Veridian Healthcare 11-100, 11-110 Instruction Manual

VH Suction Pump
Tabletop Aspirator System
Model 11-100 Model 11-110
INSTRUCTION MANUAL • ENGLISH & ESPAÑOL
READ THIS INSTRUCTION MANUAL CAREFULLY BEFORE USE
TAB LE O F C ONT ENTS
Introduction ..................................................................................................................4
Care and Safety Information.....................................................................................5
Product Features..........................................................................................................6
Set Up and Operation .............................................................................................7-8
Battery/Power Operation...........................................................................................9
Cleaning and Disinfection.................................................................................10-11
Changing the Filter ...................................................................................................12
Troubleshooting........................................................................................................13
Warnings and Cautions............................................................................................13
Product Specications .............................................................................................14
Warranty......................................................................................................................15
Instrucciones en Español...................................................................................17-32
Toll-Free Customer Care Help Line: 1-866-326-1313
Monday – Friday 8:30 a.m. – 4:30 p.m. CST
Manufactured for: Veridian Healthcare 1175 Lakeside Drive Gurnee, IL 60031 www.veridianhealthcare.com
Made in China #93-1103 12/15
©2015 Veridian Healthcare, LLC
2 • ENGLISH ENGLISH • 3
STOP! PLEASE ENSURE YOU HAVE ALL OF THE FOLLOWING COMPO NENTS BEFORE USING YOU R DEVICE
Suction Pump Motor Housing
6’ Patient Tubing
(mouthpiece catheter
NOT included)
Collection Jar
with Lid
Air Filter
2x 6” Tube
Instruction Manual Replacement Parts Order Form
IF YOU ARE MISSI NG ANY PARTS, INCLUDING INSERTS OR INSTRUCTION MANUALS, DO NOT RET URN TO PLACE OF PURCHASE. CONTACT CUSTOMER CARE AT 866-326-1313.
2-pc Power Cord
(5) Air Filters
Auto Adapter (11-110 only)
Thank you for purchasing this VH Suction Pump Tabletop Aspirator. Please read this instruction manual completely before using your unit to ensure safe and eective use, as well as long life of the product.
Federal law allows suction pumps to be sold by a physician’s prescription only. These devices should be used only at the direction of a doctor or healthcare professional and should be used only as intended and described in this manual.
INTRODUCTION
Improper care or use of your suction pump may result in injury, damage to the unit or ineective treatment. Following these instructions will ensure the suction pump’s ecacy and long life.
• Never allow the suction pump to be used without adult supervision.
• Do not use the suction pump for anything except patient treatment as directed by a physician. Do not use the suction pump in a manner inconsistent with these instructions.
• The suction pump is an electronic device, keep away from water to avoid electrical shock. Do not immerse the unit in liquid and do not use while bathing. If the unit should fall into water, do not reach for it without rst unplugging the unit.
• Do not use while driving.
• Unplug the unit before cleaning it and after each use. Do not leave the unit plugged in when not in use.
• Use only the included power cord(s).
• Always ensure that the air vents are not blocked or obstructed in any way.
• Dropping the suction pump or causing other heavy impact may damage the unit and aect proper operation.
• Do not expose the unit to direct sunlight, humid environments, extreme temperatures, strong static electricity or electromagnetic waves.
• If the suction pump is not functioning correctly, if it makes unusual sounds, or if during use the patient experiences pain or discomfort, immediately stop use.
• The patient tubing and catheter should be cleaned daily and replaced monthly during regular use or at the prescribing physician’s instructions.
• Do not use the unit if it has been dropped to the ground, exposed to extreme temperatures or high humidity, or damaged in any way.
• Do not disassemble the suction pump or attempt to repair it.
CARE AND SAFETY INFORMATION
4 • ENGLISH ENGLISH • 5
SET UP AND OPERATIONPRODUCT FEATURES
Tubing and Filter Set
Collection Canister Lid
Pressure Gauge Display Dial
Power Switch
Collection Canister
Air Vents
6’ Patient Tubing
6 • ENGLISH ENGLISH • 7
Patient Tube Connection Port
Carrying Handle
Pressure Adjustment Dial
Power Cord Connection (on back of pump)
3. Attach the tubing/lter set to the suction pump body, then securely attach the 6’ length patient tubing to the port on the collection canister lid.
Before the rst use of the suction pump if the unit has not been used for an extended period of time, it should be cleaned thoroughly as described in the Cleaning and Disinfection section of this manual.
Assemble the tubing and lter set:
• Firmly push the 6” tubing onto each end of the lter. The tubing should t tightly onto the lter. If the tubing is not snug, then the set should be replaced.
• Tubing/lter set should be left unassembled when not in use to prevent any moisture build up in the tube or lter.
Assemble the collection jar and suction pump:
1. Replace the lid on the collection canister. The jar lid will not lock or fasten to the canister but should t snugly and securely; take care when transporting liquid.
2. Place the canister into the open ring on the suction pump body
Patient Tube Connection Point
BATTERY / POWER OPERATIONSET UP AND OPERATION
8 • ENGLISH ENGLISH • 9
The Suction Pump Tabletop Aspirator System is oered with two power option models — 2 piece DC Power Adapter; model 11-110 is rechargeable and includes an auto adapter.
Use common sense electric device caution when operating this unit. Always turn o the power and unplug the device before cleaning or if the unit is not working properly.
• The included power cord should be plugged into the back of the suction pump unit.
• Always unplug the power cord from the wall and unit when the suction pump is not in use.
Rechargeable DC Power (11-110)
• This unit can operate by the internal rechargeable battery or under AC power.
• The included 2-piece power cord can be plugged into the unit for use or to charge the device. The auto-adapter is intended for charging the internal battery on the go; do not use this product while driving.
• Assemble the power cord and plug it into the unit prior to plugging into a standard wall socket. Operating the unit while plugged into a wall socket will not aect the charged battery life remaining on the internal battery.
• Charge time from expired battery: 8 hours
• Charged battery life: 60 minutes at normal use; 50 minutes at
maximum vacuum ow
Always inspect the tubing and mouthpiece (not included) for cracks or wear prior to use and replace as necessary.
Patient use:
1. Place the suction unit on a at, stable surface within operating range of the patient and an electrical outlet. Ensure that the air vents on the side of the unit housing are not blocked or obstructed.
2. Plug the power cord into the electrical outlet.
3. Press the power switch to turn the unit on and activate the suction operation.
4. Verify the suction unit is at the desired suction level prior to patient suction by turning the adjustment knob until the required level is reached. The prescribing physician can advise and explain what level is necessary for the specic patient’s care.
NOTE: This device does not include the patient catheter (commonly called suction tip or pipette) as this piece is meant to be replaced frequently. Consult your physician for advice and information on obtaining the patient catheter.
5. When complete, turn the unit o and unplug it from the outlet if it will not be used again for a period of time.
6. Dispose of the contents following the directions of the prescribing physician.
7. Follow the Cleaning and Disinfection directions as advised in this manual.
CLEANING AND DISINFECTION
Disinfection at the End of Day / Treatment Period:
The suction pump should be disinfected once daily, after the last use of the day, unless a dierent disinfection regimen is specied by a healthcare professional. Disinfect the suction pump as follows:
1. Ensure that the power cord is disconnected from the power outlet.
2. Disconnect the air tube, lter, tube set and mouthpiece (not included). Remove the collection canister and dispose of its contents. Do not wash or submerge the suction pump body.
3. Wash all parts in hot water, using a mild, non-abrasive detergent.
4. Submerge all parts in a solution of one part white vinegar and three parts distilled water for a minimum of 30 minutes.
5. Rinse all pieces thoroughly with water.
6. Allow all pieces to air dry completely, or dry them thoroughly with a clean, soft towel. Do not store any parts wet as remaining moisture creates a risk of bacterial growth.
CLEANING AND DISINFECTION
The patient tubing and mouthpiece (not included) should be thoroughly cleaned after each use and disinfected each day when being used frequently unless a dierent cleaning regimen is specied by your healthcare professional; additional hygienic precautions may be required for use with high-risk patients. Always follow the direction of your healthcare provider.
Cleaning After Each Use:
1. Ensure that the power cord is disconnected from the power outlet.
2. Disconnect the air tube, lter, tube set and mouthpiece (not included). Remove the collection canister and dispose of its contents. Do not wash or submerge the suction pump body.
3. Wash and rinse all pieces in hot water, using a mild, non-abrasive detergent.
4. Allow all pieces to air dry completely, or dry them thoroughly with a clean, soft towel. Do not store any parts wet as remaining moisture creates a risk of bacterial growth.
5. Occasionally wipe the body of the unit with a damp cloth and allow it to air dry. Only clean the body of the unit when it is unplugged and completely cool.
Never use soap pads or powder cleanser to clean any part of the unit. Never use any cleaning products that may be harmful when swallowed or inhaled.
It is recommended that air lter and tubing and collection canister be replaced regularly depending on the frequency of therapy treatments or when damaged, cracked or holes appear in the tubing. To order replacement parts, use the included order
form, call Customer Care at 866-326-1313 or contact your local dealer.