Vichy VC99 3 6/7 DMM User Manual

User’s manual for
VC99 3 6/7 DMM
1.SUMMARY
This digital multimeter is a steady performance, battery-driven 3 6/7 digital multimeter. It uses the LCD with 23mm-high figure to make the reading clear and make operation more convenient. The digital multimeter has the function of measuring DCV, ACV, DCA, ACA, resistance, capacitance, frequency, temperature, duty cycle, transistor and diode, continuity performance test .The meter can provide functions including analog bar and unit symbol display, data holding, relative value measuring, maximal/minimal value measuring, auto/manual range switching (RANGE), auto power off and warning functions .it adopts double integral A/D converter as its core. It is an ideal tool for labs, factories and radio-technology.
2.SAFTY NOTES
The instrument is designed according to IEC1010 standard (safety standard issued by International Electro technical Committee). Please read the following before operation. 1-1.Check the connection and insulation of test leads to avoid electric shock. 1-2.Do not input a limited voltage higher than DC 1000V or AC 750V when measuring ranges. 1-3.When measuring voltage higher than DC 60V, AC 40V, please be careful. 1-4. Select correct function and range to avoid fault operation.
1-5.Please move the test leads away from test points when switching the function. 1-6.Please don’t input voltage value when measuring current. 1-7.Please don’t modify the circuit. 1-8.Introduction for safety symbol:
exists high voltage; GND dual insulation “ ” must refer to manual “ ” Low battery.
3. SPECIFICATION
1. General features
1-1. Displaying: LCD; 1-2.Max display: 6000(3 6/7) digits automatic polarity, unit symbol and 61 section analog display; 1-3. Measurement method: double integral A/D conversion; 1-4. Sampling rate: approx.3 times/sec. 1-5. Over-range display: “OL” displayed in the highest digit. 1-6. Low battery display“ ” 1-7. Working environment: (040), relative humidity: <80%; 1-8.Store environment: (-1040), relative humidity: <80%; 1-9.Power supply2pcs 1.5V battery“AAA” battery); 1-10. Dimension: 185×93×35mm (length*width*height); 1-11. Weight: approx..290g(including battery) 1-12.Accessories: test leads, user manual, temperature probe, holster, gift box, and 2*1.5V battery.
2. TECHNICAL FEATURES
2-1. Accuracy: ± (a% × reading data + digits), environment temperature at (23±5), relative humidity
<75%, One year guarantee since production date.
2-2.DC VoltageDCV
Range
Accuracy
Resolution
600mV
±(0.5%+3)
0.1mV
6V
1mV
60V
10mV
600V
100mV
1000V
±(0.8%+3)
1V
Input impedance: approx. 10MΩ Overload protection1000V DC or 750V AC.
2-3.AC VoltageACV
Range
Accuracy
Resolution
6V
±(0.8%+5)
1mV
60V
10mV
600V
100mV
750V
±(1.0%+5)
1V
Input impedance: approx. 10MΩ Overload protection1000V DC or 750V AC. Frequency response:40200Hz under 750V ,other range:(40400Hz. Sine wave RMS(Average value response).
2-4.DC CurrentDCA
Range
Accuracy
Resolution
600uA
±(1.0%+3)
0.1μA
6000uA
1μA
60mA
10μA
600mA
100μA
6A
±(2.0%+5)
1mA
20A
10mA
MAX measurement voltage drop: full range mA :300mVA:100mV Max input current: 20A (within 15 seconds) Over load protection: 500mA/ 250V and 13A/250V fast molten fuse.
2-5.AC CurrentACA
Range
Accuracy
Resolution
600uA
±(1.2%+5)
0.1μA
6000uA
1μA
40mA
10μA
600mA
100μA
6A
±(2.0%+10)
1mA
20A
10mA
MAX measurement voltage dropfull range mA :400mV A:200mV Max input current: 20A (within 15 seconds) Over load protection: 500mA/ 250V and 13A/250V fast molten fuse. Frequency response: 20A range:40100Hz,other range:40 400Hz
2-6.ResistanceΩ
Range
Accuracy
Resolution
600Ω
±(0.8%+5)
0.1Ω
6
±(0.8%+3)
60kΩ
10Ω
600
100Ω
6
1kΩ
60
±(1.2%+5)
10kΩ
Open circuit voltage:600mV Over load protection: 250V DC or AC peak value; NOTE: At range 600 Ω, short-circuit the test leads to
measure the wire resistance and then subtracts it from
the real measurement. Or press “”REL” to clear the wire
resistance and read the value directly.
2-7.CAPACITANCEC
RANGE
ACCURACY
RESOLUTION
40nF
±(3.5%+10)
10pF
400nF
±(2.5%+5)
100pF
4μF
1nF
40μF
10nF
400μF
±(5.0%+8)
100nF
2000μF
1uF
Overload protection :250V DC or AC peak value.
2-8.REQUENCYF
RANGE
ACCURACY
RESOLUTIO N
10Hz
±(0.5%+4)
0.01Hz
100Hz
0.1Hz
1000Hz
1Hz
10kHz
10Hz
100kHz
100Hz
1MHz
1kHz
60MHz
10kHz
Input sensitivity:0.7V Over load protection: 250V DC or AC peak value;
2-9.Transistor(hFE)
RANGE
VALUE
TESTING
CONDITION
NPN OR PNP
01000
Basic current
approx.15μA
Vce approx.4.5V
2-10. Diode and continuity performance test
ran
ge
VALUE
TESTING
CONDITION
Forward voltage drop of diode
Forward DCA is approx. 0.5mA, the backward voltage is approx 1.5V
Buzzer makes a long sound while resistance is less than (50±10)Ω
Open circuit voltage is approx. 0.5V
Over load protection: 250V DC or AC peak value; CAUTION DO NOT INPUT VOLTAGE AT THIS RANGE!
2-10. TEMPERATURE
RANGE
VALUE
RESOLUT
ION
-40-1000
400
±(0.8%+4)
400 ±(1.5%+15)
1
0F-1832
750
±(0.8%+5)
≥750
±(1.5%+15)
SensorTP01(K type thermocouple) CAUTION DO NOT INPUT VOLTAGE AT THIS RANGE!
4. OPERATION
4-1. Panel description
1. LCD: display the measuring value and unit.
2.Function key: 2-1.HOLD key:press it, the presently measured value is held on LCD and HOLD symbol displays. Press it again, HOLD symbol disappears, and the meter is exited the holding mode. 2-2. REL keypress itreading clearturn into relative value measurement states “REL”symbol displays, Press it again,“REL” symbol disappears, and the meter is exited the relative mode. 2-3. Hz/DUTY keyWhen measuring the AC Voltage (Current) press it it will switch Frequency/duty cycle/Voltage(Current) When measuring the Frequency ,it will switch frequency/duty cycle1~99%)。 2-4.”DC/AC”key: switch DC and AC work mode.. 2-5.RANGE keyselect auto range or manual range mode, Auto range is the original states, it will display ”AUTO” symbol ,press it change to manual range .Press it more than 2 second , it will return to auto range states 2-6.MAX/MIN key :press it, turn into MAX mode, it will hold the max value of measuring, press it again ,turn into MIN mode, it will hold the min value of measuring. No auto power off and analog bar display under this mode. Press it more than 2 second , it will exit MAX/MIN states.