Watlow Electric 1500 User Manual

Series 1500
ACTUAL
ADJUST
Ramping and Profiling
Microprocessor-Based
Control
User's Manual
Watlow Controls, Inc: 1241 Bundy Blvd., Winona, MN 55987 507/454-5300, Fax: 507/452-4507
W150-MA30-9043 October, 1990
Supersedes: W150-MA20-8843
$10.00
Made in the U.S.A.
Use The Manual

How to Use the Manual

First...
Starting Out Sample Program
Install/Wire Programming
Appendix
Quick Reference
NOTE: Details of a "Note" appear here, in the narrow box on the outside of each page.
This manual will make your job easier.
way to become familiar with the Series 1500. Here's an overview. Introduction, Chapter 1, Page 4. A Sample Program in Action, Chapter 2, Page 16. Installation and Wiring, Chapter 3, page 26. Technical Reference, Chapter 4, Page 40. Calibration
Specification Ordering Information Technical Service Glossary Index Quick Reference on Back Cover
Reading it and applying the information is a good
Notes
This user's manual contains informational notes to alert you to important details. When you see a note or note icon, look for an explanation in the margin.
Safety Information
NOTE: Details of a "Warning" appear here, in the narrow box on the outside of each page.
!
NOTE: Details of a "Caution" appear here, in the narrow box on the outside of each page.
This user's manual also has boldface safety information notes to protect both you and your equipment. Please be attentive to them. Here are explanations.
(The Stop Sign or Lightning Bolt Sign in the wide text column alerts you to a "WARNING," a safety hazard which could affect you and the equipment. A full explanation is in the narrow column on the outside of the page).
!
(The Deer Crossing Sign or Exclamation Point Sign in the wide text column alerts you to a "CAUTION," a safety or functional hazard which could affect your equipment or its performance. A full explanation is in the narrow column on the outside of the page).
Technical Assistance
If you encounter a problem with your Watlow control, review all of your configuration informa­tion to verify that your selections are consistent with your application...Inputs, Outputs, Alarms, Limits, etc. If the problem persists after checking the above, you can get technical assistance by dialing: 1-507-454-5300
An Application Engineer will discuss your problem with you. Please have the following information available when calling:
Complete model number • Serial Number
All configuration information • User's Manual The model and serial numbers can be found on the outside of the case.
Your Feedback
Your comments or suggestions on this manual are welcome, please send them to: Technical Writer, Watlow Controls, 1241 Bundy Blvd., Winona, MN 55987, phone 507/454-5300, fax 507/ 452-4507. The Watlow Series 1500 User's Manual and integral software are copyrighted by Watlow Winona, Inc., © 1985, 1986, 1987, 1988 with all rights reserved.
2 blf1090
WATLOW Series 1500 User's Manual
Page Item 4 Starting Out with the Series 1500 -Chapter 1
4 General Description 5 Opening the 1500 6 Setting the DIP Switches 7 Quick Wire 9 Overview of the 1500 Modes 10 Reading the Displays and Keys 12 Overview of the Software
16 Learning the Series 1500: A Sample Program
in Action - Chapter 2
16 ApplePie Sample Program 16 Sample Program Chart 17 Setting Event Outputs 17 Modes of Operation 18 Clear Memory, Set DIP Switches 19 Programming 22 Start Your Program 22 Halt Conditions 23 Ramping Conditions 23 Wait Step 24 Jump Loop Step 24 Recycle 25 Interactive Changing Set Points & Events
26 Install and Wire the Series 1500 - Chapter 3
26 Sensor Installation Guidelines 27 Event Input and Remote Hold Input 27 Noise 29 Input Power Wiring 32 Eliminating Noise 32 Opening the 1500 34 Installation Procedure 34 Electrical Connections and Wiring
40 Technical Reference - Chapter 4 40 How to Program
40 Select the Proper DIP Switch Settings
41 Programming CHG DATA
42 Halt Conditions 43 Set Point Step Programming 44 Jump Loop Step Programming 45 Wait Step Programming 45 Next Step Programming 46 Altering a Program
46 Programming CHG PARA
47 Guarded Access Programming
49 Alarms
49 Alarm Options 49 Event Alarm Output Option 49 Latching Option 49 Alarms 50 Upper/Lower Set Point Limits
51 Event Outputs
51 Event Input 51 Remote Hold Input 52 Analog Retransmit Outputs 53-56 RTD and T/C Guarded Access/Parameter Charts 57 Programming Chart
58 Tuning
58 Initial Settings 58 Simple ON/OFF Control
59 Run and Halt a Program
59 How to Run a Program 59 How to Halt a Program
Contents
Contents
59 Restarting a Program 59 Remote Hold Input
60 Data Communications
60 Communication Parameters 60 Protocol 60 Command Rules & Words 61 Syntax 73 Data Communication Error Codes 74 Troubleshooting
75 Field Calibration Procedure 80 Specifications 82 Model # Information 83 Technical Service
83 Warranty Information 83 Returning Merchandise 84 Shipping Claims 85 Glossary 89 Index 90 Quick Reference
Figures
4 Series 1500 Input and Output Overview 1 6 DIP Switch Location and Setting 2 7 Quick Wire AC Power Connection 3 8 Quick Wire Simulated Sensor Wiring 4 9 Overview of the Series 1500 Operating Modes 5 10 Series 1500 Front Panel Components 6 13 Monitor Data (MNTR DATA) Mode Overview 7 14 Change Data (CHG DATA) Mode Overview 8 15 Change Parameter (CHG PARA) Mode 9 31 Differential Mode Filter Diagram 10 31 Common Mode Filter Diagram 11 31 Differential Common Mode Filter Diag. 12 33 Unit and Panel Cutout Dimensions 13 35 Signal Conditioner Connections and Jumpers 14 36 Event Output, Remote Hold Input, Event Input
Conn. for Dual Solid State Relay Version 15
37 Event Output, Remote Hold Input, Event Input
Conn. for Dual Solid State Switch Version A 15
36 Power Supply Input, Temp & RH Output Conn.
for Dual Solid State Relay Version 16
37 Power Supply Input, temp & RH Output Conn.
for Dual Solid State Switch Version A 16 38 Power Supply Input, CH-1 & CH-2 Output Conn. 17 38 Wiring to Disable Channel 2 18 39 RS-422/423 Pin Designations and Jumpers 19 42 Change Data Mode Flow Diagram 20 46 Change Parameter Mode Flow Diagram 21
Tables
6 DIP Switch Position and Function Table 1 17 Sample Program Event Outputs Table 2 21 Sample Program Steps Table 3 31 Noise Suppression Device Ratings Table 4
Charts
17 Series 1500 Sample Program Chart Chart 1 53 RTD Guarded Access Chart Chart 2 54 T/C Guarded Access Chart Chart 3 55 RTD General Parameter Chart Chart 4 56 T/C General Parameter Chart Chart 5 57 Programming Chart Chart 6
WATLOW Series 1500 User's Manual 3
Starting Out
Chapter 1 Starting Out with the Watlow Series 1500
"Read Me First."
Congratulations, you're about to become a fully-qualified user of the Watlow Series 1500! It is a versatile microprocessor-based control; powerful, yet simple to learn. In this chapter of the user's manual, you'll get an overview of the 1500 and its software. You'll "light" the displays and get a feel for moving through the control functions.
Figure 1 is a simplified view of the 1500's capabilities. The control has dual inputs, one per channel, and dual control outputs for each channel. There is a choice of auxiliary (event) or alarm outputs, an event input and remote hold input. An "Event Output" in the Series 1500 is an ON/OFF solid state relay or solid state switch output for activating periphered equipment or processes. Models with optional communica­tions may be connected to a computer via a serial communications interface.
Figure 1 ­Series 1500 Input and Output Overview
TEMP and %RH or TEMP See Pages 6, 8 & 35
Event Input See Pages 36 & 51
The Series 1500 will handle 51 separate control steps. Each step duration is variable from one second to 24 hours, 59 minutes, 59 seconds. The 1500 is a "profiling" control because it will control up to 51 consecutive steps with a single program; the resulting graph of all the steps makes a program "profile." The 1500 is a "ramping"
Dual Inputs See Pages 6, 35 & 82
Remote Hold Input See Pages 36 & 51
Dual Control Outputs/Channel See Pages 6 & 82
Heat/Cool See Page 36
Humidify/ De-Humidify or Heat/Cool See Page 36
8 Auxiliary (Event) Outputs, 4/Channel See Pages 36 & 51
or
RS-422A/RS-423A (RS-232C) Serial Communications Interface See Page 62
4 WATLOW Series 1500 User's Manual
3 Auxiliary (Event) Outputs/Channel and 1 Alarm Output/Channel See Pages 36, 49 & 51
Starting Out, Chapter 1
control because it will move from one process variable to another in a uniform manner.
Operator-friendly features include automatic "prompts," or entry codes, to assist you in programming and monitoring. The Series 1500 has other display codes to indicate a variety of alarm conditions about your process. The 1500 also has a "Jump Loop" option for repeating program steps or segments. A "Wait For" option makes this control responsive to particular process conditions, an external switch conditions or real times of day.
In addition, you can change the three-mode, PID control to a simple ON/OFF control by setting the proportional band parameter in each channel to zero.
When there's a power outage, the 1500 has memory protection provided by an internal Lithium battery and optional battery-backed clock.
The standard RTD version Series 1500 is factory-configured to operate in degrees Centigrade (°C) on Channel One and per cent relative humidity (%RH) on Channel Two. A software selection, the "Temp-Temp" option, is also available for the RTD unit. The Series 1500 Thermocouple version is strictly a "Temp-Temp" control. The 1500 has at least two switches you must set prior to entering your program. Informa­tion on how to set these switches follows in this chapter.
When you first apply power, the unit will "come up" with default values for set points, alarm points, control parameters, recycle, events and time. These default values provide minimum operating instructions until you enter your data. The default values are listed in Chapter 4, "Technical Reference."
Starting Out
Remove the Series 1500 carefully from its shipping container. Be sure to set this literature aside where it will not be discarded.
Packing List
Included with your Watlow Series 1500 RTD version are two 107 resistors, four jumper wires, two mounting brackets with integral screws, and this manual. The resistors and two jumpers will serve as pseudo-sensors at approximate room condi­tions while you learn the control. The other jumpers are for power connections.
How to Open the 1500
Before going further, open the Series 1500 and pull the control chassis from its case. Here's how:
The control chassis fastens to the case with a single screw located at the lower front panel. Turn the screw counterclockwise to loosen it. Three strip connector plugs, in the rear of the control chassis, feed power and signals through the back of the case to the triple terminal strip. These plugs will let go as you pull.
!
Pull firmly but gently to remove the Series 1500 from its case. The 1500 chassis will not fit into the case upside down. However, always check to see that it is oriented correctly before returning the chassis to the case. Press the unit in firmly, then turn the front panel screw clockwise to secure it. Do not overtighten the screw.
!
CAUTION: Before attempting to open a Series 1500 with data communi­cations (RS-422/423 interface), remove the two screws and DB-15 male connec­tor from the rear of the case.
WATLOW Series 1500 User's Manual 5
Starting Out
!
CAUTION: Power must be interrupted before a change in DIP Switch will take effect.
Figure 2 ­DIP Switch Loca­tion and Setting
How to Set the DIP Switches
The Watlow Series 1500 has a set of Dual In-Line Package (DIP) switches on circuit board, A007-1318. The location of the board and switches appear in Figure 2. The switches are clearly numbered; the "ON" direction is indicated by an arrow. Look at the DIP switches from the bottom of the control. You'll see them as they appear
below.
!
A007-1318 Board

DIP Switches

Table 1­DIP Switch Posi­tion and Function
Control Chassis - Bottom View
Set the DIP switches according to Table 1 below. Make your initial choices now; you may always return to change them later. Set DIP Switch #6 ON for a "Cold Start."
Sw Position Function #1 ON Auxiliary (Event) Output #4 for both Ch-1 and Ch-2 are alarm outputs.
OFF Auxiliary (Event) Output #4 for both Ch-1 and Ch-2 are event outputs.
#2 ON Alarms are latching (displayed until cleared manually).
OFF Alarms are non-latching (displayed only as long as alarm condition exists).
RTD Unit Only #3 ON The Ch-2 sensor is an RTD sensor. Jumpers W151 and W153 on the Signal
Conditioner board (A007-1316) must be installed.
OFF The Ch-2 inputs is a 0-5V signal representing 0-100% RH. Jumpers W150
and W154 on the Signal Conditioner board (A007-1316) must be installed.
T/C Unit Only #3 ON Not Used. Set in the ON position.
#4 ON Dual PID per channel - #4 OFF Single PID per channel.
#5 ON Factory use only. Must be in the ON position.
#6 ON Cold Start on power-up. (Memory cleared, parameters set to default values.)* OFF Warm Start on power up. (Programmed values are retained for all parameters.)*
#7 ON °C function after a Cold Start.
OFF °F function after a Cold Start.
#8 OFF Not Used. Set in the Off position.
*
For further Warm and Cold Start information, see the Technical Reference, Chapter
4.
6 WATLOW Series 1500 User's Manual
Starting Out, Chapter 1

Quick Wire

Starting Out
You may "power up " the Series 1500 with the following operational check-out. This procedure also prepares you for the Sample Program in Chapter 2. Or, you may skip this section and the sample program, and go directly to Chapter 3, Installation and Wiring.
1. Put the 1500 chassis back in its case. Then place the unit on a bench and connect AC power lines and the jumper wires to the Power Supply terminal strip. Use Figure 3 below.
115VAC
L2
CHASSIS GROUND
L1
For 115VAC control operation, install a jumper wire from Terminal 32 to Terminal 35, and a separate jumper wire from Terminal 33 to Terminal 36. Bring line power in to Terminals 32 and 36. Connect the line ground wire to Terminal 34.
WARNING: To avoid potential electric shock use National Electric Code safety prac­tices when wiring and when connect­ing this unit to a power source and to electrical sensors or peripheral devices.
Figure 3 ­Quick Wire AC Power Connec­tions
230VAC
L2
CHASSIS GROUND
L1
2. As soon as you apply power, the FUNCTION display will alternately flash "A1­C1, A1-C1." Press the CLEAR key to clear the flashing A1. The display will then flash another A-code.
When you apply power without sensor inputs on the Signal Conditioner terminal strip, the Series 1500 will show a flashing "A-digit" in the FUNCTION display. The digit is an alarm code The ACTUAL display will show four steady dashes, "----," meaning "open sensor." These are normal displays for a unit without sensors.
Starting Out, Chapter 1
For 230VAC control operation, install a jumper wire from Terminal 33 to Terminal 35. Bring line power in to Terminals 32 and 36. Connect the ground wire to Terminal 34.
WATLOW Series 1500 User's Manual 7
ACTUAL Display
CLEAR Key
FUNCTION Display
Starting Out
3. If you did not set DIP Switch #6 in the "ON" position before, do that now: Disconnect power from the unit. Open the 1500, set DIP Switch #6 to ON for a Cold Start. Replace the control chassis in the enclosure. Reapply power to the unit. (This clears all previously entered information from the 1500; it is a "clean" or "cold" start).
4. Now remove power again, open the unit, and set Switch #6 to OFF. (This will make the next start a "warm" one, retaining all subsequently entered informa­tion in the processor's memory). Close the 1500.
5. Locate the Signal Conditioner terminal strip on the back of the Series 1500, it's on the far left as you look at the back of the unit. The resistors, or jumper wires for the T/C version, will simulate input sensors at room temperature for this procedure. Make the correct connections using Figure 4.
For an RTD unit, connect a 107 resistor (provided) across Terminals 5 and 6, and another 107 resistor across Terminals 8 and 9. Also connect a jumper wire (provided) across Terminals 6 and 7, and another jumper wire across Terminals 9 and 10. See Figure 4.
For a T/C unit, connect a jumper wire (provided) across Terminals 5 and 7, and another jumper wire across Terminals 8 and 10.
Figure 4 ­Quick Wire Simu­lated Sensor Wiring
RTD T/C
6. Apply power to the unit. With the CLEAR key, clear the "A1" alarm code and any other "A" code that appears after it. The ACTUAL display will be reading Channel 1 at simulated room temperature. Or the ACTUAL display will show Channel 2 at a simulated %RH or temperature reading. Use the CH.SELECT key to switch from one channel to the other.
If your Series 1500 responds this way, continue learning about the control. If your unit does not respond like this, call Watlow at 507/454-5300 and ask for an applica­tions engineer.
8 WATLOW Series 1500 User's Manual
Starting Out, Chapter 1
Overview of the Three Operating Modes
Before getting into the details of the Series 1500's keys and displays, take a look at Figure 5, showing the three different modes. After you feel comfortable with the names of the modes and their functions, go ahead to learn the keys and displays.
Series 1500: Four Mode Types
Series 1500: Three Mode Types
Starting Out
See Your Program
Monitor Data Mode (MNTR DATA)
Examine current step information in a running program.
Set Up Program Steps
OR
Change Data Mode (CHG DATA)
Set up or change step type, program loops, wait for conditions, set points, auxiliary (event) outputs ON/OFF, and step duration.
Set a Non­Ramping Set Point
Monitor Data Mode (MNTR DATA)
Generate a
Set a non-ramping
non-ramping (fixed)
(fixed) set point and
set point and
manipulate event
manipulate event
outputs. At MNTR
outputs. At MNTR
DATA, halt the pro-
DATA, halt the pro-
gram and press
gram and press
RESTART. Select C1,
RESTART. Select C1,
C2, E1 or E2.
C2, E1 or E2.
Figure 5 ­Overview of the Series 1500 Operating Modes
Starting Out, Chapter 1
Set Up Your System
Change Parameter Mode (CHG PARA)
Set up or change system-wide para­meters such as real time, high/low range, alarm types, alarm points, PID settings, RS-422 address.
WATLOW Series 1500 User's Manual 9
Starting Out
Figure 6 ­Series 1500 Front Panel Components
CH SEL key: CHANNEL
SELECT--Selects either Channel 1 or Channel 2 for the ACTUAL display.
How to Read the Displays and Use the Keys
Here's an understanding of the Series 1500 displays, keys and LEDs. Use Figure 6 to locate the keys, LEDs and displays described here. Try everything! You'll learn quickly how the 1500 works. The detailed overviews in the next section will further your knowledge.
CH-2 EVENT LEDs: Indicates the status of each Channel 2 Event
ACTUAL display: Shows the actual value of the variable in Channel 1 or Channel 2, in four digits.
Output with one of four LEDs visible only when the output is energized (ON). An Event Output in the Series 1500 is an ON/OFF solid state relay or switched DC output for activating periphered
equipment or processes.
CH-1, CH-2 LEDs: Indicates the channel in the ACTUAL display.
°F LED: When lit indicates the value
in the ACTUAL display is tempera­ture in degrees Fahrenheit.
10 WATLOW Series 1500 User's Manual
% RH LED: When lit
indicates per cent relative humidity in the ACTUAL display.
°C LED: When lit, indicates tempera-
ture in degrees Celsius in the ACTUAL display.
CH-1 EVENT LEDs: Indicates the status of each Channel 1 Event Output with one of four LEDs visible only when the
output is energized (ON).
Starting Out, Chapter 1
Figure 6 (Con't) ­Series 1500 Front Panel Components
PROG END LED: PROGRAM END--
Indicates that the processor has reached the end of the program.
Starting Out
PROG HALT LED: PROGRAM HALT--When lit, indicates
that the processor is in the HALT condition. A RUN/HALT key press, or a "Blank Step" in the program causes a HALT. When flashing, indicates that the processor is in "Remote-Hold" condition. See RUN/HALT key.
CLEAR (erase) key: a. Clears alarm codes
from FUNCTION display in MONITOR DATA mode.
b. Clears the step in the STEP
display in CHANGE DATA mode when the unit is in the HALT condition.
c. Clears all 51 programmed steps
when the unit in the HALT condition, and in GUARDED ACCESS area of the CHANGE PARAMETER mode. (This prevents an operator from accidentally clearing an entire program. The GUARDED ACCESS data is not cleared).
STEP display: Indicates the current step number being programmed in the CHANGE DATA mode, or the current step being monitored in the MONITOR DATA mode.
RE-START key: Returns the controller to the initial program step only when the 1500 is in the PRO­GRAM HALT condition. See RUN/HALT key.
FUNCTION display: Shows the prompts used in the three modes. It also displays hours when TIME is selected in MONITOR DATA, CHANGE PARAMETER or CHANGE DATA modes. The FUNCTION display shows a flashing alarm code if one is activated.
RUN/HALT key: Starts or stops the program. The processor will resume a program where it was interrupted, unless the RE­START key was pressed, or the step number or set point data at the current step was changed during
the HALT condition.
DATA display: Indicates the entered data or monitored value in four digits.
CHG DATA LED: When lit indicates the processor is in the CHANGE DATA mode.
MNTR DATA LED: When lit, indicates the processor is in the MONITOR DATA mode.
MODE key: Selects either the MONITOR DATA, CHANGE DATA or CHANGE PARAMETER mode.
Starting Out, Chapter 1
FCTN key: The FUNCTION key steps through the prompts in each of the three modes.
ENTER key: Enters selected data or NEXT STEP operations in the CHANGE DATA and CHANGE PARAME­TER modes.
CHG PARA LED: When lit indicates the processor is in the CHANGE PARAMETER mode.
TIME LED: When lit, indicates that the FUNC­TION and DATA displays are showing time.
UP key: Increases the value in the DATA display. A light touch increases the value by one. Holding the key down causes the DATA display to increase rapidly.
WATLOW Series 1500 User's Manual 11
DOWN key: Acts opposite the UP key. Ramps the value in the DATA display downward. A light touch decreases the value by one. Holding the key down causes the DATA display to decrease rapidly.
Starting Out
Now that you have a good idea how the 1500's front panel works, look again at the operating modes. This time take a more detailed look.
Overview of the Mode Software
Again, the Series 1500 has three Operating Modes: Monitor Data, Change Data and Change Parameter. These modes change with the MODE key. They make up the basic software routines that run the Series 1500. Each mode has sub-routines and different prompts that appear in the FUNCTION display whenever you press the FCTN key.
Examine the three figures on the next pages, one for each of the modes. Under­standing how these modes control the Series 1500 is the key to learning the control.
Figure 7 - Monitor Data (MNTR DATA) Mode Overview
Figure 8 - Change Data (CHG DATA) Mode Overview
Figure 9 - Change Parameter (CHG PARA) Mode Overview
12 WATLOW Series 1500 User's Manual
Starting Out, Chapter 1
Monitor Data Mode (MNTR DATA):
Starting Out
Monitoring a Halted Program
You may look at any step, but must be in the CHG DATA mode to change the step number.
You'll see:
• Step type.
• Ch-1 and Ch-2 set points.
• Auxiliary outputs programming.
• Step duration.
Monitoring a Running Program
You may look at the current step. You'll see:
Figure 7 ­Monitor Data (MNTR DATA) Mode Overview
• Ch-1 and Ch-2 current set points.
• Auxiliaries programmed ON or OFF.
• Time remaining.
Set a Non-Ramping Set Point
Set a non-rampong (fixed) set point and manipulate event outputs.
Starting Out, Chapter 1
WATLOW Series 1500 User's Manual 13
Starting Out
Figure 8 ­Change Data (CHG DATA) Mode Overview
Change Data Mode (CHG DATA):
Three Step Types for Entering Program Data
Set Point (SP):
a simple, regular profile step.
Set Point Step,
Program:
This Step# is an SP step.
Set Point for Ch-1.
Set Point for Ch-2.
Auxiliary (Event) Outputs, ON/OFF.
Step Duration.
Go to any next step.
Jump Loop (JL): Wait for…(WT):
a means for jumping to a step out of numerical sequence.
Jump Loop Step,
Program:
This Step# is a JL step.
Jump to Step# __.
a means to wait for a particular condition or time.
Wait for Step,
Program:
This Step# is a WT step.
Wait for this Ch-1 value to
Repeat that
occur, __. step, __ times (Jump Counts).
Wait for
this Ch-2
value to Go to any
occur, __. next step.
Wait for
Input Event
to achieve a
specified
conditon.
Wait for
this real
time, ___.
14 WATLOW Series 1500 User's Manual
Go to any
next step.
Starting Out, Chapter 1
Change Parameter Mode (CHG PARA)
Starting Out
For Entering System-Wide Parameters
Program Real Time
Then proceed to one of three "Guarded Access" areas – code-protected system setup parameters.
Program These Parameters:
Set Ch-1 & Ch-2 alarm type and high/low points.
Leave "Guarded Access."
Program These Parameters:
Set Recycle Option (program repeat).
Set up Ch-1 PID parameters and calibration offset.
Set up Ch-2 PID parameters and calibration offset.
Leave "Guarded Access."
Figure 9 ­Change Parameter (CHG PARA) Mode Overview
Program These Parameters:
Set up Ch-1 and Ch-2 high and low ranges.
Select Series 1500 Ch-1/Ch-2 Operation, TEMP-TEMP or TEMP-RH.
Set Recycle Option (program repeat).
Select RS-422 Address #.
Select keyboard lock status.
Leave "Guarded Access."
Where To Go From Here
You are now ready to go to the Sample Program, Chapter 2, or to Installation and Wiring, Chapter 3.
If you skip the sample program, do not forget to check the position of DIP Switch 6 before you begin programming your control after installation. With DIP Switch 6 OFF, the 1500 saves your program whenever power is removed (Warm Start). With DIP Switch 6 ON, the 1500 will clear its memory of all programmed information whenever power is removed,substituting default values (Cold Start).
Starting Out, Chapter 1
WATLOW Series 1500 User's Manual 15
Sample Program
Chapter 2 Learning the Series 1500 — A Sample Program
in Action
This chapter will guide you through an easy sample program for the Series 1500. You can quickly grasp the necessary terms and concepts by entering and observing this exercise.
If you feel that your knowledge of programmable controllers does not require a sample program to learn the Series 1500, feel free to skip this chapter. In the Technical Reference section (Chapter 4), you will find details on all material here.
IMPORTANT NOTE:
If you have a Series 1500 T/C version, or are using the RTD version for 2 channels of temp­erature control (TEMP-TEMP), then simply consider all sample program references to relative humidity as a second channel of temp­erature. Use the sample values as temperatures rather than relative humidities.
The Series 1500 Wonderful-Apple-Pie Sample Program
Let's assume that Great-Grandma handed down to you her secret recipe for wonder­ful apple pie. It's not that her ingredients are so different; but she took great care to control baking temperature and humidity, and to add topping ingredients at just the right times in the baking cycle. These really are fantastic apple pies, so you've bought an oven with a Series 1500 control, and you're starting an apple pie business.
Sample Program Chart
The chart here shows the baking process. Grandma's original recipe had Grandpa turn the oven on early in the morning. Grandpa would raise and lower the oven temperature at certain times, too. That is one of Grandma's secrets. With the Series 1500 you'll be able to make your oven "loop" between 275° and 110° three times before a pie is done.
If Great-Grandma was right, this process should turn out a really fine apple pie. Since your bakery will begin baking pies at the same time every morning, you'll want your Series 1500 to automatically start the process each day.
The chart below details the "wonderful-apple-pie" sample program for the Series
1500.
16 WATLOW Series 1500 User's Manual Sample Program, Chapter 2
Temp/
h
%RH
Step 1 Step 2
Step 3
Perform Steps 4-6 T
Step 4 Step 5
Sample Program
Chart 1 ­Series 1500 Sample Program Chart
(Wait)
Temp/
Step 1 Step 2
%RH
(Wait)
*
50%RH
Setting Event Outputs
90% RH
Step 3
90% RH
400°F
74% RH
Step 4 Step 5
74% RH
275°F
To make the pie process even more automated, you have your toppings — sugar, cinnamon and nutmeg—connected to solenoid dispensers controlled by the 1500's "Event Outputs." As you enter the sample program, you'll set the Event Outputs to trigger just when you want them to. You even have an Event Output to put your pies in and take them out of the oven at just the right times. You'll need to turn the Events ON as follows:
STEP CH #1 EVENTS CH #2 EVENTS
1 Wait 0 0 2 1, 2, 4 1, 2 32 3 4 3, 4 1, 4 51 2 6 1, 2, 4 2, 3, 4 8 3, 4 1, 3
Understanding the Modes of Operation
Sample Program, Chapter 2
WATLOW Series 1500 User's Manual 17
Table 2 ­Sample Program Event Outputs
Sample Program
Before you begin entering the program, notice three small lights (LEDs) on the bottom half of the 1500's front panel. They are labeled "MNTR DATA" (Monitor Data), "CHG DATA" (Change Data), and "CHG PARA" (Change Parameters). Press the MODE key to move through these three modes as you read about each one. Another LED, labeled "TIME" indicates when time is being displayed. "MNTR DATA": While a program is running, you'll want to observe what's happen­ing. Some items can be observed in the "ACTUAL" display and in the nearby top half of the front panel: Actual Temperature or Relative Humidity, Events ON/OFF, etc. But you'll also want to know which step is being performed, and the amount of time remaining in the step.
You can monitor information in a running program by pressing the MODE key until the MNTR DATA LED is ON. Then press the FCTN key until you are observing the data you want. Right now you don't have a program entered, so no meaningful data is available. As soon as you enter the sample program, use this mode to watch things happen.
"CHG DATA": This is the mode where you'll enter program operating DATA (set points, time interval for each step and Event Outputs). In this mode, you can also change data or correct errors in the program. If you want to wait until a specific time before doing something or to set up program loops, you also enter that information in CHG DATA. With the CHG DATA LED on, pressing the FCTN key will move you through the functions you can program.
!
CAUTION: This sample pro­gram will activate some of the "Control Outputs" and "Event Outputs" on the Series 1500 terminal strips. To avoid any damage to your equipment while running the sample program, be sure all Event and Control Outputs are discon­nected from the Series 1500 termi­nals. Or remove power from any external loads attached to the Series 1500.
"CHG PARA": Some Series 1500 information is more applicable to the overall
system than to just one specific program. This includes the "real time of day," system maximum/minimum alarm limits and "PID Parameters" which tune the overall system. This system-wide information is entered or changed in this mode.
Again – pressing the FCTN key selects functions while the CHG PARA LED is on. Many of these functions are protected from accidental change by a "Guarded Ac­cess" (GA) code before they can be viewed or altered. You'll learn the GA codes soon.
Before Y ou Begin…
Before you begin to enter your Sample Program, take this precaution:
!
Clear Memory, Set DIP Switches
To clear your Series 1500's memory, do a "Cold Start." Turn the power OFF, open the 1500 and ensure that DIP Switches 2, 3, 5 and 6 are ON and that DIP Switches 1,4,7 and 8 are OFF. If you haven't done so already, attach the provided 107 resistor and a jumper (or jumpers for a T/C version) to each sensor input as de­scribed in Chapter 1 under "Quick Wire." If you want to use your sensors, refer to Chapter 3, Figure 11 for wiring. Then turn the power ON and press the "CLEAR" key to clear the flashing alarm indication(s).
18 WATLOW Series 1500 User's Manual Sample Program, Chapter 2
4.
Press the
ENTER key:
the display will flash, then it will change to “MN.”
Use the UP/DOWN keys
5.
to place the correct value for minutes into the Data
Display; then press ENTER.
The display will show “SC” (seconds): again use the UP/DOWN keys to place
6. seconds into the display and then press ENTER.
Press the FCTN key several times
7.
PARA
LED). You
should now see the correct real time on the display, with
until the TlME
LED is lit
(along with
the CHG
the seconds counting up.
Set the Recycle Option
Now, while still in the CHG that you want your program to automatically repeat each day until you decide to halt it. This “Recycle” condition is set within the “Guarded Access” (GA) section of the CHG PARA
mode. Do that now:
1.
With
the CHG
Function Display. If you
around again.
2.
Place
“0006,”
one of the Guarded Access codes, into the Data Display using
the UP/DOWN keys.
3.
Press ENTER Display.
PARA mode,
PARA
LED ON, press
pass
until
the "RC,"
“RC,”
there is one more thing to enter. Remember
the
FCTN key until “GA” appears in the
GA, keep pressing
FCTN
until GA comes
(Recycle) prompt appears in the Function
4.
Place a “0001” into the Data Display. Press ENTER.
Begin Entering Data
Now
that
you
have entered the real time and the Recycle option, go to the CHG
DATA mode to enter actual program data:
1.
Press the MODE key until the CHG DATA LED is lit. The display will read “01
SP.”
Programming Step
Refer back to the sample program chart at the beginning of this section. that while you’d like to load your program and press the start button now, the first program step must hypothetically wait until early morning before it starts running. To program Step #1 as a Wait Step,” do this:
Press the FCTN key until "WT" appears in the Function Display.
1.
Press the ENTER key,
2. Since you are waiting for a
3. FCTN
until
“WH”
1
W1”
(Wait for
specific
(Wait for Hours) appears.
Ch-1)
appears.
time, and not for input conditions, press
Remember,
Sample Program, Chapter 2
WATLOW Series
1500 Users
Manual
19
Sample Program
4. We suggest that you enter a time one hour from right now. Changing this time later is a simple matter. You go to the CHG DATA mode with "WH" displayed and then enter the new time. Select hours (0 to 23) and press
ENTER.
5. Also enter the "WM" (minutes) and "WS" (seconds). Select minutes, press ENTER. Wait a moment for the display to change, then select seconds and
press ENTER.
6. You have completed entering Step #1 as a "Wait Step." The display should
now read "01 NX 0002." This indicates that the 1500 is ready to move on to
program Step #2 unless you tell it otherwise.
Programming Step 2
1. Since you do indeed want to program Step #2 now, just press ENTER. The
display should read "02 SP", and you are ready to begin programming Step #2.
Refer again to the program chart. Note that the values you want to achieve are 400°, 90%RH. You want the 1500 to accomplish this in four minutes. You also want to have Ch-1 Events 1, 2 and 4 ON, and Ch-2 Events 1 and 2 ON during Step #2. We will now program all this data for Step #2:
1. Since the display reads "02 SP," and you do want to enter Set Point Data,
first press ENTER. The display changes to "S1" (Set Point for Ch-1).
2. Use the UP/DOWN keys to put "400.0" into the display. Some practice may be necessary to stop at the right value. Then press ENTER. The display
changes to "S2" (Set Point for Ch-2).
3. Place "090.0" into the display and press ENTER. The display changes to
"E1" (Events for Ch-1). The Data Display is blank right now.
4. In the Data Display,
the left.
display and press ENTER. The display will change to "E2" (Events for Ch-2).
5. Likewise, for Ch-2 Events #1 and 2 ON, ENTER the value "0011." The
6. Since Step #2 is only four minutes long, just press ENTER to enter "0000"
7. For Step #2 Minutes, ENTER "0004." The display changes to "SC."
8. For Step #2 Seconds, ENTER "0000." The display changes to "02 00
0400."
So, for Ch-1 Events #1, 2 and 4 ON, place "1011" into the
display changes to "HR" (The "hour" portion of the duration of Step #2).
hours. The display changes to "MN."
Event #1 is represented on the right and Event #4 is on
9. You have successfully programmed Step #2 as a "Set Point Step." The
display is indicating that Step #2 is programmed for a duration of 00 hours, 04 minutes, 00 seconds. Press either ENTER or FCTN; the display changes to "02 NX 0003."
10. Again since you do want to program Step #3 now, press ENTER. The
20 WATLOW Series 1500 User's Manual
display changes to "03 SP."
Sample Program, Chapter 2
Programming Steps 3 through 6
Now that you see how it’s done, enter the data for Set Point
be sure to stop and then return to these instructions when the display reaches SP.”
This is because you will be looping through Steps #4 through #6 three times.
Refer to the table below for
FCTN
STEP
I
#2
Steps
I
#3,4,5
STEP#3
and
I
6.
STEP
#4
I
Steps
STEP#5
#3,4,5,
&
I Step#6
6. But,
“07
Programming Step 7, a Jump Loop Step
You now
performed when Step #6 is completed. “Set Point Step.” We want Step #7 to cause the 1500 to JUMP
times
a"JUMP LOOP
have Steps #1-6 programmed. When you run the program, Step
But, you do not want Step #7 to be a normal
back to
(for a
total
of 3 times through Steps
Step.”
1.
With
the STEP and FUNCTION displays now reading “07 SP,” press the
FCTN key until “07 JL” comes up.
That’s why you’ll program Step #7 as
#4-6).
#7
will be
Step #4 two
Table
3
-
Sample Program Steps
2.
Press the ENTER key. The display changes to “JS” (Jump to Step
3.
Enter the number of the step the program will jump to (“0004”).
ENTER. The display changes to “JC” (Jump Count).
4.
Since you want to jump back to Step #4 two times (total of 3 times through
the loop), enter “0002.” Press ENTER. The display changes to “07 NX 0008.”
5.
Press ENTER (you do want to program Step #8 now). The diiplay
changes to “08 SP.”
Programming Step 8
You’re ready to enter Step below (Stop with the display reading “08 NX 0009”):
S1 S2
El E2 HR MN SC 0000
Recall that you set the Recycle (RC) option to ON, so the program will repeat
indefinitely. You have now entered the entire program.
110.0
065.0
1100
0101
0000 0006
#8.
@
It’s a normal Set Point Step, enter the information
You’re almost ready to run it.
#).
Press
f7
0
CAUTION: Notice
on the Sample Program Chart (at the beginning of this section) that the Series
1500
hoIds
the Step #8 set
points
and Step #8
Event Outputs ON as lt and waits through
Step #1 before beginning program over.
recycles
the
Sample Program, Chapter 2
WATLOW Series
1500 Users Manual
21
Almost Ready to Run
Remember that Step #1 was a “Wait Step.”
you won’t see
You may want ways to get back to Step #1 in 0009”; it is waiting for you to continue programming or monitoring. key to put “0001” in the Data display and press ENTER. The display will show “01 WI” (Step#l is a Wait Step). Now use the FCTN key to move through “SC;” enter a time real time1 0 minutes from now.
any action until the time you programmed into Step
to go
back and
adjust that time to a different value.
CHG
DATA mode. Your display is now reading “08 NX
Even though you press the “RUN” key,
#1.
There
are various
Use the DOWN “HR," “MN,”
and
Start Your Program!
The Wonderful Apple Pie
“RE-START’ “RUN/HALT’ key to start the program.
key
(upper
Sample Program
front panel) to set the program to Step
is all entered and ready. So press the
#1.
Then press the
Monitoring the Action
With the MNTR DATA LED lit, you can now use the FCTN key to check things out.
The control doesn’t seem to show much happening right now since Step #1 is a “Wait
Step.” Press the FCTN key.
programmed into Step programmed time.
In one display you’ll see the “WT” (Wait for) time you
#1. The
1500
is
watching for its real time to match that
f?
CAUTION:
You
now
have a
basic
under­standlng of how the
Series 1500
handles a control
program. Please
read and then continue
to the following
fine “Helpful Hints” sectlon will your the Series 1500 to the highest
level.
22
to refer
points.
The
raise
skills with
WATLOW
Now use the MODE You can watch as real time approaches your programmed time. When the times match, the step lights come ON to coincide with the Events you programmed for Step
As
the
program runs its course, you can use the
key to view the process. As you observe the varying data, the individual functions will become following information to gain an even more detailed understanding.
quite
number changes
clear.
f?
Helpful Hints
Here are some Series 1500 fine points:
1.
Halt Conditions
A.
A Blank Step halts the Series 1500. A “Blank Step” is any step which is not programmed. An “SP” will appear in the Function Display when you select
a Blank Step with the MODE key. The Series 1500 can enter the HALT
B.
Blank Step in a program, or you press the RUN/HALT key while a program
is running.
Series 1500 User’s Manual
key to
move to the
to
After you have
CHG PARA
02.
You will also see the Ch-1 and Ch-2 Event
observed
mode with the
MNTR DATA mode and the FCTN
the program for awhile, read the
condition
in two ways:
TlME
LED also ON.
#2.
It encounters a
Sample Program, Chapter
2
While in the HALT condition, the Series
C.
Event Output conditions from the step it was in when it was halted, or from last
step before a
“Blank
Step” halted the control. made no change to the step data, the 1500 will complete the step from the time it stopped. If you
D.
Thus, if your sample after performing Step
65%RH,
E.
If you press the RUN/HALT key at
and with Ch-1 Events 3 and 4 ON and Ch-2 Events 1 and 3 ON.
with the oven at
change data, the step begins from beginning.
program
#8,
275°, 83.2%RH,
Event 2 ON.
F.
However, you may not want the Step #8 conditions to be retained until tomorrow. Set Points of
You could add a step at the end of the program to establish
75°, 50%RH,
2. Ramping Conditions
A.
When running any Set Point step, the Series 1500 notes what the existing starting conditions are and what the desired ending conditions are.
Then it follows a linear path between the two.
Although the desired ending conditions are specifically programmed into
B.
each step, the beginning conditions depend on the step performed just previously. Know where you are going and where
1500
maintains the Set Points and
If you
restart
the control, having
did not have “Recycle” activated, it would halt
but would continue to hold the oven at 110°,
1-1/2
minutes into Step
and have
Ch-1
and with all Event
Event 1 ON and Ch-2
Cutputs
you've
OFF.
been.
#5,
it would Halt
Step #4 of your Sample Program is an example:
C.
l
As the chart on page 17 shows, you expect to start Step #4 at ramp uniformly down to
time through Step
l
But when you “loop back-from Step
will be ramping upward from 11 O° to
l
If
this is not acceptable to you, you might want to jump loop back to Step
#2
to reach a full
3. Wait Step
A.
A Wait Step maintains the same Set Points that existed at the end of the previous step. With Step #1 as a Wait Step, this means that the 1500 will use the default values for Step #1 first run. values (11 O°, 65%RH, Ch. #1 Events 3 & 4 ON and
ON). If you were to halt the program during Step #3 and then “RESTART’ and run
it, Step #1 would retain the Step #3 values. Since this is not always desirable, you could place a short (1 second) step immediately before the Wait Step to establish different Set Point values.
275°.
This certainly is what happens on the first
#4.
400°
through Step #3 before you start Step
(75°, 50%RH,
However, when it recycles
from Step #6, it will retain the Step #8
#6,
your starting point is 11
275°.
ail events OFF) during its Ch. #2 Events
400°,
#4.
1 &
O°.
3
and
You
When waiting for
B.
“53”
for Minutes, but enter nothing for Hours or Seconds, the programmed
time will be 53 minutes after midnight.
Sample Program, Chapter 2
"Time,”
program all three units (HR-MN-SC).
If you program
WATLOW
Series 1500 User's
Manual
23
In your sample program’s Wait Step, you used “Wait for time” because you
C.
aren’t connected to a variable oven yet. could be as follows:
l
In Step
90%RH.
#2,
you have
Depending upon your system, this may either be impossible
to do, or you may reach the desired conditions much earlier than that.
This could
l
lf you reprogrammed Step #2 as a Wait Step, you could measure how
result
in bad apple pies.
long it takes to achieve the desired conditions, and then react immediately. Much better apple pies.
l
But, there is a choice to make here: The Wait Step does not follow the linear ramping path to reach the new conditions as a Set Point Step
does. A Wait Step uses the previous step’s Set Point. To optimize
control (reduce the overshoot and undershoot) a Wait Step uses the “PID
Parameters” to satisfy the wait condition. the step type which best matches your needs. (The are discussed in the Technical Reference Section that follows).
4. Jump Loop Step (JL)
A.
In the sample program, you used this capability to jump from Step #6 back to Step #4 two times. Your “Jump Count” (JC) was set to
1
to 255 can be used in loops of this type.
A more common use of the Wait Step
allowed
exactly four minutes to reach
400°,
In other words, select
PID
parameters
“2.”
Any value from
B.
If you program
“0”
into
“JC,”
the JUMP will occur infinitely. This set up a never-ending loop if the jump is backwards. Or you can set up a one-time “unconditional jump” if the jump is
forward.
If you use a forward
jump, the 0 choice makes no difference. The jump will be performed once.
l
But why would you ever want to simply jump forward and skip over some steps, you ask?
Suppose your overall program has various functional sections, not all of
which you want to do every time.
By ending each section with a Jump Step, you can easily alter the program to perform various sections in diierent sequences.
During programming work, you may want to develop and test a program in
sections. You can develop the separate sections with unprogrammed
Blank Steps between them, then unite the separate sections laterwith
Jump Steps.
As you learned earlier, jumping or looping into Steps from “different
C.
directions” can sometimes cause unexpected results due to different starting conditions as you enter a new step. Watch for such possibilities.
5. Recycle
With
the Recycle option active, you might expect to see continuous activity as the
process
immediately start Step #1 after it completes Step
repeats the program sequence. This sample program will, in fact,
#8.
action
could
24
WATLOW
Series 1500
User's
Manual
Sample
Program, Chapter
2
However, remember that Step #1 is programmed as a "Wait Step," so it will wait until the same time tomorrow. Since you now know what a Wait Step does, you may want to reprogram Step #1 as a simple "Set Point" step so that you can view the Recycle action. If you do decide to reprogram Step #1, then read the following items.
You cannot alter data or parameters while the program is running. The
1500 must be in the HALT condition.
If you want to move to a different Step #, to see what is programmed there
and then modify that step, enter the CHG DATA mode. Then to re­program it from the CHG DATA mode, go to the appropriate step as de­scribed above. Then press the CLEAR key. All programmed data (for this step only) will clear, and you can now reprogram it.
6. Interactive Changing of Set Points and Events
When in the halted condition, the Series 1500 actively maintains the current set points and Event Output status. At times, you may wish to change these conditions without actually entering and running a profile step. To do this, you must be in the "MONITOR DATA" mode, halted (PROG HALT LED ON) and "RESTART" key just pressed. Then use the function key to move to the desired prompt ("C1", "C2", "E1" or "E2"), use the INCREMENT/DECREMENT keys to move to the desired data, and press the ENTER key. The Series 1500 will quickly change to controlling at the new condition.
Sample Program
Where To Go From Here
As soon as you feel comfortable with the Series 1500 sample program, move on to Installation and Wiring, Chapter 3. Or if you have already installed and wired your system, go to the Programming Chart in Chapter 4 and begin entering and docu-
menting your Series 1500 program.
Sample Program, Chapter 2
WATLOW Series 1500 User's Manual 25
Installation-Wiring
Chapter 3 How to Install and Wire the 1500
This chapter tells you how to install and wire the Series 1500. There are also some suggestions for sensor installation.
guidelines before making your panel cutout.
Sensor Installation Guidelines
Temp-RH
In this 1500 configuration, a dry bulb RTD measures temperature on Channel 1. A wet bulb RTD, in combination with the dry bulb RTD, senses relative humidity on Channel 2. The 1500 calculates the temperature difference between the two bulbs to determine percent relative humidity.
Temp-Temp
For Temp-Temp RTD operation, each dry bulb RTD senses temperature for its channel.
Also, be sure to look at the noise reduction
RTDs
The RTDs for the Series 1500 are customer-supplied items. They may be Gordon Part #S001-0133-0000, which is a wire-wound RTD inside a sealed, 18% Nickel­Silver sheath, made for fast temperature response. The detector element is platinum wire 100 at 0°C with a temperature coefficient of 0.003916Ω/Ω/°C. The sheath is
0.1875 in. (5 mm) diameter, 2.75 in. (70 mm) long. The three wire leads are 48 in. (1219 mm) long and teflon insulated.
Contact Gordon at: 5710 Kenosha St., Box 500 Richmond, IL 60071 Phone: 815/678-2211
Excessive lead length in a two-wire RTD sensor can create indication errors. To combat this, use a three wire sensor in long lead applications.
RTD - The Dry Bulb Installation
We suggest you mount the dry bulb RTD at a location in your process or system where it reads an average temperature. Air flow past this sensor should be moder­ate. The sensor should be thermally insulated from the sensor mounting.
RTD - The Wet Bulb Installation
One way to prepare the RTD wet bulb sensor for reading relative humidity on Chan­nel 2 is to slip a cotton wick over the sensor sheath, Watlow Part # 0830-0111-0000 or an equivalent. Put one end of the wick in water. Some systems use a small trough of water with a float valve to replenish the evaporating liquid.
26 WATLOW Series 1500 User's Manual
Installation - Wiring, Chapter 3
Installation-Wiring
Use distilled water to avoid mineral deposits on the wick that can change the %RH reading. The wicking distance should be as short as practical. The cotton wick must be free of any sizing and starch. The "float valve-trough" systems should also re­move the water when the ambient temperature goes below freezing or above boiling.
To obtain valid %RH readings, air flow past the %RH sensor must be sufficient to evaporate the maximum amount of water from the wick without drying it out. Evapora­tion cools the sensor; the amount of cooling relates directly to the relative humidity present.
Thermally insulate the wet bulb sensor from the sensor mounting. This prevents mounting surface heat from flowing into the sensor sheath and changing the wet bulb temperature. (The result of such heat flow will be a humidity reading higher than actual.)
Thermocouple Installation Guidelines
Locate the thermocouples in your process or system where they read average tem­peratures. Air flow past these sensors should be moderate. The sensors must be thermally insulated from the sensor mounting. Both thermocouples must be un-
grounded to obtain correct readings.
Wiring Practices for Preventing Noise
Most noise problems stem from wiring practices, the major means of coupling noise from its sources to the control circuit.
An outstanding information resource for wiring guidelines is the IEEE Standard No. 518-1982 and is available from IEEE, Inc. 345 East 47th Street, New York, NY
10017.
Noise Sources
Switches and relay contacts operating inductive loads such as motors, coils, solenoids, and relays, etc.
Thyristors or other semiconductor devices which are not zero crossover-fired (randomly-fired or phase angle-fired devices).
All welding machinery.
Heavy current carrying conductors.
Fluorescent and neon lights.
How to Decrease Noise Sensitivity
1. Our extensive noise testing has shown that the Series 1500 has a good level of noise immunity. In extremely noisy environments, however, the testing has shown that the most effective RFI techniques are:
a. An RFI filter (such as the Corcom 1ER1) in the power line mounted exter-
nal to the Series 1500 case. The RFI filter should be an RFI sealed type.
b. Shielded power cord with the shield terminated to the Series 1500 case
Installation - Wiring, Chapter 3
WATLOW Series 1500 User's Manual 27
Loading...
+ 63 hidden pages