3
Note: More detail on this experiment can be found in Experiment 1, “Using a Gas
Chromatograph: Identifying Unknown Compounds,” in the accompanying lab
manual, Gas Chromatography Investigations with the Mini GC™.
Important: The glass syringe is fragile and can be easily damaged. Be careful not to
bend the needle or bend the plunger. If the plunger is accidentally pulled out of the
glass barrel, reinserting it is extremely difficult, sometimes impossible.
3. Prepare the Vernier Mini GC for data collection.
a. Turn on the Mini GC, using its on-off switch on the left side (see Figure 5).
b. Connect the USB cable of the Mini GC to the USB port on your computer
or LabQuest.
c. Start the data-collection program, and then choose New from the File menu
(to ensure software settings are set to default values).
d. Click Collect in Logger Pro or tap ► in LabQuest 2 or LabQuest, to bring
up the Temperature-Pressure profile. This screen will look something like
this:
Figure 1 Temperature-Pressure
e. Set the Temperature-Pressure values to:
Start temperature
35C
Hold time 2 min
Ramp rate
5C/min
Final temperature
55C
Hold time 9 min
Total length 15.0 min
Pressure1 5.0 kPa
f. Select Done to initiate the Mini GC warm up. Note: A new message will
appear, “Do not inject until GC is ready,” and the LED on the Mini GC will
be red. The Mini GC will take a few minutes to warm up and stabilize.
When the Mini GC is ready for injection in Step 7, the message will read,
1
Pressure values entered here represent the pressure above ambient air pressure.
4
“Inject and select Collect simultaneously”, and the LED will turn to green.
Continue with Step 4 during warm up.
4. Follow the steps below to clean and flush the syringe with acetone. Important:
The glass syringe is fragile. Be careful not to bend the needle or bend the
plunger. Never pull the plunger back more than 50% of its total volume. Be
careful not to bend the plunger as you press it down.
a. Depress the plunger fully.
b. Submerge the tip of the syringe needle into the vial of acetone.
c. Pull back the plunger to fill the barrel about 1/3 full of acetone.
d. Expel the liquid onto a Kimwipe
or a paper towel.
e. Repeat Steps a–d at least two times, until you are comfortable pulling up a
liquid into the syringe and measuring the volume in the syringe barrel. Use
a Kimwipe or a paper towel to carefully pat around the tip of the syringe
needle.
5. Follow the Step 4 process to clean and flush the syringe with 2-butanone (or
another available ketone), the first ketone sample to be tested.
6. Collect a volume of 2-butanone for injection.
a. Submerge the needle into the vial of
2-butanone one last time.
b. Draw up approximately 0.2 L of liquid. It is not
critical that the volume be exactly 0.2 L; a tiny bit
more or less volume is all right.
c. After collecting your sample, gently wipe
the needle from barrel to tip, with a
Kimwipe.
7. Prepare for injection and the start of data
collection. It is important for you and your lab
partner to divide the tasks in this step. One
person will operate the syringe and the other
person will operate the computer controls.
a. When the Mini GC has reached the correct
start temperature and pressure, the message
reads, “Inject and select Collect
simultaneously,” and the LED on the Mini
GC is green.
b. To insert the needle of the syringe into the
injection port of the Mini GC, hold the syringe with one hand and steady
the needle with your other hand.
Important: Supporting the needle guard and the lower part of the needle
with two fingers, as shown in Figure 3, can prevent the possible bending of
the needle! Insert the needle into the injection port until the needle stop is
fully seated. If the needle sticks, rotate it slightly while inserting. Do not
move the plunger yet.
Figure 3
Figure 2