Veris PW Datasheet

PW SERIES
Wet Media Differential Pressure Transducer
Jumper-Selectable Port Swap Feature
VERIS INDUSTRIES
PW
DESCRIPTION
The PW Series wet/wet pressure transducers incorporate microprocessor proled sensors for exceptional accuracy and reliability. Easy to use and designed to provide exceptional installation savings, the PW Series is ideal for measuring pressure across pumps, lters, heat exchangers, compressors, and other non-corrosive wet media applications.
Output Clamping
F.S.
Noise Rejection
PRESSURE
Surge Damping
PRESSURE
Microproce ssor provides digita l signal condition ing
Noise rejection reduces uctuating readings due to noise or turbulence
Surge damping prevents false alarms by averaging fast peaks
0
Actual Pressure Digitally Conditioned Output
SPECIFICATIONS
Input Power 12 to 30VDC/24VAC nominal Maximum Current Draw DC: 125mA; AC: 280mA Output 3-wire transmitter; us er selectable 4-20mA (clipped and capped)/0-5V/0-10V† Accurac y @ 25°C* Range A, B, C: ±1% F.S.; Range D: ±2% F.S.** Surge Damping Electronic; 5-second averaging Tes t Mod e Overrides output to full-scale (20mA, 5V, 10V)
Pressure Ranges (Selectable):
0-50 psig 0-5/10/ 25/50 psid 0-100 psig 0-10/20/50/100 psid 0-250 psig 0-25/50/125/250 psid 0-3. 5 barg 0.35/0.7/1.75/3.5 bard 0 -7.0 ba rg 0.7/1.4/3.5/7.0 bard 0 -17.0 ba rg 1.7/3.4/8.5/17.0 bard Product Operating Environment -10° to 55°C (14° to 130°F); 0 to 90% RH noncondensing Long Term Stability ±0.25% pe r year Zero Adjust Pushbutton auto-zero and digital input (2-pos terminal block) Status Indication Dual-color LED: Green = Normal, Green Blink ing = Low > High, Red = Overrange, Red Blinking = Over pressure Housing Material White powder-coated aluminum
Sensor:
Media Compatibility Media compatible with 17-4 PH stainless steel Proof Pressure Max. 2x F.S. range Burst Pressure Max. 5x F.S. range Temperature Compensated Range 0° to 50°C (32° to 122°F); TC Zero <±1.5% of product F.S. per sensor ; TC Span<±1.5% of product F.S. per sensor, (2 sensors p er unit) Media Temperature Limits -20° to 85°C (-4° to 185°F); 0 to 90% RH non-condensing Fittin gs psig: 1/8" NPT female thread, 17-4 PH stainless; barg: 1/8” BSPT female thread, 17-4 PH stainless
†Minimum input voltage for 4-20mA operation: 250 loop (1-5V) = 12VDC; 500 loop (2-10V) = 15VDC; Minimum input voltage for volt operation: 0-5VDC output = 12VDC; 0-10VDC output = 15VDC. *Accuracy combines linearity, hysteresis, and repeatability. **FS is dened as full span of selected range in bi-directional mode. EMC Conformance: Low voltage directive 2006/95/EC; EMC directive 2004/108/EC. EMC Special Note: Connect this product to a DC distribution network or an AC/DC power adaptor with proper surge protection (EN 61000-6-1:2007 specication requirements).
800.354.8556
TIME
Year
5
Warranty
+1 503.598.4564
APPLICATIONS
Monitoring and controlling pump dierential pressure
Chiller/boiler dierential pressure drop
FEATURES
The jumper-selectable output switch for normal (4-20mA) or reverse
(20-4mA) operation provides application exibility
Rugged, die-cast enclosure provides NEMA 4 sealing
Jumper-selectable port swap feature eliminates costly replumbing when the
high and low ports are improperly plumbed...change the jumper position from normal to swap – problem solved!
Switch-selectable pressure ranges...fewer models to order and stock
Pushbutton and remote zero adjustment...maintain accuracy and prevent
callbacks with automatic zero calibration
Jumper-controlled electronic surge dampening for high stability
Pushbutton zero calibration – no trim pots to adjust
www.veris.com
HQ0001831.B 01131
800.354.8556
+1 503.598.4564
APPLICATION/WIRING DIAGRAM
www.veris.com
POWER SOURCE
12 to 30 VDC/24 VAC
DIGITAL CONTROL
Analog Input
0-5V/0-10V
or 4-20mA
Optional
Digital Output
POWER SIGNAL
COM
DIMENSIONAL DRAWING
5.8"
(147 mm)
LOW
WET DIFFERENTIAL PRESSURE
HIGH
PW SERIES
REMOTE ZERO (Dry Contact)
(57 mm)
(102 mm)
2.2"
ZERO
Range A B C D
JP8 JP7 JP6 JP3 JP2 JP1
/
Analog Reverse/Normal
Port Swap/Normal
Bidirectional/Normal
Fast/Slow Surge Damping
5V/10V Output
mA/Volts Output
Output is either mA or V
Bidirectional Operation
Input Conditions Result Outputs Read
HI PORT
100 psi 100 psi 50 psi 50 psi
0 psi
17.0 bar
17.0 bar
8.5 bar
8.5 bar 0 bar
LO PORT
0 psi 50 psi 50 psi 100 psi 100 psi
0 bar
8.5 bar
8.5 bar
17.0 bar
17.0 bar
DP +100 psi +50 psi 0 psi
-50 psi
-100 psi
+17.0 bar +8.5 bar
0 bar
-8.5 bar
-17.0 bar
4-20mA 0-10V
20mA 16mA 12mA 8mA 4mA
20mA 16mA 12mA 8mA 4mA
Use the Range switch to
select F.S. diereintial
10V
7.5V 5V
2.5V 0V
10V
7.5V 5V
2.5V 0V
pressure.
e.g. PW-04
e.g. PW-08
Range (psi)
A
B
C
Model
50
PW-03
4"
PW-04 PW-05
100 250
25 50
125
D
10
5
20
10
50
25
e.g. PW-04
Range (bar)
A
B
C
0.7
1.4
3.4
D
0.35
0.7
1.7
e.g. PW-08
pressure
Model PW-06 PW-07 PW-08
3.5
7.0
17.0
1.75
3.5
8.5
ORDERING INFORMATION
Local Display
NIST
Operational Range
PW
L = LCD Display X = No Display
1
Select operational range according to maximum gauge pressure, NOT differential pressure.
Example: High gauge pressure=90 psig, Select 100 psig model (04).
2
BARG models use BSPT threads on sensor ttings.
3
Not available with barg units
HQ0001831.B 01131
N = NIST X = None
03 = 0-50 psig 04 = 0-100 psig 05 = 0-250 psig 06 = 0-3.5 barg 07 = 0-7.0 barg 08 = 0-17 barg
1
US or EU
S = Standard C = CE
2
2
2
PW L X 04 C
3
Examp le:
ACCESSORIES
Bypass Valve assemblies (AA14A) PW installed on bypass valve manifold (AA16A) Snubbers (AA11, AA12), Steam siphon (AA13)
AA11
AA13
AA14A