Table of Contents......................................................................................................................................................3
Limitation of Warranty ........................................................................................................................................4
Restricted Rights Legend.....................................................................................................................................4
Declaration of Conformity........................................................................................................................................5
General Safety Instructions.......................................................................................................................................6
Terms and Symbols..............................................................................................................................................6
Support Resources....................................................................................................................................................8
General Description ...........................................................................................................................................12
Voltage Monitor LEDs ......................................................................................................................................15
+5 V Standby Usage ..........................................................................................................................................23
Fan Speed Switch...............................................................................................................................................24
Rack Slide Installation (20” and 24”).....................................................................................................................28
Rackmount Door Kit Installation............................................................................................................................30
Acrylic Rackmount Door Kit Installation...............................................................................................................32
Installation of VXI Modules...................................................................................................................................34
Disconnecting the Mainframe.................................................................................................................................35
Service Information................................................................................................................................................37
Replicable Parts List ..........................................................................................................................................37
CT-100C Power Supply Replacement...............................................................................................................38
Cleaning the Mainframe.....................................................................................................................................40
INDEX ......................................................................................................................................................................... 41
CT-100C Preface 3
Page 4
VXI Technology, Inc.
CERTIFICATION
VXI Technology, Inc. (VTI) certifies that this product met its published specifications at the time of shipment from
the factory. VTI further certifies that its calibration measurements are traceable to the United States National
Institute of Standards and Technology (formerly National Bureau of Standards), to the extent allowed by that
organization’s calibration facility, and to the calibration facilities of other International Standards Organization
members.
WARRANTY
The product referred to herein is warranted against defects in material and workmanship for a period of three years
from the receipt date of the product at customer’s facility. The same warranty applies to the power supply for a
period of one year. The sole and exclusive remedy for breach of any warranty concerning these goods shall be
repair or replacement of defective parts, or a refund of the purchase price, to be determined at the option of VTI.
For warranty service or repair, this product must be returned to a VXI Technology authorized service center. The
product shall be shipped prepaid to VTI and VTI shall prepay all returns of the product to the buyer. However, the
buyer shall pay all shipping charges, duties, and taxes for products returned to VTI from another country.
VTI warrants that its software and firmware designated by VTI for use with a product will execute its programming
when properly installed on that product. VTI does not however warrant that the operation of the product, or
software, or firmware will be uninterrupted or error free.
LIMITATION OF WARRANTY
The warranty shall not apply to defects resulting from improper or inadequate maintenance by the buyer, buyersupplied products or interfacing, unauthorized modification or misuse, operation outside the environmental
specifications for the product, or improper site preparation or maintenance.
VXI Technology, Inc. shall not be liable for injury to property other than the goods themselves. Other than the
limited warranty stated above, VXI Technology, Inc. makes no other warranties, express or implied, with respect to
the quality of product beyond the description of the goods on the face of the contract. VTI specifically disclaims the
implied warranties of merchantability and fitness for a particular purpose.
RESTRICTED RIGHTS LEGEND
Use, duplication, or disclosure by the Government is subject to restrictions as set forth in subdivision (b)(3)(ii) of
the Rights in Technical Data and Computer Software clause in DFARS 252.227-7013.
VXI Technology, Inc.
2031 Main Street
Irvine, CA 92614-6509 U.S.A.
4 CT-100C Preface
Page 5
www.vxitech.com
D ECLARATION OF C ONFORMITY
Declaration of Conformity According to ISO/IEC Guide 22 and EN 45014
ANUFACTURER’S NAMEVXI Technology, Inc.
M
ANUFACTURER’S ADDRESS2031 Main Street
M
Irvine, California 92614-6509
RODUCT NAMESix-Slot VXIbus Chassis
P
ODEL NUMBER(S) CT-100C
M
RODUCT OPTIONSAll
P
RODUCT CONFIGURATIONSAll
P
VXI Technology, Inc. declares that the aforementioned product conforms to the requirements of
the Low Voltage Directive 73/23/EEC and the EMC Directive 89/366/EEC (inclusive 93/68/EEC)
and carries the “CE” mark accordingly. The product has been designed and manufactured
according to the following specifications:
AFETYEN61010 (2001)
S
EMC EN61326 (1997 w/A1:98) Class A
CISPR 22 (1997) Class A
VCCI (April 2000) Class A
ICES-003 Class A (ANSI C63.4 1992)
AS/NZS 3548 (w/A1 & A2:97) Class A
FCC Part 15 Subpart B Class A
EN 61010-1:2001
I hereby declare that the aforementioned product has been designed to be in compliance with the relevant sections
of the specifications listed above as well as complying with all essential requirements of the Low Voltage Directive.
November 2007
Steve Mauga, QA Manager
CT-100C Preface 5
Page 6
Review the following safety precautions to avoid bodily injury and/or damage to the product.
These precautions must be observed during all phases of operation or service of this product.
Failure to comply with these precautions, or with specific warnings elsewhere in this manual,
violates safety standards of design, manufacture, and intended use of the product.
Service should only be performed by qualified personnel.
TERMS AND SYMBOLS
These terms may appear in this manual:
WARNING
CAUTION
These symbols may appear on the product:
VXI Technology, Inc.
GENERAL SAFETY INSTRUCTIONS
Indicates that a procedure or condition may cause bodily injury or death.
Indicates that a procedure or condition could possibly cause damage to
equipment or loss of data.
ATTENTION - Important safety instructions
WARNINGS
Frame or chassis ground
Indicates that the product was manufactured after August 13, 2005. This mark is
placed in accordance with EN 50419, Marking of electrical and electronic equipment in accordance with Article 11(2) of Directive 2002/96/EC (WEEE).
End-of-life product can be returned to VTI by obtaining an RMA number. Fees
for take-back and recycling will apply if not prohibited by national law.
Follow these precautions to avoid injury or damage to the product:
Use Proper Power Cord
Use Proper Power Source
To avoid hazard, only use the power cord specified for this product.
To avoid electrical overload, electric shock, or fire hazard, do not
use a power source that applies other than the specified voltage.
6 CT-100C Preface
Page 7
www.vxitech.com
WARNINGS (CONT.)
Avoid Electric Shock
Ground the Product
Operating Conditions
Improper Use
To avoid electric shock or fire hazard, do not operate this product
with the covers removed. Do not connect or disconnect any cable,
probes, test leads, etc. while they are connected to a voltage source.
Remove all power and unplug unit before performing any service.
Service should only be performed by qualified personnel.
This product is grounded through the grounding conductor of the
power cord. To avoid electric shock, the grounding conductor must
be connected to earth ground.
To avoid injury, electric shock or fire hazard:
- Do not operate in wet or damp conditions.
- Do not operate in an explosive atmosphere.
- Operate or store only in specified temperature range.
- Provide proper clearance for product ventilation to prevent
overheating.
- DO NOT operate if any damage to this product is suspected.
Product should be inspected or serviced only by qualified
personnel.
The operator of this instrument is advised that if equipment is
used in a manner not specified in this manual, the protection
provided by this equipment be may be impaired.
CT-100C Preface 7
Page 8
VXI Technology, Inc.
SUPPORT RESOURCES
Support resources for this product are available on the Internet and at VXI Technology customer
support centers.
VXI Technology
World Headquarters
VXI Technology, Inc.
2031 Main Street
Irvine, CA 92614-6509
Phone: (949) 955-1894
Fax: (949) 955-3041
VXI Technology
Cleveland Instrument Division
VXI Technology, Inc.
7525 Granger Road, Unit 7
Valley View, OH 44125
Phone: (216) 447-8950
Fax: (216) 447-8951
VXI Technology
Lake Stevens Instrument Division
VXI Technology, Inc.
1924 - 203 Bickford
Snohomish, WA 98290
Phone: (425) 212-2285
Fax: (425) 212-2289
Technical Support
Phone: (949) 955-1894
Fax: (949) 955-3041
E-mail:
Visit
support@vxitech.com
http://www.vxitech.com for worldwide support sites and service plan information.
8 CT-100C Preface
Page 9
www.vxitech.com
CT-100C Preface 9
Page 10
Page 11
www.vxitech.com
SECTION 1
INTRODUCTION
INTRODUCTION
The CT-100C portable C-size VXIbus mainframe provides cost-effective test situations in a small
footprint. When using VMIP™ instruments such as DMMs, waveform generators, digitizers, etc.,
complete test scenarios can easily be configured.
FIGURE 1-1:CT-100CSIX SLOT CHASSIS
CT-100C Introduction 11
Page 12
GENERAL DESCRIPTION
The CT-100C chassis is a portable, C-size, six-slot, VXIbus compatible chassis that conforms
fully to VXIbus Specification Revision 2.0. The chassis employs a multi-layer backplane to
ensure premium VXIbus and VMEbus performance and provides all power supplies required by
the VXIbus specification.
The CT-100C supports conventional existing rack designs through an optional rackmounting kit
Section 1). The six-slot design minimizes the use of precious rack space and is an economical
(see
alternative to a larger chassis when fewer slots are required.
The CT-100C chassis contains six slots in the card cage, five of which are available for use by
VXIbus compatible instruments. The sixth slot in the card cage (slot 0) is typically dedicated to
the VXIbus Resource Manager.
The CT-100C is designed to operate at line frequencies between 47 Hz and 63 Hz and utilizes a
power supply that automatically adjusts to accept line voltages between 100 V ac and 240 V ac.
There is an internal fuse that provides protection against catastrophic failures and is designed to
operate within the valid input range.
The rear panel provides a connector for 5V STANDBY. Power supplied to this connector is
passed directly to the backplane line +5V STDBY. This allows properly configured systems to
take advantage of an alternate power supply source, i.e., battery backup of memory or energizing
high stability reference oscillators.
BACKPLANE
VXI Technology, Inc.
PERFORMANCE
The backplane is a monolithic, multi-layer design, with automatic, solid-state daisy-chain
jumpering for the interrupt acknowledge and VMEbus grant lines. This eliminates the need for
manual jumpering and provides improved reliability over mechanical jumper-less backplane
designs. Instrument modules can now be added or removed without concern for the backplane
configuration.
The CT-100C uses a pressurized airflow system. As air enters the mainframe from the rear, it is
pressurized below the cards and is then evenly distributed across all slots and along the total
length of each card slot, avoiding hot spots common in other designs. The air exhausts through the
top and away from the user. This cooling approach helps increase MTBF (Mean Time Before
Failure) figures and module performance.
High-quality power supplies are used in the CT-100C that are UL, CSA, and TUV approved. The
power supplies are short-circuit, over-voltage, reverse-voltage and thermal-shutdown protected.
Auto-ranging power supplies are used to avoid any concern about the voltage source used. In
addition, all supply lines are monitored and displayed on the front panel to provide user feedback
of correct operation (see
Figure 1-4).
12 CT-100C Introduction
Page 13
www.vxitech.com
FLEXIBILITY
The CT-100C is designed to provide flexibility of use in bench-top and rackmount applications, as
well as in portable environments. The outside cover is removable for easy access to the VXIbus
modules during bench-top development, troubleshooting, or calibration. For rackmount
applications, a rackmount kit allows the CT-100C to mount flush or be recessed four inches. A
latched door is provided with the rackmount kit, which can be user-modified to accept connectors,
switches, or indicators. See
Section 1 for rackmount installation.
FIGURE 1-2:CT-100CSIX-SLOT CHASSIS –COVERS OFF
CT-100C Introduction 13
Page 14
FRONT PANEL FEATURES
POWER SWITCH
Feature 1
When elevated, the mainframe is in standby mode, where power is supplied to the
mainframe, but not to the VXI modules.
When depressed, power is supplied to both mainframe and VXI modules
VOLTAGE INDICATORS
Feature 2
Indicates whether specific backplane voltages are within specifications. See
for details.
FAN INDICATOR
Feature 3
Indicates whether the fan voltages are within specifications. See
REAR PANEL FEATURES
FAN SPEED SWITCH
Feature 4
Enables the user to vary the speed of the fan between HIGH and LOW. See
for details.
Figure 1-5 and Table 1-1 for connector pin assignments.
3
Voltag e Monitors Power
+5V +12V +24V +5V Stby
-5.2V -12V -24V -2V FAN
0 1 2 3 4 5
CT-100C
VXI Mainframe
FIGURE 1-3:MAINFRAME FEATURE LOCATIONS
VXI T ec hnology
WARNING - TO PREVENT POSSIBLE ELECTRIC SHOCK HAZARD, DISCONNECT POWER
!
CORD BEFORE REMOV ING THE POWER SUPPLY MODULE FROM THE MAINF RAME.
ATTENTION - PO UR EMP Ê CH ER LE RISQUE POSSIBLE DE D É CHARGE É LECTRIQUE,
DEBRANCHER LE CORDON DE SECTEUR AVANT D'ENLEVER LE MODULE
D'ALIMENTATI ON D' É NERGIE DE L'UNIT É CENTRALE.
WARNING - FOR PROTECTION FROM ELECTRIC SHOCK, THE GROUNDING CONNECTION
!
IN THE POWER CORD MUST BE PROPERLY CONNECTED.
ATTENTION - POUR LA PROTE CTION CONTRE L E CHOC É LECTRIQUE , LA PRISE DE
TERRE AU SOL DANS LE CORDON DE SECTEUR DOIT Ê TRE CORRECTMENT RELI É E.
WARNING - NO OPERATOR SERVICEABLE PARTS INSIDE. REFER SERVICING TO
!
TRAINED SERVICE PERSONNEL.
ATTENTION - AUCUNES PI È CES REPARABLES PAR UN OP É RATEUR À L'INT É RIEUR.
R É F É REZ L'ENTRETIEN AU PERSONNEL QUALIFI É.
FAN SPEED
J200
100-240 VAC
10A MAX.
47 - 63 Hz
HIGH / LO W
4
J201
MONITOR / CONTROL
5
6
7
14 CT-100C Introduction
Page 15
www.vxitech.com
V
VOLTAGE MONITOR LEDS
The power supply lines are monitored and displayed on the front panel, providing information
pertaining to the chassis operational status.
Voltage Monitor LEDs
Green : Within Voltage Specifications
Not Lit
Red
: Under Voltage Specifications
: Over Voltage Specifications
oltage MonitorsPower
+5V +12V+24V+5VStby
-5.2V -12V-24V-2VFAN
FAN LEDs
Green : Within Voltage Specifications
Not Lit
Red
*Note: +5 V Stby is lit only when the +5 V is supplied by the
user t o the + 5 V Stby pins (p in s 8 & 12) of the monitor
connector located on the rear panel.
: Under Voltage Specifications
: Over Voltage Specifications
0 1 2 3 4 5
CT-100C
VXI Mainframe
NOTE At power up, the voltage monitor LED may blink for the first several seconds if the fan speed is
set to “HIGH”.
FIGURE 1-4:FRONT PANEL VOLTAGE MONITORS
CT-100C Introduction 15
Page 16
TABLE 1-1:J201CONNECTOR PIN ASSIGNMENTS
Pin Number Description
1 +5 V Monitor†
2 -12 V Monitor†
3 -24 V Monitor†
4 -2 V Monitor†
5 Remote Power Switch
6 +5 V Output‡
7 +12 V Output‡
8 +5 V Standby Input
9 Ground
10 Backplane Reset I/O
11 N/C
12 N/C
13 Fan OK Output
14 +12 V Monitor†
15 +24 V Monitor†
16 -5.2 V Monitor†
17 Ground
18 Remote Power Switch Return
19 Ground
20 Ground
21 +5 V Standby
22 Ground
23 AC Fail I/O
24 Ground
25 N/C
NOTE Monitor lines function as outputs only.
Both jackscrews connect to ground.
‡
Pin 6 and Pin 7 (+5 V Output and +12 V Output, respectively) can provide 0.5 A from the
power supply.
VXI Technology, Inc.
Pin 13
Pin 25
Pin 1
Pin 14
FIGURE 1-5:J201CONNECTOR PIN DETAIL
16 CT-100C Introduction
Page 17
www.vxitech.com
CT-100CSPECIFICATIONS
GENERAL SPECIFICATIONS
SIZE
WEIGHT
VXIBUS VERSION
MTBF
MTTR
6.96" (176.78 mm) W x 15.00" (381.00 mm) H x 21.3" D (541.02 mm)
Six C-size VXIbus card slots (see
22 lb / 10 kg
2.0
100,000 hr
5 min
ENVIRONMENTAL SPECIFICATIONS
OPERATING LOCATION
TEMPERATURE
Operating
Storage
HUMIDITY
Operating
Non-operating
ALTITUDE
Operating
Non-operating
RANDOM VIBRATION*
Operating
Non-operating
* Three axis, 30 min total, 10 min per axis.
FUNCTIONAL SHOCK
Operating
This chassis should be operated indoors in a controlled environment, protected
from exposure to the elements (i.e. direct sunlight, precipitation, wind, etc.).
Pollution degree 2. Installation Category II.
0 °C to +55 °C
-40 °C to +70 °C
Up to 95% (non-condensing) at up to 30 °C; up to 45% at up to 55 °C
Up to 95% (non-condensing) at up to 55 °C
15,000 ft (4,570 m)
40,000 ft (12,190 m)
0.27 g-rms total from 5 Hz to 55 Hz
2.28 g-rms total from 5 Hz to 55 Hz
Half sine, 30 g, 11 ms duration
Meets functional shock requirements of MIL-T-28800E, Type III, Class 5
Figure 1-6 for details)
CT-100C Introduction 17
Page 18
POWER SPECIFICATIONS
USEABLE POWER
DCSUPPLY VOLTAGE
POWER INPUT
Input Voltage / Freq.
Nominal AC
Inrush Current
Input Power
Input Leakage
Input Harmonics
Fuse
POWER SUPPLIES
500 W to 50 °C, derated by 2.5%/ °C above 50 °C
Peak
Current
Voltage
+5 V 40 A 5 A +0.25 V / -0.125 V 50 mV 50 mV
-5.2 V 10 A 5 A -0.26 V / +0.125 V 50 mV 50 mV
-2 V 8 A 2 A -0.10 V / +0.72 V 50 mV 50 mV
+12 V 8 A 2 A +0.60 V / -0.36 V 50 mV 50 mV
-12 V 4 A 2 A -0.60 V / +0.36 V 50 mV 50 mV
+24 V 4 A 2 A +1.20 V / -0.72 V 150 mV 150 mV
-24 V 4 A 2 A -1.20 V / +0.74 V 150 mV 150 mV
(IMP)
Dynamic
Current
(IMD) Allowed Variation
Minimum 100 V ac to a maximum 240 V ac, 50 Hz/60 Hz
< 40 A (cold start)
10 A max
< 1.24 MA @ 264 V ac, 53 Hz
Meets EN61000-3-2
Internal and independent of line voltage (not user accessible)
UL, CSA, TUV approved, CE marked
Shout circuit, over-voltage, reverse voltage and thermal shutdown protection
COOLING SPECIFICATIONS
COOLING REQUIREMENTS*
Low Fan Speed
High Fan Speed
COOLING MODES
AIR FLOW PATH
ACOUSTIC NOISE
High Speed Fan Mode < 40 dBA
56 W/slot for a 10°C Rise
84 W/slot for a 15°C Rise
80 W/slot for a 10°C Rise
120 W/slot for a 15°C Rise
*Calculated using VXI-8 Rev. 2.0 standards
High or low speed cooling modes can be selected by moving the Fan Speed
switch at the rear of the chassis. The power supply and modules are cooled by
separate fans.
Air is drawn into the chassis from the rear and is pressurized below the cards.
The air is then distributed across all slots along the total length of each slot and
is exhausted through the top of the mainframe. When the mainframe is rack
mounted, allow approximately 2 inches (50 mm) of clearance at the top and rear
for proper airflow.
Low Speed Fan Mode < 30 dBA
VXI Technology, Inc.
Induced
Ripple/Noise
DC Load
Ripple
Noise
18 CT-100C Introduction
Page 19
www.vxitech.com
A
3.0
2.5
O
2
2.0
1.5
1.0
0.5
PressureDrop - mm H
F
a
n
a
t
H
i
g
h
S
p
e
e
d
F
a
n
a
t
L
o
w
S
p
e
e
d
0.0
02.04.06.08.010.0
* Covers all slots, unrestricted
irflow - liters/s
IGURE 1-6:COOLING CAPACITY FOR CT-100C
F
CT-100C Introduction 19
Page 20
21.000 (533.40)
V
8.100 (205.74)
14.979 (380.47)
16.425 (417.20)
VXI Technology, Inc.
oltage Monitors Power
+5V +12V +24V +5V Stby
-5.2V -12V -24V -2V FAN
Note: Dimensions in parentheses are in millimeters.
0.343 (8.71)
F
IGURE 1-7:CT-100CDIMENSIONAL DIAGRAM
0 1 2 3 4 5
CT-100C
VXI Mainframe
VXI Technology
WARNING - TO PREVENT POSSIBL E ELECTRIC SHOC K HAZARD, DISCONNEC T POWER
!
CORD BEFORE REMOVI NG THE PO WER SUPPLY MODULE FR OM THE MAI NFRAME.
ATTENTION - POUR EMP Ê CH ER LE RISQUE POSSIBLE DE D É CHARGE É LECTRIQUE,
DEBRANCHER LE CORDON DE SECTEUR AVANT D'ENLEVER LE MODULE
D'ALIMENTA TION D' É NE RGIE DE L'UNIT É CENTRALE.
WARNING - FOR PROTECT ION FROM ELECTRIC SHOCK, THE GROU NDING CONNECTION
IN THE POWER CORD MUST BE PROPERLY CONNECTED.
!
ATTENTION - P OUR LA PROTECTION CONTRE LE CHOC É LECTRIQUE , LA PRISE DE
TERRE AU SOL DANS LE CORDON DE SECTEUR DOIT Ê T RE CORRE CTMENT RE LI É E.
WARNING - NO OPERATOR SERVICEABLE PARTS INSIDE. REFER SERVICING TO
!
TRAINED SERVICE PERSONNEL.
ATTENTIO N - AUCUNES P I È CES REPARABLES PAR UN OP É RATEUR À L'INT É R IEUR.
R É F É REZ L'ENTRETIEN AU PERSONNE L QUALIFI É.
FAN SPEED
J200
HIGH / LOW
J201
100-240 VAC
10A MAX.
47 - 63 Hz
MONITOR / CONT ROL
20 CT-100C Introduction
Page 21
www.vxitech.com
SECTION 2
INSTALLATION
INTRODUCTION
When the CT-100C is unpacked from its shipping carton, the contents should include the
following items:
All components should be immediately inspected for damage upon receipt of the unit.
BACKPLANE
The CT-100C mainframe has a jumper-less, auto-configurable backplane using active-automatic
daisy chaining for the VME Interrupt acknowledge and bus grant daisy chain signal lines. This
eliminates the need to manually configure the backplane and insures that these signals are
properly configured at all times.
The power cord is the only way to disconnect the CT-100C mainframe from ac power.
Therefore, the power cord must be accessible to the operator at all times. When the CT100C mainframe is mounted in a system rack, the power cord need not be accessible since
the rack must have its own disconnect device.
Le cordon de secteur est la seule manière de démonter l’unité centrale de CT-100C du
courant alternatif Par conséquent, le cordon de secteur doit être accessible à l’opérateur à
tout moment. Quand l’unité centrale de CT-100C est montée dans un support de système, le
cordon de secteur n’a pas besoin d’être accessible puisque le support doit avoir son propre
dispositif de débranchement.
CT-100C Installation 21
Page 22
REMOTE POWER-ON OPTION
If the CT-100C mainframe is to be installed in a remote location, it is possible to apply and
remove power from the unit via the remote power pins located on connector J201. To utilize the
remote power feature, set the front panel switch to the standby position. Remove the power cord,
then remove the power supply from the chassis and place the jumper on the power interface board
to the 2 – 3 position (see
board will be located inside the mainframe at the top right. Doing this will disable the fron t panel
power switch. Pins 5 and 18 can now be used to remotely control the powering of the chassis. The
circuit is diagramed below:
VXI Technology, Inc.
Figure 2-1 for power supply removal instructions). The power interface
Jumper Location
Power Interface Board
FIGURE 1-1:REMOTE POWER-ON SWITCH WIRING
Front
Panel
Switch
Power
Supply
3
12
Power Interface Board
Pin 13Pin 1
Pin 25
Pin 5
Pin 18
Pin 5Pin 18
Pin 14
22 CT-100C Installation
Page 23
www.vxitech.com
8
+5VSTANDBY USAGE
To prevent timers and other circuits from loosing power when the CT-100C is in standby mode,
+5 V Standby pins are made available on connector J201. An external power source can be wired
to these pins with a maximum allowed current of 1 A total across Pins 8 and 21, as these pins are
wired in parallel.
Pin 13Pin 1
Pin 9
Pin
External
+5 V Standby
Power Source
Pin 25
Pin 22
Pin 21
Pin 14
Pin 8 (+5 V Standby)
Pin 21 (+5 V Standby)
Pin 9 (+5 V Standby Return)
Pin 22 (+5 V Standby Retur n)
FIGURE 1-2:+5VSTANDBY WIRING DIAGRAM
CT-100C Installation 23
Page 24
FAN SPEED SWITCH
A fan speed selector switch is located at the rear of the CT-100C chassis. To set the fans to
operate at HIGH speed, move the switch at the rear of the chassis to the left. At HIGH speed, the
chassis is provided maximum cooling to the instruments.
The fans will operate in the LOW speed mode when the fan speed switch is toggled to the right.
The low speed mode provides for quieter operation, but d ecreases the cooling capability of the
fans. To best determine which mode of operation is best in a given application, refer to
on page
installed in the chassis receive adequate cooling in the low speed mode, then this mode may be
used.
VXI Technology
WARNING - TO PREVENT POSSIBLE ELECTRIC SHOCK HAZARD, DISCONNECT POWER
!
CORD BEFORE REMOVING TH E POWER SUPPLY MODUL E FROM THE MAINFRAM E.
ATTENTION - POUR EMP Ê CHER LE RISQUE POSSIBLE DE D É CHARGE É LECTRIQU E,
DEBRANCHER LE CORDON DE SECTEUR AVANT D'ENLEVER LE MODULE
D'ALIME NTATION D' É NERGIE DE L'UNIT É CENTRALE.
WARNING - FOR PROTECTION FROM ELECTRIC SHOCK, THE GROUNDING CONNECTION
!
IN THE PO WER CORD MUST BE P ROPERLY CONNECTED.
ATTENTIO N - P OUR LA PROTECTIO N CONTRE LE CHOC É LECTRIQUE , LA PRISE DE
TERRE AU SO L DANS LE CORDON DE SECTEUR DOIT Ê TRE CORRECTMENT RELI É E.
WARNING - NO OPERATOR SERVICEABLE PARTS INSIDE. REFER SERVICING TO
!
TRAINED SERVICE PERSONN EL.
ATTENTIO N - AU CUNES PI È CES REPARABLES PAR UN OP É RATEUR À L'INT É RIEUR.
R É F É REZ L 'ENTRETIEN A U PERSONNEL QUALIFI É.
VXI Technology, Inc.
Figure 1-6
19 that delineates the cooling requirements of populated mainframes. If all instruments
J200
100-240 VAC
10A MAX.
47 - 63 Hz
FAN SPEED
HIGH / L OW
HIGH / LOW
J201
MONITOR / CONTROL
100-240 VAC
10A MAX.
47 - 63 Hz
Fan Speed Switch Loc ation
(Rear Panel of CT-100C Mainframe )
FIGURE 1-3:FAN SPEED SWITCH LOCATION
24 CT-100C Installation
Page 25
www.vxitech.com
RACKMOUNT OPTION INSTALLATION
OVERVIEW
This section contains the procedures for installing a CT-100C chassis in a standard 19” relay rack.
The available rackmounting options are:
The rackmount ear kit provides the basic hardware necessary to rackmount the CT-100C chassis.
Because the chassis is only 15” wide, standard rack support rails cannot provide mechanical
support to the chassis. If the equipment mounted below the CT-100C cannot provide mechanical
support to the chassis, either 20” or 24” slides will be required to support the chassis in the rack.
The rackmounting ears are designed to allow the chassis to be flush mounted in the rack or
recessed 4”.
The rackmount door kit provides the necessary hardware to install either a 1/8” thick aluminum
door or a 1/8” thick acrylic door in front of the rackmounted chassis. These options require that
the rackmount ear kit also be installed and configured for recessed mounting. The door may be
customer modified to hold connectors, controls, indicators, and similar components.
The 20” and 24” slide kits provide the ability to easily remove the chassis from the rack for
servicing and provides mechanical support for the chassis when installed in the rack. The 20”
slide kit is used when the chassis is flush mounted in the rack. If the chassis is to be recess
mounted, then the 24” slides are required.
CT-100C Installation 25
Page 26
RACKMOUNT KIT INSTALLATION
The rackmount kit (Option 59) provides the basic hardware necessary to rack mount the CT-100C
chassis. With the chassis being only 15” wide, standard rack support rails cannot provide
mechanical support to the chassis. If the equipment mounted below the CT-100C cannot provide
mechanical support to the chassis, either 20” or 24” rack slides will be required to support the
chassis when installed into the rack. The rackmounting ears are designed to allow the chassis to be
flush mounted in the rack or recessed 4”.
1. Remove the four (4) black plastic feet from the bottom of the chassis.
2. Lay the chassis on a protected work surface on its long side with the voltage monitor LEDs of
the chassis facing front with the power switch toward the top.
3. Locate and remove four black plastic feet on the side of the chassis and four black plastic feet
on the bottom of the chassis. Retain the feet if restorin g the ch assis to its orig inal po rtable use
is anticipated.
4. Find two threaded holes on each side of the chassis (the plastic feet were attached to two of
these holes) towards the front of the unit.
5. Line up the rackmount ears with the threaded holes selecting the flush or recessed position as
desired. Refer to the rackmount ear
6. Secure the rack ears using the supplied #8-32 hardware.
Before installing the chassis into an EIA switch rack, the chassis handle should be removed. This
can be done by doing the following:
1. Remove the four screws located near the handle.
2. Remove the two black decorative covers at each end of the carrying handle using the flat
blade screwdriver. Place a piece of paper under the screwdriver blade to prevent scratching
the chassis cover.
3. Remove the four Phillips screws holding the handle in place and then remove the handle
itself. Retain the carrying handle components if there may be some need to restore the chassis
to its original portable use in the future.
Figure 1-4 for visual assistance.
26 CT-100C Installation
Page 27
www.vxitech.com
Black Oxide
Pan Head
Phillips Screws (x4)
* Recessed Rackmount Installation Shown
IGURE 1-4:RACKMOUNT EAR INSTALLATION DIAGRAM
F
Rack Mount Ears (x2)
Flush Rac kmount Co nfiguration
CT-100C Installation 27
Page 28
RACK SLIDE INSTALLATION (20” AND 24”)
The 20” and 24” slide hardware kits (Options 63 and 64, respectively) provide standard flanges to
mount the slides into standard EIA relay racks. It should be noted that the slide flanges in the front
should be installed behind the rack’s front panel mounting rails. This will allow the rackmounting
ears to sit flush with the front of the rack. It should also be noted that although the hardware
provided in this kit will provide all the necessary components to successfully install the CT-100C
into a rack, there are many variations in how EIA relay racks are designed and additional adapter
hardware may be required to install the chassis. Please refer to the relay rack manufacture’s
catalog for additional options.
This procedure provides instructions for installing the 20” or 24” slide kits. Note that the 20” and
24” slide kits are identical except for th e length of the slide units. A 2 0” slide kit is used when the
chassis is flush mounted and a 24” slide kit is used when the chassis is recess mounted.
Required Tools
1. #2 Phillips Screwdriver
2. 1/8" Flat Blade Screwdriver
Parts List
VXI Technology, Inc.
QTY ITEM VTI P/N
2 Slide, Rack Mount, Steel, 20” 37-0054-020
- or -
2 Slide, Rack Mount, Steel, 24” 37-0054-024
1 Hardware Kit, Rack Mount Slide, Steel 37-0055-000
6 Screw, 8-32 x 3/8” Pan Head Phillips, Sems Zinc 37-0073-037
6 Screw, 8-32 x 1/4” Pan Head Phillips, Steel/Zinc 37-0074-025
4 Screw, 8-32 x 3/8”, F/H Undercut Phillips, Zinc 37-0115-037
2 Bracket, Slide, Front 41-0108-000
1 Bracket, Slide, Top Side 41-0131-000
1 Bracket, Slide, Bottom 41-0132-000
Assembly Instructions
1. Lay the chassis on a protected work surface on its long side with the voltage monitor LEDs of
the chassis facing front with the power switch toward the top.
2. If the rackmount ears were installed flush, remove the rackmount ears and install the front
slide bracket to them as shown in the assembly drawing using a flat head screw driver. If the
rackmount ears were installed recessed, simply install the slide brackets onto the ears as
shown on the following page.
3. Reinstall the rackmount ears (if necessary).
4. Install the top and bottom slide brackets using 1/4” Phillips pan head screws. Note that the
slide mounting holes should be oriented below the centerline of the chassis.
28 CT-100C Installation
Page 29
www.vxitech.com
Slide Bracket, Bottom
5. Locate the slide hardware kit provided. Install the rack flanges to the slides as required by the
application using the hardware in the kit. Note that the screws are in serted from the inside of
the slides and the nuts would show on the outside.
6. Install the slides to the mounting brackets already installed on the chassis using the provided
hardware.
Slide Bracket, T op
1/4” Pan Head
Phillips Screw (x3)
1/4” Pan Head
Phillips Screw (x3)
Rack Mount Slide (20” or 24”)
Hardware Kit
3/8” Pan Head Phillips Screws (x6)
* Recessed rack slide configuration shown
3/8” Undercut Flathead
Phillips Screws (x4)
Slide Bracket, (x2) Front
FIGURE 1-5:RACK SLIDE INSTALLATION DIAGRAM
CT-100C Installation 29
Page 30
RACKMOUNT DOOR KIT INSTALLATION
The rackmount door kit (Option 60) provides the necessary hardware to install a 1/8” thick
aluminum door in front of the rackmounted chassis. This option requires that the rackmount ear
kit also be installed and configured for recessed mounting. The door may be customer modified to
hold connectors, controls, indicators and similar components.
Required Tools
1. #2 Phillips Screwdriver
2. 1/8" Flat Blade Screwdriver
3. 11/32" Open Ended Wrench
Parts List
QTY ITEM VTI P/N
2 Screw, 6-32 x 1/2” Pan Head Ph, Sq Cone Sems Zinc 37-0028-050
2 Nut, Hex, 6-32, Zinc/Steel 37-0030-632
1 Latch, Vise Action, Know Style, Black 37-0065-000
2 Hinge, Adjustable Damping, Black 37-0066-000
8 Screw, 8-32 x 3/8” Pan Head Phil, M/S, Blk Zinc 37-0079-037
4 Screw, 8-32 x 3/8” Flat Head Phillips, Zinc 37-0080-037
1 Latch, Door, Rackmount, CT-100C 41-0133-000
1 Door, Front, Rackmount, CT-100C 41-0134-000
2 Brace, Rackmount, CT-100C 41-0136-000
VXI Technology, Inc.
Assembly Procedure
1. Lay the chassis on a protected work surface on its long side with the voltage monitor LEDs of
the chassis facing front with the power switch toward the top.
2. Locate two rackmount (cross) braces. Install one each along the front of the flanges (one on
top and one on the bottom) using four (two ea.) #8 flathead screws. The lip along the long
edge of each brace goes toward the front and pointing inside. The braces fit over the flanges
and the screws are placed from the inside going through the flange and then into the brace.
3. Locate the front door and two hinges. Install the two hinges to the front door using four black
screws provided.
4. Locate the latch kit and install it into the door from the side where the hinges are mounted.
5. Locate the door latch bracket and install it on the rackmount ear opposite the end where the
hinges will be installed. Use
6. Locate the four tapped holes on the front surface of the rackmount ear to which the door is to
be attached.
7. Install the door by its hinges to the rackmount ear using four black screws provided with the
option kit.
8. Test that the door opens and closes smoothly and adjust the latch as necessary to secure the
door when latched close.
Figure 1-6 as a reference.
30 CT-100C Installation
Page 31
www.vxitech.com
Rackmount Braces (x2)
3/8” Flat Head
Phillips Screws (x4)
Hinges (x2)
Front Door, Aluminum
3/8” Pan Head Black Oxide
Phillips Screws (x8)
Door Latch Kit
FIGURE 1-6:RACK DOOR INSTALLATION DIAGRAM
CT-100C Installation 31
Page 32
VXI Technology, Inc.
ACRYLIC RACKMOUNT DOOR KIT INSTALLATION
An acrylic rackmount door is offered as an alternative to the aluminum rackmount door. The kit
(Option 65) provides the necessary hardware to install a 1/8” thick acrylic door in front of the
rackmounted chassis. An installation diagram is provided on the following page for ease of
assembly. This option requires that the rackmount ear kit also be installed and configured for
recessed mounting.
Required Tools
1. #2 Phillips Screwdriver
2. 1/8" Flat Blade Screwdriver
3. 11/32" Open Ended Wrench
Parts List
QTY ITEM VTI P/N
2 Screw, 6-32 x 1/2" Pan Head Phillips, Sems Zinc 37-0028-050
2 Nut, Hex, 6-32, Zinc/Steel 37-0030-632
1 Latch, Vise Action, Knob Style, Black 37-0065-000
2 Hinge, Adjustable Damping, Black 37-0066-000
4 Screw, 8-32 x 3/8" Pan Head Phillips, Black Oxide 37-0079-037
1 Stop, Front Door, CT-100C 41-0133-000
1 Door, Front, CT-100, Acrylic 41-0303-000
2 Brace, Rackmount, CT-100C 41-0136-000
4 Nut, Hex, 8-32, Zinc/Steel 37-0030-832
4 Washer, Split Lock, 8-32 Zinc 37-0013-008
4 Washer, Flat 8-32, Zinc 37-0012-008
4 Screw, 8-32 x 1/2" Pan Head Phillips, Black Oxide 37-0079-050
4 Screw, 8-32 x 3/8" Flat Head Phillips, Zinc 37-0080-037
Assembly Procedure
1. Lay the chassis on a protected work surface on its long side with the voltage monitor LEDs of
the chassis facing front with the power switch toward the top.
2. Locate two rackmount (cross) braces. Install one each along the front of the flanges (one on
top and one on the bottom) using four (two ea.) #8 flathead screws. The lip along the long
edge of each brace goes toward the front and pointing inside. The braces fit over the flanges
and the screws are placed from the inside going through the flange and then into the brace.
3. Locate the front door and two hinges. Install the two hinges to the front door using four 1/2”
black oxide pan head screws through the holes of the hinge with the flat washer in contact
with the door and with the split lock washer between the hex nut and the flat washer.
4. Locate the latch kit and install it into the door from the side where the hinges are mounted.
5. Locate the door latch bracket and install it on the rackmount ear opposite the end where the
hinges will be installed. Use
6. Locate the four tapped holes on the front surface of the rackmount ear to which the door is to
be attached.
32 CT-100C Installation
Figure 1-7 as a reference.
Page 33
www.vxitech.com
7. Install the door by its hinges to the rackmount ear using four 3/8” black pan head screws
provided with the option kit.
8. Test that the door opens and closes smoothly and adjust the latch as necessary to secure the
door when latched close.
Rack Mount Braces (x2)
3/8” Flat Head
Phillips Screws (x4)
8-32 Hex Nuts (x4)
8-32 Split Lock Washer
8-32 Flat Washer
Front Door, Acrylic
Hinges (x2)
3/8” Pan Head Black Oxide
Phillips Screws (x4)
1/2” Pan Head Black O x i de
Phillips Screws (x4)
Door Latch Kit
F
IGURE 1-7:ACRYLIC RACK DOOR INSTALLATION DIAGRAM
CT-100C Installation 33
Page 34
INSTALLATION OF VXIMODULES
OVERVIEW
After the successful installation of rackmount accessories, the chassis is ready for installation of
the VXIbus base units (i.e. an SMP1100, SMP1200, etc.). It is recommended that the instruments
be installed after rackmount accessories have been installed to avoid any unnecessary physical
strains that may be incurred during the installation of the accessories. Whether single- or doublewide, the process of installation is simple.
Required Tools
1. Slotted screw driver
Installation Procedure
Install C-size modules directly into the mainframe as follows:
1. To prevent damage to the module, insure the mainframe is powered off.
2. Insert the module into the mainframe by aligning it with both the upper and lower card guide
of the desired slot. The card guide for the module is the right, longer guide-pair of each guide
set.
3. Gently push the module into the slot until it seats into the backplane connectors. The front
panel of the module should be flush with mainframe.
4. Tighten the mounting screws at the top and bottom of the module.
VXI Technology, Inc.
34 CT-100C Installation
Page 35
www.vxitech.com
DISCONNECTING THE MAINFRAME
To disconnect the CT-100C from its installation, simply follow the instructions below:
1. Place the mainframe in standby by depressing the power switch.
2. Remove the power cord from the mainframe to ensure that no power is running to the
mainframe.
3. Remove all cabling associated with the VXI modules installed in the CT-100C.
4. If the chassis is rack mounted, remove the chassis from the rack by removing the screws that
attach the mainframe to the rack.
Rackmount options can be removed from the mainframe at this time if desired.
CT-100C Installation 35
Page 36
VXI Technology, Inc.
36 CT-100C Installation
Page 37
www.vxitech.com
SECTION 3
SERVICE INFORMATION
INTRODUCTION
There are no operating instructions required for the CT-100C VXIbus chassis. After the chassis is
installed, operation is completely transparent to the operator. Just plug in the instruments then
power up the chassis.
NOTE Service should only be performed by qualified personnel.
Le service devrait seulement être assure par le personnel qualifié.
REPLICABLE PARTS LIST
The following table lists the parts that can be replaced on the CT-100C chassis.
Description Part Number
Final Assembly, Power Supply Replacement Kit, CT-100C 70-0271-000
CT-100C Service Information 37
Page 38
VXI Technology, Inc.
CT-100CPOWER SUPPLY REPLACEMENT
The CT-100C should operate without the need for service. In the event that the cooling fans or
power supply must be replaced, they are contained in a single, easy to remove module.
Replacement of the power supply can be accomplished as follows:
1. Remove the AC power cord from the chassis.
To avoid the possibility of shock, wait a minimum of ten seconds for electricity to dissipate
from the mainframe after removing power.
Pour éviter la possibilité de choc, attendez dix secondes au minimum affin que l’électricité se
dissipe l’unité centrale.
2. Locate the six screws at the rear of the chassis attaching the power supply to the mainframe
and remove all six.
3. Locate the two handles on the power supply and firmly pull the module straight back.
4. Locate the replacement power supply module and bring it up to the rear of the chassis.
5. Back the new power supply module into the rear of the chassis. Ensure that the connector on
the power supply is in line with the mating connector on the power interface board.
6. Seat the power supply module firmly in place and install the previously removed screws to
9.5 lbf/in (11.0 kgf/cm).
7. Verify that the front panel power switch is in the OFF position and reattach the power cord.
38 CT-100C Service Information
Page 39
www.vxitech.com
FIGURE 2-1:POWER SUPPLY REPLACEMENT DIAGRAM
CT-100C Service Information 39
Page 40
CLEANING THE MAINFRAME
During normal operation, dust is likely to accumulate inside the power supply of the CT-100C. To
remove this dust, power down the chassis and remove the power cord. Follow the instructions
provided on the previous page to remove the power supply from the mainframe. After removing
the power supply, use a pressurized air can to remove any dust present.