This page intentionally left blank for printing purposes
Introduction
Description and Operating Principle
1
The D3 is a non-radioactive pulsed discharge ionization detector (PDID), with
models available for “plug and play” installation on the Agilent 6890 or 7890
GC. A schematic representation of the detector is shown in Figure 1.
The D3 utilizes a stable, lo w power , pulsed DC discharge in helium as the
ionization source. Elutants from the column, flowing counter to the flow of
helium from the discharge zone, are ionized by photons from the helium
discharge above. Resulting electrons are focused toward the collector
electrode by the two bias electrodes.
The principal mode of ionization is photoionization by radiation arising from
the transition of diatomic helium
He
1
+
(
∑
)
A
u
2
to the dissociative 2He(1S
1
)
ground state. This is the well-known Hopfield emission. The photon energy
from the He2 continuum is in the range of 13.5 eV to 17.7 eV.
The D3 is essentially non-destructive (0.01 - 0.1% ionization) and highly
sensitive. The response to organic compounds is linear ov er five orders
of magnitude with minimum detectable quantities (MDQs) in the low
picogram range. The response to fix ed gases is positiv e (the standing
current increases), with MDQs in the low ppb range.
Detector response is universal except f or neon, which has an ionization
potential of 21.56 eV. Since this potential is close to the energy of the
He* metastable (19.8 eV) but greater than the photon energy from the
He2 continuum, neon exhibits a low ionization efficiency and low detector
response.
When a dopant is added to the discharge gas, the D3 also functions as a
selective photoionization detector . (Suitab le dopants include Ar for organic
compounds, Kr for unsaturated compounds, or Xe for polynuclear aromatics.)
GROUND PIN
DISCHARGE REGION
HELIUM INLET
HIGH VOLTAGE
SAPPHIRE INSULATORS
CAPILLARY COLUMN
VENT
COLUMN INLET
Figure 1: Schematic of the D-3 detector
COLLECTOR ELECTRODE
Introduction
2
Safety Notes and Information
Symbols
ATTENTION
Refer to the manual.
Installation Category
This equipment has been designed for installation category (overvoltage
category) II, pollution degree 2. It has been approved for use only in heavy
industrial environments and may not be used in the residential, commercial,
or light-industrial environment.
Safety
This instrument left the factory in a safe condition. This instruction manual
contains important information and warnings which must be followed by the
user to insure safe operation and to retain the instrument in a safe condition.
Use only with an approved mains supply cord ha ving a rating of 2A, 250V, or
greater. Do not use this equipment in a manner not specified herein.
CAUTION: During normal operation, the detector
produces ultraviolet energy (UVA, UVB), some of
which may be emitted. Do not watch the arc without
eye protection.
Introduction
Components of the Detector System
Components of the detector system are listed in T able 1 belo w. Check the
contents of the packages to verify that everything is present. Contact the
factory if anything is missing or damaged. (NOTE: damaged shipments
must remain with the original packaging for freight company inspection.)
Description QuantityProduct number
Detector assembly (includes detector, heater, and electrometer):
for Agilent 68901PD-D3-I
for Agilent 78901PD-D3-I-7890
Pulser module1PD-M2
Pulser power supply1I-23569-1
Agilent FID board: for Agilent 68901I-G1531-60020
for Agilent 78901I-G3431-60020
Helium purifier1HP2
Miniature helium purifier1HPM
Fittings kit1PD-KIT-D3
Includes:1/32" polyimide ferrule5ZF.5V
1/32" polyimide plug ferrule (no through hole)1ZF.5VX
1/16" gold-plated ferrule5ZF1GP
0.25 - 0.44 mm polyimide column ferrule5FS.4
0.4 - 0.5 mm polyimide column ferrule5FS.5
1/16" union2ZU1
1/16" tee1ZT1
1/8" external to 1/16" internal reducer2EZR21
1/8" to 1/16" reducing union2EZRU21
1/32" external nut1EN.5KN
Fused silica adapter (installed)1IZERA1.5
Packed column adapter1I-23642-D3
Restrictor, 30 cc/min @ 60 psi He1TGA-R-30F60P
Instruction manual1MAN-PDD3
3
Table 1: Components of the D3 system
Introduction
4
System Requirements
Components Not Included with the Detector System
•Helium (99.999% purity) and other support gases
•Ultra high purity grade gas pressure regulator with stainless steel
diaphragm
•Any special adapters required for connection to the gas regulator
•SS tubing to go from gas supply to GC
•Flow measuring device
System Purity
Discharge/Carrier Gas Considerations
The performance of the detector is adversely affected by the presence
of any impurities in the gas streams (carrier, discharge, or dopant). We
recommend that a quality grade of helium 5.0 (99.999% pure or better)
be used at all times. Major gas suppliers off er research g rade helium
(99.9999% pure) which is particularly low in fixed gas impurities and should
give good results in a clean system, but even the highest quality carrier
gas may contain some water vapor and fixed gas impurities; hence a
helium purifier is included as part of the detector system. The discharge
gas should
always
be run through the helium purifier.
Whenever a ne w batch of discharge gas is receiv ed, we recommend
performing a blank GC analysis of the gas in the PDHID mode to detect
and identify the presence of any impurities. Gas purity requirements
are specified on the next page.
Tubing
Standards of cleanliness that are suitable for many GC applications may
be totally inadequate for the sensitive PDHID/PDPID work. All surfaces that
contact the gas stream must be glass or stainless steel. Do not use copper
tubing or brass fittings. All tubes m ust be thoroughly cleaned and bake d
before use.
Flow Controllers
The use of valves or flow controllers in which the gas stream is exposed
to any polymer-based packing or lubricating material is to be particularly
avoided.
Pressure Regulators
We recommend commercial “ultra-pure” grade regulators with stainless
steel diaphragms. Regulators with diaphragms made of neoprene or other
elastomers should never be used.
Loading...
+ 17 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.