Voelker Kepler 5384 Series, Kepler 5384 Style, Kepler 5384 Select Instructions For Use Manual

Instructions for use
Model 5384 KeplerTM - Style - Select
Hyperlinks in interactive PDF files
Hyperlinks are cross-references in the text. They are marked with the sym­bol. The entries in the table of contents also have hyperlink functionality.
• Clicking on
takes you to the rele-
vant page.
Full-text search for a term
• Clicking
Suchen
opens the "Search"
window.
You can search for any term in this Acrobat document or optionally search all of the Acrobat files stored on your computer or those that are accessible.
Further interactive functions
The Adobe Reader and all Adobe Ac­robat program versions offer the expe­rienced user a whole range of highly useful functions for working interactively with documents. For further information, refer to the relevant user manual and/or online help.
Show the table of contents Jump to Help page (this page)
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
2 of 104
Help
Version, Imprint, Type labels
I
H
Instructions for use G155 Rev. 7 (03.2018) for Völker Bed 5384 Kepler Style - Select from year of construction 03/2016
© by Völker GmbH
Manufacturer: Völker GmbH
Wullener Feld 79, 58454 Witten GERMANY
Tel.: +49 2302 96096-0 Fax: +49 2302 96096-16 e-Mail: info@voelker.de Internet: www.voelker.de
All rights reserved. Reproduction – even in extracts – is prohibited.
We reserve the right to make changes due to technical development.
The contents of this document are sub­ject to change without prior notification.
Customers are advised to contact the responsible area sales representative before placing an order.
Type labels
Please observe the notes and instruc­tions for use for the mattresses!
The type labels are located on the in­side of the head panel.
Raise the back section to read the type labels.
The I
D No. is made up as follows:
5384-2016-12-0009
Model Year of
construc-
tion
Production
week
(calendar
Con-
secutive
number
week)
The type labels shown on this page are examples.
For further information on the type labels, see Appendix 83.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
3 of 104
Help
Table of contents 1/2
Help 2
Version, Imprint, Type labels 3 Table of contents 4 Foreword 6
Notes 7
General information 8 Designated purpose 9 General regulations, users training/
instruction, further requirements 11 General safety notes 12
Functional description 16
Overview of Style model 17 Overview of Select model 18 Manoeuvrable version (optional) 19 Hand control E2476 - Style model Hand control E2486 - Select model Hand control E2478 - Select model
(optional) 22 Mattresses 23 Trapeze bar and accessory holders 25
20
21
Securing systems 27
Rail inll panels 28
Versions and Options 29
Activation 32
General operating instructions 33 Preparation 34 Electrical activation 36 Use of the battery (optional) 37 Use of the battery and taking out of
service 38 Functional check 39
Operation 40
Guide castors - Roller carriage 4+2 41 Guide castors - Runner carriage 2+2
(optional) 42 Side rails | General safety notes 43 Continuous side rails -
Style model 44 Divided side rails - Select model 45 Hand control | Hand control
lock | Keeping safe 47
Hand control | Adjusting the upper and lower leg section 48
Hand control | Adjusting the back section 50
Hand control | Adjusting the lying surface height 51
Hand control | Setting the aid for getting out of bed 52
Hand control | Saving the lying surface height position 53
Hand control | Setting the Anti-Trendelenburg position 54
Hand control | Setting the Trendelenburg position 55
Hand control | Setting the position "Easy chair" 56
Hand control | Setting the position "Flat and horizontal" 57
Hand control | Under-bed light 58 Hand control | Functions of the
enable button 59 Rapid lowering of the back
section / CPR function
60
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
4 of 104
Help
Table of contents 2/2
Dynamic bed extension (optional for Select model) 61
Cleaning and disinfection 62
Cleaning 63 Wipe disinfection 64 Spray lances and automatic
washing systems 65
Maintenance 65
Personnel qualications, safety notes 66
Maintenance schedule 67
Technical checks 68
Visual inspection 69 Functional check of the side rails 69 Functional check of the brakes 71 Functional check of the drives 71 Power connection cable 71 Cabling 71 Housing 71
Mechanical check 71 Checking the grab handle 71 Further accessories 71
Troubleshooting 72
Table of faults 73 Service points 76
Appendix 77
Symbols used 78 Technical data (standard design) 80
Classication 82
Type labels 83 Service life / disposal 84 Manufacturer declarations, forms 85 Dimension sheets for side rails 86 Guidelines and manufacturer's
declaration – Electromagnetic compatibility 89
Available accessories for the Style and Select models 99
Requirements for electronic instructions for use 101
Trademarks 102
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
5 of 104
Help
Foreword
We are delighted that you have chosen Völker beds.
We are also grateful for the trust you have placed in our company and our products.
This step is undoubtedly the result of extensive considerations and examina­tions of the requirements that you have placed on new care or hospital beds based on your previous experiences.
You clearly have good reasons to choose a Völker bed.
We promise you: Völker beds will not disappoint you.
Völker beds do not carry a worldwide reputation as highly innovative medical aids without good reason. This refers not only to the design princi­ple, which was completely redeveloped by Völker. It also refers to the multitude of product advantages which were continually tested and improved for their practicality. These will support the comfort of the occupant/patient and also help reduce the load of daily care work.
We kept a modern design and effective functionality in mind when developing Völker beds. The beds are designed with a pleasant hotel look. The electronics allow com-
fortable adjustment of the bed using quiet motors.
Having purchased these beds, the re­sponsibility for their correct and appro­priate use is transferred to you. Con­sequently, we strongly advise that you consult the enclosed instructions for use to learn about the technical features, handling and use of all the functions.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
6 of 104
Help
Notes
The Notes section contains informa­tion on the designated purpose of the product, as well as general safety notes.
CONTENTS
General information 8 Designated purpose 9 General regulations, training/instruc-
tion of users, further requirements 11 General safety notes 12
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
7 of 104
Help
Notes | General information
General information
You have purchased a bed from Völker GmbH. This bed has been built in ac­cordance with the applicable national and international standards and regu­lations reflecting the current state of technology.
Völker beds satisfy requirements in terms of safety and functionality. They are tested in accordance with interna­tional standards and bear the CE mark, which documents the beds' compliance with essential requirements for medical products (DIN EN 60601-1 and DIN EN 60601-2-52).
Please read the general safety notes
12. Please observe (with particular
attention to any warranty claims) the fur­ther instructions on the following pages.
Variants
The bed can be designed in different variants. A description of these variants can be found in the "Versions" section
29.
You can also find information on dif­ferent operating variants or different technical data for the models at various points in these instructions for use. You can recognise the variant of your bed based on the type label 83.
Standard design
The standard design of the bed can be supplied with various options. A descrip­tion of these options can be found in the "Versions and Options" section 29.
Copyright protection
These instructions for use may only be transferred to third parties with the written consent of Völker GmbH. All documents are protected under the Copyright Act.
Guarantee and liability
Völker beds are supplied with a compre­hensive guarantee that is described in detail in the order confirmation. For more extensive information please contact Völker.
We reserve the right to make techni­cal modifications without notice as part of the further development of the beds which form the subject of these instruc­tions for use. All specifications are non binding. Printing errors are reserved.
We accept no liability for damage and operational faults caused as a result of misuse and/or non-observance of these instructions for use.
The portrayal of the accessories does not necessarily match the actual product.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
8 of 104
Help
Notes | Designated purpose 1/2
17
Intended use
The Völker beds 5384 Kepler are beds for medical use and are intended for the laying down and care of adult occu­pants/patients in care institutions and retirement homes (Application environ­ment 3 as per DIN EN 60601-2-52).
The application environment 3 involves long-term care in a medical sector in which medical surveillance is required and supervision is provided if nec­essary. This includes retirement and nursing homes, rehabilitation facilities and geriatric facilities.
The bed is intended for use by people over 12 years of age and with a height of 146 cm to 185 cm, as well as a body weight of 40 kg to 195 kg. Their Body Mass Index (BMI) must be over 17.
The Body Mass Index is an index value derived from body weight and height.
The BMI is calculated from the formula:
Body weight kg BMI = Height2
≥ 40 kg
≥ 146 cm
BMI ≥
m2
The safe working load of the bed is 230 kg. To calculate the maximum pa­tient weight, DIN EN 60601-2-52:2010 stipulates subtracting from the safe working load, when using the bed in application environment 3, 20 kg for the weight of the mattress and 15 kg for accessories as well as the load that is borne by the accessories.
Any use of the Völker bed extending beyond this intended use is excluded from a potential liability.
The safe working load and the maxi­mum patient weight can be taken from the following table:
Model
5384 Kepler 230 kg 195 kg
Safe
working load
of the bed
Maximum
patient weight
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
9 of 104
Help
Notes | Designated purpose 2/2
Inappropriate use
Inappropriate use can be dangerous.
This includes, but is not limited to:
• Incorrect actuation of electrical func­tions and uncontrolled positioning
• Operation of the bed by the occupant/ patient without having received prior instruction in how to do so
• Use of other electrical equipment on the bed
• Pulling on cables to move the bed
• Removing electrical plug connections by pulling on the cable
• Use of the bed on a slope of more than six degrees of inclination (the bed's brakes are designed for an an­gle of inclination of no more than ten degrees)
• Any attempt to move the bed while it is in braked position
• Use of the bed to transport people
• Use of the bed for transport within a vehicle
• Overloading of the bed beyond the specified safe working load
CAUTION The bed may not be
used for patients/occupants under 12 years of age or who are less than 146 cm tall.
WARNING The bed may not be
placed right next to or stacked up with other equipment. Should the operation be near to or stacked up with other equipment be nec­essary, then it must be ensured that the operation of the bed is observed, and correct use in this arrangement is checked.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
10 of 104
Help
Notes | General regulations, user training / instruction, further requirements
General Requirements
The bed must only be operated and used in accordance with its designated purpose, in line with the applicable conditions, the generally acknowledged rules of technology and the occupa­tional safety and accident prevention guidelines. The bed must not be oper­ated in a faulty state that could endan­ger its occupant/patient, care personnel or third parties.
User training
The bed may only be operated by indi­viduals whose training or understanding and experience offer surety for correct handling.
User instruction
The thorough induction of care person­nel in the operation of the bed can be provided by Völker or its representatives at the customer's request. Attendance of such training can be cert­ified and confirmed by Völker using the
form provided for this purpose, specify­ing the name, date and signature. Occupants/patients must be instructed in the use of the bed before care per­sonnel can hand over control to them.
Other requirements
Whoever is in charge of the activation, operation or preparation of the bed must have been given a copy of these instructions for use (in printed or elec­tronic form) and have read them.
To avoid operating errors and to safe­guard the smooth operation of the bed, care personnel must always have access to the safety notes below.
Flooring requirements
Increasingly overweight patients and occupants have caused a consistent increase in demands on hospital and healthcare beds in recent years. Völker has addressed this topic by increasing the "safe working load" for the beds. The higher strain requirements are not
limited to beds only, they apply also to static and floor requirements.
We therefore recommend using flooring that is designed for these strains in areas where the beds are used. This is flooring which is classified and properly laid according to DIN EN 685 minimum class 32 or 33. This is flooring for areas intended for public and industrial use with moderate or heavy traffic.
Floors corresponding to the above classifications can be marked with the following logo:
33
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
11 of 104
Help
Notes | General safety notes 1/4
Warning symbols
Information marked with this symbol must always be read and strictly observed.
CAUTION represents potential-
ly dangerous situations that can cause slight physical injuries.
DANGER represents an immediate
threat of danger that will cause serious physical injury or death.
WARNING represents potentially
dangerous situations that can lead to serious physical injury or death.
NOTE warns of potential damage
to objects or property.
Before first activation
Care personnel must read these instruc­tions for use carefully and in full before the bed is activated for the first time.
Before the bed is activated, care per­sonnel must be instructed in the han­dling of the bed using the instructions for use. In addition, the potential risks which may arise despite the proper op­eration of the bed must also be pointed out in full.
Before and during use
Before each use of the bed, the user must be sure that the bed is in a proper condition and that safe use is ensured (Functional check 71).
WARNING If other devices are
operated on the bed, which are provided with cables, air hoses or similar, make sure that these lines are routed so that they cannot become jammed in the moving parts of the bed or be damaged.
Position of the beds
CAUTION To avoid injuries
caused by falling, we recommend (except while care is being given) that the bed is generally set at its lowest position with the guide castors fixed when the patient is asleep in the bed.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
12 of 104
Help
Notes | General safety notes 2/4
Transporting the bed
CAUTION The bed is not designed
for transporting people. When moving the bed, it must always be ensured that the main connection cable does not touch the ground and that the lying surface height is at least 35 cm. The bed should only be moved over a solid surface. Never attempt to push it over ob­stacles of over 1 cm in height. The maximum angle of inclination of the floor must not exceed 6°.
Securing the bed
CAUTION "Risk of accident"
If the bed is not moved, the guide castors must always be fixed and latched into place, if necessary, since the bed may be required as a support for when the occupant/ patient stands up or lies down. If the bed rolls away without the cas­tors being braked, this can lead to a serious fall. Once the castors have been applied, check that the bed is actually properly parked.
The bed sometimes finds itself
in an unbraked position after each activation or reactivation and must therefore be checked subsequently to ensure the castors are fixed properly.
NOTE Please note that the brakes
are most effective only on dry, clean and slip-resistant flooring.
One-sided load on the bed
NOTE In order to prevent one-sid-
ed loads on the bed, it must not be used as a seat for per­sons other than the occupant/ patient (i.e. visitors must not sit on the edge of the bed).
Side rails
WARNING "Risk of entrapment"
In the case of occupants/patients whose physical or mental condition makes it necessary to use side rails to protect them from falling out of bed, the following safety measures must be observed:
The legal permissibility of using side rails must be ascertained.
The side rails may only be operated by trained care personnel.
It must be ensured that the occupant/patient does not come into contact with the side rail
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
13 of 104
Help
Notes | General safety notes 3/4
elements upon activation of the electrical lying surface adjustment. It is also important to ensure that no part of the body is sticking out through the side rails.
Ensure that the side rails (or parts thereof) are either fully raised and locked in position or completely lowered.
If the side rails are used with a person whose psychological condition makes their use necessary, then it must be ensured that the hand control is kept out of his/her reach or its functions are locked. It is also strongly recommended that side rail protection covers are used.
DANGER If these safety meas-
ures are not observed by care personnel, injuries can be caused to hands, knees, fingers, feet, legs and hips, along with haematomas and other injuries as a result of entrapment. With children or people who are less than 146 cm tall, non­observance of these guide­lines can lead to death!
Height adjustment
DANGER "Risk of entrapment
between the lower frame and/or floor and the bed frame when the bed is lowered"
It must be ensured that no peo­ple, limbs, pets, bed linen or other objects are caught between the bed frame and the lower frame and/or floor during adjustment.
DANGER "Danger of movement"
If any movement of the bed could represent a danger to the occupant /patient, all functions must be locked.
Accessories
WARNING
"Risk of injury"
Only original Völker accessories should be used! Third-party accessories must undergo testing by the operator before use.
Use of lifting devices
The lifting devices specified are appliances that can be attached to the bed for transport purposes. Patient lifters can be used.
WARNING "Risk of injury"
No lifting device must be fastened directly to the bed (patient trans- port, repair).
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
14 of 104
Help
Notes | General safety notes 4/4
Use of oxygen equipment
DANGER "Risk of fire"
Do not use any oxygen equipment other than that which is adminis­tered via nasal prongs or masks. Never use the bed in an oxygen tent or in potentially explosive atmospheres (where flammable gases or vapours are present).
Provided it has been excluded (e.g. based on information in the instructions for use of the equipment being used) that the use of the equipment or a fault may cause the O2 concentration to rise to such a degree that there is an explo­sion risk, then the equipment can be used.
Rail infill panels
When using the rail infill panel, please read the separate instructions for use for this accessory. During technical checks, the rail infill panels should also be checked to ensure they are suitable for the size of side rail used.
Cleaning and disinfection
In order to maintain consistent function­ing, the bed should be cleaned, disin­fected and tested as soon as possible following each use, so that it can be reused immediately without risk.
Incorrect cleaning/disinfection of the bed can cause danger.
Maintenance and repair
Anyone responsible for carrying out maintenance and repair work must at least have received instruction from Völker in the service tools, have read the safety notes and the service manual and be qualified in accordance with the law on medical products § 4.
After maintenance work or repairs have been carried out, a technical check must be carried out on the affected parts and/or functions 71. During this check, it must be determined whether the bed can be used in accord­ance with the specifications without risk to the occupant/patient, user or a third party.
A technical check must be carried out at least once a year and after every lengthy period of non-use.
Any discernible damage, such as signs of wear and tear, loose screws or breaks/fractures must be eliminated immediately.
Electromagnetic and electrostatic interferences
The beds in model series 5384 Kepler satisfy the EMC* requirements in accord­ance with the law on medical products (MPG). The basis for testing is standard EN 60601-1-2.
* Law concerning the Electromagnetic Compatibility of Operating Equipment
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
15 of 104
Help
Functional description
The features of the Völker care bed and its function are set out in the
Functional description section .
CONTENTS
Overview of Style model 17 Overview of Select model 18 Overview of manoeuvrable version 19 Hand control E2476 20 Hand control E2486 21 Hand control E2478 (optional) 22 Accessories - Mattresses 23 Accessories - Trapeze bar
and accessory holders 25 Securing systems 27 Accessories - Rail inll panels 28 Versions and Options 29
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
16 of 104
Help
Functional description | Overview of Style model
Plastic lying surface (standard for Style model)
Foot section
Braking guide castors with foot pedal (both sides)
Continuous side rails, both sides
Fixed seat section
Adjustable back section
Head section
Holders for trapeze bars, drip stands etc., both sides
Rapid lowering of the back section (CPR) below the seat area, both sides
Adjustable lower leg section
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
Adjustable upper leg section
17 of 104
Help
Functional description | Overview of Select model
Adjustable side rails for lower leg section, both sides
Foot section
Dynamic bed extension (optional Select model)
Braking guide castors with foot pedal (both sides)
MiS® lying surface (standard for Select model, optional for Style model)
Fixed seat section
Adjustable back section
Adjustable side rails
for back section,
both sides
Head section
Holders for trapeze bars, drip stands etc., both sides
Rapid lowering of the back section (CPR), both sides
Adjustable lower leg section
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
18 of 104
Adjustable upper leg section
Help
Functional description | Manoeuvrable version (optional)
The maneuverable version of both models (option) is fitted with two steering castors at both the head and foot ends.
(optional for Select model)
Foot pedal (both sides)
HPL lying surface
Braking guide castors (both sides)
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
19 of 104
Help
Functional description | Hand control E2476 -
Upper leg section up
Upper leg section
down
Back section up
Back section down
Style
model
WARNING When actuating mo-
torised adjustments with the side rails raised, ensure that the oc­cupant/patient does not have any contact with the side rails. Ensure that no parts of his or her body, or those of other people, are stick­ing through the side rails or are trapped between the lying surface and the lower frame and/or floor!
Lying surface up
Lying surface down
3
Aid for getting out of bed/ Position saving
1
Anti-Trendelenburg positioning
2
Enable button
Hand control lock
1, 2, 3
For use of the enable button, please refer to the
note on page  59
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
20 of 104
Help
Functional description | Hand control E2486 - Select model (optional Style model)
WARNING When actuating mo-
torised adjustments with the side
Upper leg section up
Upper leg section
down
Lying surface up
Back section down
Back section up
Aid for getting out of bed and save function
1
rails raised, ensure that the oc­cupant/patient does not have any contact with the side rails. Ensure that no parts of his or her body, or those of other people, are stick­ing through the side rails or are trapped between the lying surface and the lower frame and/or floor!
Lying surface down
Sitting position
Under-bed light
4
5
Anti-Trendelenburg positioning
2
Trendelenburg positioning
3
Enable button
Hand control lock
1, 2, 3, 4, 5
For use of the enable button, please refer to the
note on page  59
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
21 of 104
Help
Functional description | Hand control E2478 (optional)
Upper leg section up
Upper leg section
down
Back section down
Back section up
This hand control is available as an option for both models, only in con­junction with hand control E2486.
Lying surface up
Lying surface
down
Under-bed light
Hand control lock
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
Aid for getting out of bed
22 of 104
WARNING When actuating mo-
torised adjustments with the side rails raised. Ensure that the oc­cupant/patient does not have any contact with the side rails, and that no parts of his or her body or those of other people are stick­ing through the side rails or are trapped between the lying surface and the lower frame and/or floor.
Help
Functional description | Accessories 1/6 Mattresses
Mattresses
To offer the greatest degree of flexibility, Völker offers a wide range of easy­to-attach accessories. The beds are equipped as standard with holder devic­es for accessories, such as drip stands and trapeze bars.
You can find further information regard­ing the accessories in our current infor­mation brochures or on the Internet at www.voelker.de . Our staff will happily inform you about available accessories for your bed model.
To minimise the risk of injury, only use mattresses with the dimensions and properties described in the following. If you do not use Völker mattresses, please contact a dealer in whom you have confidence.
WARNING Only original Völk-
er accessories may be used!
DANGER If mattresses are used
that do not match these specifica­tions, there is a risk of suffocation!
Please read the relevant instructions for use before using the mattresses.
NOTE Air-filled mattresses are not compatible with dynamic bed extension!
Ensure that the mattresses are placed on the bed, aligned with the lying sur­face and securely attached as specified in the regulations.
The tubes and cables of the air-filled mattress must be positioned in such a way that they are not kinked, squashed or otherwise damaged by the moving parts of the bed.
To reduce the length of the power cable, the mattress system should be con­nected to the nearest possible socket so that the cable does not lie on the floor.
are not damaged during transport, e.g. by rolling over.
Cables and tubes of the air-filled mat­tresses must be checked regularly for damage.
Mattresses with damaged cables and/or tubes may not be used!
Typical representation for Völker mattress NP50 Monodensity.
Before transporting the beds, the cables and tubes of the air-filled mattresses must be secured to the bed so that they
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
23 of 104
Help
Functional description | Accessories 2/6 Mattresses
Model Surface material
Order no.
W x L x H (cm)
Dimensions
Foam mattresses
Völker NP50 ASS086 Polyurethane 88 x 200 x 12 Völker NP100 ASS087 Polyurethane 88 x 200 x 12 Völker NP150 ASS088 Polyurethane 88 x 200 x 12
Tolerance values: Length = ± 2 cm, width ± 1.5 cm, thickness = ± 1 cm
DANGER The height of the raised
side rail above the foam mattress must always be greater than or equal to 220 mm; otherwise the occupant/patient may acciden-
≥ 220 mm!
(Foam mattress in flat position)
tally fall out of bed. Please note that the height of the mattress has a direct influence on this.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
24 of 104
Help
Functional description | Accessories 3/6 Trapeze bar and accessory holders
Holders for the trapeze bar and acces­sories are located on the inside of the head panel. The trapeze bar and other accessories must slide into the holders until you hear them click into place.
Trapeze bar holders ø 34.2 mm
The trapeze bar is inserted into the holder and latched into place.
WARNING Only use the the tra -
peze bars listed in the accessories list!
WARNING "Risk of Injury"
Ensure that the trapeze bar is completely slotted into the hold­er and securely seated. Please note: The safe working load of the trapeze bar is maximum 75 kg.
WARNING "Risk of falling"
The trapeze bar must not be used by the occupant/patient as a means for climbing into the bed. The tra­peze bar must never jut out beyond the outer edge of the bed and then be used as an aid to pull oneself up (e.g. when getting out of a wheelchair).
WARNING "Risk of injury"
The trapeze bar may not be used by the patient/occupant without supervision.
Similar pictures
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
25 of 104
Help
Functional description | Accessories 4/6 Trapeze bar and accessory holders
ø 13 mm
The trapeze bar holders can be provid­ed with an adapter (ø 13 mm) for a drip stand or for other accessories.
The adapter is inserted into one of the trapeze bar holders.
The drip stand or the accessory is fastened from below with a sleeve nut.
WARNING The drip stand must
always be aligned towards the bed, as shown below, and may not point outwards.
Should the bed be moved or adjusted, the infusion tubing or cables must be closely monitored by the care personnel.
Please also note that drainage devices are able to touch the floor when lowering the bed. This also applies for the Anti-Trendelenburg and Trendelen­burg position.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
26 of 104
Help
Functional description | Accessories 5/6 Securing systems - Rail infill panels
The rail infill panels close the middle gap between the divided side rails on Völker beds so as to enable a continu­ous side rail solution.
The rail infill panels are inserted in the insertion sleeves of the upper leg section.
Check whether the pendulum locking is completely latched into place by at­tempting to pull out the rail infill panel upwards. If you succeed in doing this, press the rail infill panel fully down
WARNING Make sure that the side
rails are fully set up and latched into place. When activating the electric lying surface adjustment or side rails, it must be ensured that the patient/occupant has neither contact with the side rails, nor any part of the body protrudes through the side rails. We strong­ly recommend that you lock the functions of the hand control.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
again and adjust the pendulum until the rail infill panel can no longer be pulled out.
27 of 104
To remove the rail infill panel from the holder, move the pendulum into a verti­cal position and pull the rail infill panel up and out at the same time.
NOTE Please observe the detailed
instructions for use for the rail infill panels.
Further accessories can be found on pages 99 and 100.
Help
Functional description | Accessories 6/6 Securing systems
Use of securing systems
Securing systems such as belts should only be used exactly as specified by the manufacturer.
If securing systems in the form of ab­dominal belts are used, then it must be ensured that the side rails are complete­ly raised. The gap in the middle of the rails must always be closed using a rail infill panel with the Select model.
Devices for safeguarding patients may not be regarded as a substitute for the requisite supervision of the patient/oc­cupant by the care personnel.
Even if fastened correctly, means of safeguarding patients can become tangled and injure the patient/occupant or even cause the death of the patient/ occupant if the latter is agitated and confused.
When using safeguarding devices, the patient/occupant must be supervised corresponding to the statutory regulations and the applicable protocol of the facility. Failure to observe this note can lead to personal injury or material damage.
DANGER When using securing
systems and rail infill pan­els, please note the separate instructions for use pertain­ing to these accessories.
If the patient is restrained with the
help of restraint holders (optional for both models), the lying surfaces must never be adjusted while the patient is restrained and must always be in the lowest position!
The lying surface adjustment
functions must be locked when a patient is restrained, and the hand control must be kept out of the reach of the occupant/patient!
Restraint holder
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
28 of 104
Help
Functional description | Versions and Options 1/3
These instructions for use cover all of the versions and options listed. Precise information on the bed versions sup­plied can be found in the order specifi­cations for your bed. If the original bed specification is no longer available, please contact the Völker Customer Services. Please make a note of the serial num­ber (ID No.) on the type label 83 before you call.
Wall protector wheels
Vertical wall protector wheels are op­tionally available for both models.
Linen holder
A pull-out linen holder at the foot of the bed is optionally available for both models.
The linen holder should always be pushed in to avoid causing damage when not in use.
WARNING The linen holder may
not be used as a seat or step! The safe working load of the linen holder is 20 kg.
Accessories rail
An accessories rail is readily available for both models as an extra option.
NOTE If the lying surface is lowered below 35 cm, any object that has been attached on the accessories rail must be removed!
Under-bed light
A under-bed light is available as an option for Style model.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
29 of 104
Help
Functional description | Versions and Options 2/3
MiS® lying surface
Völker MiS® is a support system where the different elements register and promote the occupant‘s/patient‘s own movement.
Thanks to the large number of contact points, the occupant/patient is sup­ported in an anatomically correct man­ner, which also has a positive influence on sleeping behavior.
HPL lying surface
The four-part HPL lying surface (HPL = high-pressure laminate) consists of moisture-resistant high-pressure laminate.
The four elements are well ventilated via the round cut-outs, are individually removable and fulfil the highest hygienic requirements.
The HPL lying surfaces can be easily removed and can therefore be cleaned quickly and thoroughly. They also facili­tate the use of therapy mattresses.
Dynamic bed extension
The Select model can be equipped with a telescopic bed extension. Extension: approx. 28 cm.
An additional mattress extension (PA2215) must be used when using the dynamic bed extension.
Battery
A battery is available for both bed models as an option.
The battery allows the bed to be operat­ed independent of the power supply for at least two adjustment cycles under full load.
Variants divided sidesails
Select model is available with the fol­lowing siderail variants:
Extendable 34 - 35,5* cm (standard) Extendable 43,5 - 45* cm (option)
* Dimensions are from the upper edge of the side rail to the lying surface (without mattress). The different extension heights are dependent on the lying surface (HPL or MiS®), see also p. 86 - 87.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
30 of 104
Help
Loading...
+ 74 hidden pages