Hyperlinks are cross-references in the
text. They are marked with the symbol. The entries in the table of contents
also have hyperlink functionality.
• Clicking on
takes you to the rele-
vant page.
Full-text search for a term
• Clicking
Suchen
opens the "Search"
window.
You can search for any term in this
Acrobat document or optionally search
all of the Acrobat files stored on your
computer or those that are accessible.
Further interactive functions
The Adobe Reader and all Adobe Acrobat program versions offer the experienced user a whole range of highly
useful functions for working interactively
with documents. For further information,
refer to the relevant user manual and/or
online help.
Show the table of contentsJump to Help page (this page)
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
2 of 104
Help
Version, Imprint, Type labels
I
H
Instructions for use G155
Rev. 7 (03.2018)
for Völker Bed 5384 Kepler
Style - Select
from year of construction 03/2016
All rights reserved. Reproduction – even
in extracts – is prohibited.
We reserve the right to make changes
due to technical development.
The contents of this document are subject to change without prior notification.
Customers are advised to contact the
responsible area sales representative
before placing an order.
Type labels
Please observe the notes and instructions for use for the mattresses!
The type labels are located on the inside of the head panel.
Raise the back section to read the type
labels.
The I
D No. is made up as follows:
5384-2016-12-0009
ModelYear of
construc-
tion
Production
week
(calendar
Con-
secutive
number
week)
The type labels shown on this page are
examples.
For further information on the type
labels, see Appendix 83.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
3 of 104
Help
Table of contents 1/2
Help 2
Version, Imprint, Type labels 3
Table of contents 4
Foreword 6
Notes 7
General information 8
Designated purpose 9
General regulations, users training/
instruction, further requirements 11
General safety notes 12
Functional description 16
Overview of Style model 17
Overview of Select model 18
Manoeuvrable version (optional) 19
Hand control E2476 - Style model
Hand control E2486 - Select model
Hand control E2478 - Select model
(optional) 22
Mattresses 23
Trapeze bar and accessory holders 25
20
21
Securing systems 27
Rail inll panels 28
Versions and Options 29
Activation 32
General operating instructions 33
Preparation 34
Electrical activation 36
Use of thebattery(optional) 37
Use of thebatteryand taking out of
(optional) 42
Side rails | General safety notes 43
Continuous side rails -
Style model 44
Divided side rails - Select model 45
Hand control | Hand control
lock | Keeping safe 47
Hand control | Adjusting the
upper and lower leg section 48
Hand control | Adjusting the
back section 50
Hand control | Adjusting the
lying surface height 51
Hand control | Setting the
aid for getting out of bed 52
Hand control | Saving the
lying surface height position 53
Hand control | Setting the
Anti-Trendelenburg position 54
Hand control | Setting the
Trendelenburg position 55
Hand control | Setting the
position "Easy chair" 56
Hand control | Setting the
position "Flat and horizontal" 57
Hand control | Under-bed light 58
Hand control | Functions of the
enable button 59
Rapid lowering of the back
section / CPR function
60
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
4 of 104
Help
Table of contents 2/2
Dynamic bed extension
(optional for Select model) 61
Cleaning and disinfection 62
Cleaning 63
Wipe disinfection 64
Spray lances and automatic
washing systems 65
Maintenance 65
Personnel qualications,
safety notes 66
Maintenance schedule 67
Technical checks 68
Visual inspection 69
Functional check of the side rails 69
Functional check of the brakes 71
Functional check of the drives 71
Power connection cable 71
Cabling 71
Housing 71
Mechanical check 71
Checking the grab handle 71
Further accessories 71
Troubleshooting 72
Table of faults 73
Service points 76
Appendix 77
Symbols used 78
Technical data (standard design) 80
Classication 82
Type labels 83
Service life / disposal 84
Manufacturer declarations, forms 85
Dimension sheets for side rails 86
Guidelines and manufacturer's
declaration – Electromagnetic
compatibility 89
Available accessories for the
Style and Select models 99
Requirements for electronic
instructions for use 101
Trademarks 102
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
5 of 104
Help
Foreword
We are delighted that you have chosen
Völker beds.
We are also grateful for the trust you
have placed in our company and our
products.
This step is undoubtedly the result of
extensive considerations and examinations of the requirements that you have
placed on new care or hospital beds
based on your previous experiences.
You clearly have good reasons to
choose a Völker bed.
We promise you: Völker beds will not
disappoint you.
Völker beds do not carry a worldwide
reputation as highly innovative medical
aids without good reason.
This refers not only to the design principle, which was completely redeveloped
by Völker. It also refers to the multitude
of product advantages which were
continually tested and improved for
their practicality. These will support the
comfort of the occupant/patient and also
help reduce the load of daily care work.
We kept a modern design and effective
functionality in mind when developing
Völker beds.
The beds are designed with a pleasant
hotel look. The electronics allow com-
fortable adjustment of the bed using
quiet motors.
Having purchased these beds, the responsibility for their correct and appropriate use is transferred to you. Consequently, we strongly advise that you
consult the enclosed instructions for use
to learn about the technical features,
handling and use of all the functions.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
6 of 104
Help
Notes
The Notes section contains information on the designated purpose of
the product, as well as general safety
notes.
CONTENTS
General information 8
Designated purpose 9
General regulations, training/instruc-
tion of users, further requirements 11
General safety notes 12
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
7 of 104
Help
Notes | General information
General information
You have purchased a bed from Völker
GmbH. This bed has been built in accordance with the applicable national
and international standards and regulations reflecting the current state of
technology.
Völker beds satisfy requirements in
terms of safety and functionality. They
are tested in accordance with international standards and bear the CE mark,
which documents the beds' compliance
with essential requirements for medical
products (DIN EN 60601-1 and
DIN EN 60601-2-52).
Please read the general safety notes
12. Please observe (with particular
attention to any warranty claims) the further instructions on the following pages.
Variants
The bed can be designed in different
variants. A description of these variants
can be found in the "Versions" section
29.
You can also find information on different operating variants or different
technical data for the models at various
points in these instructions for use. You
can recognise the variant of your bed
based on the type label 83.
Standard design
The standard design of the bed can be
supplied with various options. A description of these options can be found in the
"Versions and Options" section 29.
Copyright protection
These instructions for use may only
be transferred to third parties with the
written consent of Völker GmbH.
All documents are protected under the
Copyright Act.
Guarantee and liability
Völker beds are supplied with a comprehensive guarantee that is described in
detail in the order confirmation.
For more extensive information please
contact Völker.
We reserve the right to make technical modifications without notice as part
of the further development of the beds
which form the subject of these instructions for use.
All specifications are non binding.
Printing errors are reserved.
We accept no liability for damage and
operational faults caused as a result of
misuse and/or non-observance of these
instructions for use.
The portrayal of the accessories
does not necessarily match the actual
product.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
8 of 104
Help
Notes | Designated purpose 1/2
17
Intended use
The Völker beds 5384 Kepler are beds
for medical use and are intended for the
laying down and care of adult occupants/patients in care institutions and
retirement homes (Application environment 3 as per DIN EN 60601-2-52).
The application environment 3 involves
long-term care in a medical sector in
which medical surveillance is required
and supervision is provided if necessary. This includes retirement and
nursing homes, rehabilitation facilities
and geriatric facilities.
The bed is intended for use by people
over 12 years of age and with a height
of 146 cm to 185 cm, as well as a body
weight of 40 kg to 195 kg. Their Body
Mass Index (BMI) must be over 17.
The Body Mass Index is an index value
derived from body weight and height.
The BMI is calculated from the formula:
Body weight kg
BMI =
Height2
≥ 40 kg
≥ 146 cm
BMI ≥
m2
The safe working load of the bed is
230 kg. To calculate the maximum patient weight, DIN EN 60601-2-52:2010
stipulates subtracting from the safe
working load, when using the bed in
application environment 3, 20 kg for the
weight of the mattress and 15 kg for
accessories as well as the load that is
borne by the accessories.
Any use of the Völker bed extending
beyond this intended use is excluded
from a potential liability.
The safe working load and the maximum patient weight can be taken from
the following table:
Model
5384 Kepler230 kg195 kg
Safe
working load
of the bed
Maximum
patient
weight
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
9 of 104
Help
Notes | Designated purpose 2/2
Inappropriate use
Inappropriate use can be dangerous.
This includes, but is not limited to:
• Incorrect actuation of electrical functions and uncontrolled positioning
• Operation of the bed by the occupant/
patient without having received prior
instruction in how to do so
• Use of other electrical equipment on
the bed
• Pulling on cables to move the bed
• Removing electrical plug connections
by pulling on the cable
• Use of the bed on a slope of more
than six degrees of inclination (the
bed's brakes are designed for an angle of inclination of no more than ten
degrees)
• Any attempt to move the bed while it
is in braked position
• Use of the bed to transport people
• Use of the bed for transport within a
vehicle
• Overloading of the bed beyond the
specified safe working load
CAUTION The bed may not be
used for patients/occupants under
12 years of age or who are less
than 146 cm tall.
WARNING The bed may not be
placed right next to or stacked up
with other equipment. Should the
operation be near to or stacked
up with other equipment be necessary, then it must be ensured
that the operation of the bed is
observed, and correct use in
this arrangement is checked.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
10 of 104
Help
Notes | General regulations, user training / instruction, further requirements
General Requirements
The bed must only be operated and
used in accordance with its designated
purpose, in line with the applicable
conditions, the generally acknowledged
rules of technology and the occupational safety and accident prevention
guidelines. The bed must not be operated in a faulty state that could endanger its occupant/patient, care personnel
or third parties.
User training
The bed may only be operated by individuals whose training or understanding
and experience offer surety for correct
handling.
User instruction
The thorough induction of care personnel in the operation of the bed can be
provided by Völker or its representatives
at the customer's request.
Attendance of such training can be certified and confirmed by Völker using the
form provided for this purpose, specifying the name, date and signature.
Occupants/patients must be instructed
in the use of the bed before care personnel can hand over control to them.
Other requirements
Whoever is in charge of the activation,
operation or preparation of the bed
must have been given a copy of these
instructions for use (in printed or electronic form) and have read them.
To avoid operating errors and to safeguard the smooth operation of the
bed, care personnel must always have
access to the safety notes below.
Flooring requirements
Increasingly overweight patients and
occupants have caused a consistent
increase in demands on hospital and
healthcare beds in recent years. Völker
has addressed this topic by increasing
the "safe working load" for the beds.
The higher strain requirements are not
limited to beds only, they apply also to
static and floor requirements.
We therefore recommend using flooring
that is designed for these strains in
areas where the beds are used. This is
flooring which is classified and properly
laid according to DIN EN 685 minimum
class 32 or 33. This is flooring for areas
intended for public and industrial use
with moderate or heavy traffic.
Floors corresponding to the above
classifications can be marked with the
following logo:
33
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
11 of 104
Help
Notes | General safety notes 1/4
Warning symbols
Information marked with this
symbol must always be read
and strictly observed.
CAUTION represents potential-
ly dangerous situations that can
cause slight physical injuries.
DANGER represents an immediate
threat of danger that will cause
serious physical injury or death.
WARNING represents potentially
dangerous situations that can lead
to serious physical injury or death.
NOTE warns of potential damage
to objects or property.
Before first activation
Care personnel must read these instructions for use carefully and in full before
the bed is activated for the first time.
Before the bed is activated, care personnel must be instructed in the handling of the bed using the instructions
for use. In addition, the potential risks
which may arise despite the proper operation of the bed must also be pointed
out in full.
Before and during use
Before each use of the bed, the user
must be sure that the bed is in a proper
condition and that safe use is ensured
(Functional check 71).
WARNING If other devices are
operated on the bed, which are
provided with cables, air hoses or
similar, make sure that these lines
are routed so that they cannot
become jammed in the moving
parts of the bed or be damaged.
Position of the beds
CAUTION To avoid injuries
caused by falling, we recommend
(except while care is being given)
that the bed is generally set at its
lowest position with the guide
castors fixed when the patient is
asleep in the bed.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
12 of 104
Help
Notes | General safety notes 2/4
Transporting the bed
CAUTION The bed is not designed
for transporting people. When
moving the bed, it must always be
ensured that the main connection
cable does not touch the ground
and that the lying surface height is
at least 35 cm.
The bed should only be moved
over a solid surface.
Never attempt to push it over obstacles of over 1 cm in height.
The maximum angle of inclination
of the floor must not exceed 6°.
Securing the bed
CAUTION "Risk of accident"
If the bed is not moved, the guide
castors must always be fixed and
latched into place, if necessary,
since the bed may be required as
a support for when the occupant/
patient stands up or lies down. If
the bed rolls away without the castors being braked, this can lead to a
serious fall.
Once the castors have been
applied, check that the bed is
actually properly parked.
The bed sometimes finds itself
in an unbraked position after
each activation or reactivation
and must therefore be checked
subsequently to ensure the
castors are fixed properly.
NOTE Please note that the brakes
are most effective only on dry,
clean and slip-resistant flooring.
One-sided load on the bed
NOTE In order to prevent one-sid-
ed loads on the bed, it must
not be used as a seat for persons other than the occupant/
patient (i.e. visitors must not
sit on the edge of the bed).
Side rails
WARNING "Risk of entrapment"
In the case of occupants/patients
whose physical or mental condition
makes it necessary to use side
rails to protect them from falling
out of bed, the following safety
measures must be observed:
• The legal permissibility of using
side rails must be ascertained.
• The side rails may only be
operated by trained care
personnel.
• It must be ensured that the
occupant/patient does not come
into contact with the side rail
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
13 of 104
Help
Notes | General safety notes 3/4
elements upon activation of the
electrical lying surface adjustment.
It is also important to ensure that
no part of the body is sticking out
through the side rails.
• Ensure that the side rails
(or parts thereof) are either
fully raised and locked in
position or completely lowered.
• If the side rails are used with
a person whose psychological
condition makes their use
necessary, then it must be
ensured that the hand control
is kept out of his/her reach or
its functions are locked. It is also
strongly recommended that
side rail protection covers
are used.
DANGER If these safety meas-
ures are not observed by care
personnel, injuries can be caused
to hands, knees, fingers, feet, legs
and hips, along with haematomas
and other injuries as a result of
entrapment.
With children or people who
are less than 146 cm tall, nonobservance of these guidelines can lead to death!
Height adjustment
DANGER "Risk of entrapment
between the lower frame and/or
floor and the bed frame when the
bed is lowered"
It must be ensured that no people, limbs, pets, bed linen or other
objects are caught between the
bed frame and the lower frame
and/or floor during adjustment.
DANGER "Danger of movement"
If any movement of the bed could
represent a danger to the occupant
/patient, all functions must be
locked.
Accessories
WARNING
"Risk of injury"
Only original Völker accessories
should be used! Third-party
accessories must undergo testing
by the operator before use.
Use of lifting devices
The lifting devices specified are
appliances that can be attached to the
bed for transport purposes. Patient
lifters can be used.
WARNING "Risk of injury"
No lifting device must be fastened
directly to the bed (patient trans-
port, repair).
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
14 of 104
Help
Notes | General safety notes 4/4
Use of oxygen equipment
DANGER "Risk of fire"
Do not use any oxygen equipment
other than that which is administered via nasal prongs or masks.
Never use the bed in an oxygen
tent or in potentially explosive
atmospheres (where flammable
gases or vapours are present).
Provided it has been excluded (e.g.
based on information in the instructions
for use of the equipment being used)
that the use of the equipment or a fault
may cause the O2 concentration to rise
to such a degree that there is an explosion risk, then the equipment can be
used.
Rail infill panels
When using the rail infill panel, please
read the separate instructions for use
for this accessory. During technical
checks, the rail infill panels should also
be checked to ensure they are suitable
for the size of side rail used.
Cleaning and disinfection
In order to maintain consistent functioning, the bed should be cleaned, disinfected and tested as soon as possible
following each use, so that it can be
reused immediately without risk.
Incorrect cleaning/disinfection of the
bed can cause danger.
Maintenance and repair
Anyone responsible for carrying out
maintenance and repair work must at
least have received instruction from
Völker in the service tools, have read
the safety notes and the service manual
and be qualified in accordance with the
law on medical products § 4.
After maintenance work or repairs have
been carried out, a technical check
must be carried out on the affected
parts and/or functions 71.
During this check, it must be determined
whether the bed can be used in accordance with the specifications without risk to
the occupant/patient, user or a third party.
A technical check must be carried out
at least once a year and after every
lengthy period of non-use.
Any discernible damage, such as
signs of wear and tear, loose screws
or breaks/fractures must be eliminated
immediately.
Electromagnetic and
electrostatic interferences
The beds in model series 5384 Kepler
satisfy the EMC* requirements in accordance with the law on medical products
(MPG). The basis for testing is standard
EN 60601-1-2.
* Law concerning the Electromagnetic
Compatibility of Operating Equipment
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
15 of 104
Help
Functional description
The features of the Völker care bed
and its function are set out in the
Functional description section .
CONTENTS
Overview of Style model 17
Overview of Select model 18
Overview of manoeuvrable version 19
Hand control E2476 20
Hand control E2486 21
Hand control E2478 (optional) 22
Accessories - Mattresses 23
Accessories - Trapeze bar
and accessory holders 25
Securing systems 27
Accessories - Rail inll panels 28
Versions and Options 29
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
16 of 104
Help
Functional description | Overview of Style model
Plastic lying surface
(standard for Style
model)
Foot section
Braking guide castors with
foot pedal (both sides)
Continuous
side rails, both sides
Fixed
seat section
Adjustable
back section
Head section
Holders for trapeze bars,
drip stands etc., both
sides
Rapid lowering
of the back section (CPR)
below the seat area,
both sides
Adjustable
lower leg section
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
Adjustable
upper leg section
17 of 104
Help
Functional description | Overview of Select model
Adjustable side rails
for lower leg section,
both sides
Foot section
Dynamic
bed extension
(optional Select model)
Braking guide castors with
foot pedal (both sides)
MiS® lying surface
(standard for Select
model, optional for
Style model)
Fixed
seat section
Adjustable
back section
Adjustable side rails
for back section,
both sides
Head section
Holders for trapeze bars,
drip stands etc., both
sides
Rapid
lowering of the
back section (CPR),
both sides
Adjustable
lower leg section
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
18 of 104
Adjustable
upper leg section
Help
Functional description | Manoeuvrable version (optional)
The maneuverable version of both
models (option) is fitted with two
steering castors at both the head and
foot ends.
(optional for Select model)
Foot pedal
(both sides)
HPL lying surface
Braking guide castors
(both sides)
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
19 of 104
Help
Functional description | Hand control E2476 -
Upper leg section up
Upper leg section
down
Back section up
Back section down
Style
model
WARNING When actuating mo-
torised adjustments with the side
rails raised, ensure that the occupant/patient does not have any
contact with the side rails. Ensure
that no parts of his or her body, or
those of other people, are sticking through the side rails or are
trapped between the lying surface
and the lower frame and/or floor!
Lying surface up
Lying surface down
3
Aid for getting out of bed/
Position saving
1
Anti-Trendelenburg
positioning
2
Enable button
Hand control lock
1, 2, 3
For use of the enable button, please refer to the
note on page 59
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
20 of 104
Help
Functional description | Hand control E2486 - Select model (optional Style model)
WARNING When actuating mo-
torised adjustments with the side
Upper leg section up
Upper leg section
down
Lying surface up
Back section down
Back section up
Aid for getting out of bed and
save function
1
rails raised, ensure that the occupant/patient does not have any
contact with the side rails. Ensure
that no parts of his or her body, or
those of other people, are sticking through the side rails or are
trapped between the lying surface
and the lower frame and/or floor!
Lying surface down
Sitting position
Under-bed light
4
5
Anti-Trendelenburg
positioning
2
Trendelenburg
positioning
3
Enable button
Hand control lock
1, 2, 3, 4, 5
For use of the enable button, please refer to the
note on page 59
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
21 of 104
Help
Functional description | Hand control E2478 (optional)
Upper leg section up
Upper leg section
down
Back section down
Back section up
This hand control is available as an
option for both models, only in conjunction with hand control E2486.
Lying surface up
Lying surface
down
Under-bed light
Hand control lock
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
Aid for getting out of bed
22 of 104
WARNING When actuating mo-
torised adjustments with the side
rails raised. Ensure that the occupant/patient does not have any
contact with the side rails, and
that no parts of his or her body or
those of other people are sticking through the side rails or are
trapped between the lying surface
and the lower frame and/or floor.
To offer the greatest degree of flexibility,
Völker offers a wide range of easyto-attach accessories. The beds are
equipped as standard with holder devices for accessories, such as drip stands
and trapeze bars.
You can find further information regarding the accessories in our current information brochures or on the Internet at
www.voelker.de . Our staff will happily
inform you about available accessories
for your bed model.
To minimise the risk of injury, only use
mattresses with the dimensions and
properties described in the following.
If you do not use Völker mattresses,
please contact a dealer in whom you
have confidence.
WARNING Only original Völk-
er accessories may be used!
DANGER If mattresses are used
that do not match these specifications, there is a risk of suffocation!
Please read the relevant instructions for
use before using the mattresses.
NOTE Air-filled mattresses are
not compatible with dynamic bed
extension!
Ensure that the mattresses are placed
on the bed, aligned with the lying surface and securely attached as specified
in the regulations.
The tubes and cables of the air-filled
mattress must be positioned in such a
way that they are not kinked, squashed
or otherwise damaged by the moving
parts of the bed.
To reduce the length of the power cable,
the mattress system should be connected to the nearest possible socket so
that the cable does not lie on the floor.
are not damaged during transport, e.g.
by rolling over.
Cables and tubes of the air-filled mattresses must be checked regularly for
damage.
Mattresses with damaged cables and/or
tubes may not be used!
Typical representation for Völker
mattress NP50 Monodensity.
Before transporting the beds, the cables
and tubes of the air-filled mattresses
must be secured to the bed so that they
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
Völker NP50 ASS086 Polyurethane 88 x 200 x 12
Völker NP100 ASS087 Polyurethane 88 x 200 x 12
Völker NP150 ASS088 Polyurethane 88 x 200 x 12
Tolerance values: Length = ± 2 cm, width ± 1.5 cm, thickness = ± 1 cm
DANGER The height of the raised
side rail above the foam mattress
must always be greater than or
equal to 220 mm; otherwise the
occupant/patient may acciden-
≥ 220 mm!
(Foam mattress
in flat position)
tally fall out of bed. Please note
that the height of the mattress
has a direct influence on this.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
24 of 104
Help
Functional description | Accessories 3/6 Trapeze bar and accessory holders
Holders for the trapeze bar and accessories are located on the inside of the
head panel.The trapeze bar and other
accessories must slide into the holders
until you hear them click into place.
Trapeze bar holders ø 34.2 mm
The trapeze bar is
inserted into the
holder and latched
into place.
WARNING Only use the the tra -
peze bars listed in the accessories
list!
WARNING "Risk of Injury"
Ensure that the trapeze bar is
completely slotted into the holder and securely seated. Please
note: The safe working load of the
trapeze bar is maximum 75 kg.
WARNING "Risk of falling"
The trapeze bar must not be used
by the occupant/patient as a means
for climbing into the bed. The trapeze bar must never jut out beyond
the outer edge of the bed and then
be used as an aid to pull oneself
up (e.g. when getting out of a
wheelchair).
WARNING "Risk of injury"
The trapeze bar may not be used
by the patient/occupant without
supervision.
Similar pictures
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
25 of 104
Help
Functional description | Accessories 4/6 Trapeze bar and accessory holders
ø 13 mm
The trapeze bar holders can be provided with an adapter (ø 13 mm) for a drip
stand or for other accessories.
The adapter is inserted into one of the
trapeze bar holders.
The drip stand or the
accessory is fastened
from below with a
sleeve nut.
WARNING The drip stand must
always be aligned towards the bed,
as shown below, and may not
point outwards.
Should the bed be moved or adjusted,
the infusion tubing or cables must be
closely monitored by the care personnel.
Please also note that drainage devices
are able to touch the floor when
lowering the bed. This also applies for
the Anti-Trendelenburg and Trendelenburg position.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
The rail infill panels close the middle
gap between the divided side rails on
Völker beds so as to enable a continuous side rail solution.
The rail infill panels are inserted in the
insertion sleeves of the upper leg
section.
Check whether the pendulum locking is
completely latched into place by attempting to pull out the rail infill panel
upwards. If you succeed in doing this,
press the rail infill panel fully down
WARNING Make sure that the side
rails are fully set up and latched
into place. When activating the
electric lying surface adjustment
or side rails, it must be ensured
that the patient/occupant has
neither contact with the side rails,
nor any part of the body protrudes
through the side rails. We strongly recommend that you lock the
functions of the hand control.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
again and adjust the pendulum until the
rail infill panel can no longer be pulled
out.
27 of 104
To remove the rail infill panel from the
holder, move the pendulum into a vertical position and pull the rail infill panel
up and out at the same time.
NOTE Please observe the detailed
instructions for use for the rail
infill panels.
Further accessories can be found on
pages 99 and 100.
Help
Functional description | Accessories 6/6 Securing systems
Use of securing systems
Securing systems such as belts should
only be used exactly as specified by the
manufacturer.
If securing systems in the form of abdominal belts are used, then it must be
ensured that the side rails are completely raised. The gap in the middle of the
rails must always be closed using a rail
infill panel with the Select model.
Devices for safeguarding patients may
not be regarded as a substitute for the
requisite supervision of the patient/occupant by the care personnel.
Even if fastened correctly, means of
safeguarding patients can become
tangled and injure the patient/occupant
or even cause the death of the patient/
occupant if the latter is agitated and
confused.
When using safeguarding devices, the
patient/occupant must be supervised
corresponding to the statutory regulations
and the applicable protocol of the facility.
Failure to observe this note can lead to
personal injury or material damage.
DANGER When using securing
systems and rail infill panels, please note the separate
instructions for use pertaining to these accessories.
If the patient is restrained with the
help of restraint holders (optional
for both models), the lying surfaces
must never be adjusted while the
patient is restrained and must
always be in the lowest position!
The lying surface adjustment
functions must be locked when
a patient is restrained, and the
hand control must be kept out of
the reach of the occupant/patient!
Restraint holder
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
28 of 104
Help
Functional description | Versions and Options 1/3
These instructions for use cover all of
the versions and options listed. Precise
information on the bed versions supplied can be found in the order specifications for your bed.
If the original bed specification is
no longer available, please contact the
Völker Customer Services.
Please make a note of the serial number (ID No.) on the type label 83
before you call.
Wall protector wheels
Vertical wall protector wheels are optionally available for both models.
Linen holder
A pull-out linen holder at the foot of
the bed is optionally available for both
models.
The linen holder should always be
pushed in to avoid causing damage
when not in use.
WARNING The linen holder may
not be used as a seat or step!
The safe working load of the linen
holder is 20 kg.
Accessories rail
An accessories rail is readily available
for both models as an extra option.
NOTE If the lying surface is
lowered below 35 cm, any object
that has been attached on the
accessories rail must be removed!
Under-bed light
A under-bed light is available as an
option for Style model.
Contents
Völker Bed 5384 Kepler - Style - Select – from year of construction 2016 – Instructions for use G155 Rev. 7 (03.2018)
29 of 104
Help
Functional description | Versions and Options 2/3
MiS® lying surface
Völker MiS® is a support system where
the different elements register and
promote the occupant‘s/patient‘s own
movement.
Thanks to the large number of contact
points, the occupant/patient is supported in an anatomically correct manner, which also has a positive influence
on sleeping behavior.
HPL lying surface
The four-part HPL lying surface (HPL =
high-pressure laminate) consists of
moisture-resistant high-pressure laminate.
The four elements are well ventilated
via the round cut-outs, are individually
removable and fulfil the highest hygienic
requirements.
The HPL lying surfaces can be easily
removed and can therefore be cleaned
quickly and thoroughly. They also facilitate the use of therapy mattresses.
Dynamic bed extension
The Select model can be equipped with
a telescopic bed extension.
Extension: approx. 28 cm.
An additional mattress extension
(PA2215) must be used when using the
dynamic bed extension.
Battery
A battery is available for both bed
models as an option.
The battery allows the bed to be operated independent of the power supply for
at least two adjustment cycles under full
load.
Variants divided sidesails
Select model is available with the following siderail variants:
Extendable 34 - 35,5* cm (standard)
Extendable 43,5 - 45* cm (option)
* Dimensions are from the upper edge of the side
rail to the lying surface (without mattress). The
different extension heights are dependent on the
lying surface (HPL or MiS®), see also p. 86 - 87.