![](/html/fb/fbfc/fbfc87112e5557c11aeaba7cfdffcfd6bfeba2e8e289cfa2479fd8a0792fbfd7/bg1.png)
Installation Guide & Owner’s Manual
Model V7501
For 1
Autoclave
Direct-to-Drain Thermal Reduction
System for Autoclave Wastewater
Model V7502
Includes CSA-Certifi ed
VistaCheck Dual Check
Backfl ow Preventer
For 2
Autoclaves
VistaCool can be used with:
Midmark M3 SciCan STATIM 2000
Midmark M9* SciCan STATIM 5000
Midmark M11*
* Requires Model V7502 and S7560 adapter kit
![](/html/fb/fbfc/fbfc87112e5557c11aeaba7cfdffcfd6bfeba2e8e289cfa2479fd8a0792fbfd7/bg2.png)
Table of Contents
Introduction ....................................................................................................................................................................3
Important Information ..................................................................................................................................................3
Preparation - V7501 .......................................................................................................................................................4
Preparation - V7502 ......................................................................................................................................................5
Typical Installation - V7501 ..........................................................................................................................................6
Typical Installation - V7502 .......................................................................................................................................... 7
Plumbing ......................................................................................................................................................................... 8
Cabinetry ..........................................................................................................................................................................9
Connecting the System ..............................................................................................................................................10
Tank Positioning ...........................................................................................................................................................12
Starting & Using the System .......................................................................................................................................12
Autoclave-Specific Instructions ................................................................................................................................13
Maintenance ..................................................................................................................................................................14
Troubleshooting ...........................................................................................................................................................14
VistaCool Limited Warranty .......................................................................................................................................15
VistaCool System Information & Specifications ....................................................................................................16
Distribution
VistaCool systems are sold through dental and medical dealers throughout North America by:
Midmark Corporation
Midmark.com
60 Vista Drive
Versailles, Ohio 45380
United States of America
PHONE (937) 526-3662
FAX (800) 365-8631
TOLL FREE (800) 643-6275
VistaResearchGroup.com
P.O. Box 321
Ashland, Ohio 44805-0321
United States of America
PHONE (419) 281-3927
FAX (419) 281-7380
SciCan Inc.
SciCan.com
701 Technology Drive
Canonsburg, Pennsylvania 15317
United States of America
PHONE (724) 820-1600
FAX (724) 820-1479
TOLL FREE (800) 572-1211
SciCan Ltd.
SciCan.com
1440 Don Mills Road
Toronto, Ontario M3B 3P9
Canada
PHONE (416) 445-1600
FAX (416) 445-2727
TOLL FREE (800) 870-7777
VistaCool is patent pending.
VistaCool is a trademark of Vista Research Group, LLC.
M3, M9 and M11 are trademarks/models of Midmark Corporation.
STATIM 2000 and STATIM 5000 are trademarks/models of SciCan Ltd.
2
Includes CSA-Certified
VistaCheck Dual Check
Backflow Preventer
TM
Installation Guide & Owner’s Manual
![](/html/fb/fbfc/fbfc87112e5557c11aeaba7cfdffcfd6bfeba2e8e289cfa2479fd8a0792fbfd7/bg3.png)
Introduction
VistaCool is a patent-pending system designed to eliminate the
need for autoclave condenser/waste bottles and condensation
tanks by cooling the exhaust water and automatically sending
it directly down the drain. With the use of a simple adapter kit,
the VistaCool Model V7502 is designed to enable the Midmark
M9 and M11 autoclaves to use fresh deionized/distilled water for
each sterilizing cycle—converting these chambered autoclaves to
operate like quick-turnaround, cassette systems. This eliminates
the repeated use of autoclave water, keeping the chamber
cleaner, while also eliminating reservoir draining and maintenance.
IMPORTANT INFORMATION
VistaCool can be used with:
Midmark M3
Midmark M9*
Midmark M11*
SciCan STATIM 2000
SciCan STATIM 5000
* Requires Model V7502 and S7560 adapter kit
For System Owners
We recommend that a professional technician or plumber familiar with dental/medical oces perform the
installation since interface with a cold water line and drain is involved. He/she should be familiar with local
plumbing codes and techniques for successful dental/medical equipment installations. Failure to install
correctly could result in water leaks, leading to possible water damage. It is the responsibility of the system
owner to validate the use of all equipment within its installed environment.
Please keep the Installation Guide & Owner’s Manual handy for future reference and ensure that anyone
responsible for operation and maintenance of the system is familiar with all details contained in this manual.
Please return the Warranty Registration form or register online immediately upon installation.
IMPORTANT INFORMATION
For System Installers
Please read this entire manual before proceeding with installation and operation, and always follow local
plumbing codes.
VistaCool systems have only been verified for installation on the compatible autoclaves (listed above).
Please check with the manufacturer of other brands of autoclaves or contact Vista Research Group before
connecting to other than SciCan or Midmark autoclaves. See page 13 for autoclave-specific instructions.
VistaCool Model V7502 has been verified for installation on Midmark M9 & M11 units with the use of
the S7560 adapter kit. Only M9, M9D, M11 and M11D (-02X, -03X) units with green displays can be
connected to VistaCool. First generation models with red displays (-0XX) are not compatible.
The VistaCool tank must be installed so that the top of the water tank is ABOVE the connection to the
drain adapter. Please see page 12 for important tank positioning details.
Check for leaks and run the autoclave(s) for at least two consecutive cycles after installation to
ensure the system is operating properly.
© 2000-2015 Vista Research Group, LLC. All rights reserved.
3
![](/html/fb/fbfc/fbfc87112e5557c11aeaba7cfdffcfd6bfeba2e8e289cfa2479fd8a0792fbfd7/bg4.png)
Preparation / V7501
Remove all contents from the VistaCool product carton to make certain all items needed
for installation are present and undamaged.
S6155P
Water Supply Valve
For 1
Autoclave
Tubing
TU751
6’
Cold Water Feed Line
⁄” OD x ⁄” ID
Blue Transparent Polyurethane
TU753
6’
Condensate Drain Line
⁄” OD x 0.170” ID
Black Opaque LLDPE
S7550
M3 Mono Plug
S7508
Inline Thermal
Sensor Assembly
S7539
x4
Cabinet Grommet
TU755
6’
Coolant Overfl ow Drain Line
⁄” OD x ⁄” ID
Clear PVC
S7506
Drain Line Assembly
VC250-A
VistaCheck Dual Check
Backfl ow Preventer
S6130P
Plug
TU748
9’
High Temperature Feed Line
⁄” OD x 0.185” ID
Tefl on®
NOTE:
A fi tting (tee, compression fi tting, etc.) for connecting to cold water supply line is not included in this VistaCool kit
and must be supplied by the installer/plumber. See Plumbing section for more details.
4
TM
Installation Guide & Owner’s Manual
![](/html/fb/fbfc/fbfc87112e5557c11aeaba7cfdffcfd6bfeba2e8e289cfa2479fd8a0792fbfd7/bg5.png)
Preparation / V7502
Remove all contents from the VistaCool product carton to make certain all items needed
for installation are present and undamaged.
S6155P
Water Supply Valve
For 2
Autoclaves
Tubing
TU751
6’
Cold Water Feed Line
⁄” OD x ⁄” ID
Blue Transparent Polyurethane
TU753
6’
Condensate Drain Line
⁄” OD x 0.170” ID
x2
Black Opaque LLDPE
S7550
x2
M3 Mono Plug
S7508
x2
Inline Thermal
Sensor Assembly
S7539
x8
Cabinet Grommet
TU755
6’
Coolant Overfl ow Drain Line
⁄” OD x ⁄” ID
Clear PVC
S7506
Drain Line Assembly
VC250-A
VistaCheck Dual Check
Backfl ow Preventer
S6130P
Plug
TU749
18’
High Temperature Feed Line
⁄” OD x 0.185” ID
Tefl on®
NOTE:
A fi tting (tee, compression fi tting, etc.) for connecting to cold water supply line is not included in this VistaCool kit
and must be supplied by the installer/plumber. See Plumbing section for more details.
© 2000-2015 Vista Research Group, LLC. All rights reserved.