Varian XGS-600 Instruction Manual

Page 1
XGS-600™ Gauge
vacuum technologies
Controller
/22/08
DRAFT 9
Manual No. 699908400
Revision A
September 2008
Page 2
XGS-600 Gauge Controller
Copyright 2008
Varian, Inc.
Page 3
XGS-600 Gauge Controller
Warranty
Products manufactured by Seller are warranted against defects in materials and workmanship for twelve (12) months from date of shipment thereof to Customer, and Seller’s liability under valid warranty claims is limited, at the option of Seller, to repair, to replace, or refund of an equitable portion of the purchase price of the Product. Items expendable in normal use are not covered by this warranty. All warranty replacement or repair of parts shall be limited to equipment malfunctions which, in the sole opinion of Seller , a re due or traceable to defects in original materials or workmanship. All obligations of Seller under this warranty shall cease in the event of abuse, accident, alteration, misuse, or neglect of the equipment. In-warranty repaired or replaced parts are warranted only for the remaining unexpired portion of the original warranty period applicable to the repaired or replaced parts. After expiration of the applicable warranty period, Customer shall be charged at the then current prices for parts, labor, and transportation.
Reasonable care must be used to avoid hazards. Seller expressly disclaims responsibility for loss or damage caused by use of its Products other than in accordance with proper operating procedures.
Except as stated herein, Seller makes no warranty, express or implied (either in fact or by operation of law), statutory or otherwise; and, except as stated herein, Seller shall have no liability under any war­ranty , express or implied (ei ther in fact or by operation of law), statutory or otherwise. Statements made by any person, including representatives of Seller, which are inconsistent or in conflict with the terms of this warranty shall not be binding upon Seller unless reduced to writing and approved by an officer of Seller.
22/08
Warranty Replacement and Adjustment
All claims under warranty must be made promptly after occurrence of circumstances giving rise thereto, and must be received within the applicable warranty period by Seller or its authorized representative. Such claims should include the Product serial number, the date of shipment, and a full description of the circumstances giving rise to the claim. Before any Products are returned for repair and/or adjust­ment, written authorization from Seller or its authorized representative for the return and instructions as to how and where these Products should be returned must be obtained. Any Product returned to Seller for examination shall be prepaid via the means of transportation indicated as acceptable by Seller. Seller reserves the right to reject any warranty claim not promptly reported and any warranty claim on any item that has been altered or has been returned by non-acceptable means of transportation. When any Product is returned for examination and inspection, or for any other reason, Customer shall be responsible for all damage resulting from improper packing or handling, and for loss in transit, notwith­standing any defect or non-conformity in the Product. In all cases, Seller has the sole responsibility for determining the cause and nature of failure, and Seller’s determination with regard thereto shall be final.
If it is found that Seller’s Product has been returned without cause and is still serviceable, Customer will be notified and the Product returned at Customer’s expense; in addition, a charge for testing and exam­ination may be made on Products so returned.
3/1/00
DRAFT 9/
Page 4
XGS-600 Gauge Controller
This page intentionally left blank.
Page 5
XGS-600 Gauge Controller
Contents
Preface .............................................................................................................................................. v
Hazard and Safety Information ................................................................................................v
EMC Warnings ....................................................................................................................... vi
Introduction ...................................................................................................................................1-1
Backwards Compatibility ......................................................................................................1-1
Part Number Scheme............................................................................................................1-1
XGS-600 Front and Back Description....................................................................................1-2
Unit Dimensions...................................................................................................................1-3
Installation .....................................................................................................................................2-1
Connect the Unit...................................................................................................................2-1
Operating Instructions ................................................................................................................. 3-1
Operational Screens..............................................................................................................3-1
Service and Troubleshooting ...................................................................................................... 4-1
Troubleshooting....................................................................................................................4-5
Appendix A. XGS-600 Gauge Specifications ............................................................................. A-1
Instrument Specifications ......................................................................................................A-1
Board Specifications/Descriptions.........................................................................................A-3
I/O Pin Assignments..............................................................................................................A-6
XGS-600 vs MultiGauge and SenTorr Backwards Compatibility Detail..................................A-8
Appendix B. Serial Commands ................................................................................................... B-1
Appendix C. Gas Correction Factor Table ................................................................................. C-1
DRAFT 9/22/08
i
Page 6
XGS-600 Gauge Controller
This page intentionally left blank.
Page 7
Declaration of Conformity Konformitätserklärung Déclaration de Conformité Declaración de Conformidad Verklaring de Overeenstemming Dichiarazione di Conformità
We Wir Nous Nosotros
Varian, Inc. 121 Hartwell A venue
Lexington, MA, 02421-3133 USA
Wij Noi
declare under our sole responsibility that the product, erklären, in alleniniger Verantwortung, daß dieses Pr odukt, déclarons sous notre seule responsabilité que le produit, declaramos, bajo nuestra sola responsabilidad, que el producto, verklaren onder onze verantwoordelijkheid, dat het product, dichiariamo sotto nostra unica responsabilità, che il prodotto,
XGS-600 Gauge Controller
to which this declaration relates is in conformity with the following standard(s) or other normative documents. auf das sich diese Erklärung bezieht, mit der/den flogenden Norm(en) oder Richtlinie(n) übereinstimmt. auquel se réfère cette déclaration est conforme à la (auz) norme(s) ou au(x) document(s) normatif(s). al que se refiere esta declaración es conforme a la(s) norma(s) u otro(s) documento(s) normativo(s). waamaar deze verklaring verwijst, aan de volende norm(en) of richtlijn(en) beantwoodt. a cui se rifersce questa dichiarazione è conforme alla/e sequen te /I no rm a/ o do cu mento/I normativo/i.
EN 55011 (1991) . . . . . . . . . . . . . . . . . . . . . . . Group 1 Class A ISM emission requirements
EN 61010-1 (1993) . . . . . . . . . . . . . . . . . . . . . Safety requirements for electrical equipment for measurement, control,
and laboratory use incorporating Amendments Nos 1 and 2.
EN 61000-3-2 (2006) . . . . . . . . . . . . . . . . . . . . Limits for harmonic current emissions
(equipment input current up to and including 16A per phase).
EN 610000-3-3 (2005) . . . . . . . . . . . . . . . . . . . Limitation of voltage fluctuations and flicker in low-voltage supply
systems for equipment with rated current up to and including 16A.
EN 61326 (1997/A1; 1998/A2; 2001/A3) . . . . . EMC requirements for Electrical equipment for measurement, control and
laboratory use — General Use.
EN 61326 (1997/A1; 1998/A2; 2001/A3) . . . . . EMC requirements for Electrical equipment for measurement, control and
laboratory use — General use for the following tsts with a deviation of
performance criteria to Criteria “C” instead of “B”
EN 61000 - 4-2 (2001) . . . . . . . . . . . . . . . . . . . Electrostatic Discharge Immunity
EN 61000 - 4-4 (2004) . . . . . . . . . . . . . . . . . . . Electrical Fast Transient Immunity
EN 61000 - 4-5 (2005) . . . . . . . . . . . . . . . . . . . AC Lightning Surge Immunity
Frederick C. Campbell Operations Manager
Varian, Inc. Lexington, Massachusetts, USA
August 27, 2008
Page 8
XGS-600 Gauge Controller
This page intentionally left blank.
Page 9
XGS-600 Gauge Controller
Preface
Hazard and Safety Information
This manual uses the following standard safety protocols:
WARNING The warning messages are for attracting the attention of the
operator to a particular procedure or practice which, if not followed correctly, could lead to serious injury.
CAUTION The caution messages are displayed before procedures, which
if not followed, could cause damage to the equipment.
NOTE The notes contain important information.
This product must only be operated and maintained by trained personnel. Board installation/replacement requires a properly trained service technician.
Before operating or servicing equipment, read and thoroughly understand all operation/ maintenance manuals provided by Varian. Be aware of the hazards associated with this equipment, know how to recognize potentially hazardous conditions, and how to avoid them. Read carefully and strictly observe all cautions and warnings. The consequences of unskilled, improper, or careless operation of the equipment can be serious.
In addition, consult local, state, and national agencies regarding specific requirements and regulations. Address any safety, operation, and/or maintenance questions to your nearest Varian office.
DRAFT 9/22/08
v
Page 10
EMC Warnings
EN 55011 Class A Warning
This is a Class A product. In a domestic environment this product may cause radio interference in which case the user may be required to take adequate measures.
The XGS-600 was tested with Varian manufactured gauge cables whose pigtails are connected to the XGS-600 ground stud, and with I/O cables (for remote control, serial communications and set points) employing combined foil-braid shields and metal shell connectors with the shields connected to the XGS-600 chassis. Compliance with FCC Part 15 rules and the European Union's EMI requirements cannot be assured unless Varian supplied gauge cables and foil-braid shielded I/O cables with metal shell connectors are used. Failure to install the unit in this way may result in the failure to meet the requirements for radiated emissions and susceptibility.
FCC
This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesirable operation.
XGS-600 Gauge Controller
NOTE The equipment has been tested and found to comply with the
limits for a Class A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generated, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is also likely to cause harmful radio communications interference in which case the user will be required to correct the interference at his own expense.
NOTE It has been observed that transient electromagnetic pheno mena
(electrostatic discharges and power line voltage spikes and/or surges) can cause the XGS-600 to interrupt a measurement cycle with an error or a blank screen. If this occurs, power cycle the XGS-600 in order to repeat the measurement.
DRAFT 9/22/08
vi
Page 11
XGS-600 Gauge Controller
Chapter 1: Introduction
Congratulations on your purchase of the XGS-600. This product represents a major advance in the simplicity, functionality and performance of vacuum gauge controllers. It was conceived after extensive interviews with end users in a variety of applications. Its design incorporates the latest in electronics and is CE and CSA certified and RoHS compliant. It is compact, taking only ½ rack and it is simply without peer in terms of its capacity, usability and capability.
You can view up to 8 gauges on one screen and can employ as many as 5 ion gauges or 12 convection gauges in one unit. With its simple user interface (designed to be operated without a manual) and its many standard features (8 set points, serial communication, universal voltage and pressure units), vacuum measurement is simplified. And where speed is necessary for certain critical applications it was designed to process a signal in less than 20 milliseconds. These are just some of the many features that make the XGS-600, a u nique and powerful device. Once you have used the XGS-600, we trust that you, too will find that it is not only a convenient and economical tool in your vacuum system but an indispensable one as well.
Read this manual carefully to discover all the capability of this device, but if you don't have time, simply turn it on and navigate through the screens to get started.
Backwards Compatibility
The XGS-600 is designed to be backward compatible with the SenTorr and MultiGauge instruments. Refer to page A-8 for a full description.
Part Number Scheme
Description Part Number
XGS-600 Vacuum Controller, without cards XGS600H0M0C0
Configure your XGS-600 Controller with cards – constructing your part number: X G S 60 0
1 Refer to the Gauge Selection Chart to help determine the gauge you need 2 Choose the number of HFIG** cards you want; place after the ”H”; will be 0 to 4
3
Choose the number of IMG* cold cathode cards you want; place after the “M”; will be 0 to 5
4
Choose the number of Dual Convection Gauge Cards; place after the “C”; each card runs two convection type gauges, will be 0 to 6
Note:
The XGS-600 has 6 gauge card slots permitting a total of 6 cards.
Up to four slots can be used for HFIG** cards. If four of these are chosen, one IMG* card or up to two
Dual Convection Gauge Cards may be selected also.
Up to five slots can be used for IMG* cards. If five of these are chosen, no additional gauge cards may be selected.
All six slots may be used for Dual Convection Gauge Cards.
** HFIG = Hot Filament Ion Gauge * IMG = Inverted Magnetron Gauge
DRAFT 9/22/08
H X M X C X
1-1
Page 12
XGS-600 Gauge Controller
5
XGS-600 Front and Back Description
The display is biased so as to be best viewed from directly in front or slightly above.
1 2
Figure 1-1 XGS-600 Front Panel
22/08
3
4
Figure 1-2 Backplate Connections
DRAFT 9/
Front/Rear Panel Components
Item Description Page
1
Display See “Screen Flow” on page 3-1.
2
Keypad/Navigation Keys See“General Navigation and Data Entry” on page 3-2.
3
Set Point Outputs See “Set-Point Connector Pin Assignments” on page A-7. Serial Communication Connector
4
5
Gauge Board Slots (6) See “Board Specifications/Descriptions” on page A-3.
Appendix A-9 “Serial Connector Pin Assignments” on page
A-7.
7
6
On/Off Power Switch and
6
AC Power Plug Receptacle
7
Ground Connection See “Connect the Unit” on page 2-1.
1-2
See “Connect the Unit” on page 2-1.
Page 13
Unit Dimensions
Figure 1-3 shows the unit dimensions with mm in brackets.
XGS-600 Gauge Controller
Figure 1-3 XGS - 600 Dimensions
DRAFT 9/22/08
1-3
Page 14
XGS-600 Gauge Controller
This page intentionally left blank.
Page 15
XGS-600 Gauge Controller
Chapter 2: Installation
Each XGS-600 unit is inspected and carefully packed prior to shipment. If the unit arrives damaged, save the packing materials and immediately notify the carrier. Because the packing materials are designed specifically for this instrument, always use them when transporting the unit. The shipping container is packed with the following contents:
XGS-600 gauge controllerAC line cordOperating manualRubber feet, adhesive
Connect the Unit
The unit comes shipped with sensor boards configured according to customer requirements.
CAUTION Do not block air vents. Allow approximately 3/4" clearance at
each side of the enclosure for proper air flow.
1. Plug the AC line cord into the receptacle on the rear panel of the unit and into the power source. Keep power off while connecting the remaining cables.
2. Connect gauge and I/O cables depending on controller configuration according to Figure 2-1:
I/O Connector
BA Gauge Power Connector
I/O Connector
Gauge Connection
Hot Filament Board Convection Board IMG Board
Gauge Connection CNV1, 3, 5, 7, 9, B
Gauge Connection CNV2, 4, 6, 8, A, C
Gauge Connection
I/O Connector
DRAFT 9/22/08
Figure 2-1 Gauge Board Rear Connections
NOTE When installing and HFIG cable, make sure to connect the pig
tails from the ground cable to the ground stud. See item 7 in Figure 1-2.
2-1
Page 16
XGS-600 Gauge Controller
3. Turn on the unit using the rear panel power switch.
NOTE Gauges and accessories are available for the XGS-600
controller such as rack mount hardware, gauge and I/O cables, and a tilt stand. See Table 2-1 and Table 2-2.
Table 2-1 XGS-600 Accessories
Cables Part Number
Cable Description 10 ft
UHV-24/UHV-24p Teflon UHV, bakeable to 250° C
UHV-24/UHV-24p Standard, non-bakeable
Glass BA Gauge Standard, non-bakeable
Ion Gauge Extension R32473065 R32473090 MBA-l00 Gauge R32483010 R32483025 R32483050 IMG-100 Gauge R03113010 R03113025 R03113050 R03113075 R03113100
22/08
IMG-UHV Gauge R03413010 R03413025 R03413065 R03413100 Thermocouple Gauge L91313010 L91313025 L91313050 L91313075 L91313100 ConvecTorr Gauge, Non-Bakeable L91223010 L91223025 L91223050 L91223100
' 25 ft. 50 ft. 65 ft 75 ft 90 ft 100 ft
R32463010 R32463025 R32463050
R32453010 R32453025 R32453050
R32443010 R32443025 R32443050
Other Accessories Part No. Power Cord Selection Part No.
HFIG** PCB, Field Install R3075301 Europe, 1OA /220-230 VAC, 2.5 Meter 656494220 IMG* PCB, Field Install R3080301 Denmark, 10A/220-230 VAC, 2.5 Meter 656494225
DRAFT 9/
Thermocouple/Convector Gauge card PCB, Field Install
R3084301 Switzerland, 10A/230 VAC, 2.5 Meter 56494235
Tilt Stand Accessory R3124301 UK/Ireland, 13A/230 VAC, 2.5 Meter 656494250 Center Rack Mount Kit L6423301 India, 10A/220-250 VAC, 2.5 Meter 656494245 Off-Center Rack Mount Kit L6422301 Israel, 10A/230 VAC, 2.5 Meter 656494230 Dual Mount Rack Mount Kit L6426301 Japan, 12A/100 VAC, 2.3 Meter 656494240
2-2
North America, 15A/125 VAC, 2.0 Meter 656458203 North America, 10A/230 VAC, 2.5 Meter 656494255
Page 17
Controller Gauge Card
Convection
Hot Filament
CHM
Vacuum Regime
-3
XGS-600 Gauge Controller
Table 2-2 Gauge Selection Chart
Cold Cathode
Gauge
Type Recommendation Gauge Characteristics
ConvecTorr Best overall convection gauge
for accuracy repeatability, secure connection
• Stainless steel
• Pipe thread, KF and CFF
• Locking bayonet connector
Rough ATM to 10
-9
10 to
-3
High 10
536 Same vacuum performance
531 Most economical rough
572 (glass) Most popular general purpose
563 (glass) Best general purpose HV
571 (glass) Same as 563 without
MBA-100 (one filament,
metal case) MBA-200
(two filaments, metal case)
as the ConvecTorr but without bayonet connector
gauge
HV gauge (hot filament)
gauge (hot filament) for overall performance and reliability
platinum coating Break resistant HV gauge
(hot filament)
• Stainless steel
• Pipe thread, KF and CFF
• Lowest cost
• Mild steel
• Pipe thread
• Measures to 1 x 10-9 Torr
• Tungsten filaments - More accurate and repeatable
• Glass - Can see when it is on
• Measures to 1 x 10-9 Torr
• Thoriated iridium filament ­More robust with better burnout resistance
• Glass - Can see when it is on
• Platinum coating - Provides higher performance in the
-4
range and a measure of EMI shielding
10
• Less expensive than the 563
• Measures to 10-9 Torr
• General purpose hot filament with metal housing (not glass) , so it will not break
• Uses less power than glass gauges.
• Well shielded
• Available in tungsten and thoriated iridium filament (see 572 and 563 above)
DRAFT 9/22/08
IMG-100
Inverted Magnetron
Continued on next page
Most durable and fastest responding HV gauge (cold cathode)
• General purpose-measures to 2 x 10-9 Torr
• Small, reliable, extremely rugged - Will not burn out
• Very fast response - < 20 milliseconds
• Less accurate and repeatable than hot filament
• Longer starting in high vacuum than a hot filament
• Can be cleaned
2-3
Page 18
Controller Gauge Card
Convection
Hot Filament
CHM
Vacuum Regime
Ultra High 10-9 to 10-12
XGS-600 Gauge Controller
Table 2-2 Gauge Selection Chart (Continued)
Cold Cathode
Gauge
Type Recommendation Gauge Characteristics
IMG-300 Inverted Magnetron
UHV-24 (nude)
UHV-24p (nude)
Most durable and fastest responding UHV gauge (cold cathode)
Most popular UHV gauge (hot filament)
UHV gauge providing the highest vacuum reading
• Measures to 1 x 10
• B akeable with magnet to 250 °C
• Bakeable without magnet to 450 °C
• Reliable, rugged, will not burn out
• Can be cleaned
• Measures to 1 x 10
• B akeable with magnet to 250 °C
• Thoriated iridium filaments run cooler - less outgassing
• Dual tungsten filaments run hotter ­More outgassing but measurements are more accurate and repeatable
• Same as UHV-24 except it measures to 2 x 10
-11
-10
Torr
Torr
-11
Torr
22/08
BA Cross Reference
The XGS-600 controller is compatible with many competitor gauges. Table 2-3 indicates which Varian gauge setting should be used for each of these gauges. See “Sensor Setup Screen” on page 3-9.
Table 2-3 BA Cross Reference
DRAFT 9/
GP ETI MDC Huntington Kurt Lesker Duniway
Select Sensor model 572 for the following:
274012 4336-TP/1 TP-100 G100TF T-075-N 274013 4336-TP TK-100 G075TK T-075-P 274015 4336-TK TK-100-F T-075-K 274016 4336-TK/1 T-100-N 274017 T-100-K 274018 T-100-P 274021 T-CFF-133
T-CFF-275
2-4
T-KF25 T-KF40
Page 19
XGS-600 Gauge Controller
Table 2-3 BA Cross Reference (Continued)
GP ETI MDC Huntington Kurt Lesker Duniway
Select Sensor model 571 for the following:
274002 4336-P IGT-075 IP-100 G075F I-075-N 274003 4336-K IGT-100 IK-100 G100F I-075-P 274005 4336-F IK-100-F G100KQF25 I-075-K 274006 4336-P/1 IP-150 G100KQF40 I-100-N 274008 4336-F/1 IK-150 G075N I-100-K 274020 4336-K/1 IK-150-F G100K I-100-P 274032 8140 I-CFF-133 274028 I-CFF-275 274043 I-KF25
I-KF40
Select Sensor model UHV24 for the following:
274022 8130 UHTT IGT-T G8130 T-NUDE-F 274023 8130T UHIR IGT-TI G8130T I-NUDE-F 274041 BATT 274042 BAIR
Select Sensor model 563 for the following:
G8140 I-075-NC G8120 I-075-KC G8140-DI I-100-NC G100K-PT I-100-KC G100F-PT IC=CFF-275
I-NUDE-BAC
Select Sensor model 564 for the following:
GX100-564F I-164-K GX100-564K I-164-N GX100-564N I-164-275
I-164-KF25 I-164-KF40
DRAFT 9/22/08
2-5
Page 20
XGS-600 Gauge Controller
This page intentionally left blank.
Page 21
XGS-600 Gauge Controller
Chapter 3: Operating Instructions
Operational Screens
Screen Flow
Figure 3-1 shows the screen flow.
Main
See page 3-3
Press EXIT
Press SETPTS Press LARGER
Press EXIT
Setpoints
See page 3-7
Press Exit
Press EXIT
Press SETUP
Sensor Setup
See page 3-9
Press SYSTEM SETUP
System Setup
See page 3-14
Press RESET SCREEN
Large Font
See page 3-16
DRAFT 9/22/08
Press EXIT
Reset
See page 3-15
Figure 3-1 Screen Flow
3-1
Page 22
XGS-600 Gauge Controller
General Navigation and Data Entry
Use navigation keys to locate cursor on an area of interest (ex. data entry, gauge control or screen navigation soft key). Only fields that are user-variable are accessible by the cursor. Hold the key pressed for faster cursor movement.
+
-
OK
Examples:
22/08
DRAFT 9/
Turning a gauge on:
a. Cursor over to the control field for a gauge. b. Press + or - keys to view control options (ON, OFF, etc). c. Press OK.
Modifying a number:
a. Cursor over to the number. b. Press + or - keys to increment/decrement number.
Press the + or - keys to increment/decrement the field value through all options. Hold the key pressed for fast increments/decrements.
Use the OK key to accept the selected option or data.
c. Press OK.
Navigating to another screen:
a. Cursor over to the screen navigation soft key (usually lower right corner of
screen).
b. Press OK.
NOTE Escape keys: While scrolling through options or data using the
3-2
+/- keys, you can escape without accepting new data by pressing any cursor navigation key. The original data is kept and displayed.
Page 23
Main Screen
Figure 3-2 shows the main screen, which appears at power up.
Use this screen to:
XGS-600 Gauge Controller
NOTE The sequence of gauge rows follows the board installation
sequence from left to right (when viewed from the front).
View a list of installed sensors.View real-time pressure readings and status by sensor.Turn an ion gauge on/off or calibrate a Convection gauge.View the setpoint status.
Auto-On Indicator
Pressure ReadingsSensor ID’s
Soft Keys
Soft keys
DRAFT 9/22/08
Figure 3-2 Main Screen
SETPTS Select to open the Set Point screen (page 3-7). SETUP Select to open the Sensor Setup screen (page 3-9) and other setup
screens.
LARGER Select to open the Large Font screen (page 3-16).
NOTE The NEXT key appears if there are more than eight gauges. Use this
key to move to the next Main screen with the additional gauges.
3-3
Page 24
XGS-600 Gauge Controller
Sensor ID’s Lists the five-character User Label configured on the “Sensor Setup
Screen” on page 3-9.
Auto-On Indicator Indicates Auto-On has been set up, using the Sensor Setup screen,
between a convection and an ion gauge(s). The same symbol appears next to both the convection and the ion gauge showing the link between the two.
In the screen shown, convection gauge CNV2 has been set up to trigger both ion gauges HFIG1 and HFIG2.
Multiple Auto-ON setups are possible, with a different symbol shape for each Convection gauge.
See “Sensor Setup Screen” on page 3-9 for Auto-ON details.
Pressure Reading Displays the gauge pressure reading. The pressure units, as selected
in the System Setup screen, appear at the top of the column:
Convection gauge readings always appear, as they are
always on.
For ion gauges, - - - appears when a gauge is first turned
on, until the software determines a valid reading.
22/08
DRAFT 9/
See Table 4-1 on page 4-5 for error messages displayed in this column.
If an ion gauge error message appears in the pressure reading column and the control column indicates OFF, clear the error as follows:
1. Press OK or reselect OFF in the control column.
2. Scroll to ON option to turn the gauge on.
3-4
Page 25
XGS-600 Gauge Controller
Control U se this column to input operational commands to the system. Th e
options available are: Hot Filament Gauges:
OFFFIL1 ONFIL2 ONDEGAS 1 ON (See section below)DEGAS 2 ONDEGAS OFF
Convection Gauges:
CAL
Inverted Magnetron Gauges:
OFFON
Degas (HFIG only): To activate the DEGAS function, select DEGAS1 ON or
DEGAS2 ON depending on which filament is currently on.
Emission has to be on before Degas starts, and the pressure reading must be below 1E-5T (or equivalent in other units). When Degas times out (approximately 25 minutes), the field reverts to FIL1 or FIL2, whichever filament is in use.
NOTE Degas Limitations: All gauges operate simultaneously, but only
two hot filament gauges can degas at the same time. Additional degasses are not allowed by the system.
To terminate degas manually, select De gas OFF. The field reverts to Fil1 or Fil2. Turning off degas momentarily turns off emission.
CAL (CNV only):
Convection TC gauges are calibrated using this control. The XGS-600 uses Varian's Smart Cal, which automatically determines whether to perform a vacuum calibration or atmosphere calibration based on the pressure presented to the sensor.
DRAFT 9/22/08
3-5
Page 26
XGS-600 Gauge Controller
NOTE Convection gauge calibration at atmosphere depends on
orientation. Be sure to install the gauge sensor in final orientation before calibrating.
To calibrate:
1. Cursor down to the CAL control for the required gauge.
2. Present either vacuum or atmospheric pressure to the gauge. (Allow the gauge to stabilize for 15 minutes after first power up.)
3. Press OK. Vacuum calibration defaults to 1E-4T and atmosphere calibration
defaults to 760T (or equivalents in mbar and Pa). See “System Setup Screen” on page 3-14 for modifying atmospheric pressure to your regional value. See “Sensor Setup Screen” on page 3-9 for selecting gas type. See “Convection Gauge Board” on page A-4 for activating the same calibration using rear-panel I/O controls, and Appendix Appendix B. “Serial Commands” on page B-1.
SET -P TS (Set Points) Displays the setpoints activated. The setpoints are numbered on the
Set Point screen (page 3-7).
22/08
DRAFT 9/
The setpoint number appears if all following conditions met:
The setpoint is assigned to the sensor,The setpoint screen is set to AUTO for that setpoint, And the setpoint has been activated.
3-6
Page 27
Set Point Screen
To open this screen select SETPTS on the Main screen.
Set Point # This number corresponds to the sequence of eight rear panel Set
XGS-600 Gauge Controller
Figure 3-3 Set Point Screen
Point outputs. This is also the number that appears on the Main screen if the setpoint is activated. These are fixed numbers and not user changeable.
SETPT Select an available gauge in this list to assign to a Setpoint #. Only
two set points are allowed per sensor. If two setpoints are already assigned to a sensor, that sensor does not appear in the list of options for this field.
NONE is available as a choice. If selected, all pressures and delays are set to zero and OFF.
ON Select a Setpoint On pressure value. If the sensor pressure falls
below this value, the rear panel output activates (voltage low) and the Setpoint # appears on the Main screen next to that sensor. The ON reading must be a lower pressure than the OFF reading or an error (ERR) warning appears in the rightmost column. Once valid settings are entered, ERR changes to AUTO.
See Appendix Appendix A. “XGS-600 Gauge Specifications” on page A-1 for setpoint output specifications and setpoint connector pin-outs.
ON DELAY If a delay is required between the time the sensor detects the
Setpoint ON pressure value and the outputs activates (rear panel, Main screen indicator and serial communication), select a delay time in seconds. Example: A setting of 1.2 results in a delay of approximately 1.2 seconds.
DRAFT 9/22/08
3-7
Page 28
XGS-600 Gauge Controller
OFF Select a Setpoint OFF pressure value. If the sensor pressure rises
above this value, the rear panel output de-activates (voltage high) and the Setpoint # clears from the Main screen. The OFF reading must be a higher pressure than the ON reading or an error (ERR) warning appears in the rightmost column. Once valid settings are entered, ERR changes to AUTO.
OFF DELAY If a delay is required between the time the sensor detects the
Setpoint OFF pressure value and the output, de-activates (rear panel, Main screen indicator and serial communication), select a delay time in seconds. Example: A setting of 1.2 results in a delay of approximately 1.2 seconds.
AUTO/ON/OFF Options include:
AUTO indicates the setpoint state is determined by
comparing the pressure to the setpoint limits. This is the default setting and is used to enable the setpoint.
OFF and ON are override values, which can also be used for
manual testing. Selecting OFF or ON does not clear pressure/time settings.
22/08
EXIT SOFT KEY Select to return to the main screen (page 3-3).
DRAFT 9/
3-8
Page 29
Sensor Setup Screen
To open this screen select SETUP on the Main screen.
Figure 3-4 Sensor Setup Screen: HFIG and IMG Sensors
XGS-600 Gauge Controller
Figure 3-5 Sensor Setup Screen: CNV (Convection) Sensors
See Table 3-1 on page 3-11 for default values. Fields that are unavailable have
DRAFT 9/22/08
- - -.
3-9
Page 30
22/08
XGS-600 Gauge Controller
Sensor ID Use this field to select the sensor to view its setup parameters.
Displays the unique system-assigned gauge identification. Scroll through this list to select a sensor and the remainder of fields on the screen populate with its present settings.
Possible sensor ID’s are:
HFIGx for Hot Cathode Ion Gauge boards using BA or UHV
sensors.
CNVx for Convection boards (TC/ConvecTorr)IMGx for IMG Gauge boards
where x indicates the number of the transducer attached. The number is system-assigned based on the sequence of board installation, left to right (when viewed from the front). Convection boards have two sensors per board. When more than 4 convection boards are installed, the 10, 11, and 12 channels are designated CNVA, CNVB, and CNVC, respectively.
User Label Use this field to enter your label for this sensor that appears on all
other screens and serial communication. This field has a maximum of five (5) characters and any alphanumeric character is available for use.
User Labels that begin with HFIG, CNV or IMG are not allowed to avoid confusion with Sensor ID’s.
Sensor Model Use this field to configure the particular model of sensor attached
DRAFT 9/
If no User Label is selected, the default is the Sensor ID. When a System Reset is initiated, this label is reset to the sensor ID.
to a board.
NOTE If Sensor Model is changed, all fields resort to default values.
3-10
Page 31
XGS-600 Gauge Controller
Table 3-1 Default Setup Values by Ion Gauge Type
HFIG
Sensor Model
Default Emis Current (ma)
Default
Sensitivity
Over-
pressure
limit (mT)+
Default
Degas Power
(mA)
Number of
Filaments
MBA100 0.1 25 1-10 mT* 1 1 MBA200 0.1 25 1-10 mT* 1 2 571 4 10 1-10 mT* 10 1 572 4101102 563 4 10 1-10 mT* 10 1 564 0.1 6 1-10 mT* 10 1 UHV24 (default) 4251102 UHV24p 4201102 IMG 100 (default) N/A 1 10 mT N/A N/A IMG 300 N/A 2.5 2 N/A N/A
+ Not a user-variable setup parameter * If Emission Current is set:
Less than 0.7 mA, overpressure is 10 mT (or equivalent in mbar or Pascal).Greater than 0.7 mA, overpressure is 1 mT (or equivalent in mbar or Pascal).
AUTO-ON IF GAUGE This field is active for HFIG and IMG gauges only. Use this field
along with IS LESS THAN to select a Convection gauge and assign a pressure level below which AUTO-ON engages.
When Auto-On is activated, you cannot manually turn on or off the ion gauge; it is controlled by the assigned convection gauge. Select
- - - to deactivate Auto On.
IS LESS THAN Use this field with Auto-ON to enter pressure (max allowed 1T) at
which to trigger the ion gauge. There is a hysteresis of 5mT built in to the software to prevent fluctuating ion gauge emission from triggering on/off. For example, if this field is set to 3E-3T, Auto-ON will turn on the ion gauge at 3E-3T and off at 8E-3T.
SENSITIVITY This field is active for HFIG and IMG gauges only. Use this field to
raise or lower the sensitivity.
EMIS CURRENT (mA) This field is active for HFIG gauges only. Use this field to configure
the EMIS Current. See Table 3-1 on page 3-11 for sensor default values.
DRAFT 9/22/08
3-11
Page 32
XGS-600 Gauge Controller
NOTE Changing the Sensor Model, defaults Sensitivity and Emission
Current settings.
GAS CORRECTION Use this field to enter a correction for various gas types. For an ion
gauge (HFIG and IMG), enter a nominal value from the Gas Correction Factor Table in Appendix C. For Convection gauges,
select Air or Argon. SYSTEM SETUP Select to open the System Setup screen (page 3-14). EXIT Select to return to the Main Screen (page 3-3).
Selecting Emission Current for Hot Filament Ion Gauges
When the gauge model number is selected, the XGS-600 automatically selects a default emission current value that works for most applications. However there are reasons why you may want to select another value based upon your application.
If you are using a gauge at pressures higher than 1.0E-4 Torr, the gauge performs
better and lasts longer at low values of emission current. In general, for this pressure range, use 0.1ma.
22/08
If you are using the gauge in the UHV range, the default emission setting may not
result in a large enough ion current to measure reliably. In this case, increase the emission current to the maximum setting of 10 ma. However, there will be more out-gassing which could adversely affect the system pressure. So, a compromise between the emission current setting and filament out-gassing must be made. Try to use the lowest emission current setting that provides good measurement results.
If you have had your gauge calibrated, it would have been done at a specific
DRAFT 9/
3-12
emission current. You should enter that value into the emission current setting of the XGS-600.
The XGS-600 employs a feature that automatically changes the emission current
based on the pressure that is measured. If the emission current is set to values greater than or equal to 1.1 ma, the controller applies the full value when the measured pressure is below 1.0E-5 Torr, and 1/10 that level. There is an algorithm to minimize hunting. However , depending upon the vacuum system, it may still be possible for the gauge to hunt for the proper emission current setting when at or near 1E-5T. This is caused by the fact that when the emission current is raised as the pressure goes down, the filament outgases at a higher rate, causing the pressure to increase a bit. Sometimes it is possible for the pressure to increase enough that the controller reduces the emission current, which eliminates the source of gas. This causes the pressure to decrease to where the controller again raises the current, and the system can get stuck in this cycle. Y ou can defeat this by setting the emission current to 1.0 ma or less, so the XGS-600 maintains that constant emission current at all pressures.
th
of that value when the pressure is above
Page 33
XGS-600 Gauge Controller
HFIG Smart Overpressure Protection
Hot filament ion gauges all exhibit non-linearity at high pressure such that they give readings lower than the actual pressure. For gauges that are rated up to 10 mTorr or higher, operating the BA gauge at low emission current can minimize this measurement error. Unfortunately, many applications require operating the gauge at emission currents that would result in poor measurement accuracy at high pressure, thereby compromising the ability to provide effective overpressure protection.
XGS-600 adjusts the overpressure limit based upon the emission current. When the emission current is 0.7ma or less, the gauge's full high pressure range can be used. If, however , the emission current is set to higher values, XGS-600 protects the gauge by turning off the filament at 1 mTorr. Overpressure Protection uses the same fast pressure signal that is used for fast set points, and is displayed by the P>MAX error message.
Setting the Auto-ON pressure level
When enabling the Auto-ON feature, the XGS-600 defaults the threshold pressure to 5 mTorr. If the convection gauge is installed so that it is measuring the same high vacuum chamber as the ion gauge, this is a good value to use. However, in many applications users may want to monitor the high vacuum pump's fore-line pressure instead. In this application, the foreline may never get down as low as 5 mTorr, especially if a large volume of gas is being pumped. The XGS-600 allows you to program a threshold pressure as high as 1 Torr. When Auto-ON is being used, the gauge overprotection feature is still active to protect the high vacuum gauge.
3-13
DRAFT 9/22/08
Page 34
System Setup Screen
Use this screen to set values for overall system operation. To open this screen select SYSTEM SETUP on the Sensor Setup screen.
XGS-600 Gauge Controller
Figure 3-6 System Setup Screen
UNITS Select: Torr (default), Pascal or mbar. Once the units are changed,
22/08
SETUP LOCKOUT Toggle this to ON to make all fields on the SYSTEM SETUP and
AUTO FILAMENT ADVANCE (HFIG Boards)
DRAFT 9/
the values throughout the system are updated to the new units.
SENSOR SETUP screen display-only (except for this field). This
provides a layer of security to prevent inadvertent changes.
For two-filament HFIG sensors, if Auto Filament Advance is
selected, Filament 2 is automatically engaged if Filament 1 has an
open filament error (NO FIL1). FIL2 ON then appears on the main
screen. Auto filament advance can occur when emission is first
turned on or during gauge operation. Auto advance does not
advance from Filament 2 to Filament 1. If an open filament error
occurs on Filament 2, instead the NO FIL2 error appears and
emission is turned off.
3-14
Page 35
XGS-600 Gauge Controller
A TMOSPHERE V ALUE (Convection Boards)
SERIAL MODE Select between RS232 or RS485 serial transceivers. RS232 and
SERIAL BAUD RATE Selects between baud rate options: 9600 (default) and 19200. RS485 ADDRESS Set the RS485 address (default 00) as used in the first two
RESET SCREEN Select to open the Software Versions screen which incudes the
EXIT Select to return to the Main Screen (page 3-3).
Software Versions Screen
This screen shows the software versions on a board-by-board basis. To open this screen select RESET SCREEN on the System Setup screen.
Sets the Atmosphere value to be used in Convection gauge calibration. See “Main Screen” on page 3-3 for the calibration procedure.
RS485 have separate outputs on the rear panel Serial connector (see Appendix Appendix A. “XGS-600 Gauge Specifications” on page A-1).
characters of all serial protocol commands (see Appendix Appendix B. “Serial Commands” on page B-1).
system reset function (page 3-15).
Figure 3-7 Software Versions Screen
SYSTEM RESET Use this to perform a full reset of user-defined variables, reverting
the system to as-shipped (factory-set) state.
EXIT Select to return to the Main Screen (page 3-3).
3-15
DRAFT 9/22/08
Page 36
Large Font Screen
To open this screen select LARGER on the Main screen.
The leftmost column shows the User Label as configured on the Sensor Setup screen. You can select which transducers appear on this screen by moving the cursor to the User Label and using the +/- modify keys.
XGS-600 Gauge Controller
Figure 3-8 Large Font Screen
22/08
EXIT Select to return to the Main Screen (page 3-3).
DRAFT 9/
3-16
Page 37
XGS-600 Gauge Controller
Chapter 4: Service and Troubleshooting
It is possible to replace sensor boards in the field for repair or to change the unit's configuration to meet new application requirements. Because the slot spacing is different for slots 5 & 6, you must follow the instructions below when installing the sensor boards.
Board Configuration Rules
6 5
4
FRONTREAR
3
2
1
Figure 4-1 XGS-600 Slot Numbers
Slots are numbered from left to right as viewed from the front of the XGS-600.Option slots 1 though 4 are “Long, High Profile Slots” and can accommodate any
board. HFIG boards are “Long, High Profile” boards and can only be installed in slots 1 through 4.
Option slots 5 and 6 are “Short, Low Profile” slots and can accommodate only
convection or IMG boards, with the following restrictions:
An IMG board may NOT be installed in slot 5 if any other board is installed in
slot 4.
DRAFT 9/22/08
An IMG board may NOT be installed in slot 6 if any other board is installed in
slot 5.
4-1
Page 38
XGS-600 Gauge Controller
Examples:
You want to install 5 IMG boards:
IMG boards are installed into slots 1, 2, 3, 4, and 6. Slot 5 remains empty. Do not
install any board into slot 5.
You want to install 4 IMG and 2 CNV boards:
IMG boards are installed in slots 1, 2, 3, and 4; CNV boards are installed in slots 5
and 6.
Sensor Board Replacement Procedure
WARNING Only trained service personnel should attempt this work.
Disconnect the XGS-600 mains connection by unplugging the IEC power cord from the back of the unit before removing the unit's cover.
CAUTION Wear an ESD bracelet and observe all appropriate precautions
to avoid damage to PCBs.
22/08
1. Turn off and unplug the line cord (AC mains).
DRAFT 9/
2. Remove the cover by unscrewing the two Philips head screws.
3. If removing an HFIG board, unplug the yellow connector near the front panel. It may
require some effort to unplug. Do not pull on the wires, pry the yellow plastic connector instead.
CAUTION When these boards are assembled certain components are bent
over intentionally. Do not straighten them.
4-2
Page 39
XGS-600 Gauge Controller
le
Board Connector (female)
4. Lift the sensor board by the tab at the rear panel (Figure 4-2). Once the board is disengaged from the board connector, pull the board out of the end guide.
Board
Connector
(male)
Tab
Figure 4-2 Tab and Main Board Connector
WARNING To ensure safety and proper operation, no slot opening can be
left unfilled. Install a filler plate by installing the two hooks into the case bottom and snapping in the foot. Contact Varian if additional fillers are needed.
5. Install the replacement board:
a. Place the end of the board into the end guide (Figure 4-3).
End Guide
HFIG Plug
HFIG Power Cab
Figure 4-3 End Guide and HFIG Connector
b. Guide the board down as straight as possible onto the connector on the main
board. Figure 4-4 shows the connector on the base of all three board types.
DRAFT 9/22/08
Figure 4-4 Sensor Board Connector: Bottom View
4-3
Page 40
c. Locate the rear panel portion of the board into its slot, press down the tab and
exert gentle pressure until it clicks into place.
CAUTION Do not force the board down. If too much pressure is required
d. Insert the yellow HFIG plug into the connector (HFIG boards only). See
Figure 4-3. This connector is keyed.
e. Inspect the installation to ensure the board is seated properly and not touching
any adjacent boards.
6. Replace the covers: a. Push the cover straight in under the lip at the XGS front bezel. b. Replace the two Philips head screws. c. The cover must be flat and fully seated. It may be necessary to adjust the boards
and filler plates slightly for the cover to fully engage. The cover should drop on front of the filler plate.
XGS-600 Gauge Controller
the connector pins are not aligned.
22/08
7. Plug the unit back in and turn on the power switch.
DRAFT 9/
NOTE Any board-specific settings must be reprogrammed.
Most Sensor setup and Set point user settings stay with the board. This means if the removed board is used elsewhere, most settings will be saved. However, if a new board is installed, you will have to reprogram the settings.
4-4
Page 41
XGS-600 Gauge Controller
Troubleshooting
Error Codes
Table 4-1 General Error Code
Error Message Meaning Cause Action
BD COM Board Comm:
Internal communication error
GRIDLO (HFIG bds only)
HITEMP
(HFIG and IMG bds
Grid voltage is too low
Internal temperature is above 75° C
only)
NOFIL1 (HFIG bds only)
NOFIL2 (HFIG bds only)
Filament is open or cable disconnected
Second filament is open or cable disconnected
Open
(CNV bds only)
P>MAX
(HFIG and IMG bds)
No Convection sensor detected
Pressure is above the maximum operating point for the gauge
Motherboard and sensor board stopped communicating
Pressure is too high
causing a glow dis­charge
HFIG board failureFan failure
XGS ventilation
openings blocked
Internal temperature
above 75° C
Filament failureIon gauge cable not
connected
Filament failureIon gauge cable not
connected
Cable not
connected
Sensor failureSystem vented from
vacuum with gauges on
Attempted to turn
gauges on with vac­uum near overpres­sure limit
1. Cycle power to unit
2. Return unit for repair
1. Check BA gauge for grid to filament short or grid to housing short
2. Replace gauge
1. Allow unit to cool
2. Check that fan is operating
3. Ensure that there is adequate ventilation around the unit
1. Use second filament
2. Replace BA gauge
3. Check that cable is installed
1. Replace BA gauge
2. Check internal sensor board connections
3. Check that cable is plugged in
1. Check cable
2. Replace sensor
1. Turn gauges off before venting system or use AUTO-ON
2. Wait for vacuum to improve before turning on ion gauges
DRAFT 9/22/08
4-5
Page 42
Tips
Varian offers reference ionization B/A gauges, which are sealed off at approximately 5E–6 Torr/6.6E–4 Pa, as well as a Convection gauge simulator which can switch between atmosphere and vacuum simulation. These devices are extremely helpful in troubleshooting a vacuum system problem by isolating the defective component.
NOTE These reference gauges and stimulators are not NIST traceable
Troubleshooting Table
Problem Cause Action
XGS-600 Gauge Controller
calibrated gauges, and should not be used for calibration.
Table 4-2 Troubleshooting
Hot filament gauge hunts or cycles around 1E-5Torr (1E-3Pa)
22/08
IMG displays readings in the –11 Torr range when the pressure is known to be higher
Sometimes after a long time period the reading becomes correct
DRAFT 9/
HFIG bd reads UHV when pressure is know to be higher
ERR appears in rightmost column of setpoint screen
Duplicate message appears when trying to enter a User Label
User Label not accepted
Auto emission adjustment occurs at this pressure. Increased filament power causes a rise in pressure. Condition will clear as system pressure decreases.
IMG cable not plugged inIMG has not startedIMG took a long time to start
Emission current not estab-
lished due to worn out Th-Ir filament
Collector cable unpluggedHFIG board failure
Setpoint OFF pressure is < than ON pressure
Label is the same as another
User Label
User input violates User
Label rules
If taking pressure measurements in this range, reduce emission current setting for the gauge to
1.0 mA or less to turn off feature
1. Check that cable is attached
2. Sometimes IMG’s are hard to start when turned on at high vacuum
3. Raise pressure
4. Clean or replace IMG
1. Check collector connection at gauge (glass BA) and at XGS-600
2. Replace gauge or filament
3. Return unit for repair/replace HFIG bd
Correct pressure entries so the setpoint ON pressure is less than the OFF pressure
Create a different entry
See “Sensor Setup Screen” on page 3-9 for User Label rules
4-6
Page 43
XGS-600 Gauge Controller
Table 4-2 Troubleshooting (Continued)
Problem Cause Action
Cannot turn HFIG or IMG on using the keypad
Cannot turn on Degas 1. Pressure is above 1E-5T
The front panel display indicates the setpoint triggers/releases at a slightly different pressure than what is programmed
Changes to baud rate, comm. mode, units, keypad lockout, or ATM value do not take effect
Auto-ON programmed for
that gauge
The keypad is locked out
2. Attempted to turn on Degas2 while running Fil1
3. Attempted to degas third gauge
Time delay = 0.0, 0.1 XGS-600 uses a less accurate
Did not exit System Setup screen after making changes
1. This is the correct response
2. If manual control is desired, turn off AUTO-ON
3. Unlock keypad
1. Pump down before degas
2. Change operation to Fil2 before attempting to degas that filament
3. XGS-600 will only degas two gauges at a time
4. Finish a degas before attempting to start a third
pressure measurement for setpoint delays of <= 0.1 secs to obtain fastest responsiveness
After making changes, exit the screen for the changes to take effect
Unit does not respond to RS232 or RS485 commands
Display goes blank followed by a refill of the screen. Operation is otherwise normal.
One or more ion gauge channels (HFIG and IMG) appear to restart, sometimes a BDCOMM message appears first
1. XGS600 in wrong serial mode or incorrect baud rate
2. Incorrect wiring
3. Incorrect terminal settings
Screen refreshes approximately every 20 minutes
A major EMI disturbance or other anomaly has caused a sensor board watchdog timeout
1. Change serial comm to RS232 or RS485, select proper baud rate, set RS485 address in General Setup screen
2. Check wiring
3. Check that terminal has correct # of data, start/stop bits, etc.
XGS-600 refreshes the entire screen periodically to update.
No action required. This is how the XGS-600
recovers from hard and soft errors. No action required.
DRAFT 9/22/08
4-7
Page 44
XGS-600 Gauge Controller
Table 4-2 Troubleshooting (Continued)
Problem Cause Action
BDCOMM message appears on one or more channels and stays
Display goes blank A major EMI has caused a
A major EMI disturbance or other anomaly has caused a non-recoverable software error
display or software error
22/08
Cycle AC (mains) power to the unit OFF and ON again to recover
Cycle AC (mains) power to the unit OFF and ON again to recover, or wait for the automatic screen refresh (every 20 minutes)
DRAFT 9/
4-8
Page 45
XGS-600 Gauge Controller
Appendix A. XGS-600 Gauge Specifications
Instrument Specifications
Table A-1 provides the XGS-600 instrument specifications.
Table A-1 General XGS-600 Gauge Specifications
Specification Description
Display Type: Backlit dot matrix monochrome LCD
Backlight: White LED Small Font mode: Up to 8 channels displayed simultaneously
Large Font mode:Up to 2 channels displayed, readable from 15 feet Keypad 5 button membrane with positive tactile feedback Serial Communication Type: RS232 or RS485
Baud Rate: 9600 or 19600 Set-Point Outputs Type: 8 open collector type, ground referenced outputs
Voltage Rating: 32 VDC max
Current Rating: 50 mA max
Output Voltage:
V
50 Gnd
out
V
@ 5 mA = 0.6 V
on
Von @ 25 mA = 0.75 V
Von @ 50 mA = 0.90 V
Programming capability:
Can assign up to 2 outputs per gauge channel
Separate "make" and "break" pressure levels
Independent time delays on "make" and "break" of 0.0 to 9.9 seconds
Manual override capability of each set point output
Response Time: 20 msec max (when delay = 0.0) Chassis Dimensions: ½ rack, 2U high, 11" deep
Material: Grey Polycarbonate
max
max
DRAFT 9/22/08
Fuse Rating 5A Fast Blow
Ventilation:Internal fan with plenum and over-temperature sensing
A-1
Page 46
XGS-600 Gauge Controller
Table A-1 General XGS-600 Gauge Specifications (Continued)
Specification Description
Mains Power and Cordage Voltage: 100-120, 200-240 VAC ±10% 50/60 HZ, 2A
Available worldwide IEC cords
Environmental/Installation Storage Temperature: -15 to 70° C
Operating Temperature: CSA certified 5 to 40° C, at 5 to 95% RH, non-condensing
Installation Category: 2 Pollution Degree: 2 Indoor use Altitude: Up to 2000 m
Regulatory Approval CE Certified:
Safety:
EN61010
22/08
DRAFT 9/
Emissions:
EN6100-3-2 : 2006 EN6100-3-2 : 2005 EN61326: 1997/A1: 1998/A2: 2001/A3: 2003
ESD, EFT, AC surge tested to criteria C
EN55011 FCC Part 15 subpart J Certified to CSA1010 for US and Canada RoHS Compliant WEEE labeled
A-2
Page 47
XGS-600 Gauge Controller
Board Specifications/Descriptions
HFIG Board
The HFIG board operates all Varian hot filament gauges and most other available Bayard-Alpert type hot filament ionization gauges including nude UHV types. The board can be set up for all Varian hot filament gauges and allows you to set gauge emission current, sensitivity value, and gas correction factor. The UHV range and E-Beam degas are standard, as is dual filament control, remote I/O control of emission and filament selection, and 0 to 10 V analog output.
Table A-2 HFIG Gauge Specifications
Item Description
Supported Gauge Heads
Displayed Measurement Range*
Varian 563, 564, 571, 572, UHV24, UHV24p, MBA-100,
MBA-200
5.0E-12 Torr to 1 mTorr (UHV24, UHV24p, 572)5.0E-11 Torr to 10 mTorr (563, 571, MBA100, MBA200)
* This represents the ability of the controller to display pressure values and not the capability of the gauge heads to accurately
measure pressure. Emission Current 0.1 to 10.0 mA Sensitivity Input Range 00.1 to 99.9/Torr Gas Correction Factor Range 0.01 to 9.99 Degas E-Beam type, 600 V at 10 mA, enabled only when pressure is
below 1.0E-05 Torr Analog Output
Error Detection
Range: 0-10 VDC
Slope: 1 V per decade log
Characteristic: output voltage = Log (Pressure) + 11
Output when gauge OFF or Error = 10V
Load Impedance: >10 kOhms
Open Filament
Shorted Grid
Over Temperature
Over Pressure (1mT or 10mT dependant on tube type selected)
DRAFT 9/22/08
Gauge Connector
Collector: SMB coax
Fil and Grid:AMP 5 pos Univ Mate-N-Lok
A-3
Page 48
XGS-600 Gauge Controller
Table A-2 HFIG Gauge Specifications
Item Description
Remote I/O
Allowable Cable Length
Convection Gauge Board
The Convection card operates two Thermocouple/Convection gauges and allows you to calibrate:
To the actual atmospheric pressure value, To vacuum pressure,and select calibration for air or argon.
22/08
E
Calibration is Varian's Smart Cal that determines whether a CAL command is for vacuum or atmosphere. Remote control of calibration and analog output is standard. Each channel has its own 9-pin Dsub connector.
Connector: 6 pin IEEE 1394
Grounding the input pin activates that function; leaving pin 1 open selects filament 1
The Status Output goes HI during an error
up to 50’ using Varian standard cableup to 300’ using under 10’ Varian standard cable + Varian exten-
sion cable
Table A-3 Convection Board Specifications
Item Description
Supported Gauge Heads Varian Model, 536, 531, and ConvecTorr
DRAFT 9/
Displayed Measurement Range 1.0E-4 Torr to 1000 Torr Gas Types Air or Argon Atmosphere Value Range 500 to 1000 Torr Analog Output
Range: 0-10 VDCSlope: 1 V per decade logCharacteristic: output voltage = Log (Pressure) + 5Output when gauge OFF or Error = 10 VLoad Impedance: >10 kOhms
Error Detection Open cable/sensor Gauge Connector Dsub9
A-4
Page 49
XGS-600 Gauge Controller
Table A-3 Convection Board Specifications
Item Description
Remote I/O Connector 6 pin Firewire type
Grounding the input pin activates that function The Status Output goes hi during an error. Note: the analog outputs for CNV1 and CNV2 are also
available at pin 7 of their respective Dsub connectors.
Allowable Cable Length up to 300’ using Varian TC or ConvecTorr cable
IMG Board
The IMG board operates Varian IMG-100 and IMG-300 gauges and allows you to set gauge sensitivity value, and gas correction factor. The UHV range is standard, as is remote I/O control of emission and 0 to 10 V analog output.
Table A-4 IMG Board Specifications
Item Description
Supported Gauge Heads Displayed Measurement Range*
Varian IMG-100, IMG-3005.0E-12 Torr to 2 mTor r (IMG-300)
5.0E-12 Torr to 5 mTor r (IMG-100)
* This represents the ability of the controller to display pressure values and not the capability of the gauge heads to accurately
measure pressure. Sensitivity Input Range 00.1 to 99.9 Gas Correction Factor Range 0.1 to 9.9 Analog Output
Error Detection
Gauge Connector Remote I/O Connector
Range: 0-10 VDC
Slope: 1 V per decade log
Characteristic: output voltage = Log (Pressure) + 11
Output when gauge OFF or Error = 10 V
Load Impedance: >10 kOhms
Over Temperature
Over Pressure (2 mT or 5 mT dependent on tube type selected)
SHV
6 pin Firewire type
Grounding the input pin activates that function
The Status Output goes HI during an error.
DRAFT 9/22/08
Allowable Cable Length
up to 300’ using RG59U
A-5
Page 50
I/O Pin Assignments
Table A-5 through Table A-9 give the I/O pin assignments for the various boards and connectors.
Table A-5 IMG Board I/O Pin Assignments
ANALOG OUT 5
STATUS OUT 4
XGS-600 Gauge Controller
Function Pin # (J26)
EMIS ON 6
GND 2 + Shield
-1
-3
22/08
DRAFT 9/
Table A-6 HFIG Board I/O Pin Assignments
Function Pin # (J26)
EMIS ON 6
ANALOG OUT 5
STATUS OUT 4
GND 2 + Shield
FIL 2 SELECT INPUT 1
DEGAS ON INPUT 3
Table A-7 Convection Board I/O Pin Assignments
Function Pin # (J1)
TC1 ANALOG OUT 5 TC2 ANALOG OUT 3
EXT TC1 CAL IN 1 EXT TC2 CAL IN 6
STATUS OUT 4
A-6
GND 2 + Shield
Page 51
XGS-600 Gauge Controller
Table A-8 Set-Point Connector Pin Assignments
Function Pin #
GND 1 SET POINT 1 OUTPUT 2 SET POINT 2 OUTPUT 3 SET POINT 3 OUTPUT 4 SET POINT 4 OUTPUT 5 SET POINT 5 OUTPUT 6 SET POINT 6 OUTPUT 7 SET POINT 7 OUTPUT 8 SET POINT 8 OUTPUT 9
Table A-9 Serial Connector Pin Assignments
Function Pin #
+RS485 XFD (DIFF INPUT/OUTPUT) 1
TXD OUTPUT RS232 2
RXD INPUT RS232 3
NOT USED 4
GND 5
NOT USED 6 NOT USED 7
RTS OUTPUT RS232 8
-RS485 XFD (DIFF INPUT/OUTPUT) 9
DRAFT 9/22/08
A-7
Page 52
XGS-600 Gauge Controller
XGS-600 vs MultiGauge and SenTorr Backwards Compatibility Detail
XGS-600 replaces both MultiGauge and SenTorr rack mount controllers. While the XGS-600 was designed to be substantially backwards compatible, there are some design features that prevent the new product from being 100% backwards compatible.
Backwards Compatible Features
There are three board options, they are:
Hot Filament Ion Gauge Board (HFIG)Inverted Magnetron Board (IMG)Two channel Convection Board (CONV)
The Hot Filament board replaces both the UHV and BA MultiGauge boards and implements the SenTorr BA and UHV models. You select the type of sensor being employed and software takes care of the settings. The ion gauge cable connectors are identical to the older products.
The Inverted Magnetron board employs the same SHV connector as MultiGauge and
22/08
DRAFT 9/
operates both of the IMG100 and the IMG300 (UHV-IMG) as does the current MultiGauge board.
The Two Channel Convection board has the same gauge cable connectors as the current MultiGauge Convection card, and the SenTorr xxTC and xx2C models. MultiGauge uses the Dsub connector for the analog output, as well as for the sensor connections, and this is supported in XGS-600.
The RS232 serial communications connector and protocol is supported in XGS-600, as well. However, the RS485 connection is now housed in the same Dsub-9 connector as for the RS232, and is not backwards compatible with the MultiGauge and SenTorr Mini-DIN connectors.
Non-Backwards Compatible Features
The significant changes with the XGS-600 vs MultiGauge and SenTorr are the following:
XGS-600 does not have a 4 channel TC card (MG)The analog outputs are housed in a firewire type connector (as menti oned above, the
two channel Convection board has its analog outputs in the Dsub's as well)
Remote I/O signals are also housed in the same fire-wire connector. There is no I/O
board option as in MultiGauge
The standard non-isolated open-collector set-point outputs are housed in a
9-position Dsub connector and only provide C and NC contacts. MultiGauge and
A-8
Page 53
XGS-600 Gauge Controller
SenTorr provide true isolated relay outputs with access to C, NC and NO contacts. SenTorr uses terminal block connectors while MultiGauge uses a 25-position Dsub.
XGS-600 is not available in a Black Box version with a remote mountable display.
Table A-10 list the features compatible for the MultiGauge and SenTorr units.
Table A-10 Feature Compatibility Table
SenTorr BA or
Feature XGS-600 MultiGauge
UHV SenTorr CC
Hot Filament
All All All n/a
Gauges Cold Cathode
Gauges
Convection Gauges
IMG100 IMG300
(UHV-IMG) 2 ch Convection
Bd using (2)
IMG100 UHV-IMG 525CCG
4 ch TC Bd using DsuB37
Dsub9
2 ch ConvecTorr Bd using (2) Dsub9
CDG’s Not supported (2) Dual Ch CDG
Bds
Rem I/O Uses Fire-wire
conn on each board
All control inputs are active
Needs optional Remote I/O Bd using Dsub37
Uses opto-isolated active HI inputs
LO, no isolation
n/a 525CCG
2 ch ConvecTorr using (2) Dsub9
2 ch TC
2 ch ConvecTorr using (2) Dsub9
2 ch TC
n/a n/a
Micro phone jack for IG control
Uses opto-isolated active HI inputs
Micro phone jack for IG control
Uses opto-isolated active HI inputs
DRAFT 9/22/08
Analog Output In Fire-wire
conn of each Bd
Conv Bd also provides access pin in Dsub9
Set Points Std 8 setpoints,
2 per ch using open collector drivers in Dsub9
Micro phone jack on each Bd
2ch and 4ch Conv bds provide access in Dsub conn
Optional 8 setpoints using relays using Dsub25, access to C,NO,NC contacts
Micro phone jack Micro phone jack
Optional 4 setpoints using relays using term block type conn, access to C,NO,NC contacts
Optional 4 setpoints using relays using term block type conn, access to C,NO,NC contacts
A-9
Page 54
XGS-600 Gauge Controller
Table A-10 Feature Compatibility Table (Continued)
Feature XGS-600 MultiGauge
SenTorr BA or
UHV SenTorr CC
RS232 Std Dsub9 IBM
pinout, supports MG
protocol
RS485 Std in Dsub9,
supports MG protocol
Black Box and Remote Display option
Table A-11 list the cable compatibility for the MultiGauge and SenTorr units.
22/08
UHV Bakeable L6440 Yes Yes Yes N/A UHV non-bakeable L6441 Yes Yes Yes N/A
No No Yes Yes
Table A-11 Cable Compatibility Table
Cable XGS-600 MultiGauge SenTorr BA or UHV SenTorr CC
Opt Dsub9, IBM pinout
Opt with choice of daisy chain using MiniDIN or Dsub9
Opt Dsub9, IBM pinout
Opt with choice of daisy chain using MiniDIN or Dsub9
Opt Dsub9, IBM pinout
Opt with choice of daisy chain using MiniDIN or Dsub9
Glass BA L6455 Yes Yes Yes N/A Ion Gauge Extension L6456 Yes Yes Yes N/A 525 CCG L5671 No Yes n/a Yes
DRAFT 9/
IMG R0341, R0311 Yes Yes n/a No MBA R1172 Yes Yes Yes n/a Quad/Dual TC L6475 No Yes No No Single TC L9131 Yes Yes Yes Yes Single Conv L9122 Yes Yes Yes Yes CDG L9153 No Yes No No
A-10
Page 55
XGS-600 Gauge Controller
Appendix B. Serial Commands
Response to serial command inquiries approximately 10 ms, response time defined as time between receipt of last query character to transmission of first response character. Serial queries more frequent than 10 per second will compromise system responsiveness.
Command Syntax and Definitions
The command format is: # {XGS-600 address} {command number} {optional data} {carriage return}
The response format is: > {optional data} {carriage return}
The following characters are used in Table B-1:
Command Description Notes
# All commands to the XGS-600 must begin with the character "#".
aa RS485 Address =00 for RS232 (system default). Max allowable: 20hex
01
c Sensor Code =U, signifies User Label is to follow.
n
Note: Alternatively, "cn" can be used to designate a short code instead of User Label, backward
compatible to Multi-Gauge/SenTorr protocol; see separate instructions below (Table B-1 on page B-2)
h Hex Digit Used to identify setpoint number or to list status of 8 bits.
Command Number
User Labels As found on Sensor Setup screen.
Each command is identified by a unique command number, 2 hex digits, 01-FF.
Examples: #00B0UION4UGATE3.2E-03 sets AUTO-ON for ion gauge
"ION4" using CNV gauge "GATE". #0030UGATE1 turns on emissions of sensor User Label Gate1. #0055UTOP26.00 sets sensitivity for ion gauge with User Label
Top. Note: If User Label is not defined by user, default User Label is
Sensor ID.
DRAFT 9/22/08
B-1
Page 56
XGS-600 Gauge Controller
Command Description Notes
x Data Digit
Termination All commands sent to the XGS-600 require < CR > termination.
< LF > is ignored so it is possible to use < CR/LF > with RS232 but
< CR >
The XGS-600 sends ?FF as a response if the command or data is invalid, or if the command length is incorrect. There is no response to a wrong address, or lack of termination character.
Description Command
not with RS485 as it would cause a collision with the response. Please note that time to response is extremely quick (10 to 20 ms) so be ready! Responses are also terminated with < CR >.
Table B-1 Serial Command Set
Response from
XGS-600 Notes
Read XGS contents #aa01 >hhhhhhhhhhhh
where “hh” sensor
22/08
DRAFT 9/
Read pressure #aa02cn >x.xxxE-xx Read
software revision
System Reset #aa06 >
#aa05 >hhhh,hhhh . . .
board codes are: 10 = Hot Filament Ion
Gauge board (HFIG) 3A = Inverted
Magnetron board (IMG) 40 = convection board
(CNV) FE = Empty slot
where hhhh represents revision hh.hh
Lists two-character codes identifying the 6 board slots in the system, following board installation sequence from left to right (front view).
Revisions shown for main board, then each sensor board following board installation sequence from left to right (front view).
Read Pressure Dump #aa0F >x.xxxE-xx,x.xxxE-xx, .
B-2
. .
All sensor readings in one command. Readings follow board installation sequence from left to right (front view).
Page 57
XGS-600 Gauge Controller
Table B-1 Serial Command Set (Continued)
Description Command
Response from
XGS-600 Notes
Set pressure units to Torr
Set pressure units to mBar
Set pressure units to Pascal
Read pressure units #aa13 >xx
Assign User Label to Sensor
Read User Label of Sensor
#aa10 >
#aa11 >
#aa12 >
where: xx = 00 is Torr xx = 01 is mBar xx = 02 is Pascal
#aa14cnLABEL where c sensor code =I for ion gauges HFIG and IMG =T for CNV's n = Sensor Count counting TC's or ion gauges from left to right from the front panel view. LABEL= user-defined 1-5 characters.
#aa15cn >LABEL
> Label can be any 5 or
where LABEL is user-defined 1-5 characters.
fewer characters (AHZ, 0-9, or space), except for IMGxx, HFIGx and CNVxxx, (where x is any character) which are reserved.
DRAFT 9/22/08
If no User Label is assigned, defaults to system-assigned Sensor
ID. Setup Lockout OFF #aa20 > Setup Lockout ON #aa21 > Read Setup Lockout
status
Set Auto-On #aaB0cn1cn2x.xE-xx
#aa22 >xx
where cn1 is the Ion Gauge and cn2 is the TC.
where: xx = 00 is OFF xx = 01 is ON
>
B-3
Page 58
XGS-600 Gauge Controller
Table B-1 Serial Command Set (Continued)
Description Command
Response from
XGS-600 Notes
Read Auto-On #aaB1cn
where cn is Ion Gauge.
>cnx.xExx where cn is the TC.
Ion Gauge Commands (HFIG and IMG gauges)
Assign Tube Type to Sensor
22/08
Read Sensor Tube Type #aa17cn >xx where xx =
#aa16cnxx where xx = 11 for IMG100 (default) 13 for IMG300 51 for MBA100 52 for MBA200 63 for 563 64 for 564 71 for 571 72 for 572 80 for UHV24 (default) 81 for UHV24p
DRAFT 9/
>
11 for IMG100 (default) 13 for IMG300 51 for MBA100 52 for MBA200 63 for 563 64 for 564 71 for 571 72 for 572 80 for UHV24 (default)
81 for UHV24p Set Emission OFF #aa30cn > Set Emission ON (Fil 1) #aa31cn > Also used for
Read Emission status #aa32cn >xx
Set Emission ON (Fil2) #aa33cn >
B-4
single-filament gauges.
where:
xx = 00 is OFF
xx = 01 is ON
Page 59
XGS-600 Gauge Controller
Table B-1 Serial Command Set (Continued)
Description Command
Read Filament Lit #aa34cn >xx
where: xx = 01 is Filament 1 xx = 02 is Filament 2
Response from
XGS-600 Notes
Set Auto Fil Advance Off
Set Auto Fil Advance On
Read Auto Fil Advance State
Set degas OFF #aa40cn > Set degas ON #aa41cn > Read degas status #aa42cn >xx
Read ion gauge gas correction
Set ion gauge gas correction
#aa35 >
#aa36 >
#aa37 >xx
where: xx = 00 is OFF xx = 01 is ON
where: xx = 00 is OFF xx = 01 is ON
#aa50cn >x.xxx
#aa51cnx.xxx >
DRAFT 9/22/08
Read Emission current #aa52cn >x.xxx Least significant two
Set Emission current #aa53cnx.xxx > Least significant two
Read Sensitivity #aa54cn >xx.xx Always reads in Torr.
digits always zero.
digits will be ignored.
Least significant digit is always zero.
B-5
Page 60
XGS-600 Gauge Controller
Table B-1 Serial Command Set (Continued)
Response from
Description Command
Set Sensitivity #aa55cnxx.xx > This sensitivity value
XGS-600 Notes
must always be in Torr due to xx.xx format restriction. Note however the value displayed on front panel is in Torr, mbar, or Pascal.
Least significant digit will be ignored.
Note: Will overwrite any factory setting.
Setpoint Commands
Read Setpoint States #aa03 >00hh
22/08
Read assigned setpoints #aa04cn >00hh
DRAFT 9/
Set Setpoint OFF/ON/Auto
#aa5Ehx where h is setpoint number 1-8 and x is state: 0 = OFF (manual override) 1 = ON (manual override) 3 = Auto (based on pressure) Example: #005E83 sets setpoint #8 to Auto.
where hh is a hex value
where each bit
represents
setpoints 1 - 8 and
value 0=Off, 1=On
where hh is a hex value
where each bit
represents
setpoints 1 - 8 and
value 0=Not assigned,
1=Assigned
>
Setpoint numbers correspond to the numbers on the Setpoint screen.
B-6
Page 61
XGS-600 Gauge Controller
Table B-1 Serial Command Set (Continued)
Description Command
Response from
XGS-600 Notes
Read Setpoint OFF/ON/Auto
Set Setpoint On pressure level
Set Setpoint ON Delay time
#aa5Fh where h is setpoint number 1-8
#aa6hcnx.xxxE-xx where h is setpoint number 1-8
#aaChx.x where h is setpoint number 1-8 and x.x is delay in seconds.
C is part of the hex command number. Example: #00C31.2 sets on delay to 1.2 seconds for setpoint # 3
>x where x is state: 0 = OFF (manual override) 1 = ON (manual override) 3 = Auto (based on pressure)
>
> Setpoint must first be
assigned to a sensor using setpoint screen or command 6h.
Set Setpoint Off pressure level
Set Setpoint Off Delay time
Read Setpoint On pressure level
Read Setpoint ON Delay time
Read Setpoint Off pressure level
#aa7hcnx.xxxE-xx where h is setpoint number 1 - 8
#aaDhx.x where h is setpoint number 1-8 and x.x is delay in seconds
#aa8h where h is setpoint number 1 - 8
#aaEh where h is setpoint number 1-8
#aa9h where h is setpoint number 1 - 8
> Setpoint ON level must
be set before OFF level.
> Setpoint must be
assigned to a sensor.
>x.xxxE-xx
>x.x where x.x is delay in seconds
>x.xxxE-xx
DRAFT 9/22/08
B-7
Page 62
XGS-600 Gauge Controller
Table B-1 Serial Command Set (Continued)
Description Command
Response from
XGS-600 Notes
Read Setpoint Off Delay time
#aaFh where h is setpoint number 1 - 8
CNV Commands
Calibrate CNV #aaA1cn
where cn is a CNV
Set Atmosphere Value #aaA3x.xxxE+xx
where pressure is local atmospheric value.
Read CNV gas type #aa4Ecn >x
22/08
Set CNV gas type #aa4Fcnx
where: x=0 for Nitrogen/Air x=1 for Argon
>xx
where x.x is delay in
seconds
> This command has the
same function as CAL on the main screen, i.e. used for both atmosphere and vacuum calibrations.
> Same as System Setup
screen Atmosphere Value.
where:
x=0 for Nitrogen/Air
x=1 for Argon
>
DRAFT 9/
Notes to owners of Multi-Gauge and SenTorr:
Most commands used by Multi-Gauge and SenTorr are still found in this protocol.Commands which apply to Multi-Gauge/SenTorr but do not apply to the XGS are
ignored by the XGS-600. There is no response and no error message.
Any commands (ex. # 56-59) referring to a gauge or sensor type that is not supp orted
by the XGS-600 result in ?FF error message.
B-8
NOTE All alpha characters must be upper case.
Page 63
XGS-600 Gauge Controller
Backward-compatible sensor code can still be used in XGS-600 as with Multi-Gauge/ SenTorr protocol:
Command
Entry Description Notes
c Sensor Code =T for TC/CNV
= I for ion gauge (HFIG or IMG) Case sensitive.
n Sensor Count Counting TC's or ion gauges from left to right from the front
panel view. Example: #0030I3 turns on the third ion gauge.
Using HyperTerminal
To see if communication has been established use a terminal emulator program like ProComm or Windows 3.xx Terminal or Hyperterminal.
Use the recommended pre-made cinch cable type MDC-6 Pxx w/mini-Din
connector.
Windows HyperTerminal Set-Up Instructions
1. Click Start.
2. Select Programs > Accessories > Communications > HyperTerminal.
3. Double-click Hypertrm.exe. The Connection Description dialog box appears.
4. Enter a name and choose an icon for the connection, for example: XGS-600 RS-232.
5. Click OK. The Connect To dialog box appears.
6. Select a COM port from the Connect Using: the pull-down list. The Com Properties/Port Setting dialog box appears.
7. Configure the following settings:
Bits per second (baud rate) – 9600 or 19200, whichever is set on XGS-600.Data bits: – 8 Parity: – NoneStop bits: – 1
DRAFT 9/22/08
Flow Control: – None
8. Click OK.
9. Select Properties from the File menu. The Properties window appears.
10. Configure the following settings on the Connect To tab:
Connect using – Verify the COM port is correct.
B-9
Page 64
XGS-600 Gauge Controller
Click Configure and verify the port settings are correct.
11. Ensure that the Settings tab is configured as follows:
Terminal keys radio button selected.Ctrl+H radio button selected.Auto detect selected from the Emulation pull-down list. ANSI entered in the Telnet terminal ID. 500 set in the Background buffer lines field.
12. Click ASCII Set-Up. The ASCII Set-Up dialog box appears.
13. Ensure that the settings are as follows:
Line delay – 0 millisecondsCharacter delay – 0 millisecondsASCII Receiving – Wrap lines that exceed terminal width
The cursor begins blinking.
14. Type some simple commands and see if there is a response:
#0001
22/08
#0005#000F
DRAFT 9/
B-10
Page 65
XGS-600 Gauge Controller
Appendix C. Gas Correction Factor Table
Table C-1 on page C-2 lists relative gauge gas correction factors for various gases. The table is reproduced for convenience only.
WARNING The XGS-6000 used with any ion or convection gauge is NOT
considered intrinsically safe, and should not be used with potentially flammable or explosive gas mixtures.
The values are derived by empirical methods substantiated by measurements reported in literature. This table was compiled and published by Robert L. Summers of Lewis Research Center, NASA Technical Note TND-5285, National Aeronautics and Space Administration, Washington, DC, June 1969.
To automatically convert readings of the XGS-600 Controller, normally calibrated for nitrogen, to read pressures of the other gasses:
1. Use Sensor Setup to access the GAS CORRECTION function.
2. Enter the relative gas correction constant. The gauge divides the result by the gas correction constant and displays the correct
adjusted value. The default for Gas Correction is 1.
The correction for different gas species is purely arithmetic. The sensitivity of the tube is affected by different gases which, in turn, are responsible for the tube output being manipulated by the pressure equation. There is some loss in resolution of the instrument when gas correction constants are used. The loss in resolution becomes more apparent as the correction constants approach 0.5 from either direction.
When the correction constants are 0.1 or 10, the tube output is 1/10 or 10 times normal. This causes the instrument to lose the high vacuu m dec ade or the ne ar atmo sphere deca de, respectively.
DRAFT 9/22/08
C-1
Page 66
XGS-600 Gauge Controller
Table C-1 Gas Correction Factor Table
22/08
DRAFT 9/
C-2
Page 67
XGS-600 Gauge Controller
Table C-1 Gas Correction Factor Table (Continued)
C-3
DRAFT 9/22/08
Page 68
XGS-600 Gauge Controller
This page intentionally left blank.
Page 69
Page 70
Sales and Service Offices
Canada Central coordination through: Vari a n , Inc.
121 Hartwell Avenue Lexington, MA 02421 USA Tel.: +1 781 861 7200 Toll-Free: +1 800 882 7426 Fax: +1 781 860 5437
China Varian Technologies China, Ltd.
Room 1648 Central Tower South Wing Beijing Junefield Plaza No. 10 XuanWuMenWai Street Beijing 100052 P. R. China Tel.: +86 (10) 6310 8550 Toll-Free: 800 820 6556 Fax: +86 (10) 6310 0141
France Varian s.a.
7 avenue des Tropiques Z.A. de Courtaboeuf - B.P. 12 91941 Les Ulis cedex France Tel.: +33 (0) 1 69 86 38 84 Fax: +33 (0) 1 69 86 29 88
Benelux Varian Vacuum Technologies
Herculesweg 8 4338 PL Middelburg The Netherlands Tel.: +31 118 671570 Fax: +31 118 671569
Germany & Austria Varian Deutschland GmbH
Alsfelder Strasse 6 Postfach 11 14 35 64289 Darmstadt Germany Tel.: +49 (0) 6151 703 353 Fax: +49 (0) 6151 703 302
Italy Vari a n , Inc.
via F.lli Varian 54 10040 Leini, (Torino) Italy Tel.: +39 011 997 9111 Toll-Free: 00 800 234 234 00 Fax: +39 011 997 9350
Japan Varian Technologies Japan, Ltd.
8th Floor Sumitomo Shibaura Building 4-16-36 Shibaura Minato-ku Tokyo 108 Japan Tel.: +81 3 5232 1253 Toll-Free: 0120 655 040 Fax: +81 3 5232 1710
Korea Varian Technologies Korea, Ltd.
Shinsa 2nd Bldg. 2F 966-5 Daechi-dong Kangnam-gu, Seoul Korea 135-280 Tel.: +82 2 3452 2452 Toll-Free: 080 222 2452 Fax: +82 2 3452 2451
Mexico Varian, S. de R.L. de C.V.
Concepcion Beistegui No 109 Col Del Valle C.P. 03100 Mexico, D.F. Tel.: +52 5 523 9465 Fax: +52 5 523 9472
Ta i w a n Varian Technologies Asia, Ltd.
14F-6, No. 77, Hsin Tai Wu Road, Sec. 1 Hsi chih, Taipei Hsien Taiwan, R.O.C. Tel.: +886 2 2698 9555 Toll Free: 0800 051 342 Fax: +886 2 2698 96782
United States Vari a n , Inc
121 Hartwell Avenue Lexington, MA 02421 USA Tel.: +1 781 861 7200 Toll-Free: +1 800 882 7426 Fax: +1 781 860 5437
Other Countries Varian Vacuum Technologies
via F.lli Varian 54 10040 Leini, (Torino) Italy Tel: (39) 011 997 9 111 Fax: (39) 011 997 9 350
Customer Support and Service:
North America
Tel: 1 (800) 882-7426 (toll-free) vtl.technical.support@varianinc.com
Europe
Tel: 00 (800) 234 234 00 (toll-free) vtl.technical.support@varianinc.com
Japan
Tel: (81) 3 5232 1253 (dedicated line) vtj.technical.support@varianinc.com
Korea
Tel (82) 2 3452 2452 (dedicated line) vtk.technical.support@varianinc.com
Ta i w a n
Tel: 0 (800) 051 342 (toll-free) vtw.technical.support@varianinc.com
Worl d w ide Web Site , Catalog and On-line Orders:
www.varianinc.com
Representatives in most countries
India Varian India Pvt. Ltd.
205-A, "A" wing of Galleria 2nd floor, Hiranandani Gardens Powai, Mumbai-400 076 India Tel.: +91 22 2570 8595/8597 Fax: +91 22 2570 8599 Mobile: +91 98 679 55969
04/08
UK & Ireland Vari a n L td.
6 Mead Road Oxford Industrial Park Yarnton, Oxford OX5 1QU UK Tel.: +44 (0) 1865 291570 Fax: +44 (0) 1865 291571
Loading...