Veris pH Manager Reference Guide

Field Reference GuidepH Manager
Keep with pH System and refer to daily.
(Refer to Operating Instruction Manual for complete instructions)
PRIOR TO MAPPING: PERFORM ROUTINE DAILY MAINTENANCE
3. In Manual Mode test each function
(washing, raising and lowering):
1. Inspect unit for loose or missing bolts, other obvious problems
4. Raise and lower sampling shoecheck scraper for appropriate contact:
5. Inspect prox sensor light during manual cycling:
Adjust hydraulic flow so cycling time is 1.5-2 seconds
6. Raise sampling shoe—check for ¼” to ½” clearance between electrodes and trough; make sure this is where
the shoe will stop in automatic mode (see prox sensor
adjustment section in Operating Instructions)
7. Operate wash jets and check alignment; does water bubble out of top when electrodes are removed?
9. Calibrate electrodes
8. Raise unit and drive slowly in semi-circle with unit cycling to check automated functions
Insert electrodes into holder and finger­tighten set screws. Follow menu
prompts to “pH Setup”. The instrument
will prompt for the electrodes to be inserted into pH buffer 4 solution; Slide pH 4 buffer solution container onto electrode holder. Press 1 to continue with calibration or 2 to exit. Repeat with pH 7 buffer solution.
Loading...
+ 1 hidden pages