All rights reserved. Copying and reproduction either wholly or in part prohibited.
S-3
2001
oelker Nursing Bed
V
aintenance Manual
M
M
odels 2080/ 3010/ 3080
Foreword
We would like to take this opportunity of congratulating you on your choice of a Voelker hospital
bed and of thanking you for your confidence in our company and its products.
We are sure that you received detailed information and advice on our nursing beds that convinced
you of their many benefits ― benefits that really are only to be found in this configuration in these
beds. For this reason, your choice was not only good for us, but also right for you.
The fact that a Voelker nursing bed does not look like a sterile piece of institutional equiment
despite ist many functional features and technical refinements undoubtedly made your choice all
the easier.
We have produced this maintenance manual so that you and your staff never have any problems
with this nursing bed. To this end, the manual:
describes all technical features,
−
provides tips for caring for the bed,
−
informs you about maintenance work and
−
desrcibes trouble-shooting measures.
−
We are certain that with this information you will quickly become familiar with your new nursing
bed, and that it will give you the greatest possible satisfaction in use.
Völker GmbH
3
Foreword
oelker Nursing Bed
V
aintenance Manual
M
odel 2080/ 3010/ 3080
M
Table of contents
Page
1. General remarks.....................................................................................5
7. Tool list ................................................................................................44
8. Service centres.....................................................................................45
9. Spare parts order form ........................................................................45
4
oelker Nursing Bed
V
aintenance Manual
M
M
odels 2080/ 3010/ 3080
1. General remarks
You have purchased a Voelker nursing bed. This nursing bed has been manufactured in
accordance with national and international standards and regulations, and the current
state of technical knowledge.
Voelker nursing beds meet all safety and functional requirements. They have been
inspected in accordance with international standards and have been awarded the CE
Safety Seal for medical products.
Despite this, please read the basic safety instructions (see Operating Instructions). Please
pay special attention to the further instructions in this manual, especially with regard to
possible warranty claims.
This maintenance manual is intended to give technicians practical information on the safe
and correct servicing of the bed.
All individuals responsible for carrying out the maintenance, servicing or repair of the bed
must have read
the safety regulations and
−
the maintenance manual
−
To avoid mistakes and to ensure the trouble-free operation of our nursing beds, this
documentation must always be available to maintenance staff.
1.1 Copyright
This maintenance manual may only be made available to third parties with the written
permission of Voelker GmbH. All documentation is copyright protected.
Unless expressly permitted, transfer and copying of documentation, either whole or in part,
or ist exploitation and/or divulgence are fobidden. Infringements are punishable at law and
require the payment of damages. We reseve the right to exercise all our rights in
commercial law.
1.2 Warranty and liability
We accept liability for possible defects and omissions within the terms of the warranty set
out in the main contract of sale. All further claims are excluded. Claims for damages, from
whatever legal ground they may be held to arise, are also expressly excluded.
We reseve the right to alter the nursing bed described in this maintenance manual in
accordance with technical developments.
.
5
oelker Nursing Bed
V
aintenance Manual
M
odel 2080/ 3010/ 3080
M
We accept no liability whatsoever for damage and/or disruption caused by faulty operation
or by disregarding these maintenance instructions.
Actual accessories do not necessarily accord with drawings of them in every detail.
Exceptionally important information in the technical description is highlighted by the
following symbol:
= Caution
This information is to be read without fail and strictly obeyed.
2. Safety regulations
These regulations are to be read and strictly obeyed!
Additionally, the safety regulations to be found in the operating instructions are to be
read and obeyed!
2.1 Before beginning maintenance work
Those responsible for carrying out maintenance work must have read this maintenance
manual in detail before beginning work.
Before beginning servicing or repair work, those responsible must be thoroughly instructed
in the operation of the bed and made expressly aware of the risks involved in assembling
and disassembling each subassembly, as described in this maintenance manual.
The nursing beds are not explosion-protected. For this reason, they may only be serviced in
environments free of inflammable materials and/or substances.
2.2 Checking the operational safety and condition of the bed
The bed must be checked for operational safety after the completion of each maintenance
procedure and/or repair. This is to ensure that the bed can be used as intended without
endangering residents and staff (functional test: see operating instructions).
Should no servicing or repair work have been necessary, then it is sufficient to carry out a
functional test once per year.
All electrical components (particularly mains power supply cable, manual control unit,
power plug) must be regularly inspected for damage and safe routeing (should damage be
found, then the bed is to be taken out of use and the damaged parts replaced without
delay)!
6
oelker Nursing Bed
V
aintenance Manual
M
M
odels 2080/ 3010/ 3080
3. Technical specification
3.1
Technical da ta, Models 2080, 3010, 3080
2080 3010 3080
Length 204.5/224.5 cm 212.5/232.5 cm 204.5/224.5 cm
Width 98 – 102 cm
Height adjustment 34.5 cm 40.5 cm
Top edge, mattress-frame 35.5 - 70 cm 39.5 – 80 cm
Mattress-frame (4-sectional) 90 x 200 cm
100 x 200 cm
90 x 220 cm
100 x 220 cm
Special versions
Net weight approx. 97 kg
Safe working load 170 kg
Castors, 4X Type: K-100 / 2x1
Fa. Halver Rollen bzw. Fa. Blickle Rollen
Castors, safe running load 100 kg, tires: Vulkolan
Mains voltage AC 120 V
Rated output 350 W
Rated frequency 50 Hz
Transformer 150 VA
Secondary voltage 24 Volt
Primary fuse Type: fine-wire fuse, built-in, 3,15 A T
Fuse, manual control unit Type: polyswitch, built-in, 400 mA
Fuse, mattress-frame motors Type: polyswitch, built-in, 2,5 A
Fuse, lifting jack motor Type: polyswitch, built-in, each motor 3 A
Battery Type 9 V compound battery
(Alkali-Mangan – handelsüblich)
Standard manual control unit Type: Voelker
Special manual control unit Type: Voelker
Carers’ keyboard Type: Voelker
Drive unit for
back + upper leg sections Type: OKIMAT 2+2
Order No. 3080-04-00-001
OKIN
Lifting jack motor Type: 0 390 257 651 or 653
BOSCH
Operating temperature +10° to +40° Celsius
Transport/storage temperature -20° to +60° Celsius
Humidity 30% to 75%
Atmospheric pressure 700hPa to 1060hPa
7
3.2
3.2 Classification
Classification
3.23.2
ClassificationClassification
oelker Nursing Bed
V
aintenance Manual
M
odel 2080/ 3010/ 3080
M
Protection against electric shock Safety class II
(Equipment with double insulation and/or internal
electrical power source)
Protected by casing according to EN 60259 IP X4
Protection of parts accessible to user
against electric shock
according to DIN EN 60 601-1 Type B
Protection against explosive materials
and substances The bed is not explosion-protected and may not be
used in environments in which inflammable
anaesthetics or inflammable cleaning agents are
present
(see brochure ZH 1/200).
Duty cycle ED 10%
continuous switched operation max. 2 min.
max. 5 operations / 1 min.
MPG – classification Class 1
3.3 Functions
Height adjustment electric motor
Back section adjustment electric motor; 76°
Upper leg section adjustment electric motor; 45°
Lower leg dection adjustment manual
Raised lower leg positioning electric motor/manual
8
oelker Nursing Bed
V
aintenance Manual
M
M
odels 2080/ 3010/ 3080
4. Description of subassemblies, Nursing bed Models 2080, 30 10, 3080
Exploded drawings of the individual subassemblies are provided in overview on the
following pages.
With their help, you can quickly find the part you are looking for.
You can get the name of the part
drwing.
- parts marked with **** depend on length, width and model