WalkMed 350VL Quick Reference

Page 1
350VL Patient Quick Reference
This guide is meant as a complement to the Operation Manual. Please refer to the Operation Manual for complete instructions for use.
Display Message
Sounds / Lights Condition Corrective Action
Sounds: Slow beeping
Blinking yellow
Not delivering
Press
key to start delivery.
Press
key to silence alert for 5 minutes.
Sounds: Slow beeping
Blinking yellow
Low battery
Press
key to start delivery.
Replace battery within 30 minutes.
Sounds: Continuous tone
Continuous red
Depleted battery Remove depleted battery to silence alarm.
Replace battery.
Press
key to turh pump on.
Press
key to restart.
Sounds: Continuous tone
Blinking red
Clamp bar is open
Press
key to silence alert.
Close clamp bar.
Press
key to restart.
Sounds: Continuous tone
Blinking red
IV tubing occlusion Check tubing system for kinks and correct.
Pump will resume delivery when line is cleared.
Press
key to silence alert.
Check tubing system for kinks and correct.
Press key to restart.
Continued on following page
Page 2
© 2018 WalkMed
Sounds: Continuous tone
Continuous red
System malfunction Write down Error message from display.
Remove battery to silence alarm. Take pump out of service. Contact support with Error message.
Sounds: Slow beeping
Blinking yellow
Total volume limit reached The pump will continue delivery at the KVO rate for
appromiately 4 hours. An audible alert will beep every 30 minutes.
Sounds: Fast beeping
Blinking red
End of KVO
Press
key to silence alert.
Clear END alert by: Reset volume delivered.
- or - Increase total volume limit amount.
- or - Set total volume limit to OFF.
Failed key in keypad
Press
key to turn pump off.
Take pump out of service.
Sounds: Continuous tone
Continuous red
System malfunction during battery install
Remove battery and wait 15 seconds before re­installing.
www.walkmed.com/resources/wm-350vl-patient-quick-reference/