WATANABE ELECTRIC INDUSTRY CO., LTD AP-244A-11 series, AP-244A-12 series, AP-244A-13 series, AP-244A-14 series Instruction Manual

(1)The application of voltage or current exceeding its maximum
allowable value to the input terminals may result in instrument damage.
(2)The supply of power out of its allowable range may cause fire,
electric shock or instrument failure.
(3)The content of this manual may subject to change without prior
(4)This manual is carefully prepared. However, if any question
arises, or any mistake, omission or suggestion is found in the content of this manual, contact your nearest our sales agent.
1.Before operation 1-1 Checking Model No.
Each of AP-244A Model Nos. is configured as shown in the following,
check that the Model No. ordered matches that delivered.
1.2 Checking accessories The following accessories will come with the AP-244A. Lower 10-pin connector・・・・・1 pc. Instruction manual・・・・・・・1 copy Upper 26-pin connector・・・・・1 pc. (Only with BCD output) If you have any query, please contact our sales agent or Watanabe
directly.
1.3 Mounting
1.3.1 Panel cutout dimensions Make the panel cutout for mounting the AP-244A according to the following Fig.
1.3.2 Mounting the AP-244A When mounting the AP-244A in the panel, push the meter in the panel
from the panel front while pressing down the snaps located at the sides of the case.
1.4 Terminal connection diagram
*Do not gradually apply the power supply voltage to the meter but from 0V quickly to the rated voltage. (If the voltage is gradually increased, the meter may not start.)
2. Various functions
2.1 Setting the position of decimal point
The position of the decimal point of this meter can be externally controlled from the lower connector terminals. Short the terminal corresponding to the digit to be lit with the COM terminal.
2.2 Start/hold function
The start/hold function is used to control display hold
(measurement stop) and start (measurement start) from the external control terminals.
Measurement is held with the START/HOLD terminal shorted with the
COM terminal (or level 0) or started with the START/HOLD terminal opened from the COM terminal (or level 1).
In addition, a positive pulse within one-sampling time exceeding
1 ms enables one-shot measurement.
The function of the START/HOLD terminal wired on the upper side
when the BCD output is optionally provided is the same as that described above.
However in this case, the control line is isolated from the input
signal line.
2.3 BCD output
This meter can provide the parallel BCD output as option. In addition, for the BCD output of this meter, the input circuit is isolated from the power circuit.
2.3.1 Terminal connection diagram
2.3.2 Output status[( );For TTL]
Measured data; Negative logic (Positive logic) Polarity signal; Transistor turned on at plus (Logic 1) OVER signal; Transistor turned on at OVER (Logic 1) Printing command output (PC); Transistor turned on (Logic 1) for about 10 ms every time
measurement is complete. Output capacity; Max. voltage 30V Max. current 15 mA (Fanout 2)
INSTRUCTION MANUAL
DIGITAL PANEL METER
MODEL AP-244A SERIES
Caution
WATANABE ELECTRIC INDUSTRY CO., LTD.
2.3.3 ENABLE
All of the transistors are turned off for the open -collector output
or they are set to the high impendance state for the TTL output with the ENABLE terminal shorted with the D.COM terminal or set to level 0.
2.4 Selecting sampling speed Select a sampling speed of 12.5 times/sec (50Hz) or 15 times/sec
(60Hz) depending on the commercial power frequency used. (If the incorrect frequency is selected, the meter operates as usual but its display may widely fluctuate.)
3. Calibration
In order to calibrate this meter, a standard voltage generator with
an accuracy of ±0.01% or higher is required. In addition, for calibration, see the following Fig.
4.Specifications
Model No./
range codes
Measurem­ent range
Max.
resolution
Input
impedance
Max.allowable Input Voltage
AP-244A-11-
XX-X
±
199.99mV
10μV
100MΩ
±100V
AP-244A-12-
XX-X
±
1.9999V
100μV
100MΩ
±100V
AP-242A-13-
XX-X
±
19.999V
1mV
1MΩ
±250V
AP-242A-14-
XX-X
±
199.99V
10mV
1MΩ
±250V
Accuracy: ±(0.03% of rdg. +1 digit) at 23℃±5
General specifications
Measurement
function
DC voltage measurement
Operation method
Double integral
Input circuit
Single-ended type
Max. display
: 19999
Sampling speed
2.5 times/sec.
12.5 times/sec(50Hz) or 15 times/sec(60Hz)
Noise rejection ratio
NMR, More than 50dB(50/60Hz)
Overrange warning
:”0000 or -0000flashes for an
input signal of more than 19999.
Display
7-segment LED
(Red light emitting diode numeric
element) character height, 8mm
Polarity display
:”-is automatically displayed when
the computation result is minus.
External control
HOLD・・・ Shorting START/HOLD terminal with
COM terminal(or level 0)
START・・・
Opening START/HOLD terminal from COM terminal (or level 1)
Operating temp/ humidity range
: 0 to 50/
35 to 85%RH
Power supply
: 5 to 12V DC ±5%
12 to 24V DC ±10%
Consuming current
At 5V DC・・・
Approx.100mA(BCD not provided) Approx. 220mA(BCD provided) At 12V DC・・・ Approx.50mA(BCD not provided) Approx. 100mA(BCD provided) At 24V DC・・・ Approx.30mA(BCD not provided) Approx. 60mA(BCD provided)
Dielectric strength
Between LO input and 0V power
terminals・・・ For 1 min. at 500V DC
Between LO input and D.COM output
terminals・・・ For 1 min. at 500V DC
Between 0V power and D.COM output terminals・・・
For 1 min. at 500V DC
Between LO input terminal and case・・・
For 1 min. at 1500V AC
Insulation resistance
Between each terminal described
above・・・
More than 100 MΩ at 500V DC
Dimensions
48(W)×24(H)×87(D)mm
Weight
Approx. 50g
Accessories
: Lower 10-pin connector・・・1pc.
Instruction manual・・・・・ 1 copy Upper 26-pin connector・・・1pc. (Only for BCD output)
5. Dimensions
6. Timing chart
7. Warranty and after-sale service
1) Warranty This meter is warranted for a period of one year form date of delivery. Any defect which occurs in this period and is undoubtedly caused by Watanabe's faults will be remedied free of charge. This warranty does not apply to the meter showing abuse or damage which has been altered or repaired by others except as authorized by Watanabe Co.,Ltd.
2) After-sale service
This meter is delivered after being manufactured, tested and
inspected under strict quality control. However, if any problem does occur, contact your nearest Watanabe sales agent or Watanabe directly giving as much information on problem as possibl e.
6-16-19, Jingumae, Shibuya-ku, Tokyo 150-0001, Japan Phone: (81)3-3400-6140
Homepage http://en.watanabe-electric.co.jp/
WATANABE ELECTRIC INDUSTRY CO., LTD.
ASAHI KEIKI DIVISION
6-16-19, Jingumae, Shibuya-ku, Tokyo 150-0001, Japan Phone: (81)3-3400-6141
Homepage http://www.watanabe-electric.co.jp/en/
WATANABE ELECTRIC INDUSTRY CO., LTD.
MODEL AP-244A Series ED-50092c (2/2)
Loading...