Victor VC9808plus Operation Manual

DIGITAL MULTIMETER OPERATION MANUAL
1. GENERAL
The instrument is a stable and good performance digital multimeter driven by battery. It uses the LCD with 42mm-high figure to make the reading clear. Unit symbol displaying, backlight and the function of overload protection make operation is convenient. The instrument has the function of measuring DCV, ACV, DCA, ACA, resistance, capacitance, inductance, temperature and frequency, and diode, triode and continuity test. The instrument takes dual-integral A/D converter as key point, is an excellent tool.
2. SAFETY NOTES
This series meter meets the standard of IEC1010. Please read below notes carefully before operation. 1Do not input a limited voltage which RMS is over DC 1000V or AC 700V when measuring voltage. 2Voltage less than 36V is a safety voltage. When measuring voltage higher than DC 36V, AC 25V, check the connection and insulation of test leads to avoid electric shock. 3Be sure to keep the test leads off the testing point when converting function and range. 4Select correct function and range to avoid fault operation. 5When measuring current, do not input current over 20A. 6Safety symbols
“ ” exists high voltage,“ ”GND,“ ” dual insulation,“ ” must refer to manual “ ” low battery
3. SPECIFICATION
1) GENERAL
1-1Displaying : LCD displaying 1-2Max. indication: 19993 1/2, auto polarity indication 1-3Measuring method: dual slope A/D transfer 1-4Sampling rate: approx. 3 times/sec 1-5Over range indication: MSD displays “OL 1-6Low battery indication: “ ” symbol displays 1-7Operation: 040relative humidity <80% 1-8Powerone 9V battery (NEDA1604/6F22 or equivalent 1-9Size: 190×88.5×27.5 mm 1-10Weight: approx.422g (including 9V battery)
2) TECHNICAL DATA 2-1Accuracy: ±a% × reading + dgts at (23 ± 5), relative humidity<75%. One
year guarantee since production date.
2-2. Function:
·DCV ·Resistance Ω ·Temperature
·ACV ·Diode continuity testing ·Frequency f
·DCA ·Triode hFE ·Inductance L
·ACA ·Capacitance C ·Auto power-off
·True RMS · Backlight
2-3. TECHNICAL DATA
2-3-1. DC Voltage
Input resistanceAll range is 10MΩ
Overload protectionRange 200mV: DC 250V or AC peak value
Other ranges: DC 1000V or AC peak value
2-3-2. AC Voltage
Input impedanceAll range is 10MΩ The frequency response of the standard sine wave and triangular wave is 40Hz ~1kHz. The frequency response of other waves is 40Hz ~200Hz
2-3-3. DCA
Max. measuring voltage drop: 200mV Max. input current: 20A (within 10 seconds) Overload protection: 0.2A / 250V fuse ;20A/250V fast-melt fused.
2-3-4. ACA
Range
Accuracy
Resolution
200μA
±(0.8%+10)
0.1μA
2mA
0.001mA
20mA
0.01mA
200mA
0.1A
20A
±(2.0%+5)
0.01A
Max. measuring voltage drop: 200mV Max. input current: 20A (within 10 second Overload protection0.2A / 250V fuse ;20A/250V fast-melt fused. Frequency responsesine and triangular wave: 40Hz~1kHz Other wave: 40Hz~200Hz Displaying: True RMS
2-3-5. ResistanceΩ
Range
Accuracy
Resolution
200Ω
± (0.8%+5)
0.1Ω
2kΩ
± (0.8%+3)
20kΩ
10Ω
200kΩ
100Ω
2MΩ
1kΩ
20MΩ
± (1.0%+25)
10kΩ
200MΩ
± (5.0% + 30)
100kΩ
Open voltage:less than 3V Overload protectionDC 250V or AC peak value NOTE: AAt range 200 Ω, short-circuit the test leads to measure the wire resistance, then, subtracts it from the real measurement. BThe reading be displayed slowly when the measurement is more than 1 .
Please wait it to be stable.
2-3-6. Capacitance
Range
Accuracy
Resolution
20nF
±(3.5%+20)
10pF
200nF
100pF
2uF
1nF
20uF
10nF
200uF
±(5.0%+10)
100nF
2000uF
1uF
Overload protectionDC 250V or AC peak value
2-3-7. Inductance (L)
Range
Accuracy
Resolution
2mH
±(2.5%+30)
1uH
20mH
10uH
200mH
100uH
2H
1mH
20H
10mH
Overload protectionDC 36V DO NOT INPUT VOLTAGE AT THIS RANGE
2-3-8. Temperature
Range
Accuracy
Resolution
(-20 ~ 1000)
±(1.0%+5) < 400
±(1.5%+15) ≥ 400
1
(0~1832)°F
±(0.75%+5) < 750°F
±(1.5%+15) ≥ 750°F
1°F
2-3-9. Frequency
Range
Accuracy
Resolution
10Hz
±(1.0%+3)
0.01Hz
100Hz
0.1Hz
1kHz
1Hz
10kHz
10Hz
100kHz
100Hz
1MHz/10MHz
1kHz/10kHz
Input sensitivity1V RMS Overload protectionDC 250V or AC peak value (within 10 seconds)
2-3-10. Diode and continuity testing
Range
Reading
Condition
Forward voltage drop of diode
Forward DCA is approx.
1mA, the backward
voltage is approx 3V
Buzzer makes a long sound while resistance is less than 70Ω±20Ω
Open voltage is approx. 3V
Overload protectionDC 250V or AC peak value
Range
Accuracy
Resolution
200mV
±(0.5%+3)
0.1mV
2V
0.001V
20V
0.01V
200V
0.1V
1000V
±(0.8%+10)
1V
Range
Accuracy
Resolution
200mV
±(0.8%+5)
0.1mV
2V
1mV
20V
10mV
200V
100mV
750V
±(1.2%+10)
1V
Range
Accuracy
Resolution
200μA
±(0.8%+10)
0.1μA
2mA
0.001mA
20mA
0.01mA
200mA
0.1A
20A
±(2.0%+5)
0.01A
CAUTIONDO NOT INPUT VOLTAGE AT THIS RANGE! 2-3-11. Transistor hFE DATA TEST
Range
Displaying range
Test condition
hFE NPN or PNP
0 ~ 1000β
Basic current is approx.
10uA Vce is about 3V
4. OPERATION 4-1Panel description
1. LCD: display the measuring value and unit.
2. Press REL/MAX/MIN, then you can test the relative magnitude. If you hold the REL/MAX/MIN for more than 2s, you can test the Max and Min value.
3. Transistor input jack
4. Backlight and function key; Diode/buzzer and frequency range, when press HOLD/B/L, you can do the function change; When under other range, this is the hold key; press HOLD/B/L key for more than 2s, the backlight will on/off.
5. Power on/off
6. Switch: change the function or the range.
7. 200mA current and inductance input jack
8. 20A/2A current test jack
9. Inductance, temperature, -common input jack
10. Voltage, resistance, diode, capacitance, frequency, temperature, + input jack. SEE THE FIG.
4-2Voltage measurement
1. Apply the black test lead to “COM” terminal and the red one to “V/Ω/Hz” terminal.
2. Setup the Knob on a prope r range “V”. If the voltage range is unsure beforehand, please set it on the max., then measure diminishingly to reach a resolute value.
3. Press the DC/AC key down to AC mode to measure AC voltage. Oppositely, resile the key to DC mode to measure DC voltage.
4. Connecting the test leads reliably with the tested circuit, the voltage value will be displayed on LCD. While testing a DC voltage, the reading is voltage and polarity of the point connected by the red lead.
NOTE:
1. While the reading is OL,the voltage is beyond the present range. Now you need to set the knob to the higher.
2. Do not input a voltage over DC 1000V or AC 750V. Please keep the test leads off the circuit while switching the function or range, or after finished your testing.
3. Be carefully while measuring a high voltage. DO NOT touch the circuit.
4-3Current measurement
1. Apply the black test lead to “COM” terminal and the red one to “mA” (Max 200mA) or “20A”(Max 20A) terminal.
2. Set the knob to a proper range A. If the current under tested is unsure beforehand, please set it on the max., then measure diminishingly to reach a resolute value.
3. Press the DC/AC key down to AC mode to measure AC current. Oppositely, resiling the key to DC mode to measure DC current.
4. Connecting the leads with the tested circuit in series, the current value is displayed on LCD. While testing a DC current, the reading is the value of current and polarity of the point connected by the red lead.
NOTE:
1.If the LCD displays OL,it means the current is over range. Now you need to set
the knob to the higher. 2.Max. input current is 200mA or 20A(subject to where the red test lead apply to), too large current will blow the fuse. Be sure the test is less than 10 seconds when test 20A current. Please keep the leads off the circuit while switching the function and range knob.
3. Do not input more than DC 36V, AC 25V between the current jack and COM jack.
4-4RESISTANCE MEASUREMENT
1. Apply the black lest lead to COM terminal and the red one to “V/Ω/Hz” terminal.
2. Set the knob to a proper resistance range, and connect the leads crossly with the resistor under tested.
NOTE:
1. The LCD displays OL while the resistance is over the selected range. The knob should be adjusted to a higher range.
2. When test low resistance, test leads will bring internal resistance, in order to get a accurate readings, you can mark the short circuit value of the test leads, then subtract it from the value of the low resistance.
3. When measuring in line resistor, be sure that the power is off and all capacitors
are released completely.
4. Do not input any volt at resistance range.
5. W hen measuring value is over 1MΩ, the reading will take a few seconds to be
stable. It’s normal for high resistance measuring.
4-5CAPACITANCE MEASUREMENT
Apply the knob to proper capacitance range, and insert the capacitor under tested into “V/Ω/Hz” and COM terminal, be wary of polarity if necessary.
NOTE:
1. If the capacitance under tested is over the max. value of selected range, LCD
displays OL, thus, should set the knob to a higher range.
2. It’s normal that there is a remained value on LCD before capacitance
measurement when at the range 20nF and you can subtract it from the value.
3. When measuring at large capacitance range, if capacitor is crept badly or broken,
LCD displays a value and it’s unstable.
4. Release the capacitor completely before measuring.
4-6INDUCTANCE MEASUREMENT
Set the knob to a proper inductance range and insert the inductor to mAand COM terminal.
NOTE:
1. The LCD displays OL while the tested inductance is over the selected range.
Thus, the knob should be set to a higher range.
2. Range mH is auto change of 2mH/20mH/200mH, and range H is the auto range of 2H/20H.
3. Do not input voltage at this range.
4-7. FREQUENCY MEASUREMENT
1. Apply test leads or shield cable to COM and “V/Ω/Hz” terminal.
2. Set the knob to frequency range, connect test leads or cable crossly to signal
source or tested load.
NOTE:
1. When input over 10V RMS, reading is workable but accuracy is not guaranteed.
2. It is better to use shield cable to measure small signal at noisy environment.
3. Be careful when measuring high volt circuit.
4. Do not input voltage over DC 250V or AC peak value.
4-8. TRANSISTOR hFE
1. Set the knob to hFE range.
2. Verify the transistor under tested is NPN or PNP, insert emitter, base and
collector to proper jack.
4-9. DIODE AND CONTINUITY TEST
1. Apply the bl ack test lead to “COM” termi nal and the red one to “V/Ω/Hz” terminal (the polarity of red lead is “+”.
2. Set the knob to range, connect test leads to the diode under tested, the red test connect to diode positive polarity, the reading is the approx. value of diode forward volt drop.
3. Apply test leads to two points of tested circuit, if the inner buzzer sounds, the resistance is less than (50 ± 20) Ω. NOTE: Do not input voltage at this function.
4-10DATA HOLD/ BACKLIGHT
Except Diode, Buzzer and Frequency function, at all other functions, if you press HOLD B/L, it will show HOLD symbol on the display, and the present value will be keep on it too, and if you press the key again, the symbol will disappear. If you press the key for more than 2s, it will be the backlight on/off.
4-11AUTO POWER OFF
After stop working for 15 minutes, the meter will be into sleep mode. Press “POWER” key can back to work. Press “REL/MAX/MIN” key and at the same time press the “POWER” the “APO symbol will disappear, now you already closed the function of auto power off.
5MAINTENANCE Do not try to modify the electric circuit.
5-1. Keep the meter away from water, dust and shock. 5-2. Do not store and operate the meter under the condition of high temperature, high humidity, combustible, explosive and strong magnetic place. 5-3. Wipe the case with a damp cloth and detergent, do not use abrasives and alcohol. 5-4. If do not operate for a long time, should take out the battery to avoid leakage When signal displays, should replace the battery following the steps 5-4-1 Unlock the button and remove the battery case. 5-4-2 Take out the old battery and replace the new one. Its better to use alkaline
battery for longer life. 5-4-3 Fit on the battery case, lock the button and put on the holster. 5-5 Replace fuse When replacing fuce, please use another same type and
specification fuse.
6. If the meter does not work properly, check the meter as following:
Fault
Solution
No reading on LCD
Power off--- Pls turn on the power
Holding key---PLs set a correct mode
Replace battery
The signal appears
Replace battery
No current or temperature input
Replace fuse
Error Value
Replace battery
601E-9808-002E