VET-PRO VIP 2000 Operator's Manual

Page 1
VET-PRO
VIP 2000
VETERINARY INFUSION PUMP
OPERATORS MANUAL
0473
Distributed in the U.K by: Burtons Medical Equipment Ltd
Units 1 - 4 Guardian Industrial Estate Pattenden Lane Marden Kent TN12 9QD UK
Email: [email protected] Tel (SALES) : 01622 834300 Fax: 01622 834330
URL:WWW.burtons.uk.com
EDITION MARCH 2002
Ref.: 100-090VX
Page 2
VET-PRO Operator’s Manual
2
LIMITED WARRANTY
The VET-PRO Infusion Pump has been carefully manufactured from the highest quality components.
The pump is guaranteed against defects in material and workmanship for twelve (12) months from date of purchase by the original purchaser.
The obligation under this Limited Warranty, shall be limited, at the manufacturer’s option, or that of its designated representative, to repairing or replacing pumps, which upon examination, are found to be defective in material or workmanship. The repair or replacement of any product under this Limited Warranty shall not extend the above-mentioned Warranty period.
All repairs under this Limited Warranty should be undertaken only by qualified, trained service personnel. In the event that a pump is found to be defective during the warranty period, the purchaser shall notify the manufactuer or its designated representative within thirty (30) days after such defect is discovered.
The defective pump should be sent immediately to the manufacturer or its designated representative for inspection, repair or replacement. Shipping costs are the purchaser’s responsibility.
Material returned to the manufactuer or its designated representative should be properly packaged. Inadequate packaging may result in severe pump damage.
This Limited Warranty shall not apply to defects or damage caused, wholly or in part, by negligence, spilt fluids, dropping of the pump, misuse, abuse, improper installation or alteration by anyone other than qualified, trained personnel; or to damage resulting from inadequate packaging in shipping the pump to the manufacturer or its designated representative.
If, after inspection, the manufactuer or its designated representative is unable to identify a problem, the manufactuer or its designated representative reserves the right to invoice purchaser for such inspection.
This Limited Warranty is the sole and entire warranty pertaining to the manufacturer’s products and is in lieu of and excludes all other warranties of any nature whatsoever, whether stated, or implied or arising by operation of law, trade, usage or course of dealing, including but not limited to, warranties of merchantability and warranties of fitness for a particular purpose. Purchaser expressly agrees that the remedies granted to it under this limited warranty are purchaser’s sole and exclusive remedies with respect to any claim of purchaser arising under this Limited Warranty.
Page 3
VET-PRO Operator’s Manual
3
Contents
Introduction …………………………………………………………………………
4
Glossary ……………………………………………………………………………..
5
Safety Precautions ……………………………………………………………………
6-7
Specifications ………………………………………………………………………..
8-9
Ordering Information 9
Trumpet and Flow Rate ………………………………………………………………
10-11
VET-PRO Front view ………………………………………………………………
12-13
VET-PRO with Open Door ………………………………………………………….
14
The Charger ………………………………………………………………………….
15
Optional Cage Mounting Bracket 16
Unpacking …………………………………………………………………………..
17
Loading the Administration Set …………………………………………………………….
17
Flush Function. …………………………………………………………………….
18
VET-PRO Operating ………………………………………………………………..
19
Pump Operation Test. ………………………………………………………….……
19
Air in Line Alarm Test. ……………………………………………………………..
20
Door Open Alarm Test. ……………………………………………………………..
21
Down Occlusion Alarm Test. ………………………………………………………
22
Charge Indicator Test. ………………………………………………………………
22
Programs …………………………………………………………………………….
23
Changing Current Program………………………………………………………………... 24-25
Down Occlusion Setting ………………………………………………………..……
26
CONTINUOUS Program……………………………………………………………..
27-30
25
STEPS PROTOCOL……………………………………………………………………..
30-31
TPN Program ………………………………………………………………………..
32-34
INTERMITTENT Program …………………………………………………………..
35-37
PCA Program – with protocols ………………………………..……………………..
37-44
PCA Program – Epidural Mode ………………………………..……………………..
45-46
PCA Program – without protocols ……………………………………………………
46
INFO Mode …………………………………………………………………………
47-48
Program Delay ………………………………………………………………………
48-49
Program Completed …………………………………………………………………
49-50
Locking ……………………………………………………………………………..
50-51
How to Lock …………………………………………………………………………
51-53
Battery Operation. ……………………………………………………………………
53-54
Cleaning and Storage ……………………………………………………………………….
54-55
Alarm Conditions……………………………………………………………………..
55
Troubleshooting
……………………………………………………………………..
55-58
Show Setting
………………………………………………………………………..
58-59
Adjust Setting ……………………………………………………………………….
59-60
Page 4
VET-PRO Operator’s Manual
4
Introduction
This Manual Provides directions for the setup and operation of the VET-PRO Pump
VET-PRO Features
S Small, light and compact pump. S Multi Programs:
CONTINUOUS
Continuous program. Infusion Rate: 0.1 to 1200 ml/hr. Volume: 0.1 to 9999 ml.
User can design a specific protocol, for drug delivery, which contains up to 25 steps. Infusion Rate: 0.1 to 1200 ml/hr. Volume: up to 9999ml., For each step. Accumulated Volume for all steps – up to 10 liters.
Rate Taper program, with programmable up
and down times
INTERMITTENT
A set Dose delivered in set intervals. Between Intervals the pump will allow a base rate or KVO.
PCA
Patient Control Analgesia, continuous delivery plus /or programmed boluses. Bolus Rate: 0.1-1000 ml/h
S
Highly accurate fluid delivery. The piston mechanism ensures an accurate volume delivery over more than 24 hours using the same administration set segment
S Operated with a customized safe administration set S Rechargeable internal battery S Can be latched into a Charger, which is mounted on an IV pole S A.B.S. Anti-Bolus System S Silent operation
25 STEPS
TPN
Page 5
VET-PRO Operator’s Manual
5
Glossary
Terms used in manual:
F
Note:
Essential operating procedure or condition.
G
Caution:
A precaution which, if not followed, could result in extensive damage to the equipment.
I
Warning
A precaution, which, if not followed, can result in personal injury or loss of life.
Page 6
VET-PRO Operator’s Manual
6
Safety Precautions
I
Warning
F
To avoid possible personal injury or loss of life, observe the following: Note: Read the entire Operation Manual before using the pump, since the
text includes important precautions.
•
Disconnect the Charger from the mains, prior to opening the casing. Voltage present on internal components may cause severe shock or death on contact. Only trained service personnel should open the pump cover.
•
Before servicing pumps connected to mains via the Charger - remove rings, watches, and other jewelry, which may present a shock / burn hazard.
•
To prevent a fire hazard, replace blown fuses on the Charger only with fuses of the same type and rating (see fuse values on the Charger PCB).
•
The equipment is not suitable for use in the presence of flammable gases – e.g. air/oxygen/nitrous oxide mixture.
•
Watch your fingers / nails when opening the pump door.
To avoid possible injury to patient, observe the following:
I
Warning
Battery charging is enabled as long as the charger cord is connected to the mains and the pump is in the charger.
Switching the pump off does not disconnect it from the mains. In order to disconnect from the mains, take the
charger cable out of the mains. In order to disconnect pump from the mains, take it out of the charger.
F Note:
To avoid battery damage do not leave the pump battery in a temperature more than 500C
•
Make sure the pump is attached securely to the Charger, which is connected snugly to an IV pole.
•
If the pump is dropped, return the pump for inspection by qualified service personnel.
•
Before operation, verify that administration set tubing is not kinked or occluded.
Page 7
VET-PRO Operator’s Manual
7
•
Although the pump will operate using a variety of giving sets, use of other administration sets may impair the operation of the pump and the accuracy of the infusion.
The maximum volume that may be infused under SINGLE FAULT
CONDITION
is 0.1 ml
G CAUTION
To avoid possible damage to the equipment, observe the following:
•
Do not immerse the pump in liquids.
•
Clean solution spills on the pump immediately. Use a damp cloth or sponge. A cleaning disinfectant may be used. Wipe thoroughly with a dry cloth.
•
Do not clean pump with chemicals such as Xylene, Acetone or similar solvents. These chemicals cause damage to plastic components and paint.
IWarning:
Any adjustments, maintenance, or repair of the uncovered pump while connected to the power should be avoided. If necessary, these repairs should be carried out by skilled technicians who are aware of the hazards involved.
IWarning:
The pump should be operated within an ambient temperature range of 15°C to 35°C. and at up to 85% humidity
IWarning
:
Do not operate the pump near high-energy radio-frequency emitting equipment, such as electro-surgical cauterizing equipment, and cellular telephones. False alarm signals may occur.
Page 8
VET-PRO Operator’s Manual
8
Specifications
Pumping Mechanism: Piston Mechanism Flow Rate: PCA: 0 to 100 ml/hr in 0.1ml increments
All other programs: 0.1 to 1200 ml/hr in 0.1ml or 1 ml increments.
Total Infused Volume: PCA: 0.1 to 1,000 ml.
Continuous, TPN, Intermittent: 0.1-9999 ml 25 Steps: 0.1 to 9999 ml. for each step Accumulated volume for all steps- up to 10 liters.
Total Time Setting 100 hours Accuracy: ± 3 %. KVO rate: 0.1 to 5 ml/hr. (to be set) Air Sensor: Ultrasonic, adjustable air bubble size Maximum Pressure: 0.7 bar or 10 psi Adjustable (high/normal/low) Power Supply: 110-240 VAC, 50/60 Hz. 20 W Battery: Li-Ion 7.2V, 1450mA. (Rechargeable)
Battery Operation at 125 ml/hr 20 hours
Battery Charging:
Automatic when latched into the Charger connected to an AC power source (4 hours)
Alarms: End Program
Low & End Battery Air in line Down Occlusion Door Open Pump Unattended Internal Malfunction Locking state Missing Key
Dimensions:
112 x 89 x 32mm. (L x w x h).
Symbol Attention (Consult accompanying documents)
Page 9
VET-PRO Operator’s Manual
9
Specifications
Classification
Type CF Equipment (degree of protection against electrical shock)
Housing: ABS (fire retardant)
Weight
280 gr. without battery 390 gr. With battery
Electrical Safety Complies with:
IEC 601-1 (Medical Electrical Equipment Safety), IEC 601-2-24 (Infusion pumps and controllers), IEC 601-1-4 (Programmable Electrical Medical
System), UL 2601- 1 and CAN/CSA C22.2 No
601.1.
Standards Manufacture in accordance to ISO 9001 and
EN 46001 CE marked (In accordance with the Medical
Devices Directive 93/42/EEC)
Ordering Information
129-000 VetPro 2000 Infusion pump c/w mains cable, carry case and
operators manual.
129-010 Veterinary Administration Set – Sold in packs of 25. 129-050 Cage mounting bracket – fits Suburban, Shor, Technik, & HC two
cages. 129-051 Cage mounting bracket – fits Geeling cages and others with
50mm x 50mm mesh doors.
Page 10
VET-PRO Operator’s Manual
10
Trumpet and Flow Rate
In VET-PRO, as with all infusion systems, the action of the pumping mechanism and variations cause short – term fluctuations in rate accuracy.
The following curves show typical performance of the system in two ways:
1. The accuracy of fluid delivery over various time periods is measured (Trumpet curves).
2. The delay in onset of fluid flow when infusion commences (Start up curves).
Trumpet curves are named for their characteristic shape. They display discrete data averaged over particular time periods or “Observation windows”, not continuous data versus operating time. Over long observation windows, short­term fluctuation has little effect on accuracy as represented by the flat part of the curve. As the observation window is reduced, short-term fluctuations have greater effects as represented by the “mouth” of the trumpet.
Knowledge of system accuracy over various observation windows may be of interest when certain drugs are being administered. Short term fluctuations in rate accuracy may have clinical impact depending on the shelf-life of the drug being infused and the degree of intravascular integration, the clinical effect cannot be determined from the trumpet curves alone.
The Start-up curves represent continuous flow versus operating time for two hours from the start of the infusion. They exhibit the delay in onset of delivery due to mechanical compliance and provide a visual representation of uniformity. Trumpet curves are derived from the second hour of this data. Tests performed according to IEC 601-2-24 standard.
Page 11
VET-PRO Operator’s Manual
11
Start-Up Graph at 1.0 ml/h (Initial Period)
Using Human BodySet
Time (minutes)
Trumpet Graph at 1.0 ml/h (Initial Period)
Using Human BodySet
Overall Mean Error = -0.1% Max Rate Error = 8%
Trumpet Graph at 1.0 ml/h (After 4 Hours)
Using Human BodySet
Overall Mean Error = -1.1% Max Rate Error = 6%
Page 12
VET-PRO Operator’s Manual
12
VET-PRO Front View
No.
Description
Function
1.
Display
Displays: Selected Infusion Rate, Total Time, Time Delay, Volume To Be Infused and Volume Infused.
2.
START/ OK Key START
OK
Starts pump operation and delivery.
Confirms selection and setting.
3.
STOP/NO key STOP
NO
• Silences an alarm condition and pauses
operation.
• Zeroes the displayed value during setting
•
Press to return to previous screen
4.
Numeric Keyboard
Used to set the RATE, TIME limit and VTBI. Dial the required number. Pressing NO will erase the last digit.
Rate: 125 ml /h
Press OK for Volume
8
7
3
4
9
Page 13
VET-PRO Operator’s Manual
13
5.
Arrow keys
2£ 0¤
Enable scrolling through options.
Pressing Syringe key during data setting: Enable Flushing procedure. Flushing default volume: 4 ml.
6.
Syringe key
Pressing Syringe key during PCA program operation:
Enable bolus procedure.
7.
On / Off key
Turns the pump on and off.
§ To turn the pump on press until the self- test screen appears
§ To turn the pump off press until the OFF Graph is all black and a beep is sounded.
8.
Operation LED
Flashing of the LED indicates that the set program is running.
9.
Door Latch
Opens the pump door. Lift to open the door.
10.
Info key
i
Pressing Info key supplies information about the pump and its programs (see info mode chapter).
FLUSH 2 ml
0
4
B Volume, ml.
O L U
0 2.8 5
Page 14
VET-PRO Operator’s Manual
14
VET-PRO –with door open
No. Description
Function
1.
Latch holder
Holds the door closed when latch is in vertical position.
2.
Pressure sensor
Detects downstream tubing restriction and occlusion. Alarm level can be adjusted to suit patient needs.
3.
Tubing Guide
Guide the administration set
4.
Pressing Plate
Connected to the door by 2 springs.
5.
Covers the pressing plate.
6.
Air sensor – front
Ultrasonic air detector, which is mounted on the front housing.
7.
Air sensor
-
door
Mounted on the door.
2
1
3
4
5
Page 15
VET-PRO Operator’s Manual
15
The Charger
IV Set Requirements
The VET-PRO pump should be operated only with the Bodyset.
The Bodyset administration set is supplied with an integrated valve (see drawings below).
No
. Description Function
1.
Flow counter
Check the Flow
2.
Release Handle
Push to release the pump from the Charger.
F
Note:
The pump is protected against overcharging. In order to keep your battery fully charged, connect the pump to the mains via the charger whenever possible.
2
1
Page 16
VET-PRO Operator’s Manual
16
Optional Cage Bracket
Fig.1
Fig.2
Fig.3
Fig.4
Engage the pole of the bracket in the slot on the rear of the charging base (Fig.1). Rotate the clamp wheel until the pole is held firmly. Ensure that the horizontal arms of the bracket face outward and are just above the top of the pole slot.
Locate the arms of the bracket through the grille of the cage door (Fig.2 & 3). The arms should be rested on a horizontal grille brace (Fig.4).
Stock code 129-050 is suitable for use on most popular cage manufacturers doors such as Suburban, Shorline, Technik and HC Two. Stock code 129-051 is suitable for use on cage doors with a mesh size of 50mm x 50mm such as Geeling.
Please contact our sales team if you need guidance on the suitability of these brackets for your cage doors.
Page 17
VET-PRO Operator’s Manual
17
Unpacking:
1.
Carefully remove the pump and charger from the box.
2.
Make sure no items were damaged during shipment
3.
Make sure you have the following items:
•
VET-PRO Infusion Pump
•
Charger
•
Battery Pack
•
Operator’s Manual
•
If any items are missing or damaged, contact your VET-PRO supplier.
Loading the Administration Set
F
Note:
Follow the Instructions supplied with the individual administration set. Use of any administration set, other than the original manufacturers, may impair the operation of the pump and the accuracy of the infusion.
1.
Lift the door latch to open the pump door.
2.
Insert the IV tubing into the tubing guide, following the direction flow label located above the tubing slot. Insert the tubing from left to right, and avoid stretching or pulling the tubing. Check key is located in correct position.
3.
Ensure that the distal flow valve is on the right hand side of the pump.
4.
Close the pump door. Press on the door to close firmly, ensure that the latch is correctly applied.
F
Note:
Ensure that the tubing is inserted completely into the pumping channel.
ADMINISTRATION SET RE-ORDER CODE 129-010 SUPPLIED IN PACKS OF 25
Page 18
VET-PRO Operator’s Manual
18
Flush Function (Priming)
Flush Function is used:
1.
Before starting a program (see instruction below);
2.
After the air in line alarm, as follows: Press STOP/NO key to mute the alarm. Disconnect IV tubing from patient; press the SYRINGE key to push air bubbles through. When flushing is completed press START/OK key to resume program.
Flush operation
1.
Switch on the pump using the ON/OFF key. (Press until the self-test screen appears). After a short self-test, the pump will display the program setting and status followed by the last setting rate.
FLUSH
2.
Press the SYRINGE key to flush the administration set.
ENSURE THE SET IS NOT CONNECTED TO THE PATIENT !!
Press OK to start
3.
Press the START/OK key to start flushing.
4.
The display will show a graph indicating the flush operation.
FLUSH 2 ml
0 4
5.
Once the flushing is completed, the display will change back to the previous screen the pump was in, before flushing.
F
Note:
The flushing rate can be adjust, in adjust setting options, up to 1200 ml/h. The default is 600 ml/hr. During flushing the air in line alarm is disabled.
I
Warning
Always disconnect IV tubing from patient before starting flushing procedure.
Page 19
VET-PRO Operator’s Manual
19
VET-PRO – Operation
Before attaching the VET-PRO to a patient, run the following tests to verify that all indicators and alarms work properly.
When an alarm is activated the following will occur:
1.
An alarm message will appear on the display.
2.
An audible alarm will be sounded.
3.
Infusion will cease.
Pump Operation Test:
1.
Click the pump into a Charger connected to AC power. Verify that the CHARGE indicator is on (on charger).
2.
Load an administration set in the pump (as described on page 17).
3.
Switch on the pump using the ON/OFF key (press until the self-test screen appears).
4.
The pump will display "Self Test" followed by the program selected and its status. During the self-test a long beep will be performed. This procedure verifies that the pump’s acoustic and visual features are working properly. After overview, the display will show the last rate setting.
FLUSH
5.
Press the SYRINGE key to flush the administration set.
Press OK to start
6.
Press the START/OK key to start flushing.
7.
The display will show a graph indicating the flushing operation.
FLUSH 2 ml
0 4
Pump Unattended
Press start key
8.
When flushing is completed, wait without pressing any key. Verify that when two minutes have passed without any key being pressed an audible alarm is activated and a “Pump Unattended” message appears on the display.
9.
Press the START/OK key to silence the alarm.
Page 20
VET-PRO Operator’s Manual
20
Air in Line - Alarm Test:
1.
Lift the door latch to open the pump door. Insert an administration set in the tubing guide of the pump (as described on page 17).
2.
Close the door. Press the door until a click is heard.
Rate 500 ml/h
3.
Set the infusion rate at 500ml/hr, using the Numeric keyboard.
Press OK for Volume
Volume 20 ml
4.
Set Volume at 20 ml.
Press OK to start
5.
Press the START/OK key to start operation. Turn the drip chamber upside down, to allow an air bubble greater than 5 mm to enter into the administration set.
Air in Line
6.
AIR alarm will be activated as soon as the air bubble enters the set segment located behind the pump door.
Clean air
7.
Press the STOP/NO key to turn the alarm off.
STOP
2 min
8.
STOP appears on the display.
Press OK to start
Page 21
VET-PRO Operator’s Manual
21
Door Open - Alarm Test:
1.
Insert an administration set into the pump (as described in page 17) and flush it (as described on page 18).
Rate 500 ml/h
2.
Set the infusion rate at 500ml/hr, using the numeric keyboard.
Press OK for Volume
Volume 20 ml
3.
Set Volume at 20 ml.
Press OK to start
3.
Press the START/OK key to start operation.
4. Open the pump door.
Door Open
5
Door Open alarm will be activated at once.
Close door
6.
Close the door.
STOP
2 min
7.
STOP appears on the display.
Press OK to start
Page 22
VET-PRO Operator’s Manual
22
Down Occlusion - Alarm Test:
1.
Insert an administration set into the pump (as described in page 17) and flush it (as described on page 18).
Rate 500 ml/h
2.
Set the infusion rate at 500ml/hr.
Press OK for Volume
Volume 20 ml
3.
Set Volume at 20 ml using the numeric keyboard.
Press OK to start
4.
Occlude the administration set (between the pump and the patient).
5.
After 10 seconds Down Occlusion alarm will be activated (At a rate of 60 ml/hr the Down Occlusion alarm will occur after 60 sec. at a rate
of 1 ml/hr after 13 min.)
Down Occlusion
Press STOP to mute
6.
Press STOP/NO key to silence the alarm.
STOP
2 min
7.
STOP appears on the display.
Press OK to start
F
Note:
The pressure threshold can be changed to suit patient need (see Changing the current default Pressure Limit, page 26). Always ensure that the tubing and fluid are at least at room temperature (>19 deg C ) Giving set tubing is effected by temperature and the pump may not be able to operate is the tubing is too stiff. This would cause a “Down Occlusion Alarm”
Charge Indicator Test:
1.
Click the pump into the Charger, connect the Charger to the mains and check on the Charger that the pump's battery is being charged (red led for charging and green led if the battery is full). Disconnect the power cord from the AC power outlet.
2.
Verify that the CHARGE indicator is off.
3.
Connect the power cord to the AC power outlet. Verify that the CHARGE indicator on the Charger is on.
Page 23
VET-PRO Operator’s Manual
23
Programs
The VET-PRO infusion pump features 5 different programs. Following are the specifications for each program:
CONTINUOUS program
Rate: 0.1 to 1200 ml/hr Total volume to be infused: 0.1 to 9999 ml.
25 STEPS program
User can design a specific protocol, for drug delivery, which contains up to 25 steps. Infusion Rate: 0.1 to 1200 ml/hr. Volume: 9999 ml. For each step. Accumulated volume for all steps- up to 10 liters.
TPN program
To be used for Total Parenteral Nutrition The program has a taper up/taper down pattern. The parameters to be set are: Total Volume, Time Up, Time Down and Total Time. The pump will calculate the maximal rate for verification.
INTERMITENT program
To be used for setting a Dose regiment, at set intervals.
PCA program
To be used for Pain Relief. The PCA program combines basil rate and pre-programmed Boluses. Rate: 0 to 100 ml/hr. Total volume to be infused: up to 1000 ml
Page 24
VET-PRO Operator’s Manual
24
Changing Current Program:
1.
Switch the pump on by pressing ON/OFF key.
2. When the screen of Rate appears, press info key twice.
Volume Infused Show Setting
3. Use ARROW keys to scroll to "Adjust Setting"
option.
=> Adjust Setting
4. Choose "Adjust Setting" with START/OK key.
Buzzer level
=> More …
5. Use ARROW keys (2?, 0?) to scroll to "More".
Show Setting
6. Confirm with START/OK key.
Technician
7. Set technician Code (Code 901), using the numeric keyboard and confirm with START/OK key.
Code
901
Pressure Default
=> Select Program
8. Scroll through the presetting parameters, using the ARROW keys (2?, 0?), until the "Select Program " option. Confirm with START/OK key.
Flushing Volume Select program:
=> CONTINUOUS
9. Choose the required program, using the
ARROW keys, and confirm with the START/OK key.
TPN
History
=> Restart Pump
10.
Choose “Restart Pump” Option and confirm with the START/OK key.
Pressure Default
F
Note:
The selected program will remain in the pump’s memory until
changed by the user.
Page 25
VET-PRO Operator’s Manual
25
F
Note:
User may change programs using the Short Cut:
CONTINUOUS
PROGRAM
1.
C
O N T
Press NO to change
When switching the pump on, after a short Self Test, the pump will display the last selected program for 5 seconds.
During this time, press the STOP/NO key for 2 sec. (Pressing any other key will elapse the pump to the Rate screen).
Select program:
=> CONTINUOUS
2. The screen will show the “Select Program”
menu.
TPN
3. Choose the required program, using the ARROW keys,
and confirm with the START/OK key.
F
Note:
If the current selected program is PCA, the Short Cut option is not available.
Page 26
VET-PRO Operator’s Manual
26
Down Occlusion Setting:
1.
Switch the pump on, by pressing ON/OFF key
2. When the screen of Rate appears, press info key twice.
Volume Infused Show Setting
3. Use ARROW keys (2?, 0?) to scroll to "Adjust
Setting" option.
=> Adjust Setting
4. Choose "Adjust Setting" with START/ OK key.
Adjust setting:
=> Pressure Level
5. The option selected is “pressure level”.
Buzzer Level
6. Confirm with START/OK key.
Pressure setting
High
10 psi
7. Choose the desired pressure level, using the
arrow keys (2?, 0?).
Adjust ?? & press OK
8. Press the START/OK
key to store the parameter in the pump’s memory.
F
Note:
The following pressure parameters can be set:
Low = 5 psi (0.3 bar)
Normal = 7.5 psi (0.5 bar)
High = 10 psi (0.7 bar)
F
Note:
The selected pressure setting will remain in the pump’s memory
only until turning off (and then return to the default - HIGH).
F
Note:
In the event a down occlusion situation occurs, the patient is
protected by an anti-bolus feature, which prevents bolus delivery to the patient when the downstream occlusion is released.
Page 27
VET-PRO Operator’s Manual
27
Connection of Pump to IV Pole
1. Connect the giving set to the medication bag. Mount the bag on an IV pole.
2. Click the pump into the Charger. Mount the Charger on an IV pole.
3. Connect the Charger via a standard power cord to a grounded AC outlet.
4. Open the door of the pump and insert the administration set into the pump’s tubing guide (as described on page 17)
5. Switch the pump ON using the ON/OFF
key
(press until the self-test screen appears).
CONTINUOUS
PROGRAM
6. The program and status will appear on the display followed by the last rate setting.
C O N T
Press NO to change
7. Flush the administration set by pressing the SYRINGE key, and confirm with START/OK.
8. Set the required program (see program select option on page 24) and continue as described for every program setting.
9.
Verify that infusion is proceeding normally before leaving the pump unattended.
Continuous Program
Switch the pump ON using the ON/OFF key.
CONTINUOUS
PROGRAM
C O
N
T
Press NO to change
CONTINUOUS
Pump displays program name and status. The
status is either "Lock out" or "program" (locking and unlocking the pump – see page
60).
C O
N
T
LOCK OUT
Page 28
VET-PRO Operator’s Manual
28
Program Setting
F
Note:
Whenever you decide to set a new program, follow instructions at the bottom of screen.
Rate over Volume
Rate 500 ml/h
1.
The pump will display the last set rate. Use the numeric keys to change or press START/OK to confirm.
Press OK for Volume
Rate: 500 ml/h
During operation the pump allows rate titration, by pressing a numeric or digital point key.
Adjust & press OK
F
Note:
The rate on the display will change during titration, however operation rate will change only if the START/OK key is pressed within 10 seconds confirming the rate change. If the rate change is not confirmed within 10 seconds, the pump will continue operation on the original rate, and the display will return to the previous rate set.
F
Note:
From the STOP mode, the user may return to the Rate screen by pressing twice on one of the numeric keys.
F
Note:
During stop or data setting mode, use the STOP/NO key to return one step backwards: First press will erase the last digit
displayed, once all the digits are erased, pressing the
STOP/NO key will display the previous screen .
Volume 456 ml
Press OK to start
2.
When the Rate is confirmed, the pump will display the last set volume (if the time option is set off). Use the numeric keys to change the Volume or press START/OK to confirm and start the program operation. Setting Program Delay– Refer to page 48.
Page 29
VET-PRO Operator’s Manual
29
.
During program operation, the pump will display the rate in which the program is operating.
Rate
250 ml/h
Note:
-If one of the values is changed during the programming mode, the pump will not store the new value, unless changes are confirmed by pressing the START/OK key.
-If the flush key is activated during data setting, the pump will return to the original (previous) screen, once flushing is completed.
Resume Press OK
In the event the program is in lock mode or if the time option is on and the previous program has been interrupted the pump will display the Repeat/Resume screen.
C
O N T
Repeat Press NO
To continue the previous program (from where it stopped) press the
START/OK key.
To repeat the previous program from the beginning, or start a new
program (if pump is not in lock mode) press the STOP/NO key.
Volume over Time
Volume 456 ml
C
O N
T
Press OK key
1.
If time option is on (adjust setting – see p. 60) after displaying program name and status, pump will display last set volume, set or confirm the displayed number by pressing the START/OK key.
otal Time 10:30
2.
C
O N
T
Press OK key
The display will show the time screen, set the time, for a volume
over time program, and press START/OK to start operation.
Setting Program Delay– Refer to page 48
Page 30
VET-PRO Operator’s Manual
30
Time setting options: A.setting 4 digits: 2 first digits will represent hours and the other 2 will represent minutes B.setting 2 digits for full hours C.setting 1 digit for hours and using the digital point to set minutes. D.using the digital point to set only minutes.
.
During program operation, the pump will display the rate in which program is operating.
Rate
250 ml/h
Note:
-If one of the values is changed during the programming mode, the pump will not recall the new value, unless changes are confirmed by pressing the START/OK key.
-If the flush key is activated during data setting, the pump will return to the original (previous) screen, once flushing is completed.
25 Steps Protocol
In the 25 steps program, the user can set up to twenty-five different steps.Each step has its own rate and volume.The steps accumulation is limited to a Volume of 10 liters. Protocol steps are delivered one after another, sequentially, until all steps are completed.
Switch the pump ON using the ON/OFF key.
25 - Steps
PROGRAM
S T E P
Press NO to change
25 - Steps
Pump displays program name and status. The
status is either "Lock out" or "program". (locking and unlocking the pump – see page
50).
S T E P
LOCK OUT
Page 31
VET-PRO Operator’s Manual
31
Program Setting
F
Note:
Whenever you decide to set a new program, follow the instructions at the bottom of the screen.
Note:
-If one of the values is changed during the programming mode, the pump will not store the new value, unless changes are confirmed by pressing the START/OK key.
-If the flush key is activated during data setting, the pump will return to the original (previous) screen, once flushing is completed.
Step 1 Rate
100 ml/h
.
The pump will display the previous rate for the first step. Set a new rate, using numeric keys, or confirm pressing START/OK.
Press OK for Volume Step 1 Volume
1000 ml
.
Pump will display the previous volume for first step. Set a new volume, or confirm pressing START/OK key.
Press OK to continue
3.
Pump will display the rate for step 2. Repeat directions as for step 1 and repeat through all the steps required, up to 25.
Step 11 Rate
0 ml/h
4.
Press OK for Volume
After last required step, set rate to zero. This indicates that all data required for the protocol has been set. All protocols following the last set protocol, will be automatically zeroed .Press START/OK to confirm.
Total Steps 10
Total Volume 5000ml
5.
Press OK to start
After last step rate was set to zero or after 25 steps are completed, the pump will display the accumulated data setting, to enable comparison and verification. Press START/OK to start the program or STOP/NO to scroll through all the protocol settings again. Setting Program Delay– Refer to page 48.
Rate Step 1 of 10
6.
During the program operation, the display updates the user on which step the pump is running.
S T E p
100 ml/h
F
Note:
Pressing the STOP/NO key shall erase the last digit. Once no digit is displayed, pressing the STOP/NO key will change the screen to show the previous display.
Page 32
VET-PRO Operator’s Manual
32
TPN Program
Total Parenteral Nutrition - This mode delivers an infusion while
automatically tapering up and down. The pattern of the program is a trapezoid.
Rate
Time
On a TPN Program the user may set the total volume which corresponds to the size of the bag to be infused, the total time, and the up and down time. The pump displays the maximal rate, as calculated, from the set parameters. The maximal rate display is used to verify that the setting corresponds to the requested data, and to ensure that the maximal rate is suitable to the patient.
Up and down time can be set to zero. A TPN program with up and down time set to zero, is a continuous program.
Switch the pump ON using the ON/OFF key.
TPN
PROGRAM
T P
N
Press NO to change
TPN
Pump displays program name and status. The
status is either "Lock out" or "program". (locking and unlocking the pump – see page
50).
T P
N
LOCK OUT
Resume Press OK
In the event a TPN program was interrupted and was not completed, the pump will display the Resume/Repeat screen.
T P
N
Repeat Press NO
T1 T2 T3
Max Rate
T1 = Time UP
T2= Time FLAT T3= Time DOWN
Page 33
VET-PRO Operator’s Manual
33
To continue the interrupted program, press the START/OK key. To repeat the previous program or start a new
Volume 1256 ml
Max Rate: 234 ml/h
T P
N
Press OK to start
If you choose to complete an interrupted
program, the pump will display interrupted program data. Scroll with the ARROW keys to check program data. Pressing the START/OK key will start the program operation.
Program Setting
F
Note:
Whenever you decide to set a new program, follow the instructions at the bottom of the screen.
Note:
-If one of the values is changed during the programing mode, the pump will not store the new value, unless the change is confirmed by pressing the START/OK key.
-If the flush key is activated during data setting, the pump will return to the original (previous) screen, once flushing has been completed.
Volume 1256 ml
1.
The first parameter to set is the volume of the TPN bag.
T
P
N
Press OK key
Total Time 12:30
T
P
N
Press OK key
2.
Set the total time for the program to run: Options are: a. Setting 4 digits - 2 first digits will represent hours and the other 2 will represent minutes b. Setting 2 digits for full hours c. Setting 1 digit for hours and using the digital point to set minutes. d. Using the digital point to set only minutes.
Up Time: 0:30
3.
Set the time for the rate to taper up (up to 4:15 h).
T
P
N
Press OK key
Down Time: 0:30
4.
Set the time for the rate to taper down
(up to 4:15 h).
T
P
N
Press OK key
Page 34
VET-PRO Operator’s Manual
34
Volume 1256 ml
Max Rate: 108 ml/h
T
P
N
Press OK to start
5.
The pump will calculate all the set data and display the Maximal rate for the program (to be infused during Flat period). Press the START/OK key to start the program. Setting Program Delay– Refer to page 48.
F
Note:
During the stop or data setting mode, use the STOP/NO key to return one step backwards: First press will erase the last digit displayed, once all the digits are erased pressing the STOP/NO key will display the previous screen .
Rate U
P
2 ml/h
6.
During the TPN program operation, the pump will display the actual rate and program direction (Up/Flat/ Down). Program operation will start from 2 ml/h and will increase to the maximal rate.
Rate
. During the Flat session the rate will not change
and the pump will show the following message:
F
L A T
108 ml/h
Rate
.
During the last part of the program (T3), the program will start to decrease the infusion rate according to the set taper down time.
D
O W N
25 ml/h
F
Note:
T P
N
Taper Down: 0:30
Press OK key
If the program must be interrupted early, and the planned taper down time must commence immediately press the STOP/NO key continually (2 sec) to stop the pump operation and display Taper Down screen.
Set a new time or confirm the default setting. Press the START/OK
key to start the taper down process immediately.
Page 35
VET-PRO Operator’s Manual
35
Intermittent Program
This is a special program, which enables the design of a Dose delivery protocol at set intervals. Each Dose will be given at a set time. Between Doses the pump will Keep the Vein Open. The infusion program continues until the total infusion bag is infused.
This “Launch and Forget” program enables the operator to program medication treatment for long treatments. For treatments like, IV Antibiotics, giving every 4 hours a dose of 100 ml, the intermittent program is saving 5 administration sets, during 24 hours, and essential time of the operator.
Example: Programs setting
Program
Rate Volume Time
Period-1
100 ml/hr 100 ml. 1:00
Period-2 - KVO
5 ml/hr. 20 ml. 4:00
Period-3
100 ml/hr 100 ml. 1:00
Period-4 - KVO
5 ml/hr. 20 ml. 4:00
Period-5
100 ml/hr 100 ml. 1:00
Period-6 - KVO
5 ml/hr. 20 ml. 4:00
Switch the pump ON using the ON/OFF key.
Intermittent
PROGRAM
I N T R
Press NO to change
Intermittent
Pump displays the program name and status.
The status is either "Lock out" or "program". (locking and unlocking the pump – see page
50).
I N T R
LOCK OUT
Page 36
VET-PRO Operator’s Manual
36
Program Setting
F
Note:
Whenever you decide to set a new program, follow the instructions at the bottom of the screen.
Note:
-If one of the values is changed during the programing mode, the pump will not store the new value, unless the change is confirmed by pressing the START/OK key.
-If the flush key is activated during data setting, the pump will return to the original (previous) screen, once flushing has been completed.
Rate 4 ml/h
1.
The pump will display the last set rate. Change or press START/OK to confirm.
Press OK for Volume
Dose Vol: 100 ml
2.
The pump will display the last set Dose Volume. Change or press START/OK to confirm
Press OK to continue
Pause 04:00
3.
Adjust the time pause between Doses (up to 20:00 hours), or confirm the last setting by pressing the START/OK key.
Press OK to continue
Volume 1000 ml
4.
The pump will display the last set Volume (total volume to be infused). Use the numeric keys to change or press START/OK to confirm and start program operation.
Press OK to start
5.
During the intermittent program the display will show continually the actual rate of infusion.
Rate
5 ml/h
6.
During a pause between the doses, the display will show the KVO screen.
Delay 4:00
KVO 5 ml/h
Page 37
VET-PRO Operator’s Manual
37
Rate: 500 ml/h
During operation the pump allows rate titration, by pressing a numeric or digital point key.
Adjust & press OK
F
Note:
The rate on the display will change during titration, however the operation rate will change only if the START/OK key is pressed within 10 seconds confirming the rate change. If the rate change is not confirmed within 10 seconds, the pump will continue operation on the original rate, and the display will return to the previous rate setting.
F
Note:
During stop or data setting mode, use the STOP/NO key to return one step backwards: First press will erase the last digit displayed, once all the digits are errsed pressing the STOP/NO key will display the previous screen .
F
Note:
Delay Before start option is cancelled in the Intermittent program.
PCA Program –
With protocols
Patient Control Analgesia is a program used in subcutaneous, IV or
epidural pain relief. Pain management is undertaken by delivering a continuous basal rate, in combination with pre-programmed Boluses. The pump will deliver only the boluses corresponding to the preset time intervals. All attempts to obtain a bolus are stored in the pump memory. The operator can read these statistics and adjust the settings as required.
It is possible to set and store up to 5 different protocols for PCA, including names of drugs.
Switch the pump ON using the ON/OFF key.
Page 38
VET-PRO Operator’s Manual
38
PCA
P C A
PROGRAM
PCA
P C A
PCA MODE
PCA
P C A
Epidural Mode
Pump displays program name and status. The status is either " Program ", "PCA Mode" or “Epidural Mode”. The PCA program operates only in a lock mode.
Program mode will be displayed if no protocol has been locked. PCA mode will be displayed if a protocol has been locked and Epidural mode is Off. Epidural mode will be displayed if a protocol has been locked and Epidural mode is On – see epidural mode chapter. In all cases entering the PCA parameters and/or changing them, requires a code.
Program Setting
F
Note:
Whenever you decide to set a new program, follow instructions at the bottom of the screen.
1.
To enter PCA menu, pump will require a code. Use numeric keys to set code 007 and confirm with START/OK key.
P C A
Enter Code
007
Press OK key
.
P C A
=> Flush
MD Bolus History
Pump will display the PCA menu. Use the arrow keys to scroll through the different options, and select the required option, using the START/OK key. Selecting the “More..” option will display the Adjust Setting second selection menu.
3. Flushing is disabled during operation, therfore, always flush the set before starting a program.
Selecting the flush option from the PCA menu, will display the flush previous screen. Pressing the START/OK key will activate the flush procedure. If no key will be pressed within 10 sec. the pump will return to the PCA menu screen. When the flush is completed the pump will go back to the PCA menu screen.
F
Note:
The flushing rate can be adjusted, in adjust setting options, up to 1200 ml/h. The default is 600 ml/hr. During flushing the air in line alarm is disabled.
Page 39
VET-PRO Operator’s Manual
39
I
Warning
Always disconnect IV tubing from patient before starting flushing procedure.
4.
Select the “Modify Program” option and press the START/OK key.
P C A
Select Program
=> Modify Program
More
5.
Use the numeric key to set the code 008 and press the START/OK key.
P C A
Enter Code
008
Press OK key
Epidural Mode
OFF
6.
Adjust ?? & press OK
If Epidural mode is required (see p.53) use arrows key to change the setting to ON and approve using the START/OK key. To continue with epidural mode OFF, press START/OK key and move to step 8.
Air sensor alarm
ON
Adjust ?? & press OK
7.
If Epidural Mode was set ON, the user will have to decide if to set the air alarm on or off. Use arrows key to change the setting as required and approve using the START/OK key.
8.
P C A
Clear History & Start New Bag
Press the START/OK key to start new bag and clear previous program data. Pump will display bag volume screen - continue to step 9. Pressing the STOP/NO key will display program accumulation screen. Pressing START/OK key from accumulation screen will take the VTBI left from previous program as total vtbi for the new program. continue to step 10.
9.
Pump will display the Bag Volume screen (with total volume of previous program). Use the numeric keys to set a new value or approve the previous value, using the START/OK key.
P C A
Bag Vol: 500 ml
Press OK key
A:MORPHINE
B: METHADONE
10.
C: PETHIDINE
Pump will display the list of 5 protocols with drug names. Use the arrow keys to scroll through the options. Pressing the START/OK key will display the rate screen of the chosen protocol. Pressing the STOP/NO key twice will enable the user to change the drug name.
Page 40
VET-PRO Operator’s Manual
40
Use the STOP/NO key to erase the displayed drug name.Use the arrow keys and keys of 4 and 6 to move between letters. Select the letters using the START/OK key. Once the protocol name is completed, return the select key to the empty space, before the letter A and repeatedly press the START/OK key until the display returns to the Protocol List.
11.
P
A
Rate 50 ml/h
Adjust & press OK
Pump will display the rate screen (the chosen protocol will be displayed on the left column of the screen). Use the numeric keys to set a new value (up to 100 ml/hr) or approve the previous value, using the START/OK key. Rate can be displayed in mg/ml, as well. Using this option – go to step 12, if not, continue to step 13.
12.
P C A
Concentration
0.5 mg/ml
From the rate screen press the STOP/NO key 3 times to display the concentration screen. Use numeric keys to set concentration and approve by pressing the START/OK. Pump will go back to the protocols list. To return from mg to ml, press STOP/NO key twice, from rate screen, enter a concentration of 0 mg/ml, and press the START/OK key for approval.
13.
Pump will display the Bolus screen. Use the numeric keys to set a new value or approve the previous value, using the START/OK key.
P
A
BOLUS 5 ml
Adjust & press OK
14.
P
A
LockTime 60 Min
Adjust & press OK
Lock time is the minimal time between boluses. When a bolus is required during the lock time, a beep will be sounded but no bolus will be given. Use the numeric keys to set a new value or approve the previous value, using the START/OK key.
15.
P
A
4 H Limit 10 ml
Adjust & press OK
4 H Limit is the maximum volume to be given in 4 hours. Once this volume will be infused the pump will operate in the KVO mode, to complete the 4 hours. Use the numeric keys to set a new value or approve the previous value, using START/OK key.
16.
PCA program operates only in a lock mode. Press the START/OK to lock the program or STOP/NO to go back to protocols list and scroll again through the data.
P C A
Lock:Press OK Scroll:Press NO
Page 41
VET-PRO Operator’s Manual
41
Bag Vol: 500 ml Volume Left 500 Infused 0
17.
P
A
Press OK to start
The pump will display the protocol accumultion screen. Use the arrow keys to scroll through the data. Press the START/OK to start the program operation. Pressing the
STOP/NO twice will go back to menu screen.
Note:
-If one of the values is changed during the programing mode, the pump will not store the new value, unless the change is confirmed by pressing the START/OK key.
MORPHINE
Rate 50 ml/h
18.
During the PCA program the display will show the selected protocol, the name of the drug, the rate of operating and the bolus volume.
P
A
BOLUS 5 ml Volume 5 ml
19.
B O L U
0 2.1 5
Pressing the Syringe key and approving with
START/OK key will activate the Bolus Operation. When the Bolus is given, the pump
will display the progress of the procedure.
A bolus can also be given by pressing twice on the bolus cable.
20.
P C A
Resume:Press OK Menu:Press NO
After a program has been locked, whenever the pump will be turn on, resume-menu screen will be displayed. Pressing START/OK key will display accumulation screen, with Volume to be infused, as volume left from previous program. User can only continue previous program by pressing again START/OK key. Pressing STOP/NO key, will display code setting screen. Set code 007, to enter the PCA menu.
Start new bag
1.
Selecting “Start new Bag”, from the PCA menu, will display another screen for approval.
P C A
History
=> Start New Bag
Select Program
Page 42
VET-PRO Operator’s Manual
42
2.
Pressing the START/OK key will start a new bag (Set total bag as Volume to be infused with no regards to previous Volume Infused). All other parameters are locked and can not be changed. Pressing the STOP/NO key will go back to the menu.
Start New Bag
Resume – Press OK
Air Sensor Alarm
ON
3.
Adjust ?? & press OK
Pump will display Air sensor alarm message, if epidural mode is On. If epidural mode is off, air alarm is automatic on and pump will display accumulation screen – continue to step 4. If epidural mode is on, user can choose to set air sensor to OFF (use arrow keys to change air sensor status and press START/OK key for approval).
Bag Vol: 100ml Volume Left 100 Infused 5
4.
P C A
Press OK to Start
Data includes previous Volume Infused. Pressing the START/OK key, from accumulation screen, will start program operation. Pressing the STOP/NO key will go back to display resume-menu screen.
F
Note:
After end program, user can start a new bag without a code. Pump will display the option to start a new bag or going back to menu. Pressing START/OK key will display program accumulation screen. Approving with START/OK will start the program.
Selecting a Program
1.
Selecting “Select Program”, from the PCA menu, will display 2 options – to start new bag or to continue with previous bag.
P C A
Start New Bag
=> Select Program
Modify Program
2.
P C A
Clear History & Start New Bag
Pressing the START/OK key will clear the
volume infused history and start a new bag.
Pressing the STOP/NO key will continue the program with previous Volume Infused. In both cases pump will continue to display accumulation screen.
Page 43
VET-PRO Operator’s Manual
43
Bag Vol: 500ml Volume Left 500 Infused 0
3.
P C A
Resume - Press OK
Pressing the START/OK key will display the protocols list. Pressing the STOP/NO key will display again PCA main menu.
A:MORPHINE
B: METHADONE
4.
Use the arrow keys to scroll through different protocols and select with the START/OK key.
C: PETHIDINE
Rate 50 ml/h BOLUS 5 ML LockTime 60 Min
5.
Pump will display the protocol accumulation screen. If Epidural mode was set to OFF, pressing START/OK key will start program operation.
P
A
4 H Limit 100 ml
Air sensor alarm
ON
Adjust ?? & press OK
6.
If epidural mode is On, pump will display Air sensor alarm message. User can choose to set air sensor to OFF (use arrow keys to change air sensor status and press START/OK key for approval). Press the START/OK key to start operation.
MD Bolus
P C A
Flush
=> MD Bolus
History
P C
A
MD Bolus Volume 2 ml
Press OK to start
Selecting the MD Bolus, from PCA menu, will display the MD Bolus setting screen. If no key is pressed within 30 sec. the pump will go back to the code screen.
Use the numeric keys to set the Bolus Volume, up to 25.5 ml. When the Bolus is completed the pump will display the Resume/Menu screen. Press the STOP/NO key to go back to the menu code screen or the START/OK key to display the program accumulation screen.
Page 44
VET-PRO Operator’s Manual
44
History
P C A
MD Bolus
=> History
Select Program
H
2 4
From 12.01.02 11:10 To 13.01.02 11:10 Given: 200 ml Bol. Given: 4
Selecting “History“ will display the PCA operation history for the last 24 hours or last program, if started within the last 24 hours. Pump will display: Volume Infused, Bolus Given and Bolus attempted. Use the ARROW keys to scroll through data.
Use the down ARROW key to scroll to the information regarding the last 8 hours. Use the ARROW keys to scroll through the data.
H
8
From To 13.01.02 11:10 Given: 80 ml
Bol. Given: 2
Press the info key for details per hour. Use the ARROW keys to scroll through the data.
Boluses Last 24 Hr
Press the info key once more to display the Boluses given and attempted for the last 24 hours or last program, in a graphic way. Every column in the graph represents 1 hour. The black areas in each column are the boluses given and the white areas are the patient attempts for boluses.
Volume Last 24 Hr
Press the info key once more to display the Volume infused for the last 24 hours or last program, in a graphic way. Every column in the graph represents 1 hour.
F
Note:
History can also be displayed during operation, by
pressing the STOP/NO key and then the info key.
Page 45
VET-PRO Operator’s Manual
45
PCA Program – Epidural Mode
The epidural mode is a variation of the PCA Protocol mode with 2 main differences: the Air in line and back pressure detection operate differently than in other programs.
Air in Line Alarm
Epidural Mode Air Sensor Disabled
There is a choice in air detection system to disable the air in line alarm when Epidural mode is ON. When pump operates in Epidural mode, the display will show every 8 seconds either the message “Epidural mode, air sensor disabled” or “Epidural mode, air sensor enabled”.
Pressure Sensor
In both Bolus mode (bolus request by user and MD bolus) and during flushing, the operation rate will vary as a result of the pressure sensor reading. In case that the back pressure will rise above a preset border, the Bolus and flushing operation rate will go down to maintain a preset backpressure. This system may increase operation time, but protect the patient from high back pressure infusion.
F
Note:
During Epidural mode, pressure limit will be 20 psi, with no regards to the pressure setting. In case that although the rate was adjusted the pressure reach 20 psi an alarm will be activated.
After exit from Epidural mode, pump will automatically return to operate according to the pressure setting.
Setting Epidural mode to ON
Epidural Mode
OFF
Adjust ?? & press OK
Whenever the user chooses to modify a program (using code 008), the epidural mode will be set to OFF. Use arrows key to change the setting to ON and approve using the
Page 46
VET-PRO Operator’s Manual
46
Air sensor alarm
ON
Adjust ?? & press OK
After setting the Epidural Mode to ON, the user will have to decide if to set the air alarm on or off. Use arrows key to change the setting to ON and approve using the START/OK key.
If epidural mode is on, whenever the user chooses the option to start a new bag or select a program, the pump will display the option to set air alarm on/off (if epidural mode is off, air alarm is automatically on).
F
Note:
If the user chooses to resume the program from PCA first screen (before entering PCA menu) all epidural setting will remain without confirming it with the user.
PCA Program – Without protocols
There are models, in which the PCA program has no protocol options. In those models no code is required to set or change the PCA data. The pump should not be in the lock mode (if required, unlock the pump as described on page 51).
In this type of PCA, the user should only set or approve the Bag Volume, Program Rate, Bolus Volume and Locking time (between boluses). Setting is as described in the PCA Program with Protocols.
Page 47
VET-PRO Operator’s Manual
47
Info Mode
Pressing on the info key during operation will display the total Volume To Be Infused and the Infused Volume. The screen will be displayed for 7 seconds.
Infused
Pressing twice on the info key during operation will display the battery status.The screen will be displayed for 7 seconds.
Battery Level
t + 25c
Pressing 3 times on the info key during operation will display the administration set temperature.The screen will be displayed for 7 seconds. This option is not available in the 25 steps program.
Boluses
A
ttempted: 3
Given: 1
Pressing 3 times on the info key, during PCA program, will display Boluses History, from beginning of the program operation.
P
C A
Delay 60Min
Pressing 4 times on the info key during operation will display the time left until the end of the program. This Option is not available in the 25 steps program and in the PCA program.
Time Left
0:59
Time
Pressing 4 times on the info key during TPN operation will display the program status using a graph. The performed part of the trapeze will be painted black. The screen will be displayed for 7 seconds.
Rate: 50 ml/h
BOLUS 1 ml
Lock Time 60Min
Pressing 4 times on the info key during PCA operation will display the data of the present protocol operating.
P
A
4 H Limit
200 ml
ml
Empty Full
0 6:30 12
Page 48
VET-PRO Operator’s Manual
48
Pressing 5 times on the info key during PCA operation will display the time and date.
Time 8:30
Date 24.03.02
To clear the volume infused (VI): press the info key to display the volume infused (from stop or setting mode) and then the STOP/NO key. The screen will change to show volume infused zero.
Infused
Pressing the STOP/NO
key after the program is completed , will erased the
VI. The option to erase the volume infused
through info, is used in case the
last program was not completed.
F
Note:
Exit from the info mode is automatic after 10 sec, or can be achieved by pressing the info key.
Program Delay
Delay of the program before the start is enabled in the Continuous program (both in rate over volume and in volume over time), TPN and the 25 steps program.
C
O N T
Volume 250ml
Press OK for Delay
To enable delay before start option, set delay before start to on, as described on page 48 (Adjust Setting). After setting the program data the pump will enable the delay setting.
C
O N T
Rate 25 ml/h Volume 250ml Time 10:00 Press OK for Delay
Time Delay 15:00
Press OK to start
Use the numeric keys to set the time delay (up to 20 hours). At this stage, if no value was set, pressing the START/OK key wil start the program
operation with no delay.
ml
Page 49
VET-PRO Operator’s Manual
49
Delay 12:20
During the time delay, the KVO session will run. Time left will count down, until 00:00. At this stage the set program will automatically start operation.
KVO 5 ml/h
Program Completed
End PROGRAM
When the program is completed, the following message will appear on the display: press the STOP/NO key to mute the alarm.
Press STOP to mute
KVO 5 ml/h
The pump will automatically turn to a KVO mode when the program is completed. Press the STOP/NO key to reset the pump and set a new program.
Press STOP to renew
If the main program rate is higher than the KVO rate, set in the technician mode, the rate will remain as set in the technician mode. If the main program rate is lower than the KVO set rate in the technician mode, the pump will operate at the program rate. KVO volume is limited to 5 ml.
F
Note:
Before operating the pump, make sure the IV bag contains 5 mls more than the VTBI– for the KVO (and delay time, if applicable)
When the program is completed a special sound algorithm will be performed: 4 single beeps followed by a 4 minute pause, then again 4 beeps, another 4 minute pause and a continuous alarm is activated. When the audible alarm is activated, the display shows the “end program” screen. Between the beeps and alarm pause the display shows the KVO screen.
STOP
2 min
Press the STOP/NO key to stop the pump. This ceases the pump operation and the display shows STOP.
Press OK to start
Stop mode is limited to 2 min. After 2 min. an audiable alarm is sounded. Press the STOP/NO key again, to mute the alarm.
Page 50
VET-PRO Operator’s Manual
50
From STOP mode, switch the pump off by pressing the ON/OFF key (continuously until a beep is sounded).
OFF
WAIT FOR BEEP
F
Note:
The pump cannot be switched off during operation.
Important: Use the technique dictated by the IV container provider.
1. Ensure that the pump is not infusing by placing the pump in STOP mode or turning the pump OFF.
2. Remove the bag spike on the IV set from the empty / used container. Discard
the empty/used container according to the providers instructions.
3. Prepare a new container as per the providers instructions and insert the spike into the new container.
4. Squeeze the drip chamber (if applicable) to fill approximately halfway with fluid.
5. Restart infusion.
Locking
The VET-PRO enables the user to lock the programs, on three different levels.
Level 1 (keys locking)
Level 2 (lock out program)
Level 3 (presetting maximal rates)
Level 1 – key locking
The VET-PRO keys locking function disables all but the info key,
START/OK (during setting mode) and STOP/NO (during operation mode)
keys.
Level – program locking
Page 51
VET-PRO Operator’s Manual
51
The VET-PRO has a unique option, which enables the user to lock out the
numeric keys, so that entered parameters, or set programs cannot be changed once they are set.
This option is important, for example, to prevent inexperienced staff unknowingly changing the program; or for operators repeating the same program every day, it's important to have a lock-out program, to prevent mistakes in setting.
Level 3 – maximal rate locking
The VET-PRO provides the option of setting a maximal rate in the technician mode. The program rate can be adjusted up to a preset limit. For home patient (mainly for TPN or PCA treatments), who must repeat the same program daily, the option of adjusting the delivery regiment is very important.
How to Lock ?
Level 1 – key locking
LOCK Mode
Pressing continuously on the info key, untill the graph is all black and a beep is sounded will lock the program.
OFF ON
“1” will be shown on the right side of the display, indicating that the program is first level locked.
LOCK Mode
Pressing continuously on the info key, untill the graph is all empty and a beep is sounded will unlock the program.
OFF ON
Level – program locking
Press info key twice, from stop or setting mode.
Volume Infused Show Setting
Use ARROW keys to scroll to "Adjust Setting" option.
=> Adjust Setting
Page 52
VET-PRO Operator’s Manual
52
Choose "Adjust Setting" with START/OK key.
Buzzer level
=> More …
Use ARROW keys to scroll to "More" and confirm with START/OK key.
Show Setting
Technician
Set technician Code (Code 902), and confirm with START/OK key.
Code
902
Lock Out
ON
Use one of ARROW keys to change from OFF to ON. Confirm with START/OK key.
Adjust ?? & press OK
The default setting is “Lock Out - NO”. Unlock in the same procedure. “2” will be shown on the right side of the display, indicating that the program is second level locked.
Level 3 – maximal rate locking
The operator can adjust the Maximal rate acceptable according to patient needs:
Press info key twice.
Volume Infused Show Setting
Use ARROW keys to scroll to "Adjust Setting" option.
=> Adjust Setting
Enter "Adjust Setting" pressing START/OK key.
Buzzer level
=> More …
Use ARROW keys to scroll to "More".
Adjust Setting
Confirm selection with START/OK key.
Technician
Set technician Code (Code 901), using the numeric keys, and confirm with START/OK key.
Code
901
Page 53
VET-PRO Operator’s Manual
53
Bolus Rate
=> Maximal Rate
Scroll through the setting parameters, using the ARROW keys, until the "Maximal Rate " option. Confirm with START/OK key.
KVO Rate
Maximal Rate
500 ml/h
Set the maximal rate, using the ARROW keys and confirm with START/OK key.
Adjust & press OK
F
Note:
The default setting for a maximal rate is 500 ml/h. “ ” will be shown on the right side of the display, indicating that the program is in the third lock level only when the max. rate will be set to a rate different than 500 ml/h.
F
Note:
§ The selected Locking Option will remain in the pump’s memory until changed by the user.
§ Whenever the pump is in more than one locking level, the higher level will be displayed on the right side of the screen.
Battery Operation
The VET-PRO can operate on battery power, enabling operation when the patient is being moved or during electrical power failure. When the pump runs on the battery power, the AC icon is off. At full charge, the battery provides 9
hours of operation at an infusion rate of 250ml/hr.
I
Warning
Don’t operate the pump on AC power if the battery is not loaded in the pump for back up.
F Note:
When the pump is not in use, click the pump into the Charger and plug the system into an AC wall outlet (if possible) to charge battery.
F Note:
After the ³end battery“ signal has been activated or following long periods of storage, wait 2 minutes after the pump has been connected to an AC power supply before operating.
Page 54
VET-PRO Operator’s Manual
54
There are two battery alarm conditions:
1.
The pump will warn that the battery is low a minimum of 30 minutes before the “end battery” alarm activates (when the battery power reaches less than 7.1v).
Low Battery
Connect to charger
2.
After the ³battery low´ warnings, when the battery power reaches less than 6.05v, the pump will cease operation and the "end
battery" alarm will be activated.
End Battery
Connect to charger
•
If the battery runs out, End Battery appears on the display. The pump stops the infusion.
•
From the "end battery" state the user cannot restart the pump, until it has been connected to mains.
G
Caution:
Leaving the battery in a discharged state for a long period of time may damage the battery.
F
Note:
Whenever possible, use the pump connected to an AC power supply via the charging unit. This preserves the battery power supply for emergency use or for situations where the AC power is not available.
F
Note:
Replace the battery once every four years.
Cleaning and Storage
Before the transfer of a pump to a new patient and periodically during use, clean the pump by wiping over with a lint-free cloth lightly dampened with warm water and a standard disinfectant / detergent solution.
IWarning:
Before cleaning always switch off and disconnect from the AC power supply.
G CAUTION Do not clean pump with chemicals such as Xylene, Acetone or
similar solvents. These chemicals cause damage to plastic components and paint.
Page 55
VET-PRO Operator’s Manual
55
If the pump is to be stored for an extended period it should be first cleaned and the battery fully charged. Store in a clean, dry atmosphere at room temperature and, if available, employ the original packaging for protection.
Once every 3 months during storage, carry out functional tests as described in p.23-26 and ensure that the battery is fully charged.
Alarms Conditions
At any time, when the pump detects a problem, an alarm is activated and the following will occur:
•
An alarm message appears on the display, stating the cause of the alarm and indicates instructions for continued use.
•
An audible alarm sounds.
•
The infusion stops.
Alarms
•
Air in Line
•
Door Open
•
Down Occlusion
•
Pump Unattended
•
Low Battery
•
End Battery
•
Internal Malfunction
•
End Program
•
Missing Key
The “Troubleshooting Table” (on the next pages) describes the pump’s alarm conditions and remedies.
Page 56
VET-PRO Operator’s Manual
56
Troubleshooting
Description Result Possible Cause Remedy
Air in Line
Air in Line
Clean air
Air is present in administration set
Disconnect patient, press the STOP/NO key and remove the air from Administration set, as described on page 17. After flushing, press the START/OK key to resume the
Infusion ceases
The roller clamp above the pump is closed.
Open Roler clamp
Door Open
The door of the pump was
not closed prior to operation
Door Open
Close door
Infusion ceases
T
he door has accidentally
opened during operation.
Close Door
Down stream occlusion
Roller clamp closed
Open the Roller clamp
Set kinked
Straight out the
set
Down Occlusion
Press STOP to mute
Administration set loaded
incorrectly
Reload administration set, as described on page 17
Cannula blocked
Change Cannula
Infusion too cold, affecting the elasticity of the tubing
Bring the tubing up to room temperature >19deg C. Warm the liquid.
Infusion ceases
**User may change pressure setting
Adjust pressure setting, as described on page 26
No key has been pressed for 2 minutes, during setting
Pump Unattended
Press start key
Pump operation has been stopped
N
o key was pressed for 2
minutes.
Press the START/OK key
Page 57
VET-PRO Operator’s Manual
57
Troubleshooting
Description Result Possible Cause Remedy
Message displayed showing ± 30 minutes
left of battery operation
Low Battery
Connect to charger
If pump won’t be connected to mains, battery will be depleted
Message will
be displayed as
long as the internal battery
voltage is less than 7.1v.
Connect to Mains
End Battery
End Battery
Connect to charger
Infusion ceases
Battery is depleted.
The pump cannot be used
before being connected to
mains.
Connect to Mains and wait 2 minutes before operating.
Message displayed in the TPN programs, that the parameters entered are incorrect and cannot be calculated by the pump
Wrong TPN Date
Check Data
Program will not start operating
The set parameters are
impossible to perform.
Volume or time parameters
are incorrect
Check data and change accordingly.
Pump is in lock-mode
Lock out
Program parameters can not be changed
The program is
set in
second-
level locking, which
enables daily operation of
the set program, without the
ability to change parameters.
No action required.
A fatal error is detected
Error XXX
Send for Service
Infusion ceases
The pump will advise the
user to send
the pump for
service.
Restart the pump. If the alarm does not stop pack the pump properly, in its original packaging, and send it for service.
Page 58
VET-PRO Operator’s Manual
58
Troubleshooting
Description Result Possible Cause Remedy
Program completed
End PROGRAM
Press STOP to mute
The program was completed
KVO 5 ml/h
Press STOP to renew
Pump will turn to KVO mode
KVO (Keep Vein Open)
program starts.
Press the STOP/NO key to restart
Show Setting
The program was designed to disable unauthourised user access to technician settings. Access to show setting mode, enable the user to
inspect all settings, but not change them.
Press info key twice.
Volume Infused
=>Show Setting
Choose "Show Setting" to display all software adjustable parameters.
Adjust setting
Show Setting:
=>Battery Level
Scroll down the followung parameters on the Show Setting mode:
Lock Out
Show Setting Options
Screen no. Setting Screen
1 Battery Level (Empty ? Full) Battery Level 2 Lock Out Option (ON/OFF) Lock Out 3 Time & Date display Time, Date 4 Tube Temperature Tube Temperature 5 Buzzer Level Buzzer Level 6 Pressure level
(High/Normal/Low)
Pressure level
7 Pressure Default Pressure Default 8 Select Program Select Program 9 Display Flushing Volume (4 ml. Flushing Volume
Page 59
VET-PRO Operator’s Manual
59
Is the default- up to 20 ml)
10 Showing Flushing Rate
(up to 1200 ml/hr)
Flushing Rate
11 Display BOLUS rate Bolus Rate
Maximal Rate setting Maximal Rate Display KVO rate KVO Rate Air bubbles size Air Size Store last program Store last Program Operation led option (on/off) Operation Led
Time option (on/off) Time Option 18 Delay before start option (on/off) Delay Before Start 19 Language Language 20 Pump Serial Number Serial Number 21 Production Date Production Date 22 Operation time from service
(hours)
Operation Time
23 Software Version Software Version 24 Volume Calibration Volume Calibration 25 Pressure calibration - delta Pressure Delta 26 Pressure calibration CAP Pressure CAP 27 EXIT EXIT
Adjust Setting
Press info key twice.
Volume Infused Show Setting
Choose "Adjust Setting" to display all software adjustable parameters.
=>Adjust Setting
Confirm with START/OK key.
Adjust Setting:
=> Pressure Level
See list of parameters that can be set on "Adjust Setting" selection on the next page.
Buzzer Level
The first 2 parameters, Pressure level and Buzzer level, can be adjusted using the arrow keys- no secret code required.
Buzzer level
=> More …
Access to “More…” will required a code. Use ARROW keys to scroll to "More".
Show Setting
Page 60
VET-PRO Operator’s Manual
60
Confirm with START/OK key.
Technician
Set technician Code (Code ), using the numeric keys and confirm with START/OK key.
Code
** For program locking (lock level 2) – set code 902
Scroll through the setting parameters, using the ARROW keys and change them if necessary.
Adjust Setting Options
Screen no. Adjust Screen
A Lock Out Option (Code 902 ) Lock Out B Buzzer Level (no code required) Buzzer Level (1 to 7) C Pressure level (no code required) Pressure level (high,
normal, low)
Use code 901 to enter
1 Restart Pump Restart Pump 2 Pressure Default Pressure Default 3 Select Program Select Program
Setting Flushing volume
(up to 20 ml)
Flushing Volume
5 Setting Flushing Rate
(up to 1200 ml/hr)
Flushing Rate
6 Setting BOLUS rate (different
from default)
Bolus Rate
7 Maximal Rate setting Maximal Rate 8 Setting KVO rate KVO Rate 9 Air bubbles size Air Size
11 Store last program Store Last Program
2 Setting Operation led (on/off) Operation Led 3 Time Option On/ Off Time Option 4 Time Delay Option On/Off Delay Before Start 5 Language Language
16 SetTime & Date Set Time & Date
7 View History Events History
Loading...