VICI D-4-I Quick Start

Analytical
Components
Systems
&
Pulsed Discharge Detector Model D-4-I Quick Reference Guide
General Precautions:
Never run the detector without discharge gas. To insure a continuous supply of discharge gas, always
change the gas bottle when it reaches 500 psi.
Never bake out a column while it is connected to the detector.
Detector lifetime is shortened in the Ar/Kr PID mode. To maximize detector lifetime, turn off the discharge
power when the GC is not actually analyzing samples.
Gas connections
Detector performance is adversely affected by the presence of impurities in the gas stream. To minimize the chance of this occurring:
1. Always use the helium purifier.
2. Never use copper or plastic tubes, even before the helium pur ifier.
3. Never use flow controllers containing polymers or lubricant.
4. Never use pipe fittings or teflon tape downstream of the purifier.
You will need:
Helium
99.9999% purity (“six nines”)
Air Liquide part number:
Gas regulator
Ultrahigh purity, with stainless steel diaphragm
Valco part number:
Valco Helium Purifier
(included with the detector system)
Valco part number:
Stainless steel tubing and
Fittings with gold ferrules
(consult the Valco catalog)
Restrictor
(included with the detector system)
Valco part number:
ValGas 4
TGA-422-580
HP2
TGA-R-30F60P
VENT
COLUMN
INLET
Leave column inlet plugged until flow rate is set. (See “Initial Power-Up” on next page.)
DISCHARGE GAS
COLUMN
GAS CHROMATOGRAPH
(99.9999% He)
TGA-R-30F60P
RESTRICTOR
(30 mL/min minimum)
DISCHARGE GAS INLET
SAMPLE
IN
HELIUM
PURIFIER
VENT
TEE (ZT1)
Electrical Connections
HIGH
VOLT AGE
PULSER
MODULE
(no connection)
HEATER CABLE FROM DETECTOR
BIAS-B ELECTROMETER BIAS-A
CONTROLLER
(Rear panel)
RECORDER
ATTENUATED
OUTPUT
SHIELD
+ –
RED BLACK GREEN WHITE
Initial Power-Up
Caution:
Always make sure that discharge gas is flowing before heating and powering up the detector.
1. Before installing the column, set the gas flow to 30 ml/min (measured at the detector vent). Let it flow for 15 minutes so that all air is purged from the helium purifier.
2. Plug in the helium purifier.
3. Install the column as illustrated at right.
4. Turn on the MAINS switch on the back of the controller. Set the MODE switch on the front panel of the controller to PDHID.
5. Set the detector temperature and allow time for the detector and helium purifier to reach the set temperature.
6. Turn on the DISCHARGE switch on the front panel of the controller.
TO
11.4 cm
TO COMPUTER OR INTEGRATOR
RED BLACK WHITE GREEN BROWN BLUE ORANGE YELLOW
COLUMN INLET
CAPILLARY COLUMN ADAPTER (IZERA1.5)
COLUMN FERRULE
NUT CAPILLARY
COLUMN
0 - 10 V Common 0 - 1 V Shield COM SET
UNATTENUATED OUTPUT
AUTO ZERO
9.8 cm
(set at the
factory)
COLUMN INLET
PACKED COLUMN ADAPTER (I-23642-D4RU)
The discharge should start within five minutes. In a clean system, the discharge will have a peach/pink color. A purple discharge indicates leakage or impurities in the system.
7. Check the standing/background current on the controller display. Optimum current is 1.0 - 2.5 nA.
PACKED
COLUMN
Troubleshooting
High Background Current
Disconnect the column and cap the column inlet
Pink
Check discharge color
Low Sensitivity
Check background current with column removed and column inlet plugged
Purple/
blue
Do a hydrogen leak test (described on last page)
Baseline current falls to normal range (£ 2.5 nA)
Baseline current is still high (> 2.5 nA)
Leaks in the fittings of the PDD plumbing
Leaks inside the detector cell itself
No detectable leaks.
The detector isn’t leaking. Check for column bleed and/or leaks in the GC setup.
Tighten leaking fittings
Consult
Valco
Possible bad carrier gas or empty He bottle. Install a new bottle.
Baseline current falls to normal range (£ 2.5 nA)
Baseline current is still high (> 2.5 nA)
Baseline current falls to normal range (£ 2.5 nA)
Baseline current is still high (> 2.5 nA)
Baseline current falls to normal range (£ 2.5 nA)
Baseline current is still high (> 2.5 nA)
Consult
Valco
Consult
Valco
Consult
Valco
Sensitivity is in acceptable range
High background current (> 2.5 nA)
Refer to “Troubleshooting: High Background Current”
No Peaks
Normal background current (£ 2.5 nA)
Reinstall column and check background current
Background current increases
Background current remains in normal range
If the background current is stable but there are no peaks:
1. Check column flow.
2. Check the column insertion position (11.4 cm for capillary columns, 8.9 cm for packed columns).
3. If there are still no peaks, check to make sure that the GC is actually making an injection.
4. If there are still no peaks, consult Valco.
Check column inlet for leaks. Tighten fittings.
Check column position according to drawing on page 2, and run a sample.
Background current remains high
Background current drops to normal range
Sensitivity is in acceptable range
Sensitivity is still too low
Remove column and condition it, then run a sample
Run a sample
Remove column and bake detector at 300°C for four hours. Install column and run a sample.
Sensitivity is still too low
Sensitivity is in acceptable range
Sensitivity is still too low
Sensitivity is in acceptable range
Sensitivity is still too low
Consult
Valco
High Noise Level
If the noise level is high:
1. See if it improves with the GC fan turned off.
2. Check the column insertion position (11.4 cm for capillary columns, 8.9 cm for packed columns).
3. If the noise level is still high, do a hydrogen leak test (described on the next page).
4. If there are no leaks found, or leaks are found and repaired and the noise is still high, consult Valco.
Hydrogen Leak Test
A hydrogen leak test allows you to pinpoint loose fittings and leaks inside the detector.
1. Connect a length of plastic tubing to a regulated hydrogen bottle. Establish a flow of 5 -10 ml/min.
2. Hold the hydrogen outlet tube at a fitting connec­tion for ten seconds while monitoring the baseline. (Flip the mode switch to PDECD and back to PDHID to un-zero the baseline.)
If the baseline stays the same, that connection is
leak-free. If the baseline goes up, you have located a leak.
3. Tighten the fitting and test it again, repeating as necessary until every connection has passed the ten second test.
4. If the current remains high, hold the hydrogen outlet close to the detector.
tube into any detector holes.)
detector is indicated, contact Valco.
(Do not insert the
If a leak in the
Valco Instruments Co. Inc. Valco International
tel: 800 367-8424 fax: 713 688-8106
valco@vici.com
tel: Int + 41 41 925-6200 fax: Int + 41 41 925-6201 vici@vici.com
Printed in USA © 2003
VICI® and Cheminert
are registered trademarks of
Valco Instruments Co. Inc. and VICI AG
7/03
®
Loading...