Veris Industries PW SERIES Data Sheet

Pressure Monitoring | Wet Media Differential Pressure Transducer
PW SERIES
Jumper-Selectable Port Swap Feature
PW
The PW Series wet pressure transducers incorporate microprocessor profiled sensors for exceptional accuracy and reliability. Easy to use and designed to provide exceptional installation savings, the PW Series is ideal for measuring pressure across pumps, filters, heat exchangers, compressors, and other non-corrosive wet media applications.
The jumper-selectable port swap feature eliminates costly replumbing when the high and low ports are improperly plumbed, allowing the jumper position to be changed from normal to swap.
SPECIFICATIONS
GENERAL
Input Power Class 2; 12 to 30 Vdc or 24 Vac nominal, 50/60 Hz
Max. Current Draw DC: 125 mA; AC: 280 mA
Output 3-wire transmitter; user selectable 4 to 20 mA
(clipped & capped)/0-5 V/0-10 V*
Surge Damping Electronic; 5-second averaging
Test Mode Overrides output to full-scale (20 mA, 5 V, 10 V)
Zero Adjust Pushbutton auto-zero & digital input (2-pos
Status Indication Dual-color LED:
Green = Normal, Green Blinking = Low > High, Red = Overrange, Red Blinking = Overpressure
Housing Material White powder-coated aluminum
Fittings psig: 1/8” NPT female thread, 17 to 4 PH
stainless; barg: 1/8” BSPT female thread, 17 to 4 PH stainless
PRESSURE RANGES (SELECTABLE)
0 to 50 psig (Gauge) 0 to 5/10/25/50 psid (Differential)
0 to 100 psig (Gauge) 0 to 10/20/50/100 psid (Differential)
0 to 250 psig (Gauge) 0 to 25/50/125/250 psid (Differential)
0 to 3.5 barg (Gauge) 0.35/0.7/1.75/3.5 bard (Differential)
0 to 7.0 barg (Gauge) 0.7/1.4/3.5/7.0 bard (Differential)
0 to 17.0 barg (Gauge) 1.7/3.4/8.5/17.0 bard (Differential)
SENSOR
Accuracy @ 25 °C** Range A, B, C: ±1% F.S.; Range D: ±2% F.S.***
Long Term Stability ±0.25% per year
Media Compatibility Media compatible with 17 to 4 PH stainless steel
Proof Pressure Max. 2x F.S. range
Burst Pressure Max. 5x F.S. range
Jumper-selectable
The jumper-selectable output switch for normal (4 to 20 mA) or reverse (20 to 4 mA) operation
Switch-selectable
Switch-selectable pressure ranges
fewer models to order
and stock
provides application flexibility
Rugged
Rugged, die-cast enclosure provides NEMA 4 sealing
Zero calibration
Pushbutton and remote zero adjustment
maintain accuracy
and reduce callbacks with automatic zero calibration
High stability
Jumper-controlled electronic surge dampening for high stability
APPLICATIONS
• Monitoring and controlling pump differential pressure
• Chiller/boiler differential pressure drop
Temperature Compensated Range
Media Temp Limits -20 to 85 °C (-4 to 185 °F); 0 to 90% RH
Product Operating Environment
WARRANTY
Limited Warranty 5 years
AGENCY APPROVALS
Available
*Minimum input voltage for 4 to 20 mA operation: 250 Ω loop (1 to 5 V) = 12 Vdc; 500 Ω loop (2 to 10 V) = 15 Vdc; Minimum input voltage for volt operation: 0 to 5 Vdc output = 12 Vdc; 0 to 10 Vdc output = 15 Vdc. **Accuracy combines linearity, hysteresis, and repeatability. *** FS is defined as full span of selected range in bi-directional mode. EMC Conformance - CE option: Low voltage directive 2014/35/EU; EMC directive 2014/30/EU. EMC Special Note: Connect this product to a DC distribution network or an AC/DC power adaptor with proper surge protec tion (EN 61000-6-1 specification requirements).
† The CE mark indicates RoHS2 compliance. Please refer to the CE Declaration of Conformity for additional details.
0 to 50 °C (32 to 122 °F); TC Zero <±1.5% of product F.S. per sensor ; TC Span<±1.5% of product F.S. per sensor, (2 sensors per unit)
non-condensing
-10 to 55 °C (14 to 130 °F); 0 to 90% RH non-condensing
• CW/HW system differential pressure
TM
HQ0001831.G 0117
WIRING DIAGRAM
POWER SOURCE
12 to 30 Vdc/24 Vac
DIGITAL CONTROL
Analog Input
0-5 V/0-10 V
or 4 to 20 mA
Optional
Digital Output
DIMENSIONAL DRAWING
POWER SIGNAL
COM
WET DIFFERENTIAL PRESSURE
PW SERIES
REMOTE ZERO (Dry Contact)
2.2"
(57 mm)
4"
(102 mm)
ZERO
Range
JP8
A
JP7
B
JP6
C
JP3
D
JP2 JP1
Bidirectional Operation
Input Conditions Result Outputs Read
HI PORT
100 psi 100 psi 50 psi 50 psi
0 psi
17.0 bar
17.0 bar
8.5 bar
8.5 bar 0 bar
Model PW-03 PW-04 PW-05
Model PW-06 PW-07 PW-08
Analog Reverse/Normal
Port Swap/Normal
Bidirectional/Normal
Fast/Slow Surge Damping
5 V/10 V Output
mA/Volts Output
LO PORT
0 psi 50 psi 50 psi 100 psi 100 psi
0 bar
8.5 bar
8.5 bar
17.0 bar
17.0 bar
A
50
100 250
A
3.5
7.0
17.0
Pressure Monitoring | PW Series
/
Output is either mA or V
0.7
1.4
3.4
C 10 20 50
C
0.35
0.7
1.7
4-20mA 0-10V
20mA 16mA 12mA 8mA 4mA
20mA 16mA 12mA 8mA 4mA
Use the Range switch to
select F.S. differeintial
pressure.
D 5
e.g. PW-04
10 25
D
e.g. PW-08
10V
7.5V 5V
2.5V 0V
10V
7.5V 5V
2.5V 0V
DP +100 psi +50 psi 0 psi
-50 psi
-100 psi
+17.0 bar +8.5 bar
0 bar
-8.5 bar
-17.0 bar
Range (psi)
B 25 50
125
Range (bar)
B
1.75
3.5
8.5
e.g. PW-04
e.g. PW-08
5.8"
LOW
HIGH
(147 mm)
ORDERING INFORMATION
Local Display
PW
L = LCD Display X = No Display
Example:
XL 04 CPW
* Select operational range according to maximum gauge pressure, NOT differential
pressure.
Example: High gauge pressure=90 psig, Select 100 psig model (04). **Barg models use BSPT threads on sensor fittings ***Not available with barg units
HQ0001831.G 0117
NIST
N = NIST X = None
Operational Range*
03 = 0-50 psi g 04 = 0-100 psig 05 = 0-250 psig 06 = 0-3. 5 barg** 07 = 0-7.0 barg** 08 = 0-17 barg**
US or EU
S = Standard*** C = CE
TM
MICROPROCESSOR PROVIDES DIGITAL SIGNAL CONDITIONING
• Noise rejection reduces fluctuating readings due to noise or turbulence
F.S.
Noise Rejection
PRESSURE
0
Actual Pressure Digitally Conditioned Output
• Surge damping prevents false alarms by averaging fast peaks
Output Clamping
Surge Damping
TIME
Loading...