VisionAire 2, 3 Service manual

MN138-1 Rev H
02/14
Oxygen Concentrator Service Manual
AirSep Corporation · 401 Creekside Drive · Buffalo, NY 14228-2085 USA
Telephone: (716) 691-0202 · 24-Hour Fax: (716) 691-4141 · www.airsep.com
ii MN138-1 Rev H
02/14
____________________________________________________________________________________
Expedited Product Warranty Check service is always at your fingertips with AirSep:
http://www.airsep.com/medical/equipmentproviders.html
OR
In the US or Canada,
dial 866-873-9277
AirSep® is a registered trademark of AirSep Corporation.
VisionAire™ is a trademark of AirSep Corporation.
iii MN138-1 Rev H
02/14
Table of Contents
Section 1.0 Introduction
1.1 Equipment Provider Responsibility 1
1.2 Important Notice and Symbol Explanations 2
1.3 Functional Specifications 5
Section 2.0
Operation Check and Oxygen Concentration Test
2.1 Description of Operation 6
2.2 Operation Check 6
2.3 Alarm System 6
2.3.1 Start Up 7
2.3.2 Power Failure Alarm Test 7
2.3.3 No Flow Alarm Test 7
2.4 Oxygen Concentration Test and Specification 7
Section 3.0
General Instructions
3.1 Instructions 8
3.2 Patient, Provider, and Routine Maintenance 9
3.2.1 Product Filter – No Maintenance Required 9
3.2.2 Preparing for New Patient Use/Method 9 of Cleaning and Infection Control
Section 4.0
Main Components
4.1 Components 10
4.2 Cabinet Removal 10
4.2.1 Removing Front Panel 10
4.2.2 Removing Back Panel 10
4.3 Compressor 10
4.3.1 Compressor Replacement 11
iv MN138-1 Rev H
02/14
4.3.2 Capacitor Replacement 12
4.4 Solenoid Valves 12
4.4.1 Valve Manifold Replacement 12
4.4.2 Solenoid Valve Coil Replacement 14
4.5 Sieve Beds 14
4.5.1 Sieve Bed Replacement 14
4.6 Cabinet Fan 15
4.6.1 Cabinet Fan Replacement 15
4.7 Circuit Board 16
4.7.1 Circuit Board Replacement 17
4.8 Product Regulator Check and Setting 18
4.8.1 Setting Product regulator for Normal 18 Operation
4.8.2 Product Regulator Lockout 18
4.8.3 Back Pressure Correction 19
4.8.4 Product Regulator Cleaning or Rebuilding 19
4.9 Circuit Breaker Replacement 20
4.10 l/0 Power Switch Replacement 20
4.11 Hour Meter Replacement 21
4.12 Flow Meter Replacement 21
4.13 Power Cord Replacement 22
4.14 Oxygen Monitor Circuit Board Replacement 23
4.15 Mixing Tank Replacement 23
4.16 Exhaust Muffler Replacement 23
4.17 Caster Replacement 24
4.18 Roller Base Replacement 25
4.19 Air Outlet Option (if equipped) 25
v MN138-1 Rev H
02/14 02/14
Section 5.0
Troubleshooting
5.1 Operation Pressure Test 26
5.1.1 High Operating Pressure 26
5.1.2 Low Operating Pressure 26
5.2 General Troubleshooting 27
5.3 Troubleshooting Chart 28
5.4 Tool Kit and Pressure Test Gauge 31
vi MN138-1 Rev H
02/14
Appendix
Exploded Drawings
Wiring Diagram - VisionAire A
Wiring Diagram – VisionAire 2, VisionAire 3 B
Valve Wire Harness – WH105-1 C
Main Assembly D
Center Section Assembly E
Compressor Enclosure Assembly F
Compressor Assembly G
Valve Manifold Assembly H
1
MN138-1 Rev H
02/14
1.0 Introduction
1.1 Equipment Provider Responsibility
All Equipment Providers of the VisionAire™ Oxygen Concentrator must assume responsibilities for handling, operational check-out, patient instruction, and oxygen concentration checks. These responsibilities are outlined below and throughout this manual.
As an Equipment Provider, you must do all of the following:
! Inspect the condition of each VisionAire unit immediately upon delivery to your business
location. Note any sign of damage on the delivery receipt, and report it directly to both the freight company and AirSep Corporation immediately.
! Check the operation of each VisionAire unit before delivery to a patient. Confirm the
oxygen concentration level is within specifications as referred to in Section 2.4. (Test the power disconnect alarm as described in Section 2.3 of this manual.)
! Deliver VisionAire units only to patients authorized by a licensed health care provider or
physician’s prescription. The VisionAire oxygen concentrator must not be used as a life­supporting device. A backup supply of oxygen must be available.
! Instruct patients how to use VisionAire in conjunction with the Patient Manual. ! Instruct patients to notify their licensed health care provider/physician if they experience
any signs of discomfort.
! Be available to provide service to each patient at any time. ! Establish and implement a protocol to check oxygen concentration.
This unit is not a life-support device. Geriatric, pediatric, or any other patient unable to communicate discomfort while using this oxygen concentrator may require additional monitoring. Patients with hearing and/or sight impairments may need assistance with monitoring the alarms.
This device supplies high-concentration oxygen that promotes rapid burning. Do not allow smoking or open flames within the same room of (1) this device, or (2) any oxygen-carrying accessory. Failure to observe this warning can result in severe fire, property damage, and/or cause physical injury or death.
Use no oil, grease, or petroleum-based or other flammable products on or near nasal end of cannula or on the VisionAire unit. Oxygen accelerates the combustion of flammable substances.
2
MN138-1 Rev H
02/14
1.2 Important Notice and Symbol Explanations
As you read the manual, pay special attention to the WARNING, CAUTION, and NOTE messages. They identify safety guidelines or other important information as follows:
Describes a hazard or unsafe practice that can result in severe bodily injury or death.
Describes a hazard or unsafe practice that can result in minor bodily injury or property damage.
Provides information important enough to emphasize or repeat.
Symbols/Abbreviations
Symbols are frequently used on equipment in preference to words with the intention of lessening any possibility of misunderstanding caused by language differences. Symbols can also permit easier comprehension of a concept within a restricted space. The following table is a list of symbols and definitions that may be used with AirSep’s VisionAire Oxygen Concentrators. These symbols are referenced from the appropriate International Electro-technical Commission (IEC) standards:
Symbol
Description
Symbol Description
ON (power switch on) OFF (power switch off)
No smoking
Do not disassemble
Type B equipment
Consult instructions for use
Warning – Describes a hazard or
unsafe practice that if not avoided can result in severe bodily injury, death or property damage
Class II Device
Caution – Describes a hazard or
unsafe practice that if not avoided can result in minor bodily injury or property damage
Complies with the 93/42/EEC directive drawn up by the approved organization No. 0459
Note – Provides information important enough to emphasize or repeat
Safety agency for CAN/CSA C22.2 No.
601.1 M90 for medical electrical equipment
3
MN138-1 Rev H
02/14
Symbol
Description
Symbol Description
Consult the accompanying documents
Keep unit and accessories dry Use no oil or grease
Proper disposal of waste of electrical and electronic equipment required
Oxygen outlet connection to the cannula
Do not expose to open flames
Caution: Federal law (USA) restricts this for sale or rental by or on the order of a physician or licensed health care provider.
Keep in the vertical position Fragile – handle with care
Oxygen concentration warning LED
See Instructions
Method disposing of waste: All waste from the VisionAire (patient circuit, etc.) must be disposed of using appropriate methods.
Method for disposing of the device: In order to preserve the environment, the concentrator must only be disposed of using the appropriate methods.
Conformity with EN 60-601 (§ 6.8.2 b):
“The manufacturer, assembler, installer or importer is not considered to be responsible themselves for the consequences on the safety, reliability and characteristics of a device unless:
- The assembly, fitting, extensions, adjustments, modifications or repairs have been performed by persons authorized by the party in question.
- The electrical installation of the corresponding premises complies with IEC regulations.
- The device is used in accordance with the instructions for use.” If the replacement parts used for the periodic servicing by an approved technician do not comply with the manufacturer’s specifications, the latter is absolved from all responsibility in the event of an accident. Do not open the device while in operation: risk of electrical shock. This device complies with the requirements of the 93/42/EEC European directive but its operation may be affected by other devices being used close by, such as diathermy and high frequency electro-surgical equipment, defibrillators, short wave therapy equipment, mobile telephones, CB and other portable devices, microwave ovens, induction plates or even remote control toys and more generally electromagnetic interferences which exceed the levels specified by the EN 60601-1-2 standard.
4
MN138-1 Rev H
02/14
Classification
Type of protection against electric shock: Class II Protection from electric shock is achieved by DOUBLE INSULATION. Protective earthing or reliance upon installation conditions is not required.
Degree of protection against electric shock:
Type B Equipment providing a particular degree of protection against electric shock,
particularly regarding:
1) allowable leakage current;
2) Reliability of protective earth connection (if present).
Not intended for direct cardiac application.
Degree of protection against harmful ingress of water:
Equipment provided with an enclosure preventing entry of such an amount of falling liquid as might interfere with the satisfactory and safe operation of the equipment.
Method of cleaning and infection control allowed:
Please refer to the General Instructions section in the VisionAire Service Manual.
Degree of safety of application in the presence of flammable anesthetic gases: Equipment not suitable for such application.
Mode of operation: Continuous duty.
5
MN138-1 Rev H
02/14
1.3 Functional Specifications
Oxygen Concentration: 1-5 liters per minute at 90% +5.5/-3%
1-3 liters per minute at 90% +5.5/-3% (VisionAire 3) -2 liters per minute at 90% +5.5/-3% (VisionAire 2) (Based on 70°F [21°C] at sea level)
Accuracy: ± 10% of indicated flow setting, or ± 200 ml, whichever is greater.
Response Time: Allow 5 minutes to attain maximum oxygen concentration.
Positioning: Operate the unit in an upright position, maintaining at least 12 inches
(30.5cm) of open space on all sides for ventilation.
Dimensions: 20.8 in. high, 14.1 in. wide, 11.5 in. deep
(52.8 cm high, 35.8 cm wide, 29.2 cm deep)
Weight: 30 lb; shipping weight – 37 lb
(13.6 kg; shipping weight – 16.7 kg)
Electrical (+/- 10%):
115VAC, 60 Hz, 3 amps, 290 watts
115VAC, 60 Hz, 3 amps, 190 watts (VisionAire 3) 115VAC, 60 Hz, 3 amps, 200 watts (VisionAire 2) 220-240 VAC, 50 Hz, 1.5 amps, 290 watts 220 VAC, 60 Hz, 1.5 amps, 290 watts 230 VAC, 50 Hz, 1.5 amps, 180 watts (VisionAire 2) 230 VAC, 50 Hz, 1.5 amps, 175 watts (VisionAire 3)
Alarms: Power Failure
Low Oxygen Concentration (with optional Oxygen Monitor)
Low Pressure
High Pressure High Temperature No Flow
Electrical Shock Class II, Type B Protection:
Operating Temperature 41 to 105°F (5 to 40°C) Range:
Storage Temperature -4 to 140°F (-20 to 60°C) Range:
Humidity: 15 – 95% (non-condensing)
6
MN138-1 Rev H
02/14
2.0 Operational Check and Oxygen Concentration Test
2.1 Description of Operation
Air is drawn into the VisionAire Oxygen Concentrator. For overall, sound efficiency, before air enters the compressor, it passes through the unit’s intake muffler. Pressurized air then exits the compressor and moves to the valve manifold. The valve manifold incorporates the use of five two-way solenoid valves which control the flow of air and oxygen in and out of the molecular sieve beds. The unique property of molecular sieve enables it to physically attract (adsorb) nitrogen when air passes through this material, thus producing high concentration oxygen.
There are two sieve beds: while one produces high concentration oxygen, the other is purged of the nitrogen it adsorbed (collected) during this pressure swing adsorption (PSA) cycle. Each adsorber produces oxygen and delivers it to the product tank. Oxygen exits the product tank through a pressure regulator, flow control valve, flow meter, and finally exits the unit. The VisionAire unit delivers up to 95.5% oxygen at the prescribed flow rate.
2.2 Operation Check
AirSep tests every VisionAire Oxygen Concentrator thoroughly after manufacture. You must perform the following test to ensure that no damage occurred in shipping or handling.
Do not use extension cords with this unit or connect too many plugs into the same electrical outlet. The use of extension cords could adversely affect the performance of the device. Too many plugs into one outlet can result in an overload to the electrical panel causing the breaker/fuse to activate or fire if the breaker or fuse fails to operate.
1. Open and inspect all cartons (that contain units) upon delivery. Unpack the unit and
remove it from the carton. Inspect the unit itself for damage. If the exterior of a unit’s carton is damaged, or the unit itself is damaged, note it on the freight bill signed by the driver.
2. Plug in the power cord of the unit, and set the I/0 power switch to the “I” position. Check
to see that the following occurs:
a. A continuous alarm sounds for approximately four seconds. See the
troubleshooting chart in Section 5.0 of this manual if the unit’s alarm does anything other than sound for four seconds.
b. The compressor runs and flow is indicated in flow meter. c. OPTIONAL: The Oxygen Monitor’s amber light will remain on for four
minutes upon startup regardless of unit’s concentration. After four minutes, the light will go out if the concentration level has reached minimum concentration requirements.
d. Perform an oxygen concentration test, as described in Section 2.4.
2.3 Alarm System
The VisionAire Oxygen Concentrator is equipped with an alarm system, which sounds an
intermittent alarm when a power failure or other alarm condition occurs.
The intermittent alarm sounds are displayed in the following table:
7
MN138-1 Rev H
02/14
VisionAire Intermittent Alarm Indicators
Condition
No Power
Low O2
Low Pressure
High Pressure
High Temperature
No Flow
Audible Alarm
One beep
Two beeps
Three beeps
Four beeps
Five beeps
Six beeps
If the high pressure alarm is activated, the compressor and valves will shut down, and the audible four–beep alarm will continue. This alarm remains on until you set the I/0 power switch to the “0” position. Refer to Section 5.0 for alarm identifications, and a list of probable alarm causes.
2.3.1 Start Up
Each time the VisionAire unit is turned on, an alarm should sound for approximately four seconds.
2.3.2 Power Failure Alarm Test
To test the power failure alarm, disconnect the unit from power, set the l/0 power switch to the “I” position. This should activate the audible one-beep alarm.
If the unit has not been used for an extended period of time, it needs to operate for several minutes to recharge the power failure alarm.
2.3.3 No Flow Alarm Test
To test the no flow alarm, place thumb over oxygen outlet to completely block flow for approximately 10-15 seconds. The flow meter ball should drop to zero and the 6-beep alarm should be activated. (Note: not all units were originally manufactured with a no flow alarm. Units equipped with main circuit boards prior to rev. R will not have this feature.)
2.4 Oxygen Concentration Test and Specification
To ensure that the unit’s output of oxygen is within specification, you must perform a test of the oxygen concentration. Test the unit upon delivery to a patient and at periodic intervals. Equipment Providers need to establish and implement a protocol to check oxygen concentration.
1. If a humidifier bottle is used, disconnect it.
2. Connect a calibrated oxygen concentration analyzer to the oxygen outlet.
3. Verify that the product flow rate delivered by the unit matches the patient’s
prescription and does not exceed the capacity of the unit.
4. Set the unit’s l/0 power switch to the “I” position. Allow approximately five minutes
for the oxygen concentration to stabilize. Take oxygen concentration readings and verify levels are within specification at the liter flow being tested.
5. Disconnect the oxygen analyzer, and reconnect the humidifier bottle.
Do not measure oxygen concentration output after the product stream passes through a humidifier bottle, or erroneous readings will result.
8
MN138-1 Rev H
02/14
3.0 General Instructions
3.1 Instructions
It is important that patients thoroughly understand how to operate the AirSep VisionAire unit. This enables proper treatment as prescribed by a qualified, licensed health care provider/physician. If patients experience any discomfort or the unit alarms, they must notify their licensed health care provider/physician immediately.
You, as the Equipment Provider, are responsible to see that each patient receives the Patient Manual. Explain each step in the operation of the unit to the patient in reference to this manual.
Smoking while using oxygen is the number one cause of fire, injury, and death. You must follow these safety warnings:
Do not allow smoking, candles, or open flames within the same room of the device or oxygen-carrying accessories.
Smoking while wearing an oxygen cannula may result in facial burns and possibly death.
Removing the cannula and placing it on surfaces such as bedding, sofas, or other cushion material will cause a flash fire when exposed to a cigarette, heat source, or flame.
If you smoke, you must follow these 3 life-saving steps: turn off the oxygen concentrator, take off the cannula, and leave the room where this device is located.
“No Smoking – Oxygen in Use” signs must be prominently displayed in the home or where the oxygen concentrator is in use. Patient and their caregivers must be informed about the dangers of smoking in the present of, or while using, medical oxygen
This device supplies high-concentration oxygen that promotes rapid burning. Do not allow smoking or open flames within the same room of (1) this device, or (2) any oxygen-carrying accessory. Failure to observe this warning can result in severe fire, property damage, and /or cause physical injury or death.
Electrical shock hazard. Disconnect the power cord from the electric outlet before you clean the unit to prevent accidental electrical shock and burn hazard. Only your Equipment Provider or a qualified service technician should remove the covers or service the unit.
Care should be taken to prevent the VisionAire from getting wet or allowing any liquid to enter the unit. This can cause the unit to malfunction or shut down, and cause an increased risk for electrical shock or burns.
Loading...
+ 32 hidden pages