Viasensor G100 Series, G100, G110, G150 Operating Manual

LANDTEC North America, Inc., DBA VIASENSOR
G100 Range
Operating Manual
G100 CO2 0-20%
G110 CO2 0-100%
G150 CO2 0-10,000ppm
VIASENSOR
850 South Via Lata Suite 112 Colton, CA 92324 USA Toll Free: 855 VIASENSOR Tel: +1 (909) 783-9472 Fax: +1 (909) 825-0591 Email: sales@viasensor.com Website: www.viasensor.com
OMG100N1.17 G100 Range
Page 2 LANDTEC North America, Inc., DBA VIASENSOR
This page is intentionally left blank
G100 Range OMG100N1.17
Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 3
Table of Contents
1.0 MANUAL GUIDELINES ....................................................................................... 5
1.1 Document History ............................................................................................. 5
1.2 Safety Related Information ................................................................................ 5
1.3 Notes .............................................................................................................. 5
2.0 INTRODUCTION .................................................................................................. 6
2.1 The G100 Analyzer Range.................................................................................. 6
2.2 Instrument Components - Standard Product ........................................................ 7
3.0 G100 RANGE OPTIONAL PRODUCTS AND ACCESSORIES ................................... 8
3.1 Optional Products ............................................................................................. 8
3.1.1 Analyzer Data Manager (Optional) ................................................................ 8
3.1.1.1 Event Log ............................................................................................. 8
3.1.2 Temperature Probe Reading (Optional) .......................................................... 8
3.1.3 Humidity Probe Reading (Optional) ............................................................... 9
3.1.4 Oxygen Reading (Optional) .......................................................................... 9
3.2 Instrument Accessory Products......................................................................... 10
4.0 INSTRUMENT FEATURES ................................................................................. 11
4.1 Physical Characteristics of the Instrument Panel ................................................. 11
4.2 Panel Key Functions ........................................................................................ 12
4.3 Instrument Connection Points .......................................................................... 13
5.0 GENERAL OPERATIONAL INSTRUCTIONS .......................................................... 14
5.1 Switching the Instrument On ........................................................................... 14
5.2 Switching the Instrument Off ........................................................................... 14
5.3 Instrument Main Read Screen .......................................................................... 14
5.4 Instrument Status Icons .................................................................................. 16
5.5 Entering Data ................................................................................................ 16
5.6 Changing Between Parameters ......................................................................... 17
5.7 Memory ......................................................................................................... 17
5.8 Storage ......................................................................................................... 17
5.9 Main Menu ..................................................................................................... 17
5.9.1 Information .............................................................................................. 18
5.9.2 Utilities .................................................................................................... 18
5.9.2.1 Time & Date ....................................................................................... 18
5.9.2.2 Contrast ............................................................................................. 19
5.9.2.3 Alarms ............................................................................................... 20
5.9.2.4 Settings ............................................................................................. 20
5.9.2.5 Flow Fail ............................................................................................ 21
5.9.2.6 Logging .............................................................................................. 23
5.9.2.7 Reset ................................................................................................. 24
5.9.3 Calibration ............................................................................................... 24
5.9.4 View Data ................................................................................................ 24
5.9.4.1 Clear Reading Memory ......................................................................... 25
5.9.5 Diagnostics .............................................................................................. 25
5.10 Warning and Error Codes .............................................................................. 25
5.11 Battery/Charging ......................................................................................... 26
6.0 TAKING READINGS ......................................................................................... 27
6.1 Preliminary Checks - Best Practice .................................................................... 27
6.2 Gas Measurement Process – Best Practice .......................................................... 28
6.3 Alternative Reading Methods ............................................................................ 29
6.3.1 Logged Reading ........................................................................................ 29
OMG100N1.17 G100 Range
Page 4 LANDTEC North America, Inc., DBA VIASENSOR
6.3.2 Peak Reading ........................................................................................... 29
6.3.3 Hold Reading ............................................................................................ 30
7.0 CALIBRATION ................................................................................................. 31
7.1 User Calibration ............................................................................................. 31
7.2 Calibration Gases ........................................................................................... 31
7.3 Calibration Set-up .......................................................................................... 31
7.4 Calibration Equipment ..................................................................................... 32
7.5 Calibration Method ......................................................................................... 32
7.5.1 Zero CO2 Channel: .................................................................................... 33
7.5.2 Span CO2 Channel: ................................................................................... 33
7.5.3 Zero O2 Channel: ...................................................................................... 34
7.5.4 Span O2 Channel: ..................................................................................... 34
7.5.5 Reset Factory Settings ............................................................................... 34
7.6 Last Field Calibration ...................................................................................... 35
7.7 Calibration Record .......................................................................................... 35
8.0 PROBLEM SOLVING ........................................................................................ 36
8.1 Warnings and Errors ....................................................................................... 36
8.2 Flow Fail Warning ........................................................................................... 36
8.3 Self-test Warning Messages ............................................................................. 37
8.4 User Calibration Trouble Shooting ..................................................................... 39
8.5 Cross-Gas Effects ........................................................................................... 41
8.6 Error Due to CO2 Solubility in Water .................................................................. 41
8.7 Hardware Reset .............................................................................................. 41
9.0 SERVICE ......................................................................................................... 43
10.0 WARRANTY POLICY ..................................................................................... 44
11.0 TECHNICAL SPECIFICATIONS ...................................................................... 45
11.1 Technical Specification - G100 ....................................................................... 45
11.2 Technical Specification - G110 ....................................................................... 47
11.3 Technical Specification - G150 ....................................................................... 49
12.0 EVENT LOG .................................................................................................. 51
13.0 SAMPLE CERTIFICATE OF CALIBRATION ...................................................... 52
14.0 EC DECLARATION OF CONFORMITY ............................................................. 53
15.0 GLOSSARY OF TERMS – G100 RANGE ........................................................... 54
G100 Range OMG100N1.17
Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 5
1.0 Manual Guidelines
1.1 Document History
Issued
By
Issue
Date
Change
Control ID
Issue
No.
Reason for
Change
LA
Nov 2009
OMG100UK1.1
4
New Instructions
LA
Nov 2009
OMG100N1.1
5
Update Instructions
LA
June 2010
OMG100N1.1
6
Update Technical Specifications
AV
May 2012
OMG100N1.17
7
Update Part Numbers and Product Images
1.2 Safety Related Information
Information in this manual that may affect the safety of users and others is preceded by the following symbol:
Warning
Failure to follow this information may result in physical injury which in some cases could be fatal.
1.3 Notes
Important/useful information and instructions are shown clearly throughout the manual in a note format.
For example:
Note: For further information, please contact Sales at VIASENSOR Toll Free
(855) VIASENSOR, + 1 (909) 783-9472 or email Sales@viasensor.com.
OMG100N1.17 G100 Range
Page 6 LANDTEC North America, Inc., DBA VIASENSOR
2.0 Introduction
This manual explains how to use the instrument model types listed below:
G100 CO
2
0-20%
G110 CO
2
0-100%
G150 CO
2
0-10,000ppm
Note: The instrument is a sensitive piece of scientific equipment, and should be treated as
such.
The G100 range has been developed to incorporate the latest technology and specification requirements, which provide the user with a fast, simple-to-use and accurate analyzer. Each
model has been specifically designed to meet specific application requirements. Apparatus intended for use in residential, commercial, pharmaceutical and light industrial environments.
2.1 The G100 Analyzer Range
The analyzer has the following features:
CO
2
0-20% - G100
CO
2
0-100% - G110
CO
2
0-10,000ppm – G150
Options for:
- O2 0-100%
- Dual temperature probes 0-122o F
- Data storage and download
- Humidity sensor 0-100%
Improved accuracy on CO
2
readings
Quick verification of CO
2
Time saving with dual temperature probes Large data storage and user friendly software and download Easy-to-read, large well lit display Built-in gas moisture removal
Applications:
IVF/Medical Research  Laboratories IAQ  Brewing Atmosphere Control  Incubators Gas mixing
G100 Range OMG100N1.17
Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 7
2.2 Instrument Components - Standard Product
G100 Analyzer
Reference:
A
Analyzer
B
Operating Manual on Digital Media
C
Battery Charger
D
Battery Charger Adaptors: U.S.A. Europe Australia
E
Sample Tube Kit: Sample Filter Sample Tube
OMG100N1.17 G100 Range
Page 8 LANDTEC North America, Inc., DBA VIASENSOR
3.0 G100 Range Optional Products and Accessories
3.1 Optional Products
The G100 analyzer range has a number of optional products for purchase which enhance the usability and enable further analysis of data and reading information.
Note: For further information, please contact Sales at VIASENSOR Toll Free
(855) VIASENSOR, + 1 (909) 783-9472 or email Sales@viasensor.com.
3.1.1 Analyzer Data Manager (Optional)
Analyzer Data Manager enables the user to maximise the operation of the incubator analyzer. Instrument readings and event log data may be downloaded to a PC for further analysis and exported to other applications such as MS Excel. It enables direct communication with the unit, features a simple download facility and is fully compatible with the latest Microsoft operating systems.
3.1.1.1 Event Log
The G100 instruments incorporate the facility to log significant events via the ‘Event Log’. This can be used as an aid to monitoring the use of the instrument. It can also be used as a diagnostic tool if there is a problem with the instrument.
The event log can only be viewed via the optional Analyzer Data Manager software. It cannot be viewed on the analyzer screen. Applicable events are stored in the event log automatically. No user intervention is required.
The event log can hold approximately 270 events. If the log becomes full then it begins to over-write the older events. This can be identified by the index field which starts from event number 1. The log is cleared when the instrument is reset.
Note: Please refer to section ‘12.0 Event Log’ of this operating manual for further
information.
3.1.2 Temperature Probe Reading (Optional)
The G100 range of instruments has the facility to read and display two temperature readings via optional temperature probes. When a temperature probe is fitted to one of the two temperature ports on the top of the instrument, the display will automatically show the current reading and a temperature probe icon will be displayed.
The display can also be changed to show a T1 – T2 calculation by pressing the appropriate soft-key. The current mode of operation can be identified by normal or inverse status of the soft-key, where inverse indicates that the option is active.
Note: The T
1
– T2 result is not stored as part of the reading.
The operator can also choose to display the reading in either Fahrenheit or Centigrade using the temperature option accessed from the ‘Settings’ menu.
G100 Range OMG100N1.17
Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 9
3.1.3 Humidity Probe Reading (Optional)
The instrument has the optional facility to use a humidity probe (specified at the time of manufacture). This allows the instrument to read and display humidity readings from an optional probe.
When a humidity probe is fitted, the display will change automatically to show the
current reading and a humidity probe icon will be displayed. The ‘Scroll’ keys on the
instrument panel are used to switch between the reading screens.
The humidity kit comprises of:
Humidity lead Humidity sensor
Connection of the Humidity Sensor:
Plug the humidity sensor onto the mating connector of the transmitter or connection cable. Make sure that the catches are aligned correctly. Tighten the knurled nut by hand.
Note: The humidity sensor can take 30 minutes to stabilize and special handling is
required for optimum performance and stability. Please refer to the instruction pamphlet included with the humidity sensor packaging for the Humidity Standards.
3.1.4 Oxygen Reading (Optional)
The instrument has the optional facility to use an internal oxygen cell (specified at the time of manufacture). This allows the instrument to read and display oxygen readings along with CO2.
Note: Oxygen sensor stability; as the sensor is a partial pressure sensor its output
will be affected by changes in relative humidity. Although the percentage of O2 in the air is relatively constant, the relative humidity in air is variable. A unit calibrated with dry air could cause the readout to read low by up to 0.5% by volume.
OMG100N1.17 G100 Range
Page 10 LANDTEC North America, Inc., DBA
VIASENSOR
3.2 Instrument Accessory Products
Optional accessory and replacement parts may be purchased for the G100 from VIASENSOR direct. Please refer to the website www.viasensor.com for further information.
Ref
Description
Part Number
A
USB Communication Cable
G-USB
B
Filters (Pack of 5)
G-PTFE
C
Sample Tube Kit with Filter
G-TKIT
D
Hard Carry Case
G-HCASE
E
Temperature Probe 100mm SS Tip
G-TP100
F
Temperature Probe 5mm Tip
G-TP5
G
Relative Humidity Probe 15mm Diameter
G-RH15
H
Relative Humidity Probe 4mm Diameter
G-HR4
J
Moisture Trap Filter (Pack of 2)
G-MT
K
150PSI-2.8PSI - Step Down Pressure Regulator
G-STEPPR
L
Battery Charger
G-BC
M
Soft Carry Case
G-SCASE
N
Soda Lime Filter Kit
G-SLKIT
O
G Series Download Software for Managing Data
G-SW
P
Calibration Kit - Regulator, tubing & filter. No Calibration Gas
G-CALKIT
G100 Range OMG100N1.17
Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 11
4.0 Instrument Features
4.1 Physical Characteristics of the Instrument Panel
Front View: Reference:
Back View: Reference:
A
Main Read Screen
B
Soft-Keys
C
On/Off Key
D
Pump Key
E
Key 4 – Scroll Left
F
Key 8 – Scroll Down
G
Menu Key
H
Enter Key
I
Key 2 – Scroll Up
J
Key 6 – Scroll Right
K
Moisture Removal Tube
L
Serial Number
M
Instrument Stand
OMG100N1.17 G100 Range
Page 12 LANDTEC North America, Inc., DBA
VIASENSOR
4.2 Panel Key Functions Front Panel:
Back Panel:
K
Moisture Removal Tube
Removes the moisture from the sample gas.
L Serial Number
Unique Identification for the instrument. Verification of the serial number will be required if Technical Support assistance is needed.
M
Instrument Stand
Instrument stand.
Note: Do NOT attempt to remove the cover off the back of the analyzer which houses
the moisture removal tube.
Do NOT cover the moisture removal tube with your hand when holding the
analyzer to take readings.
Key
Description
Function
A
Main Read Screen
Start and end screen when using the instrument.
B Soft Keys
The function of the three ‘soft-keys’ on the front of the instrument panel are determined by menu options taken. Functions vary from screen to screen.
C
On/Off Key
Press the ‘On/Off’ key briefly to switch the
instrument on and off. D
Pump Key
Press the ‘Pump’ key to start or stop the pump.
E Scroll Left Key
Also ‘Key 4’. Enables the operator to scroll left to
display more information.
F
Scroll Down Key
Also ‘Key 8’. Enables the operator to scroll down
to display more information.
G
Menu Key
Press the ‘Menu’ key to go to the ‘Main’ menu.
Enables the operator to pre-set values and
settings. Select options from the ‘Main’ menu to
also view data and readings stored or held.
H
Enter Key
The ‘Enter’ key accepts/confirms choices made by the operator to various functions and operations. Also, required to confirm numeric data entry.
I
Scroll Up Key
Also ‘Key 2’. Press scroll up to view further
information on the instrument read screen.
J
Scroll Right Key
Also ‘Key 6’. Press scroll right to view further
information on the instrument read screen.
G100 Range OMG100N1.17
Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 13
4.3 Instrument Connection Points
Top View: Side View:
N.B. Depending on the configuration purchased, certain connectors may not be present.
Note: Temperature connectors are fitted with snap rivets to prevent dust ingress,
remove before use. Grip back cap and pull upwards to release. Refit when not in use.
A
Top View:
Temperature 1
Temperature 1 connector measures differential temperature – incubator and ambient temperatures.
B Temperature 2
Temperature 2 connector measures differential temperature – incubator and ambient temperatures.
C
Humidity Probe (Optional)
Humidity probe attachment point.
D
Gas Outlet
Gas outlet port used to exhaust the gas.
E
Gas Inlet
Gas inlet point used to attach the sample tube and filter in order to take the gas reading.
F
Side View:
USB Cable Attachment Point
Used to connect the analyzer to a PC via a USB cable to download data.
G Battery Charger connection
Used to attach the battery charger to the analyzer for charging.
5V ± 0.5V(max 1000mA)
+-
5V ± 0.5V(max 1000mA)
+- +-
OMG100N1.17 G100 Range
Page 14 LANDTEC North America, Inc., DBA
VIASENSOR
5.0 General Operational Instructions
Note: Fully charge the unit before use when the instrument is first received or if the
instrument has been in storage for six months or more.
5.1 Switching the Instrument On
1) To switch on the instrument, press the ‘On/Off’ key briefly. There will be a short
beep and a slight pause followed by the VIASENSOR logo.
2) The power on self-test will then commence (approximately 15 seconds) including
warm-up time.
3) Assuming there are no warnings to display the instrument will continue to the ‘Main
Read Screen’.
5.2 Switching the Instrument Off
1) Purge with fresh air. Run the pump for approximately 30 seconds or until the
readings have returned to normal levels.
Note: Before the instrument is switched off a clean air purge should be
performed. This ensures that the instrument is free from gas and ready for the next measurement. This final purge is especially important for the oxygen sensor as it may degrade if stored when contaminated with gas.
2) To switch off the instrument, press the ‘On/Off’ key briefly.
Note: If the ‘Auto Off’ utilities setting is set to ‘Yes’, the analyzer will switch off
automatically after 10 minutes if not in use.
5.3 Instrument Main Read Screen
After the analyzer has been switched on and the warm-up self-tests completed the analyzer will display the ‘Main Read Screen’.
G150 Screen Shown:
Main Read Screen
Soft-Keys:
Peak - Enables the operator to display the peak reading. Store - Enables the operator to store the displayed reading for viewing/download later. Hold - Enables the operator to hold the current reading being taken.
Note: The pump is turned off when a reading is stored.
Battery Power Indicator
Pump (On) Indicator Symbol
Alarm Set
Temperature Probe Connected Indicator
G100 Range OMG100N1.17
Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 15
OMG100N1.17 G100 Range
Page 16 LANDTEC North America, Inc., DBA
VIASENSOR
5.4 Instrument Status Icons
The following icons may be displayed on the instrument read screens:
Icon
Description
Battery charge state
(flashing)
< 1 hour remaining
(flashing)
Battery charging
Charged
Pump running
(flashing)
Pump stalled (Backlight turns red)
Alarm set
(flashing)
Alarm active (Backlight turns red)
USB connected to PC (flickers when transferring data)
Logging mode active (flashes when memory nearly full)
Temperature probe(s) connected
Humidity probe connected
(flashing)
Service due (every 12 months)
Service overdue
Fault/repair
Waiting
Note: A red backlight is displayed if the pump is stalled or the alarm is activated. For
further information, please refer to ‘section 5.9.2.3 Alarms’ or ‘section 5.9.2.5 Flow Fail’.
5.5 Entering Data
During normal operation the user may be prompted to enter data or information via the keypad, i.e. entering an ID code or setting an alarm level.
When entering data into the instrument all fields are fixed format and are populated from the right. For example, to enter a new time 09:25:00 the user would type in 092500 using the numeric keypad in the following sequence:-
* : : 0 * : :09 * : 0:92 * :09:25
G100 Range OMG100N1.17
Copyright 2009 LANDTEC North America, Inc., DBA VIASENSOR Page 17
* 0:92:50 * 09:25:00
Press the ‘Enter’ key to confirm/accept data keyed.
Any mistakes can be corrected using the soft-key ‘Delete’ which will delete the last character typed. Alternatively, the sequence can be re-typed before the ‘Enter’ key is pressed and the existing numbers will be pushed off the screen.
Note: The instrument will not allow invalid data to be entered; this should be deleted
and re-entered.
5.6 Changing Between Parameters
By default, the instrument displays the ‘Main Read Screen’ (for gas measurement). This shows the CO2 reading along with the optional O2 reading. The instrument will return to
this screen after power on or when returning from the menus. The ‘Scroll’ keys can be
used to switch to another measurement screen, i.e. temperature or humidity. Continue to press the ‘Scroll’ key to return to the ‘Main Read Screen’.
5.7 Memory
The memory should not be used as a permanent storage medium and any important data should be transferred to a more permanent storage medium as soon as possible. The instrument should not be stored for prolonged periods with valuable data in its memory.
5.8 Storage
When not in use the instrument should be kept in a clean, dry and warm environment, such as an office. It should be stored flat with the stand folded away which helps prolong the life of the O2 cell.
Note: Fully charge the instrument before use if instrument has been stored for six
months or more.
5.9 Main Menu
The ‘Main Menu’ enables the operator to select options to set up specific parameters and perform operational tasks prior to sample readings being taken or to view data/information stored in the instrument.
1) Press the ‘Menu’ key on the front of the instrument panel and the following screen is
displayed:
Main Menu
2) Press the soft-key ‘Exit’ to exit the ‘Main’ menu.
OMG100N1.17 G100 Range
Page 18 LANDTEC North America, Inc., DBA
VIASENSOR
5.9.1 Information
The ‘Information’ option enables the operator to display information such as instrument
type, serial number, current software version, service due date and the dates of the last factory and user calibrations.
1) From the ‘Main Read Screen’ press the ‘Menu’ key on the instrument panel.
2) Press ‘Key 1’ to display general information about the instrument.
Information Screen
5.9.2 Utilities
The ‘Utilities’ option enables the operator to configure instrument settings prior to taking readings.
1) From the ‘Main Read Screen’ press the ‘Menu’ key on the instrument panel.
2) Press ‘Key 2’ to display the ‘Utilities’ menu and the following screen is displayed:
Utilities Menu
5.9.2.1 Time & Date
The ‘Time and Date’ option enables the operator to check or set the instrument’s
internal clock. The current time/date are appended to every stored reading.
1) From the ‘Main Read Screen’ press the ‘Menu’ key on the instrument panel.
2) Press ‘Key 2’ to display the ‘Utilities’ menu.
3) Press ‘Key 1 – Time & Date’ and the ‘Set Time & Date’ menu is displayed:
Loading...
+ 39 hidden pages