Walz PHYTO-PAM-II User Manual

Printed in Germany
Phytoplankton Analyzer
PHYTO-PAM-II
and
Phyto-Win 3
Software V 3
Principles of Operation
PhytoPamII_2.doc
Heinz Walz GmbH, 2016
Heinz Walz GmbH Eichenring 6 91090 Effeltrich Germany
Phone +49-(0)9133/7765-0  Telefax +49-(0)9133/5395
Email info@walz.com Internet www.walz.com
MAY 2002 CONTENTS
1Safety instructions ........................................................................ 1
1.1General safety instructions .......................................................... 1
1.2Special safety instructions ........................................................... 2
2PHYTO-PAM-II ........................................................................... 3
2.1Compact Version ......................................................................... 5
Spherical Micro Quantum Sensor US-SQS/WB .................... 82.1.1
Stirring Device WATER-S (optional) .................................... 92.1.2
Battery Charger MINI-PAM/L ............................................ 102.1.3
3PhytoWin Software installation and Phyto-PAM-II setup ..... 11
3.1Measure PAR Lists .................................................................... 13
4First fluorescence measurements ............................................. 16
4.1Principle of distinguishing between different groups of
phytoplankton ............................................................................ 21
5Saturation Pulse Analysis .......................................................... 23
5.1Measurements with Dark-Acclimated Samples ........................ 23
5.2Measurements with Illuminated Samples .................................. 24
5.3Fluorescence Ratio Parameters ................................................. 25
6Features of PhytoWin Software ................................................ 30
6.1Elements for system operation and display of instrument status33
6.2Channels-window ...................................................................... 39
Ft, F, Fm’, Fm’-F, Y(II), Fv/Fm and Mean Value display ... 396.2.1
Zero Offset and noise N(t) ................................................... 406.2.2
6.3Algae-window ........................................................................... 42
6.4Settings-window ........................................................................ 44
6.5Slow Kinetics window .............................................................. 49
6.6Light Curve window .................................................................. 51
Edit Light Curves ................................................................. 536.6.1
Light Curve Fit-parameters .................................................. 566.6.2
CONTENTS
Comments on Light curves .................................................. 626.6.3
6.7Report Window ......................................................................... 64
6.8Reference-window deconvolution and Chlorophyll
determination calibration values ............................................... 68
How to generate Reference Spectra ..................................... 726.8.1
6.9Fluo Spec -window .................................................................... 74
6.10 Fast Kinetik - window .......................................................... 75
O-I
1
Fit window .............................................................. 776.10.1
Parameters and Output of Fo-I
1
Rise Analysis ............... 816.10.2
6.11 VIEW - MODE .................................................................... 82
6.12 Main Menu Bar .................................................................... 85
Options: ETR Parameter ................................................. 876.12.1
Al current/PAR list ......................................................... 886.12.2
6.13 Script files ............................................................................ 90
Data Management ........................................................... 916.13.1
Editing Tools ................................................................... 926.13.2
List of Script File Commands ......................................... 936.13.3
7Technical Specifications ........................................................... 101
PHYTO-PAM-II Compact Version ................................... 1017.1.1
System Control and Data Acquisition ................................ 1027.1.2
Battery Charger MINI-PAM/L .......................................... 1037.1.3
Spherical Micro Quantum Sensor US-SQS/WB ................ 1037.1.4
Transport Box PHYTO-T .................................................. 1047.1.5
Accessory Stirring Device WATER-S ............................... 1047.1.6
8Rechargeable battery ............................................................... 105
9References and further reading .............................................. 106
10Index .......................................................................................... 112
11Warranty .................................................................................. 115
MAY 2002 CONTENTS
11.1Conditions .......................................................................... 115
11.2Instructions ......................................................................... 116
CHAPTER 1MAY 2002 SAFETY INSTRUCTIONS
1
1 Safety instructions
1.1
General safety instructions
1. Read the safety instructions and the operating instructions
first.
2. Pay attention to all the safety warnings.
3. Keep the device away from water or high moisture areas.
4. Keep the device away from dust, sand and dirt.
5. Always ensure there is sufficient ventilation.
6. Do not put the device anywhere near sources of heat.
7. Connect the device only to the power source indicated in the
operating instructions or on the device.
8. Clean the device only according to the manufacturer’s
recommendations.
9. If the device is not in use, remove the mains plug from the
socket.
10. Ensure that no liquids or other foreign bodies can find their
way inside the device.
11. The device should only be repaired by qualified personnel.
CHAPTER 1 SAFETY INSTRUCTIONS
2
1.2 Special safety instructions
1. PHYTO-PAM-II Multiple Excitation Wavelength Phyto­plankton and Photosynthesis Analyzer is a highly sensitive research instrument which should be used only for research purposes, as specified in this manual. Please follow the instructions of this manual in order to avoid potential harm to the user and damage to the instrument.
2. PHYTO-PAM-II employs high intensity LED-array light sources which may cause damage to the eye. Avoid looking directly into these light sources during continuous illumination or saturation pulses.
3. Do not cover the ventilation grille rear side of the instrument.
CHAPTER 2 PHYTO-PAM-II
3
2 PHYTO-PAM-II
PHYTO-PAM-II represents the progress of established WALZ fluorometers dedicated to aquatic research like XE-PAM, WATER- PAM, previous PHYTO-PAM and MULTI-COLOR-PAM. Choice of several measuring light wavelengths at variable measuring light settings (intensity and frequency) offered by XE- and 4-wavelengths PHYTO-PAM are combined with the high time resolution of MULTI-COLOR-PAM. Thus enabling classical PAM analysis, deconvolution of phytoplankton and analysis of fast induction kinetics.
The PHYTO-PAM-II incorporates a multi-color Chip-On-Board (COB) LED Array featuring 5 measuring light colors and 6 colors for actinic light illumination. The 440 nm, 480 nm, 540 nm, 590 nm, 625 nm measuring light facilitates online differentiation of 4 different pigment types therefor deconvolution of green algae, cyanobacteria, diatoms/dinoflagellates and phycoerythrin containing organisms like cryptophytes. As in the first generation PHYTO-PAM deconvolution bases on fluorescence excitation reference spectra. But for the first time these spectra are not instrument specific but universal. Due to spectral calibration of PHYTO-PAM-II instruments reference spectra can be shared between users and instruments.
The PHYTO-PAM-II user surface is based on the proven PhytoWin­software featuring classical PAM analysis like the estimation of the effective photochemical quantum yield of PS II and the determination of photochemical and non-photochemical quenching parameters. For extended applications a special Fast Kinetics mode of operation is added for measuring the wavelength-dependent O-I1 fluorescence rise kinetics upon onset of pulses of strong actinic light information. Thus enabling evaluation of the functional absorption cross-section of PS II, σPSII, determined by light color and the
CHAPTER 2 PHYTO-PAM-II
4
pigment composition of photosynthetic organisms (Klughammer and Schreiber, 2015)
Two versions of the PHYTO-PAM-II Phytoplankton Analyzer are available. A PHYTO-PAM-II Compact Version with integrated emitter-detector unit and a Modular PHYTO-PAM-II Version featuring different emitter-detector units connectable to a Power-and­Control-Unit.
Both versions of the PHYTO-PAM-II are operated by PhytoWin-3 software.
CHAPTER 2 PHYTO-PAM-II
5
2.1 Compact Version
The PHYTO-PAM-II Compact Version is a lightweight and easily portable instrument. The splash-proof cast aluminium housing incorporates the control unit as well as all essentials of the emitter­detector unit: the optical unit, the measuring and actinic LED-array, the photomultiplier-detector and the cuvette retainer.
Fig. 1: PHYTO-PAM Compact Version
Front side:
Power switch and green power indicator lamp
USB socket to connect external PC for instrument control
Charge socket, to connect Battery Charger MINI-PAM/L
(100 to 240V AC)
Fuse plug, containing 3.15 AT main fuse of internal power
circuit
CHAPTER 2 PHYTO-PAM-II
6
Light Sensor socket, to connect the Spherical Micro
Quantum Sensor (US-SQS/WB) for PAR list calibration.
Top :
Measuring Head with optical port for inserting sample
cuvette
PVC centering ring with o-ring sealing against the inner
wall of the Measuring Head, serving as a guide for the cuvette and as an adapter for mounting the optional Miniature Stirring Motor Water-S and the optional Spherical Micro Quantum Sensor US-SQS/WB.
Cup-shaped perspex inset sealing against an inner o-ring of
the Measuring Head, thus protecting the opto-electronical components from spilled water samples.
A green LED shows the status of the internal
Photomultiplier Detector. The photomultiplier automatically is switched off at excessive light impact.
Quartz Cuvette with 15 mm outer and 13 mm inner
diameter; height 46 mm.
Darkening Hood covering the part of the Measuring Head
protruding from the housing.
Photomultiplier Detector The Photomultiplier supply
voltage is automatically switched off when it sees too much light (red status LED on housing lights up).
COB LED-Array The light emitting diodes (LED) are
densely arranged on a 10 x 10 mm area. PHYTO-PAM-II provides 5 differently coloured measuring lights (440 nm, 480 nm, 540 nm, 590 nm, 625 nm), and 5 actinic light
CHAPTER 2 PHYTO-PAM-II
7
sources (440 nm, 480 nm, 540 nm, 590 nm, 625 nm); the latter are complemented by white (440-640 nm) and far red (735 nm) LEDs (for measuring light and actinic light spectra see Fig. 2). To compensate for varying brightness of the different types of light emission diodes, more LEDs are used when the intensity of the individual LED is lower.
Fig. 2: Typical normalized emission spectra of PHYTO-PAM-II light sources. Spectra are not corrected for spectral response of the spectrometer.
CHAPTER 2 PHYTO-PAM-II
8
Spherical Micro Quantum Sensor US-SQS/WB 2.1.1
Fig. 3: Spherical Micro Quantum Sensor US-SQS/WB
The Spherical Micro Quantum Sensor US-SQS/WB is included in the system package as calibration of the PAR-List is essential for correct assessment of all PHYTO-PAM-II functions.
The US-SQS/WB device consists of a submersible spherical quantum sensor, US-SQS/L, and a small amplifier unit which is connected to the light sensor socket. The amplifier is tuned to yield correct PAR readings within the automated PAR-List calibration (see chapter
3.1) and needs to be set to 1x.
The 3.7 mm diffusing sphere of the US-SQS/WB sensor picks up photosynthetically active radiation (PAR) with an angular response error of less than ± 5% from -100° to 100° angle. Hence, the sensor is ideally suited to measure light conditions in the suspension cuvette where reflections and scattering results in randomization of light direction. For stability of calibration, it is important to keep the
CHAPTER 2 PHYTO-PAM-II
9
diffuser clean. If needed the diffuser can be gently cleaned by using a cotton tip applicator moistened with some ethanol.
The position of the light sensor is adjustable and will be determined for PHYTO-PAM-II applications by a 15.3 mm spacer above the sensor hood.
Stirring Device WATER-S (optional) 2.1.2
Fig. 4: Stirring Device WATER-S
WATER-S stirring device can be particularly useful for measurements of rapidly settling samples.
A special adapter ring is provided for mounting the optional Stirring Device WATER-S to the PHYTO-PAM-II Compact version.
WATER-S runs on a long life 3 V Lithium battery (size CR 123A). It features an on/off switch and a potentiometer knob for stirring rate adjustment. In the bottom center, a stirring paddle is mounted on the
CHAPTER 2 PHYTO-PAM-II
10
motor axis (via split brass-tube adapter). The disposable paddle can be removed by gentle pulling. The other way around, a replacement paddle can be mounted by pushing its cylindrical end all the way into the holder. For replacement of the battery the housing has to be opened by pulling the white and grey halves apart. Separation of the two halves is facilitated by forcing gently a thin flat body into the slit (like finger nail or thin screw driver).
It should be noted that at high photomultiplier gain the paddle of the WATER-S will cause some increase of noise. This is due to the fact that some measuring light is reflected from the paddle towards the photodetector, such that the background signal is approximately doubled and the electronic noise is correspondingly increased. Furthermore, there is an increase of sample noise caused by the movement of cells or cell groups.
Battery Charger MINI-PAM/L 2.1.3
The Battery Charger MINI-PAM/L is provided for recharging the internal lead-acid battery (12V/2Ah) of PHYTO-PAM-II. It is connected to the Charge-socket on the front panel of the Power- and-Control-Unit. The charger, which operates at input voltages between 100 and 240V AC, features overload protection. Battery voltage is displayed on the Settings-window of PhytoWin Software.
CHAPTER 3 INSTALLATION
11
3 PhytoWin Software installation and Phyto-PAM-II setup
PhytoWin-3 and manual versions can be downloaded from Walz web page http://www.walz.com/downloads/overview.html. The PhytoWin-3 program needs be installed on the PC going to be used in conjunction with the PHYTO-PAM-II. At the end of the guided installation procedure a Phyto-PAM_3 folder is created on the PC with Data-directories of three different types of Phyto-PAM Measuring Heads (Phyto US, Phyto ED and Phyto Compact Unit). Into these directories all measured data will be written. Next to these directories a script folder, the Phyto.exe file and after definition of the used measuring head, a Phyto.cfg is located within the Phyto­PAM_3 folder.
After download double click the file Setup.exe
After start of Setup.exe the Install Wizard is called up.
This wizard guides through the installation, at the end of which the PhytoPAM-3 folder will be installed on the PC.
Links to PhytoPAM-3 Folder and to PhytoWin-3.exe are established on the desktop.
CHAPTER 3 INSTALLATION
12
Connect PHYTO-PAM-II to computer via USB-cable and start PhytoWin program by clicking the PhytoWin-3 icon on the desktop. The Start-window is displayed, showing the number of the current PhytoWin version.
Following query appears:
to search for Com-Port and enter Measure Mode
to view and analyse data acquired by the PHYTO-PAM-II
and start the program in the viewing mode.
to quit the program.
CHAPTER 3 INSTALLATION
13
In both software operation modes (Measure Mode and Viewing Mode) the selection of
the applied measuring head is essential, as each measuring head features individual parameters (e.g. relating to photomultiplier sensitivity) that are stored in separate Data
directories. These parameters are essential for correct storage and analysis of the measured data. After definition of the measuring head the actual program is started with the 5­channels excitation window being displayed.
NOTE:
When a computer installs the PhytoWin-3 user software first time for a selected PHYTO-PAM-II measuring head, the internal PAR list needs to be measured before bringing into operation. Calibrations of the PAR-Lists and a valid .par file in the measuring head folder is essential for correct assessment of PHYTO-PAM-II functions.
The default.par file is only for startup and not suitable for functional operation.
The features of the PhytoWin-3 user software described in the following sections apply to all measuring heads. Exceptions are indicated.
3.1
Measure PAR Lists
To bring the PHYTO-PAM-II to operation first time the LED array light parameters (PAR lists) need to be calibrated. The light list measurements include calibration of the light sensor offset and
CHAPTER 3 INSTALLATION
14
readout of the complete LED array done in an automated calibration procedure.
Switch on PHYTO-PAM-II and start PhytoWin-3
software. Please plug in the Spherical Micro Quantum Sensor US-SQS/WB to the PHYTO-PAM-II light sensor connector, the light sensor will be recognized and the Measure PAR Lists button in the Settings Window enabled.
At first use the light sensor might show an offset. This offset will be eliminated during light sensor calibration within the Measure Par Lists routine.
Fill distilled water into the quartz glass cuvette and
insert it in the PHYTO-PAM-II measuring head.
Insert the light sensor. The position of the light sensor is
critical for correct detection of PAR values. Therefor the light sensor/darkening hood geometry is fixed by a 15.3 mm spacer and should not be altered.
Open the measure PAR routine by clicking the Measure
PAR Lists button (Settings window) and start Light calibration process.
CHAPTER 3 INSTALLATION
15
The light sensor offset and all LED array parameters will be
calibrated. When the Measure PAR List routine is finished a .par file will be stored within the Data-directories of the related measuring head folder.
Disconnect light sensor. The functionality of PHYTO-
PAM-II is established.
CHAPTER 4 FIRST MEASUREMENTS
16
4 First fluorescence measurements
Fig. 5 Channels-Window
The 5-channels excitation mode is the standard mode of operation of the PHYTO-PAM-II. After start of the program, on the PC monitor screen the "Channels"-window is displayed. This shows the current Chl fluorescence yield, Ft, measured continuously with 5 different excitation wavelengths. In addition, also the mean value of the 5 fluorescence signals is displayed. Normally, after program start the displayed Ft-values are close to zero, as the Gain (photomultiplier voltage) is set to a low setting by default, in order to avoid unintended damage. As indicated by the status of the ML-switch (bottom, left), the measuring light is switched on upon program start. It is applied in LED-pulses, such that its actinic effect is relatively weak. This means that no electrons accumulate at the acceptor side of PSII and, hence, the minimal fluorescence yield, Fo, of a dark­adapted sample is assessed. When the AL-switch is activated also the intensity of the ML-LEDs is increased. This is due to an automatic increase of the frequency of ML-pulses during actinic illumination. In this way, the signal/noise ratio is increased and the fluorescence changes induced by the actinic light are assessed. At the
CHAPTER 4 FIRST MEASUREMENTS
17
same time the ML at high frequency contributes to overall actinic intensity, which is displayed in the PAR-field in units of µmol quanta m
-2s-1
(photosynthetically active radiation). A third type of illumination is triggered by the "Sat-Pulse" button. But, please avoid looking directly into the LED-array source, as this light is very strong and may harm your eyes. This so-called "saturation pulse" can cause complete reduction of the PSII acceptor pool and, hence, induce an increase of fluorescence yield from its current level (Ft) to its maximal value (Fm). Based on such measurements, the effective quantum yield of photosynthetic energy conversion in PSII can be determined, using the simple relationship:
Yield = (Fm-F)/Fm = Fm-Fo/Fm
For some first fluorescence measurements fill the cuvette with a sample (which first may be pure water) and make sure that the Photomultiplier-Detector is switched on (green indicator LED on the right hand side of the elements of system operation bar). Underneath the photomultiplier button is the Gain control box, showing setting 5 of photomultiplier gain upon program start. This gain is by far too low to show any fluorescence signal with a pure water sample. After clicking the Gain-button the Gain-setting is automatically increased until the channel with the largest signal shows about 700 units (Auto-Gain function). Even pure water samples will show a fluorescence signal, if the Gain is sufficiently high (ca. setting 26 by Automatic Gain control). This unavoidable "background signal" is due to stray fluorescence originating from various system components like the LED-array, cuvette and filters. You will find that an empty cuvette gives a much larger background signal than a cuvette filled with pure water. The unavoidable background signal can be digitally suppressed by the automatic
CHAPTER 4 FIRST MEASUREMENTS
18
Zero-offset function (Zoff). But, please note that it will always cause a decrease in the signal/noise ratio.
In practice, natural surface waters often contain (besides phytoplankton) other fluorescing substances (like humic acids) in solution. In order to get rid of this contribution, together with the small background signal caused by system fluorescence, it is recommended to proceed as follows:
Make sure that the cuvette is clean, e.g. by washing with
ethanol and rinsing with water.
Make sure that the cuvette is placed correctly into the
Measuring Head. If the cuvette is not all the way down, this will cause an increased background signal.
Fill the cuvette with ca. 4 ml of the sample to be
investigated and apply Auto-Gain to define the Gain-setting at which the measurements will be carried out.
Prepare a filtrate of the sample using a 0.2 µm millipore
filter that will retain all phytoplankton.
Exchange the cuvette by a cuvette containing ca. 4 ml of the
filtrate and measure its fluorescence using the same Gain­setting as found appropriate for the unfiltered sample. By giving a saturation pulse, you may convince yourself that the signal displayed by the filtrate really is not originating from active Chl. The Fm`-F and Yield-values will be zero or close to zero. Please note the little indicator lamp below the PAR-box, which lights up red as long as the signal is unstable. All measurements, including Zoff-determination, should be preferentially carried out after this lamp lights up green. After Zoff-determination, the signals of the 5 channels are close to zero. Fluctuating values of up to ca. 5 units may occur due to digital noise and are of no concern.
CHAPTER 4 FIRST MEASUREMENTS
19
After Zoff-determination the filtrate is substituted by the sample and now the proper fluorescence measurements can start, as the fluorescence yields displayed for the 5 channels now are only due to the phytoplankton. The most fundamental measurement is the assessment of the quantum yield of photochemical energy conversion in PSII by application of a saturation pulse. With an active sample, the 5 channels will show values of maximal PSII quantum yield (Yield under quasi-dark adapted conditions) in the order of 0.5 - 0.8. You may have a look at the saturation pulse kinetics (View Pulse check-box). For appropriate Yield- determination, it is important that the maximal fluorescence yield is reached during the saturation pulse, which is the case when a distinct plateau is observed
Another fundamental measurement is the recording of the fluorescence changes upon transition from darkness to continuous light. Just switch on the actinic light (AL-button) and follow the changes of fluorescence yield with time, Ft. You will observe that fluorescence yield first rises to a peak level and then slowly declines towards a steady state level. This is the famous Kautsky-effect, which reflects the dark-light induction kinetics of photosynthesis. In the slow kinetic window you can see the graphical displayed of the recorded values.
When a saturation pulse is applied during actinic illumination, the observed Yield-values are distinctly lower than after dark-adaptation. This reflects a decrease in the efficiency of energy transformation at PSII reaction centers due to two major factors: first, partial reaction center closure (primary acceptor Q
A
reduced) and, second, increase
of nonradiative energy dissipation.
Chl fluorescence carries information on the effective quantum yield of PSII under quasi-dark and light conditions. The product of quantum yield and quantum flux density of incident
CHAPTER 4 FIRST MEASUREMENTS
20
photosynthetically active radiation (PAR) provides a relative measure of electron transport rate (ETR). Plots of quantum yield and ETR versus PAR (so-called light response curves) give valuable information on the photosynthetic performance and light saturation characteristics of a sample.
The PhytoWin-program provides a routine for automated recording of light response curves. For a first demonstration, open the "Light Curve"-window and click the "Start"-button. There is an immediate Yield-determination of the sample adapted to the Measuring Light (at the given frequency of Measuring Light pulses). Then light intensity automatically is increased in a first step (see increase in displayed PAR-value) that extends over a defined time period, at the end of which Yield again is determined. Further steps of increased light intensity follow and at the end of each the Yield is determined, thus resulting in light response curves of Yield and of the derived ETR. The PAR-values of the various steps, the illumination time during each step and the total number of steps can be defined by the user (via Edit, see
6.6.1). These "Rapid Light
Curves" provide relevant information as outlined in more detail below (see
6.6).
The fluorescence information obtained will be automatically stored in the Report-file which can be accessed by clicking the corresponding "register card" (Report see
6.7). All data stored in the
Report-file can be also recalled on the Channels- and Algae- windows for further inspection with the help of the VIEW-mode (see
6.11). In order to continue with measurements, the user must
return to the MEASURE-mode.
CHAPTER 4 FIRST MEASUREMENTS
21
4.1 Principle of distinguishing between different
groups of phytoplankton
Well proven by the 1st generation PHYTO-PAM is the deconvolution of different groups of phytoplankton. The reliable assessment of fluorescence parameters using a number of different excitation wavelengths is the basis for this distinction and characterization. PHYTO-PAM-II features now a new spectral composition using five excitation wavelengths that are chosen for optimal differentiation between cyanobacteria, green algae, diatoms/dinoflagellates and phycoerythrin-containing organisms, which differ substantially in the absorbance spectra of their antenna pigments.
An essential prerequisite for the differentiation between various types of phytoplankton is that the 5-channels fluorescence responses of the pure cultures are known. The measurements of such "Reference Excitation Spectra" are automated and, hence, quite simple to be performed with the PHYTO-PAM-II, (see
6.7).
PHYTO-PAM-II normalizes measured reference spectra. Reference spectra measured with PYTHO-PAM II are universal and can be shared among users. More information about reference spectra are given in chapter
6.8.
Based on the "Reference Spectra" the PhytoWin-program deconvolutes the original 5-channels signals into the contributions of the corresponding algal classes. It should be emphasized that contrary to the unbiased fluorescence information displayed in the "Channels"-window, the information on the "Algae"-window is strongly biased by the information contained in the applied References. Hence, the quality of the obtained results depends on previous work invested by the user into the measurement of the References. Such work will profit from background knowledge on the likely presence of particular phytoplankton species in the
CHAPTER 4 FIRST MEASUREMENTS
22
investigated water sample. In this sense, the success of practical applications to a considerable extent depends on close interaction with basic research, not only using Chlorophyll fluorescence, but also alternative methods, like microscopy, flow cytometry and analysis by HPLC.
Deconvolution of the phytoplankton classes can be applied to data obtained by classical PAM analysis as outlined in chapter
5.
CHAPTER 5 SATURATION PULSE ANALYSIS
23
5 Saturation Pulse Analysis
Typically, five different types of fluorescence levels are acquired by Saturation Pulse analyses named Fo, Fm Fo’ Fm’ and F. In most cases, the PAM fluorescence signal is proportionally related the yield for chlorophyll fluorescence. Therefore, differences between these five fluorescence levels reflect variations in chlorophyll fluorescence yields.
Two of these levels (Fo and Fm) need to be determined with a dark­acclimated sample. The three remaining levels (Fo’, F, and Fm’) are repeatedly measured during subsequent sample treatments (e.g., exposure to actinic light; see Fig. 6).
5.1
Measurements with Dark-Acclimated Samples
Fo Minimum fluorescence level excited by very low intensity of
measuring light to keep PS II reaction centers open.
Fm Maximum fluorescence level elicited by a pulse of saturating
light (Saturation Pulse) which closes all PS II reaction centers.
CHAPTER 5 SATURATION PULSE ANALYSIS
24
5.2 Measurements with Illuminated Samples
Fo’ Minimum fluorescence level of illuminated sample which is
lowered with respect to Fo by non-photochemical quenching. When the measuring routine for Fo’ is active, the Fo’ level is determined during a dark interval following the Saturation Pulse. In the dark interval, far­red light is applied to selectively drive PS I and to quickly remove electrons accumulated in the intersystem electron transport chain, thus reopening PS II reaction centers (see Fig. 6, time point 75 s). Alternatively, the Fo’
Fig. 6: Measurements for Saturation Pulse Analysis. AL, Actinic
Light; D, dark; SP, Saturation Pulse; FR, Far-red illumination.
Loading...
+ 92 hidden pages