VistaLab Ovation Operator's Manual

Operator’s Guide for
0.2-10mL
0.2-10mL
Electronic Single Channel
and Macro Volume Pipettes
L065-0001-001 Rev. F
Introduction
TABLE OF CONTENTS
roduct Description & Specifications 1-1
P
ipette Components 1-2
P
BioNatural Pipetting
Operating Procedures
Calibration
Maintenance & Troubleshooting
Introduction & Ergonomics 2-1
Ovation pipetting 2-3
Picking Up & Holding 2-3 Selecting Liquid Handling Functions 2-3 Acquiring Tips 2-3 Aspirating and Dispensing 2-4 Ejecting Tips 2-4 Using serological or volumetric pipettes (macro models) 2-5
Getting Started, Keypad & Display 3-1 Function Indicators and Setting Indicators 3-2 Set-Up of Liquid Handling Functions 3-2
Set-Up of Pipette Settings 3-9
Introduction 4-1 Determining a New Calibration Factor 4-1 Entering a Known Calibration Factor 4-2 Calibration with Serological or Volumetric pipets 4-2
Cleaning Exterior 5-1 Disinfecting Nozzles 5-1 Replacing a Nozzle (Single Channel) 5-1 Replacing a Nozzle (Macro 5mL and 10mL) 5-2 Replacing a Nozzle Filter (Single Channel) 5-3 Replacing a Nozzle Filter (Macro 5mL) 5-4 Replacing a Nozzle Filter (Macro 10mL) 5-5 Replacing a Nozzle Insert (Macro 10mL) 5-5 Battery Information 5-6 Troubleshooting 5-7
Appendix A
Appendix B
Ovation BioNatural Pipette – Electronic Single Channel Operator’s Guide
Warranty, Safety Compliance & Contact Information A-1 Hazards & Precautions A-2 Operating Temperature & Environment Conditions A-2 Power Supply and Battery Specifications A-2 Chemical Compatibility A-2
Tips & Accessories B-1
ELECTRONIC SINGLE CHANNEL
1
AND MACRO OVATION PIPETTES
Ovation BioNatural™pipettes are the ideal ergonomic solution for handling repetitive and complex liquid handling tasks that contribute to repetitive stress injuries in
today's laboratory. Ovation pipettes increase laboratory efficiency through automated
liquid handling routines, while reducing fatigue and the effect of poor postures.
This manual describes how to use and care for your Ovation electronic single channel or macro pipette. As you can see from its appearance, it is different from every
pipette that you have used before, therefore please read the instructions carefully.
Upon initial receipt of the pipette, the battery must be charged for 90 minutes prior to use. (See instructions on page 5–5). Also, please activate the warranty on the VistaLab Technologies web site – www.vistalab.com. The required on-line form can be found in the “support” menu. Retain all packing materials in the event that the pipette is to be sent back to VistaLab Technologies for calibration verification or service.
Product Description
Electronic single channel and macro Ovation BioNatural Pipettes are adjustable volume, air displacement,
fully electronic motor-driven
dispense precise fluid volumes.
Catalog# Color Range Increments Accuracy* Precision*
20L
10L 0.50% at 10L
1.80% at 2L
125L
62.5L 0.24% at 62.5L
12.5L 1.00% at 12.5L
250L
125L 0.25% at 125L
25L 1.00% at 25L
1250L
625L 0.20% at 625L
Ovation BioNatural Pipette – Electronic Single Channel and Macro Operator’s Guide
1065-0020
1065-0020L
1065-0125
1065-0125L
1065-0250
1065-0250L
Note: Catalog numbers ending in “L” are left-handed models –
1065-1250
the LCD screen is on the opposite side of the pipette.
Yello
Gree
Blue
Purp
Grey
Red
Volume Dispensing
pipettes intended to aspirate and
–1.0% at 20L 0.30% at
0.5 — 20L 0.05L –1.5% at
–3.0% at 2L
–0.8% at 125L 0.15% at
2 — 125L 0.5L –0.8% at
–2.9% at
–0.8% at 250L 0.15% at
5 — 250L 1L –0.8% at
–2.5% at
–0.75% at 1250L 0.15% at
25 — 1250L 5L –0.8% at
1 - 1
Pipette Components
100-
0.2-10mL
Tip Eject
Ejector
Pipetting
Adjustabl
e
Bod
Nozzle
Keypa
Recharging Port
Volume
Description Function
Pipetting Trigger Initiates pipetting action
Tip Eject Button Ejects disposable pipette tips with minimal thumb force
LCD
Captive
Serial
Adjustable Hook Accommodates left and right handed users
Ejector Sleeve Stored energy from tip installation releases tips
Nozzle T
Body Larger grip surface for comfort, eliminating tight hand grip
Keypad Selects and sets up liquid handling functions, settings and calibration
LCD
Battery Lithium ion rechargeable, located in base of pipette (not shown)
Volume Label Identifies volume range
Recharging Port Power supply connection for battery recharging
Power Supply AC adapter for recharging battery (not shown)
Captive Screw When loosened, allows access to inside of pipette for battery replacement
Serial Number Provides positive identification for the pipette
Ovation BioNatural Pipette – Electronic Single Channel and Macro Operator’s Guide
apered, chemically resistant coupling that secure pipette tips
Indicates liquid handling functions, status, settings,
and remaining battery power
1 - 2
BIONATURAL PIPETTING
2
Introduction
Posture
Studies have shown that pipetting is the #1 cause of musculoskeletal disorders in the laboratory – it is, by nature, a repetitive process that puts strain on the body. While we can’t change how repetitive pipetting is, we can and did change how comfortable you are while doing it.
The Ovation BioNatural Pipette is the only pipette designed to keep your hand in the neutral position recommended by ergonomics experts. We call this BioNatural™ pipetting – it allows a lower hand location to ease stress in the shoulder, and a relaxed wrist angle eliminates uncomfortable extension and radial deviation movements in the arm. Force, velocity and exertion from repetition or duration have also been neutralized because of the Ovation pipette’s unique working position and reduced forces required during operation.
Some practice may be required to change years of posture and habits developed using standard axial pipettes; the physical benefits of BioNatural pipetting are worth the practice! When using the Ovation pipette, arm/hand elevation should remain low to the work surface, wrist rotation should not exceed 90º, and hand posture should remain relaxed with the wrist and back of hand slightly flexed.
Arm elevation remains low, minimizing stress to elbow, shoulder and neck
Ovation BioNatural Pipette – Electronic Single Channel and Macro Operator’s Guide
A loose, relaxed grip increases available strength in the hand, improving endurance and productivity during pipetting
Wrist remains in a neutral range of motion throughout all pipetting operations
2 - 1
Keep a “comfort zone” for pipetting
Maximum
Reach
“Comfort Zone”
14”
24”
14”
24”
Ergonomics experts recommend establishing a “comfort zone” of movement for the task of pipetting. Position your body within 9” of the counter surface, and keep the items needed within easy reach. Ideally this should be 14” or less, and should not exceed 24” of occasional reaching. Wrist rotation should not exceed 90˚ from the work surface, and arm/hand elevations should remain low to the counter surface.
1. Relaxed shoulders and upper arms
2. Low thumb forces
3. Relaxed hand grip
Ovation BioNatural Pipette – Electronic Single Channel and Macro Operator’s Guide
4. Neutral forearm and wrist rotation
5. Low arm and elbow height
Ovation comfort-zone benefits:
1. Minimized tension and fatigue
2. Minimized exertion and contact stress
3. Low contact pressure
4. Lowest pressure in carpal tunnel and to median nerve
5. Minimized tension to shoulder and elbow
2 - 2
icking Up the Pipette
P
Selecting Liquid Handling Functions and Changing Settings
The Ovation pipette has been designed
o allow the body of the pipette to fill
t your palm. Rotate the adjustable rest comfortably on your texture of the unit’s back
hook to
forefinger. The
helps reinforce
correct hand positioning.
The unique ergonomic design and adjustable hook is compatible for both right and left-handed use.
Keep hand posture relaxed – there’s no need for a “firm grip” when using an Ovation pipette.
The Ovation pipette’s keypad is easy to use and provides access to all liquid handling functions and settings. Simply press the buttons on the keypad, monitoring the selection of functions or changes to settings on the LCD display.
Acquiring Tips
Complete information and examples of all liquid handling functions and settings are included in “Operating Procedures” - Chapter 3 of this guide.
While maintaining a flat wrist-hand posture, insert the Ovation pipette nozzles into a tip. Press down until you hear or feel a “click”. This “click” indicates that the tip is properly seated and ready for use.
Ovation BioNatural Pipette – Electronic Single Channel and Macro Operator’s Guide
2 - 3
Aspirating and Dispensing
Smooth gliding movements of the upper extremities with proper posture and minimal stress is the key to ergonomically
orrect pipetting. Maintain a low forearm
c elevation and neutral wrist posture throughout the pipetting cycle. With Electronic Single Channel models, Immerse pipette tip to an appropriate depth (1mm
for 0.5–20µL, 2 to 3mm for 2µL–125µL & 5-
50µL, and 2 to 4mm for 25–1250µL).
2
Press & release the pipetting trigger to aspirate sample. Press pipetting trigger again to dispense or mix sample.
Aspirating and dispensing large sample volumes traditionally create ergonomic challenges because of an axial pipette’s design, elongated macro tips, or the overall length of typical serological pipets
The Ovation macro pipette allows the operator to minimize wrist rotation
and
arm elevation.
.
Ejecting Tips
Immerse pipette tip into sample, then press and release the pipetting trigger to aspirate. Wait one second after beep tone, then remove tip and place it into dispensing container at same angle used during aspiration. Press pipetting trigger again to dispense or mix sample.
Note: To interrupt aspirating cycle and/or prevent accidental overfilling of a pipet (as when set volume exceeds capacity of serological pipet), quickly press and release pipetting trigger. PURGE will be on display. Pressing pipetting trigger again then expels any liquid or air bubbles in the tip or pipet.
Ovation pipettes store energy captured during the acquisition of a tip, and use it to discard tip when the tip eject button is pressed. Point pipette tip into a suitable waste container and press the tip eject button to discard the tip.
To minimize forearm elevation and “reaching”, position the waste container below worksurface level or use shallow containers. Raising the arm to discard tips into a tall receptical on the counter may negate some of the benefits of Ovation’s low-profile design.
Ovation BioNatural Pipette – Electronic Single Channel and Macro Operator’s Guide
2 - 4
Using serological or volumetric pipets
(macro models only)
Note: Due to the additional length of the air
olumn or variations in
c markings on serological or volumetric pipets, it may be necessary to adjust the Ovation’s calibration factor. To optimize performance, review the calibration instructions in chapter 4.
When extra length is needed to reach the bottom of tall vessels, glass or
lastic serological or volumetric pipets
p can be used with Ovation macro volume models.
The nozzle on the 0.2-10mL model accommodates both tips and pipets.
nsert the end of a pipet into the nozzle
I until it stops. To remove the pipet, manually pull it out of the nozzle.
The 100-5000mL model requires installation of a Sero-Adapter prior to use with serological pipets. The adapter easily attaches to the pipette in place its standard nozzle (See instructions
below).
Insert the end of a pipet into the Sero-Adapter until it stops. To remove the pipet, manually pull it out of the Sero-Adapter. Re-install the standard nozzle to use pipette tips.
of
Installing a Sero-Adapter
(100-5000µL model only)
1.
Place a macro pipette tip onto the nozzle and push on ejector sleeve until click is heard. Grip tip firmly near the nozzle, and turn it in a counterclockwise direction to unscrew and remove nozzle from pipette. Store nozzle in a safe place.
2.
Carefully screw on adapter (clockwise). Ensure adapter is square to threads for proper installation.
Nozzle
Ovation BioNatural Pipette – Electronic Single Channel and Macro Operator’s Guide
2 - 5
OPERATING PROCEDURES
3
etting Started
G
Pipette Keypad
he general steps to set up and operate an electronic single channel or macro model
T Ovation pipette are:
1. Select a liquid handling function
2. Review and/or change settings as required
3. Acquire tip
4. Run the liquid handling function according to its protocol
5. Eject disposable tip into a proper waste container.
Familiarize yourself with the use of the keypad, display prompts, set-up and use of the pipette’s liquid handling functions in this chapter before use.
Button Description
Advances display to next liquid handling function in the run mode. When in set-up, the button advances display to the next be changed.
setting within the liquid handling function that can
When changing the setting(s) within a liquid handling
UP DOWN
Pipette Display
Displays volume and values during set-up, e.g. # of mix
example
Ovation BioNatural Pipette – Electronic Single Channel and Macro Operator’s Guide
cycles, # of multiple dispenses
Displayed when cal factor been changed from factory setting of 1.000
function, pressing the or button adjusts the settings. When pressed together and held, they are used to access pipette settings.
Description
has
Icon indicates remaining battery capacity
Symbol indicates next portion of aspirate (up) or dispense (down) cycle to be performed
Displays
function
that is active
liquid handling
, setting or prompt
3 - 1