VERITAS T-302, T-602, T-1002, T-2001, T-3001 Instruction Manual

...
SPECIFICATIONS
Model number Capacity Division
T 302 300g 0.01g
T -602 600g 0.01g T -1002 1000g 0.01g T -2001 2000g 0.1g T -3001 3000g 0.1g T -5201 5200g 0.1g
Platter 136x140mm
Net/gross weight 3.5lb/4.5lb
Package
Standard carton: 33.5×23×14(cm
3
)
6 Units in one box: 48.5×39.5×46(cm
3
)
Operating Temperature
0-40(32-104)
Power source
6XAAor AC/DC
Adapter 12V/500mA
ERROR CODE
COMPACT SCALE
/BALANCE
INSTRUCTION MANUAL
Error Code
Reason Trouble Shooting
Err-O Overload Use within capacity. Err-Z 1).Exceeded the zero
tracking range when powering on.
2).Forgot to install pan when performing linearity calibration.
1). Pan should be empty on power up. Remove any weight from pan.
2). Install pan and then recalibrate scale linearity calibration.
Err-S Inner code unstable
when powering on.
1) Make sure the scale is stable when powering on.
2) Low battery , ch a ng e or cha rg e ba tter y.
3) Allow for longer warm up period.
Err-L Calibrating without a
weight or the weight is not the correct value.
Place a calibration weight of the value shown on the display and then press calibration button.
Err-P Counting setting error. Place the correct size parts sample when
setting up counting mode.
SAFETY PRECAUTIONS
All safety messages are identified by the following the words "WARNING" and "CAUTION". These words mean the following:
WARNING
Important information to alert you to a situation that might cause serious injury and damage to your property if instructions are not followed.
CAUTION
Important information that tells how to prevent damage to equipment.
When using the scale/balance, the following safety precautions should always be
followed.
WARNING
Use only the AC adaptor with the scale/balance. Other adaptor may cause damage. (AC adaptor is optional)
CAUTION
Avoid installing the scale/balance in direct sunlight, this may cause
discoloration or malfunctions.
Keep scale at room temperature, away from heat sources.
Avoid shock impact and loading over capacity. Do not keep heavy object
permanently on the platform.
If scale is not used for an extended period, keep it clean and stored in
plastic package.
Matter charged by static electricity will influence the weighing.
1
CALIBRATION
When to calibrate
1. Press [ON/OFF] to turn the power on for 15 minutes for warm up.
2. With scale turned off, press [ON/OFF] to turn the power on.
3. Display will show flashing digits and then “--------”,
4. Press [ZERO], the display will show “CAL”.
5. Press [ZERO] again , the display will show “X0000 g” where X is the flashing digit,
6. Press [UNIT] key to move the flashing digit to the right.
7. Press [PCS] key to increase the flashing digit. (X is the calibration weight which can be set according to your requirement – we recommend a minimum of 50% balance capacity).
8. Press the [ZERO] key, the display will show a series of digits known as the A/D value, these numbers will continuously move, once the display shows the stable indicator (please see page 16 for help locat ing the stable indicator), press the [UNIT] key, the display will then show --------, then “X00.00” or “X000.0” depending on the model and capacity.
9. Place the calibration weight on the center of the platform.
10. Press [ZERO] key to return to weighing mode.
11. Now the Calibration is completed.
12. Press [ON/OFF] to turn off the power.
13. Press [ON/OFF] to turn on the power again.
14. Place the weight on the platform to make sure weighing is correct, if not, repeat above steps.