The information in this document is subject to change without notice and should not be
construed as a commitment by Waters Corporation. Waters Corporation assumes no
responsibility for any errors that may appear in this document. This manual is believed to
be complete and accurate at the time of publication. In no event shall Waters Corporation
be liable for incidental or consequential damages in connection with or arising from the
use of this manual.
1993 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA.
ALL RIGHTS RESERVED. THIS BOOK OR PARTS THEREOF MAY NOT BE
REPRODUCED IN ANY FORM WITHOUT THE WRITTEN PERMISSION OF THE
PUBLISHER.
LAC/E, PowerLine, and Waters are trademarks of Waters Corporation.
Millex is a trademark of Millipore Corporation.
Torx is a trademark of Camcar, Division of Textron, Inc.
The quality management systems of Waters manufacturing facilities comply with
the International Organization for Standardization International Standard ISO 9001
Quality Management and Quality Assurance Standards. Waters quality
management systems are periodically audited by the registering body to ensure
compliance.
Attention: The Waters 717plus Autosampler may be used for IN VITRO diagnostic
STOP
applications. This is a highly sensitive instrument. Read the accompanying operator's
manual before using.
When using the instrument, follow generally accepted procedures for quality control and
methods development.
If you observe a change in the retention of a particular compound, in the resolution
between two compounds, or in peak shape, immediately take steps to determine the
reason for the changes. Until you determine the cause of a change, do not rely upon the
results of the separations.
Attention: Changes or modifications to this unit not expressly approved by the party
STOP
responsible for compliance could void the user’s authority to operate the equipment.
STOP
Attention:
This equipment has been tested and found to comply with the limits for a Class
A digital device, pursuant to P art 15 of the FCC rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses, and can radiate radio
frequency energy, and, if not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications . Operation of this equipment in a
residential area is likely to cause harmful interference, in which case you must correct the
interference at your own expense.
Shielded cables must be used with this unit to ensure compliance with the Class A FCC
limits.
Caution: For continued protection against fire hazard, replace fuses with those of the
same type and rating.
Caution:
To avoid the possibility of electric shock, do not remove the power supply co vers .
The power supply does not contain user-serviceable components.
Attention: To meet the regulatory requirements of immunity from external electrical
disturbances that may affect the performance of this instrument, do not use cables longer
than 9.8 feet (3 meters) when you make connections to the screw-type barrier terminal
strips. In addition, ensure you always connect the shield of the cable to chassis ground at
one instrument only.
The Installation Category (Overvoltage Category) for this instrument is Level II. The Level
II category pertains to equipment that receiv es its electrical power from a local level, such
as an electrical wall outlet.
This digital apparatus does not exceed the Class A limits for radio noise
emissions from digital apparatus set forth in the Radio Interference
Regulations of the Canadian Department of Communications.
Le présent appareil numérique n’émet pas de bruits radioélectriques
dépassant les limites applicables aux appareils numériques de la classe A
prescrites dans les règlements sur le brouillage radioélectrique édictés par le
Ministère des Communications du Canada.
717plus Autosampler. It also includes installation, maintenance, and troubleshooting
procedures.
Audience
The audience for this operator’s manual includes chemists and technicians who will use
the autosampler to make injections for chromatographic analysis.
describes the features and use of the Waters
Structure of This Manual
This manual is divided into chapters. Each chapter page is marked with a tab and a footer
facilitating access to information within the chapter.
The table below describes the material covered in each chapter.
Chapter 1, IntroductionDescribes the Waters 717plus Autosampler.
Chapter 2, InstallationProvides procedures for installing the Waters
717plus and making fluidic and electrical
connections.
Chapter 3, Startup and
Configuration
Chapter 4, Operating the
Autosampler
Chapter 5, MaintenanceProvides procedures for defrosting the
Provides procedures for starting up, calibrating,
configuring, verifying, and powering down.
Provides procedures for preparing and loading
samples, programming and using the automatic
features, and making a stat run.
heater/cooler and for replacing the syringe, seal
pak, and in-line filters.
For detailed information about using the Adobe Acrobat Reader, refer to the
Acrobat Reader Online Guide
navigating and printing electronic documentation from Adobe Acrobat Reader.
. This Online Guide covers procedures such as viewing,
Adobe
Printing From This Electronic Document
Adobe Acrobat Reader lets you easily print pages, pages ranges, or the entire electronic
document by selecting Print from the File menu. For optimum print quantity, Waters
recommends that you specify a Postscript printer driver for your printer. Ideally, use a
printer that supports 600 dpi print resolution.
Conventions Used in This Guide
This guide uses the following conventions to make text easier to understand.
•Purple Te xt indicates user action. For example:
Press 0, then press Enter for the remaining fields.
•
Italic
text denotes new or important words, and is also used for emphasis. For
• Underlined, Blue Color textindicates hypertext cross-references to a specific
chapter, section, subsection, or sidehead. Clicking this topic using the hand symbol
automatically brings you to this topic within the electronic document. Right-clicking
and selecting Go Back from the popup context menu brings you back to the
originating topic. For example:
For details on selecting a needle wash solution for a specific sample, refer to
Section 4.3, Selecting a Needle Wash Solvent.
Notes, Attentions, and Cautions
• Notes call out information that is important to the operator. For example:
Note:Record your results before you proceed to the next step.
STOP
• Attentions provide information about preventing possible damage to the system or
equipment. For example:
Attention: To avoid damaging the detector flow cell, do not touch the flow cell
window.
• Cautions provide information essential to the safety of the operator. For example:
Caution: To avoid chemical or electrical hazards, always observe safe laboratory
practices when operating the system.
Caution:To avoid the possibility of electrical shock and possible injury, always turn
off the detector and unplug the power cord before performing maintenance
procedures.
This chapter introduces you to the Waters 717plus Autosampler. The autosampler’s
operational features are summarized and its components described.
1.1 Description
The Waters 717plus Autosampler injects samples into a high performance liquid
chromatography (HPLC) system. The Waters 717plus automates sampling by storing up
to 96 sampling methods or “steps”. Each method controls all injection parameters for a
sequence of vials.
The Waters 717plus (Figure 1-1) may be used as either a stand-alone unit or as a
component in a Waters IEEE-488 based HPLC system. As part of an IEEE-488 based
system, the autosampler may be controlled by Waters PowerLine System Controllers or
Waters Data Workstations.
Inject Stop - Occurs when the programmed run time expires. Signals data
integrators that the chromatography run is complete. Contact closure is initiated by
the autosampler.
•Hold - Halts the autosampler until the signal changes. Can be used to prevent
sample injections if an LC system module fails or while a data acquisition system is
preparing for the next run. Signal is received from external devices.
•Report Delay - Delays the next injection until a data integrator can complete a
report for the previous run. The delay is a programmable interval that occurs after
the Inject Stop signal.
•Programmable Event - The 717plus includes a programmable event output that
can control external devices during a run.
IEEE-488 Bus Communication
As part of a Waters IEEE-488 system, the Waters 717plus autosampler accepts
programming and control signals from an LC system controller.
RS-232 Communications
The Waters 717plus autosampler sends data over an RS-232 communication line. The
data is in ASCII format and contains:
VIAL # nn INJ # n OF n
INJ VOL nnnn RUN TIME nnnnn
Purge and Compression
Purge and compression testing capability includes:
•Purge - Fills the sample loop and injection syringe with fresh mobile phase.
•Compression - Tests the sample loop for gas content (compressibility).
Compressibility limits may be set for specific mobile phases. Testing may be
automatic or operator initiated.
Needle Wash System
The needle wash system:
• Prevents cross-contamination of samples
• Extends seal life by removing the mobile phase and the sample from the sampling
This chapter describes 717plus installation and setup. It includes procedures for making
connections, voltage selection, and configuring optional hardware.
2.1 Site Requirements
2
2
•
•
•
Required Space
Before beginning the installation procedure, note the following autosampler site and utility
requirements (Figure 2-1):
Note: The autosampler site requires room for a waste bottle and a needle wash solvent
reservoir.
Bench space - Allow 21 x 23.5 inches (53 x 72 cm) for the autosampler. This space
includes ventilation sides 2.5 inches (6.3 cm) and back 3.5 inches (8.75 cm).
Note: Ensure the bench top must be level to ensure proper drainage of the solvent
spill lines.
Bench space with heater/cooler installed - Allow an additional three inches on
the back and on each side next to the air vents.
Note: To avoid heater/cooler performance degradation, do not place the rear of the
unit in an enclosed space, under a shelf, or near a heat source.
Column location - Ensure the red fluid line reaches the column. Lengthening this
line may result in peak broadening.
•Waste and solvent container location - Place the needle wash reservoir at bench
Figure 2-1 Waters 717plus Autosampler Site Requirements
Operating Conditions
Locate the Waters 717plus in a clean area free from extremes in temperature, humidity,
strong radio frequency radiation, appreciable shock or vibration. The Waters 717plus
operates within the following ranges:
Temperature
Humidity
Do not locate the Waters 717plus in direct sunlight, near heat registers, or under air
conditioning vents.
4 to 40 °C
Note that cooler performance is derated above 27 °C
80% maximum relative humidity, non-condensing
Weight
Requires a bench capable of supporting at least 75 lbs.
Requires 5 Amps at 110 Vac (nominal) or 4 Amps at 220 Vac (nominal) from a grounded
outlet.
2.2 Unpacking and Inspection
Unpacking
To unpack your unit:
1. Remove the straps and open the top of the shipping carton.
2. Remove the Startup Kit.
3. Remove the packing material from the carton.
4. Lift the carton sides from the base of the carton to expose the autosampler.
Caution: The autosampler weighs approximately 63 pounds (28.6 kg). Two
people are required to safely lift the autosampler; use care when lifting.
5. Grasp the autosampler from the bottom to lift. Do not lift by holding the front
panel.
6. Place the 717plus securely on the bench top. Allow clearance for the power cord
and ventilation.
List of Components
The autosampler shipping carton contains:
• Waters 717plus autosampler
• Startup Kit
Reporting Damage
Check the packing list against the contents as you unpack, and confirm that all items are
included. If you find any damage or discrepancy, immediately report any damage to both
the shipping carrier and to your Technical Service representative. You can contact Waters
Technical Service at (800) 252-4752,
customers, call your local Waters subsidiary or your local Waters Technical Service
Representative, or call Waters corporate headquarters for assistance at (508) 478-2000
(U.S.).
U.S. and Canadian customers only
. Other
2
If any items are damaged, use the shipping container for subsequent claim purposes.
Attention: Do not use the voltage selector card to compensate for low line
STOP
voltage. Always use a selection that equals the nominal line voltage, or power
surges may damage the instrument.
4. The edge of the voltage card is labeled with four voltages (Figure 2-3). Position
the voltage indicator pin on the opposite side of the card from the required
voltage. Seat the indicator in the notch on the edge of the card.
Indicator pin
2
4
0
0
1
240 V
2
0
0
2
0
100V120 V220 V
1
0
0
1
2
2
2
0
0
1
2
0
0
2
2
2
2
4
2
2
1
0
2
0
0
4
Notch for indicator pin
Figure 2-3 Voltage Selector Card
5. Insert the voltage selector card into the housing until it is firmly seated.
Changing Fuses
Change the fuse block arrangement if the fuse requirements change from single 1/4-inch x
1 1/4-inch to dual 5 mm x 20 mm.
To change the fuse configuration:
1. Loosen the Phillips screw two turns and remove the fuse block by tilting it up and
sliding it away from the screw (Figure 2-4).
2. Or ient the fuse block for the desired voltage (Figure 2-5) as follows:
• For single 1/4-inch x 1 1/4-inch fuse arrangements, install the fuse block with
the fuse and jumper bar facing away from the cover.
• For dual 5 mm x 20 mm fuse arrangements, install the fuse block with the two
fuses facing away from the cover.
100/120 Fuse
220/240 Fuse
Dual
Configuration
Figure 2-5 Fuse Installation
3. Install the appropriate fuse or fuses as labeled on the rear panel near the power
cord connector.
4. Ensure that the assembly is seated against the shoulder of the screw mount and
tighten the Phillips screw.
5. Install the fuse block and cover in the rear panel of the autosampler. Make sure
that the indicator pin fits through the appropriate opening on the left end of the
cover.
Caution: Observe safe laboratory practices when handling solvents. Refer to the
Material Safety Data sheets for the solvents in use.
The Waters 717plus Autosampler fluid lines are:
• Green - Plastic tubing for needle wash inlet
• Yellow - Plastic tubing for needle wash waste
• Clear - Plastic tubing for purging the sample loop waste
• Blue - High pressure steel tubing, 0.04-inch (1 mm) I.D., connects the autosampler
to the pump
• Red - High pressure steel tubing, 0.009-inch (0.2 mm) I.D., connects the
autosampler to the column
• Drain - Two 1/4-inch clear lines attached to the drip trays on bottom of the
autosampler. One line drains solvent and sample spills. The other line drains water
when the heater/cooler is defrosted.
1. Cut the plastic tie securing the coil of tubing located at the rear of the autosampler.
The coil consists of yellow, clear, and green tubing.
2. Fill a flask with a filtered solution of 90 percent methanol/10 percent water. Use
this solution for the needle wash during installation. Place the flask at bench
level, and put the end of the green line in the solution.
Note: For details on selecting a needle wash solution for a specific sample, refer to
Section 4.3, Selecting a Needle Wash Solvent.
3. Place the ends of the yellow (needle wash waste), and clear (purge waste) lines
in appropriate containers.
4. Connect the blue line to the pump. Do not connect the red line to the column
inlet until you flush the system (see Section 3.3, Flushing the Autosampler).
Attention: Ensure that the red and blue fluid lines are in the side panel slots and not
STOP
pinched between the cover and side panels.
Installing Drain Lines
Using parts from the 717plus Startup Kit, connect the two drain lines to the drain pans
located on the bottom of the autosampler (Figure 2-7) as follows:
1. Place an appropriate waste container for solvent and sample leaks below bench
level.
2. Cut a length of 1/4-inch tubing long enough to reach the waste container from
the autosampler.
3. Slip the tubing strain relief onto the tubing and push it onto the fitting for the
solvent drain pan (located on left side).
4. Use the elbows to facilitate bends in the tubing.
5. If you have a heater/cooler, connect a drain line for condensation/defrosting
following steps 1 through 4.
The signals available on the back panel terminal strip include:
Table 2-1 Terminal Strip Signals
2
Terminal
Numbers
1, 2HoldInput signals the autosampler to hold (pause) the
4, 5HoldInput signals the autosampler to hold (pause) the
6, 7Inject StartOutput signals the start of an injection
9, 10Inject StartOutput signals the start of an injection
11, 12Prog EventProgrammable event output signal
14, 15Inject StopOutput signals the end of the programmed run
3, 8, 13GroundSignal ground for cable shields
For specific information on signal use with other Waters equipment, refer to the
appropriate manual.
Attention: Do not connect two cables to the same autosampler terminal on the terminal
strip.
Attention: To meet the regulatory requirements of immunity from external electrical
disturbances that may affect the performance of this instrument, do not use cables longer
than 9.8 feet (3 meters) when you make connections to the screw-type barrier terminal
strips. In addition, ensure you always connect the shield of the cable to chassis ground at
one instrument only.
Using a signal (contact closure) from the Waters 717plus autosampler, a Waters system
controller can coordinate the start of detectors, pumps, and computers. F or specific details
on programming a system controller, refer to the system controller operator’s manual.
To connect a system controller, connect the Inject Startterminal on the autosampler as
outlined in Table 2-2.
Table 2-2 System Controller Connections
Autosampler
Terminal
Inject StartMillennium Chromatography ManagerEvent In for
Inject StartExpertEase SAT/INEvent In for
Inject StartExpertEase SIMStart Inject and
Inject Start600 SeriesINJ and GND
Inject StartMaxima Watch Dog Event BoxIN 1 or IN 2
Controller
Controller
channel 1 or 2
channel 1 or 2
COM
Making LAC/E Connections
Use the Inject Start signal from the Waters 717plus:
• If the LAC/E system contains components such as detectors that are not connected
to the IEEE-488 bus.
• If the ExpertEase multi-method is set to “Start by Injector.”
Terminal
2
Making Detector Connections
Waters detectors use the start of the injection for:
• Auto zeroing
• Start of data transfer
• Detector chart marking
For specific details on programming a detector for an auto zero, refer to the detector
operator’s manual.
Waters 600 series gradient controllers provide a Hold signal to stop the next injection.
Connect the Waters 717plus autosampler to the 600 series pump as follows:
Autosampler600 Series
HOLDHOLD and GRN
Making Event Output Connections
Connect a signal cable to the Prog Event terminals (pins 11 and 12, pin 13 for ground) on
the 717plus autosampler terminal strip (Figure 2-9).
2.5.2 Making IEEE-488 Network Connections
2
The IEEE-488 bus provides a communication link between the Waters 717plus
autosampler and system controllers such as the Millennium Chromatography Manager,
ExpertEase data systems (through a LAC/E module), and 600 series system controllers.
Making IEEE-488 Bus Connections
With all instruments powered down, make the following cable connections (Figure 2-10):
Attention: Power down all instruments on the IEEE-488 bus before making connections.
1. Connect one end of the IEEE-488 cable to the IEEE connector of the 717plus
autosampler.
2. Connect the other end of the IEEE-488 cable to a system controller or to any
one of your IEEE-488 components.
Ensure that all IEEE cable screws are fastened securely.
In order to interact with the PowerLine system controller or LAC/E, each device on the
IEEE-488 communications bus must have a unique individual address.
Note: To use the autosampler on the IEEE-488 network, change the autosamplers
IEEE-488 address (refer to Section 3.6.2, Setting Parameters for LC System Operation).
2.5.3 Connecting RS-232 Output to Data Module
The RS-232 connector on the rear panel allows the autosampler to transfer injection
parameters to a data integrator such as the Waters 746 Data Integrator.
You may use the RS-232 port with any device that can read ASCII code.
To connect to the RS-232 port, plug the cable into the modified phone jack connector on
Figure 2-13 Connecting the Heater/Cooler Power Cord and Temperature Probe
Opening For
Heater/Cooler
Power Cable
6. Push the temperature probe through the grommet located on the floor of the unit
behind the carousel drive (Figure 2-14). Push the probe down until it reaches the
stop.
7. Attach the temperature probe connector to the main circuit board. The
temperature probe connects to the main circuit board at J18, located to the lower
right of the battery (see Figure 2-13).
8. Replace the autosampler cover (see Section 2.6.1, Removing/Replacing the
717plus Cover) and reconnect the power cord.
2.6.3 Installing the Large Sample Loop
To inject sample volumes greater than 200 µL, you must install the auxiliary loop.
Caution: To avoid the possibility of electric shock, disconnect the autosampler pow er cord
Before calibrating the seal pak, run a purge to ensure that there is no air in the
autosampler, the mobile phase is fresh, and there are no bubbles in the syringe.
1. Using a degassed solution of 100 percent methanol, set the solvent delivery pump
to 1 mL/min.
2. Remove the syringe cover panel to examine the syringe for bubbles.
3. From the Main Menu, press Edit Purge. The Purge and Compressibility menu
appears (Figure 3-13), listing the default parameter values.
4. Use the arrow keys to set Compression Check to No, then press Enter. Press
Start Purge.
5. Run at least 2 purges before calibrating the seal pak.
6. If the purging fails to remove bubbles from the syringe, perform a manual purge
(Section 6.4.1, Man
ual Syringe Purge).
3.4.2 Auto Zeroing the Transducer
Auto zeroing the transducer ensures that pressure readings are accurate.
3
To zero the transducer:
1. Ensure that the system pressure is at 0 psi.
2. Set pump flow to 0 mL/min.
3. To auto zero the transducer, press the keys as outlined in the flowchart below
(Figure 3-5).
After pressing Auto zero Pressure, the system requires approximately 2 minutes to auto
zero the pressure.
If the system is unable to auto zero , manually adjust the transducer (Section 6.4.3, Man ual
4. The Inject Start default pulse width is 0.5 seconds. Change the pulse width if the
instrument connected to the autosampler requires a different setting.
Inject Stop Pulse
5. The Inject Stop Pulse signal default pulse width is 0.5 seconds. Change the
pulse width if the instrument connected to the autosampler requires a different
setting.
Setting IEEE-488 Address
To ensure that the autosampler is correctly identified when it is part of an IEEE-488
network, verify that the address is a unique number from 2 to 29. The default IEEE-488
address is set to 30.
Note: To use the Waters 717plus Autosampler in an IEEE-488 network, you must change
the address.
To change the address:
1. From the 717plus Main Menu, press Config Page.
2. From Configuration Page 1, press Next Cfg Page three times to view Page 4 of
the Configuration Pages.
3. On Configuration Page 4 (Figure 3-11), highlight the desired IEEE Address field,
type the new address, and press Enter.
3
Note: The address must be unique to the system and must be a number between 2
and 29.
4. Press Reboot Systemto configure and store the new address.
Note: If the system controller is powered up, rescan the system to view the new
717plus Autosampler address. Refer to the system controller operator's manual for
details.
Press REBOOT SYSTEM softkey to
install desired IEEE address.
REBOOT
SYSTEM
PrevCfg
Page
Figure 3-11 Setting IEEE-488 Address
3.6.3 Entering Hardware Changes
To enable the autosampler to adjust to changes in hardware configuration (syringe and
sample loop size), you must enter the changes on Configuration Page 3 (Figure 3-12).
6. If the compression exceeds the default compressible volume, degas the mobile
phase and run the compression check again.
7. If the compression check still fails, perform a leak test (Section 6.4.2, Valve Leak
Test).
3.8 Power Down and Storage
If the mobile phase contains buffers, remove the buffers before power down and storage.
This section describes:
• Removing the buffered mobile phase
• Powering down the 717plus Autosampler
• Long-term storage
3.8.1 Removing Buffered Mobile Phase
Remove the mobile phase containing buffer salts as described below:
1. Replace the buffered mobile phase with a solution of 100 percent water, and flush
the sampling system for 10 minutes at 1 mL/min.
2. Run three purges to ensure that the syr inge is well-purged (see Section 3.4.1,
Purging the Autosampler).
3. Replace the needle wash solvent with 90 percent methanol 10 percent water,
and perform a needle wash prime (Section 3.5, Priming the Needle Wash Pump).
4. Replace the water with a solution of 90 percent methanol and 10 percent water.
Flush the sampling system for 3 minutes at 1 mL/min.
5. Purge the autosampler.
3
3.8.2 Powering Down
This section covers procedures for powering down the Waters 717plus autosampler for 1
to 14 days.
1. If the current mobile phase is water, replace the water with a solution of 90 percent
methanol and 10 percent water. Flush the sampling system for 3 minutes at
1 mL/min.
2. Purge the autosampler (see Section 3.4.1, Purging the Autosampler).
3. To power down the autosampler, press 1/0 (on/off).
This chapter covers 717plus autosampler operation. Provided are procedures for
preparing and loading samples, programming and using the automatic features, and
making a stat run.
4.1 Overview of Operation
Figure 4-1 provides an overview of autosampling.
Note: If the Waters 717plus autosampler is part of an IEEE-488 network, power up the
autosampler before powering up the system controller.
Note: Prevent precipitation of buffered salts or of sample through careful consideration of
solvent concentration and pH. Whenever possible, use mobile phase as the solvent for
dissolving samples.
Observe the following guidelines when selecting a sample solvent:
• Normal phase column - Use the same solvent that is used for the mobile phase.
• Reverse phase column - Use the same or weaker concentration of the mobile
phase organic component.
• Ion exchange - Use a low salt concentration.
• GPC - Dissolve in the mobile phase (preferably from the same bottle).
Filtering Samples
Unless restricted by sample chemistry, filter prepared samples before placing them in the
sample vials. Filtering with a 0.5 µ filter, such as a Millipore Millex filter unit, prevents
particulate accumulation on internal filters or the column inlet fitting which would result in
an increase in backpressure.
4.2.2 Selecting Sample Vials
The Waters 717plus autosampler is compatible with the complete line of Waters
chromatography sample vials.
Attention: Use only sample vials and supplies from Waters. Use of non-Waters sample
vials, low-volume inserts, or septa may result in damage to the needle.
Selecting Sample Vials
Select a sample vial based on:
• Chemistry of the sample
• Volume of sample
• Volume of injection
• Number of injections
• Volatility of sample
For detail on sample vials, see Appendix E, Selecting Sample Vials.
dimples in) may cause the cap to fit incorrectly. Also, the use of self-sealing septa may
cause variable injector volumes on multiple injections due to vacuum buildup.
Do not overtighten the cap with a self-sealing septum. Ov ertightening (septum
Screw Cap
Limited Volume
Insert
Snap-on Cap
Septum
Figure 4-3 Sample Vial Assembly
4
4.3 Selecting a Needle Wash Solvent
The needle wash prevents cross-contamination of samples by wiping the needle with
porous, wetted frits. Select a needle wash solvent based on the sample and mobile phase
chemistry . Table 4-1 lists some suggested needle wash solvents. High sample
concentrations may require stronger needle wash solvents.
Attention: Do not use buffers in the needle wash solvent. Buffers decreases seal life.
• Compressibility of the mobile phase due to dissolved gas
• Integrity of the autoinjector sample loop
Ensuring Reproducibility
1. Run a purge and a compression check:
• To remove mobile phase that has been sitting in the system during a shutdown.
• To check the system on a daily basis for air or solvent leaks.
• To remove the previous mobile phase when changing to a new mobile phase.
• To establish the compressibility for a new mobile phase.
2. Run a purge and a compression check (see Section 4.5.3, Performing a
Compression Check) on the degassed mobile phase.
3. If the mobile phase fails the compression check, degas the solvent (see
Appendix D, Reproducibility and Solvent Degassing).
4. If the degassed mobile phase again fails the compression check, run a leak test
(see Section 6.4.2, Valve Leak Test) to verify the integrity of the sample loop.
4.5.2 Performing a Purge
Perform a purge from the Purge and Compressibility Page.
Purging the Autosampler
To perform a purge using the default values, press the Waters 717plus autosampler keys
as indicated below (Figure 4-6):
Purge Time (min):
Compression Check?
Compression Pressure (psi):
Pressure Decay Time (min):
Compressible Volume ( l):
Pressure Change (psi)
Compressed Volume (psi)
Start
Purge
1.0
Yes
30
0.2
20
Main
Page
No
Figure 4-7 Purge and Compressibility Page
2. Calculate the purge time based on the flow rate and sample loop size. The time
must be sufficient to allow three volume changes of the autosampler. The
autosampler volume in Table 4-2 includes the inlet and outlet lines.
Table 4-3 lists sample purge times for the two sample loops at selected flow rates.
Table 4-3 Purge Time Examples
Sample
Loop
Normal0.212.5
Auxiliary0.519.4
Flow Rate
(mL/min)
0.55.0
>12.5
19.7
24.8
Purge Time
(min)
4.5.3 Performing a Compression Check
The compression check tests for gas in the mobile phase, and, when enabled, occurs
immediately after a purge. For most mobile phases, use the default parameters.
Note: The compression check is only valid using the standard 250 µL syringe and 200 µL
sample loop.
Change the default parameters if:
• Mobile phase is highly compressible (see Appendix D, Reproducibility and Solvent
Degassing).
• Testing for leaks (see Section 6.4.2, Valve Leak Test).
4
Compression Parameters
When enabled, the compression check is executed immediately after a purge. The
parameters are:
For testing the compressibility of the mobile phase, use 30 psi. For leak testing, use the
maximum pressure.
Range30 to 350 psi
Default30
Pressure Decay Time
Sets the length of time the system pauses before recompressing the mobile phase in the
sample loop. The default value is sufficient for most applications. The longer the decay
time, the more rigorous the test.
Range0.0 to 3.0 min
Default0.2 min
Compressible V olume
4
The default value is sufficient for most applications. Changes to the compression limit are
dependent on the solvent used for the mobile phase. If the mobile phase is highly
compressible, increase the compression limit setting (see Appendix D, Reproducibility and
Use Auto Run mode to program, run, and monitor sample injections. Auto Run mode
allows:
• Multi-method programming - Allows programming of up to 96 different steps
(methods). Each step can use different numbers of injections, volumes, run times,
and optional functions.
• Unattended operation - Provides injection start and stop output signals for
controlling external devices.
• Programmable output signal - Allows the 717plus to control external devices
during a run.
• Auto Standard, Auto Addition, and Auto Transfer functions - Allow automated
operation for standards, pre-injection addition of samples, and automatic
pre-injection transfer and mixing.
Overview of Programming
Figure 4-9 is an overview of the steps involved in programming and running an auto run.
Display the Edit: Auto Run screen by pressing the Auto Page key on the Main Menu.
The Edit: Auto Run screen (see Figure 4-10) lets you program up to 96 steps.
You must enter a value for each parameter in a step. Parameters you enter are saved in
permanent memory.
Before Programming
Load the required vials in the carousel, noting the carousel positions.
Note: Standard vials and common vials do not ha ve to be loaded in consecutive positions.
Programming
To program:
Step
1. Place the cursor on a blank line, then press Enter.
The 717plus inserts a new line and copies the information, including step
function and option settings, from the previous line.
You can program up to 96 different steps. Each vial included in the step shares
the same number of injections, injection volume, run time, step functions, and
options. By changing steps and step parameters, you can run a different method
using the same set of vials.
4
Vial From/To
2. Enter a vial number in the From Vial and To Vial fields. Vial number defaults to
the values you entered on the previous line. Valid entries are 1 to 48 or 1 to 96,
depending on the type of carousel in use.
When injecting from a single vial, enter the same vial number in the From and To
fields. When injecting from a sequence of vials, enter the starting vial in the From
Vial field and the last vial in the To Vial field.
STOP
# Inj
91Operating the Autosampler
Attention: The vial injection sequence entered in the From/To field is continuous. If
a vial is missing, the autosampler will inject air into the LC system.
3. Enter the number of injections to make from each vial. Valid entries are 1 to 99.
4. Enter the injection volume. The maximum injection volume depends on the size
of the sample loop.
Sample Loop SizeRange (µL)
Standard 0 to 200
Auxiliary0 to 2000
Run Time
5. Enter the r un time in minutes. Valid entries are 0 to 999.9.
The run time clock starts when the sample enters the mobile phase stream. At
the same time, the autosampler closes the Inject Start contact. When the run
time clock expires, the autosampler closes the Inject Stop contact and begins
another injection sequence unless a Report Delay has been configured.
Note: You can delay the next injection in order to complete an integration printout.
Use the Report Delay or External Hold input signal (see Section 3.6.2, Setting
Parameters for LC System Operation).
Step Functions
6. Program step functions (described in Section 4.6.2, Programming Step
Functions).
Note: You can program only one set of step functions and one option per step.
4.6.2 Programming Step Functions
Use the Step Function screen to select the following parameters for each Auto Run step:
• Needle Position
• Purge After Step
• Options for an Auto Run step
• Programmable Event outputs
This section describes the Step Function screen (Figure 4-11).
Editing Step: 1
Needle Position (%)? 25 50 75 100
Purge After Step? Yes No
Option? None Stds Addn Trns
Run Event (0=Off 1=On)
Time Time Prog Event
10.01.0
2.0
3.0
4.0
1
0
1
0
Trns
Page
Auto
Page
Figure 4-11 Edit: Step Function Screen
Note: Figure 4-11 is an example of the Edit: Step Function screen with the Auto Transfer
option selected. When other options are selected, keys for the selected function appear.
4
If no option is enabled, the option field displays “None”, and no option key appears.
Programming Step Functions
To program step functions:
1. Display the Edit: Step Function screen by pressing Step Function on the Edit:A uto
Run screen.
The currently selected step number appears on the Edit: Step Function screen.
Editing Step
2. To change the step number, enter a new number on the keypad and press Enter .
Needle Position
3. Move the cursor to the Needle Position% line. This determines needle
withdrawal position. Use the right/left arrow keys to select the desired position of
1. From the Edit: Auto Run screen (see Figure 4-10), select a step, then press Step
Function to display the Edit: Step Function screen.
Note: For instructions on setting Needle Position and Purge, refer to Section 4.6.2,
Programming Step Functions.
2. From the Edit: Step Function screen (see Figure 4-11), press the down-arrow
key to move the cursor to the Option? field.
3. Press the right-arrowkey to move the cursor to the Stds selection. The Stds
Page key appears on the screen.
4. Press Stds Page to display the Edit: Auto Standards screen (Figure 4-12).
IDLE
Edit: Auto Standards 1 of 1 Stds
Editing Step:
Run Standards? Yes No
Vials Between: 1
Vial
Delete
Table
1
Delete
Inj
Vol
15.0
Line
#
Inj
1
Run
Time
1.0
Step
Function
Figure 4-12 Edit: Auto Standards Screen
5. At the Run Standards? field, select Yes (default).
6. Move the cursor to the Vials Between field. Enter the number of sample vials to
run between standards (1 to 48 or 1 to 96, depending on carousel type).
7. Press the down-arrow key to move the cursor to the Vial field. Enter the
following parameters for each standard vial. You can enter up to five standard
vials for each Auto Run step.
ParameterDescription/Range
Vial Carousel number, 1 to 96
4
Inj V ol0.0 to 200.0 µL or
0.0 to 2000.0 µL
# Inj Number of injections for standard vial,
1 to 99
Run Time Total run time from injection to Inject
Stop signal, 0.0 to 999.9 min
1
Dependent on sample loop installed in Waters 717plus
1
Running Auto Standards
To start the Auto Run:
1. When programming is complete, press Step Function, then Auto Page to return to
the Edit: Auto Run screen.
2. Press Start Auto. The Auto Standards routine runs as part of each programmed
step. A message appears indicating that Auto Standards is running. You cannot
change screen settings of the currently running step.
Turning off Auto Standards
To turn off Auto Addition on the Edit Auto Addition screen, select No at the Run Addition?
field. The information programmed f or A uto Standards is sa ved, b ut A uto Standards do not
run.
When Auto Standards are turned off, the Stds notation is not displayed on the Auto Run
screen.
Note: If you turn off Auto Standards on the Edit: Step Function screen (by setting the
Option field to None), you delete the information programmed for Auto Standards.
In an Auto Standards run, standards always run at the beginning of a step, and may run
during or at the end of a step.
Whether standards run during or at the end of a step is determined by the number of vials
in a step and the number of vials specified to run between standards.
Note: If you run a stat sample during an Auto Standard run, the stat sample is not
considered to be one of the sample vials running between standards.
Standards During or at End of Run
If the number of vials in the step is
equal
to the Vials Between value, standards run at the
beginning and at the end of the step.
If the number of vials in the step is a
multiple
of the Vials Between v alue , standards run at
the beginning, during, and at the end of the step (Figure 4-13).
Step
1
ial
1
ial
1
Auto Run
From
2
ial
2
3
To
7
ial
1
ial
ial
ial
4
1
Auto Standards
Vials Between = 3
Standard Vial = 1
ial
5
6
Figure 4-13 Example Auto Standard Run 1
No Standards at End of Run
Standards do not run after the last sample if (Figure 4-14):
Auto Addition withdraws specified volumes from common vials, then withdraws from a
sample vial. The common solutions and the sample are all contained
before injection. You can set a time delay for the injection of the sample and common
solutions.
Auto Addition is useful for applications requiring automatic addition of derivatizing agents ,
standards, or reference peaks.
ial
1
ial
5
1
ial
ial
ial
6
7
in the sample loop
The Auto Addition table is saved in permanent memory.
Common Vials
You can specify up to five common vials for each Auto Run step. Common vials:
• Can be unique or repeated
• Do not have to be in consecutive positions in the carousel