non-warranty repairs.
Contact any Welch Allyn regional service center for pricing and service options.
Welch Allyn offers modular repair parts for sale to support non-warranty service. This
service must be performed only by qualified end-user biomedical/clinical engineers using
this service manual.
The Welch Allyn Monitor Service Utility supports certain service functions. For
information, see “Welch Allyn monitor service utility” on page 37.
Related documents
TitleReorder number
Vital Signs Monitor 300 Series Directions for Use (Masimo, multilanguage)810-2250-XX
Vital Signs Monitor 300 Series Directions for Use (Nellcor, multilanguage)810-2252-XX
Service menu
To see the service menu, first power the monitor off. Press and holdfor 3 seconds.
Whileis depressed, pressuntil the monitor displays the message Service Mode.
8Service overviewWelch Allyn VSM 300 Series
The monitor runs a self-test and then displays the main software version number
(M: X.XX.XX). Pressrepeatedly to cycle to the menu selection of interest.
SYS
DIA (mmHg)
/min
SpO2 %
3 seconds
SERVICE MODE
Power-up Self-test
RESET TO DEFAULT
Main SW version
M: X.XX.XX
YES
BP CYCLES:XXXXXX
RUN TIME:XXXXX
NO
NIBP SW version
N:XX.XX.XX
POWER OFF
BATTERY:X.XX V
NIBP TEST
Exit Service Mode
SpO
SW version
2
S:X.X.X.X
Temperature SW version
T:X.X
9
3
Functional verification
Functional verification overview
This section describes the procedure for a complete functional test to support
recommended preventive-maintenance schedules.
The verification includes tests for a monitor configured with the printer, temperature, and
SpO2options. Perform only the tests applicable to the actual configuration.
A checklist of the functional tests is provided on “Checklist and test results report form”
on page 31. It is recommended that you print a copy of the checklist each time you
perform the functional verification procedure, so that you can record and save the test
results. If the monitor ever requires service, the records of test results can often facilitate
troubleshooting.
Functional verification does not require opening the monitor case.
Equipment required
This equipment is required for functional verification of a fully configured monitor.
Commercially available general-purpose/medical test equipment
ItemManufacturer part number/specification
Power supplyVariable, 0-8 VDC, 0.75 amperes (minimum), with voltage and current
Digital pressure meterNetech Digimano 1000 or equivalent
AC withstand voltage (hi-pot) testerAssociated Research 3605 or equivalent
functional tester (Nellcor, for testing
SpO
2
the monitor only)
SpO2functional tester (Masimo, for testing
the monitor only)
SpO2extension cable (required for SRCMAX)
SpO2simulator (for testing the monitor and
the SpO
Hi-pot cable connectorsSee “Test connections” on page 29.
Timer (to display elapsed time, in seconds)
sensor)
2
indicators (for 1mA current measurement)
Nellcor SRC-MAX
Masimo Tester REF 1593
Nellcor DEC-8
Fluke (Biotek) Index2 XL/XLFE or equivalent
10Functional verificationWelch Allyn VSM 300 Series
Welch Allyn accessories and test equipment
Order numberDescription
06138-000Temperature test key
(for testing the monitor only)
01802-110, 110 V
01802-220, 220 V
9600 Plus Calibration Tester
(for testing the monitor and the temperature probe)
008-0265-XXNeonatal cuff hose, 96-inch
008-0620-XXNeonatal #1 cuff, disposable, box of 10
020-0702-XXCuff simulator (or equivalent 500 cc test volume)
5200-101ADC power adapter
407590Battery substitution connector
810-1784-XXWelch Allyn Monitor Service Utility
(required for NIBP repair or replacement;
not required for functional testing)
008-0842-XXService Serial Cable (for use with the Welch Allyn Monitor
Service Utility 810-1784-XX)
WA 5088-01Tycos air release valve and small bulb
600-0021-XXNIBP tubing connector, threaded
600-0179-XXTubing, 1’
600-0043-XXTee, plastic
Service manualFunctional verification11
Fah
Functional verification procedure
Reference illustration
renheit
Systolic
window
Diastolic
window
Adult
Pediatric
Neonatal
Message
window
SYS
DIA
SpO2 %
/min
Celsius
Print patient data
Change settings
Monitored
temperature
Increase/Decrease
Forward/Back
Battery charged
Battery charging
Battery low
System/power
Note
Setup
•If the monitor is configured with the temperature option, connect the temperature
•If you are using an optional Welch Allyn Model Calibration Tester, plug it in and set it to
•If the monitor is configured with the SpO2option, connect the SpO2sensor.
Power
On/Off
Review
patient
data
Set
NIBP
interval
Start/
stop
NIBP
Set
alarm
limits
Silence
alarms
Other than the optional NIBP overpressure test (page 18), the tests described
here must be performed as part of a complete functional verification procedure.
probe and insert it into the probe well.
the values that match your tester model:
ModelTemperature
960096.4 °F(35.8 °C)
9600 Plus96.8 °F(36.0 °C)
Battery charge and beeper
1. With the monitor turned off, disconnect the power adapter from the monitor.
2. Verify that the charge LEDis off.
3. Connect the power adapter. The monitor emits a single beep tone.
4. Verify that the charge LEDis on.
12Functional verificationWelch Allyn VSM 300 Series
Note
Depending on the charge level of the battery, the charge LED may be either
flashing or steady.
flashing indicates that the monitor is running on AC, the battery is charging,
and the battery is charged to less than 90% capacity.
steady indicates that the monitor is running on AC, the battery may or may
not be charging, and the battery is charged to at least 90% capacity.
Battery substitution cable setup
1. Disconnect the power adapter.
2. Remove the battery cover and remove and disconnect the battery.
3. Connect the open-ended red (+) and black (–) wires of the battery substitution cable
(407590) to the variable DC power supply.
4. Set the power supply to 6.0 V ± 0.25 V.
WARNING Do not exceed 8.0 V.
5. Connect the test power cable to the battery connector on the monitor.
Monitor-off current
With the monitor powered down, verify that the current draw from the power supply is
less than 1 mA.
Power-on self-test
1. Power the monitor on.
Note
2. Verify that the start-up tone (double beep) is audible.
3. Verify that all front-panel lights (background indicators, LCD pixels, and LED segments
If the monitor displays error E38, power the monitor off and then power it on
again.
and periods) come on in the proper order: left, center, and right.
Initialization/idle mode current
Note
1. If the temperature option is present:
If your monitor is configured without the temperature option and without the
SpO2option, skip these steps and proceed to “Baseline current draw” on
page 13.
a. Verify that the temperature probe is in the probe well.
b. Set the temperature mode to MONITOR.
c. Remove the temperature probe from the probe well.
d. Verify that the temperature reading appears within 4 seconds.
Service manualFunctional verification13
e. Do not return the probe to the probe well.
2. If the SpO2option is present:
a. Verify that the SpO2sensor cable is connected to the monitor.
b. Verify that the current draw from the bench power supply is less than 800 mA.
3. Disconnect the SpO2sensor (if equipped).
4. Insert the temperature probe (if the monitor is so equipped) into the probe well.
Baseline current draw
1. With the monitor powered on, wait for the monitor LEDs to blank. In this state, the
SpO2 % reads - -, the time of day is displayed in the message window, and the rest of
the displays are blank.
2. Note and record the exact current from the power supply. (This value will be used in
the NIBP and printer tests.)
Battery voltage
1. Power the monitor off.
2. Simultaneously press and holdandto bring up the monitor in SERVICE MODE.
(When the monitor completes the power-on self-test in service mode, the main
software version number appears in the message display.)
3. Pressrepeatedly until BATTERY VOLTAGE appears in the message display.
4. Verify that the displayed battery voltage is within 0.1 volt of the DC power supply
input.
5. Exit Service Mode by turning off the monitor and then turning it on again.
14Functional verificationWelch Allyn VSM 300 Series
NIBP
Note
Characterization test
1. Attach a neonate hose (part 008-0265-01) to the NIBP fitting on the monitor.
2. Prepare the 60-mL syringe as follows, with reference to the illustration below:
a. Move the syringe plunger to the 35 mL line.
b. Drill a small hole (for example, 9/64 inch) through the syringe and the plunger
c. Insert a rod or bolt (for example, a 6-32 screw) through the hole so that the
The tests described in this section are to be performed only as part of a complete
functional verification procedure.
shaft, at a location between the plunger and the top of the syringe.
plunger cannot move, creating a constant volume in the syringe of 35 mL ± 2 mL.
Secure the rod or bolt so that it cannot fall out of the hole.
-- 60
--
-- 50
--
-- 40
--
-- 30
--
-- 20
--
-- 10
--
ml
3. Insert the tip of the syringe into the open end of the neonate hose. Verify that the fit
is tight and secure.
4. Set the monitor patient type to Adult, as follows:
a. Pressonce.
b. If the monitor is not already in Adult mode, pressonce or twice until Adult
appears in the message display.
c. Press.
5. Press.
6. Verify that the error code C03 appears in the message display within a few seconds.
WARNING If the error code C03 does not appear, characterize NIBP according
to the instructions presented on page 36, and then repeat the NIBP
characterization test.
WARNING Do not use the monitor if it does not pass the NIBP characterization
test. If the NIBP module is not properly characterized, the monitor could
overinflate a neonatal cuff, which could create a hazard for neonatal patients.
WARNING If you cannot characterize the NIBP module, remove the monitor
from service immediately and return it to Welch Allyn for service. (See “Returning
products” on page 6.)
Service manualFunctional verification15
Leak test
This tests the NIBP system for abnormal leakage.
If the NIBP system is leaking, check for leaks in the external plumbing before opening the
case to look for internal leaks.
If you determine that any NIBP module component is causing abnormal leakage, you
must replace the NIBP module before returning the monitor to clinical use.
Caution Do not attempt component-level repair of the NIBP module. Any
change to the NIBP module requires that the monitor be returned to the factory
for NIBP module calibration.
1. Put the monitor in Service mode:
a. Power the monitor off.
b. Power the monitor on while pressing.
WARNING Do not connect the monitor to a patient while the monitor is in
Service mode. Overpressure detection is disabled while the monitor is in NIBP
test mode.
2. Attach a #1 neonatal cuff and hose to the monitor
Wrap the cuff securely around a solid cylindrical object of circumference between 1.6
and 1.9 inches (4.1 and 4.8 cm).
3. Pressrepeatedly until NIBP TEST appears in the message window and 0 is
displayed in the SYS and DIA windows.
Note
4. Pressonce to select 80 mmHg (10.7 kPa). The cuff inflates to approximately 115
Note
When you first enter the NIBP test mode, give the monitor about a minute to
initialize NIBP before you change the target test pressure.
When switching from one target pressure to the next, give the monitor time to
fully inflate and stop before you select the next target pressure.
In the NIBP test mode, pressrepeatedly to select the target NIBP test
pressure. The target pressure is displayed on the DIA LEDs. The measured
instantaneous pressure determined by the monitor is displayed on the SYS LEDs.
mmHg (15.3 kPa).
In the NIBP test mode, and especially at small test volumes, the pressure
achieved can vary significantly (30-40 mmHg or 4-5.3 kPa) from the target
pressure.
5. Wait 15 seconds, and note the current pressure.
6. Wait another 10 seconds and verify that the pressure has not dropped more than
8 mmHg (1.1 kPa) below the pressure noted in step 5.
If the pressure drop is greater than 8 mmHg (1.1 kPa), check the cuff, the hose, and all
external connections for leaks, and then repeat from step 3.
16Functional verificationWelch Allyn VSM 300 Series
7. Pressseveral times to select 0 mmHg (0 kPa). The valve opens to release
pressure.
8. Disconnect the neonate cuff.
Pressure calibration verification
This tests pressure readings on the monitor against a calibrated external pressure meter.
1. Put the monitor in Service mode:
a. Power the monitor off.
b. Power the monitor on while pressing.
WARNING Do not connect the monitor to a patient while the monitor is in
Service mode. Overpressure detection is disabled while the monitor is in NIBP
test mode.
2. Connect the monitor to an adult cuff or a 500 cc test volume, a pressure meter or a
manometer, and a pump bulb, as shown.
3. Pressrepeatedly until NIBP TEST appears in the message window and 0 is
displayed in the SYS and DIA windows.
Service manualFunctional verification17
4. Pressonce to select 80 mmHg (10.7 kPa). The cuff quickly inflates to
approximately 80 mmHg (10.7 kPa), and then settles at a slightly lower pressure level.
Wait a few seconds for the pressure to stabilize.
5. Verify that the value displayed in SYS is within 3 mmHg (0.4 kPa) of the value
displayed on the digital pressure meter.
6. Pressto select 150 mmHg (20 kPa) target pressure. The cuff quickly inflates to
approximately 150 mmHg (20 kPa), and then settles at a slightly lower pressure level.
Wait a few seconds for the pressure to stabilize.
7. Verify that the value displayed in SYS is within 3 mmHg (0.4 kPa) of the value on the
digital pressure meter.
8. Pressto select 300 mmHg (40 kPa). The cuff quickly inflates to approximately 300
mmHg (40 kPa), and then settles at a slightly lower pressure level. Wait a few
seconds for the pressure to stabilize.
9. Verify that the value displayed in SYS is within 6 mmHg (0.8 kPa) of the value on the
digital pressure meter.
Valve and pump current, inflation, and deflation tests
Note
To test pump current
1. Put the monitor in Service mode:
2. Pressrepeatedly until NIBP TEST appears in the message window and 0 is
3. Pressto select 0 mmHg (0 kPa) target pressure.
4. While watching the current meter, pressto select 80 mmHg (10.7 kPa) target
5. Note the highest current reading during inflation.
6. While the pump is running, verify that the reading on the current meter is not more
Replace the internal battery with an external power supply. (See “Battery
substitution cable setup” on page 12.)
a. Power the monitor off.
b. Power the monitor on while pressing.
WARNING Do not connect the monitor to a patient while the monitor is in
Service mode. Overpressure detection is disabled while the monitor is in NIBP
test mode.
displayed in the SYS and DIA windows.
pressure.
than 750 mA above the current level noted in step 2 of the verification test (“Baseline
current draw” on page 13).
7. Pressthree times to select 0 mmHg (0 kPa) target pressure.
To test inflation
1. Pressrepeatedly until NIBP TEST appears in the message window and 0 is
displayed in the SYS and DIA windows.
18Functional verificationWelch Allyn VSM 300 Series
2. Pressonce to select 80 mmHg (10.7 kPa).
3. Wait for the pump to start and stop.
4. Pressonce to select 150 mmHg (20 kPa).
5. Wait for the pump to start and stop.
6. Bleed the pressure to 0 by opening the relief valve on the bulb.
7. Close the relief valve on the bulb.
8. Have a timer ready.
9. Pressonce to select 300 mmHg (40 kPa), and immediately observe the
manometer.
10. As soon as the manometer reads 5 mmHg (0.67 kPa), start the timer.
11. When the manometer reaches 250 mmHg (33.3 kPa), stop the timer.
12. Verify that the elapsed time (step 11 minus step 10) is less than 8 seconds.
To test deflation
1. If the pressure has dropped more than 10 mmHg, use the pump bulb to raise it to 300
mmHg ± 5 mmHg.
2. While the pressure is still at approximately 300 mmHg (as shown by the manometer
and the SYS window), pressonce to select 0 mmHg, and immediately start the
timer.
3. After 10 seconds, verify that the manometer reads less than 15 mmHg.
If you are doing the NIBP overpressure test, skip to “Overpressure tests (optional)”
on page 18.
4. Disconnect the hose from the monitor.
Overpressure tests (optional)
Note
These tests demonstrate that:
Redundant circuitry in the VSM 300 series monitor guarantees that the bloodpressure cuff cannot overinflate.
The allowable cuff pressure and the overpressure cutoff are controlled by
software. A software failure (a defective PROM) would generate a checksum
error, disabling monitor operation and setting it in a safe mode.
•the monitor cannot exceed the maximum allowable cuff pressure for adults (280
mmHg), pediatrics (200 mmHg), and neonates (141 mmHg)
•the overpressure cutoff feature shuts down the pump and dumps pressure before
the pressure reaches 330 mmHg (44 kPa)
To verify maximum allowable cuff pressure for NIBP version 3.20.00
1. Assemble one of the test setups illustrated on page 16, using an adult cuff or cuff
simulator.
Service manualFunctional verification19
2. Turn on the monitor.
3. Set the monitor to ADULT mode.
a. Pressonce.
b. Pressrepeatedly until ADULT appears.
4. Set the inflation target pressure to 270 mmHg (36 kPa).
a. Pressrepeatedly until TARGET PRESSURE appears in the display window.
b. Pressrepeatedly as needed to set the target pressure (displayed in the SYS
window) to 270 mmHg (36 kPa).
5. Pressto start the pump.
The pressure reaches approximately 270 mmHg (36 kPa), the pump shuts off, and the
pressure begins stepping down.
6. Carefully squeeze the pump bulb to raise the pressure to 280 mmHg (37.3 kPa).
•The dump valve opens and quickly drops the pressure to approximately 0.
•The monitor displays C10 to indicate that an overpressure event has occurred.
7. Set the monitor to PEDIATRIC mode.
8. Set the inflation target pressure to 170 mmHg (22.7 kPa).
9. Raise the pressure to 200 mmHg (26.7 kPa).
a. Pressand wait for the pressure to reach 170 mmHg (22.7 kPa). The pump
shuts off and the pressure reading starts stepping down.
b. Carefully squeeze the pump bulb to raise the pressure to 200 mmHg (26.7 kPa).
•The dump valve opens and quickly drops the pressure to approximately 0.
•The monitor displays C10 to indicate that an overpressure event has occurred.
10. Replace the adult cuff with a #1 neonate cuff, and wrap the neonate cuff securely
around a solid cylindrical object with circumference between 1.6 and 1.9 inches (4.1
and 4.8 cm).
11. Set the monitor to NEONATE mode.
12. Set the inflation target pressure to 132 mmHg (17.6 kPa).
13. Raise the pressure to 141 mmHg.
a. Pressand wait for the pressure to reach 132 mmHg (17.6 kPa). The pump
shuts off and the pressure reading starts stepping down.
b. Carefully squeeze the pump bulb to raise the pressure to 141 mmHg (18.8 kPa).
•The dump valve opens and quickly drops the pressure to approximately 0.
•The monitor displays C10 to indicate that an overpressure event has occurred.
To verify overpressure cutoff for NIBP versions 4.20.00 and later
This test demonstrates that the overpressure cutoff feature prevents pressure from rising
fast enough to exceed the cutoff pressure (295 to 330 mmHg, or 39.3 to 44.0 kPa) before
the pump stops and the dump valve and the bleed valve open.
20Functional verificationWelch Allyn VSM 300 Series
When the overpressure cutoff triggers, the pump stops, the dump and bleed valves open,
and the monitor displays E40. Any subsequent button press shuts off the monitor.
Note
Due to NIBP implementation differences, this test does not verify overpressure
cutoff in earlier NIBP versions.
In those versions, due to a longer software delay in activating the cutoff function,
the normal overpressure cuff pressure is exceeded before the overpressure
cutoff safety triggers. If the normal overpressure-detection algorithm were to fail,
the overpressure cutoff safety function would trigger immediately.
1. Restart the monitor.
2. Set the monitor to ADULT mode.
3. Set the inflation target pressure to 270 mmHg (36.0 kPa).
4. Increase the inflation rate, causing the overpressure cutoff to stop the pump and
open the bleed and dump valves.
a. Start the pump.
b. When the pressure (SYS window) reaches approximately 230 mmHg (32.4 kPa),
very rapidly squeeze the pump bulb while observing the digital pressure meter.
Assuming that you squeeze the bulb rapidly enough to reach the trigger threshold
and activate the overpressure cutoff:
•The pressure does not exceed 280 mmHg.
•The dump valve and bleed valve open, dropping the pressure to
approximately 0.
•E40 appears in the Sys window.
•Pressing any button causes the monitor to shut down.
5. Disconnect the test setup from the monitor.
If you executed this test as part of the functional verification procedure, proceed to...
•“Printer” on page 21.
•If the monitor has no printer, “SpO2” on page 23.
•If the monitor has no SpO2 option, “Temperature” on page 25.
•If the monitor has no temperature option, “Nurse call” on page 27.
Service manualFunctional verification21
Printer
Note
1. Put the monitor into Service Mode.
2. Verify that the printer has paper.
3. Press. Verify that a settings report prints, and that it contains no printed anomalies
Note
Note
This test is to be performed only as part of a complete functional verification
procedure.
and no missing or faded sections.
With a new roll of paper, the first line might be faded. This does not indicate a
problem.
The settings report (as shown in the example below) contains a calibration record,
user record, hardware status record, and software versions record.
•The calibration record includes manufacturing configuration data: monitor
serial number, set parameters, and language.
•The user record includes user-configurable settings: alarm limits, patient type,
parameter modes/units, auto interval, tone volume, and others.
•The hardware status record shows the hardware revision number, the battery
voltage level, the total number of NIBP monitoring cycles completed on the
monitor, and the total number of hours of operation of the monitor.
•The software versions record indicates the software version numbers of the
main board, SpO2and Temperature options (if present), and NIBP.
4. After 2 seconds of printing, verify that the current draw is no more than 2.5A above
the baseline current recorded in step 2 of the verification test “Baseline current
draw” on page 13.
5. Exit the Service Mode.
22Functional verificationWelch Allyn VSM 300 Series
Service manualFunctional verification23
SpO
2
Note
Perform one of the following three SpO2tests.
The tests described in this section are to be performed only as part of a complete
functional verification procedure.
Testing the monitor only (Nellcor SpO2)
Use this procedure to test only the monitor SpO2function.
1. Power the monitor off.
2. Connect the Nellcor SRC-MAX SpO2functional tester to the SpO2input connector
through a Nellcor DEC-8 extension cable.
3. Power the monitor on.
Note
4. Verify the following on the SRC-MAX:
In the following tests, if the SRC-MAX defaults are outside the monitor alarm
limits, readjust the limits or silence the alarms.
•All of the monitor LEDs flash: left panel, center panel, right panel.
•SRC-MAX initializes to a default condition where the four test parameter LEDs are
lit closest to their selector buttons
•The default pulse rate is 60 bpm and the default SpO2is 75%
5. Give the monitor up to 30 seconds to stabilize, and verify a displayed pulse rate of
60 ± 3 bpm and a displayed SpO2of 75 ± 3%.
6. Set the SRC-MAX pulse rate to 200 bpm.
7. Give the monitor up to 30 seconds to stabilize, and verify a displayed pulse rate of
200 ± 3 bpm.
8. Switch the SRC-MAX SpO2saturation percentage to 90.
9. Give the monitor up to 30 seconds to stabilize, and verify a displayed SpO2saturation
level of 90 ± 3%.
10. Disconnect the SRC-MAX.
Testing the monitor only (Masimo SpO2)
Use this procedure to test only the monitor SpO2function.
1. Power the monitor off.
2. Connect the Masimo Tester REF 1593.
3. Power the monitor on.
4. Give the monitor up to 30 seconds to stabilize, and then verify a displayed pulse rate
(from SpO2) of 61 bpm ± 1 bpm and a displayed saturation of 81% ± 3%.
24Functional verificationWelch Allyn VSM 300 Series
Testing the sensor with the monitor
Use this procedure to test the monitor SpO2function while verifying a sensor.
1. Connect an SpO
2. Set the BioTek Index2 SpO
sensor to the simulator optical finger and to the monitor.
2
simulator as follows:
2
SpO2manufacturer typeNellcor (MP-506A or Nell-3A)
Masimo (MS-11)
% SpO
2
94
Rate60 (Motion Artifact not selected)
3. Wait for the monitor display to stabilize.
4. Verify an SpO2saturation level of 94 ± 4.
5. Verify that the monitor displays a pulse rate of 60 ± 4.
Service manualFunctional verification25
Temperature
Note
Perform one of the following temperature tests.
The tests described in this section are to be performed only as part of a complete
functional verification procedure.
Testing the monitor only
Use this procedure to test only the temperature function.
1. With monitor power on and with the temperature probe in the well, disconnect the
probe cable from the temperature input connector on the rear of the unit.
2. Connect the Welch Allyn Sure Temp Plus temperature test key (06138-000) to the
temperature input connector.
3. Remove the probe from the well.
4. Verify that the displayed temperature is 97.3 ± 0.2 F (36.3 ± 0.1 C).
Testing the probe with the monitor
Use this procedure to test the temperature function while verifying the
temperature probe.
1. If the Welch Allyn Calibration Tester is not already warmed up, warm it up as follows:
a. Set the tester temperature to the following temperature:
ModelTemperature
960096.4 °F(35.8 °C)
9600 Plus96.8 °F(36.0 °C)
b. Plug the power adapter into the tester; wait (up to 15 minutes) until the tester
‘ready’ indicator light is on continuously.
2. Set the monitor temperature mode to Monitored.
3. Insert the temperature probe (without probe cover) into the small thermometer hole
on the dry-heat well of the tester.
4. Wait two minutes and verify that the reading displayed on the monitor is as follows:
ModelTemperature
960096.4 °F ±0.4 °F(35.8 °C ±0.2 °C)
9600 Plus96.8 °F ±0.4 °F(36.0 °C ±0.2 °C)
5. Switch the tester to the following temperature and wait about five minutes for the
‘ready’ indicator to light.
ModelTemperature
9600106.0 °F(41.1 °C)
9600 Plus105.8 °F(41.0 °C)
26Functional verificationWelch Allyn VSM 300 Series
6. Verify that the reading displayed on the monitor is as follows:
ModelTemperature
9600106.0 °F ±0.4 °F(41.1 °C ±0.2 °C)
9600 Plus105.8 °F ±0.4 °F(41.0 °C ±0.2 °C)
7. Remove the probe from the tester and replace it in the probe well.
8. Disconnect the tester.
Loading...
+ 76 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.