Vishay G4, G4-PM, G4-HE, G4-DT Technical Manual

G4 Multi Channel
Force Instrument
Program version 101.1.0.0
Technical Manual
RM type
G4 Multi Channel Force Instrument
1. Introduction
Manual updates ..................................... 1-1
General .................................................. 1-1
Functions ............................................... 1-2
Maintenance .......................................... 1-4
Safety information .................................. 1-4
Technical data ....................................... 1-5
Ordering information .............................. 1-12
2. Installation
Mechanical installation ........................... 2-1
Electrical installation .............................. 2-2
CPU unit ................................................. 2-3
VIEWPAN module ................................. 2-5
WF IN, WF IN2 and HS WF2 ................. 2-6
AOUT1 and AOUT4 ............................... 2-9
DIO8 ...................................................... 2-10
Profibus-DP Fieldbus Adaptor ................ 2-11
DeviceNet Fieldbus Adaptor .................. 2-13
3. Instrument Functionality
General ................................................... 3-1
Function Block ........................................ 3-1
4. Set-up
General .................................................. 4-1
Operators interface ................................ 4-2
Menu structure ....................................... 4-6
Parameters ............................................ 4-8
5. Calibration
General .................................................. 5-1
Common parameters ............................. 5-2
Calibration using datasheet ................... 5-3
Using parameters values from
a previous calibration ............................. 5-4
Calibration with known forces ................. 5-5
6. Operation
General .................................................. 6-1
Power supply ......................................... 6-1
Power-up sequence ............................... 6-1
Operating display ................................... 6-2
Security locks ........................................ 6-3
Zero setting ............................................ 6-3
Main menu ............................................. 6-4
Level supervision ................................... 6-5
Use of inputs and outputs ...................... 6-6
Filter function ......................................... 6-7
7. Communication
General .................................................. 7-1
Serial interface ........................................ 7-1
Modbus RTU Slave ................................ 7-1
Modbus TCP Slave ................................ 7-2
Modbus protocol .................................... 7-3
Fieldbus interface .................................... 7-28
8. Remote Access
General .................................................. 8-1
Browser requirements ............................. 8-1
Using the Remote Access ...................... 8-2
Security ................................................... 8-2
Remote Access Login and Logout .......... 8-3
Remote / Local Access ........................... 8-6
Remote Set-up ........................................ 8-7
Remote Access Maintenance ................. 8-10
9. Maintenance
General .................................................. 9-1
Diagnostics ............................................. 9-1
File handling ........................................... 9-4
Create Backup ....................................... 9-5
Restore Backup ..................................... 9-5
Set Default .............................................. 9-5
Program Upgrade ................................... 9-5
Instrument Restart .................................. 9-6
10. Troubleshooting
General ................................................... 10-1
Error codes ............................................ 10-1
Appendix
Declaration of Conformity ........................ App.1
Contents
Technical Manual
PRECAUTIONS
READ this manual BEFORE operating or servicing this instrument. FOLLOW these instructions carefully. SAVE this manual for future reference.
DO NOT allow untrained personnel to operate, clean, inspect, maintain, service, or tamper with this instrument.
WARNING Only permit qualified personnel to install and service this instrument.
Exercise care when making checks, tests and adjustments that must be made with power on. Failing to observe these precautions can result in bodily harm.
!
G4 Multi Channel Force Instrument
1-1
1. Introduction
Manual updates
This manual is updated for the 101.1.0.0 program version. In this program is a web server added which enables for the user to access the instrument using a PC running a web browser. This manual can also be used for the previous program version (101.0.0.0) where the only differences are the ‘Network Config.’ menu, the Instrument restart functionality and the remote access part.
General
The G4 Instrument is a high performance multi channel force indicator intended for industrial systems. Its basic function is to convert the signals from strain gauge transducers to useful force information. Transducer excitation is included as well as parameter controlled signal processing, indication of output levels, error supervision and operation of optional external equipment. The instrument can be equipped with up to 6 synchronized force channels.
The instrument is modular and can be equipped with different types of I/O units to match the demands in the specific applications. There are strain gauge transducer interface modules, a digital input/output module, analog output modules. Internal solid-state outputs in the instrument can be used for output functions from level supervision or ‘In process’ indication, reporting the operating status of the G4 Instrument.
The CPU unit in the instrument has several communication interfaces. It has two serial communication ports, an Ethernet port, an USB port and a fieldbus slot. Several G4 Instruments can be controlled from a master computer or PLC. Serial communication interfaces are RS-485 and RS-232 using Modbus RTU protocol. The Ethernet interface is using Modbus TCP and the optional fieldbus interface uses Profibus or DeviceNet.
CPU
unit
I/O
unit
1
I/O
unit
2
I/O
unit
3
Service panel with
Power supply unit
Backplane
DIN rail mount version
Enclosure & mounting accessories
Block structure of a G4 Instrument
Technical Manual
1-2
It is possible to load new software into the instrument using the USB port. All functions in the G4 Instrument are controlled by set-up parameters. Setting of
parameter values can be performed via a service panel with display and 4 keys (VIEWPAN module).
The Rail mount instrument type is powered with 24 V DC. All input and output signals are galvanically isolated from the power supply by operational insulation.
Functions
Measurement with strain gauge transducers.
Both excitation voltage and output signal are measured at the transducer to avoid influence from voltage drop in the connection cable. Excitation to the transducer, from the G4 Instrument is provided over separate wires. A shielded 6-wire cable must be used to connect a distant transducer to the instrument.
A/D conversion. The analog signals from the transducer are converted to digital form and filtered to give an internal transducer signal with high resolution.
Calculation. The transducer excitation and signal values are combined to form an internal transducer signal, representing the load on the transducer. Influenced by calibration data, this signal is converted to a digital measurement value, the force value, which can be presented at the local display window and at external equipment.
Error supervision. As long as the error supervision detects no error, the signal ‘In process’ is present but if an error is detected, ‘In process’ will be off and a specific error message will be displayed. ‘In process’ can be set to control any digital output. Note that there are force channel specific and instrument specific error detection.
Levels. 32 level comparators in the instrument can be set to switch at defined signal levels with any selected hysteresis added, meaning that the switch level can be different for increasing and decreasing signal. It’s possible to set level switching delays. Output signals from these comparators are available on the serial communication. The level comparator outputs can also be set to control digital outputs from the instrument.
Communication. The G4 Instrument utilizes the serial interface, Ethernet and a fieldbus interface for communication with control computer. The serial interface consists of a RS-232 (COM1) connection and a RS-485/RS-422 (COM2) connection. COM2 can be used with 2- or 4-wire connection. Force values, web tension values, level status, error status etc. can be collected and commands given through the communication interfaces. Modbus RTU protocol is used for the serial interfaces and Modbus TCP for the Ethernet connection. For the optional fieldbus interface Profibus or DeviceNet can be used.
G4 Multi Channel Force Instrument
1-3
Instrument modes.
In normal operation mode the G4 Instrument is presenting the measurement values on the front panel (VIEWPAN) alphanumerical display. Values for one Function Block at a time can be shown. During parameter set-up the instrument will continue normal operation. However if hardware set-up parameters have been changed the instrument will be restarted. The operator will always be notified before the instrument is restarted.
Parameter setting.
In the instrument all operating functions are controlled by set-up parameters with numerical values, string values, or pre-selected values from a list of alternatives. Parameter set-up is performed by the keys at the service panel (VIEWPAN).
Presentation.
The instrument can present measured or calculated values, parameter settings etc. at the front panel. An extensive system of menus gives the possibility to present various information about the instrument. The VIEWPAN module can present one or two values at a time.
Measured or calculated values, status of levels and so on, can be transferred to external equipment via the different communication interfaces (some are optional).
Technical Manual
1-4
Maintenance
The G4 instrument needs no maintenance, performed by the end-user. Any service or repair work must be performed by qualified personnel. Contact your supplier.
Cleaning
Before cleaning the G4, break the power connection to the instrument. Use a soft cloth to clean the exterior of the instrument.
Safety information
Utilization.
Before connecting power to the instrument, check that all fixation screws at the modules are tightened so that the instruments functional grounding by the housing is maintained.
The instrument may only be utilized for the measurement and control functions, described in this Technical Manual. It is especially important to adhere to the load limits of the input/output connectors. We accept no responsibility for any damage arising from improper operation.
Any changes to the instrument, which causes any function changes, may only be carried out by the manufacturer, or after discussion with and permission by the manufacturer.
Meaning of symbols, used in this manual
Direct current.
!
Caution, risk of danger. Documentation needs to be consulted.
G4 Multi Channel Force Instrument
1-5
Technical data
Enclosure type RM – Rail Mount enclosure
Dimensions
Enclosure design Aluminium housing with plastic frames Rail mount DIN 46 277/3 and DIN EN 50022 (w=35 mm, h=7.5 mm) Display, keyboard See VIEWPAN module Environmental Temperature range
Rated performance: -10 to +50 °C Storage: -25 to +85 °C
Relative humidity
Max. 85% up to 40 °C, decreasing linear to 50% at 50 °C Rated pollution Pollution degree 2 Protection IP20. For indoor use. Altitude Up to 2000 m EMC, RF CE (Industrial), OIML
Technical Manual
1-6
VIEWPAN
Module type Display/keypad interface with integrated power supply for the
complete instrument.
Input voltage 24 V ±15% including fluctuations, 40W
Impulse withstand (overvoltage) category I of IEC 60364-4-443
Output voltage 24 V output 0.1 A
The same voltage as input voltage
Display
2 x 16 character LCD with backlight
Keyboard 4 keys
CPU
Module type CPU module
RTC backup battery
Manufacturer Type Lithium battery Panasonic-BSG CR2032 CR2032 3V GP Batteries CR2032 Varta CR2032 (V)
COM1 (RS232) and COM2 (RS485)
For process data and control
Isolated by operational insulation Protocol Modbus RTU Baud rate Up to 115 kbaud
Fieldbus
For process data and control (optional) Types Profibus or DeviceNet
USB
Version 1 Keyboard USB keyboard for PC USB Memory USB type for PC
For backup and restore of set-up parameters
For change to a new program version
Ethernet
10/100BASE-T. For process data and control Protocols Modbus TCP, ftp, http RJ45 Indications Green LED: Transmitting. Yellow LED: Receiving.
G4 Multi Channel Force Instrument
1-7
WF IN / WF IN2
Module type Weight/Force input module Max. # of transducers 8 (350 ohm) per channel
Maximum 48 transducers per instrument Excitation voltage: 5 VDC A/D conversion: 3.9 kHz, 16 000000 units (24 bits) Input range +/- 7 mV/V Update rate:
9.3 − 300 readings per second
No. of channels: WF IN has 1 Weight/Force channel
WF IN2 has 2 Weight/Force channels Sensitivity:
0.1 µV Zero drift: <10 nV/V/K Span drift: <2 ppm/K Digital I/O 4 inputs, 24 V ±15%, 5 mA from external power supply,
isolated by operational insulation and with common return 2 outputs, 24 V ±15%, max 100 mA from external power supply, isolated by operational insulation and with common return
Technical Manual
1-8
HS WF2
Module type High Speed Weight/Force input module Max. # of transducers 4 (350 ohm) per channel Excitation voltage: 10 VDC A/D conversion: 20 kHz, 16 000000 units (24 bits) Input range +/- 4.5 mV/V Update rate:
25 − 800 readings per second
No. of channels: HS WF2 has 2 Weight/Force channels,
separately isolated by operational insulation
Sensitivity:
0.1 µV Zero drift: <10 nV/V/K Span drift: <2 ppm/K Digital I/O 4 inputs, 24 V ±15%, 5 mA from external power supply,
isolated by operational insulation and with common return 2 outputs, 24 V ±15%, max 100 mA from external power supply, isolated by operational insulation and with common return
G4 Multi Channel Force Instrument
1-9
DIO8
Module type Digital input/output module Separate I/O module 2 units can be used Digital I/O 8 inputs, 24 V ±15%, 5 mA from external power supply,
isolated by operational insulation and with common return 8 outputs, 24 V ±15%, max 100 mA from external power supply, isolated by operational insulation and with common return
AOUT1 / AOUT4
Module type Analog output module Number of channels 1 or 4 channels,
separately isolated by operational insulation Resolution 65000 units, 16 bits Voltage output 0 – 10 V, -10 – 10 V, > 1 kohm load Current output 4 – 20 mA, 0 – 20 mA, -12 – 20 mA, -20 – 20 mA,
< 500 ohm load Update rate Function Block update rate, adjustable smoothing filter
Technical Manual
1-10
Profibus-DP
Module type Profibus-DP fieldbus adaptor Connector Profibus 9-pin, female D-sub (DB9F) Baudrate Auto setting 9.6 kbps – 12 Mbps Address 1 – 125, set by parameter Fieldbus data 16 bytes from fieldbus to instrument.
32 – 244 bytes from instrument to fieldbus (may be limited by the master).
See chapter ‘7 Communication’ section ‘Fieldbus communication interface’ for details on fieldbus data mapping.
Mounting The fieldbus adaptor is mounted through the front of
the CPU module with LED’s and connector accessible through the CPU front panel.
Remove the plastic cover from the fieldbus slot in the CPU module front panel. Insert the adaptor very carefully and make absolutely sure that the adaptor slides correctly into the guides in the connector on the CPU PCB. Tighten the two fastening screws at the adaptor front and check that the two securing hooks locks into the CPU PCB.
Settings All fieldbus settings are done with setup parameters in
the instrument. No settings are done on the module itself.
G4 Multi Channel Force Instrument
1-11
DeviceNet
Module type DeviceNet fieldbus adaptor Connector 5 pin male connector. Baudrate 125, 250, 500 kbps or Auto. Set by parameter. Address 0 – 63, set by parameter Fieldbus data 16 bytes from fieldbus to instrument.
32 – 244 bytes from instrument to fieldbus (may be limited by
the master).
See chapter ‘7 Communication’ section
‘Fieldbus communication interface’ for details on fieldbus data
mapping. Mounting The fieldbus adaptor is mounted through the front of
the CPU module with LED’s and connector accessible through
the CPU front panel.
Remove the plastic cover from the fieldbus slot in the CPU
module front panel. Insert the adaptor very carefully and make
absolutely sure that the adaptor slides correctly into the guides
in the connector on the CPU PCB. Tighten
the two fastening screws at the adaptor front and check
that the two securing hooks locks into the CPU PCB. Settings All fieldbus settings are done with setup parameters in
the instrument. No settings are done on the module itself. Bus Supply Voltage According to the DeviceNet (Node) Specification:
nominal 24 VDC, range 11 – 25 VDC.
Technical Manual
1-12
Ordering information
For Rail mount (RM) instruments
G4-RM-FB-S1-S2-S3-V-SW G4 Instrument type G4 RM Enclosure type RM Rail mount FB Fieldbus interface 0
P D
None Profibus DeviceNet
Si Slot 1 to 3 type 0
2 3 4 6 7 8
Blank HS WF2 High speed weight/force input module, 2 ch. WF IN Weight/force input module, 1 channel. WF IN2 Weight/force input module, 2 channels. AOUT1 Analog Output module, 1 channel AOUT4 Analog Output module, 4 channels DIO8 Digital Input and output module
V User interface and
power
V VIEWPAN, 24VDC
SW Software (note 1)
W F S
(No code specified) Standard Weighing Program Standard Weighing Program Standard Force Program Special Program (note 2)
Example: G4-RM-0-4-8-0-V-F
G4 instrument (G4)
Rail mount (RM)
No field bus (0)
Slot 1 = WF IN2 (4)
Slot 2 = DIO8 (8)
Slot 3 = Blank (0)
Power =VIEWPAN unit (V)
Software = Force program (F)
Note 1: The Software (SW) code specifies the program function that should be installed in the instrument. If no code is specified the standard Weighing program will be supplied.
Note 2: If a special program option is desired, the number of the special program must be specified when ordering.
G4 Multi Channel Force Instrument
1-13
Separate modules Spec.no
Module type Module name
110 544 CPU CPU unit 110 546 HS WF2 High speed dual Weight/Force input module 110 547 WF IN Single Weight/Force input module 110 548 WF IN2 Dual Weight/Force input module 110 549 AOUT1 Single channel analog output module 110 550 AOUT4 Four channel analog output module 110 551 DIO8 Digital input/output module 110 552 BLANK Blank panel 110 554 VIEWPAN Rail mount service panel 110 559 PROFIBUS-DP Profibus DP fieldbus adaptor 110 560 DEVICENET DeviceNet fieldbus adaptor
Module selection rules
Every system needs 1 VIEWPAN module and 1 CPU module (can be equipped with one fieldbus adaptor).
Limitations on number of I/O modules that can be used in one instrument:
Maximum 3 modules (not counting the VIEWPAN module).
Maximum 3 WF IN / WF IN2 or 2 HS WF2 modules.
HS WF2 and WF IN(2) cannot be mixed in the same system
Maximum 1 AOUT1 or 1 AOUT4 module.
Maximum 2 DIO8 modules.
Technical Manual
1-14
G4 Multi Channel Force Instrument
2-1
2. Installation
Mechanical installation
See section ‘1. Introduction – Technical data’ for references to RM mechanical measures.
G4, RM type, should be used in a dry, clean environment or it must be mounted in a cabinet, protecting from water, dust etc. The instrument is prepared for mounting on a flat surface by DIN rail according to Technical data. The instrument needs minimum 200 mm space above this surface. To remove the instrument from the rail, the black lever at the lower bottom side of the instrument must be engaged.
Each instrument should have a free space of minimum 30 mm to the left and to right, 100 mm above and below the instrument. Cable ducts may be mounted in this free space.
All electrical connections to the instrument modules are made to terminals at the front side so enough room for these terminals should be arranged.
CABLE DUCT
LOCK LEVER
TERMINALS
100
100
200
RM instrument.
Recommended free space at the instrument.
Technical Manual
2-2
Electrical installation
The field wiring of the instrument shall be suitable to the environment (e.g. chemically) in the end-user application.
Field wiring installation shall comply with any national regulations, hereunder National Electrical Code (NEC) for US and/or Canadian Electrical Code for Canada.
A switch or circuit-breaker shall be included in the building installation.
The switch shall be in close proximity to the equipment and
within easy reach of the operator.
The switch shall be marked as the disconnecting device for the equipment.
The equipment switch or circuit-breaker employed as disconnecting device shall
comply with relevant requirements of IEC 60947-1 and IEC 60947-3.
For electrical installation with an external dc supply, see page 2-5.
WARNING Make sure that that the power to the instrument is turned off before:
- any modules are removed from or inserted in the instrument.
- any connections are connected to or disconnected from the instrument.
All modules should be regarded as ESD sensitive. Make sure that an ESD safe environment is maintained when inserting modules, removing modules and when handling modules separately from the instrument. Modules must be kept in metallised ESD bag when not mounted in the instrument.
!
G4 Multi Channel Force Instrument
2-3
!
CPU unit
External computing devices connected to the CPU communication interfaces of the instrument have to comply with the standard, UL 60950.
The internal battery in the CPU module is to be used only in the equipment where servicing of the battery circuit and replacement of the lithium battery will be done by a trained technician.
COM1
RS232 Serial communication. This is a SELV/SELV-E circuit. COM1 can be used for serial
communication with computer/PLC (Modbus RTU).
Point to point communication, only one G4 unit connected to the computer/PLC.
Connections are made to terminals 7 – 9. Shielded cable must be used. Connect shield to terminal 10.
COM2
RS485 Serial communication for 2-wire or 4-wire with common 0 V.
This is a SELV/SELV-E circuit. COM2 can be used for serial
communication to computer/PLC (Modbus RTU).
Connections are made to terminals 1 – 5. Shielded cable must be used. Connect shield to terminal 6.
The communication lines must be terminated in both ends. If G4 is connected at the end of the communication line switches must be set as:
2-wire termination: Both T2 ON, both R2 OFF. 4-wire termination: Both T2 ON, both R2 ON.
TX RX 0V Shield
COM1 RS-232
10
9
8
7
(DTE) RS-232 9p 25p
2 3 3 2 5 7
RS485 2-wire
1
T­T+ R­R+ 0 V Shield
COM2
6
2 3 4 5
+5
T2
T2
0
+5
R2
R2
0
RS485 4-wire
1
T­T+ R­R+ 0 V Shield
COM2
6
2 3 4 5
+5
R2
R2
0
+5
T2
T2
0
Technical Manual
2-4
Field Bus
Slot for optional Fieldbus interface. Profibus DP-V1 and DeviceNet are available.
See section Profibus-DP Fieldbus Adaptor or DeviceNet Fieldbus Adaptor later in this chapter for details.
USB
Connector for USB device(s). This port has no operational insulation and should be considered as a
SELV/SELV-E circuit. Allows connection of following devices:
1 - USB Memory 2 - USB Keyboard An USB hub can be used to allow connection of more than one device.
Ethernet
This is a SELV circuit. Uses crossover category 5 cable from RJ-45 Ethernet port on the CPU front panel to PC (point to point connection) or standard cable to connect to other equipment through a switch, hub or router which is isolating the circuit from the public network.
G4 Multi Channel Force Instrument
2-5
VIEWPAN module
The output of the external dc supply must be rated 24 V , ±15% including fluctuations, min. 40 W. The supply must provide Double Insulation between Mains parts and 24 V SELV or SELV-E Circuit, and a limited-energy circuit (maximum available current of 8 A). For the US market this energy limit can be achieved with an ANSI/UL248-14 fuse rated 5A. For other markets an IEC 60127 T type fuse rated 4A may also be used.
With integrated 24 V
power supply unit.
The integrated power supply is used to supply the complete instrument.
24 V input
Terminals 1, 2 and 3. The G4 instrument should be powered by 24 V
, connected according to the diagram below. To achieve functional grounding, terminal 3 should be connected to ground.
See Technical data for input voltage ratings.
24 V output
Terminals 4 and 5 can be used to supply max. 100 mA to digital inputs and outputs. Connection should be according to the diagram below. See Technical data for ratings.
1
DC SUPPLY
2
0 V
3
Shield
4 5
0 V
6
Shield
0 V
Ground
0 V
+24 V In
+24 VDC In
+24 VDC Out+24 V Out
!
Technical Manual
2-6
WF IN, WF IN2 and HS WF2
The voltage levels on connectors of I/O modules shall not exceed hazardous voltage levels of 30 Vrms, 42.4 Vpeak or 60 Vdc under normal conditions. In wet locations these voltage levels shall not exceed 16 Vrms, 22.6 Vpeak or 35 Vdc.
Transducer inputs
Terminals 17 – 23 (channel 1), 10 – 16 (channel 2). Transducer connection should be handled with great care to achieve good measuring data. Transducer integrated cables may not be shortened.
NOTE! Transducer cables must be routed at least 200 mm away from 230/400 V, 50/60 Hz power cables. By cables with other frequencies or high power, an even wider distance is preferable.
4-wire connection can be used if the transducer integrated cable is long enough to be connected directly to a transducer input. At 4-wire connection, some terminals must be interconnected as shown in figure on next page.
6-wire connection should be used if the integrated cable must be lengthened or if several transducers should be connected to one transducer input.
The channel 1 cable shield must be connected to terminal 21 and the channel 2 cable shield must be connected to terminal 14.
In WF IN and WF IN2 the shield terminals are internally connected to the G4 housing, which is internally connected to earth via the power supply connector pin 3 (shield). The shield shall not be connected at any other point.
In HS WF2 the transducer input channels are separately insulated by operational insulation and the shields can be connected to the most convenient ground/earth point. This can be the junction box when using multiple transducers or at the barrier ground when using Ex zener barriers.
In the junction box SL-4 from Nobel Weighing Systems, see figure, all necessary terminals and interconnections are provided.
!
G4 Multi Channel Force Instrument
2-7
10
15
13
12
11
14
16
Exc.-
Exc.+
Sign.+ Sign.-
4-wire connection
Transducer
Transducer input, ch. 2
Exc+ Sense+
Exc­Sense­Sign+ Sign­Shield
Junction box
17
22
20
19
18
21
23
Transducer input, ch. 1
Exc.-
Exc.+
Sign.+ Sign.-
Transducer
Transducer
Exc.-
Exc.+
Sign.+ Sign.-
6-wire connection
Exc+ Sense+
Exc­Sense­Sign+ Sign­Shield
A transducer may be connected directly to terminals at the transducer input.
For several transducers or long distance, a junction box
and lengthening cable is needed.
For a HS WF2 channel, the shield can be connected to
ground/earth at any point
Technical Manual
2-8
Digital inputs
Terminals 6 – 9 and terminal 5 (ICom) as a common connection. Four digital inputs are provided, with functions that can be set in the G4 set-up.
External sourcing (24 V
) from the instrument power supply (max. 100 mA) or from a separate DC supply must be used. Note that either the positive or the negative pole of the voltage source (24 V
) can be connected to ICom (5).
Shielded cable/cables must be used, with the shield connected to terminal 4.
Solid state relay outputs
Terminals 1 and 2 with terminal 3 (OCom) as a common connection. Two digital (relay) outputs are provided with contact rating given in Technical data.
External sourcing (24 V
) from the instrument power supply (max. 100 mA) or from a separate DC supply should be used. Note that either the positive or the negative pole of the voltage source (24 V
) can be connected to OCom (3).
Shielded cable/cables must be used, with the shield connected to terminal 4.
O1 O2 OCom Shield ICom I1 I2 I3 I4
1
8
5
4
3
2
9
7
6
Load
Load
+
-
-
+
-
+
-
+
24 V
24 V
G4 Multi Channel Force Instrument
2-9
AOUT1 and AOUT4
The voltage levels on connectors of I/O modules shall not exceed hazardous voltage levels of 30 Vrms, 42.4 Vpeak or 60 Vdc under normal conditions. In wet locations these voltage levels shall not exceed 16 Vrms,
22.6 Vpeak or 35 Vdc.
Analog output unit
The AOUT4 unit has 4 analog output channels, independently isolated by operational insulation. The AOUT1 has 1 output channel, isolated by operational insulation.
The analog output signal will be connected to terminals 10, 11 (channel 1) 7, 8 (channel 2) 4, 5 (channel 3) 1, 2 (channel 4).
Shielded cable/cables must be used and the shield(s) must be connected to the shield terminal(s) 12, 9, 6 and 3. The shields can be connected to the most convenient ground/earth point.
OUT 0V Shield
AOUT1 Channel 1
10 11 12
+
OUT 0V Shield
Channel 1
10 11 12
OUT 0V Shield
Channel 2
7 8 9
OUT 0V Shield
Channel 3
4 5 6
OUT 0V Shield
AOUT4 Channel 4
1 2 3
+
+
+
+
!
Technical Manual
2-10
DIO8
The voltage levels on connectors of I/O modules shall not exceed hazardous voltage levels of 30 Vrms, 42.4 Vpeak or 60 Vdc under normal conditions. In wet locations these voltage levels shall not exceed 16 Vrms, 22.6 Vpeak or 35 Vdc.
Digital inputs
Terminals 11 – 18 and terminal 19 (ICom) as a common connection. Eight digital inputs are provided, with functions that can be set in the G4 set-up. External sourcing (24 V
) from the instrument power supply (max. 100 mA) or from a separate DC supply must be used. Note that either the positive or the negative pole of the voltage source (24 V
) can be connected to ICom (19). Shielded cable/cables must be used, with the shield connected to terminal 20.
I1 I2 I3 I4 I5 I6 I7 I8 ICom Shield
-
­+
+
11 12 13 14 15 16 17 18 19 20
24 V
Solid state relay outputs
Terminals 1 - 8 with terminal 9 (OCom) as a common connection. Eight digital (relay) outputs are provided with contact rating given in Technical data.
External sourcing (24 V
) from the instrument power supply (max. 100 mA) or from a separate DC supply must be used. Note that either the positive or the negative pole of the voltage source (24 V
) can be connected to OCom (9).
Shielded cable/cables must be used, with the shield connected to terminal 10.
O1 O2 O3 O4 O5 O6 O7 O8 OCom Shield
Load
Load
Load
Load
Load
Load
Load
Load
+
-
+
-
1 2
4
3
5 6 7 8 9
10
24 V
!
G4 Multi Channel Force Instrument
2-11
Profibus-DP Fieldbus Adaptor
Profibus module front view
(1) Operation mode LED. (2) Status LED. (3) Profibus connector.
Operation mode LED
State Indication Off Not online / No power Green On-line, data exchange Flashing Green On-line, clear Flashing Red (1 flash) Parameterization error Flashing Red (2 flashes) Profibus configuration error
Status LED
State Indication Off No power or not initialised Green Initialised Flashing Green Initialised, diagnostic event(s) present Red Exception error
OP ST
PROFIBUS DP-V1
(1)
(2)
(3)
Technical Manual
2-12
Profibus connector (DB9F)
Pin Signal Description 1 - -
2 - ­3 B line Positive RxD/TxD, RS485 level 4 RTS Request to send 5 GND Bus Ground (isolated) 6 + 5V Bus Output +5V termination power (isolated) 7 - ­8 A line Negative RxD/TxD, RS485 level 9 - ­Housing Cable shield Internally connected to the Anybus protective
earth via cable shield filters according to the Profibus standard.
For connection of the adaptor to the Profibus master, use a Profibus standard cable and connector according to the diagram below.
For reliable fieldbus function, line termination must be arranged in both ends of the transmission line. For a G4 instrument, at the end of the cable, a connector with line termination should be used. For all other G4 Instruments connection without line termination should be used.
For configuration of the adaptor, a GSD file (VISH0AB3.GSD) is available and should be installed in the master.
G4 Modular
Instrument
D-sub
A-line B-line
8 3
shield
Loading...
+ 136 hidden pages