This document describes the configuration of the Varian ProStar
210, 215 and Varian PrepStar 218, SD-1 pump driver. It also
explains how to build a control method for this driver, how to start
and view a running acquisition and what actions are available
during an analysis.
The ProStar 210, 215 and PrepStar 218 Solvent Delivery
Modules are identical in appearance and operation. The ProStar
215 and PrepStar 218 have more powerful motors, larger power
supplies and can operate at higher pressures than the ProStar
210. The pressures are identified in the Appendix.
The manual will refer to the pumps as 210/215/218/SD-1
throughout.
To install the Galaxie Drivers software refer to the Galaxie Data
System Installation Guide.
Configuring the Communication Server
To communicate with the ProStar 210-215-218-SD-1 module a
RS232_Interface or a RS232_PC communication bus must be
configured on the acquisition server.
The parameters of the bus to must be the following:
Stops 1 Stop
Baud 19200
Bits 8 Bits
Buffer Tx 4096
NOTE: For more details about the configuration of a RS232_Interface or a RS232_PC
bus, refer to the Galaxie Chromatography Data System Installation’s Guide.
The ID entered here
must match the ID of
the pump.
Select from the list the type of the pump: isocratic, binary,
ternary, quaternary, isocratic + injector, binary + injector, ternary
+ injector. A new tab (Pump B, Pump C, Pump D or Injector)
appears depending on the selected type of pump.
NOTE:
- The system must be equipped by at least two ProStar / PrepStar modules to work as a
binary pump, three modules for a ternary one and four to work in quaternary mode.
- With this type of pump, a module can be used as an injector. In this case a new
Injection screen becomes active in the method. To learn more about this option, refer to
the chapter
Control method for the ProStar / PrepStar pump.
- When an Injector module is configured, in the lower part of the Galaxie Configuration
Manager an injector (of the new system) called ProStar 210/215/218 pump is created.
Enter the identification number of the pump module in the RS-422 unit ID field. The RS-422 unit ID field must match the ID
specified on the pump itself. To check the ID on the pump, press
Setup on the pump (no flow), scroll left until the ID is displayed.
Press Enter when finished.
Select from the list the model of the pump module. Four choices
are available: ProStar 210, 215, PrepStar 218 and SD-1.
The 210/215/218 modules are identical in appearance and
operation. They differ only in one respect: the ProStar 215 and
PrepStar 218 have more powerful motors, larger power supplies
and can operate at higher pressures than the ProStar 210.
NOTE: The pump head size must be configured on the instrument. Maximum flow rates
authorized in Galaxie will then depend on that configuration.
- With ProStar 210/215 and PrepStar 218: press the ~ button to access the setup menu
and enter the correct head size under Hdsz section using the ▲ and ▼ buttons.
- With the PrepStar SD-1: press the ~ button to access the setup menu. Select YES
under Change Pump using the ▲ and ▼ buttons and follow the procedure displayed in
the LCD screen. At the end of the procedure, new pump head size will be automatically
configured.
Please refer to the instrument manual for more details.
3. Go back to the System Configuration screen (Properties
wizard). A graphic of the devices that comprise the system is
displayed in the synopsis screen. To configure the synopsis
as the real instrument configuration, move each module, and
then connect it appropriately. To do this, select one anchor
for each of the two modules (the selected anchors become
red) and click on the
button; a connection tube will
appear between the two anchors.
Click on the
the
button to delete all the tubes.
Click on
button to delete a selected tube, click on
to define a background color, or on to
make the background transparent.
Click on the
icon to change the grid color.
4. Once the synopsis configuration is correctly done, it will be
displayed in Galaxie Chromatography Data System, in the
Systems tab. In this tab, the Overview part will show the
synopsis of the system with some parameters displayed
(temperature, flow, signal, etc.) depending on the installed
devices.
For example, the following pictures represent the synopsis of
the ProStar / PrepStar pump:
NOTE: Once the system has been created and correctly configured, it must be
associated with one or more projects in order to start acquisitions. For more
details about the creation and configuration of a system, refer to the Galaxie
Configuration Manager User’s Guide (chapter The System).
The method contains parameters for instrument control, data
acquisition, chromatogram processing, and editing of final
results. To initially set up a method, it is not necessary to define
all of the sub-method sections; however, you must fill inthe
instrument control section prior to starting an acquisition.
1. To create a new method, select the FILE / NEW METHOD
option from the Galaxie Chromatography Data System main
menu.
A wizard guide is displayed to provide assistance during the
first steps of method creation:
2. Choose the system (chromatograph) associated with the
method. A method is created for a particular system. When
starting an acquisition, the name of the system that performs
the acquisition must be correct, since method access is
limited only to those methods associated with this system.
3. Once this field is completed click on the Next button to move
to the second step of the method creation:
4. Enter the New Method Name in the first field. In the Description field, enter information concerning the method
(optional).
5. Once these two fields have been completed, click on the OK
button. The new method is created and is opened with
default parameters in the Galaxie Chromatography Data
System. The next step is the definition of each of the method
sections.
The instrument control section must be defined in order to
start an acquisition. In the following part of this guide, only
the instrument control section of the method for the ProStar /
PrepStar pump is detailed. Other parameters such as
Injection Volume, Vial and Rack number could be necessary.
They are defined in the acquisition sub-method part.
NOTE: For more details about the programming of each method section, refer to the
Press the button to display the System Control Method Advanced Tools window:
to send the current method to the instrument (not
applicable to this driver)
to get the active method from the instrument (not
applicable to this driver)
to print the current method
Click on the button to display the ProStar / PrepStar
pump control method section. It contains up to four sub-sections:
Elution, Miscellaneous, Pressure and Injection. Click on the
required sub-section to define the corresponding parameters that
will be used during the chromatographic run. The toolbar allows
you to easily scroll through the control method windows:
to select the first window of the control method.
to move back to the previous window of the control
method.
to move forward to the next window of the control
method.
A compressibility factor and a pressure constant are associated
with each solvent (Compressibility and Pressure Constant fields).
When the Compressibility compensation field is unchecked, the
pump will not take into account the corresponding factors and
the factors will be concealed.
Refer to the Compressibility and Pressure Constant section in
the appendix of this manual to find the list of these values.
The refill time parameter is the total time that it takes to refill the
pump head with mobile phase. This value may be changed
when a very viscous mobile phase is used. In this case, a rapid
refill might lead to cavitations or bubble formation in the pump
head causing a lower than expected flow rate. If the refill time is
too long, excessive pulsations in the flow may be observed. This
parameter is programmable by pump module.
NOTE: The activation of the scrolling lists A, B…D depends on the type of the pump
previously configured in the Galaxie Configuration Manager (isocratic, binary,
ternary or quaternary pump).
Program a flow and/or a composition gradient. Each step of the
program includes time, flow and/or composition new values. The
Pumps Gradient Chart represents the evolution of these
parameters.
Use the following buttons to edit the table:
to add a line to the table.
to delete a line in the table.
to clear the entire table.
Below the pump gradient chart, select the display type of the
chart. Two choices are available Solvent A on top or Solvent A
on bottom.
Enter the minimum and maximum pressure limits of the pump.
Outside these limits, the pump will stop and an alarm will sound.
NOTE:
- The maximum pressure limit of a ProStar / PrepStar module depends on the type of its
pump head. This limit is automatically adapted according to the connected pump head.
To know these limits, refer to the Pressure Limits chapter in the appendix of this manual.
- If a higher value of pressure is entered by error, a message appears indicating the
pressure limits for the connected pump head and the last entered valid value is
displayed.
- If several ProStar / PrepStar modules are connected (in case of binary, ternary or
quaternary pump), the maximum pressure limit is the lowest value among the different
pump ones.
This window becomes active only when a pump module is
configured as an injector in the system (for more details refer to
the Configuration of the ProStar / PrepStar pump chapter).
Thru Pump Injection Parameter Group:
Enter the time when the injection will start in the Injection time
field.
Enter the flow rate of the injection in the Injection Rate field. The
end time of the injection is determined according to the flow rate
and the injection volume entered in the Acquisition method
section (or in the Quick Start window or in the Sequence).
NOTE: The Injection Volume entered in the Acquisition part of the method must be
Once the instrument control method has been correctly
programmed, an analysis can be started. The next section briefly
describes how to start an acquisition.
Quick Start (Single Acquisition)
1. Press the button.
In the Quick Start window, select the system, the method,
and optionally the project if connected in "All project" mode.
In the Acquisition parameters tab, required fields like the
Run Name, the Run ID and the Acquisition length must be
filled in before starting an analysis. If the system is equipped
with an autosampler controlled by the Galaxie
Chromatography Data System, the Injection Volume, the Vial
position and the Rack number (if required) must be entered.
Other parameters for each acquisition channel can be
entered in the Channel X tab.
Click on the Start button to start the acquisition if the system
is a single-injector system or click on the Next button if the
system is a multiple injector system. In a multiple injector
system, the same parameters as the ones entered for the
first injector must be entered before starting the acquisition.
Sequence
1. To create a sequence, select the FILE / NEW SEQUENCE
menu, complete the displayed creation wizard screen (refer
to Galaxie Chromatography Data System User’s Guide), by
entering the system name that will perform the analysis, the
number of lines (corresponding to the number of samples)
and the sequence name.
2. The sequence is displayed: complete the information
sequence columns (refer to Galaxie Chromatography Data
System User’s Guide).
The Method Name, the Run Name, the Run ID and the
Acquisition length must be filled in before starting the
analysis. All these fields are automatically imported from the
acquisition part if defined in the method.
3. Start the sequence by clicking on the
NOTE: For more details about how to start an acquisition, refer to the Galaxie
Chromatography Data System User’s Guide.
button.
Running Acquisition Status
Viewing a Running Acquisition
To view a running acquisition, click on the Systems tab then
select the running system. A window is activated which is
composed of two parts: the running chromatogram and the
status window where the driver modules are listed.
Status of the ProStar / PrepStar Pump
Click on the button corresponding to the ProStar /
PrepStar pump driver, and the pump status screens will appear.
It contains four sub-sections: Elution, Alarm, About and Overview. Click on the required sub-section to display the
corresponding parameters that will be used during the
chromatographic run.
A toolbar allows you to easily scroll through the status windows:
to select the first status window.
to move back to the previous status window.
to move forward to the next status window.
to select the final status window.
Elution Window
In this window, the elution parameters (solvent composition,
pressure, flow, etc.) are shown with their current values,
depending on the control method program.
Most of these values can be modified during the acquisition or in
order to start the pump for equilibration for example.
The
, … buttons can be activated
depending on the pump configuration previously done in the
Galaxie Configuration Manager (isocratic, binary, ternary or
quaternary pump).
Click on one of these buttons and the following screen appears:
The Flow and/or the solvent composition (%x) values can be set.
The Ramp parameter is the time for each module to join linearly
the new setting values. Click on OK to validate.
Click on the
button and the same screen as above
will be displayed.
Click on
Click on the
or to start or stop the pump.
button when the pump is running to stop the
progression of the gradient at its current value.
Click on the
button and the progression of the gradient
starts again from its current value to the ramp end point.
NOTE: The chart shown on the right panel of the Elution window represents the progress
When the pressure is out of the limits, the pump stops
automatically and the pump status can be read in this screen.
The light becomes red indicating the nature of the error. It is
possible to define new pressure limits directly from this status
screen, by clicking on the
button.
The same screen than above appears:
Enter the new values of pressure limits. Outside these limits,
the pump stops and the alarm sounds. Click on OK to validate.