VICI D-3-I-7890 User Manual

Valco Instruments Co. Inc.
Pulsed Discharge Detector Models D-3-I-HP and D-3-I-7890 Instruction Manual
North America, South America, and Australia/Oceania contact:
® ®
Valco Instruments Co. Inc.
tel: 800 367-8424 fax: 713 688-8106
valco@vici.com
D3.P65
Rev. 9/07
Printed in USA
Europe, Asia, and Africa contact:
VICI AG International
tel: + 41 41 925-6200 fax: + 41 41 925-6201 info@vici.ch
This page intentionally left blank for printing purposes
Table of Contents
Introduction
Description and Operating Principles ............................................................ 1
Safety Notes and Information ........................................................................ 2
Components of the Detector System............................................................. 3
System Requirements
Components not Included with the Detector System ..................................... 4
System Purity ................................................................................................ 4
Gas Specifications......................................................................................... 5
Installation
General Precautions...................................................................................... 6
Checking the Firmware V ersion
Mounting the Detector.................................................................................... 6
Installing and Configuring the FID Logic Board
Gas Connections ..........................................................................................1 1
Initial Pow er-Up .............................................................................................1 3
Capillary Column Connection ........................................................................1 4
Pac ked Column Connection ..........................................................................14
Testing for Leaks...........................................................................................15
Symbols Installation Category Safety
Agilent 6890 GC..................................................................................... 6
Agilent 7890 GC..................................................................................... 6
Agilent 6890 GC..................................................................................... 8
Agilent 7890 GC..................................................................................... 8
Installing and Purging the Gas Regulator...............................................1 1
Installing and Purging the Helium Purifier ..............................................12
Connecting the Discharge Gas to the Detector......................................1 2
Mode Selection and Setup
Helium Ionization Mode.................................................................................1 6
Selective Photoionization Mode ....................................................................1 6
T roubleshooting ...................................................................................................1 8
Warranty .............................................................................................................20
Detector Perf ormance Log ...................................................................................21
This page intentionally left blank for printing purposes
Introduction
Description and Operating Principle
1
The D3 is a non-radioactive pulsed discharge ionization detector (PDID), with models available for “plug and play” installation on the Agilent 6890 or 7890 GC. A schematic representation of the detector is shown in Figure 1.
The D3 utilizes a stable, lo w power , pulsed DC discharge in helium as the ionization source. Elutants from the column, flowing counter to the flow of helium from the discharge zone, are ionized by photons from the helium discharge above. Resulting electrons are focused toward the collector electrode by the two bias electrodes.
The principal mode of ionization is photoionization by radiation arising from the transition of diatomic helium
He
1
+
(
)
A
u
2
to the dissociative 2He(1S
1
) ground state. This is the well-known Hopfield emission. The photon energy from the He2 continuum is in the range of 13.5 eV to 17.7 eV.
The D3 is essentially non-destructive (0.01 - 0.1% ionization) and highly sensitive. The response to organic compounds is linear ov er five orders of magnitude with minimum detectable quantities (MDQs) in the low picogram range. The response to fix ed gases is positiv e (the standing current increases), with MDQs in the low ppb range.
Detector response is universal except f or neon, which has an ionization potential of 21.56 eV. Since this potential is close to the energy of the He* metastable (19.8 eV) but greater than the photon energy from the He2 continuum, neon exhibits a low ionization efficiency and low detector response.
When a dopant is added to the discharge gas, the D3 also functions as a selective photoionization detector . (Suitab le dopants include Ar for organic compounds, Kr for unsaturated compounds, or Xe for polynuclear aromatics.)
GROUND PIN
DISCHARGE REGION
HELIUM INLET
HIGH VOLTAGE
SAPPHIRE INSULATORS
CAPILLARY COLUMN
VENT
COLUMN INLET
Figure 1: Schematic of the D-3 detector
COLLECTOR ELECTRODE
Introduction
2
Safety Notes and Information
Symbols
ATTENTION
Refer to the manual.
Installation Category
This equipment has been designed for installation category (overvoltage category) II, pollution degree 2. It has been approved for use only in heavy industrial environments and may not be used in the residential, commercial, or light-industrial environment.
Safety
This instrument left the factory in a safe condition. This instruction manual contains important information and warnings which must be followed by the user to insure safe operation and to retain the instrument in a safe condition. Use only with an approved mains supply cord ha ving a rating of 2A, 250V, or greater. Do not use this equipment in a manner not specified herein.
CAUTION: During normal operation, the detector produces ultraviolet energy (UVA, UVB), some of which may be emitted. Do not watch the arc without eye protection.
Introduction
Components of the Detector System
Components of the detector system are listed in T able 1 belo w. Check the contents of the packages to verify that everything is present. Contact the factory if anything is missing or damaged. (NOTE: damaged shipments must remain with the original packaging for freight company inspection.)
Description Quantity Product number
Detector assembly (includes detector, heater, and electrometer):
for Agilent 6890 1 PD-D3-I
for Agilent 7890 1 PD-D3-I-7890
Pulser module 1 PD-M2
Pulser power supply 1 I-23569-1
Agilent FID board: for Agilent 6890 1 I-G1531-60020
for Agilent 7890 1 I-G3431-60020
Helium purifier 1 HP2
Miniature helium purifier 1 HPM
Fittings kit 1 PD-KIT-D3
Includes: 1/32" polyimide ferrule 5 ZF.5V
1/32" polyimide plug ferrule (no through hole) 1 ZF.5VX
1/16" gold-plated ferrule 5 ZF1GP
0.25 - 0.44 mm polyimide column ferrule 5 FS.4
0.4 - 0.5 mm polyimide column ferrule 5 FS.5
1/16" union 2 ZU1
1/16" tee 1 ZT1
1/8" external to 1/16" internal reducer 2 EZR21
1/8" to 1/16" reducing union 2 EZRU21
1/32" external nut 1 EN.5KN
Fused silica adapter (installed) 1 IZERA1.5
Packed column adapter 1 I-23642-D3
Restrictor, 30 cc/min @ 60 psi He 1 TGA-R-30F60P
Instruction manual 1 MAN-PDD3
3
Table 1: Components of the D3 system
Introduction
4
System Requirements
Components Not Included with the Detector System
Helium (99.999% purity) and other support gases
Ultra high purity grade gas pressure regulator with stainless steel
diaphragm
Any special adapters required for connection to the gas regulator
SS tubing to go from gas supply to GC
Flow measuring device
System Purity
Discharge/Carrier Gas Considerations
The performance of the detector is adversely affected by the presence of any impurities in the gas streams (carrier, discharge, or dopant). We recommend that a quality grade of helium 5.0 (99.999% pure or better) be used at all times. Major gas suppliers off er research g rade helium (99.9999% pure) which is particularly low in fixed gas impurities and should give good results in a clean system, but even the highest quality carrier gas may contain some water vapor and fixed gas impurities; hence a helium purifier is included as part of the detector system. The discharge gas should
always
be run through the helium purifier.
Whenever a ne w batch of discharge gas is receiv ed, we recommend performing a blank GC analysis of the gas in the PDHID mode to detect and identify the presence of any impurities. Gas purity requirements are specified on the next page.
Tubing
Standards of cleanliness that are suitable for many GC applications may be totally inadequate for the sensitive PDHID/PDPID work. All surfaces that contact the gas stream must be glass or stainless steel. Do not use copper tubing or brass fittings. All tubes m ust be thoroughly cleaned and bake d before use.
Flow Controllers
The use of valves or flow controllers in which the gas stream is exposed to any polymer-based packing or lubricating material is to be particularly avoided.
Pressure Regulators
We recommend commercial “ultra-pure” grade regulators with stainless steel diaphragms. Regulators with diaphragms made of neoprene or other elastomers should never be used.
Loading...
+ 17 hidden pages