20 Compliance with local laws and regulations . . . . . . . 94
6
ba77077e0209/2014
MU 6100 HSafety
1Safety
1.1Safety information
1.1.1Safety information in the operating manual
This operating manual provides important information on the safe operation of
the meter. Read this operating manual thoroughly and make yourself familiar
with the meter before putting it into operation or working with it. The operating
manual must be kept in the vicinity of the meter so you can always find the information you need.
Important safety instructions are highlighted in this operating manual. They are
indicated by the warning symbol (triangle) in the left column. The signal word
(e.g. "CAUTION") indicates the level of danger:
WARNING
indicates a possibly dangerous situation that can lead to serious (irreversible) injury or death if the safety instruction is not
followed.
CAUTION
indicates a possibly dangerous situation that can lead to
slight (reversible) injury if the safety instruction is not followed.
NOTE
indicates a possibly dangerous situation where goods might be damaged
if the actions mentioned are not taken.
1.1.2Safety signs on the meter
Note all labels, information signs and safety symbols on the meter and in the
battery compartment. A warning symbol (triangle) without text refers to safety
information in this operating manual.
1.1.3Further documents providing safety information
The following documents provide additional information, which you should
observe for your safety when working with the measuring system:
•Operating manuals of sensors and other accessories
•Safety datasheets of calibration or maintenance accessories (such as buffer
solutions, electrolyte solutions, etc.)
ba77077e0209/20147
SafetyMU 6100 H
1.2Safe operation
1.2.1Authorized use
This meter is authorized exclusively for pH, ISE, ORP, dissolved oxygen and
conductivity measurements in a field and laboratory environment.
Only the operation and running of the meter according to the instructions and
technical specifications given in this operating manual is authorized (see
section 3 T
Any other use is considered unauthorized.
1.2.2Requirements for safe operation
Note the following points for safe operation:
•The meter may only be operated according to the authorized use specified
above.
•The meter may only be supplied with power by the energy sources
mentioned in this operating manual.
•The meter may only be operated under the environmental conditions
mentioned in this operating manual.
•The meter may only be opened if this is explicitly described in this operating
manual (example: Inserting the batteries).
ECHNICALDATA, page 10).
1.2.3Unauthorized use
The meter must not be put into operation if:
•it is visibly damaged (e.g. after being transported)
•it was stored under adverse conditions for a lengthy period of time (storing
conditions, see section 3 T
ECHNICALDATA, page 10).
8ba77077e0209/2014
MU 6100 HOverview
1
2
3
6100 H
MU
2Overview
2.1MU 6100 H meter
The MU 6100 H meter enables you to perform measurements (pH, U, conductivity, dissolved oxygen) ) quickly and reliably.
The MU 6100 H provides the maximum degree of operating comfort, reliability
and measuring certainty for all applications.
1 Keypad
2 Display
3 Connectors
2.2Sensors
A measuring system ready to measure consists of the MU 6100 H meter and a
suitable sensor.
Suitable sensors are pH electrodes, ORP electrodes, D.O. sensors and
conductivity measuring cells.
ba77077e0209/20149
Technical dataMU 6100 H
3Technical data
3.1Measuring ranges, resolution, accuracy
3.1.1pH/ORP
Measuring ranges,
resolution
VariableMeasuring rangeResolution
pH-2.0 ... +20.00.1
-2.00 ... +20.000.01
- 2.000 ... + 19.9990.001
U [mV]-2500 ... +25001
-1200.0 ... +1200.0.1
T [°C]-5.0 ... +105.00.1
T [°F]23.0 ... +221.00.1
Manual
temperature input
VariableRangeIncrement
T
T
[°C]-25 ... +1301
manual
[°F]-13 ... +2661
manual
Accuracy (± 1 digit)VariableAccuracyTemperature of the test
sample
pH / range *
-2.0 ... +20.0± 0.1+15 °C ... +35 °C
-2.00 ... +20.00± 0.01+15 °C ... +35 °C
- 2.000 ... + 19.999± 0.005+15 °C ... +35 °C
U [mV] / range
-2500 ... +2500± 1+15 °C ... +35 °C
-1200.0 ... +1200.± 0.3+15 °C ... +35 °C
T [°C] / temperature sensor
NTC 30± 0.1
PT 1000± 0.1
* when measuring in a range of ± 2 pH around a calibration point
The accuracy values specified here apply exclusively to the meter.
The accuracy of the electrodes and buffer solutions has to be taken
into account additionally.
10ba77077e0209/2014
MU 6100 HTechnical data
3.1.2Oxi
Measuring ranges,
resolution
(depending on the
sensor)
VariableMeasuring rangeResolution
Concentration [mg/l]0 ... 20.00 (0 ... 20.0)
20.0 ... 90.0
(20 ... 90)
Saturation [%]0 ... 200.0 (0 ... 600)
0 ... 600
D.O. partial pressure
[mbar]
0 ... 200.0 (0 ... 1250)
0 ... 1250
T [°C]0 ... 50,00.1
Accuracy(± 1 digit)VariableAccuracy
Concentration [mg/l]± 0.5 % of measured value
at ambient temperature of +5 °C ... +30 °C
Saturation [%]± 0.5% of measured value
when measuring in the range of ± 10 K
around the calibration temperature
D.O. partial pressure
[mbar]
± 0.5 % of measured value
at ambient temperature of +5 °C ... +30 °C
T [°C] / temperature sensor
0.01 (0.1)
0.1 (1)
0.1 (1)
1
0.1 (1)
1
NTC 30± 0.1
PT 1000± 0.1
Correction functionsTemperature
compensation
Salinity correction0 ... 70.0 SAL
Air pressure correction
The accuracy values specified here apply exclusively to the meter.
The accuracy of the D.O. sensors has also to be taken into
account.
Accuracy better than 2 % at 0 ... +40 °C
Automatic through integrated
pressure sensor in the range of 500 ... 1100
mbar
ba77077e0209/201411
Technical dataMU 6100 H
3.1.3Cond
Measuring ranges,
resolution
(depending on the sensor)
VariableMeasuring rangeResolution
ϰ[µS/cm]0.000 ... 1.999*
0.00 ... 19.99**
0.0 ... 199.9
200 ... 1999
ϰ [mS/cm]2.00 ... 19.99
20.0 ... 199.9
200 ... 1000
(Resistivity)
ρ
[Ohm*cm]
(Resistivity)
ρ
[kOhm*cm]
(Resistivity)
ρ
[MOhm*cm]
SAL0.0 ... 70.0
1.000 ... 1.999
2.00 ... 19.99
20.0 ... 199.9
200 ... 1999
2.00 ... 19.99
20.0 ... 199.9
200 ... 1999
2.00 ... 19.99**
20.0 ... 199.9*
0.001
0.01
0.1
1
0.01
0.1
1
0.001
0.01
0.1
1
0.01
0.1
1
0.01
0.1
0.1
according to the IOT
table
TDS0 ... 1999 mg/l
2.00 ... 19.99 g/l
20.0 ... 199.9 g/l
T [°C]-5.0 ... +105.00.1
T [°F]+23.0 ... +221.00.1
* only possible with cells of the cell constant, 0.010 cm-1
** only possible with cells of the cell constant, 0.010 cm
0.110 cm
-1
Cell constants Cell constant CValues
Can be calibrated in
the ranges
0.450 ... 0.500 cm
0.800 ... 0.880 cm
Adjustable0.090 ... 0.110 cm
0.010 cm-1 (fixed)
0,250 ... 1,000 cm
Reference temperatureReference tempera-
Values
ture
1
0.01
0.1
-1
or 0.090 ...
-1
-1
-1
-1
Adjustable20 °C (Tref20)
25 °C (Tref25)
12ba77077e0209/2014
MU 6100 HTechnical data
Accuracy (± 1 digit)VariableAccuracyTemperature of the test
sample
ϰand ρ / temperature compensation
None (Off)± 0.5 %
Nonlinear (nLF)± 0.5 %0 °C ... +35 °C
according to EN 27 888
± 0.5 %+35 °C ... +50 °C
enhanced nLF function
Linear (lin)± 0.5 %+10 °C ... +75 °C
SAL / range
0.0 ... 42.0± 0.1+5 °C ... +25 °C
± 0.2+25 °C ... +30 °C
TDS [mg/l]
± 0.5 %
T [°C] / temperature sensor
NTC 30± 0.1
PT 1000± 0.1
The accuracy values specified here apply exclusively to the meter.
The accuracy of the measuring cell has also to be taken into
account.
Further data are given in the documentation of your sensor.
Batteries4 x 1.5 V alkali-manganese batteries, type
AA
Rechargeable batteries
4 x 1.2 V NiMH rechargeable batteries,
type AA
(no charging function)
Operational lifeUp to 1000 h without / 150 h with illumination
pH sensor inputInput resistance> 5 * 10
Input current< 1 * 10
USB interfaceTypeUSB 1.1
USB B (device), data output
Baud rateAdjustable:
1200, 2400, 4800, 9600, 19200 Baud
Data bits8
Stop bits2
ParityNone
HandshakeRTS/CTS
Cable length Max. 3 m
12
-12
ohm
A
Guidelines
and norms used
EMCEC directive 2004/108/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A
Meter safetyEC directive 2006/95/EC
EN 61010-1
UL 61010-1
CAN/CSA-C22.2#61010-1
IP protection classEN 60529
14ba77077e0209/2014
MU 6100 HCommissioning
4Commissioning
4.1Scope of delivery
MeterMU 6100 H
4 batteries 1.5 V Mignon type AA
Short instructions
CD-ROM with
– USB drivers
– detailed operating manual
– Software MultiLab Importer
4.2Power supply
The MU 6100 H is supplied with power in the following ways:
Battery operation (4 x alkaline manganese batteries, type AA)
USB operation via a connected USB-B cable
4.3Initial commissioning
Perform the following activities:
Insert the supplied batteries
Switch on the meter
(see section 5.2 S
Set the date and time
(see section 5.4.6 E
page 24)
WITCHINGONTHEMETER, page 19)
XAMPLE 2 ONNAVIGATION: SETTINGTHEDATEAND TIME,
ba77077e0209/201415
CommissioningMU 6100 H
1
2
4.3.1Inserting the batteries
1Screws
2 Battery compartment
CAUTION
Make sure that the poles of the batteries are positioned correctly.
The ± signs on the batteries must correspond to the
the battery compartment.
You can operate the meter either with normal batteries or with
rechargeable batteries (Ni-MH). In order to charge the batteries, an
external charging device is required.
1.Place four batteries (type Mignon AA) in the battery compartment.
2.Close the battery compartment.
3.Set the date and time
(see section 5.4.6 E
XAMPLE 2 ONNAVIGATION: SETTINGTHEDATEAND TIME,
page 24).
± signs in
16ba77077e0209/2014
MU 6100 HOperation
5Operation
5.1General operating principles
5.1.1Keypad
In this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. <OK>) generally indicates a short keystroke (under
2 sec) in this operating manual. A long keystroke (approx. 2 sec) is indicated
by the underscore behind the key symbol (e.g. <OK
_>).
<F1>:
<F1
_>:
<F2>:
<F2
_>:
<On/Off>:Switches the meter on or off
<MODE>:Selects the measured parameter / Quits the settings
Configures the automatic data output to the USB inter-
face
Calls up the calibration procedure
Displays the calibration data
Saves a measured value manually
Opens the menu for the automatic save function
Displays the manually stored measured values
Displays the automatically stored measured values
Menu control, navigation
Increments, decrements values
Increments, decrements values continuously
Opens the menu for measurement settings / confirms
entries
Opens the menu for system settings
<HOLD>Freezes the measured value (HOLD function)
ba77077e0209/201417
OperationMU 6100 H
HOLD AR
25.0
6.093
pH
°C
4
3
2
5
6
7
8
1
AutoCal TEC
15.03.2014
08:00
Menu
5.1.2Display
1 Status information (sensor)
2 Measured value (with unit)
3 Measured parameter
4 Channel display: Plug position of the sensor
5 Sensor symbol (calibration evaluation, calibration interval)
6 Measured temperature (with unit)
7 Status information (meter)
8 Softkeys and date + time
5.1.3Status information (meter)
ARStability control (AutoRead) is active (calibration)
HOLDMeasured value is frozen (<HOLD> key)
Batteries are almost empty
Data are automatically output to the USB-B interface at
intervals
18ba77077e0209/2014
MU 6100 HOperation
1
2
3
4
5
pH
6.949
25.0
°C
15.02.2014
14:15
Menu
5.1.4Connectors
1 pH electrode
2 Temperature sensor
3 D.O. sensor or conductivity measuring cell
4 Mini USB-B interface
5 Service interface
CAUTION
Only connect sensors to the meter that cannot return any voltages or currents that are not allowed (> SELV and > current
circuit with current limiting).
Almost all customary sensors fulfill these conditions.
5.2Switching on the meter
1.Switch the meter on with <On/Off>.
The meter performs a self-test.
The display shows the manufacturer's logo while the self-test is being
performed.
The measured value display appears.
ba77077e0209/201419
OperationMU 6100 H
5.3Switching off
1.Switch the printer off with <On/Off>.
Automatic shut-off
function
The instrument has an automatic shut-off function in order to save the batteries
(see section 10.2.1 S
YSTEM, page 62). The automatic shut-off function switches
off the meter if no key is pressed for an adjustable period.
The automatic shut-off function is not active
if the communication cable is connected
if the Automatic data storage function is active, or with automatic data trans-
mission
Display illuminationThe meter automatically switches off the display illumination if no key is
pressed for 30 seconds. The illumination is switched on with the next keystroke
again.
You can also generally switch the display illumination on or off (see section
10.2.1 S
YSTEM, page 62).
5.4Navigation
The principles of navigation in menus and dialogs are explained in the following
sections.
5.4.1Operating modes
The instrument has the following operating modes:
Operating
Explanation
mode
MeasuringThe measurement data of the connected sensor are shown
in the measured value display
CalibrationThe course of a calibration with calibration information, func-
tions and settings is displayed
Storing in
The meter stores measuring data automatically or manually
memory
Transmitting data
The meter transmits measuring data and calibration records
to a USB-B interface automatically or manually.
SettingThe system menu or a sensor menu with submenus, set-
tings and functions is displayed
Only those displays and functions are available in the active operating mode
that are currently being required.
20ba77077e0209/2014
MU 6100 HOperation
General
Interface
Clock
Service information
Reset
System
15.02.2014
14:15
Back
5.4.2Measured value display
In the measured value display, open the setting menus with <OK>. The current
functions of the softkeys are shown on the display.
Use <OK> (short
pressure) to open the menu for calibration and measure-
ment settings for the displayed measured parameter.
Use <OK
_> (long keystroke (approx. 2 s) to open the Storage & config
menu with the sensor-independent settings.
Use the keys of the keypad to carry out further functions such as storage or calibration (see section 5.1.1 K
EYPAD, page 17). These functions are not available
in other operating situations.
5.4.3Menus and dialogs
The menus for settings and dialogs in procedures contain further subelements.
To select a subelement, use the <><> keys. The current selection is dis-
played with a frame.
To make further settings, switch to the next higher menu level with
<F1>[Back].
Use <MODE> to return to the measured value display.
5.4.4Elements in menus and dialogs
Submenus
The name of the submenu is displayed at the upper edge of the frame. Submenus are opened by confirming with <OK>. Example:
Settings
Settings are indicated by a colon. The current setting is displayed on the
right-hand side. The setting mode is opened with <OK>. Subsequently, the
setting can be changed with <><> and <OK>. Example:
ba77077e0209/201421
OperationMU 6100 H
Language:Deutsch
Beep:Off
Illumination:On
Contrast:50 %
Switchoff time:1 h
Temperature unit:°C
Stability control:On
General
15.02.2014
14:15
Back
Calibration record
Calibration data storage
Buffer:TEC
One point calibration:Yes
Calibration interval:7 d
Unit for slope:mV/pH
i
4.00 7.00 10.00 (20 °C)
pH
15.02.2014
14:15
Back
Calibration record
Calibration data storage
Buffer:TEC
One point calibration:Yes
Calibration interval:7 d
Unit for slope:mV/pH
i
4.00 7.00 10.00 (20 °C)
pH
15.02.2014
14:15
Back
Functions
Functions are designated by the name of the function. They are immediately
carried out by confirming with <OK>.
Example: Display the Calibration record function.
22ba77077e0209/2014
Messages
Information is marked by the i symbol. It cannot be selected. Example:
5.4.5Navigation example 1:Setting the language
1.Press the <On/Off> key.
The measured value display appears.
The instrument is in the measuring mode.
MU 6100 HOperation
pH
6.949
25.0
°C
15.02.2014
14:15
Menu
System
Data storage
Storage & config
15.02.2014
14:15
Back
General
Interface
Clock
Service information
Reset
System
15.02.2014
14:15
Back
2.Open the Storage & config menu with <OK_>.
The instrument is in the setting mode.
3.Select the System submenu with <><>.
The current selection is displayed with a frame.
4.Open the System submenu with <OK>.
5.Select the General submenu with <><>.
The current selection is displayed with a frame.
6.Open the General
submenu with <OK>.
ba77077e0209/201423
OperationMU 6100 H
Language:Deutsch
Beep:Off
Illumination:On
Contrast:50 %
Switchoff time:1 h
Temperature unit:°C
Stability control:On
General
15.02.2014
14:15
Back
Language:Deutsch
Beep:Off
Illumination:On
Contrast:50 %
Switchoff time:1 h
Temperature unit:°C
Stability control:On
General
15.02.2014
14:15
Back
7.Open the setting mode for the Language with <OK>.
24ba77077e0209/2014
8.Select the required language with <><>.
9.Confirm the setting with <OK>.
The meter switches to the measuring mode.
The selected language is active.
5.4.6Example 2 on navigation: Setting the date and time
The meter has a clock with a date function. The date and time are shown in the
measured value display.
When storing measured values and calibrating, the current date and time are
automatically stored as well.
The correct setting of the date and time and date format is important for the following functions and displays:
Current date and time
Calibration date
Identification of stored measured values.
Therefore, check the time at regular intervals.
After a fall of the supply voltage (empty batteries), the date and time
are reset.
MU 6100 HOperation
Date format:dd.mm.yyyy
Date:15.02.2014
Time:14:15:25
Clock
15.02.2014
14:15
Back
The date format can be switched from the display of day, month, year
(dd.mm.yyyy) to the display of month, day, year (mm/dd/yyyy or mm.dd.yyyy).
1.In the measured value display:
Open the Storage & config menu with <OK
_>.
The instrument is in the setting mode.
2.Select and confirm the System / Clock menu with <><> and <OK>.
The setting menu for the date and time opens up.
3.Select and confirm the Time menu with <><> and <OK>.
The hours are highlighted.
4.Change and confirm the setting with <><> and <OK>.
The minutes are highlighted.
5.Change and confirm the setting with <><> and <OK>.
The seconds are highlighted.
6.Change and confirm the setting with <><> and <OK>.
The time is set.
7.If necessary, set the Date and Date format. The setting is made similarly
to that of the time.
8.To make further settings, switch to the next higher menu level with
[Back]<F1>.
or
Switch to the measured value display with <MODE>.
The instrument is in the measuring mode.
ba77077e0209/201425
OperationMU 6100 H
Channel display
Display of the plug position
for the respective parameter
The filled bar indicates for
each connected sensor to
which plug position of the
meter it is connected.
pH
6.949
24.8 °C
15.02.2014
14:15
USB output
pH6.966
25.0 °C
ϰ
0.840 1/cm
25.0 °C
15.02.2014
14:15
USB output
pH
6.966
25.0
°C
15.02.2014
14:15
USB output
5.5Channel display
The MU 6100 H manages the connected sensors and displays which sensor is
plugged to which connection.
5.5.1Display of several sensors in the measuring mode
The measured values of the connected sensors can be displayed in the following ways:
Clear display of all connected sensors
Detailed display of one sensor
With the softkey you can very easily switch between the two display types. The
suitable softkey is displayed depending on the operating situation.
26ba77077e0209/2014
MU 6100 HpH value
pH
6.949
24.8 °C
15.02.2014
14:15
Menu
6pH value
6.1Measuring
6.1.1Measuring the pH value
NOTE
When connecting an grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect.
The USB interface is not galvanically isolated.
To ensure the high measurement accuracy of the measuring system, always measure with a calibrated electrode (see section 6.2
C
ALIBRATION, page 29).
1.Connect the pH electrode to the meter.
2.If necessary, select the measured parameter with <MODE>.
3.When measuring without temperature sensor:
– Temper the test sample, or measure the current temperature.
– Enter the temperature value in the menu.
4.Immerse the pH electrode in the test sample.
The measured value is checked for stability (automatic stability control).
The display of the measured parameter flashes.
5.Wait for a stable measured value.
The display of the measured parameter no longer flashes.
Stability control
(AutoRead )
Criteria for a stable
measured value
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproducibility of measured values (see section 10.2.3 A
page 63).
The Stability control function checks whether the measured values are stable
UTOMATIC STABILITYCONTROL,
ba77077e0209/201427
pH valueMU 6100 H
within the monitored time interval.
Freezes the mea-
sured value (HOLD
function)
Measured
Time intervalStability in the time interval
parameter
pH value15 seconds
Temperature15 seconds
Δ : better than 0.01 pH
Δ : better than 0.5 °C
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
With the HOLD function, you can freeze the current measured value. The displayed measured value stops changing until you switch the HOLD function off.
1.Freeze the measured value with <HOLD>.
The [HOLD] status indicator is displayed.
2.Release the frozen measured value again with <HOLD>.
The HOLD function is switched off.
The [HOLD] status display disappears.
6.1.2Measuring the temperature
For reproducible pH measurements, it is essential to measure the temperature
of the test sample.
You have the following options to measure the temperature:
Automatic measurement of the temperature with the temperature sensor
(NTC30 or Pt1000) integrated in the sensor.
Measurement by an external temperature sensor.
Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
Tempera-
ture sensor
Resolution of the
temp. display
Temp. measurement
yes0.1 °CAutomatic with temperature sensor
-1 °CManual
If you wish to measure (or calibrate) without temperature sensor, proceed as
follows:
1.Measure the current temperature of the test sample.
2.In the <OK>/pH/Man. temperature menu, set the temperature value with
<><>.
28ba77077e0209/2014
MU 6100 HpH value
6.2Calibration
6.2.1Why calibrate?
pH electrodes age. This changes the zero point (asymmetry) and slope of the
pH electrode. As a result, an inexact measured value is displayed. Calibration
determines and stores the current values of the zero point and slope of the
electrode.
Thus, you should calibrate at regular intervals.
6.2.2When do you have to calibrate?
After connecting a sensor
Routinely within the framework of the company quality assurance
When the calibration interval has expired
6.2.3Automatic calibration (AutoCal)
Make sure that in the sensor menu, Buffer menu, the buffer set is correctly
selected (see 10.1.1 S
ETTINGSFORPH MEASUREMENTS, PAGE 52).
Use any one to five buffer solutions of the selected buffer set in ascending or
descending order.
Below, calibration with Technical buffers (TEC) is described. When other buffer
sets are used, other nominal buffer values are displayed. Apart from that, the
procedure is identical.
If single-point calibration was set in the menu, the calibration procedure is automatically finished with the measurement of buffer solution 1 and the calibration record is displayed.
1.Connect the pH electrode to the meter.
2.When measuring without temperature sensor:
– Temper the buffer solution, or measure the current temperature.
– Enter the temperature value in the menu.
3.In the measured value display, select the measured parameter pH or mV
with <MODE>.
4.Start the calibration with <CAL>.
The calibration display for the first buffer appears (voltage display).
ba77077e0209/201429
Loading...
+ 67 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.