Please read all instructions carefully before use. Failure to understand and
follow instructions and cautions may lead to damage to your equipment
This manual contains important operating and safety information.
MicroSonde MPL7-33
The serial number of your device is indicated on your MicroSonde and also on the
storage case. The serial number may wear off, enter the serial number below and
always refer to this information when you need to contact the manufacturer.
VERSION NUMBER X00-B
SERIAL NUMBER ___________________________
IMPORTANT SAFETY INFORMATION
The following symbol is used throughout this manual mark important instructions.
Please take care when meeting this symbol:
GENERAL INFORM A T I O N
The MicroSonde MPL7-33 is small battery operated transmitter that emits a 33 kHz
signal, which can be located by a receiver operating on the same frequency. The
Sonde is intended to be used in non-conductive ducts, cavities and structures. The
device is intended for professional use.
ITEMS SUPPLIED
1 MicroSonde Transmitter MPL7-33
2 Lithium Battery BR535
1 Case with Quick Reference Guide SB7
BATTERY INSTA LLA T I O N , T U R N I N G O N A N D OFF
Figure 1: Inserting battery
Insert a BR535 battery. Observe polarity (Figure 1). To turn the MPL7-33 on, tighten
the battery compartment. The LED at the tip of the MicroSonde will start flashing.
Take care to fully tighten to avoid water getting into the device. To turn the Sonde
off, unscrew battery compartment. Always remove the battery completely when
you are not using the Sonde.
When the battery has insufficient capacity left for carrying out measurements, the
LED will turn off. Replace the battery. Low temperatures reduce the operating time
of the battery.
1 (6)
MicroSonde MPL7-33
High mechanical stress, high ambient temperatures or immersion into liquids
can cause leakage, fire or explosions of the batteries.
Precautions: Protect the batteries from mechanical influences and high ambient
temperatures. Do not drop or immerse batteries into liquids.
OPERATING THE DE V I C E
When the battery is inserted and the battery compartment attached, the
MicroSonde will start emitting a 33 kHz Signal. The location of the MPL7-33 can now
be detected with a receiver that operates at the same frequency, such as the
Vesala CL43. Before use, ensure the MicroSonde is fully functional by locating the
signal with a corresponding receiver. Refer to the user manual of the receiver on
pinpointing the location of the MicroSonde.
You can attach the MicroSonde to a duct rod, cable rodder, snake rod, push or
pull rod using the female M6 thread at the end of the battery compartment. An Oring provides added grip when tightening.
When using the device for blowing with compressed air, always blow with battery
compartment end first. Refer to the user manual of the blowing unit and follow safe
work practices.
Use of compressed air can lead to the MicroSonde exiting the duct at great
speed.
Precaution: Always use a catcher at the end of the Duct. If changing the direction
of the blowing due to the device getting stuck, remember to move the catcher to
the other end of the duct. Do not blow the MicroSonde with pressure exceeding 6
bars. Always use safety goggles, when using the MPL7-33 with compressed air.
MAINTENANCE
CLEANING
Clean with clean soft cloth. Do not use abrasive agents or solvents for cleaning.
Take care to avoid dirt getting into the unit. Regularly inspect O-ring and replace if
damaged. If water gets into the battery compartment, allow it to dry at room
temperature.
STORAGE
Store inside storage box. Respect the temperature limits stated in the technical
datawhen storing the device. Always remove the batteries from the product
during storage to avoid leakage. After a long time of storage, check the function
of the device.
2 (6)
MicroSonde MPL7-33
TROUBLESHOO TIN G
Problem Possible cause Remedy
MicroSonde does not
turn on.
MicroSonde
unintentionally turns off.
Battery is weak or
wrongly placed.
MicroSonde is damaged
Battery is weak. Replace battery
Replace battery, ensure
polarity.
There are no serviceable
parts in the device.
Contact your distributor
or manufacturer.
No signal is detected. Battery is weak.
Device is used in
conductive duct.
Replace battery.
DISPOSAL
This product is an electric device and must be collected separately
for disposal according to the European Directive on waste electrical
and electronic equipment.
For business users in the European Union:
If you wish to discard electrical and electronic equipment, please contact your
supplier or manufacturer for further information.
Information on Disposal in other Countries outside the European Union:
This symbol is only valid in the European Union. If you wish to discard this product
please contact your local authorities or dealer and ask for the correct method of
disposal.
3 (6)
MicroSonde MPL7-33
TECHNICAL SPE CI F I C A T I O N S
Dimensions; Length x Diameter 114 mm x 7.5 mm
Weight including BR535 Battery 15 g
Typical Range in air* 5.1 m
Operating Frequency 32.768 kHz
Frequency Tolerance ± 3 Hz
Transmitting Power <10 mW
Signal Type Continuous
Battery Type 1 x BR535
Typical operating time 6 hrs at 20°C
Typical power consumption 8 mA at 3.0 V
Operating Temperature -10° C to +50° C
Storage Temperature -20° C to +60° C
Protection IP67 IK07
Max. ambient pressure 6 bar
Mounting Thread M6 Female
Max. Speed 10 m/s
Max. Speed w. Shock Absorber 20 m/s
*will vary with choice of receiver, battery voltage, ambient noise and interference