VWR International 89231-656, 89231-658, 89231-660 User Manual

DOC022.53.80319
sympHony
Multisensor Probes
07/2012, Edition 1
User Manual
Safety information on page 3 Calibration on page 6
Specifications on page 3 Sample requirements on page 8
Product overview on page 4 Sample measurement on page 9
Meter compatibility on page 5 Maintenance on page 10
Preparation for use on page 5 Troubleshooting on page 12
Polarize the dissolved oxygen probe (89231-660 multisensor) on page 5
Safety information
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end­of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
Specifications
Specifications are subject to change without notice.
Specification 89231-656 89231-658 89231-660
Parameter pH, conductivity and ORP pH and conductivity pH, conductivity and
dissolved oxygen
Measurement range 0 to 14 pH
5 to 30,000 µS/cm
±2000 mV
0 to 14 pH
5 to 30,000 µS/cm
0 to 14 pH
5 to 30,000 µS/cm
0.03 mg/L oxygen to saturation
Temperature range 0 to 80 °C
(32 to 176 °F)
0 to 80 °C
(32 to 176 °F)
0 to 50 °C
(32 to 122 °F)
Temperature sensor Yes Yes Yes
Sensor pH: glass (standard)
Conductivity: platinum
pH: glass (standard)
Conductivity: platinum
pH: glass (standard)
Conductivity: platinum
Dissolved oxygen: replaceable dissolved oxygen membrane
Reference type Ag/AgCl Ag/AgCl Ag/AgCl
Junction type Porous pin Porous pin Porous pin
Electrolyte Non-refillable gel Non-refillable gel Non-refillable gel
Storage solution KCl 0.01 M (conductivity
standard 1412 µS/cm)
KCl 0.01 M (conductivity standard 1412 µS/cm)
3M KCl
Number of poles 2 2 2
English 3
Specification 89231-656 89231-658 89231-660
Conductivity cell constant (cm-1)
1.0 1.0 1.0
Platinized No No No
Meter compatibility H30PCO H30PCO and H30PCD H30PCD
Body material Polycarbonate Polycarbonate Polycarbonate
Length 170 mm 170 mm 170 mm
Diameter 12 mm 12 mm 40 mm total diameter with
12 mm individual sensors
Cable 1 m 1 m 1 m
Connector MP8
1
MP8
1
MP8
1
1
For use with sympHony handheld meters only
Product overview
The multisensor probes are combination gel-filled electrodes for pH, conductivity and ORP or dissolved oxygen measurements in general aqueous samples. The multisensor probes have a built­in temperature sensor. Refer to Figure 1 and Figure 2. Refer to Specifications on page 3 for the probe parameters. Refer to Sample requirements on page 8.
The multisensor probes are used with sympHony meters. Refer to Meter compatibility on page 5.
Figure 1 Probe overview – single-probe
1 Probe 2 Probe storage tube
4 English
Figure 2 Probe overview – multl-probe (89231-660)
1 Protective shroud 4 Conductivity probe 7 Anode
2 Probe storage tube 5 mini-MP connector
1
8 Cathode
3 Dissolved oxygen probe 6 pH probe 9 Membrane module
1
Each probe has a unique mini-MP connector.
Meter compatibility
The multisensor probes are compatible with the sympHony handheld meter models: H30PCO and H30PCD. Refer to Specifications on page 3 for specific compatibility.
Preparation for use
C A U T I O N
Personal injury hazard. Broken glass can cause cuts. Use tools and personal protective equipment to remove broken glass.
To prepare the probe for calibration or sample measurement:
1. Turn the probe storage tube(s) and remove from the probe. Keep the probe storage tube(s).
2. Rinse the probe with deionized water. Blot dry with a lint-free cloth.
Polarize the dissolved oxygen probe (89231-660 multisensor)
Before calibration or measurement of dissolved oxygen with the 89231-660 multisensor, polarize the dissolved oxygen probe.
The dissolved oxygen probe must be polarized every time the multisensor is disconnected from the meter or the batteries are removed from the meter. The probe is continuously polarized while it is connected to the meter.
1. Fill the dissolved oxygen probe storage tube with deionized water to the mark.
English
5