Installation & Maintenance Checklist
CHECK OK / NOTES MAINTENANCE ITEM
Delivery Supplied complete no transit damage
Position Level and stable on sound, waterproof, self draining floor
Benchwork secure with smooth rack path in and out
Water Isolator valve fitted accessible, all fittings sound, no leaks
DCVs fitted correct direction
Supply Temperature 65 ± 5°C
Flow rate minimum 15 litres per minute
Pressure 200-350kpa, limiter fitted if above this
range, booster fitted if below this range
Cold water if required pressure not above 350 kpa
Hardness filter or softening if required,
excessive hardness or chlorides can
damage
Power Isolating switch fitted, functional and accessible
Supply as specified voltage, current, circuit breaker
Waste Connections 40 mm sound, no leaks
Waste line 50mm
Ventilation
Appropriate/not excessive
refer separate section in manual,
curtains fitted correctly
Commissioning
Commissioning switch
break seal and turn on (if fitted)
Pump Rotation CONFIRM CORRECT DIRECTION
as shown by arrow on pump
if necessary rotate supply phases
Chemicals – chemicals and chemical injectors to be supplied and adjusted by professional supplier
Product