VOLTECH PM1000 DATA SHEET

Voltech
PM1000+
Precision Power Analyzer
• Easy to use.
• 0.1% basic accuracy.
• Special standby power, inrush, and W-hr integration modes.
• Comprehensive high-speed interfacing for automated use.
• Ideal for the design and test of all electrical products.
The No Compromise Power Analyzer
Driven by consumer demand and energy efficiency legislation, tomorrow’s electrical and electronic products must operate with ever-greater efficiency and employ increasingly complex control methods such as a low-power standby operation. The accurate measurement of electrical power has never been more important than it is today.
The Voltech PM1000+ is the first power analyzer to combine bench instrument accuracy with sophisticated energy consumption features and low-power standby measurements at an affordable price. The PM1000+ measures power consumption from milliwatts to megawatts, providing accurate power and harmonic data on products ranging from the tiniest cell phone charger to the latest electric hybrid bus.
Designed and built using over 20 years of Voltech know-how. The PM1000+ is the most powerful, accurate, no­compromise power and energy analysis tool for the design and test of tomorrow’s products.
Features and benefits…
• Direct connection – no CT errors
• Accurate up to Crest Factors of 20 – no compromise specification on distorted waveforms
• Rugged analog design-stands overloads up to 5kV
• Discrete Fourier Transform provides harmonics more accurately than FFT
• Voltech proprietary frequency detection avoids problems with zero crossing detection
• Built with low power standby in mind- no special accessories or channels required
• High sample rate captures all the data and avoids aliasing problems
• Full color clear and versatile display
• Great interfacing USB, IEEE, RS232
1986 The world’s first digital power analyzer, the Voltech PM1000.
When you really want to be sure,
you can trust Voltech.
When you really want to be sure,
you can trust Voltech.
For more information or to request a no-obligation
trial of a PM1000+, please see our website at
www.voltech.com
Measurement Functions - Selectable From Menu
Direct Input Ratings
660Vrms. 1500Vpk
20Arms, 100Apk
External Shunt Input Ideal for current transducers with a voltage output as well as resistive shunts
Bright 320 X 240 color display
Easy-to-use menu keys
On-board help and character entry
Volts 0 to 600 RMS & +/-1500V Peak Direct Input
Current 0 to 20A RMS & +/-100A Peak Direct Input
Power Watts
Apparent Power
Reactive Power
Frequency DC + 10Hz to 1MHz
Power Factor
Crest Factor
Ballast Mode
Up to 500kHz Output
Harmonics 0 to 50 Voltage and Current
THD 0 to 999%
Energy Whr
Inrush Current
0 to 90kW Direct Input
0 to 90kW Direct Input
90kW Direct Input
-1.000 to + 1.000
0 to 20.00
50/60/400Hz Input
Programmable Timer
Up to 100A peak
4 Parameter Measure Mode
14 Parameter Measure Mode
Low Power Standby
Impedance 5m to 5M
1 to 300 second window
Basic Accuracy 0.1% Reading +0.1% of Range
Measurement Selection
Modes of Operation - Applications
• Harmonics Display
• Energy Integrator
• Wave Form Display
• Standby Power
• Lighting Ballast
• Normal
• Inrush Current
Measurement Parameter
Harmonics and Distortion
Main menu, with modes selected
Modes with standby selected
• Up to the 50th harmonic
• Amplitude and phase from trouble-free DFT
• THD Total Harmonic Distortion
• Accurate DC measurements in the presence of AC
• 450kHz bandwidth for harmonics
• Harmonic bargraph display with cursor selection Harmonic barchart
Modes of Operation - Applications
Energy Measurement
The power consumption of everyday home and office
electrical appliances is of importance to consumers
and generators of electricity alike.
When the power consumption varies over time, then integration of the power ( W-hr integration) is required. The PM1000+ provides comprehensive integration features suitable for Energy Star measurements and for low-power measurements in accordance with international directives, eg. IEC 62301 which also requires crest factor measurements up to 8 and 50 harmonics.
• Precision Graphical Watt-Hour and VA-Hour
Measurements
• Clock
Waveform Display
• Display voltage & current waveforms
• Cursor readout: Volts RMS
& Amps RMS
Modes of Operation - Applications
Low-Power Standby
The first power analyzer developed with low power
measurements built-in as standard.
• Accurate low power measurements without special input channels or transducers.
• Special mode averages and captures power supplies in burst mode to provide
accurate measurements in the shortest possible time
• Current Range: From 10mA to 20A fully autoranging
• Crest Factor: (peak/rms) up to 20
Current
10mA 20mA
Burst Mode
Time 1 sec 2 sec
Lighting Ballasts
Special operating mode measures the output of electronic ballasts.
For 50Hz, 60Hz, 400Hz lighting systems
Voltech Ballast Current Transformer
• Isolates common mode switching voltages
• 5mA to 1A RMS in 2 ranges
• 5kHz to 1MHz bandwidth
• Accuracy (70 kHz to 500kHz) 1%
• Current phase better than 1
• Voltage phase better than 3
0
0
DC to
DC
conver tor
PFC
circuit
(if
required)
AC
Line
C
Regulated
DC
Outputs
DC Bus
AC Line Inp ut
V V
i i
Modes of Operation - Applications
Power Supplies
Ideal for measurements on power supplies,
from wall chargers to UPS and high-power
converters, the PM1000+ makes accurate
measurements on all waveforms including
those heavily distorted by the rectification and
smoothing at power supply inputs.
Measurements
W Input and Output power
Vrms Line regulation, drop-out
voltage, testing power fail
circuits
Arms Conductor and fuse rating
VA Apparent power for supply
rating
Apk MAX Inrush Current
Verification of inrush
limiting circuit design
Qualified fuse rating
PF Power Factor (W/ VA) for
verification of power factor
control circuits
A harm Amps harmonics for testing
to harmonic standards
A THD Distortion of input current
V THD Distortion of supply or
AC output
Integrator Low-power standby
measurements
Wiring Connections
Load
Single Phase Direct Connection up to 20Arms
Load
Single Phase Current Clamp Connection up to 3000Arms
Single Phase with Current Transformer Connection up to 1000 Arms
Load
C o n n e c t i o n s
Connectivity/Accessories
All interface options are standard features
Universal Power Input 85-264 VAC 45-65Hz 25VA
USB Peripheral
Aux/Trigger Port
(future release)
Clamp-on Current Transformers
CT1000 –Dual Ratio Precision Current Transformer
• Accuracy (23°C ± 5°C): ± 0.2% of specified ratio
• Frequency range: 45Hz to 1kHz
• Current range: 100:1 ratio: 10A to 120A rms 1000:1 ratio: 100A to 1200A rms
• Maximum input current: 1000A continuous 2000A for 1 hour
• Phase error (23°C ± 5°C): Better than ± 0.1° at 50Hz
Printer Port 25 Pin D Sub Connector
• Accuracy better than 1%
• Connect to the PM1000+ via safety leads and 1A shunt
• CL100 100A:1A ratio: 1A to 100VA rms range
• CL1000 1000A:1A ratio: 1A to 1000VA rms range
• CL3000 ratio: 1A to 3600VA rms range
RS232 9600, 38400, 19200 Baud Rates
IEEE 488.2
PS1000 - Inrush Switch
Solid-state switch for energizing loads (up to 200Apk) at either the peak or the zero crossing of AC voltage. Ideal for inrush current testing.
Ballast CT
Purposely designed for lighting applications, this device overcomes problems that are usually found when using conventional or Hall effect CTs.
• Convenient: No need to feed cables through a CT core.
• Better than 1% accuracy: Trifilar wound toroidal core.
• 5kHz to 1MHz bandwidth .
A c c e s s o r i e s
• 5mA to 1A measurement range
Specification
VOLTAGE RMS RANGES 900, 215, 46, 10 Vpk
Vrms + VDC Frequency range 10Hz to 1MHz
Peak continuous 1500 Vpk
Peak < 1 second 5000 Vpk
Input Impedance 1 M
Display 4.5 Digits
Crest Factor 20 (Peak/RMS)
Accuracy 0.1% of Reading + 0.1%of range + 5mV + 0.02%/kHz
VDC Accuracy 0.1% of Reading + 0.2% of range + 5mV + 0.02%/kHz
VOLTAGE +/- PEAK Accuracy 0.5% of Reading + 0.5% of Range + 0.02%/kHz
CURRENT RANGES 100, 25, 6.25, 1.6, 0.4, 0.1 Apk
RMS Frequency range 10Hz to 1MHz
ARMS Peak continuous 20Arms
Peak < 1 second Apk
Input resistance 12.5 m
Crest Factor 20 (Peak/RMS)
Accuracy 0.1% of Reading + 0.1% of range + 1mA + 0.02%/kHz
ADC Accuracy 0.1% of Reading + 0.2% of range + 1mA + 0.02%/kHz
CURRENT +/- PEAK Accuracy 0.5% of Reading + 0.5% of range + 0.02%/kHz
WATTS RANGES 1Wpk to 90kW
Frequency range 10Hz to 1MHz
Accuracy 0.2% of Reading + 0.1% of range + 5mW + (0.05/PF)%/kHz
VA RANGES 1 VA to 90VA
Frequency range 10Hz to 1MHz
Accuracy 0.2% Reading + 0.1 % of range +5mVA + 0.05%/kHz
VAr RANGES 1 VA to 90VAr
Frequency range 10Hz to 1MHz
Accuracy
POWER FACTOR Range +/- 0.000 to 1.000
Accuracy +/-(0.002+/-(0.001/PF)/kHz)
+ indicates leading PF and - indicates lagging PF
FREQUENCY Range DC and 10Hz to 1MHz
Accuracy 0.1%
0.2 of Reading + 0.1% of range + 5mVAr + (0.05/1-PF)%/kHz
Specification
VOLTAGE CREST FACTOR RANGE 1.00 to 20.0
Accuracy %Vpk error + % Vrms error
CURRENT CREST FACTOR RANGE 1.00 to 20.0
Accuracy %Apk error + % Arms error
PEAK INRUSH CURRENT RANGE 100Apk
Accuracy 2% of range +/- 20mA
HARMONIC ANAYLSIS Number of Voltage & 50
Current Harmonics
Maximum Harmonics 450kHz
Frequency
Accuracy 0.2% of Reading + 0.1% of Range +0.04% per
kHz of Harmonics
Frequency Range 10Hz to 450kHz
THD
Total Harmonic Distortion Range & Accuracy Range 0-999%
Accuracy 0.4% + 0.1%/kHz
Formula H2
STANDBY POWER Time Window 1-300 sec
Resolution 1 second
IMPEDANCE Range 0.005 to 1M
Accuracy 0.2% of Reading +0.1% of range
+5m
0.2% of Reading +0.1% of range
RESISTANCE Range 0.005 to 1M
Accuracy 0.2% of reading + 0.1% of range
+5m
REACTANCE Range 0.005 to 1M
Accuracy 0.2% of Reading + 0.1% of range
+5m
EXTERNAL SHUNT Input Range +/- 1250 mVpk
Scaling 0.0001 to 100000
MECHANICAL 1/2 rack size. Rack height 85mm
W = 224mm. Height including feet 103mm D = 285mm. Weight 3.21Kg (7lbs.)
thd =
2
+ H32 + H42 + H52 +...+H50
H1
+ (0.05/PF)%/kHz
+ (0.05/PF)%/kHz
+ (0.05/1-PF)%/kHz
2
Distributor
Voltech
www.voltech.com
Voltech Instruments Inc. 11637 Kelly Road, Suite 306 Fort Myers, FL 33908 USA Tel: (239) 437 0494 Fax: (239) 437 3841 Email sales@voltech.com
Voltech has an intensive program of design and development and reserves the right to alter product specification without notice. Although every care has been taken in compiling this information Voltech does not accept any liability for errors or omissions. ©2007 Voltech.
Voltech Instruments Ltd. 148 Harwell Int’l Business Centre Didcot, Oxon. OX11 0RA UK Tel: +44 (0) 1235 834555 Fax: +44 (0) 1235 835016 Email: sales@voltech.co.uk
P12 VPN 86-618/1