VWR International 634-6002, UV-3100PC Instruction Manual

VWR Collection Manual ver 1, rel 15.05.2007
UV-3100PC Spectrophotometer
INSTRUCTION MANUAL
European Catalogue Number(s): 634-6002
Version: 1.0.0 Issued: 07 , April 2010
Legal Address of Manufacturer
Europe
VWR International Europe bvba Researchpark Haasrode 2020 Geldenaaksebaan 464 B 3001 Leuven Tel.:+ 32 16 385 011 http://www.be.vwr.com
Made in China
Table of Contents
Part 1: Spectrophotometer Safety Information 1 Package Contents 1 Unpacking 2 Installation 2 Intended Use 2 Symbols and Conventions 3 Specifications 3 Overview 4 Description of Buttons and Switches 4 Operational Keys 6 Getting Started 8 Important Guidelines 9 General Operating Instructions 10 Operation 11
Basic Mode 12 Quantitative Mode 14 Wavelength Scan 19 Kinetics 21 DNA/Protein Mode 24 Multi Wavelength Mode 26
System Utility 28
Troubleshooting 34 Repair and Maintenance 35 Daily Maintain 35 Spare Parts Replacement 36 Accessories and Spares Parts 41 Technical Service 43
Warranty 43 Equipment Disposal 44
Part 2: Software Functions 45
Main Functions 45 Spectrum Processing Function 46 System Check and Calibration Function 46
Installation 47 Introduction 50 Operation 55
Single Wavelength Photometric Measurement 55 Fixed Point Measurement 55 Wavelength Scanning 61 Time Scanning (Kinetic Analysis) 73 DNA/Protein Measurement 75
Appendix 1 78 Appendix 2 78
Part 1: Spectrophotometer
1
Safety Information
Please follow the guidelines below, and read this manual in its entirety to ensure safe operation of the unit.
VWR recommends against the use of UV-3100PC Spectrophotometer.
Do not open the device.  Disconnect the device from the mains supply before carrying out
maintenance work or changing the fuses.
The inside of the device is a high-voltage area Danger!  Do not use the device if it is damaged, especially if the main power
cable is in any way damaged or defective.
Repairs may only be carried out by the service technicians from your
local VWR office and authorized contractual partners.
The device must be connected to a power outlet that has a protective
ground connection.
If the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment may be impaired.
Do not allow any liquid to enter into the device.
Do not operate the device in a hazardous location or potentially explosive environment.
Package Contents
Description Quantity Spectrophotometer 1PC Glass cuvette 4PCS Quartz cuvette 2PCS Power Cord ( Euro Plug ) 1PC Power Cord ( UK Plug ) 1PC
2
Power Cord ( SW Plug ) 1PC USB Cable 1PC CD-ROM 1PC Manual 1PC
Unpacking
Open the package, according to carefully check the packaging packing list items, if found inside the packaging are missing or damaged items please your local VWR office and authorized contractual partners.
Installation
Placement
Place the instrument on the stable table carefully.
Install printer (Optional)
Check to confirm instrument power switch is turned off, connect the printer’s data cable to
the Instrument’s parallel port.
Link the power cord
Check to confirm instrument power switch is turned off, the power cord plug into two separate power interface and power supply socket apparatus.
Intended Use
UV-3100PC Spectrophotometer used in Chemistry, Pharmaceuticals, Biochemical, metallurgy, Light Industry, Textile, Material, Environments, Medical, Education and some other fields. It is one of the most important instruments in Quality Control and an essential in normal laboratories.
3
Symbols and Conventions
The following chart is an illustrated glossary of the symbols that are used in this manual.
CAUTION This symbol indicates a potential risk and alerts you to proceed with caution
CAUTION This symbol indicates the presence of high voltage and warns the user to proceed with caution
CAUTION This symbol indicates risks associated with hot surfaces
Specifications
Optical System Single Beam Wavelength Range 190-1100nm Band Width 2nm Stray Light ≤0.05%T @ 220nm & 360nm Photometric Range 0-200%T,-0.3-3.0A, 0-9999C Wavelength Accuracy ±0.5nm Photometric Accuracy ≤±0.5%T or 0.005A@1A Stability 0.002A/h @ 500nm Memory 32K Language English, French, German, Spanish Display 320×240 Dots Matrix LCD Interface USB, Parallel Measuring Procedure Photometry, Quantitation, Wavelength Scan,
Kinetics, DNA/Protein, Multi-wavelength
Power Supply AC 110/220V, 50/60Hz Dimension 490×360×240mm Weight 14kg
This instrument is compliant to the European Directives on
4
Low Voltage Directive 2006/95/EC Electromagnetic compatibility 2004/108/EC Restriction on use of Hazardous Substances RoHS 2002/95/EC and theirs amandements.
Overview
UV-3100PC Spectrophotometer has the characters of wide range of wavelength, high sensitivity, powerful function, easy to use, simple structure and pretty figure. Besides these, the large LCD, High Precise A/D and easy to store RAM makes the instrument much more superior than other originals. It is widely used in Chemistry, Pharmaceuticals, Biochemical, metallurgy, Light Industry, Textile, Material, Environments, Medical, Education and some other fields. It is one of the most important instruments in Quality Control and an essential in normal laboratories.
Description of Buttons and Switches
Front View
5
Rear View
6
Operational Keys
Key Functions
7
Function Keys: Functions on-screen prompts
,
Numeric Keys: Enter numbers and letters
CLEAR Key: Delete the input value or stored data
ESC Key: Return to previous Interface
100%T/0Abs Key: Blank
OPEN Key: Open files stored in internal memory
SAVE Key: Save files to internal memory
START/STOP Key: Start/Stop testing
GOTO ʎ Keys: Set wavelength
PRINT Key: Print measuring result
8
ENTER Key: Confirm operation
CELL Key: Select/Deselect Auto-cell Holder
,
RIGHT, LEFT Keys: Search peak/valley and set X
scale
,
UP, DOWN Keys: Scroll menu/data and set Y
scale
Getting Started
The following chart describes the basic operation of the instrument.
Turn On and Self-check
Switch on the power. Then the instrument begins to self-check and 15 minutes’ warm up. Self-check includes the following steps:
WL: 500.0nm 10:00:00
D2
W
VWR International bvba
Wait until EasyRTOS booted:
Turn on lamps RAM Check Start RTOS kernel Initialize Comm. Port Initialize Printer Initialize AD System position Warm up.
9
D2
W
Press ESC to skip
WL: 500.0nm 10:00:00
Wait until EasyRTOS booted: Warm up 15 minutes…………?
After warm up, the instrument will ask the user to re-calibrate the system. Users can decide if they need to re-calibrate the system or not. After this step, the instrument can work normally.
Wait until EasyRTOS booted: Warm up 15 minutes…………?
D2
W
WL: 500.0nm 10:00:00
System calibration? No
D2
W
Cancel Move Dn Select
WL: 500.0nm 10:00:00
Main Menu
Basic Mode
Quantitative ModeWavelength ScanKinetics ModeDNA/Protein ModeMulti-Wavelength ModeSystem Utility
Important Guidelines
Reagents and dilution buffers can cause cauterization and other damage to health.  Samples (nucleic acids, proteins, bacteria cultures) can be infectious and cause
serious damage to health.
During sample preparation, measuring procedures and maintenance and cleaning
work, observe all local laboratory safety precautions (e.g. wear protective clothing and gloves, use of disinfectant) regarding the handling of sample material.
10
Dispose of measuring solutions and cleaning and disinfectant materials in
accordance with the relevant local laboratory regulations.
General Operating Instructions
Select Application
Main menu, press numeric keys or use the key , to choose corresponding menu,
then press to enter into.
Set Wavelength
Press to set wavelength, use numeric keys to input the values, press to confirm and go to the point you set, then do blank automatically.
Set Parameters
In different application, press function key to set parameters, press , to choose or
input the values by numeric keys, press to enter into, press to return.
Set Auto-cell Holder
Press to active the auto cell holder and press the numeric key (1-8)to make
corresponding cell position at the light path. Press again to inactive the auto cell holder.
Delete the Input Value
Press to delete a character, press to delete all the characters.
Delete the Test Results and Stored data
Press to delete the test result or stored data.
Blank
11
Put the Reference in the light path, press to do blank.
Measure Samples
Put the samples in the light path, press to measure.
Print the Test Results
Press to print the test results.
Store the Test Results
Press to store the test results, input the file name by numeric keys and press
to save.
Load the Stored File
In the test interface, press to go into file selecting interface, press , to
choose the file you want, press to open.
Operation
Self-check
Remove all the blocks in the light path and close the lid of the compartment; Switch on the power supply to begin the self-test.
Warm up
After self-test, the instrument goes into pre-warm state. For accurate test, at least 30 minutes of warm up is required.
12
Check the cuvettes
The cuvettes must be clear and there’s no remains of the samples on the surface of it. Only Silicon (Quartz) cuvettes are permitted to be used in the range of UV area.
Basic Mode
1. Enter into Basic Mode
Main menu, press numeric key or , to choose “Basic Mode”, then press
to go into.
2. Set Photometric Mode
Press to set photometric mode. Press , to choose “Abs.”, “T%” or
“Conc./Factor” and press to confirm. If users choose “Abs.” or “T%”, please go to step 5 directly.
3. Set Concentration Unit
Press to set concentration unit. Press , to choose unit followed with pressed to confirm. You can also choose “other” to input the self defined unit.
4. Set “Factor” or “Standard”
Two methods are under your choice:
Method 1::Input Factor F
13
Press to set F. Input the value of F by numeric keypad, press to confirm. Then the F value would display on the screen.
Method 2::Standards Mark
Put the reference sample in the light path and calibrate 100%T/0Abs; Put the
standard sample in the light pathpress to start the mark. Input the
concentration value of the standard and press to confirm, then it displays on the screen.
5. Set Wavelength
Press to set wavelength, input the value by the numeric keypad followed with
pressed to confirm.
6. Blank
Put the Reference in the light path and press to do blank.
7. Measurement samples
Put the sample to be measured in the light path, then the result displays on the screen automatically.
8. Print the Test Results
Press to print the Test Results.
14
Basic Mode Test Report
Wavelength: 500.0nm Result: 0.000 Abs Date & Time: mm-dd-yyyy, hh:mm:ss
Model: UV-3100PC SN: UEFXXXXXXX Version: A1.176 VWR International bvba.
Quantitative Mode
1. Enter into Quantitative Mode
Main menu, press or press , to choose “Quantitative Mode” followed with
pressed to confirm.
2. Set Unit
Press to set concentration unit, press , to choose and press to confirm.
3. Set up Standard Curve or load the stored curves
Press to go into set up interface, 2 methods are under your choice.
15
Set up Standard Curve:
Method 1: Input Regression Equation
1) Set Fit Curve Method. Press to set Fit methoduse , to
choose the method and press to confirm.
2) Set Wavelength. Press to set wavelength. Use , to choose
measure method, then press to confirm. Input the wavelength value
you need and press to confirm.
3) Input the Factor of the Regression Equation. Press and input the
factors, press to confirm.
Method 2: Use Standard Samples
1) Set Fit Method. Press to set fit method, press , to choose fit
method, then press to confirm.
16
2) Set Wavelength. Press to go into wavelength setting interface, press
, to select measure method and press to confirm. Input the
value of the wavelength and press to confirm.
3) Blank. Put the Reference Sample in the light path and press to do blank.
4) Setup Standard Samples. Press to setup standard, input the concentrations of corresponding standard samples according the indication
and press to confirm. Users can use , to choose the value
you just input and press to delete, then input a new value, press
to confirm. Press to cancel after all the input.
5) Calibrate Standard Samples. Put the corresponding standard samples in the
light path as the screen indicates and press to measure. Then the Abs. value would appear in the corresponding table.
17
Load the Stored Curves
In the “Calibration Table” interface, press to go into files select interface. Use
, to select the curve you need and press to load. Users can press
to view the curve, press to cancel.
0 Conc. 4
4. Return the sample measurement interface
In the “Calibration Table” interface, press to return the Quantitative Test interface.
5. Blank
18
Put the Reference Sample in the light path, press to do blank.
6. Measure Samples
Place the sample to be tested in the light path, press to measure. Then the test result will display in the data sheet. Repeat this step to finish measuring all the samples.
7. Print the Test Results
Press to print the Test Results.
Quantitative Test Report
File Name: Abc Date & Time: mm-dd-yyyy, hh:mm:ss
No. 500.0nm Abs(eff) C(mg/L)
1 0.212 0.212 0.212 2 0.210 0.210 0.210 3 0.209 0.209 0.209
Fitting Params:
C=1.000*A^1
r=1.000 Model: UV-3100PC SN: UEFXXXXXXX Version: A1.176 VWR International bvba.
19
Wavelength Scan
1. Enter into Wavelength Scan
Main menu, press numeric key or , to choose “Wavelength Scan” and
press to enter.
0.000 Abs 1.200
2. Parameters Setup
Press to set parameters, set “Scan From” , “scan to”, “scan step” and “scan speed”,
press to confirm.
3. Set Photometric Mode
Press to set photometric modechoose “T%”, “Abs.” or “E” and press to confirm.
4. Scan Baseline
Put Reference Sample in the light path, press to scan the baseline, press to cancel.
5. Scan Samples
Put the samples in the light pathpress to scan the sample, press to cancel.
Loading...
+ 63 hidden pages