Veris Precision Rate Controller Operating Instructions

Pub.# OM17-PRC
Operating Instruction Manual
Precision Rate Controller
601 N. Broadway
Salina KS 67401
(785) 825-1978
www.veristech.com
Pub.# OM17-PRC
WARNING !
Return all hydraulic valves to neutral position before exiting tractor cab. Inadvertent drive rotation could cause serious injury.
WARNING!
Escaping fluid under pressure can have sufficient pressure to penetrate the skin. Check all hydraulic lines and fittings before applying pressure. Fluid escaping from a very small hole can be almost invisible. Use paper or cardboard, not body parts, and wear heavy gloves to check for suspected leaks. If injured, seek medical assistance from a doctor that is familiar with this type of injury. Foreign fluids in the tissue must be surgically removed within a few hours or gangrene will result.
NOTICE !
Do not weld on drill unless electronic components are removed.
Table of Contents
Operational RequirementsHydraulic and Electrical 1-1 Tractor Hookup 1-2 Console Functions 1-3
Rate Calibration 2-1
Speed Calibration 3-1
Operations 4-1
Varying Rates with Pre-set Function 5-1 Varying Rates with GPS-based maps 6-1 Maintenance 7-1
Troubleshooting 8-1
Important Information
!
Operating Instructions
Drive Operational Requirements:
Hydraulic System:
Closed center, pressure compensated or load sensed systems only Drive will not operate on open-centered
hydraulic systems Minimum Hyd. Pressure: 2250 psi Maximum Hyd. Pressure: 3000 psi Maximum Required Flow: 8.5 gpm Electrical System: Voltage: 12 volt DC
Amperage: 4 amperes
Tractor Hookup
Hydraulics:
1) Connect pressure hose (P) to retraction outlet
2) Connect motor return hose (T) to motor return port (if available) or to
extension outlet.
3) Set flow rate at maximum.
4) If tractor is equipped with electro-hydraulic valves, set timer to “constant” flow.
Electrical:
1) Power can be connected in two different manners:
a) Power port adaptor (PN 19676)
1-1
Pub.# OM17-PRC
1-2
Connect to power port Connect to console
Connect to power wire from implement
Figure 1 b) To battery. Make sure that eyelets are properly connected (red to positive,
Black to negative). Connect female socket to power port adapter.
Figure 2
Pub.# OM17-PRC
1-3
On/VR keyturns drive system on. Press twice for VR mode. Note: Engage key must also be pressed to start drive operation.
Function key: used in calibration mode and to program Console for VR recipes.
Up/down arrow keys: used to change rates manually, to set calibration numbers, and VR controller options.
Engage key: press to start drive. Note: drive will not run unless light above key is illuminated
Off key: used to shut off Console.
Console Functions
Shown below in Figure 3 is the Precision Population Controller Console. The uses of each key are as follows:
Figure 3. Console Set to Main Operating Screen, in manual mode
Pub.# OM17-PRC
2-1
Calibration--Rate
Turn on Controller Console by pressing On/VR key. Green light above On/VR key will illuminate when power is on. Adjust desired planting rate by using the Up/Down arrow keys to change the Set rate. (See Figure 4)
Figure 4. Set the desired Rate by using the Up/Down arrow keys. Press Function key until Calibration Number window appears. (See Figure 5)
Select drive calibration number from chart if available. If calibration number is not known, enter 1000. Use Up/Down arrow keys to set Calibration number. Press Function key until the Calibration window appears as shown in Figure 5.
Figure 5. Example of calibration number. Enter new calibration number using Up/Down arrow keys. Press Function key to advance to Calibration Mode. Figure 6. Rate calibration window. Press up arrow key to enter calibration mode.
Pub.# OM17-PRC
2-2
Figure 7. This is the first window that appears in Calibration mode. Press Function key to accept, or Up key to change to metric settings.
Figure 8. Enter implement width here using Up/Down arrow keys. Press Function key to advance to next window. Figure 9. Enter number of rows here using Up/Down arrow keys. Press Function key to advance to next window.
Figure 10. Enter your planned planting speed using Up/Down arrow keys. Calibration mode will not accept a speed higher than 10 mph. Press Function key to advance to next window. Figure 11. Enter the number of seeds you want to count. For singulated metering, at least 100 seeds should be selected. Press Function key to advance to next window.
Pub.# OM17-PRC
2-3
Figure 12. The cab console is informing you the length of time that the drive will be operating, in order to meter the amount of seed you have requested, at the calibration number, planter or application width, and number of rows you have selected. If the screen reads TIME TOO LOW, you will need to increase the amount of seed that you will count or measure. Press Function key to advance to next window.
Figure 13. WARNING: The drive is about to operate, meaning there is a
danger of entanglement if anyone is in the drive area. Be sure to verify that no one is in or near the drive area before advancing to the next step.
Figure 14. Only after you have verified the drive area is clear, press the up arrow
key to initiate calibration mode. If drive area isn’t clear, press Down arrow key to
exit calibration mode. Figure 15. Continuing to verify that the drive area is clear, press the Engage key to start drive rotation. Pressing Function key will exit calibration mode.
Pub.# OM17-PRC
2-4
Figure 16. While the drive is rotating, the display window shows the time
remaining and the “Out” rate. CONTINUE TO MONITOR DRIVE AREA DURING
DRIVE ROTATION. PRESS THE ENGAGE KEY (OR OFF KEY) TO STOP DRIVE DURING CALIBRATION
Figure 17. After Calibration meter rotation has ended, the screen above left will appear. Enter the actual amount metered using the Up/Down arrow keys as shown in Figure 18, above right. Press Function key to advance to next window.
Figure 19. A New calibration number is suggested, along with the Old calibration number. Press Function key to advance to window shown in Figure 20. Figure 20. If you wish to keep the Old number, perhaps to re-run the calibration procedure, press the Down arrow key. If you want to accept the New calibration number, press the Up arrow key.
Loading...
+ 18 hidden pages