HSICBB-P Portable Calibration Black Body User Manual
1. Application
The Wahl HSICBB-P is a portable ambient temperature black body used for single point calibration and checking of Infrared Thermal Imaging Cameras and Point Infrared Thermometers. It is housed in a watertight IP67 carrying case.
Case Latches
2. User Controls
Scale switch – front panel switch is used to select between °F and °C scale. Changing the switch position will result in a change of the scale and displayed temperature on the next display update cycle. Power switch – front panel switch is used to power on or off the HSICBB-P.
3. Operation
1. Open enclosure by pulling up on the two side latches.
2. Set scale switch to desired scale.
3. Set power switch to on position, indicated by “I”.
4. Unit will go through self-check and momentarily display all display segments.
5. Unit will momentarily blank the display followed by displaying the blackbody’s current temperature.
6. The unit will update the black body temperature at approximately ten-second intervals.
Scale Switch
Ambient Temperature Target Radiator
Power Switch
2 of 4
Page 3
HSICBB-P Portable Calibration Black Body User Manual
4. Typical Calibration Technique and procedure
Stabilization – It is important that both the IR measuring device (Unit Under Test ­UUT) being calibrated and the HSICBB-P Calibration Black Body are thermally stable. They should be in a stable (± 4°F) ambient temperature for a minimum period of 30 minutes. This will result in a more accurate calibration check.
1) Turn on IR imager or thermometer and HSICBB-P and allow them to stabilize 30 minutes minimum.
2) Make certain that if the UUT has variable emissivity, it should be set to .95.
2) Aim the UUT at the blackbody target at a distance so that it fills the spot size or measurement area of the UUT.
3) Compare and record the Blackbody’s display reading to the UUT reading.
4) Document readings as required.
5. Error Codes
Cbl1, Cbl2, Cbl3 and Cbl4 indicate open cable or open sensor. Low Batt – Icon in bottom right corner of display indicates battery voltage is low. Replace battery with Wahl p/n DSA3060. “-----“ Indicates battery below level for reliable reading. Replace battery. For S/N prior to 37105.xxx, use DSA3062 “C” size battery. For S/N after 37105.xxx, use battery #12234-03, “AA” size battery.
6. Specifications
Operating Range
System Accuracy
Display Resolution
Display Update Rate
Radiator
Meter
Scale
Display
3” diameter concentric rings with integral 4-wire RTD sensor
4-wire RTD thermometer
-40° to 158°F (-40° to 70°C)
± 0.3°F or ± 0.2°C over entire range
°F or °C user selectable
4-digit 1” high LCD
0.1°
10 seconds
Repeatability
Heating Method
Emissivity
Power
0.1° F or C
None, assumes the ambient temperature
0.95 @ ± 0.02
1 – 3.6 volt Lithium Thionyl Chloride Battery, size “AA”
Battery Life
Case/ Case Rating
Size / Weight
Approximately 3 years, with “Low Bat” indicator (turning off unit
when not in use)
Pelican 1150 / IP67 (1 meter submersion for 30 minutes) with case closed
9.12” x 7.56” x 4.32” (23.2 x 19.2 x 11.1 cm) / 2.85 lbs. (1.29 kg)
Included
NIST Traceable Report of Calibration, User Manual
3 of 4
Page 4
HSICBB-P Portable Calibration Black Body User Manual
234 Old Weaverville Road, Asheville, NC 28804
7. Service
For calibration, service or technical support, contact: Wahl Instruments, Inc. 234 Old Weaverville Road Asheville, NC 28804 Ph.: 800-421-2853 (US only)
828.658.3131 Fax: 828.658.0728 Web: www.palmerwahl.com