These Directions for Use (DFU) are kept up-to-date by VDW GmbH. The most current version can
be accessed at www.vdw-dental.com and is also available from VDW.CONNECT® App. We
strongly recommend that you periodically check this website for updates of this DFU.
The DFU is available in other languages upon request. If a printed version in your language is not
enclosed, VDW GmbH will send a free copy to any address in Europe. To order a free copy of the
DFU, send a request to service@vdw-dental.com.
Some images in the DFU may vary slightly from the actual device. We would appreciate your
feedback. If any errors are noted, please contact VDW GmbH Customer Service.
VDW.CONNECT Locate® Apex Locator supports the dentist in the determination of the working
length during the endodontic treatment. The use of this product is destined exclusively to duly
qualified dental practitioners.
The use of this product by unqualified, or untrained personnel can result in patient injury, personal
injury and / or material damage.
Disposable Barrier Sleeves and Covers are intended to serve as a disposable barrier for dental
instruments and equipment. VDW.CONNECT Locate® Disposable Barrier Sleeve is non-sterile
and intended for single patient use only.
To guarantee the safety, intended functionalities and good maintenance of VDW.CONNECT
Locate® Apex Locator, follow the indications given in the present Directions for Use (DFU). Keep
the DFU updated as per described in the Introduction section. VDW GmbH will not accept any
liability if VDW.CONNECT Locate® Apex Locator is used, stored or manipulated outside of the
indications given in the DFU.
2CONTRAINDICATIONS
Read the following contraindications before use.
•Do not use VDW.CONNECT Locate® Apex Locator on patients wearing active implantable
devices such as a heart pacemaker or defibrillator.
•The use of VDW.CONNECT Locate
to metals. Check Section 8 Technical Characteristics for a list of materials used.
•Do not use VDW.CONNECT Locate
procedure outside endodontics.
•Safety and effectiveness of use have not been established in pregnant or breastfeeding
women or in children.
•The end user of the device should use clinical judgment for its proper use.
•Color-blind users must ensure that they always operate the device with the sound ON, and
that they are familiar with the various sound signals before using the device for the first time.
Check section 6.7 Description of the Sounds before using the device.
•Hearing-impaired users must ensure familiarity with the various visual signals before using
the device for the first time. Check section 6.6 Description of the LED Colors before using
the device.
•The device is not indicated for users both color-blind and hearing-impaired unless they have
these difficulties balanced out.
5
®
Apex Locator is not recommended in patients allergic
®
Apex Locator for implants or any other dental
3WARNINGS
Read the following warnings before use.
VDW GmbH accepts no liability in the event that any of these warnings are disregarded.
Check the outer packaging seal on reception. Do not use the device if the packaging seal has
been broken. Contact customer service.
Report discrepancies between labeling and packaging, or issues with either, to customer service.
Dental Practitioner
•Read the full DFU before using VDW.CONNECT Locate® Apex Locator on a patient.
Operate VDW.CONNECT Locate
safe and effective use.
•Ensure that you have a suitable working distance to see the display from your working
position.
•VDW.CONNECT Locate
- Endodontic treatment.
- Use by trained and qualified professionals only, such as:
- dentists in dental practice, and
- dental assistants for cleaning and preparation of the device only, and in accordance
with this DFU.
Environmental conditions
•Do not use VDW.CONNECT Locate® Apex Locator in the presence of free oxygen,
anesthetics, flammable substances / gases or in a potentially explosive environment.
•VDW.CONNECT Locate® Apex Locator must be operated and stored in a safe, dry and
clean environment.
•Place VDW.CONNECT Locate
the clamp) or firmly attached to the lab coat (when used with the clamp).
•VDW.CONNECT Locate
locations or near liquids.
•Do not expose VDW.CONNECT Locate® Apex Locator to direct or indirect heat sources.
•Avoid collisions and falls. After a fall, please check that VDW.CONNECT Locate
Locator works properly (refer to section 6.9 Testing VDW.CONNECT Locate
Locator Functionality).
•VDW.CONNECT Locate® Apex Locator contains a Bluetooth® module and may cause radio
interference or disrupt the operation of equipment in its vicinity. If this happens, the
interference should be reduced by reorienting or repositioning VDW.CONNECT Locate
Apex Locator or by screening off its immediate environment. For further information refer to
Appendix A: Electromagnetic Emissions and Immunity.
6
®
Apex Locator only in accordance with this DFU to ensure
®
Apex Locator and the VDW.CONNECT® App are intended for:
®
Apex Locator on a flat and stable surface (if used without
®
Apex Locator must not be placed on damp surfaces, in damp
®
Apex
®
Apex
®
7
®
•VDW.CONNECT Locate
Apex Locator requires special precautions with regard to
electromagnetic compatibility (EMC) and must be installed and operated in strict compliance
with the EMC Directive. In particular, VDW.CONNECT Locate
®
Apex Locator must not be
used in the vicinity of fluorescent lamps, radio transmitters, remote controls or portable /
mobile RF communication devices, even if they meet CISPR 11 requirements. See
Appendix A: Electromagnetic Emissions and Immunity.
•Make sure cables do not limit mobility during a procedure.
•Do not kink the cables or the charger.
•Do not pull directly on the cables when plugging in or unplugging components. Make sure the
cables are not obstructed or have limited mobility.
•Refer also to section 8 Technical Characteristics for general Environmental Conditions
specifications.
•Use of VDW.CONNECT Locate
®
Apex Locator adjacent to, or stacked with other equipment
should be avoided because it could result in improper operation. If such use is necessary,
VDW.CONNECT Locate
®
Apex Locator and the other equipment should be observed to
verify that they are operating normally.
•Do not expose VDW.CONNECT Locate® Apex Locator to infrasound or ultrasound sources.
During treatment
•Ensure that contact point between file and file clip are kept dry to avoid inaccurate readings.
®
•Make sure that the VDW.CONNECT Locate
Lip Clip and the VDW.CONNECT Locate®
File Clip do not come into contact with an electric power source such as an electrical socket.
This could result in severe electrical shock.
•The cable functionality must be checked with a closed-loop test before each use. See 6.9.1
Closed-loop test.
•If the VDW.CONNECT Locate
®
File Clip is not properly connected to the endodontic
instrument, VDW.CONNECT Locate® Apex Locator will not provide correct information and
the file clip could be damaged.
•Use of gloves and of a rubber dam are mandatory during treatment.
®
•Do not pour water or any other liquid over VDW.CONNECT Locate
•The use of VDW GmbH VDW.CONNECT Locate
®
Disposable Barrier Sleeves is required
Apex Locator.
(see section 7 Cleaning, Disinfection and Sterilization). Use only custom barriers provided
by VDW GmbH made specifically for VDW.CONNECT Locate® Apex Locator.
•Always perform an X-ray and set the stopper to the determined working length before using
®
VDW.CONNECT Locate
Apex Locator.
•Before starting the reading, dry the access cavity with a cotton pellet and ensure that there is
no excess of liquid inside the root canal targeted for the treatment. An excess of liquid can
hinder the reading process. Take into account that the canal should not be completely dry
otherwise no reading will occur. Moisten an extremely dry canal with Glyde™ file prep to help
provide a stable reading.
8
•Before starting the treatment, make sure that the device is in a stable position (over a flat
surface or firmly attached to the lab coat if using the clamp) to ensure that the treatment can
be done comfortably and that any visual / auditory signals emitted by VDW.CONNECT
®
Locate
•If any irregularities with VDW.CONNECT Locate
Apex Locator can be received.
®
Apex Locator should occur during
treatment, switch it off, consult Section 10 Troubleshooting or Contact VDW GmbH
Customer Service.
®
•The scale indication on VDW.CONNECT Locate
Apex Locator and on VDW.CONNECT®
App does not represent a distinct length or distance in [mm] or other linear units. It only
indicates an estimated proximity of the file tip to a reference point for working length
determination.
•Make sure to use only fully metallic endodontic files whether in combined mode or in
uncombined mode.
•Only use undamaged endodontic instruments, meaning the instruments shall be in good
condition and within their useful life (as defined by the manufacturer). Please, refer to the
information provided by the manufacturer of the instrument.
•An accurate working length determination cannot be made using a damaged
VDW.CONNECT Locate
®
File Clip.
•Only use original cables and adaptors.
•The following patient-related factors may prevent accurate readings:
- Blocked root canal.
- Teeth with large apices.
- Root fracture or perforation.
- Metal crowns or bridges.
- Endodonic file not matching canal size (canal size much larger than endodontic file
size).
•If there is no audible sound or no visible flash of the LEDs when VDW.CONNECT Locate®
Apex Locator is switched on, do not use it.
®
•Always switch off VDW.CONNECT Locate
•VDW GmbH cannot be held liable if VDW.CONNECT Locate
Apex Locator before cleaning it.
®
Apex Locator is not installed,
operated and maintained in accordance with the instructions in this DFU. No modifications to
VDW.CONNECT Locate® Apex Locator are allowed, except battery changes (see following
Section).
•Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the
®
VDW.CONNECT Locate
Apex Locator (IEC 60601-1), including cables specified by VDW
GmbH. Failure to observe this could result in a degradation of the performance of this
equipment. For further information, refer to Appendix A: Electromagnetic Emissions and
Immunity.
9
After Treatment
•Never leave VDW.CONNECT Locate® Lip Clip or VDW.CONNECT Locate® File Clip in
contact with a patient if either of them is not plugged into the measurement cable. This could
result in severe electrical shock if they come into contact with an electrical source.
•Follow instructions given in section 7 Cleaning, Disinfection and Sterilization.
Battery
•Always use the original charger with the charging cable to charge VDW.CONNECT Locate®
Apex Locator. Use of non-original chargers may result in serious risk for the safety of the
user.
•Regularly check for signs of battery leakage.
®
•If fluid leaks from VDW.CONNECT Locate
Apex Locator, the battery may be damaged.
Stop using the device immediately. If the leakage occurs while charging, disconnect the
charger immediately. Avoid skin contact with the leaking fluid and contact the VDW GmbH
Customer Service as soon as possible.
•Do not recharge the battery in case of suspected faulty/ leaking battery.
•Do not use batteries past the expiry date (when provided).
•Never use a battery that is deformed, discolored or otherwise abnormal.
•If the batteries do not charge fully after 6 hours, they should be replaced.
•To charge the device, connect the charger to electrical mains compliant with the regulations
in each country.
•Be sure to purchase and use only the original battery (part reference M00CLOC2ACBAT).
®
•Only charge VDW.CONNECT Locate
Apex Locator out of patients reach.
•Only perform VDW.CONNECT Locate® Apex Locator test out of patients reach. Use new
gloves to avoid contamination of the external tester.
®
•Only open VDW.CONNECT Locate
Apex Locator to change the batteries (see section
6.12.1 Battery Replacement). Removing more than the bottom cover of VDW.CONNECT
®
Locate
Apex Locator voids the warranty. Contact VDW GmbH Customer Service for
assistance.
•Fully charge VDW.CONNECT Locate® Apex Locator prior to its first use (6 hours charging
time).
®
•Always ensure that both VDW.CONNECT Locate
Apex Locator and the iPad® have
sufficient battery charge before starting treatment. At low battery level, VDW.CONNECT
®
Locate
Apex Locator should not be used. See section 6.12 Battery Care.
•VDW.CONNECT Locate® Apex Locator must not be charged during treatment.
®
•VDW.CONNECT Locate
Apex Locator with an empty battery will not provide any audio or
visual signals.
10
Combined mode
•VDW.CONNECT Locate® Apex Locator can be used in combined mode with
®
VDW.CONNECT Drive
•Only use VDW.CONNECT Drive
Handpiece.
®
Handpiece for working in combined mode. Using
VDW.CONNECT Locate® Apex Locator with another Handpiece can lead to severe
electrical problems (including electrical shock).
•Only use VDW GmbH accessories and spare parts.
®
•Do not plug the VDW.CONNECT Locate
original cables into any other devices with
compatible plugs.
®
•The use of the VDW.CONNECT Drive
Handpiece Sleeve is mandatory to ensure proper
functioning of the system in combined mode. The barrier is designed for single use and must
be disposed of and changed after each patient.
®
•Please ensure that you have the latest version of VDW.CONNECT Drive
Handpiece
firmware and of VDW.CONNECT Locate® Apex Locator firmware to work in combined
mode.
For further information on VDW.CONNECT Drive® Handpiece Sleeve, please refer to
VDW.CONNECT Drive® Handpiece Directions For Use.
Please read VDW.CONNECT Drive® Handpiece Directions For Use - 3 Warnings before using
VDW.CONNECT Drive® Handpiece.
Please read VDW.CONNECT® App User Manual - 3 Warnings before using VDW.CONNECT®
App.
Accessories
•VDW.CONNECT Locate® Lip Clip and VDW.CONNECT Locate® File Clip are supplied
unsterilized. You must reprocess them prior to first use and between patients (Follow
instructions given in section 7 Cleaning, Disinfection and Sterilization).
•Use VDW GmbH components / accessories and spare parts only. Using non-original
accessories / spare parts can lead to increased emission of electromagnetic interferences or
to reduced electromagnetic interference immunity.
•Examine the state of the connectors before use.
•Never leave VDW.CONNECT Locate
®
Lip Clip or VDW.CONNECT Locate® File Clip
unplugged to the VDW.CONNECT Locate® Measurement Cable while in contact with a
patient.
11
Repairs and Defects
•Check if VDW.CONNECT Locate® Apex Locator works properly, if it has been subjected to
mechanical shock or impact (e.g. if it has been dropped). Do not use it in case of malfunction,
apparent breakage or cracks. The safety and effectiveness of the device can be seriously
®
compromised. In cases of major shock, VDW.CONNECT Locate
Apex Locator needs to be
inspected by authorized technical personnel. Contact VDW GmbH Customer Service for
assistance.
•VDW.CONNECT Locate
®
Apex Locator does not contain user serviceable parts (except
batteries). The service and repair should be provided by factory trained service personnel
only. Therefore, repairs, alterations and modifications to VDW.CONNECT Locate® Apex
Locator are not permitted. VDW GmbH cannot be held liable if VDW.CONNECT Locate
®
Apex Locator has been altered or modified. If a defect should occur, contact VDW GmbH
Customer Service.
®
•It is mandatory to reprocess VDW.CONNECT Locate
Apex Locator and its accessories
(including cables and tester) before shipping them to the service centre (follow instructions
given in section 7 Cleaning, Disinfection and Sterilization).
®
•Protect VDW.CONNECT Locate
Apex Locator with appropriate packaging when sending it
for service.
Transport
•Intact VDW.CONNECT Locate® Apex Locator can be transported by land freight or airfreight
in the original packaging. The applicable requirements must be met (see table below).
®
•Defective VDW.CONNECT Locate
Apex Locator with intact battery can also be transported
by airfreight or land freight in the original packaging.
•If the battery is defective (refer to Battery section above to identify defective batteries),
®
VDW.CONNECT Locate
Apex Locator must not be transported by airfreight under any
circumstances.
Applicable standards and regulations for the transport of VDW.CONNECT Locate® Apex Locator
Intact
VDW.CONNECT
Locate® Apex
Locator or defective
VDW.CONNECT
®
Locate
Apex
Locator with intact
battery.
VDW.CONNECT
Locate® Apex
Locator with
defective battery.
Electromagnetic emissions and immunity
•HF (high-frequency) surgical equipment can influence the operation of the device and may
not be operated in combination with VDW.CONNECT Locate
•Portable wireless communications equipment such as wireless home network devices,
mobile phones, cordless telephones and their base stations, walkie-talkie etc. can affect
VDW.CONNECT Locate
to any part of VDW.CONNECT Locate® Apex Locator.
•The use of accessories, transducers and cables other than those specified, with the
exception of transducers and cables sold by the manufacturer of the VDW.CONNECT
Locate
emissions or decreased immunity of the VDW.CONNECT Locate
•Use of VDW.CONNECT Locate® Apex Locator adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If such use is necessary,
VDW.CONNECT Locate
verify that they are operating normally.
12
Air freightLand freight
•IATA: The batteries are
considered “Not restricted” as long
as they are shipped in compliance
with SP A199.
•UN 3496 NiMH (Nickel-metal hydride)
batteries in equipment.
•EU (ADR / RID) Identified damaged or
defective batteries shall be carried in
Not possible.
accordance with especial provision 376
and packaged in accordance with
packaging instructions P908 of 4.1.4.1 or
LP904 of 4.1.4.3 as applicable.
®
Apex Locator.
®
Apex Locator and should be kept at least at a distance of 30 cm
®
Apex Locator or replacement parts for internal components, may result in increased
®
Apex Locator and the other equipment should be observed to
®
Apex Locator.
13
4PRECAUTIONS
Read the following precautions before use.
•The manufacturer declines any responsibility in the case of using VDW.CONNECT Locate®
Apex Locator outside of its intended use or indications specified in this DFU. Please refer to
sections 1 Indications for Use and 3 Warnings for further information on the indications for
use.
•VDW.CONNECT Locate
(see Sections 2 Contraindications and 3 Warnings).
•It is mandatory to take preoperative radiographs prior to using VDW.CONNECT Locate
Apex Locator and to compare the information given by both methods. In case of clinically
relevant discrepancy, it is recommended to take a second radiograph with an endodontic file
in place at the selected working length.
®
Apex Locator may not be able to work properly in all conditions
5ADVERSE REACTIONS
The use of VDW.CONNECT Locate® Apex Locator is not recommended with patients allergic to
metals (see Section 8 Technical Characteristics).
If VDW.CONNECT Locate® Apex Locator provides an incorrect reading and there is no
radiographic data to compare to, the following adverse reactions may occur:
•Inadequate root canal treatment due to under- or over-instrumentation.
6STEP BY STEP INSTRUCTIONS
®
Refer to the section 3 Warnings to verify any special care to exercise before starting to use the
device.
Refer to the section 13 Identification of Symbols to verify the description of all symbols before
starting to use the device.
Refer to VDW.CONNECT® App User Manual for all information regarding VDW.CONNECT
Locate® Apex Locator connection with VDW.CONNECT® App.
Before use, check the exact contents of the package. See 6.3 Package Contents.
14
6.1Document Symbols
SymbolIdentification
Consult VDW.CONNECT® App User Manual or VDW.CONNECT Locate® Apex Locator
Directions For Use
If the instructions are not followed properly, operation may carry risks for the product or the
user / patient
Additional information, explanation on operation and performance
Suggestion or advice
6.2Device Compatibility
SpecificationDescription
•iPad Mini™ 2
•iPad Mini™ 3
®
iPad
VDW.CONNECT® App operating system
VDW.CONNECT Drive® Handpiece
*Apple®, the Apple logo, iPad®, iPad Pro® and iPad Mini™ are trademarks of Apple Inc., registered in the U.S. and other
countries. App Store
®
is a service mark of Apple Inc.
•iPad Mini™ 4
•iPad® (6th gen)
®
•iPad Pro
•iPad Pro
10.5"
®
12.9" (2nd gen)
•iOS
•Original VDW GmbH VDW.CONNECT Drive®
Handpiece
6.3Package Contents
N°Action
Remove carefully the device and the accessories from their packaging and place them on a flat
A
surface.
Check that VDW.CONNECT Locate® Apex Locator is supplied with all the components shown
B
below and that all components are in good condition.
List of components with VDW GmbH Stock Keeping Unit
VDW.CONNECT Locate® Apex
Locator M00CLOC000000
(supplied with rechargeable
battery M00CLOC2ACBAT)
VDW.CONNECT Locate®
Measurement Cable
M00CLOC1ACCBL
15
R
VDW.CONNECT Drive®
Measurement Cable
M00CLOC1ACHPA
VDW.CONNECT Locate®
External Tester
M00CLOC1ACTST (There may
be pins attached to the back of
the tester depending on the
tester version. These pins have
no function)
List of components with VDW GmbH Stock Keeping Unit
VDW.CONNECT Locate® Lip
Clip (x 5) M00CLOC5ACLCL
VDW.CONNECT Locate® File
Clip (x 2) M00CLOC2ACFCL
VDW.CONNECT Locate®
Clamp* M00CLOCCLAMP0
16
AC Adapter Plugs
(*EU, *UK, US, *ANZ and
universal adapter)
M00CLOC1ACCHG
Getting Started Guide(s) / DFU-
*These components are not available in all kits
17
Optional VDW.CONNECT Locate® Clamp
The device is delivered with an optional clamp (if available in your country), which can be installed
on the back of the device with a screw (included in the package). The clamp can be installed with
a standard flat / slot drive screwdriver (not included in the package). Remove the grub screw
before attaching the clamp.
Components not included
The following components are not included in the kit:
iPad® with iOS operating system.
6.3.1Assembling the charger
N°Action
ASelect the charger plug adapter that matches your power outlet.
BPlace the plug adapter into the slots and slide it downwards until it locks with a click.
6.4Modes of Operation
R
R
VDW.CONNECT Locate® Apex Locator can be used in 2 different modes:
Uncombined modeCombined mode
18
Working only with VDW.CONNECT Locate®
Apex Locator.
For cable connection see 6.8.1 Plugging in the
VDW.CONNECT Locate® Measurement Cable
in uncombined mode.
Working with VDW.CONNECT Locate® Apex Locator
®
plugged to VDW.CONNECT Drive
Handpiece for
shaping towards the reference point while receiving
endodontic file progression information.
®
VDW.CONNECT Locate
Apex Locator can
be used while VDW.CONNECT Drive®
Handpiece motor is rotating.
®
The use of VDW.CONNECT Drive
Handpiece Sleeve is mandatory to ensure
the proper functioning of the system in
combined mode.
The VDW.CONNECT Drive® Handpiece sleeve supplied
in your country may look different from the
®
VDW.CONNECT Drive
Handpiece sleeve shown on
the illustration
R
For cable connection see 6.8.2 Plugging in the
VDW.CONNECT Drive® Measurement Cable in
combined mode.
On both operating modes, the device can be connected to the VDW.CONNECT® App.
Refer to VDW.CONNECT® App User Manual.
19
6.5Description of the VDW.CONNECT Locate® Apex Locator
6.5.1General overview
Connector for
VDW.CONNECT
Locate®
Measurement
Cable,
VDW.CONNECT
Drive®
Measurement
Cable, charger
cable and external
tester
ON/OFF/Standby button
File progression
indication
R
Screws of battery
compartment
Rubber pad
Bluetooth
®
button
Screw thread
Battery compartment
The screw thread provides the ability to connect a VDW.CONNECT Locate® Clamp. Check with
your distributor for more information about the clamp (if not included in the kit). The illustration
above might slightly differ from the actual product.
20
6.6Description of the LED Colors
Button / LEDFunctionStatusDescription
All LEDs will light up white for a short period of time at the startup of
VDW.CONNECT Locate
®
Apex Locator.
VDW.CONNECT Locate® Apex Locator is off.
Device is powering on.
Power status
Device is on.
Stand-by.
LED flashing every 2.5 seconds, with a color
corresponding to battery status.
Device is charging.
R
LED flashing every second.
Device is sufficiently charged.
Enough autonomy for several procedures.
Low battery level.
Battery status
Enough autonomy for approximately 15 minutes.
Very low battery level.
Enough autonomy for approximately 5 minutes.
Do not start a new treatment at this battery level.
Critically low battery level.
There is no autonomy left. Please recharge your
®
VDW.CONNECT Locate
Apex Locator.
21
Button / LEDFunctionStatusDescription
Bluetooth® not activated.
Bluetooth® active but device not connected.
Connection of
VDW.CONNECT
R
Locate® Apex
Locator with the
VDW.CONNECT
®
App
Device connected to VDW.CONNECT® App.
See VDW.CONNECT® App User
Manual - 6 Step by Step Instructions.
Connection lost.
See VDW.CONNECT® App User
Manual - 10 Troubleshooting.
Shaping target reverted to default setting 5
minutes after connection loss (if no file progress
information was received during that time).
File progress information.
Endodontic file
progress
information
The color (blue, green flashing, green, red
flashing) varies according to the progression of
the file in the root canal. See 6.10.3 Working
length determination.
R
Closed-loop test
(See 6.9.1
Closed-loop test)
Closed-loop test is successful.
See 6.9.1 Closed-loop test.
Closed-loop-test failed.
Any indication other than white steady or no
indication at all.
See 10 Troubleshooting.
Device test is on-going.
Device test
R
(see 6.9.2 Device
test)
Device test is successful.
The LED is ON until the VDW.CONNECT
Locate® External Tester is unplugged.
Device test failed.
The LED is ON until the VDW.CONNECT
®
Locate
External Tester is unplugged.
6.7Description of the Sounds
Sound volume can only be adjusted through the VDW.CONNECT® App.
See VDW.CONNECT® App User Manual - 6 Step by Step Instructions.
SituationSignal
Any button pressed
VDW.CONNECT Locate® External Tester
connected
Combination of VDW.CONNECT Locate®
Apex Locator with VDW.CONNECT Drive
Handpiece
Device test
Working length determination
•One short beep.
•Two short beeps
•Double beep if VDW.CONNECT Locate® Apex
•No beep if an unknown cable is plugged or if
®
•Three longer beeps if the combination is lost.
See 6.10.5 Combined mode.
•Double beep when VDW.CONNECT Locate® Apex
•Alternating beeps when VDW.CONNECT Locate
See 6.9.2 Device test.
•Beeping intervals become shorter as endodontic file
•At reference point, continuous tone until apical
•Fast pulsating tones (alarm condition) when file
See 6.10.3 Working length determination.
22
Locator is correctly combined with VDW.CONNECT
®
Handpiece.
Drive
®
VDW.CONNECT Locate
Apex Locator is
incorrectly combined with VDW.CONNECT Drive®
Handpiece.
Locator device test is successful.
Apex Locator device test failed.
progresses towards reference point for working
length determination.
foramen (major apical foramen/diameter) is passed.
passes the apical foramen.
®
VDW.CONNECT Drive® Handpiece will rotate in reverse direction when VDW.CONNECT
®
Locate
Apex Locator is uncombined during a treatment and the Apical Reverse function is
activated.
23
6.8Plugging in the VDW.CONNECT Locate® Measurement
Cable
6.8.1Plugging in the VDW.CONNECT Locate® Measurement Cable in
uncombined mode
N°Action
A
Insert VDW.CONNECT Locate® Apex Locator inside its proprietary single use protective sleeve.
Plug in the VDW.CONNECT Locate® Measurement Cable (2) to VDW.CONNECT Locate®
B
Apex Locator (1).
Connect the VDW.CONNECT Locate® Lip Clip (3) to the VDW.CONNECT Locate®
C
Measurement Cable (2).
Connect the VDW.CONNECT Locate® File Clip (4) to the VDW.CONNECT Locate®
D
Measurement Cable (2).
1
2
3
4
24
R
R
6.8.2Plugging in the VDW.CONNECT Drive® Measurement Cable in
combined mode
N°Action
Only use VDW.CONNECT Drive® Handpiece for working in combined mode. Using
VDW.CONNECT Locate
®
Apex Locator with another Handpiece can lead to severe electrical
problems (including electrical shock).
A
Insert VDW.CONNECT Drive® Handpiece inside its proprietary single-use protective sleeve.
The use of the VDW.CONNECT Drive® Handpiece Sleeve is mandatory to avoid contact
between the contra-angle and the patient’s mouth, which might cause wrong readings on
This feature checks the cables. This test should be performed before each treatment.
N°Action
Turn ON VDW.CONNECT Locate® Apex Locator (1).
A
See 6.6 Description of the LED Colors.
Connect the VDW.CONNECT Locate® or VDW.CONNECT Drive® Measurement Cable (2 or 5)
B
to VDW.CONNECT Locate
Connect the VDW.CONNECT Locate® Lip Clip (3) and the VDW.CONNECT Locate® File
Clip (4) or VDW.CONNECT Drive
C
VDW.CONNECT Drive
in the VDW.CONNECT Drive
Make a closed-loop circuit with the VDW.CONNECT Locate® or VDW.CONNECT Drive®
Measurement Cable:
Hook the VDW.CONNECT Locate
D
with the endodontic file inserted in the VDW.CONNECT Drive
VDW.CONNECT Locate
functional.
Any result other than white indicates a test fail.
®
Apex Locator (1).
®
®
Measurement Cable (2 or 5). Ensure that an endodontic file is inserted
®
Apex Locator (1) central LED will turn white indicating that cables are
Handpiece (6) to the VDW.CONNECT Locate® or
®
Contra-angle.
®
File Clip (4) to the lip clip (3) or put the lip clip (3) in contact
®
Contra-angle (7).
If the closed-loop test failed, please refer to 10 Troubleshooting.
R
2
3
1
4
R
1
The VDW.CONNECT Drive® Handpiece sleeve
supplied in your country may look different from
the VDW.CONNECT Drive
shown on the illustration.
®
5
Handpiece sleeve
3
R
6
7
26
6.9.2Device test
Use the VDW.CONNECT Locate® External Tester to test the basic functionality of the
VDW.CONNECT Locate® Apex Locator. Run this test periodically (e.g. once a week), after the
device has not been used for a long period of time or for troubleshooting reasons (as per Section
10 Troubleshooting).
The VDW.CONNECT® App will guide you through the different steps of the device test.
See VDW.CONNECT® App User Manual - 6 Step by Step Instructions.
N°Action
Turn ON VDW.CONNECT Locate® Apex Locator.
A
See 6.6 Description of the LED Colors.
Plug the VDW.CONNECT Locate® External Tester (1) in the device (2).
B
The device test will automatically start. The central LED will flash white and a double beep will be
emitted.
R
1
2
27
If device test is successful:
N°Action
Central LED will light up green and a double beep will be emitted simultaneously.
You can now unplug the VDW.CONNECT Locate® External Tester from VDW.CONNECT
C
Locate® Apex Locator.
R
If device test fails:
N°Action
Central LED will light up red and alternating beeps will be emitted simultaneously.
Unplug the VDW.CONNECT Locate® External Tester from VDW.CONNECT Locate® Apex
D
Locator.
Repeat the above procedure (A to B one more time).
E
If the error persists DO NOT USE THE DEVICE and contact your VDW GmbH Customer
Service.
R
28
6.10Determining the Working Length
6.10.1Definition
The anatomical terminology is defined as follows:
N°Description
(1)Coronal and medial regions.
(2)Region coronal to the reference point for working length determination.
(3)Reference point for working length determination.
(4)Region beyond reference point.
(5)Apical foramen.
(6)Apical foramen (Major apical foramen or diameter) passed/over-instrumentation.
29
6.10.2Start working length determination
As detailed in section 6.4 Modes of Operation, VDW.CONNECT Locate® Apex Locator can work
in different modes. Please also refer to Section 3 Warnings.
Regardless of the mode of operation, the following steps and principles apply.
Prior to using VDW.CONNECT Locate® Apex Locator on a patient, the access cavity must be
dried with a cotton pellet to prevent an electric short circuit through another canal. However, the
canal being treated must be humid to allow electric conduction. Both conditions need to be met to
ensure correct file progression information.
Moisten an extremely dry canal with Glyde™ file prep to help provide conditions for a stable
reading.
Select the size of the file matching the root canal diameter.
Before starting to work with the VDW.CONNECT Locate® File Clip, check if corrosion is visible
on the metal part of the hook. In case corrosion is present, use a new file clip (see Section 10
Troubleshooting).
For detailed steps on how to work with VDW.CONNECT Locate® Apex Locator in uncombined
and combined modes, refer to sections 6.10.4 Uncombined mode and 6.10.5 Combined mode.
6.10.3Working length determination
It is mandatory to read carefully Section 3 Warnings before reading this section. Important
safety information is contained there.
When working in connected mode, the working length is graphically illustrated in
VDW.CONNECT® App. Please refer to VDW.CONNECT® App User Manual - 6 Step by Step
Instructions.
Color indication on
VDW.CONNECT Locate
Apex Locator
R
®
Endodontic file progression
Coronal and medial regions:
Move the endodontic file apically.
LED: Solid blue.
Sound: Slow pulsating tone.
Color indication on
VDW.CONNECT Locate®
Apex Locator
30
Endodontic file progression
Audio signals (from the device) serve as indication of the file tip progression
towards reference point. The interval between the beeps becomes shorter
when the endodontic file approaches the reference point for working length
determination.
Region coronal to the reference point for working length determination:
R
The endodontic file can be moved further apically.
LED: Flashing green.
Sound: Fast pulsating tone.
Reference point for working length determination:
R
This is the relevant position to determine the working length.
LED: Solid green.
Sound: Solid tone.
In combined mode, the Apical Reverse function can be activated. Refer to
®
VDW.CONNECT
App User Manual – 6 Step by Step Instructions.
The position of the reference point can be adjusted with the “Shaping Target”
function, which is only available when VDW.CONNECT Locate
®
Apex Locator
is connected to the VDW.CONNECT® App. Refer to the VDW.CONNECT®
App User Manual for instructions on that function.
Region beyond reference point:
R
Further file progression beyond the reference point is visualized.
LED: Solid green.
Sound: Solid tone.
Color indication on
VDW.CONNECT Locate®
Apex Locator
31
Endodontic file progression
Apical foramen (Major apical foramen or diameter) passed / overinstrumentation.
R
Alarm condition:
LED: Flashing red.
Sound: Fast pulsating tones.
Potential harm to the patient can be caused when this region is reached.
6.10.4Uncombined mode
If you have an iPad® and you prefer to work in connected mode, see VDW.CONNECT® App User
Manual.
The following step-by-step instructions refer to the uncombined use (without VDW.CONNECT
Drive® Handpiece).
Ensure that a rubber dam/dental dam is securely placed before you start endodontic treatment.
Avoid touching the device with contaminated gloves to prevent contamination near the connector
or underneath the clamp.
32
N°Action
After reprocessing of VDW.CONNECT Locate® Apex Locator and its accessories (see Section 7
®
Cleaning, Disinfection and Sterilization), place VDW.CONNECT Locate
single-use VDW.CONNECT Locate
®
Disposable Barrier Sleeve, with the long end of the barrier
Apex Locator in a
sleeve at the top. If you are using the optional VDW.CONNECT Locate® Clamp on the back of
the device, pull a second sleeve over the clamp as follows:
A
B
Press the ON/OFF/Standby button on VDW.CONNECT Locate® Apex Locator (1).
C
Connect the VDW.CONNECT Locate® Measurement Cable (2) to the device (1).
Connect the VDW.CONNECT Locate® Lip Clip (3) to the VDW.CONNECT Locate®
D
Measurement cable (2).
Connect the VDW.CONNECT Locate® File Clip (4) to the VDW.CONNECT Locate®
E
Measurement Cable (2).
FPerform a closed-loop test. See 6.9.1 Closed-loop test.
GPut the lip clip (3) on the patient’s lip on the opposite site of the tooth to be treated.
Insert the endodontic file (5) into the root canal and attach the file clip (4) to the metallic part of
H
the endodontic file (5) directly below the plastic handle and above the stopper (6).
Move the endodontic file apically until reference point (default reference point or reference point
I
defined as per Shaping Target function) is reached.
JSet the silicon stopper at the coronal reference.
Remove the endodontic file from the root canal and use an external ruler to measure the distance
K
between the tip of the endodontic file and the silicon stopper. This information will be relevant for
subsequent files in your sequence.
Compare this distance information with the corresponding information given by pre-treatment
L
radiographs.
Proceed with endodontic treatment. Carefully monitor file progression to avoid over-
M
instrumentation.
33
2
R
1
Please note that the mandatory rubber dam is not
shown in the schematic image below, for clarity’s
sake.
3
5
4
6
3
34
6.10.5Combined mode
If you own VDW.CONNECT Drive® Handpiece, you can operate VDW.CONNECT Locate® Apex
Locator in combined mode for simultaneous canal shaping and file progression monitoring.
It is mandatory to read carefully Section 3 Warnings before reading this section. Important safety
information is contained there.
The use of VDW.CONNECT Drive® Handpiece Sleeve is mandatory to ensure the correct
functioning of VDW.CONNECT Locate
Ensure that a rubber dam/dental dam is securely placed before starting endodontic treatment.
Always establish the working length manually (radiographically and/or using uncombined mode)
before starting an endodontic treatment in combined mode.
Avoid touching the device with contaminated gloves to prevent contamination near the connector
or underneath the clamp.
Please refer to VDW.CONNECT Drive® Handpiece Directions For Use for its use and specific
hygiene measures.
N°Action
®
Apex Locator in combined mode.
After reprocessing of VDW.CONNECT Locate® Apex Locator and its accessories (see Section 7
®
Cleaning, Disinfection and Sterilization), place VDW.CONNECT Locate
single-use VDW.CONNECT Locate® Disposable Barrier Sleeve, with the long end of the barrier
A
sleeve at the top. If the optional VDW.CONNECT Locate
®
VDW.CONNECT Locate
Disposable Barrier Sleeve over the clamp as illustrated in Section
®
Clamp is used, pull a second
Apex Locator in a
6.10.4 Uncombined mode.
B
Press the ON/OFF/Standby button on VDW.CONNECT Locate® Apex Locator (1).
C
Connect the VDW.CONNECT Drive® Measurement Cable (2) to the device (1).
Connect the VDW.CONNECT Locate® Lip Clip (3) to the VDW.CONNECT Drive® Measurement
D
Cable (2).
Put a VDW.CONNECT Drive® Handpiece single-use barrier (8) over the Handpiece (5) to isolate
E
it.
Plug the VDW.CONNECT Drive® Measurement Cable (4) into the Handpiece (5) and insert an
F
endodontic file (7) into the VDW.CONNECT Drive
®
Contra-angle (6).
GPerform a closed-loop test. See 6.9.1 Closed-loop test.
Put the VDW.CONNECT Locate® Lip Clip (3) on the patient’s lip on the opposite side of the tooth
H
to be treated.
Move the endodontic file apically until reference point (default reference point or reference point
I
defined as per Shaping Target function) is reached.
Proceed with endodontic treatment. Carefully monitor file progression to avoid over-
J
instrumentation.
In combined mode, you can use the Apical Reverse function. This function is only available when
®
the VDW.CONNECT Drive
to the VDW.CONNECT
Handpiece is connected to the VDW.CONNECT® App. Please refer
®
App User Manual for description of that function.
35
R
R
Please note that the mandatory rubber dam is not shown in the
schematic image below, for clarity’s sake.
The Handpiece sleeve supplied in your country may look
different from the Handpiece sleeve shown on the illustration.
1
R
2
R
4
3
5
7
6
3
36
6.10.6Interruption of file progression indication
During the treatment, the VDW.CONNECT Locate® File Clip may be disconnected from the
endodontic file and reconnected at any time (e.g. when the file is changed to a larger size or when
the length of another canal should be determined).
6.10.7Tips for successful file progression indication
Implausible, rapidly changing indications from VDW.CONNECT Locate® Apex Locator may have
various causes. Should you experience either situation, please review the below checklist to
identify possible causes and take appropriate actions.
IssueSolution
Excess liquid in the pulp chamber or root canal
(i.e. rinsing solution, blood or saliva) creating
wrong conductive path and incorrect length
determination.
Gingival proliferation can lead to direct contact
with the endodontic file causing a short circuit
and incorrect length determination.
The endodontic file contacting metallic dental
restorations (crown, intracanal, amalgam filling)
may cause a short circuit and incorrect length
determination.
No rubber dam or no VDW.CONNECT Drive®
Handpiece Sleeve is used. This may cause a
distortion of the electrical signal for the reading.
Dry the access cavity with a cotton pellet / air-blower.
Wait until any bleeding has stopped.
Isolate the access cavity by:
•Adequate preparation filling;
•Placing a dental dam;
•Electro-cauterizing.
Carefully enlarge the access cavity and isolate with flow
composite.
Widen carefully the opening at the top of the crown.
Use a rubber dam and the VDW.CONNECT Drive®
Handpiece Sleeve.
37
If the device indications change too slowly, please review the following check-list:
IssueSolution
Blocked root canal impeding the conductive path
and preventing normal device functioning.
Retreatment: blockage by old canal filling
material residue, impeding the conductive path
and preventing normal device functioning.
Blockage by remnants of a medicated substance
(e.g. calcium hydroxide) impeding the
conductive path and preventing normal device
functioning.
Extremely dry root canal impeding the
conductive path and preventing normal device
functioning.
Special condition symptom: exceptionally large
apical foramen due to lesion or incomplete
formation.
Root fracture or perforation.
Check the comparative radiographs for support
information on canal length; Catheterize with ISO 06 / 08
endodontic file until the working length is reached.
Take a radiograph to re-check and try to completely
remove the old canal filling material prior to determining
length.
Completely remove the remnants prior to determining
length.
Rinse root canal with irrigating solution such as NaCl or
NaOCl and dry the access cavity with a cotton pellet / airblower.
May lead to incorrect feedback from VDW.CONNECT
®
Locate
May lead to incorrect feedback from VDW.CONNECT
Locate® Apex Locator. Perform radiography.
Apex Locator. Perform radiography.
6.10.8Comparison of electronic length determination versus
radiography
Radiography represents a two-dimensional projection of a three-dimensional root canal system.
There are some cases when the radiographic length and the electronic length do not match. In
case of discrepancy, please refer to GCP standards applicable in your country as well as your
clinical expertise.
6.11Stand-by Mode
When VDW.CONNECT Locate® Apex Locator is not used for a certain period of time, it enters
stand-by mode as described below.
ModeLED descriptionWake-up action
38
Device active
Stand-by mode
Shut-down
(inactive for 20 min)
R
ON/OFF/Standby LED emits a color
corresponding to the battery status.
R
ON/OFF/Standby LED flashes and emits a
color corresponding to the battery status.
Bluetooth
®
and Central LEDs are turned
OFF.
R
None.
(VDW.CONNECT Locate® Apex
Locator ready to be used).
Start using VDW.CONNECT
Locate® Apex Locator.
Press the ON/OFF/Standby
button.
All LEDs are turned OFF.
39
6.12Battery Care
It is recommended to read Section 3 Warnings (Battery) before reading this section. Important
safety information is located there.
VDW.CONNECT Locate® Apex Locator is powered by two AAA (1.2 V) Ni-MH batteries. During
the operation, the battery status is shown on the device front panel as the color of the ON/OFF/
Standby button. See 6.6 Description of the LED Colors.
The following notes will help you to ensure the battery provides a long service life:
•If the battery LED on the device flashes red, there is less than 5 min of device autonomy left.
You must charge the device.
•Do not charge the battery more than once a day. It is recommended that you charge the
device at the end of each day.
•Charge the battery overnight after using VDW.CONNECT Locate
battery is fully charged after up to 6 hours of charging.
•Always fully charge the battery, i.e. until the battery LED stops flashing green and turns
steady green.
•When the charger is unplugged, the device will go in active mode.
•When the device is in standby mode and the charger is plugged, the device will start
charging.
•When the device is turned off and the charger is plugged, the device will start charging.
N°Action
AConnect the charger (2) to the device (1) and plug it into the power outlet.
®
Apex Locator. The
1
R
2
40
6.12.1Battery Replacement
It is recommended to read section 3 Warnings (Battery) before reading this section. Important
safety information is located there.
Before replacing the battery, disconnect the charger from the device. Then switch the device OFF.
Always replace the batteries with two new batteries if they cannot be recharged correctly. The old
batteries must be disposed of in accordance with local regulations.
Fully charge the batteries before first use. Only use VDW GmbH batteries (part reference
M00CLOC2ACBAT).
The battery compartment is located at the rear of VDW.CONNECT Locate® Apex Locator and its
cover is secured by two 6-lobe screws (screw type: KN6041; screwdriver type: Torx T5).
N°Action
AUsing an Torx T5 screwdriver, unscrew the two screws (1) and remove the battery cover (2).
BRemove the two batteries (3) from the battery compartment (4).
CInsert two new batteries, replace the battery cover (2) and screw the two screws (1).
In order to avoid ingress of liquids during reprocessing procedures, make sure that after closing
the battery compartment no gaps are visible in the housing between the battery cover and the
lower housing.
1
2
+
-
-
+
3
4
41
7CLEANING, DISINFECTION AND STERILIZATION
7.1General
You must consult the special VDW.CONNECT Locate® Cleaning, Disinfection and Sterilization
Guide before proceeding with cleaning disinfection and sterilization.
Disposable (single-use) VDW.CONNECT Locate® Barrier Sleeves do not replace reprocessing
of dental equipment / accessories or endodontic instruments. Always reprocess VDW.CONNECT
Locate® Apex Locator.
The following parts shall be reprocessed before and after each treatment.
The reprocessing procedure might damage the (USB) plugs connected to the VDW.CONNECT
Locate® Apex Locator / VDW.CONNECT Drive® Handpiece ports. They are unlikely to be
contaminated during normal (plugged) use. Avoid reprocessing these plugs.
PartReprocessing steps
•VDW.CONNECT Locate® Apex Locator
•VDW.CONNECT Locate® Measurement Cable
•VDW.CONNECT Drive
®
Measurement Cable
Clean and disinfect.
Do not sterilize.
•VDW.CONNECT Locate® File Clip
•VDW.CONNECT Locate® Lip Clip
VDW GmbH accepts no liability in the event that these instructions are disregarded or
unvalidated processes are used to reprocess the accessories.
Use only properly maintained equipment and approved materials according to national laws and/
or regulations.
Always wear protective clothes for your own safety (gloves, mask, glasses and a waterproof
gown) for the reprocessing process.
Clean, Disinfect and Sterilize.
7.2Cleaning Procedure
Cleaning is intended to remove all the visible soiling (e.g. organic and inorganic) from the surface
of the device and accessories so that the subsequent disinfection process can be efficient.
N°Cleaning steps
42
VDW.CONNECT Locate® Apex Locator (and clamp, if used), VDW.CONNECT Locate®
Measurement Cable and VDW.CONNECT Drive
Switch off VDW.CONNECT Locate® Apex Locator, then remove and discard the
A
single-use VDW.CONNECT Locate® Disposable Barrier Sleeve (sleeves if the
VDW.CONNECT Locate
Pay attention that no liquid enters the device connecting port, e.g. by keeping the
cable plugged-in.
Clean the devices by using a single-use wipe or soft cloth soaked in cleaning fluid
B
containing Alcohol Quaternary Ammonium active ingredients (e.g. CaviWipes™ or
CaviCide™).
For the VDW.CONNECT Locate® Apex Locator, clean the grooves and device
screws by using a clean brush with soft bristle (e.g. nylon) slightly soaked in cleaning
C
solution containing Alcohol Quaternary Ammonium active ingredients (e.g.
CaviWipes™ or CaviCides™).
Continue steps B and C until no debris appear on the wipe / brush. Replace the wipe
D
and clean the brush as necessary.
VDW.CONNECT Locate® File Clip and Lip Clip
Pre-treatment of the file clip and lip clip – for personal protection
After use, place the file clip and lip clip directly into a dish filled with a suitable
cleaning and disinfection solution (Cidezyme, Johnson & Johnson Medical, 0.8% for
E
at least 1 minute, maximum 2 hours) for cleaning, pre-disinfection and interim
storage. Make sure that the devices are fully immersed.
®
clamp is used).
®
Measurement Cable
43
N°Cleaning steps
Use clean, soft brushes or a clean, soft cloth or single-use wipe to remove
F
contamination and remnants manually. Do not use metal brushes or wire wool.
Squeeze the VDW.CONNECT Locate® File Clip open and release it 5 times during
cleaning to allow the inner parts to be cleaned effectively. Note that disinfectant used
G
for pre-treatment is only for personal protection and does not obviate the need for
disinfection once cleaning has been completed. Pre-treatment should never be
omitted.
Rinse the devices under flowing sterile, deionized water at least 3 times for 1 minute
H
each time, in order to remove all visible traces of contamination and remnants.
Check that no visible contamination or remnants remain, and repeat E – H if
I
necessary. Replace the wipe and clean the brush when necessary.
Cleaning of the file clip and lip clip – Cleaning and ultrasonic bath
Place the pre-treated VDW.CONNECT Locate® File Clip and Lip Clip into the
cleaning bath for the prescribed contact time (e.g. CIDEZYME
Detergent Solutions, Johnson & Johnson Medical, 0.8% for 1 minute); make sure
J
that the devices are fully immersed (if necessary, use a soft brush to make sure they
®
, ENZOL® Enzymatic
remain fully immersed). Squeeze the file clip open and release it 5 times during
cleaning to allow the inner parts to be cleaned effectively.
Remove the devices from the cleaning bath and rinse them thoroughly at least 3
K
times for 1 minute each time with sterile, deionized water; while rinsing, squeeze the
file clip open and release it 5 times.
Place the devices in an ultrasonic bath with a cleaning agent (e.g. CIDEZYME®,
ENZOL® Enzymatic Detergent Solutions, Johnson & Johnson Medical, 0.8% for
L
20 minutes); make sure that the devices are fully immersed (if necessary, use a soft
brush to make sure they remain fully immersed). Squeeze the file clip open and
release it 5 times during cleaning to allow the inner parts to be cleaned effectively.
Then, remove the devices from the ultrasonic bath and rinse them thoroughly at least
M
3 times for 1 minute each time with sterile, deionized water; while rinsing, squeeze
the file clip open and release it 5 times.
7.3Disinfection
Disinfection shall be done immediately after cleaning using the following procedure:
N°Disinfection steps
44
VDW.CONNECT Locate® Apex Locator (and clamp, if used), VDW.CONNECT Locate®
®
Measurement Cable and VDW.CONNECT Drive
Measurement Cable
Disinfect the surface of devices paying special attention to grooves and screws (and
clamp, if used). Pay attention that no disinfectant enters the VDW.CONNECT
Locate
®
Apex Locator connector, e.g. by having the cable plugged-in. Use a single-
A
use wipe or soft clean cloth lightly soaked in disinfectant containing Alcohol
Quaternary Ammonium active ingredients (e.g. CaviWipes™ or CaviCide™).
Remove all disinfectant residue. No visible impurities or liquid should remain on the
B
devices after disinfection.
Insert VDW.CONNECT Locate® Apex Locator in a new and unused single-use
C
VDW.CONNECT Locate
®
Disposable Barrier Sleeve before starting the next
treatment.
If the device is not being used immediately after the reprocessing process, keep it in
D
a safe and clean environment and follow the instructions for storage and transport.
VDW.CONNECT Locate® File Clip and Lip Clip
Once the devices have been cleaned and inspected, place them into the disinfection
bath for the prescribed contact time (Cidex OPA, Johnson & Johnson Medical, 100%
E
for 20 minutes); the devices must be fully immersed in the solution. Squeeze the file
clip open and release it 5 times during disinfection to allow the inner parts to be
disinfected effectively.
Remove the devices from the disinfection bath and rinse them off thoroughly at least
F
5 times for 1 minute each time with water; when you rinse them, squeeze the file clip
open and release it 5 times.
Dry the file clip and lip clip with oil-free, filtered compressed air and leave them to dry
G
further in a clean place for at least 20 minutes.
Once the devices are dry, inspect them to ensure they are clean, dry and
H
undamaged.
Use only disinfectants containing Alcohol Quaternary Ammonium active ingredients (e.g.
®
CaviWipes™ or CaviCides™), except for the VDW.CONNECT Locate
File Clip and Lip Clip
(see above). Use only approved disinfectants compatible with the device and accessories (see 8
Technical Characteristics).
Leave the disinfectant to act in accordance to manufacturer’s instructions. Avoid leaving the
disinfectant in contact with the device surfaces for a time longer than that recommended by the
manufacturer of the disinfectant product. It can result in damage to the device.
45
7.4Sterilization of the File Clip and Lip Clip
The sterilization process applies only to the lip clip and file clip. The maximum number of
sterilization cycles is:
•File clip: 120;
•Lip clip: 120.
Proceed with the sterilization as follows:
N°Sterilization steps for file clip and lip clip
Pack and seal accessories into single-use, undamaged sterilization packaging (single-layer
packaging) that complies with the following requirements:
A
•DIN EN ISO / ANSI AAMI ISO 11607;
•Suitable for steam sterilization (withstands temperatures of up to 142 °C or more, sufficient
vapor permeability).
Sterilize the file clip and lip clip by Pre-Vacuum Displacement (3 times pre-vacuum < 60 mbar).
•At 134 °C for 3 minutes, drying time 20min.
Please note:
B
•Do not exceed a sterilization temperature of 135°C and a sterilization duration of 18min;
•Always observe the operating instructions of the manufacturer for your sterilizer, especially
about the loading weight, the operating time and functional testing;
•Use steam sterilizers in accordance with DIN EN 13060 or DIN EN 285, ANSI / AAMI ST 79.
Inspect the file clip and the lip clip before use after they have been reprocessed. If visible signs of
moisture are present (damp spots on sterile packaging, pooled water in the load) at the end of
the sterilization cycle, repackage and re-sterilize using a longer drying time. Dispose of any
C
defective accessories immediately. If the press-release mechanism of the file clip does not
function properly, repeat the complete reprocessing procedure. Discard the file clip if the pressrelease mechanism does not work properly.
The sterilization process must be validated in accordance with DIN EN ISO 17665 (valid installation and operational
qualification (IQ and OQ) and product-specific performance qualification (PQ)).
This equipment does not have Essential Performance (Ref.
IEC 60601-1:2005 + A1:2012)
Medical Device ClassClass IIa, Rule 10, according to MDD 93/42/EEC (Annex IX)
49
9VDW.CONNECT LOCATE® APEX LOCATOR ERROR
CODES
If a failure occurs on VDW.CONNECT Locate® Apex Locator, the nature of the failure is given by
a combination of colors and sounds as described in the following table.
StateFailureSolution
Charge VDW.CONNECT Locate® Apex Locator
R
Sound: Audio signal
R
Critical / Device
unrecoverable
Device warning recoverable
(see 6.12 Battery Care).
It the problem persists, return the instrument to your service
center.
Connect to VDW.CONNECT® App for further information /
instructions.
See VDW.CONNECT® App User Manual - 6 Step by
Step instructions.
Sound: triple beep
R
Sound: off
R
Sound: triple beep
Empty battery
Bluetooth®
module failure
Charge VDW.CONNECT Locate® Apex Locator
(see 6.12 Battery Care).
Connect to VDW.CONNECT® App for further information /
instructions.
See VDW.CONNECT® App User Manual - 6 Step by
Step instructions.
50
StateFailureSolution
R
Sound: triple beep
R
Sound: triple beep
R
Sound: triple beep
Battery charge
not sufficient to
perform a
firmware upgrade
General firmware
failure
LED failure
Connect to VDW.CONNECT® App for further information /
instructions.
®
See VDW.CONNECT
App User Manual - 6 Step by
Step instructions.
Connect to VDW.CONNECT® App for further information /
instructions.
®
See VDW.CONNECT
App User Manual - 6 Step by
Step instructions.
If NONE of the LEDs lights up in white color during boot-up and
a triple beep sound is heard, there is a LED failure.
Return the instrument to your service center.
10TROUBLESHOOTING
If VDW.CONNECT Locate® Apex Locator is not operating properly, check the possible cause
using the following checklist. If you are unable to solve the problem by referring to the checklist,
contact VDW GmbH Customer Service.
Also refer to the error codes, which may be indicated on VDW.CONNECT Locate® Apex Locator,
as described in section 9 VDW.CONNECT Locate® Apex Locator Error Codes.
Refer to VDW.CONNECT Drive® Directions For Use for any troubleshooting related to
VDW.CONNECT Drive® Handpiece in combined mode.
51
Refer to VDW.CONNECT® App User Manual for any troubleshooting related to Bluetooth
connection.
VDW GmbH requires that you reprocess the product before shipping it to a service center.
To remedy certain problems, it is necessary to perform a factory reset (delete user settings and
restore factory settings) on the device. In such case proceed as described below.
To perform a factory reset:
N°KeyAction
Hold down the ON/OFF/Standby button for approximately 20 seconds.
A
BRelease the ON/OFF/Standby button.
The factory reset takes 20 seconds in total. After the first 10 seconds, there is a
brief white flash of all LEDs. At the end of the process, another white flash and a
long beep indicate a successful factory reset.
®
52
VDW.CONNECT Locate® Apex Locator does not switch on or it shuts off during treatment.
Possible causesSolutions
•The battery is completely empty (all LEDs are
OFF).
•The charger is not connected properly into the
power socket.
•The battery is wrongly inserted.
•The original battery has been replaced with a
non-original one.
•ON/OFF/Standby button malfunction.•Try pressing the ON/OFF/Standby button
•Electronic malfunction.•Perform a factory reset.
•Connect the device to the charger and charge
the device.
•Ensure that the charger is plugged in correctly
and the power socket is under tension/switched
ON (if applicable).
•Ensure that the original adapter is being used.
•Open the battery compartment and check the
battery orientation.
several times.
•Replace the batteries with a new original battery.
•Contact VDW GmbH Customer Service.
•Send the device to VDW GmbH Customer
Service.
53
No feedback from VDW.CONNECT Locate® Apex Locator during treatment.
Possible causesSolutions
•The battery is completely empty (all LEDs are
OFF).
•The cables are damaged or not connected
correctly.
•Non-original cables are used.
•Connect the device to the charger and charge
the device.
•Ensure that original cables have been used.
•Ensure that the cables are undamaged and
connected correctly.
•Perform closed-loop to test to check functionality
of cables and device test to check functionality
of device.
•If in combined mode, ensure that the original
VDW.CONNECT Drive
used.
•If in combined mode, ensure that the endodontic
file has been correctly inserted into the
VDW.CONNECT Drive
•Clean the lip clip and the file clip according to the
procedure.
•If all else fails.•Perform a factory reset.
If cables are damaged or if the ClosedLoop test fails after a second attempt,
DO NOT USE VDW.CONNECT Locate
Apex Locator with the damaged cables
and contact your VDW GmbH
Customer Service.
®
Contra-angle is being
®
Contra-angle.
®
•If requested to send the device and/or
components to VDW GmbH Customer Service,
reprocess the device, file clip, lip clip and cables
before sending them back to customer services
(Refer to Section 7 Cleaning, Disinfection and
Sterilization).
54
No sound or no light signal emitted during treatment.
Possible causesSolutions
•The battery is completely empty (all LEDs are
OFF).
•Connect the device to the charger and charge
the device.
•Electronic malfunction.•Perform a factory reset.
•Send the device to VDW GmbH Customer
Service.
Endodontic file progression indication is unstable or frozen.
Possible causesSolutions
•Calcified or obliterated root canal.
•Extremely dry root canal.
•Old root filling residues creating a blockage due
to insufficient removal of intra-canal medicament
(i.e. calcium hydroxide).
•Check comparative X-rays for more information
and repeat reading on VDW.CONNECT
Locate® Apex Locator to verify the result.
•Rinse with NaOCl solution. Dry access cavity
with cotton ball / air blow.
•Do a comparative X-ray, and fully remove old
gutta-percha / intra-canal medicament.
•Endodontic file too narrow for large root canal.
•Large lateral canal.
•Select a larger endodontic file.
•Block the external conductive path.
•Deep caries provide a conductive path outside
the canal.
•Perforation.
•VDW.CONNECT Locate® Lip Clip incorrectly
attached to patient mouth. Insufficient contact
between the lip clip and the oral mucosa.
•Endodontic file incorrectly connected to the file
clip (i.e. connected to the metal shaft).
®
•Soiled VDW.CONNECT Locate
File Clip.
•Place the lip clip in the patient mouth and ensure
a good contact between the mucosa and lip clip
(place the lip clip in the labial angle opposite to
the tooth to be treated).
•Ensure that the endodontic file and the file clip
are properly in contact with each other.
•Perform a closed-loop test.
•Clean the file clip according to the reprocessing
instructions in 7 Cleaning, Disinfection and
Sterilization.
•Defective connection cable.•Check cables and sockets for visible damage
and ensure the connections are correct.
•Perform a closed-loop test to test cable
functionality.
•LEDs damaged or not working properly.
•Electronic malfunction.
•Perform a factory reset.
•Contact VDW GmbH Customer Service.
55
Endodontic file progression indication is too sensitive (i.e. it displays the relevant position for
working length determination too quickly).
Possible causesSolutions
•Short-circuit due to excess fluid in the pulp
chamber (rinsing solution, saliva, blood).
•Dry the access cavity using an air blower or a
cotton ball. In case of excess bleeding wait until
it has stopped.
•If heavy bleeding is observed, wait until bleeding
has been successfully stopped.
•Direct contact between endodontic file and
gingiva / gingival proliferation (i.e. in cases with
fractured metal crowns).
•Direct contact between endodontic file and metal
restoration work (crowns, parapulpal pins;
amalgam fillings).
•For isolation:
- Adequate tooth build-up;
- Crown lengthening;
- Electro-cauterisation.
•Use a dental dam. Isolate the endodontic file by
placing 2-3 silicone stoppers on it or insert the
endodontic file in a small polyvinyl tube before
use.
•Use the protective VDW.CONNECT Drive®
Handpiece Sleeve on the VDW.CONNECT
®
Handpiece (if in combined mode).
Drive
The use of VDW.CONNECT Drive
Handpiece Sleeve on the
®
VDW.CONNECT Drive
Handpiece
(when working in combined mode)
is mandatory to ensure the correct
functioning of the system in
combined mode.
•Lateral root canal.
•Immature canal with large apex.
•Carefully widen the access cavity. If necessary
add flow composites to isolate.
•Repeat length determination process.
®
56
Battery does not recharge when the power supply is connected
(the battery LED remains OFF when power supply is connected).
Possible causesSolutions
•Bad charger connection.
•Power socket is not under tension/ switched off
•Disconnect the charger from the device and
reconnect it again.
(if applicable).
•Charger malfunction.•Replace the charger with an original charger.
Device test and quick closed-loop test do not start.
Possible causesSolutions
•VDW.CONNECT Locate® External tester is not
working properly.
•Perform a factory reset.
•Contact VDW GmbH Customer Service.
•Electronic malfunction.
VDW.CONNECT Drive® Handpiece undetected in combined mode.
Possible causesSolutions
•VDW.CONNECT Drive® Measurement Cable or
one of the connectors are damaged.
•Check the connection of the cable between
®
VDW.CONNECT Locate
VDW.CONNECT Drive
Apex Locator and the
®
Handpiece.
•Perform a closed- loop test in combined mode.
•Firmware of the VDW.CONNECT Drive®
Handpiece not up to date.
•Check if VDW.CONNECT Drive® Handpiece
firmware is the latest version available (updated
for VDW.CONNECT Locate
®
Apex Locator
feature).
•Electronic malfunction.•Perform a factory reset.
•Contact VDW GmbH Customer Service.
57
Apical Reverse function cannot be switched on.
Possible causesSolutions
•No connection between VDW.CONNECT
Locate® Apex Locator and VDW.CONNECT
®
Handpiece.
Drive
•Wrong cable connection.
•Cable or connector damaged.
®
•VDW.CONNECT Drive
Handpiece not
connected to VDW.CONNECT® App.
•Ensure proper connection of the cables between
VDW.CONNECT Locate® Apex Locator and the
VDW.CONNECT Drive
•Ensure that all cables are original and
undamaged.
•Connect VDW.CONNECT Drive
VDW.CONNECT® App to benefit from Apical
Reverse function.
•Firmware of VDW.CONNECT Drive® Handpiece
not up to date.
•Ensure that VDW.CONNECT Drive® Handpiece
firmware is the latest version available (updated
for VDW.CONNECT Locate
feature).
•Apical Reverse setting on the VDW.CONNECT®
App accidentally set to OFF.
•Switch ON the feature on the VDW.CONNECT®
App and retry.
•Electronic malfunction.•Perform a factory reset.
•Contact VDW GmbH Customer Service.
VDW.CONNECT Locate® Apex Locator is overheating.
®
Handpiece.
®
®
Handpiece to
Apex Locator
Possible causesSolutions
•Electronic malfunction.
•VDW.CONNECT Locate
overheated and it has switched off.
®
Apex Locator has
•Allow the device to cool down and start again.
•Check for possible leakage from
VDW.CONNECT Locate
from defective battery). In case of suspected
leakage, refer to Section 3 Warnings.
®
Apex Locator (liquids
11WARRANTY
VDW GmbH offers customers the following product warranty:
•VDW GmbH guarantees proper manufacturing of the product, the use of top quality
materials, performance of all required tests and compliance with to all applicable relevant
laws and regulations relating to the product. The full functionality of VDW.CONNECT
Locate® Apex Locator is covered by a 12-month warranty. Cables, accessories and battery
are covered by a 6-month warranty. Warranty period begins on the date of equipment
purchase. The customer is only entitled to warranty services within the warranty period and
under the condition that VDW GmbH has been informed of the defect within a one-month
period from the date of discovery of the defect.
ComponentWarranty period
58
VDW.CONNECT Locate® Apex Locator unit
VDW.CONNECT Locate® Charger
VDW.CONNECT Locate® Lip Clip
VDW.CONNECT Locate® File Clip
VDW.CONNECT Locate® External tester
VDW.CONNECT Locate® Measurement Cable
VDW.CONNECT Drive® Measurement Cable
VDW.CONNECT Locate® Clamp
•This warranty covers only the exchange or repair of individual components or parts affected
by manufacturing defects. The cost for personnel to be dispatched for purposes of technical
assistance from the dealer to the customer, and /or packaging expenses of the customer will
not be covered by VDW GmbH. Any customer claims beyond the realm of repair, such as
claims for damages, will not be covered. This warranty does not include any compensation
for direct or indirect personal injuries or material damage of any kind. The customer will not
be entitled to damages covering downtime of the equipment.
•The warranty does not extend to damage for which evidence can be provided by VDW
GmbH that such damage has been caused by negligence on the part of the user as related
to Directions For Use, particularly during battery charging or battery replacement. The
warranty explicitly excludes any defects arising on the basis of:
- damage occurring during transportation to VDW GmbH for the purposes of repair;
- damage occurring through environmental events such as lightning strikes, fire and /
or humidity. This warranty shall automatically become cancelled and void if the
product has been improperly repaired, modified or manipulated in any manner by the
user, by non-authorized persons or by third party personnel.
•Legal claims, such as under product liability law remain unaffected.
12 months
6 months
6 months
6 months
6 months
6 months
6 months
59
12DISPOSAL OF THE PRODUCT
Comply with your national and regional regulations, guidelines and requirements for the disposal
of end-of-life electrical equipment and batteries.
Make sure the product or battery are not mixed with other types of waste when being disposed of.
Prior to disassembly and disposal, your device must not be contaminated and must have been
completely reprocessed (Refer to Section 7 Cleaning, Disinfection and Sterilization).
13IDENTIFICATION OF SYMBOLS
13.1Normative Symbols
SymbolIdentification
SN
Serial number
Batch code
Catalogue number
Refer to Instruction Manual / Booklet
Consult Instructions for Use
Electronic Instructions for Use (relevant information for use of the product is available in
electronic format from vdw-dental.com).
Manufacturer
Date of manufacture
Class II equipment
Type BF applied part
Direct current (connection for power supply)
Alternating current
60
䢵
䢶
SymbolIdentification
Sterilizable in a steam sterilizer (autoclave) at temperature specified
Opened packages are not replaced
Temperature limit
Humidity limitation
Keep dry
Fragile; handle with care
Accessory
Use by date
xxx
C
°
Do not steam sterilize
Do not re-use
Caution
Bluetooth
P
Plastic
®
Non-sterile
CE Mark
0123
Recycling: PLEASE DO NOT THROW AWAY! This product and all its components must
absolutely be recycled through your distributor
61
Appendix A: ELECTROMAGNETIC EMISSIONS AND
IMMUNITY
Refer to the section 3 Warnings to verify any special care to exercise before starting to use the
device.
EMC (general)
VDW.CONNECT Locate® Apex Locator is to be used in a dental practice, in a hospital or a private
practice.
VDW.CONNECT Locate® Apex Locator is EMC-tested in conformity with the requirements of IEC
60601-1-2:2007 3rd (see the following tables) and IEC 60601-1-2:2014 4th Edition (according to
clause 7 and 8.9, tables 4 to 9). VDW.CONNECT Locate® Apex Locator is a medical device that
requires special safety precautions and must be installed and placed in operation in accordance
with the below EMC information.
Compliant Cables and Accessories
The use of accessories and cables other than specified by the manufacturer may result in increased emissions or decreased immunity of this product.
Information about charger
The charger must be used while the VDW.CONNECT Locate® Apex Locator is in charging mode
only. VDW.CONNECT Locate® Apex Locator cannot be used for treatments when charging.
The charger for VDW.CONNECT Locate® Apex Locator complies with all relevant standards for
battery charger devices.
Information about radio communications
•Bluetooth® LE 4.0 Communication.
•Frequency band: 2.400 to 2.4835 GHz.
•The effective radiated power is below 10 mW.
•Europe: EU EN 300 328 V2.1.1.
62
This device complies with part 15C of the FCC:
Operation is subject to the following two conditions:
•This device may not cause harmful interference;
•This device must accept any interference received, including interference that may cause
undesired operation of the device.
This Class B digital apparatus complies with Canadian ICES-0003.
Compliance with Emission and Immunity Standard
This section provides guidance on compliance to Edition 3 (2007) and 4 (2014) of standard IEC
60601-1-2. Please refer to the relevant subsection according to the applicable regulation in your
territory
Compliance to IEC 60601-1-2:2007
Guidance and manufacturer’s declaration - electromagnetic emissions
VDW.CONNECT Locate® Apex Locator is intended for use in the electromagnetic environment specified
below. The customer or the user of VDW.CONNECT Locate
such an environment.
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Group 1
Class B
Not applicable
Not applicable
VDW.CONNECT Locate® Apex Locator uses RF energy
only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in
nearby electronic equipment.
VDW.CONNECT Locate® Apex Locator is suitable for use
in all establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
®
Apex Locator should ensure that it is used in
The following tables are guidelines according to the 3rd edition of the medical standard IEC
60601-1-2.
63
Guidance and manufacturer's declaration - electromagnetic immunity
VDW.CONNECT Locate® Apex Locator is intended for use in the electromagnetic environment specified
below. The customer or the user of VDW.CONNECT Locate
®
Apex Locator should ensure that it is used in
such an environment.
Immunity test
standard
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IEC 60601 test levelCompliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input / output
lines
± 1 kV line(s) to
Line(s)mode
± 2 kV line(s) to earth
± 8 kV contact
± 15 kV air
± 2 kV for power supply
lines
Not applicable (no input/
output lines)
± 1 kV differential mode
Not applicable (no earth)
Electromagnetic environment -
guidance
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
0% UT
(> 95% dip in UT) for 0.5
< 5% UT
(> 95% dip in UT) for 0.5
cycle
cycle
(at 0, 45, 90, 135,
180, 225 and 315°
phase angle)
Voltage dips,
short interruptions
and voltage
variations on
power supply
lines
IEC 61000-4-11
< 40% UT
(60% dip in UT) for 5
cycles
70% UT
(30% dip in UT) for 25
cycles.
0% UT
(> 95% dip in UT) for 1
cycle
(at 0 phase angle)
70% UT
(30% dip in UT) for 25/30
cycles (at 0 phase angle)
Mains power quality should be
that of a typical commercial or
hospital environment.
< 5% UT
(> 95% dip in UT) for 5 s
0% UT
(> 95% dip in UT) for
250/300 cycles (at 0
phase angle)
64
Power frequency
(50 / 60 Hz)
magnetic field
IEC 61000-4-8
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 A/m30 A/m
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.7 GHz
Power frequency magnetic fields
should be at levels characteristic
of a location in a typical
commercial or hospital
environment.
Portable and mobile RF
communications equipment
should be used no closer to any
part of the VDW.CONNECT
Locate® Apex Locator, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance
d1.2P=
150 kHz to 80 MHz
d1.2P=
80 MHz to 800 MHz
d2.3P=
800 MHz to 2.5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey,
should be less than the
compliance level in each
frequency range.
b
a
Interference may occur in the
vicinity of equipment marked
with the following symbol:
65
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular /cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
VDW.CONNECT Locate® Apex Locator is used exceeds the applicable RF compliance level above, the
VDW.CONNECT Locate
®
Apex Locator should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the
VDW.CONNECT Locate
®
Apex Locator.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
66
Recommended separation distances between portable and mobile RF communications equipment and the
VDW.CONNECT Locate® Apex Locator
The VDW.CONNECT Locate® Apex Locator is intended for use in the electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the VDW.CONNECT Locate
®
Apex Locator can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the VDW.CONNECT Locate® Apex
Locator as recommended below, according to the maximum output power of the communication
equipment.
Rated maximum
output power of
transmitter
(W)
150 kHz to 80 MHz80 MHz to 800 MHz800 MHz to 2.5 GHz
d1.2P=d1.2P=d2.3P=
Separation distance according to frequency of transmitter
(m)
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
10012.012.023.0
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P
is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Note 3: An additional factor of 10/3 is used in calculating the recommended separation distance to
decrease the likelihood that mobile / portable communications equipment could cause interference if it is
inadvertently brought into patient areas.
67
Compliance with IEC 60601-1-2:2014
Use of VDW.CONNECT Locate® Apex Locator adjacent to, or stacked with other equipment
should be avoided because it could result in improper operation. If such use is necessary,
VDW.CONNECT Locate
that they are operating normally.
WARNING: Portable RF communication equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm from any part of VDW.CONNECT
Locate® Apex Locator. Otherwise, degradation of the performance of this equipment could result.
VDW.CONNECT Locate® Apex Locator does not have Essential Performance.
Guidance and manufacturer’s declaration - electromagnetic emissions
VDW.CONNECT Locate® Apex Locator is intended for use in the electromagnetic environment specified
below. The customer or the user of VDW.CONNECT Locate® Apex Locator should ensure that it is used in
such an environment.
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Group 1
Class B
Not applicable-
Not applicable-
®
Apex Locator and the other equipment should be observed to verify
VDW.CONNECT Locate® Apex Locator generates and
uses conductively coupled radio-frequency energy only for
its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
VDW.CONNECT Locate® Apex Locator is suitable for use
in establishments directly connected to low voltage power
supply network which supplies buildings used for domestic
purposes.
a) The interface between the PATIENT physiological signal simulation, if used, and the ME EQUIPMENT or
ME SYSTEM shall be located within 0,1 m of the vertical plane of the uniform field area in one orientation of
the ME EQUIPMENT or ME SYSTEM.
b) ME EQUIPMENT and ME SYSTEMS that intentionally receive RF electromagnetic energy for the
purpose of their operation shall be tested at the frequency of reception. Testing may be performed at other
modulation frequencies identified by the RISK MANAGEMENT PROCESS. This test assesses the BASIC
SAFETY and ESSENTIAL PERFORMANCE of an intentional receiver when an ambient signal is in the
passband. It is understood that the receiver might not achieve normal reception during the test.
c) Testing may be performed at other modulation frequencies identified by the RISK MANAGEMENT
PROCESS.
d) Applies only to ME EQUIPMENT and ME SYSTEMS with magnetically sensitive components or circuitry.
e) During the test, the ME EQUIPMENT or ME SYSTEM may be powered at any NOMINAL input voltage,
but with the same frequency as the test signal (see Table 1).
f) Before modulation is applied.
g) This test level assumes a minimum distance between the ME EQUIPMENT or ME SYSTEM and sources
of power frequency magnetic field of at least 15 cm. If the RISK ANALYSIS shows that the ME
EQUIPMENT or ME SYSTEM will be used closer than 15 cm to sources of power frequency magnetic field,
the IMMUNITY TEST LEVEL shall be adjusted as appropriate for the minimum expected distance.
Input a.c. power port
69
PHENOMENON
Electrical fast
transients / bursts
a) l) o)
a) b) j) o)
Surges
Line-to-line
a) b) j) k) o)
Surges
Line-to-ground
Conducted
disturbances induced
by RF fields
Voltage dips
c) d) o)
f) p) r)
Voltage interruptions
f) i) o) r)
BASIC EMC
IMMUNITY TEST LEVELS
STANDARD OR TEST
METHOD
PROFESSIONAL
HEALTHCARE FACILITY
ENVIRONMENT
IEC 61000-4-4
± 2 kV
100 kHz repetition frequency
Pass
IEC 61000-4-5± 0.5 kV, ±1 kVPass
IEC 61000-4-5± 0.5 kV, ± 1 kV, ± 2 kVPass
m)
3 V
0.15 MHz - 80 MHz
IEC 61000-4-6
IEC 61000-4-11
m)
in ISM bands between
6 V
0.15 MHz and 80 MHz
80% AM at 1 kHz
0% UT; 0.5 cycle
n)
e)
g)
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
q)
Pass
Pass
0% UT; 1 cycle and
h)
70% UT; 25/30 cycles
Pass
Single phase: at 0°
IEC 61000-4-11Pass
TEST RESULTS
70
a) The test may be performed at any one power input voltage within the ME EQUIPMENT or ME SYSTEM
RATED voltage range. If the ME EQUIPMENT or ME SYSTEM is tested at one power input voltage, it is not
necessary to re-test at additional voltages.
b) All ME EQUIPMENT and ME SYSTEM cables are attached during the test.
c) Calibration for current injection clamps shall be performed in a 150 Ω system.
d) If the frequency stepping skips over an ISM or amateur band, as applicable, an additional test frequency
shall be used in the ISM or amateur radio band. This applies to each ISM and amateur radio band within the
specified frequency range.
e) Testing may be performed at other modulation frequencies identified by the RISK MANAGEMENT
PROCESS.
f) ME EQUIPMENT and ME SYSTEMS with a d.c. power input intended for use with a.c.-to-d.c. converters
shall be tested using a converter that meets the specifications of the MANUFACTURER of the ME
EQUIPMENT or ME SYSTEM. The IMMUNITY TEST LEVELS are applied to the a.c. power input of the
converter.
g) Applicable only to ME EQUIPMENT and ME SYSTEMS connected to single-phase a.c. mains.
h) E.g. 10/12 means 10 periods at 50 Hz or 12 periods at 60 Hz.
i) ME EQUIPMENT and ME SYSTEMS with RATED input current greater than 16 A / phase shall be
interrupted once for 250/300 cycles at any angle and at all phases at the same time (if applicable). ME
EQUIPMENT and ME SYSTEMS with battery backup shall resume line power operation after the test. For
ME EQUIPMENT and ME SYSTEMS with RATED input current not exceeding 16 A, all phases shall be
interrupted simultaneously.
j) ME EQUIPMENT and ME SYSTEMS that do not have a surge protection device in the primary power
circuit may be tested only at ± 2 kV line(s) to earth and ± 1 kV line(s) to line(s).
k) Not applicable to CLASS II ME EQUIPMENT and ME SYSTEMS.
l) Direct coupling shall be used.
m) r.m.s., before modulation is applied.
n) The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to
6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The
amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3
MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17
MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
o) Applicable to ME EQUIPMENT and ME SYSTEMS with RATED input current less than or equal to 16 A /
phase and ME EQUIPMENT and ME SYSTEMS with RATED input current greater than 16 A / phase.
p) Applicable to ME EQUIPMENT and ME SYSTEMS with RATED input current less than or equal to 16 A /
phase.
q) At some phase angles, applying this test to ME EQUIPMENT with transformer mains power input might
cause an overcurrent protection device to open. This can occur due to magnetic flux saturation of the
transformer core after the voltage dip. If this occurs, the ME EQUIPMENT or ME SYSTEM shall provide
BASIC SAFETY during and after the test.
r) For ME EQUIPMENT and ME SYSTEMS that have multiple voltage settings or auto ranging voltage
capability, the test shall be performed at the minimum and maximum RATED input voltage. ME
EQUIPMENT and ME SYSTEMS with a RATED input voltage range of less than 25 % of the highest
RATED input voltage shall be tested at one RATED input voltage within the range. See Table 1 Note c) for
examples calculations.