VC837 DIGTAL MULTIMETER Operation Manual
1. General Description
This is a 3 1/2 digital multimeter with high stablility and performance. It uses a LCD with
20mm high figure, which makes the reading clearer and the operation more convenient.
It can test DCV, ACV, DCA, ACA, resistance, capacitance, frequency, NC
temperature, diode, and continuity. This meter also designed with functions including
unit symbol display, data hold, lighting, auto range, auto power off and warning
functions. To assure high accuracy and resolution, it adopts integrated circuit 8‐bit
microprocessor and a dual integral A/D conversion as LCD driver, giving high resolution
and high accurac
. It is an ideal tool for labs, factories and radio‐technology.
2. Safety Instructions
The instrument is designed in compliance with IEC1010 standard (safety standard issued
by International Electro technical Committee). Please read the followingsafety
instructions before operation.
2.1 Check the connection and insulation of test leads to avoid electric shock.
2.2 To avoid electric shock and damage to the meter, do not input voltage exceeding
rated value.
2.3 When measuring voltage higher than DC 60V or AC 40V, please be careful andavoid
electric shock.
2.4 Select correct function and range to avoid wrong operation.
2.5 Move the test leads away from test points when switching to other function.
2.6 Don’t input voltage in current terminal.
2.7 Don’t make any modification to the circuit. It may damage the meter or jeopardize
safety.
2.8 Safety symbols:
”Highvoltage,“ ”GND,“ ” Dual insulation, “ ” Refer to manual, “ ”Low
“
battery indication.
,dutycycle,
3.1.2 Max display: 6000 (3 5/6 digits, automatic polarity, and unit symbol display);
3.1.3 Measurement method: Analog to digital converter (in micro processorADC+MCU);
3.1.4 Sampling rate: approx.3 times/sec.
3.1.5 Over‐range display: “OL” displayed.
3.1.6 Low battery indicator: “
”
3.1.7 Working environment: (0~40)℃, relative humidity: <80%;
3.1.8 Store condition: (‐10~50)℃, relative humidity: <80%
3.1.9 Battery: 2 pieces 1.5V battery (“AAA” 7# battery);
3.1.10 Dimension: 140×72×37mm (length x width x height);
3.1.11 Weight: approx. 195g (including battery);
3.1.12 Accessories: test leads, TP01 thermocouple, user manual, holster,giftbox,and
2*1.5V batteries.
3.2 Technical Features
3.2.1 Accuracy: ± (a% × reading + digits). To assure accuracy, the ambient temperature
should be (23±5) ℃, relative humidity <75%. One year accuracy is guaranteed since
production date.
3.2.2 DC Voltage(DCV)
Range Accuracy Resolution
600mV
6V 1mV
60V 10mV
600V ±(1.0%+4) 100mV
±(0.5%+4)
0.1mV
Input impedance: at 200mv range >40MΩ, at other ranges is 10MΩ.
Overload protection: 600V DC/AC rms.
3. Features
3.1 General Characteristics
3.1.1 Display: LCD;
3.2.3. AC Voltage(ACV
Range Accuracy Resolution
)
![](/html/1a/1ae2/1ae275093a25fccdfeaba52d54f2f82498a8a3fbc2138422a29d79da89435c9d/bg2.png)
6V
±(0.8%+6)
60V 10mV
1mV
60mA 10μA
600mA 100μA
600V ±(1.0%+6) 100mV
Input impedance: 10MΩ.
Overload protection: 600V DC/ AC rms.
Frequency response: at 600V range: (40~1000)Hz, at other ranges: (40~2000)Hz
Displaying: True RMS response (calibration based on sine wave RMS)
3.2.4 DC Current(DCA
Range Accuracy Resolution
600uA
6000uA 1μA
60mA 10μA
600mA 100μA
6A 1mA
10A ±(2.0%+5) 10mA
)
0.1μA
±(1.0%+5)
Maximum voltage drop: 200 mV for full range.
Maximum input current: 10A (within 10 seconds).
Overload protection: 0.5A/250V fuse and 10A/250V fuse.
6A 1mA
10A ±(2.0%+10) 10mA
Maximum voltage drop: 200 mV for full range.
Maximum input current: 10A (within 10 seconds).
Overload protection: 0.5A/250V fuse and 10A/250V fuse.
Frequency response: 10A range: 40~100Hz, other ranges: 40~400Hz.
3.2.6 Resistance(Ω)
Range Accuracy Resolution
600Ω ±(0.8%+5) 0.1Ω
6kΩ
60kΩ 10Ω
±(0.8%+1)
600kΩ 100Ω
6MΩ 1kΩ
60MΩ ±(1.2%+5) 10kΩ
1Ω
Open circuit voltage: 500mV
Overload protection: 250V DC/AC peak value.
3.2.5 AC Current (ACA)
Range Accuracy Resolution
600uA
±(1.5%+5)
6000uA 1μA
0.1μA
3.2.7 Capacitance (C)
Range Accuracy Resolution
60nF ±(2.5%+6) 10pF
600nF ±(2.5%+5) 100pF
![](/html/1a/1ae2/1ae275093a25fccdfeaba52d54f2f82498a8a3fbc2138422a29d79da89435c9d/bg3.png)
6uF 1nF
60uF 10nF
600uF
6mF 1uF
±(5.0%+8)
60mF 10uF
Overload protection: 250V DC/AC peak value.
3.2.8 Frequency
Range Accuracy Resolution
100nF
‐40℃~1000℃
0F~1832℉
<400℃ ±(1.0%+5)
≥400℃ ±(1.5%+15)
<750℉ ±(1.0%+5)
≥750℉ ±(1.5%+15)
Thermocouple: K type.
WARNING: do not input voltage at this range!
Operation
4.1 Panel Description
①
. LCD: Display the data and unit symbol
1℃
℉
10Hz
0.01Hz
100Hz 0.1Hz
1000Hz 1Hz
10kHz 10Hz
±(0.5%+4)
100kHz 100Hz
1MHz 1kHz
10MHz 10kHz
Input sensitivity: 0.7V rms.
Overload protection: 250V DC/AC peak value.
3.2.9 Diode and Continuity Test
Measurement Range Test conditions
Forward DC current is
Diode forward voltage drop.
approx 0.8mA,
reverse voltage is
approx 2.2V.
When the resistance under test
is less than 50Ω, buzzer sounds
continuously.
Open circuit voltage:
2.2V
Overload protection: 250V DC/AC peak value.
WARNING: Do not input voltage at this range!
3.2.10 Temperature (℃/
Range Accuracy Resolution
℉
)
Number Feature Description
1△
2
3
4DC
5AC
6AUTO
7
8
9
10 MAX MIN
11
12
13
14
-
APO
FE
h
%,℃,℉
MΩ,kΩ,Ω
Hz,kHz,MHz
mV, V
uA, mA, A
Relative (REL) mode is active
Data Hold is active.
Indicates negative readings.
DC voltage or current measurement.
AC voltage or current measurement.
Auto range mode.
Null
Diode test mode.
The continuity beeper is on.
Null
Auto power off symbol.
The symbol of high voltage risk indication.
Low battery indication. Warning: To avoid error readings, which could leadto
possible electric shock or personal injury, please replace the battery in time.
Null;
Null;
Duty cycle, Degrees Celsius, Degrees Fahrenheit;
Megohm, Kilohm, Ohm;
Hertz, Kilohertz, Megahertz;
Millivolts, Volts;
Microamp, Milliamp, Amperes (A).
. Rotary switch: it is used to change the range and choose functions.