Vitlab micropipette Operating Manual [de, en]

Page 1
VITLAB
®
micropipette
VITLAB GmbH
Linus-Pauling-Str.1 63762 Grossostheim Germany tel: +49 6026 97799-0 fax: +49 6026 97799-30 [email protected] www.vitlab.com
Gebrauchsanweisung
Operating Manual
Pipette / Pipette
Page 2
Page 3
3
Inhalt
1. Sicherheitsbestimmungen 4
2. Verwendungszweck 5
3. Einsatzgrenzen 5
4. Einsatzbeschränkungen 5
5. Einsatzausschlüsse 5
6. Funktions- und Bedienelemente 6
7. Pipettieren 7
8. Volumen kontrollieren 9
9. Genauigkeitstabelle 10
10. Justieren 11
11. Autoklavieren 12
12. UV Entkeimung 12
13. Filter Pipette 5 ml / 10 ml 12
14. Wartung und Reinigung 13
14.1. VITLAB® micropipette bis 1000 µl 13
14.2. VITLAB® micropipette 5 ml / 10 ml 14
15. Bestelldaten · Zubehör 15
15.1 Bestelldaten 15
15.2 Zubehör 15
16. Ersatzteile 16
16.1. VITLAB® micropipette bis 1000 µl 16
16.2. VITLAB® micropipette 5 ml / 10 ml 17
17. Störung - was tun? 18
18. Reparatur · Kalibrierservice 19
18.1. Reparatur 19
18.2. Kalibrierservice 19
19. Mängelhaftung 20
20. Entsorgung 20
Page 4
4
Bitte unbedingt sorgfältig durchlesen!
Dieses Gerät kann in Kombination mit gefährlichen Materialien, Arbeitsvorgängen und Apparaturen verwendet werden. Die Gebrauchsanleitung kann jedoch nicht alle Sicherheitsprobleme aufzeigen, die hierbei eventuell aufteten. Es liegt in der Verantwortung des Anwenders, die Einhaltung der Sicherheits- und Gesundheitsvorschriften sicherzustellen und die entsprechenden Einschränkungen vor Gebrauch festzulegen.
1. Sicherheitsbestimmungen
1. Jeder Anwender muss diese Gebrauchsan-
leitung vor Gebrauch des Gerätes gelesen haben und beachten.
2. Allgemeine Gefahrenhinweise und Sicher-
heits-vorschriften befolgen (z. B. Schutzklei­dung, Augenschutz und Schutzhandschuhe tragen). Beim Arbeiten mit infektiösen oder gefährlichen Proben müssen die Standard­laborvorschriften und -vorkehrungen einge­halten werden.
3. Angaben der Reagenzienhersteller beachten.
4. Gerät nur zum Pipettieren von Flüssigkeiten
im Rahmen der definierten Einsatzgrenzen und -beschränkungen einsetzen. Einsatzaus­schlüsse beachten (s. Seite 5)! Bei Zweifel unbedingt an den Hersteller oder Händler wenden.
5. Stets so arbeiten, dass weder Anwender
noch andere Personen gefährdet werden. Spritzer vermeiden. Nur geeignete Gefäße verwenden.
6. Die Berührung der Spitzenöffnung ist beim Arbeiten mit aggressiven Medien zu vermei­den.
7. Nie Gewalt anwenden.
8. Nur Original-Ersatzteile verwenden. Keine technischen Veränderungen vornehmen. Das Gerät nicht weiter zerlegen, als in der Gebrauchsanleitung beschrieben ist!
9. Vor Verwendung stets den ordnungsge­mäßen Zustand des Gerätes prüfen. Sollten sich Störungen des Gerätes ankündigen (z.B. schwergängiger Kolben, Undichtigkeit), so­fort aufhören zu pipettieren und das Kapitel ‚Störung – was tun‘ befolgen (s. Seite 18). Ggf. an den Hersteller wenden.
Page 5
5
3. Einsatzgrenzen
Das Gerät dient zum Pipettieren von Proben unter Beachtung folgender Grenzen:
- Einsatz zwischen +15 °C und +40 °C von Gerät und Reagenz – andere Temperaturen auf Anfrage
- Dampfdruck bis 500 mbar
- Viskosität: 260 mPa s
2. Verwendungszweck
Luftpolsterpipette zum Pipettieren von wässerigen Lösungen mittlerer Dichte und geringer bis mittlerer Viskosität.
4. Einsatzbeschränkungen
Viskose und benetzende Flüssigkeiten können die Genauigkeit des Volumens beeinträchtigen. Ebenso Flüssigkeiten, deren Tempera­tur mehr als ± 1 °C von der Raumtemperatur abweicht.
5. Einsatzausschlüsse
Der Anwender muss die Eignung des Geräts für den Verwen­dungszweck selbst überprüfen.
Das Gerät kann nicht eingesetzt werden:
- für Flüssigkeiten, die Polypropylen angreifen
- für Flüssigkeiten, die Polycarbonat angreifen (Sichtfenster)
- für Flüssigkeiten mit sehr hohem Dampfdruck
- für Flüssigkeiten, die FKM und Polyetheretherketon (PEEK) angreifen
Page 6
6
6. Funktions- und Bedienelemente
Volumeneinstellrad
Spitzenabwurftaste*
Volumenanzeige**
**Volumenanzeige Die Ziffern in der Anzeige werden von oben nach unten gelesen, der Strich entspricht dem Dezimal­punkt.
*Spitzenabwurftaste Die Seriennummer befindet sich hinter der Spitzen­abwurftaste.
Hinweis:
Einwandfreie Analysenergebnisse sind nur mit Qualitäts-Spitzen zu erreichen. Wir empfehlen VITLAB®-Pipettenspitzen.
Pipettierknopf
Kalibrierfunktion
Fingerbügel
Griffteil
Spitzenaufnahmekonus
(Abb.: VITLAB® micropipette 100 µl)
Pipettenschaft
Page 7
7
- 5 ml- und 10 ml-Geräte sollten nur mit eingebautem PE-Filter verwendet werden (s. Seite 12)!
- Pipettenspitzen sind Einmalartikel!
2. Volumen einstellen
Volumeneinstellrad zur Auswahl des gewünschten Volumens drehen. Dabei gleichmäßig drehen und abrupte Drehbewegungen vermeiden.
3. Probe aufnehmen
a) Pipettierknopf bis zum ersten Anschlag drücken.
b) Gerät senkrecht halten und Spitze in die Flüssigkeit eintauchen.
c) Pipettierknopf gleichmäßig zurückgleiten lassen. Damit die
Flüssigkeit ihre Endposition erreicht, Spitzen noch ca. 1-2 s eingetaucht lassen.
d) Spitzen kurz an die Gefäßwand führen.
1. Spitzen aufstecken
Richtige Spitzen entsprechend dem Volumenbereich bzw. Color­Code verwenden! Auf dichten und festen Spitzensitz achten.
7. Pipettieren
Volumen­bereich
Eintauchtiefe
in mm
Wartezeit
in s
> 1 µl - 100 µl 2 - 3 1 > 100 µl - 1000 µl 2 - 4 1 > 1000 µl 3 - 6 3
Page 8
8
5. Spitzen abwerfen
Pipettenschaft über einen geeigneten Entsorgungsbehälter halten und die Spitzenabwurftaste bis zum Anschlag niederdrücken.
Hinweis:
Die ISO 8655 schreibt vor, die Pipettenspitzen vor dem ei­gentlichen Pipettiervorgang einmal mit der Probenflüssigkeit vorzuspülen.
Wichtig!
Gerät mit gefüllten Spitzen nicht hinlegen, da sonst Medium in das Gerät fließen und dieses kontaminieren kann! Gerät stets aufrecht und ohne Spitze im separat zu bestellenden Regalhalter bzw. Tischständer aufbewahren.
c) Spitzen durch Überhub völlig entleeren: Pipettierknopf bis zum
zweiten Anschlag drücken.
d) Pipettenspitze dabei an der Gefäßwand abstreifen.
e) Pipettenspitze von der Gefäßwand zurücknehmen und Pipet-
tierknopf zurückgleiten lassen.
4. Probe abgeben
a) Pipettenspitze an Gefäßwand anlegen. Pipette im Winkel von
30-45° zur Gefäßwand halten.
b) Pipettierknopf mit gleichmäßiger Geschwindigkeit bis zum er-
sten Anschlag drücken und festhalten. Bei Seren, hochviskosen oder entspannten Medien entsprechende Wartezeit einhalten, um Genauigkeit zu verbessern.
Page 9
9
8. Volumen kontrollieren
Wir empfehlen, je nach Einsatz, alle 3-12 Monate eine Prüfung des Gerätes. Der Zyklus kann aber den individuellen Anforderungen angepasst werden. Die gravimetrische Volumenprüfung der Pipette erfolgt durch nachfolgende Schritte und entspricht der DIN EN ISO 8655, Teil 6.
1. Nennvolumen einstellen
Maximales angegebenes Gerätevolumen
einstellen (Vorgehensweise siehe Seite 7).
2. Pipette konditionieren
Pipette vor der Prüfung konditionieren,
indem mit einer Pipettenspitze fünfmal die Prüfflüssigkeit (H2O dest.) aufgenommen und abgegeben wird.
3. Prüfung durchführen
a) Neue Pipettenspitze aufstecken und einmal
mit Prüfflüssigkeit vorspülen.
b) Prüfflüssigkeit aufnehmen und in das Wäge-
gefäß pipettieren.
c) Pipettierte Menge mit einer Analysenwaage
wägen. Beachten Sie bitte die Gebrauchsan­leitung des Waagenherstellers.
d) Pipettiertes Volumen berechnen. Dabei die
Temperatur der Prüfflüssigkeit berücksichti­gen.
e) Mindestens 10 Pipettierungen und Wägun-
gen in 3 Volumenbereichen (100%, 50%, 10%) werden empfohlen.
Berechnung für Nennvolumen V
0
xi = Wäge-Ergebnisse n = Anzahl der Wägungen Z = Korrekturfaktor (z. B. 1,0029 µl/mg
bei 20 °C, 1013 hPa)
Mittelwert
x =
Σ x
i
n
Mittleres Volumen
Richtigkeit*
Variationskoeffizient*
Standardabweichung
V = x · Z
100 s
V
VK% =
Σ (xi – x)
2
n – 1
s = Z ·
*) = Berechnung von Richtigkeit (R%) und Vari­ationskoeffizient (VK%): R% und VK% werden nach den Formeln der statistischen Qualitäts­kontrolle berechnet.
V – V
0
V
0
R% = · 100
Hinweis:
Prüfanweisungen (SOPs) stehen unter www.vitlab.com zum Download zur Verfü­gung.
Page 10
10
9. Genauigkeitstabelle
VITLAB® micropipette, Typ variabel
Volumenbe-
reich µl
Teilvolumen
µl
R*
≤ ± %
VK*
≤ %
Teilschritte
µl
Empfohlener
Spitzentyp, µl
0,5 - 10
10
5 1
1 1,6 7
0,5 1 4
0,01 0,5 - 20
2 - 20
20 10
2
0,8 1,2 5
0,4 0,7 2
0,02 2 - 200
10 - 100
100
50 10
0,6 0,8 3
0,2 0,4 1
0,1 2 - 200
20 - 200
200 100
20
0,6 0,8 3
0,2 0,3 0,6
0,2 2 - 200
100 - 1000
1000
500 100
0,6 0,8 3
0,2 0,3 0,6
1 50 - 1000
500 - 5000
5000 2500
500
0,6 0,8 3
0,2 0,3 0,6
5 500 - 5000
1000 - 10000
10000
5000 1000
0,6 0,8 3
0,2 0,3 0,6
10 1000 - 10000
Endprüfwerte bezogen auf das auf dem Gerät aufgedruckte Nennvolumen (= max. Volu­men) bei gleicher Temperatur (20 °C) von Gerät, Umgebung und aqua dest., gemäß der DIN EN ISO 8655.
* R = Richtigkeit * VK = Variationskoeffizient
20 °C Ex
Page 11
11
10. Justieren
Das Gerät ist permanent justiert für wässerige Lösungen. Sollte einwandfrei feststehen, dass die Pipette ungenau arbeitet oder, um das Gerät auf Lösungen unterschiedlicher Dichte und Viskosität oder speziell geformte Pipettenspitzen einzustellen, kann es justiert werden.
1. Volumenkontrolle durchführen, Ist-Wert ermitteln (s. Seite 9).
2. Abdeckung entfernen: Haken nach vorn drücken, leicht anhe-
ben und nach hinten ziehen.
3. Mit einer Büroklammer oder einer unbenutzten Pipettenspitze
die Schutzfolie entfernen (die Schutzfolie wird nicht weiter benötigt).
4. Roten Justageschieber vollständig nach hinten schieben, Volu-
meneinstellrad hochziehen (Entkopplung) und Justageschieber loslassen.
5. Justagewert einstellen:
Den zuvor ermittelten Ist-Wert mit dem Volumeneinstellrad
einstellen. Es wird eine Volumenkontrolle nach jeder Justage empfohlen.
6. Justageschieber erneut vollständig nach hinten schieben, das
Volumeneinstellrad nach unten drücken und den Justageschie­ber loslassen. Abdeckung wieder montieren.
Hinweis:
Die Änderung der Werkseinstellung wird durch den dann sichtbaren roten Justageschieber angezeigt.
Page 12
12
13. Filter Pipette 5 ml / 10 ml
Der hydrophobe PE-Filter dient als Schutz vor dem Eindringen von Flüssigkeit in die Pipette. Filter wechseln, sobald der Filter benetzt oder verschmutzt ist.
- Flachen Gegenstand, z. B. Schraubendreher verwenden.
- Filter vorsichtig herausziehen, ohne den Spitzenkonus zu
beschädigen.
Vor dem Autoklavieren Filter entfernen! Das Gerät kann auch ohne Filter betrieben werden.
11. Autoklavieren
Die Pipette ist komplett autoklavierbar bei 121 °C (2 bar) und einer Haltezeit von mindestens 15 Minuten nach DIN EN 285.
1. Pipettenspitze abwerfen.
2. Ohne weitere Demontage die komplette Pipette autoklavieren. Bei Pipette 5 ml und 10 ml den Filter vor dem Autoklavieren entfernen.
3. Pipette vollständig abkühlen und trocknen lassen.
Hinweis:
Die Wirksamkeit des Autoklavierens ist vom Anwender selbst zu prüfen. Höchste Sicherheit wird durch Vakuumsterilisation erreicht. Wir empfehlen die Verwendung von Sterilisationsbeu­teln.
Achtung!
Vor dem Autoklavieren muss die Volumeneinstellung auf einen mit Ziffern versehenen Wert eingestellt werden (z.B. auf 11,25 oder 11,26, aber nicht dazwischen).
Bei häufigem Autoklavieren sollten Kolben und Dichtung zur besseren Gängigkeit mit Silikonfett gefettet werden. Bitte nur empfohlenes Silikonfett verwenden, s. Zubehör Seite
15. Gegebenenfalls nach dem Autoklavieren Schraubverbindung
zwischen Griffteil und Pipettiereinheit festziehen.
12. UV-Entkeimung
Das Gerät ist gegen die übliche Belastung einer UV-Entkeimungs­lampe beständig. Infolge der UV-Einwirkung ist eine Farbände­rung möglich.
Page 13
A
B
D
E
S
C
13
14. Wartung und Reinigung
14.1. VITLAB® micropipette bis 1000 µl
1. Wartung
Pipettenaufnahmekonus auf Beschädigung prüfen. Kolben und Dichtung auf Verschmutzung untersuchen. Dichtheit des Geräts prüfen: Dazu Probe aufsaugen und Gerät ca. 10 s senkrecht halten. Falls sich an der Pipettenspitze ein Tropfen bildet: Störung – was tun? Seite 18.
2. Demontage und Reinigung
1. Pipettenschaft (S) vom Griffteil durch Abschrauben lösen.
2. Abwerferoberteil (A) aus dem Pipettenschaft herausschrauben.
3. Schaft (B, C u. D) aus dem Abwerferunterteil (E) herausziehen.
4. Kolbeneinheit (B) herausschrauben.
Hinweis:
Kolben bleibt mit Kolbeneinheit (B) ver­bunden!
5. Dichtung mit Feder (C) entnehmen (bei Pipette 10 µl nicht möglich!).
6. Abgebildete Teile mit Seifenlösung oder Isopropanol reinigen, anschließend mit aqua dest. spülen.
7. Teile trocknen (max. 120 °C).
8. Kolben und Dichtung hauchdünn mit Silikonfett nachfetten. Bitte nur empfohlenes Silikonfett verwenden, s. Zubehör Seite 15.
9. Abgekühlte Teile wieder in umgekehrter Reihenfolge montie­ren. Kolbeneinheit und Abwerferoberteil (A, B) nur handfest anziehen.
(Abbildung beispielhaft)
Page 14
S
K
F
G
I
H
F´
14
14.2. VITLAB® micropipette 5 ml und 10 ml
1. Wartung
Pipettenaufnahmekonus auf Beschädigung prüfen. Kolben und O-Ring-Dichtung auf Verschmutzung untersuchen. Dichtheit des Geräts prüfen. Dazu Probe aufsaugen, Gerät ca. 10 s senkrecht halten. Falls sich an der Pipettenspitze ein Tropfen bildet: „Störung – was tun?“, Seite 18.
2. Demontage und Reinigung
1. Kompletten Schaft (S) durch Drehen am Abwerferoberteil (F) vom Griffteil lösen und Filter (K) aus Schaftunterteil (H) heraus­ziehen.
2. Abwerferunterteil (F‘) durch Abschrauben vom Abwerferober­teil (F) trennen.
3. Kolbeneinheit (G) mit Abwerferfeder
( I ) und Schaftunterteil (H) auseinander-schrauben.
4. O-Ring von Kolbeneinheit abziehen und reinigen.
Hinweis:
Die Kolbeneinheit (G) nicht weiter de­montieren!
5. Kolbeneinheit (G) und Schaftunterteil (H) mit Seifenlösung
oder Isopropanol reingen, anschließend mit aqua dest. spülen.
6. Teile trocknen (max. 120 °C) und abkühlen lassen.
7. O-Ring sorgfältig innen und außen fetten und auf Kolben
aufziehen.
8. Die Einzelkomponenten wieder in umgekehrter Reihenfolge
montieren.
(Abbildung beispielhaft!)
Page 15
15
15. Bestelldaten und Zubehör
Tischständer für 6 VITLAB® micropipette
Best.-Nr. 1672002
Regalhalter für 1 VITLAB® micropipette
Best.-Nr. 1672000
VITLAB® micropipette
Volumen Best.-Nr.
0,5 - 10 µl 1641000 2 - 20 µl 1641002
10 - 100 µl 1641004 20 - 200 µl 1641006
100 - 1000 µl 1641008
0,5 - 5 ml 1641010
1,0 - 10 ml 1641012
15.2. Zubehör
15.1. Bestelldaten
Filter für VITLAB® micropipette 5 ml
VE 25 Stk. Best.-Nr. 1672010
Filter für VITLAB® micropipette 10 ml VE 25 Stk. Best.-Nr. 1672012
Silikonfett für VITLAB® micropipette bis 1000 µl VE 1 St. Best.-Nr. 1672015
Silikonfett für VITLAB® micropipette 5 ml/10 ml
VE 1 St. Best.-Nr. 1672016
Page 16
A
B
E
C
D
16
16. Ersatzteile
16.1. VITLAB® micropipette bis 1000 µl
Abwerfer­oberteil
Kolben­einheit
Dichtung mit Feder
Schaft mit Ab­werferfeder
Abwerfer­unterteil
Volumen A B C D E
0,5 - 10 µl 1671400 1671411 - 1671431* 1671441 2 - 20 µl 1671401 1671412 1671420 1671432 1671442 10 - 100 µl 1671402 1671416 1671424 1671433 1671443 20 - 200 µl 1671403 1671417 1671425 1671434 1671444 100 - 1000 µl 1671404 1671418 1671426 1671435 1671445
* 0,5-10 µl inkl. Dichtung
Vor dem Nachbestellen einer Kolbeneinheit oder Dichtung mit Feder für Ihre VITLAB® micropipette mit Glaskolben (bis Serien­nummer 08N) wenden Sie sich bitte an [email protected]. (Abbil­dung beispielhaft, Aussehen und Abmessungen der Ersatzteile entsprechen dem jeweiligen Nennvolumen.)
Page 17
I
H
G
F´
F
17
16.2. VITLAB® micropipette 5 ml und 10 ml
Abwerfer­oberteil
Abwerfer­unterteil
Kolben­einheit
Schaft­unterteil
Abwerfer­feder
(Abbildung beispielhaft, Aussehen und Abmessungen der Ersatzteile entsprechen dem jeweiligen Nennvolumen.)
Volumen F + F‘ G H I
0,5 - 5 ml 1671451 1671455 1671461 1671465 1 - 10 ml 1671452 1671456 1671462 1671465
Page 18
18
17. Störung - was tun?
Störung Mögliche Ursache Was tun?
Spitze tropft (Gerät undicht)
- Ungeeignete Spitze - Nur Qualitätsspitzen verwen­den
- Spitze sitzt nicht fest - Spitze fester aufdrücken
Gerät saugt nicht oder zu we­nig auf, abgegebenes Volumen zu klein
- Dichtung verunreinigt - Dichtung reinigen
- Dichtung oder Konus
beschädigt
- Dichtung oder Schaft ersetzen
- Kolben verunreinigt oder
beschädigt
- Kolben reinigen oder ersetzen
Ansaugen sehr langsam - Schaft verstopft - Schaft reinigen
- Bei 5 ml und 10 ml
Geräten Filter verschmutzt
- Filter wechseln
Abgegebenes Volumen zu groß - Pipettierknopf vor dem
Ansaugen zu weit bis in den Überhub gedrückt
- Auf korrekte Handhabung achten. Siehe ‚Pipettieren‘, Seite 7.
Kolben schwergängig - Kolben verschmutzt oder
ohne Fett
- Kolben reinigen und fetten
Page 19
19
18. Reparatur - Kalibrierservice
Sollte eine evtl. Funktionsstörung nicht im eigenen Labor durch einfachen Austausch von Ersatzteilen zu beheben sein, muss das Gerät zur Reparatur eingesandt werden.
Dabei ist zu beachten, dass aus Sicherheitsgründen nur saubere und dekontaminierte Geräte geprüft und repariert werden können!
18.1. Zur Reparatur einsenden
a) Gerät gründlich reinigen und dekontaminieren. b) Formular „Erklärungen zur gesundheitlichen Unbedenklichkeit“
ausfüllen (Vordrucke können beim Händler oder Hersteller angefordert werden bzw. stehen unter www.vitlab.com zum Download bereit).
c) Ausgefülltes Formular gemeinsam mit dem Gerät an den Her-
steller bzw. Händler senden mit genauer Beschreibung der Art der Störung und der verwendeten Medien.
Der Rücktransport geschieht auf Gefahr und Kosten des Einsen­ders.
18.2. Kalibrierservice
Die ISO 9001 und GLP-Richtlinien fordern die regelmäßige Überprüfung Ihrer Volumenmessgeräte. Wir empfehlen, alle 3-12 Monate eine Volumenkontrolle vorzunehmen. Der Zyklus ist abhängig von den individuellen Anforderungen an das Gerät. Bei hoher Gebrauchshäufigkeit oder aggressiven Medien sollte häufiger geprüft werden. Die ausführliche Prüfanweisung steht unter www.vitlab.com zum Download bereit. VITLAB bietet Ihnen darüber hinaus die Möglichkeit, Ihre Geräte durch unseren Kalibrierservice kalibrieren zu lassen. Schicken Sie uns einfach die zu kalibrierenden Geräte mit der Angabe, welche Art der Kalibrierung Sie wünschen. Sie erhalten die Geräte nach wenigen Tagen zusammen mit einem Prüfbericht (Werkskalibrierung) bzw. mit einem DAkkS-Kalibrierschein zurück. Nähere Informationen erhalten Sie von Ihrem Fachhändler oder direkt von VITLAB.
Page 20
20
19. Mängelhaftung
Wir haften nicht für Folgen unsachgemäßer Behandlung, Verwen­dung, Wartung, Bedienung oder nicht autorisierter Reparatur des Gerätes oder für Folgen normaler Abnutzung, insbesondere von Verschleißteilen wie z.B. Kolben, Dichtungen, Ventilen sowie bei Glasbruch. Gleiches gilt für die Nichtbeachtung der Gebrauchs­anleitung. Insbesondere übernehmen wir keine Haftung für entstandene Schäden, wenn das Gerät weiter zerlegt wurde als in der Gebrauchsanleitung beschrieben oder wenn fremde Zubehör­bzw. Ersatzteile eingebaut wurden.
20. Entsorgung
Bei der Entsorgung der Geräte und der Spitzen bitte die jeweiligen nationalen Entsorgungsvorschriften beachten.
Technische Änderungen, Irrtum und Druckfehler vorbehalten.
Page 21
Contents
1. Safety Instructions 22
2. Purpose 23
3. Limitations of use 23
4. Operating Limitations 23
5. Eperating Exclusions 23
6. Operating and Control Elements 24
7. Pipetting 25
8. Checking the Volume 27
9. Accuracy Table 28
10. Adjustment 29
11. Autoclaving 30
12. UV sterilisation 30
13. Filter pipette 5 ml / 10 ml 30
14. Servicing and Cleaning 31
14.1. VITLAB® micropipette up to 1000 µl 31
14.2. VITLAB® micropipette 5 ml / 10 ml 32
15. Ordering Information · Accessories 33
15.1. Ordering Information 33
15.2. Accessories 33
16. Spare Parts 34
16.1. VITLAB® micropipette up to 1000 µl 34
16.2. VITLAB® micropipette 5 ml / 10 ml 35
17. Troubleshooting 36
18. Repairs · Calibration Service 37
18.1. Return for repair 37
18.2. Calibration Service 37
19. Warranty Information 38
20. Disposal 38
21
Page 22
Please read the following carefully!
This instrument may sometimes be used with hazardous materials, operations, and equipment. It is
beyond the scope of this manual to address all of the potential safety risks associated with its use in such applications. It is the responsibility of the user of this pipette to consult and establish appropri­ate safety and health practices and determine the applicability of regulatory limitations prior to use.
1. Safety Instructions
1. Every user must read and understand this operating manual prior to using the instru­ment and observe these instructions during use.
2. Follow general instructions for hazard pre­vention and safety instructions; e.g., wear protective clothing, eye protection and gloves. When working with infectious or other hazardous samples, all appropriate re­gulations and precautions must be followed.
3. Observe all specifications provided by rea­gent manufacturers.
4. Only use the instrument for pipetting liquids that conform to the specifications defined
in the limitations of use and operating limi­tations. Observe operating exclusions (see page 23). If in doubt, contact the manufac­turer or supplier.
5. Always use the instrument in such a way that neither the user nor any other person is endangered. Avoid splashes. Use only suita-
ble vessels.
6. Avoid touching the tip orifices when wor­king with hazardous samples.
7. Never use force on the instrument!
8. Use only original spare parts. Do not at-
tempt to make any technical alterations. Do not dismantle the instrument any further than is described in the operating manual!
9. Before use check the instrument for visible
damages. If there is a sign of a potential malfunction (e.g., piston difficult to move,
leakage), immediately stop pipetting. Con-
sult the ‘Troubleshooting’ section of this manual (see page 36), and contact the manufacturer if needed.
22
Page 23
2. Purpose
Air-displacement pipette for pipetting aqueous solutions of medi­um density and low to medium viscosity.
3. Limitations of Use
The instrument is intended for the pipetting of liquids within the following limitations:
- Temperature of both the instrument and solution should be between 15 °C to 40 °C (59 °F to 104 °F). Consult the ma­ nufacturer for use in temperatures outside of this range.
- Vapor pressure up to 500 mbar
- Viscosity: 260 mPa s (260 cps)
4. Operating Limitations
Viscous and highly adhesive liquids may impair volumetric accu­racy. Volumetric accuracy may also be impaired when pipetting liquids that differ from ambient temperature by more than ± 1 °C/ ± 1.8 °F.
5. Operating Exclusions
The user has to ensure the compatibility of the instrument with the intended application. This instrument cannot be used:
- for liquids incompatible with polypropylene
- for liquids incompatible with polycarbonate (inspection window)
- for liquids of a very high vapor pressure
- for liquids incompatible with FKM and polyetheretherketone (PEEK)
23
Page 24
6. Operating and Control Elements
Volume-setting wheel
Tip ejection key *
Volume display **
Pipetting button
Calibration function
Finger rest
Handgrip
Tip cone
Pipette shaft
**Volume display The figures in the display are read from top to bottom, the dash represents the decimal point.
*Tip ejection key The serial number is behind the tip ejection key.
Note:
Optimum analysis results can only be ob­tained with quality tips. We recommend VITLAB® pipette tips.
(Fig. shows VITLAB® micropipette 100 µl)
24
Page 25
Use the correct tips according to the volume range or the color code. Ensure that the tips is securely seated.
a) Press pipetting button to the first stop.
b) Hold the pipette vertically and immerse the tip into the liquid.
c) Let the pipetting button slide back slowly. In order for the
liquid to reach its end position, leave the tip immersed for another 1-2 s.
d) Touch the tip against the container wall.
3. Aspirate sample
2. Volume setting
Select the desired volume by rotating the volumes setting wheel. Avoid twisting and abrupt rotating motions during this adjust­ment.
7. Pipetting
1. Fitting the tip
- 5 ml and 10 ml instruments should only be used with the PE
filter installed (see page 30).
- Pipette tips are disposable items!
25
Volume range
Immersion
depth in mm
Waiting
time in s
> 1 µl - 100 µl 2 - 3 1 > 100 µl - 1000 µl 2 - 4 1 > 1000 µl 3 - 6 3
Page 26
5. Ejecting the tip
Hold the pipette shaft over a suitable disposal container and press the tip ejection key to the stop.
Note:
ISO 8655 prescribes rinsing the pipette tip once with the sam­ple liquid prior to the actual pipetting process.
Important!
Don‘t lay the instrument horizontal when the tip is filled. Liquid may enter and contaminate the instrument. The instru­ment should be stored without tips, placed upright in the shelf/rack mount or bench top rack which can be ordered seperately.
c) The blow-out stroke empties the tip completely: Press the pi-
petting button down to the second stop.
d) While doing this, wipe the pipette tip against the wall of the
container.
e) Remove the pipette tip from the container wall and let the
pipetting button slide back.
a) Place the pipette tip against the wall of the vessel. Hold the
pipette at an angle of 30-45° relative to the container wall.
b) Press the pipetting button slowly to the first stop and hold it
down. For serum and liquids of high viscosity or low surface tension, observe adequate waiting time to improve accuracy.
4. Discharge sample
26
Page 27
Mean value
x =
Σ x
i
n
Mean volume
Accuracy*
Coefficient of Variation*
Standard Deviation
V = x · Z
100 s
V
CV% =
Σ (xi – x)
2
n – 1
s = Z ·
V – V
0
V
0
A% = · 100
8. Checking the Volume
Depending on use, we recommend inspection of the instrument every 3 to 12 months. The cycle can, however, be adjusted to individual requirements. The gravimetric testing of the pipette volume is performed according to the following steps and is in accordance with DIN EN ISO 8655, Part 6.
1. Set nominal volume
Set volume to the maximum volume indi-
cated on the instrument (see page 25 for procedure).
2. Condition the pipette
Condition the pipette before testing by using
a pipette tip to aspirate and discharge the test liquid (distilled H2O) five times.
3. Carry out the test
a) Attach new pipette tip and prerinse one time
with test liquid.
b) Aspirate liquid and pipette it into the weigh-
ing vessel.
c) Weigh the pipetted quantity with an
analytical balance. Please follow the operat­ing manual instructions from the balance manufacturer.
d) Calculate the volume, taking the tempera-
ture of test liquid into account.
e) At least 10 pipettings and weighings in three
volume ranges (100%, 50%, 10% of nomi­nal volume) are recommended for statistical analysis.
Calculation for nominal volume V
0
xi = Weighing results n = Number of weighings Z = Correction factor (e.g. 1,0029 µl/mg at 20 °C, 1013 hPa)
*) = Calculation of accuracy (A%) and variation coefficient (CV%): A% and CV% are calculated according to the formulas for statistical control.
Note:
Testing instructions (SOPs) are available for download at www.vitlab.com.
27
Page 28
Volume range
µl
Volume step
µl
A*
≤ ± %
CV* ≤ %
Increment
µl
Recommended
type of tip, µl
0.5 - 10
10
5 1
1
1.6 7
0.5 1 4
0.01 0.5 - 20
2 - 20
20 10
2
0.8
1.2 5
0.4
0.7 2
0.02 2 - 200
10 - 100
100
50 10
0.6
0.8 3
0.2
0.4 1
0.1 2 - 200
20 - 200
200 100
20
0.6
0.8 3
0.2
0.3
0.6
0.2 2 - 200
100 - 1000
1000
500 100
0.6
0.8 3
0.2
0.3
0.6
1 50 - 1000
500 - 5000
5000 2500
500
0.6
0.8 3
0.2
0.3
0.6
5 500 - 5000
1000 - 10000
10000
5000 1000
0.6
0.8 3
0.2
0.3
0.6
10 1000 - 10000
9. Accuracy Table
VITLAB® micropipette, Digital adjustable
Final test values related to the nominal capacity (maximum volume) indicated on the instrument, obtained when instrument and distilled water are equilibrated at ambient temperature (20 °C/68 °F) and with smooth operation. According to DIN EN ISO 8655.
* A = Accuracy * CV = Coefficient of Variation
28
20 °C Ex
Page 29
10. Adjustment
The instrument is permanently adjusted for aqueous solutions. If the pipette operation is clearly inaccurate, or if the instrument must be adjusted for solutions of different densities and viscosities or specially-shaped pipette tips, adjustments can be made.
1. Check the volume, determine actual value (see page 27).
2. Remove the cover: Push the hook forward, raise it slightly and then pull it back.
3. Using a paperclip or an unused pipette tip, remove the protec­tive film (this protective film can be discarded).
4. Push the red adjustment slider completely back, raise the volume-setting wheel (decoupling) and release the adjustment slider.
5. Set the adjustment value:
set the volume-setting wheel to the previously determined
actual value. A volume check is recommended after every adjustment.
6. Push the adjustment slider completely back again, push the volume-setting wheel downwards and release the adjustment slider. Re-insert the cover.
Note:
The change to the factory settings is indicated by the red adjustment slider now visible in the lable window.
29
Page 30
13. Filter pipette 5 ml / 10 ml
A hydrophobic PE filter is used as a safeguard against liquid entering the pipette. Change the filter if it becomes wet or contaminated.
- Use a flat object such as a screwdriver
- Remove the filter without damaging the tip cone.
Remove the filter before autoclaving! The instrument can be operated without a filter.
11. Autoclaving
The pipette is completely autoclavable at 121 °C (250 °F), 2 bar absolute (30 psi) with a holding time of at least 15 minutes, ac­cording to DIN EN 285.
1. Eject the pipette tip.
2. Autoclave the complete pipette without any further disas­sembling. Remove the filter of pipette 5 ml and 10 ml before autoclaving.
3. Allow the pipette to completely cool and dry.
Note:
The effectiveness of the autoclaving must be verified by the user. Maximum reliability is obtained with vacu­um sterilization. We recommend the use of sterilization bags.
Attention!
Prior to autoclaving, the volume adjustment must be set on an available numbered volume (e.g., 11.25 or
11.26 but not between).
If the pipette is autoclaved frequently, the piston and the seal should be greased with silicone grease in order to preserve smooth movement. Please use only the recommended silicone grease, see accessories page 33. If necessary after sterilization, tighten the connection between the hand grip and the pipette shaft.
12. UV sterilization
The unit can withstand the usual output of a UV sterilization lamp. The effects of the UV may cause some color change.
30
Page 31
A
B
D
E
S
C
14. Servicing and Cleaning
14.1. VITLAB® micropipette up to 1000 µl
2. Disassembly and cleaning
1. Unscrew the pipette shaft (S) from the hand grip.
2. Unscrew the upper part of the ejector (A) from the pipette shaft.
3. Pull the shaft (B, C and D) out of the lower part (E) of the ejec­tor.
4. Unscrew the piston unit (B).
Note:
Piston remains connected with piston unit (B)!
5. Remove the seal with spring (C) (this is non-removable on 10
µl pipette models).
6. Clean the parts shown with a mild soap solution or isopropa-
nol and then rinse with distilled water.
7. Allow the parts to dry (max.120 °C/ 248 °F).
8. Grease piston and seal with a very thin layer of silicone grease.
Please use only the recommended silicone grease, see acces­sories page 33.
9. Assemble the ambient temperature parts in reverse order from
above. Piston unit and upper part of the ejector (A, B) should only be hand-tight.
(For illustration purposes only)
1. Servicing
Inspect the pipette tip cone for damage. Inspect the piston and seal for contamination. Test the instrument‘s piston seal. To do this affix a tip, and aspirate a sample. Hold the instrument vertically, with the sample in the tip for approximately 10 s. If a drop forms at the tip orifice, see the troubleshooting guide on page 36.
31
Page 32
S
K
F
G
I
H
F´
14.2. VITLAB® micropipette 5 ml and 10 ml
1. Servicing
Inspect the pipette tip cone for damage. Inspect the piston and O-Ring-seal for contamination. Test the instrument‘s piston seal. To do this, affix a tip, and aspirate a sample. Hold the instrument vertically, with the sample in the tip for approximately 10 s. If a drop forms at the tip orifice, see the troubleshooting guide on page 36.
2. Disassembly and cleaning
1. Remove the entire shaft (S) from the handgrip by rotating at the upper end of the ejector (F) and remove the filter (K) from the bottom part of the shaft (H).
2. Separate the bottom part of the ejector (F‘) by unscrewing it from the upper part of the ejector (F).
3. Unscrew and dismantle the piston unit (G) with the ejector spring ( I ) and the bottom part of the shaft (H).
4. Remove the O-Ring-seal from the piston unit and clean it.
Note:
Do not disassemble piston unit (G) any further!
5. Clean piston unit (G) and lower part of pipette shaft (H) with a soap solution or isopropanol and then rinse with distilled water.
6. Allow the parts to dry (max. 120 °C / 248 °F) and to cool down.
7. Carefully lubricate the inside and outside of the O-ring and mount it on the piston.
8. Assemble the individual components in the reverse order from that described above.
(For illustration purposes only)
32
Page 33
15. Ordering Information and Accessories
Bench-top rack for 6 pipettes
Cat. No. 1672002
Shelf/rack mount for 1 pipette
Cat. No. 1672000
VITLAB® micropipette
Volume Cat. No.
0.5 - 10 µl 1641000 2 - 20 µl 1641002
10 - 100 µl 1641004 20 - 200 µl 1641006
100 - 1000 µl 1641008
0.5 - 5 ml 1641010
1.0 - 10 ml 1641012
15.2. Accessories
15.1. Ordering Information
Filter for VITLAB® micropipette 5 ml,
pack of 25 Cat. No.1672010
Filter for VITLAB® micropipette 10 ml,
pack of 25 Cat. No. 1672012
Silicone grease for VITLAB® micropipette up to 1000 µl
pack of 1 Cat. No. 1672015
Silicone grease for VITLAB® micropipette 5 ml/10 ml
pack of 1 Cat. No. 1672016
33
Page 34
A
B
E
C
D
16. Spare parts
16.1. VITLAB® micropipette up to 1000 µl
Upper part of ejector
Piston unit
Seal with spring
Shaft with ejector spring
Lower Part of ejector
Volume A B C D E
0.5 - 10 µl 1671400 1671411 - 1671431* 1671441 2 - 20 µl 1671401 1671412 1671420 1671432 1671442 10 - 100 µl 1671402 1671416 1671424 1671433 1671443 20 - 200 µl 1671403 1671417 1671425 1671434 1671444 100 - 1000 µl 1671404 1671418 1671426 1671435 1671445
* 0.5-10 µl including seal
Before reordering a piston unit or seal and spring for your VITLAB® micropipette with glass piston (up to serial number 08N), please contact [email protected]. (For illustration purposes only, parts will differ slightly de­pending on nominal volume of instrument.)
34
Page 35
I
H
G
F´
F
Upper part of ejector
Lower part of ejector
Piston unit
Lower part of pipette shaft
Ejector spring
16.2. VITLAB® micropipette 5 ml and 10 ml
(For illustration purposes only, parts will differ slightly depen­ding on nominal volume of instrument.)
Volume F + F‘ G H I
0.5 - 5 ml 1671451 1671455 1671461 1671465 1 - 10 ml 1671452 1671456 1671462 1671465
35
Page 36
17. Troubleshooting
Problem Possible cause Corrective action
Tip dripping (instrument leaks)
- Unsuitable tip - Only use high-quality tips
- Tip not seated tightly - Press tip on firmly
The instrument does not aspi­rate or aspirates too little; the discharged volume is too low
- Seal contaminated - Clean seal
- The seal or cone is damaged - Replace seal or shaft
- The piston is contaminated or damaged
- Clean or replace piston
Aspiration is too slow - Shaft clogged - Clean shaft
- The filter in the 5 ml and 10 ml models is contaminated
- Change the filter
Discharged volume is too large - Pipetting button pressed too
far into the blow-out position before sample uptake
- Operate properly. See ‚Pipetting‘, page 25.
Piston is difficult to move - The piston is contaminated or
needs grease
- Clean and grease piston
36
Page 37
18. Repairs - Calibration Service
If a problem cannot be fixed by following the troubleshooting guide, or by replacing spare parts, then the instrument must be sent in for repair.
For safety reasons, instruments returned for checks and repairs must be clean and decontaminated!
18.1. Return for repair
a) Clean and decontaminate the instrument carefully. b) Complete the „Declaration on Absence of Health Hazards“
(ask your supplier or manufacturer for the form. The form can also be downloaded from www.vitlab.com).
c) Send the completed form along with the instrument to the
manufacturer or to the dealer with an exact description of the type of malfunction and the media used.
The return transport of the instrument is at risk and cost of the sender.
18.2. Calibration Service
ISO 9001 and GLP-guidelines require regular examinations of your volumetric instruments. We recommend checking the volume every 3-12 months. The interval depends on the specific require­ments on the instrument. For instruments frequently used or in use with aggressive media, the interval should be shorter. The detailed testing instruction can be downloaded on www.vitlab.com. VITLAB also offers you the possibility to have your instruments calibrated by the VITLAB Calibration Service. Just send in the instruments to be calibrated, accompanied by an indication of which kind of calibration you wish. Your instru­ments will be returned within a few days together with a test report (VITLAB calibration service) or with a DAkkS Calibration Certificate. For further information, please contact your dealer or VITLAB.
37
Page 38
19. Warranty
We shall not be liable for the consequences of improper handling, use, servicing, operating or unauthorized repairs of the instru­ment or the consequences of normal wear and tear especially of wearing parts such as pistons, seals, valves and the breakage of glass as well as the failure to follow the instructions of the operat­ing manual. We are not liable for damage resulting from any actions not described in the operating manual or if non-original parts have been used.
20. Disposal
For the disposal of instruments and tips, please observe the rel­evant national disposal regulations.
Subject to technical modification without notice. We will not be held responsible for printing or typographical errors.
38
Page 39
Page 40
13 / 1695043 / 0718 / 2
Loading...