Varian ProStar Wiring Manual

Page 1
ProStar
Wiring Guide
©Varian, Inc. 2001 - 2004 Printed in U.S.A. 03-914825-00:Rev. 6
Star Workstation Control
Galaxie Control
Stand-alone Control
Page 2

Star Workstation Control Contents

2
Pumps
210/215/218 X X X X X X X X X 220/230/240 X X X X
SD-1
Detectors
310 X X 320 X X X X X X X
325/335 X
330 X X 340 345 X
350/352
363 X X
AutoSampler
400 410 X X
420/430
3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
See Operation Manual
3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
X X X X X X X X X X X X X
X
X X X
500 510 701 X X X X 704 WS
800 MIB
3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
See Operation Manual
X
X
X
X
X
X X X X X X X
X
X
X
X
Page 3
Star WS Control
3
of ProStar 210/218 Pumps, 310 UV Detector, and 410 AutoSampler
comm2
comm1
connected to
comm1 or
comm2
GPIB Interface
CPU
P1 cable P2 cable
ProStar 410 AutoSampler
ProStar
210/218
Pumps
9-Pin connectors
serial communication ribbon
03-935462-91
1v 10V
Data system
output 2AUFS
1mV 10mV
Chart recorder
Output
Syn Signal
Relay 1
Relay 2
Relay 3
Relay 4
Peak sensing relay
Device Address
J9
J10
ProStar 310 UV Detector
7725i
Connected to GPIB
slot in the CPU using
cable P/N 81-
839700-00
Back connectors for ProStar 210/218
ProStar 701
(to internal CIM)
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSI
S GRD
ANALOG
OUT
ANALOG
OUT GRD
1
2
3
AUX
+5V AUX GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
1 2 3 4 5
INJECT
RACK
OUTPUTS INPUTS
PEAK
ADVANCE
+13VDC
EVT MARK
A-GND
ANALOG
GND
CHASSIS
Page 4
Star WS Control of ProStar 210/218 Pumps,
4
320 UV Detector, and 410 AutoSampler
ProStar210/218
Pumps
comm2
comm1
connected to
comm1 or
comm2
CPU
P1 cable P2 cable
ProStar 410 AutoSampler
9-Pin connectors
serial communication ribbon
03-935462-91
320 UV
ProStar
Detector Back Plate
REC HI
REC LOW
REC GRD
INT HI
INIT
LOW
INIT GRD
EXT
WAVE HI
EXT WAVE
LOW
LAMP
ON/OFF
AUTO ZERO
EVENT
MARK
D GRD
7725i
PN R007200141
To Contact 2 of 210/218 back connector
To Contact 1 of 210/ 218 back connector
ProStar 701
Back connectors for 210/218
for 210/218
(To internal CIM)
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
1
2
3
AUX +5V
AUX GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
1 2 3 4 5
INJECT
RACK
OUTPUTS INPUTS
PEAK
ADVANCE
+13VDC
EVT MARK
A-GND
ANALOG
GND
CHASSIS
Page 5
Star WS Control of
5
ProStar 210/218 Pumps, 320 UV Detector, and 430 AutoSampler with a 510 Mistral Oven
comm1
comm2
connected to
comm1 or
comm2
CPU
COM
PORT
P4
MARKER
RELAY
INPUTS
ProStar 430 AutoSampler
P6
510 Mistral Oven
ProStar 210/218 Pumps
serial communication
ribbon
9-Pin connectors
serial communication ribbon
ProStar 320 UV
Detector Back Plate
REC HI
REC
LOW
REC GRD
INT HI
INIT
LOW
INIT
GRD
EXT
WAVE HI
EXT WAVE
LOW
LAMP
ON/OFF
AUTO ZERO
EVENT
MARK
D GRD
03-935462-91
PN R007200141
To Contact 2 of 210/218 back connector
7725i
ProStar 701
To Contact 1 of 210/218 back connector
Back connectors for 210/218
(To internal CIM)
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
1
2
3
AUX +5V
AUX GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
1 2 3 4 5
PEAK
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
A-GND
ANALOG
GND
+13VDC
CHASSIS
Page 6
Star WS Control of
6
ProStar 210/218 Pumps, 325 UV Detector, 701 Fraction Collector and 410 AutoSampler
comm1
comm2
GPIB Interface
CPU
Ethernet
Cable
ProStar 210/218
Pumps
7725i
9-Pin connectors
serial communication ribbon
03-935462-91
ProStar 701
Back connectors for 210/218
(To internal CIM)
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
1
2
3
AUX
+5V
AUX
GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
R007200141
P1 cable P2 cable
ProStar 410 AutoSampler
ProStar 325
Analog Cable 110744200
Channel A or B
1 2 3 4
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
5
PEAK
A-GND
ANALOG
GND
+13VDC
CHASSIS
Page 7
Star WS Control of
7
ProStar 210/218 Pumps, 330 PDA Detector, and 410 AutoSampler
ProStar 210/218 Pumps
comm1
comm2
GPIB Interface
CPU
9-Pin connectors
serial communication ribbon
03-935462-91
7725i
To Inject & ground
of the 210/218 CIM
back connector
Back connectors for 210/218
(To internal CIM)
INTGR
1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
START (+)
1
2
GROUND (-)
3
AUX +5V
AUX
GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
D-line
connectors
p/n 03-935716-01
P1 cable P2 cable
ProStar 410 AutoSampler
ProStar 330 PDA Detector
p/n 81-839700-00
Page 8
Star WS Control of
8
ProStar 210/218 Pumps, 335 PDA Detector, 701 Fraction Collector and 410 AutoSampler
ProStar 210/218 Pumps
comm1
comm2
GPIB Interface
CPU
Ethernet
Cable
7725i
9-Pin connectors
serial communication ribbon
03-935462-91
ProStar
701
Back connectors for 210/218
(To internal CIM)
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
1
2
3
AUX +5 V
AUX GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
R007200141
1 2 3 4 5
PEAK
P1 cable P2 cable
ProStar 410
AutoSampler
Pro Star 335
Analog Cable 10744200
Channel A or B
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
A-GND
ANALOG
GND
+13VDC
CHASSIS
Page 9
Star WS Control of
9
ProStar 210/218 Pumps, 345 UV Detector, 800 MIB, and 410 AutoSampler
comm1
comm2
connected to
comm1 or
comm2
CPU
P1 cable P2 cable
ProStar 410 AutoSampler
ProStar 210/215 Pumps
serial communication ribbon
03-935462-91
Crossover cable
393320601
Analog cable to 800MIB
CH .1
CH .2
9-Pin connectors
Communication
Ethernet
800 MIB
0390793802
.01 V F /S
.10 V F /S
1.0 V F /S
ANALOG
GND
.01 V F /S
.10 V F /S
1.0 V F /S
ANALOG
GND
ProStar 345 Back Plate
Back connectors for 210/218
(To internal CIM)
7725i
STOP
GND
RUN /
START
GND
ZERO
GND
GND
GND
NOT
READY
N.O.
COMM
N.C.
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
1
2
3
AUX
+5V
AUX GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
Note: In 6.2, both CIM can be
activated and used for receiving analog signals, so the CH1 and 2 data can be input into INT hi & lo of the two
available CIMs
Page 10
Star WS Control of
10
ProStar 210/218 Pumps, 363 Fluorescence Detector, and 410 AutoSampler
comm1
comm2
connected to
comm1 or
comm2
GPIB Interface
CPU
P1 cable P2 cable
ProStar 410
AutoSampler
ProStar 210/218 Pumps
9 -Pin connectors
serial communication ribbon
03-935462-91
0393592201
PROCESSOR
RECORDER
READY/BUSY OUT
Fan
Power
ERROR OUT
START IN
AUTO ZERO OUT
LAMP OFF
MARKER IN
ProStar 363
Fluorescence Detector
Serial out
7725i
R007200141
Can send data to 800 MIB using the cable PN 390793806
ProStar 701
Back connectors for 210/218
(To internal CIM)
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
1
2
3
AUX
+5V
AUX GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
1 2 3 4 5
INJECT
RACK
OUTPUTS INPUTS
PEAK
ADVANCE
+13VDC
EVT MARK
A-GND
ANALOG
GND
CHASSIS
Page 11
Star WS Control of
11
11
ProStar 210/218 Pumps, with 500 CVM, 320 UV Detector, and 410 AutoSampler
comm1
comm2
connect to comm1
or comm2
CPU
P1 cable P2 cable
ProStar 410
AutoSampler
ProStar 210/218 Pumps
R S 4 2 2
9 -Pin connectors
serial communication ribbon
03-935462-91
ProStar 320 UV Detector Back Plate
REC HI
REC LOW
REC
GRD
INT HI
INIT
LOW
INIT
GRD
EXT
WAVE HI
EXT WAVE
LOW
LAMP
ON/OFF
AUTO ZERO
EVENT
MARK
D GRD
500 CVM
PN R007200141
To Contact 2 of 210/ 218 back connector
ProStar 701
Back connectors for 210/218
(internal CIM)
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT ANALOG OUT GRD
1
2
3
AUX +5V
AUX
GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
1 2 3 4 5
PEAK
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
A-GND
ANALOG
GND
+13VDC
CHASSIS
Page 12
Star WS Control of
12
ProStar 230 Pump, 310 UV Detector, and 410 AutoSampler
comm1
comm2
connect to comm1
or comm2
GPIB Interface
CPU
Position sensing
switch of Rheodyne
manual injector is
connected to Start In
Start In ­Start In +
Ready In -
Ready In +
Fault In -
Fault In +
Gnd
Gnd Start Out ­Start Out +
Ready Out ­Ready Out +
Fault Out ­Fault Out +
Ready In enable
+5V
J9
J10
serial communication ribbon
03-935462-91
ProStar 230 Pump
1v 10V (switch)
Data system
output 2AUFS
1mV 10mV (switch)
Chart recorder
Output
Relay 1
Relay 2
Relay 3
Relay 4
Syn Signal
Peak sensing relay
Device Address
J9
J10
GPIB cable
PN 81-839700-00
-
P1 cable P2 cable
ProStar 410
AutoSampler
1v 10V
Data system
output 2AUFS
1mV 10mV
Chart recorder
Output
Relay 1
Relay 2
Relay 3
Relay 4
Syn Signal
Peak sensing relay
Device Address
J9
J10
ProStar 310 UV Detector
ProStar 701
Connected to GPIB
slot in the CPU using
cable PN 81-839700-00
1 2 3 4 5
INJECT
RACK
OUTPUTS INPUTS
PEAK
ADVANCE
+13VDC
EVT MARK
A-GND
ANALOG
GND
CHASSIS
Page 13
Star WS Control of
13
ProStar 230 Pump, 320 UV Detector, 800 MIB, and 410 AutoSampler
comm1
comm2
connect to comm1
or comm2
GPIB Interface
serial communication ribbon
CPU
P1 cable P2 cable
ProStar 410
AutoSampler
Position sensing
switch of Rheodyne
manual injector is
connected to Start In
P/N 81-839700-00
03-935462-91
GPIB cable
Start In -
Start In + Ready In ­Ready In +
Fault In -
Fault In +
Gnd
Gnd Start Out ­Start Out +
Ready Out ­Ready Out +
Fault Out -
Fault Out +
Ready In enable
+5V
J9
J10
REC HI
REC LOW
REC GRD
INT HI
INIT
LOW
INIT GRD
EXT
WAVE HI
EXT WAVE
LOW
LAMP
ON/OFF
AUTO ZERO
EVENT
MARK
D GRD
ProStar 320 UV Detector
back plate
ProStar 230 Pump
1v 10V (switch)
Data system
output 2AUFS
1mV 10mV (switch)
Chart recorder
Output
Relay 1
Relay 2
Relay 3
Relay 4
Syn Signal
Peak sensing relay
Device Address
J9
J10
Ethernet Communication
Crossover Cable
p/n 393320601
Analog Cable
390793802
ProStar 701
NOTE: Lamp ON/OFF,
autozero can be connected
to Relay 2 & 3 respectively
of the 230 pump using the
cable P/N 02-001594-00.
800 MIB
1 2 3 4 5
PEAK
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
A-GND
ANALOG
GND
+13VDC
CHASSIS
Page 14
Star WS Control of ProStar 230 Pump,
14
330 PDA Detector, 410 AutoSampler and 701 Fraction Collector
comm1
comm2
connect to comm1
or comm2
Position sensing
switch of Rheodyne
manual injector is
connected to Start In
Start In -
Start In + Ready In ­Ready In +
Fault In -
Fault In +
Gnd
Gnd Start Out ­Start Out +
Ready Out -
Ready Out +
Fault Out -
Fault Out +
Ready In enable
+5V
J10
ProStar 230 Pump
1v 10V (switch)
Data system
output 2AUFS
1mV 10mV (switch)
Chart recorder
Output
J9
Relay 1
Relay 2
Relay 3
Relay 4
Syn Signal
Peak sensing relay
Device Address
J9
J10
GPIB cable
P/N 81-839700-00
CPU
P1 cable P2 cable
ProStar 410
AutoSampler
GPIB Interface 2 at
address 1
GPIB Interface 1 at
address 0
serial communication ribbon
03-935462-91
Analog I/O board
p/n 0393592191
ProStar 330
ProStar 701
Connected to GPIB
slot in the CPU using
cable P/N 81-
839700-00
D-line
Connectors use
cable p/n
0393571701
1 2 3 4 5
INJECT
RACK
OUTPUTS INPUTS
PEAK
ADVANCE
+13VDC
EVT MARK
A-GND
ANALOG
GND
CHASSIS
Page 15
Star WS Control of
15
ProStar 230 Pump, 363 Fluorescence Detector, 800 MIB, and 410 AutoSampler
comm1
comm2
connect to comm1
or comm2
GPIB Interface
CPU
Position sensing
switch of Rheodyne
manual injector is
connected to Start In
Start In ­Start In +
Ready In -
Ready In +
Fault In ­Fault In +
Gnd
Gnd Start Out -
Start Out + Ready Out ­Ready Out +
Ready In enable
PN 81-839700-00
GPIB cable
Fault Out -
Fault Out +
+5V
J10
J9
serial communication ribbon
03-935462-91
ProStar 230 Pump
1v
10
(switch)
Data system
output 2AUFS
Chart recorder
Output
Relay 1
Relay 2
Relay 3
Relay 4
Syn Signal
V
1mV 10mV (switch)
Peak sensing relay
Device Address
J9
J10
Ethernet Communication
Crossover Cable
393320601
800 MIB
Analog Cable for 363 only
390793806
P1 cable P2 cable
ProStar 410
AutoSampler
PROCESSOR
RECORDER
READY/BUSY OUT
Fan
ERROR OUT
AUTO ZERO OUT
MARKER IN
Power
ProStar 363
Fluorescence Detector
START IN
LAMP OFF
Serial out
ProStar 701
1 2 3 4 5
INJECT
RACK
OUTPUTS INPUTS
PEAK
ADVANCE
+13VDC
EVT MARK
A-GND
ANALOG
GND
CHASSIS
Page 16
Back Connector
16
7725i
Back connectors for SD-1
(To internal CIM)
Star WS Control of PrepStar SD-1 Pumps,
ProStar SD-1
Pumps
320 UV Detector, and 410 AutoSampler
comm1
comm2
connected to
comm1 or
comm2
CPU
P1 cable P 2 cable
ProStar 410 AutoSampler
9-Pin connectors
serial communication
ribbon
03-935462-91
320
INIT
LOW
EXT
LOW
LAMP
UV
ProStar
Detector Back Plate
REC HI
REC
LOW
REC GRD
INT HI
INIT
GRD
WAVE HI
EXT WAVE
ON/OFF
AUTO ZERO
EVENT
MARK
D GRD
PN R007200141
To Contact 2 of SD-1
back connector
To Contact 1 of SD-1
back connector
TB 3 RELAY PUTPUT
ProStar 701
TB 2 DIGITAL INPUT
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
AUX
+5V
CHASSIS
GRD
1 2 3 4 5
INJECT
RACK
OUTPUTS INPUTS
TB1 ANALOG
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
INTGR 2
HI
INTGR 2
LOW
2
INTGR
GRD
CHASSIS
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
CHASSIS
GRD
PEAK
A-GND
ADVANCE
+13VDC
EVT MARK
ANALOG
GND
CHASSIS
Page 17
Star WS Control of
17
PrepStar SD-1 Pumps, 500 CMV, 320 UV Detector, and 410 AutoSampler
comm1
comm2
connected to
comm1 or
comm2
CPU
P1 cable P2 cable
ProStar 410
AutoSampler
Back Connector
ProStar SD-1
Pumps
9-Pin connectors
serial communication
03-935462-91
Pro Star 320 UV
Detector Back Plate
REC HI
REC LOW
REC GRD
INT HI
INIT
LOW
INIT
GRD
EXT
WAVE HI
EXT WAVE
LOW
LAMP
ON/OFF
AUTO ZERO
EVENT
MARK
D GRD
500 CVM
ribbon
PN R007200141
To Contact 2 of SD-1 back connector
TB 3 RELAY PUTPUT
ProStar 701
-
Back connectors for SD-1
(To internal CIM)
TB 2 DIGITAL INPUT
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
AUX
+5V
CHASSIS
GRD
1 2 3 4 5
INJECT
RACK
OUTPUTS INPUTS
PEAK
ADVANCE
+13VDC
EVT MARK
TB1 ANALOG
INTGR 1
INTGR 1
INTGR 1
CHASSIS
INTGR 2
INTGR 2
INTGR 2
CHASSIS
CHASSIS
ANALOG
ANALOG
OUT GRD
CHASSIS
A-GND
GND
HI
LOW
GRD
GRD
HI
LOW
GRD
S GRD
GRD
OUT
GRD
ANALOG
CHASSIS
Page 18
Star WS Control of
18
PrepStar SD-1 Pumps, 320 UV Detector, 430 AutoSampler and 510 Mistral Oven
comm1
comm2
connected to
COMM1
ProStar 430 AutoSampler
P4
MARKER
RELAY
CPU
COM PORT
P6
INPUTS
Mistral ProStar 510 Oven
Back Connector
Model SD-1
Pumps
serial communication
ribbon
9-Pin connectors
serial communication
ribbon
03-935462-91
ProStar 320 UV
Detector Back Plate
REC HI
REC LOW
REC
GRD
INT HI
INIT
LOW
INIT GRD
EXT
WAVE HI
EXT WAVE
LOW
LAMP
ON/OFF
AUTO ZERO
EVENT
MARK
D GRD
7725i
TB 3 RELAY PUTPUT
PN R007200141
To Contact 2 of SD-1
back connector
To Contact 1 of SD- 1
back connector
Back connectors for SD-1
ProStar 701
(To internal CIM)
TB 2 DIGITAL INPUT
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D
GRD
MARK
D GRD
AUX
+5V
CHASSIS
GRD
1 2 3 4 5
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
TB1 ANALOG
PEAK
+13VDC
INTGR 1
HI
INTGR
LOW
INTGR 1
GRD
CHASSIS
GRD
INTGR 2
HI
INTGR 2
LOW
INTGR 2
GRD
CHASSIS
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD CHASSIS
GRD
A-GND
GND
1
ANALOG
CHASSIS
Page 19

Galaxie Control Contents

19
Pumps
210/215/218 X X X X X 220/230/240 X X X X
SD-1
20 21 22 23 24 25 26 27 28
Detectors 20 21 22 23 24 25 26 27 28
310 X X 320 X X X
325/335 X
330 340 345 X
350/352 See Operation Manual
363 X X
AutoSampler 20 21 22 23 24 25 26 27 28
400 410 X X
420/430
X
X X X X X X
500 510 701 X X X X X X X X 704 See Operation Manual WS
800 MIB X X X X X X X X X
1 2 3 4 5 6 7 8 9
X
Page 20
Galaxie CDS Control through 800 MIB of
20
ProStar 210/218 Pumps, 310 UV Detector, and 410 AutoSampler
ProStar
210/218
Pumps
9-Pin connectors
7725i
Back connectors for ProStar 210/218
(to internal CIM)
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSI
S GRD
ANALOG
OUT ANALOG OUT GRD
1
2
3
Network
800 MIB
03-907938-14 Serial Cable RJ45 to 9 Pin
P1 cable
ProStar 410 AutoSampler
P2 cable
serial communication ribbon
03-935462-91
1v 10V
Data system
output 2AUFS
1mV 10mV
Chart recorder
Output
Syn Signal
Relay 1
Relay 2
Relay 3
Relay 4
Peak sensing relay
Device Address
J9
J10
ProStar 310 UV Detector
ProStar 701
Connected to GPIB
slot in the CPU using
cable P/N 81-
839700-00
AUX +5V
AUX
GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
1 2 3 4 5
INJECT
RACK
OUTPUTS INPUTS
PEAK
ADVANCE
+13VDC
EVT MARK
A-GND
ANALOG
GND
CHASSIS
Page 21
Galaxie CDS Control through 800 MIB
21
of ProStar 210/218 Pumps, 320 UV Detector, 410 AutoSampler, and 701 FC
ProStar210/218
Pumps
9-Pin connectors
7725i
Back connectors for 210/218
for 210/218
(To internal CIM)
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG OUT GRD
1
2
3
Network
800 MIB
Connect to comm 1, 2, 3 or 4
03-907938-14
P1 cable P2 cable
ProStar 410 AutoSampler
serial communication ribbon
03-935462-91
320 UV
ProStar
Detector Back Plate
REC HI
REC LOW
REC GRD
INT HI
INIT
LOW
INIT GRD
EXT
WAVE HI
EXT WAVE
LOW
LAMP
ON/OFF
AUTO ZERO
EVENT
MARK
D GRD
PN R007200141
To Contact 2 of 210/218 back connector
To Contact 1 of 210/ 218 back connector
ProStar 701
AUX +5V
AUX GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
1 2 3 4 5
INJECT
RACK
OUTPUTS INPUTS
PEAK
ADVANCE
+13VDC
EVT MARK
A-GND
ANALOG
GND
CHASSIS
Page 22
Galaxie CDS Control through 800 MIB
22
of ProStar 210/218 Pumps, 320 UV Detector, and 430 AutoSampler with a 510 Mistral Oven
800 MIB
Network
Connect to comm 1, 2, 3 or 4
connected to
comm1 or
MARKER
RELAY
comm2
P4
03-907938-14
COM
PORT
P6
INPUTS
ProStar 430 AutoSampler
ProStar 210/218 Pumps
serial communication
510 Mistral Oven
ribbon
Detector Back Plate
9-Pin connectors
serial communication ribbon
03-935462-91
ProStar 320 UV
REC HI
REC LOW
REC
GRD
INT HI
INIT
LOW
INIT
GRD
EXT
WAVE HI
EXT WAVE
LOW
LAMP
ON/OFF
AUTO ZERO
EVENT MARK
D GRD
7725i
PN R007200141
To Contact 2 of 210/218 back connector
To Contact 1 of 210/218 back connector
ProStar 701
Back connectors for 210/218
(To internal CIM)
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
1
2
3
AUX +5V
AUX GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
1 2 3 4 5
PEAK
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
A-GND
ANALOG
GND
+13VDC
CHASSIS
Page 23
Galaxie CDS Control through 800 MIB of
23
ProStar 210/218 Pumps, 345 UV Detector, and 410 AutoSampler
ProStar 210/215 Pumps
serial communication ribbon
P1 cable P2 cable
ProStar 410 AutoSampler
03-935462-91
9-Pin connectors
03-907938-14
Analog cable to 800MIB
CH .1
CH .2
7725i
800 MIB
0390793802
.01 V F /S
.10 V F /S
1.0 V F /S
ANALOG
GND
.01 V F /S
.10 V F /S
1.0 V F /S
ANALOG
GND
ProStar 345 Back Plate
STOP
GND
RUN /
START
GND
ZERO
GND
GND
GND
NOT
READY
N.O.
COMM
N.C.
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
1
2
3
AUX
+5V
AUX
GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
Page 24
Galaxie CDS Control through 800 MIB
24
of ProStar 210/218 Pumps, 363 Fluorescence Detector, and 410 AutoSampler
ProStar 210/218 Pumps
7725i
9 -Pin connectors
Network
800 MIB
Connect to comm 1, 2, 3 or 4
03-907938-14
03-907938-06
P1 cable P2 cable
ProStar 410
AutoSampler
serial communication ribbon
03-935462-91
PROCESSOR
RECORDER
READY/BUSY OUT
Fan
Power
ERROR OUT
START IN
AUTO ZERO OUT
LAMP OFF
MARKER IN
ProStar 363
Fluorescence Detector
Serial out
R007200141
ProStar 701
AUX
+5V
AUX GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
1 2 3 4 5
INJECT
RACK
OUTPUTS INPUTS
PEAK
ADVANCE
+13VDC
EVT MARK
A-GND
ANALOG
GND
CHASSIS
Page 25
Galaxie CDS Control through 800 MIB
25
of ProStar 230 Pump, 310 UV Detector and 410 AutoSampler
GPIB Cable Extender
81-839794-00
800 MIB
Connect to
Network
comm 1, 2, 3 or 4
03-907938-14
Position sensing
switch of Rheodyne
manual injector is
connected to Start In
Start In ­Start In +
Ready In -
Ready In +
Fault In ­Fault In +
Gnd
Gnd Start Out ­Start Out +
Ready Out -
Ready Out +
Fault Out ­Fault Out +
Ready In enable
+5V
J9
J10
serial communication ribbon
03-935462-91
ProStar 230 Pump
1v 10V (switch)
Data system
output 2AUFS
1mV 10mV (switch)
Chart recorder
Output
Relay 1
Relay 2
Relay 3
Relay 4
Syn Signal
Peak sensing relay
Device Address
J9
J10
GPIB cable
PN 81-839700-00
-
P1 cable P2 cable
ProStar 410
AutoSampler
1v 10V
Data system
output 2AUFS
1mV 10mV
Chart recorder
Output
Relay 1
Relay 2
Relay 3
Relay 4
Syn Signal
Peak sensing relay
Device Address
J9
J10
ProStar 310 UV Detector
ProStar 701
Connected to GPIB
slot in the CPU using
cable PN 81-839700-00
1 2 3 4 5
INJECT
RACK
OUTPUTS INPUTS
PEAK
ADVANCE
+13VDC
EVT MARK
A-GND
ANALOG
GND
CHASSIS
Page 26
Galaxie CDS Control through 800 MIB
26
of ProStar 230 Pump, 320 UV Detector, and 410 AutoSampler
Position sensing
switch of Rheodyne
manual injector is
connected to Start In
Start In -
Start In + Ready In ­Ready In +
Fault In -
Fault In +
Gnd
Gnd Start Out ­Start Out +
Ready Out -
Ready Out +
Fault Out -
Fault Out +
Ready In enable
+5V
J10
ProStar 230 Pump
1v 10V (switch)
Data system
output 2AUFS
1mV 10mV (switch)
Chart recorder
Output
J9
Relay 1
Relay 2
Relay 3
Relay 4
Syn Signal
Peak sensing relay
Device Address
J9
J10
03-907938-14
03-907938-14
GPIB Cable Extender 81-839704-00
800 MIB
Network
P1 cable P2 cable
ProStar 410
AutoSampler
serial communication ribbon
03-935462-91
REC HI
REC LOW
REC GRD
INT HI
INIT
LOW
INIT GRD
EXT
WAVE HI
EXT WAVE
LOW
LAMP
ON/OFF
AUTO ZERO
EVENT
MARK
D GRD
ProStar 320 UV Detector
back plate
NOTE: Lamp ON
autozero can be connected to Relay 2 & 3 respectively
of the 230 pump using the
cable P/N 02-001594-00.
/OFF,
Analog Cable 390793802
ProStar 701
OUTPUTS INPUTS
1 2 3 4 5
PEAK
INJECT
ADVANCE
RACK
EVT MARK
A-GND
GND
+13VDC
ANALOG
CHASSIS
Page 27
Galaxie CDS Control through 800 MIB
27
of ProStar 230 Pump, 325/335 PDA Detector, 410 AutoSampler and Fraction Collector
GPIB Cable Extender
81-839704-00
800 MIB
Connect to comm 1, 2, 3 or 4
Position sensing
switch of Rheodyne
manual injector is
connected to Start In
Start In -
Start In +
Ready In -
Ready In +
Fault In -
Fault In +
Gnd Gnd
Start Out -
Start Out + Ready Out ­Ready Out +
Fault Out -
Fault Out +
Ready In enable
+5V
J10
ProStar 230 Pump
1v 10V (switch)
Data system
output 2AUFS
1mV 10mV (switch)
Chart recorder
Output
J9
Syn Signal
Relay 1
Relay 2
Relay 3
Relay 4
Peak sensing relay
Device Address
J9
J10
GPIB cable
P/N 81-839700-00
Network
(required for 335)
Acquistion Server
Ethernet
Ethernet Switch
03-907938-14
P1 cable P2 cable
ProStar 410
AutoSampler
serial communication ribbon
03-935462-91
ProStar 325/335
ProStar 701
D-line
Connectors use
cable p/n
0393571701
1 2 3 4 5
INJECT
RACK
OUTPUTS INPUTS
PEAK
ADVANCE
+13VDC
EVT MARK
A-GND
ANALOG
GND
CHASSIS
Page 28
Galaxie CDS Control of ProStar 230 Pump,
28
363 Fluorescence Detector, 800 MIB, and 410 AutoSampler
Position sensing
switch of Rheodyne
manual injector is
connected to Start In
Start In -
Start In + Ready In ­Ready In +
Fault In +
Start Out ­Start Out +
Ready Out -
Ready Out +
Fault Out -
Fault Out +
Ready In enable
PN 81-839700-00 GPIB cable
Fault In -
Gnd Gnd
+5V
J9
J10
ProStar 230 Pump
Data system
output 2AUFS
Chart recorder
serial communication ribbon
03-935462-91
1v
1mV 10mV (switch)
Output
Relay 1
Relay 2
Relay 3
Relay 4
Syn Signal
10
(switch)
V
Peak sensing relay
Device Address
J9
J10
800 MIB
Network
Analog Cable for 363 only
390793806
P1 cable P2 cable
ProStar 410
AutoSampler
PROCESSOR
RECORDER
READY/BUSY OUT
Fan
ERROR OUT
AUTO ZERO OUT
MARKER IN
Power
ProStar 363
Fluorescence Detector
START IN
LAMP OFF
Serial out
ProStar 701
1 2 3 4 5
INJECT
RACK
OUTPUTS INPUTS
PEAK
ADVANCE
+13VDC
EVT MARK
A-GND
ANALOG
GND
CHASSIS
Page 29
Stand-alone Control Contents
29
Pumps 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46
210/215/218 X X X X X 220/230/240 X X X X X X X
SD-1
Detectors 30 31 32 33 34 35 36 37 38 39 41 41 42 43 44 45 46
310 X X X X X X 320 X X X X X X
325/335
330 340 X X X 345 X X
350/353 See Operation Manual
363
AutoSampler 30 31 32 33 34 35 36 37 38 39 41 41 42 43 44 45 46
400 410 X X
420/430
30 31 32 33 34 35 36 37 38 39 41 41 42 43 44 45 46
X
X X
X X
X X
X
X X
X X X X
X
X
X
X
X
X
500 510 701 X X X X X X X X X X X X X X X X X 704 See Operation Manual WS
800 MIB
Page 30
ProStar 220/230/240 Pump (Master),
30
310 UV Detector, and 410 AutoSampler
Start In -
Start In + Ready In ­Ready In +
J9
Fault In -
Fault In +
Gnd
Sync Cable Kit 03-91994890
Position sensing switch of Rheodyne manual injector is connected to Start In
Sync cable for ProStar 410 AutoSampler 0393516301
ProStar 410 AutoSampler
Gnd Start Out ­Start Out +
Ready Out ­Ready Out +
Fault Out ­Fault Out +
Ready In enable
+5V
P1 cable P2 cable
J10
Start In -
Start In + Ready In ­Ready In +
Fault In -
Fault In +
Gnd
Gnd Start Out ­Start Out +
Ready Out -
Ready Out +
Fault Out ­Fault Out +
Ready In enable
+5V
ProStar 220/230/240 Pump
1v 10V (switch)
Data system
output 2AUFS
1mV 10mV (switch)
Chart recorder
Output
Relay 1 Relay 2
Peak sensing relay
Device Address
Relay 3 Relay 4
J9
Syn Signal
J10
J9
1v 10V (switch)
Data system
J10
output 2AUFS Chart recorder
Output
Relay 1 Relay 2
1mV
10mV (switch)
Peak sensing relay
Device Address
Relay 3 Relay 4
Syn Signal
J9
J10
ProStar 310 UV-Detector
ProStar 701
Detail
Signal Cable 02-001954-00
Clear +ve
12345
PEAK
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
A-GND
GND
+13VDC
ANALOG
CHASSIS
Black -ve
Page 31
210/215/218 Pump (Master),
31
320 UV Detector, and 410 AutoSampler
Extension cable R000815107
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
1 2
3
Sync cable for ProStar 410 AutoSamplers 0393516301
P1 cable P2 cable
ProStar 410 AutoSam
pler
210/215/218 Pump
Shielded cable R007200141
ProStar 320 UV-Detector
REC HI
REC
LOW
REC GRD
INT HI
INIT
LOW
INIT
GRD
EXT
WAVE HI
EXT WAVE
LOW
LAMP
ON/OFF
AUTO ZERO
EVENT
MARK
D GRD
Position sensing switch of Rheodyne manual injector is connected to Start In
Shielded cable R007200141
NOTE
: Lamp ON/OFF, autozero can be connected to Relay 2 & 3 respectively of the ProStar 220/230/240 pump using the cable 02-001594-00.
AUX +5V
AUX GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
ProStar 701
Detail
Shielded cable R007200141
12345
PEAK
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
A-GND
GND
+13VDC
ANALOG
CHASSIS
Page 32
210/215/218 Pump (Master), 320 UV
32
Detector, and 400 AutoSampler
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
1 2
3
Sync cable for ProStar 400 AutoSamplers 0393516201
COMM
PORT
CONTROL I/O
ProStar 400 AutoSampler
Extension cable R000815107
210/215/218 Pump
Shielded cable R007200141
REC HI
REC
LOW
REC GRD
INT HI
INIT
LOW INIT
GRD
EXT
WAVE HI EXT WAVE
LOW
LAMP
ON/OFF
AUTO ZERO
EVENT
MARK
D GRD
ProStar 320 UV-Detector
Position sensing switch of Rheodyne manual injector is connected to Start In
Shielded cable R007200141
NOTE
: Lamp ON/OFF, autozero can be connected to Relay 2 & 3 respectively of the ProStar 220/230/240 pump using the cable 02-001594-00.
AUX
+5V
AUX GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD INJECT
D
GRD
MARK
D GRD
ProStar 701
Detail
Shielded cable R007200141
12345
PEAK
A-GND
GND
+13VDC
OUTPUTS INPUTS
INJECT
ADVANCE
RACK
EVT MARK
ANALOG
CHASSIS
Page 33
210/215/218 Pump (Master), 320 UV
33
Detector, and 420/430 AutoSampler
Extension cable
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
R000815107
1 2
3
Sync cable for ProStar 420/430 AutoSamplers 0393516101
COM PORT
P4
MARKER
RELAY
P6
INPUTS
ProStar 420/430 AutoSampler
210/215/218 Pump
Shielded cable R007200141
REC HI
REC
LOW
REC GRD
INT HI
INIT
LOW INIT
GRD
EXT
WAVE HI
EXT WAVE
LOW
LAMP
ON/OFF
AUTO ZERO
EVENT MARK
D GRD
ProStar 320 UV-Detector
Position sensing switch of Rheodyne manual injector is connected to Start In
Shielded cable R007200141
NOTE
: Lamp ON/OFF, autozero can be connected to Relay 2 & 3 respectively of the ProStar 220/230/240 pump using the cable 02-001594-00.
AUX
+5V
AUX GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
ProStar 701
Detail
Shielded cable R007200141
12345
PEAK
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
A-GND
GND
+13VDC
ANALOG
CHASSIS
Page 34
210/215/218 Pump (Master), 340 UV
34
Detector, and 420/430 AutoSampler
NOTE:
Please refer to the other detector diagrams for connection with the other AutoSamplers
Sync cable for ProStar 420/430 AutoSampler 0393516101
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
1
2
3
COM
PORT
Extension cable R000815107
210/215/218 Pump
Shielded cable R007200141
OUTPUT REMOTE
AUTO
INT
RECORDER
+ - + -
1AU/V
ZERO
+ -
1 2 3 4
Position sensing switch of Rheodyne manual injector is connected to Start In
Shielded cable R007200141
EVENT
+ - + -
5 6
AUX +5V
AUX
GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK
D GRD
ProStar 701
Shielded cable R007200141
P4
MARKER
RELAY
P6
INPUTS
ProStar 420/430 AutoSampler
ProStar 340 UV-Detector
Detail
1 2 3 4 5
INJECT
RACK
OUTPUTS INPUTS
PEAK
ADVANCE
+13VDC
EVT MARK
A-GND
ANALOG
GND
CHASSIS
Page 35
ProStar 220/230/240 Pump (Master),
35
340 UV Detector, and 410 AutoSampler
Position sensing switch of Rheodyne manual injector is connected to Start In
Start In -
Start In +
Ready In -
Ready In +
Fault In ­Fault In +
Gnd Gnd
Start Out -
Start Out +
Ready Out -
Ready Out +
Fault Out ­Fault Out +
Ready In enable
+5V
J9
J10
ProStar 220/230/240 Pump
10V
1v
Data system
output 2AUFS Chart recorder
Output
Relay 1 Relay 2 Relay 3 Relay 4
Syn Signal
(switch)
1mV 10mV (switch)
Peak sensing relay
Device Address
J9
J10
Shielded cable, R007200141
Signal Cable 02-001954-00
NOTE:
Please refer the appropriate diagram for connections to the other AutoSamplers
Sync cable for ProStar 410 AutoSampler 0393516301
P1 cable P2 cable
ProStar 410 AutoSampler
OUTPUT REMOTE
RECORDER
+ - + -
12 3 4
INT
1AU/V
ProStar 340 UV-Detector back plate
To the appropriate
Integrator
AUTO ZERO
+ -
EVENT
+ - + -
56
Detail
ProStar 701
Clear +ve
12345
PEAK
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
A-GND
GND
+13VDC
Black +ve
ANALOG
CHASSIS
Page 36
ProStar 220/230/240 Pump (Master), 345 UV and
36
360 Fluorescence Detectors, and 410 AutoSampler
Position sensing switch of Rheodyne manual injector is connected to Start In
NOTE:
Please refer the appropriate diagram for connections to the other AutoSamplers
Start In ­Start In +
Ready In -
Ready In +
J9
Fault In ­Fault In +
Gnd Gnd
Start Out -
Start Out +
Ready Out -
J10
Ready Out +
Fault Out -
Fault Out +
Ready In enable
+5V
Shielded cable R007200141
Sync cable for ProStar 410 AutoSampler 0393516301
P1 cable P2 cable
ProStar 410 Autosampler
ProStar 220/230/240 Pump
1v 10V (switch)
Data system
output 2AUFS Chart recorder
Output
Relay 1 Relay 2 Relay 3 Relay 4
Syn Signal
1mV
10mV (switch)
Peak sensing relay
Device Address
J9
J10
STOP
GND
RUN/START
GND
ZERO
GND
GND
GND
NOT READY
N.O.
COMM
N.C.
ProStar 345/360 Detector
CH.1
CH.2
Signal Cable 02-001954-00
To the appropriate
integrator
.01 V F/S
.10 V F/S
1.0 V F/S ANALOG
GND
.01 V F/S .10 V F/S
1.0 V F/S
ANALOG
GND
ProStar 701
Detail
Signal Cable 02-001954-00
Shielded cable R007200141
Clear +ve
12345
PEAK
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
A-GND
GND
+13VDC
ANALOG
CHASSIS
Black +ve
Page 37
210/215/218 Pump (Master), 345 UV & 360
37
Fluorescence Detectors, and 420/430 AutoSampler
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
1
P4
MARKER
RELAY
2 3
NOTE:
Please refer to the other detector diagrams for connection with the other AutoSamplers
Sync cable for ProStar 420/430 AutoSamplers 0393516101
ProStar 420/430 AutoSampler
Extension cable R000815107
COM
PORT
P6
INPUTS
210/215/218 Pump
Shielded cable R007200141
STOP
GND
RUN/START
GND
ZERO
GND
GND
GND
NOT READY
N.O.
COMM
N.C.
ProStar 345/360 Detector
CH.1
CH.2
Position sensing switch of Rheodyne manual injector is connected to Start In
.01 V F/S .10 V F/S
1.0 V F/S ANALOG
GND
.01 V F/S
.10 V F/S
1.0 V F/S ANALOG
GND
AUX
+5V AUX GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD
MARK D GRD
Shielded cable R007200141
ProStar 701
Detail
Shielded cable R007200141
12345
PEAK
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
A-GND
GND
+13VDC
ANALOG
CHASSIS
Page 38
ProStar 220/230/240 Pump
38
(Master), 310 UV Detector, and 400 AutoSampler
Start In -
Start In + Ready In ­Ready In +
J9
Fault In -
Fault In +
Gnd
Sync Cable Kit 03-91994890
Position sensing switch of Rheodyne manual injector is connected to Start In
Sync cable for ProStar 400 AutoSampler 0393516201
COMM
PORT
Gnd Start Out ­Start Out +
Ready Out ­Ready Out +
Fault Out -
Fault Out +
Ready In enable
+5V
CONTROL I/O
J10
Start In -
Start In + Ready In ­Ready In +
Fault In -
Fault In +
Gnd
Gnd Start Out ­Start Out +
Ready Out -
Ready Out +
Fault Out ­Fault Out +
Ready In enable
+5V
ProStar 220/230/240 Pump
1v
Data system
output 2AUFS Chart recorder
Output
Relay 1 Relay 2
10V (switch)
1mV 10mV (switch)
Peak sensing relay
Device Address
Relay 3 Relay 4
Syn Signal
J9
J10
J9
10V (switch)
1v
Data system
J10
output 2AUFS Chart recorder
Output
Relay 1 Relay 2
1mV
10mV (switch)
Peak sensing relay
Device Address
Relay 3 Relay 4
Syn Signal
J9
J10
ProStar 310 UV-Detector
ProStar 701
Detail
Signal Cable 02-001954-00
Clear +ve
12345
PEAK
INJECT
ADVANCE
A-GND
ANALOG
OUTPUTS INPUTS
RACK
EVT MARK
GND
+13VDC
CHASSIS
ProStar 400 AutoSampler
Black -ve
Page 39
ProStar 220/230/240 Pump (Master), 320
39
UV Detector, and 400 AutoSampler
Position sensing switch of Rheodyne manual injector is connected to Start In
Start In ­Start In +
Ready In -
Ready In +
Fault In ­Fault In +
Gnd Gnd
Start Out -
Start Out +
Ready Out -
Ready Out +
Fault Out -
Fault Out +
Ready In enable
+5V
Sync cable for ProStar 400 AutoSampler 0393516201
COMM PORT
CONTROL I/O
ProStar 400 AutoSampler
J10
ProStar 220/230/240 Pump
1v 10V (switch)
Data system
output 2AUFS
1mV 10mV (switch)
Chart recorder
Output
J9
Relay 1 Relay 2 Relay 3 Relay 4
Syn Signal
Peak sensing relay
Device Address
J9
J10
REC HI
REC
LOW
REC GRD
INT HI
INIT
LOW
INIT
GRD
EXT
WAVE HI EXT WAVE
LOW
LAMP
ON/OFF
AUTO ZERO
EVENT
MARK
D GRD
ProStar 320 UV-Detector
back plate
ProStar 701
NOTE: Lamp ON/OFF,
autozero can be connected to Relay 2 & 3 respectively of the ProStar 220/230/240 pump using the cable 02-001594-00.
Detail
Signal Cable 02-001954-00
Clear +ve
12345
PEAK
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
A-GND
GND
+13VDC
ANALOG
CHASSIS
Black -ve
Page 40
ProStar 220/230/240 Pump (Master), 320
40
UV Detector, and 420/430 AutoSampler
Position sensing switch of Rheodyne manual injector is connected to Start In
Start In -
Start In + Ready In ­Ready In +
Fault In -
Fault In +
Gnd
Gnd Start Out ­Start Out +
Ready Out ­Ready Out +
Fault Out -
Fault Out +
Ready In enable
+5V
Sync cable for ProStar 420/430 AutoSampler 0393516101
COM
PORT
P4
MARKER
RELAY
P6
INPUTS
ProStar 420/430 AutoSampler
J10
ProStar 220/230/240 Pump
1v
Data system
output 2AUFS
Chart recorder
Output
J9
Relay 1 Relay 2
10V (switch)
1mV 10mV (switch)
Peak sensing relay
Device Address
Relay 3 Relay 4
Syn Signal
J9
J10
Signal Cable 02-001954-00
REC HI
REC LOW
REC GRD
INT HI
INIT
LOW INIT
GRD
EXT
WAVE HI
EXT WAVE
LOW
LAMP
ON/OFF
AUTO ZERO
EVENT
MARK
D GRD
NOTE: Lamp ON/OFF, autozero can be connected to Relay 2 & 3 respectively of the ProStar 220/230/240 pump using the cable 02-001594-00.
ProStar 701
Detail
Clear +ve
12345
PEAK
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
A-GND
GND
+13VDC
ANALOG
CHASSIS
ProStar 320 UV-Detector
back plate
Black -ve
Page 41
ProStar 220/230/240 Pump (Master),
41
310 UV Detector, and 420/430 AutoSampler
Start In -
Start In +
Ready In -
Sync Cable Kit 03-91994890
switch of Rheodyne manual injector is connected to Start In
P4
MARKER
RELAY
Ready In +
Start Out + Ready Out ­Ready Out +
Fault Out +
Ready In enable
COM
POR
P6
INPUTS
Fault In ­Fault In +
Gnd Gnd
Start Out -
Fault Out -
+5V
T
J9
J10
Start In -
Start In + Ready In ­Ready In +
Fault In -
Fault In +
Gnd
Gnd Start Out ­Start Out +
Ready Out -
Ready Out +
Fault Out ­Fault Out +
Ready In enable
+5V
Sync cable for ProStar 420/430 AutoSampler 0393516101
ProStar 220/230/240 Pump
1v 10V (switch)
Data system
output 2AUFS
1mV 10mV (switch)
Chart recorder
Output
Relay 1 Relay 2
Peak sensing relay
Device Address
Relay 3 Relay 4
Syn Signal
J9
J10
J9
10V (switch)
1v
Data system
J10
output 2AUFS Chart recorder
Output
Relay 1 Relay 2
1mV
10mV (switch)
Peak sensing relay
Device Address
Relay 3 Relay 4
Syn Signal
J9
J10
ProStar 310 UV-Detector
ProStar 701
Detail
Signal Cable 02-001954-00Position sensing
Clear +ve
12345
PEAK
INJECT
ADVANCE
A-GND
ANALOG
ProStar 420/430 AutoSampler
OUTPUTS INPUTS
RACK
EVT MARK
GND
+13VDC
CHASSIS
Black -ve
Page 42
210/215/218 Pump (Master),
42
310 UV Detector, and 400 AutoSampler
210/215/218 Pump
Start In -
Start In + Ready In ­Ready In +
Start Out ­Start Out +
Ready Out -
Ready Out +
Fault Out ­Fault Out +
Ready In enable
Fault In -
Fault In +
Gnd Gnd
+5V
J9
J10
Sync cable for ProStar 400 AutoSampler 0393516201
COMM
PORT
CONTROL
INTGR
HI
INTGR
LOW
INTGR
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT
1 2
3
I/O
Extension R000815107
Shielded R007200141
Data system
output 2AUFS Chart recorder
SyN Signal
ProStar 310 UV-Detectorr
Output
Relay Relay Relay Relay
1v
1mV
1 2 3 4
Position sensing switch of Rheodyne manual injector is connected to Start In
10V (switch)
10mV (switch)
Sensing relay
Peak
Address
Device
J9
J10
Shielded R007200141
ProStar 701
AUX +5V
AUX
GND STOP
GND
D
HOLD
GND
D
TRANSFER
GND
D
INJECT
GND
D
MARK
GND
D
Detail
Shielded R007200141
1234 5
PEAK
INJECT
ADVANCE
A-GND
ANALOG
ProStar 400 AutoSampler
RACK
OUTPUTS INPUTS
EVT MARK
GND
+13VDC
CHASSIS
Page 43
210/215/218 Pump (Master), 310 UV
43
Detector, and 420/430 AutoSampler
COM
PORT
P6
INPUTS
Extension cable R000815107
Sync cable for ProStar 420/430 AutoSampler 0393516101
P4
MARKER
RELAY
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
1 2
3
210/215/218 Pump
Start In -
Start In + Ready In ­Ready In +
J9
Fault In -
Fault In +
Gnd
Gnd Start Out ­Start Out +
Ready Out -
J10
Ready Out +
Fault Out -
Fault Out +
Ready In enable
+5V
Position sensing switch of Rheodyne manual injector is connected to Start In
Shielded cable R007200141
1v 10V (switch)
Data system
output 2AUFS
1mV 10mV (switch)
Chart recorder
Output
Syn Signal
Relay 1 Relay 2 Relay 3 Relay 4
Peak sensing relay
Device Address
J9
J10
ProStar 310 UV-Detector
Shielded cable R007200141
ProStar 701
AUX
+5V AUX
GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD INJECT
D GRD
MARK
D GRD
Detail
Shielded cable
R007200141
1234
PEAK
INJECT
ADVANCE
A-GND
5
ANALOG
ProStar 420/430 AutoSampler
RACK
OUTPUTS INPUTS
EVT MARK
GND
+13VDC
CHASSIS
Page 44
210/215/218 Pump (Master), 310
44
UV Detector, and 410 AutoSampler
INTGR 1 INTGR 1 INTGR 1
CHASSIS
ANALOG
ANALOG
OUT GRD
1 2
3
Sync cable for ProStar 410 AutoSamplers 0393516301
P1 cable P2 cable
ProStar 410 AutoSampler
LOW GRD GRD
OUT
Extension
HI
cable R000815107
210/215/218 Pump
Position sensing switch of Rheodyne manual injector is connected to Start In
AUX
+5V
AUX GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD
INJECT
D GRD MARK D GRD
Shielded cable
R007200141
Shielded cable R007200141
Start In -
Start In + Ready In ­Ready In +
Fault In -
Fault In +
Gnd
Gnd Start Out ­Start Out +
Ready Out -
Ready Out +
Fault Out -
Fault Out +
Ready In enable
+5V
J9
J10
Data system
output 2AUFS
1mV
Chart recorder
Output
Relay 1 Relay 2 Relay 3
Relay 4
Syn
Signal
1v 10V (switch)
10mV (switch)
Peak sensing relay
Device Address
J9
J10
ProStar 701
Shielded cable R007200141
1234 5
ProStar 310 UV-Detector
PEAK
+13VDC
A-GND
GND
ANALOG
CHASSIS
Detail
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
Page 45
210/215/218 Pump (Master), 340 UV
45
Detector, and 420/430 AutoSampler
INTGR 1
HI
INTGR 1
LOW
INTGR 1
GRD
CHASSIS
GRD
ANALOG
OUT
ANALOG
OUT GRD
1
P4
MARKER
RELAY
2 3
NOTE:
Please refer to the other detector diagrams for connection with the other AutoSamplers
Sync cable for ProStar 420/430 AutoSampler 0393516101
ProStar 420/430 AutoSampler
COM
PORT
P6
INPUTS
Extension cable R000815107
210/215/218 Pump
Shielded cable R007200141
OUTPUT REMOTE
AUTO
INT
RECORDER
+ - + -
1AU/V
ZERO
+ -
12 3 4
ProStar 340 UV-Detector
Position sensing switch of Rheodyne manual injector is connected to Start In
Shielded cable R007200141
EVENT
+ - + -
56
AUX
+5V
AUX GND
STOP
D GRD
HOLD
D GRD
TRANSFER
D GRD INJECT
D GRD MARK
D GRD
ProStar 701
Detail
Shielded cable R007200141
12345
PEAK
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
A-GND
GND
+13VDC
ANALOG
CHASSIS
Page 46
ProStar 220/230/240 Pump (Master), 320
46
UV Detector, and 410 AutoSampler
Position sensing switch of Rheodyne manual injector is connected to Start In
ProStar410 AutoSampler
Start In -
Start In + Ready In ­Ready In +
Fault In -
Fault In +
Gnd
Gnd Start Out ­Start Out +
Ready Out ­Ready Out +
Fault Out -
Fault Out +
Ready In enable
+5V
Sync cable for ProStar 410 0393516301
P1
cable
P2
cable
J10
ProStar 220/230/240 Pump
1v
Data system
output 2AUFS
Chart recorder
Output
J9
Relay 1 Relay 2
10V (switch)
1mV 10mV (switch)
Peak sensing relay
Device Address
Relay 3 Relay 4
Syn Signal
J9
J10
Signal Cable 02-001954-00
REC HI
REC LOW
REC GRD
INT HI
INIT
LOW INIT
GRD
EXT
WAVE HI EXT WAVE
LOW
LAMP
ON/OFF
AUTO ZERO
EVENT
MARK
D GRD
NOTE: Lamp ON/OFF, autozero can be connected to Relay 2 & 3 respectively of the ProStar 220/230/240 pump using the cable 02-001594-00.
ProStar 701
Detail
Clear +ve
12345
PEAK
INJECT
ADVANCE
RACK
OUTPUTS INPUTS
EVT MARK
A-GND
GND
+13VDC
ANALOG
CHASSIS
ProStar 320 UV-Detector
back plate
Black -ve