Varian FRG-700 Short Operating Instructions

Page 1
vacuum technologies
FRG-700
Inverted
Magnetron Pirani
Gauge
Manual No. tqma48e1 Revision -­November 2008
SHORT OPERATING INSTRUCTIONS
Page 2
FRG-700 Inverted Magnetron Pirani
Gauge
Copyright 2008
Vacuum Technologies
2
Page 3
FRG-700 Inverted Magnetron Pirani Gauge
Warranty
Products manufactured by Seller are warranted against defects in materials and workmanship for twelve (12) months from date of shipment thereof to Customer, and Seller’s liability under valid warranty claims is limited, at the option of Seller, to repair, to re­place, or refund of an equitable portion of the purchase price of the Product. Items expendable in normal use are not covered by this warranty. All warranty replacement or repair of parts shall be limited to equipment malfunctions which, in the sole opinion of Seller, are due or traceable to defects in original materials or workmanship. All obligations of Seller under this warranty shall cease in the event of abuse, accident, alteration, misuse, or neglect of the equipment. In-warranty repaired or replaced parts are warranted only for the remaining unexpired portion of the original warranty period applicable to the repaired or replaced parts. After expiration of the applicable warranty period, Custo­mer shall be charged at the then current prices for parts, labor, and transportation.
Reasonable care must be used to avoid hazards. Seller expres­sly disclaims responsibility for loss or damage caused by use of its Products other than in accordance with proper operating pro­cedures.
Except as stated herein, Seller makes no warranty, express or implied (either in fact or by operation of law), statutory or other­wise; and, except as stated herein, Seller shall have no liability under any warranty, express or implied (either in fact or by oper­ation of law), statutory or otherwise. Statements made by any person, including representatives of Seller, which are inconsis­tent or in conflict with the terms of this warranty shall not be bin­ding upon Seller unless reduced to writing and approved by an officer of Seller.
3
Page 4
FRG-700 Inverted Magnetron Pirani Gauge
Warranty Replacement and Adjustment
All claims under warranty must be made promptly after occur­rence of circumstances giving rise thereto, and must be received within the applicable warranty period by Seller or its authorized representative. Such claims should include the Product serial number, the date of shipment, and a full description of the cir­cumstances giving rise to the claim. Before any Products are returned for repair and/or adjustment, written authorization from Seller or its authorized representative for the return and instruc­tions as to how and where these Products should be returned must be obtained. Any Product returned to Seller for examination shall be prepaid via the means of transportation indicated as acceptable by Seller. Seller reserves the right to reject any war­ranty claim not promptly reported and any warranty claim on any item that has been altered or has been returned by non-accepta­ble means of transportation. When any Product is returned for examination and inspection, or for any other reason, Customer shall be responsible for all damage resulting from improper packing or handling, and for loss in transit, not withstanding any defect or non-conformity in the Product. In all cases, Seller has the sole responsibility for determining the cause and nature of failure, and Seller’s determination with regard thereto shall be final.
If it is found that Seller’s Product has been returned without cause and is still serviceable, Customer will be notified and the Product returned at Customer’s expense; in addition, a charge for testing and examination may be made on Products so returned.
3/1/00
4
Page 5
FRG-700 Inverted Magnetron Pirani Gauge
Product Identification
In all communications with VARIAN, please specify the infor­mation given on the product nameplate. For convenient refer­ence copy that information into the space provided below.
VARIAN Lexington MA 02421 USA
Model: PN: SN: V W
Validity
This document applies to products with the following part num­bers:
FRG700KF25 FRG700CF35 (DN 40 CF-F)
The part number (PN) can be taken from the product nameplate. If not indicated otherwise in the legends, the illustrations in this
document correspond to the gauge with vacuum connection DN 25 ISO-KF. They apply to the gauge with other vacuum connection by analogy.
We reserve the right to make technical changes without prior notice.
All dimensions in mm.
5
(DN 25 ISO-KF)
Page 6
FRG-700 Inverted Magnetron Pirani Gauge
Intended Use
The Inverted Magnetron Pirani Gauge FRG-700 has been designed for vacuum measurement in the pressure range of
-9
5×10
… 1000 mbar.
The Inverted Magnetron Pirani Gauge must not be used for measuring flammable or combustible gases which react in air.
The gauge can be operated in connection with a VARIAN FRG Control Unit, a VARIAN Turbo AG Rack Controller, or with another controller.
Functional Principle
The gauge consists of two separate measurement systems (Pirani and cold cathode system) the signals of which are combined in such a way that one measurement signal is output. The Pirani measurement circuit is always on.
6
Page 7
FRG-700 Inverted Magnetron Pirani Gauge
Contents
Warranty 3 Warranty Replacement and Adjustment 4 Product Identification 5 Validity 5 Intended Use 6 Functional Principle 6
1
Safety 8
1.1 Symbols Used 8
1.2 Personnel Qualifications 8
1.3 General Safety Instructions 9
2 Technical Data 10 3 Installation 17
3.1 Vacuum Connection 17
3.2 Electrical Connection 21
4 Operation 22
4.1 Gas Type Dependence 22
4.2 Ignition Delay 22
4.3 Adjusting the Gauge 23
5 Deinstallation 26 6 Maintenance, Troubleshooting 27 7 Returning the Product 28 8 Disposal 28 9 Literature 29 EC Declaration of Conformity 30
For cross-references within this document, the symbol ( XY) is used, for cross-references to further documents, listed under literature, the symbol ( [Z]).
7
Page 8
FRG-700 Inverted Magnetron Pirani Gauge
1 Safety
1.1 Symbols Used
DANGER
Information on preventing any kind of physical injury.
WARNING
Information on preventing extensive equipment and environ­mental damage.
Caution
Information on correct handling or use. Disregard can lead to malfunctions or minor equipment damage.
Notice
1.2 Personnel Qualifications
Skilled personnel
All work described in this document may only be carried out by persons who have suitable technical training and the neces­sary experience or who have been instructed by the end-user of the product.
8
Page 9
FRG-700 Inverted Magnetron Pirani Gauge
1.3 General Safety Instructions
Adhere to the applicable regulations and take the necessary precautions for the process media used.
Consider possible reactions with the product materials. Consider possible reactions (e.g. explosion) of the process
media due to the heat generated by the product.
Adhere to the applicable regulations and take the necessary precautions for all work you are going to do and consider the safety instructions in this document.
Before beginning to work, find out whether any vacuum com­ponents are contaminated. Adhere to the relevant regulations and take the necessary precautions when handling contamin­ated parts.
DANGER
DANGER: magnetic fields Strong magnetic fields can disturb electronic
devices like heart pacemakers or impair their function.
Maintain a safety distance of 10 cm between the magnet and the heart pacemaker or prevent the influence of strong magnetic fields by antimagnetic
Communicate the safety instructions to all other users.
shielding.
9
Page 10
FRG-700 Inverted Magnetron Pirani Gauge
2 Technical Data
Measuring range (air, N2) 5×10-9 … 1000 mbar Accuracy
Repeatability
Output signal (measuring signal)
Voltage range 0 … +10.5 V Measurement range 1.82 … 8.6 V Voltage vs. pressure logarithmic , 0.6 V/decade
Error signal <0.5 V no supply
Output impedance 2×10 Ω Minimum loaded impedance 10 k Response time
-6
p > 10
mbar
-8
mbar 1000 ms
p = 10
Identification gauge
Status pin 6
-2
p > 10
mbar
Pirani-only mode
-2
p < 10
mbar Cold cathode not ignited Pirani-only mode
-2
mbar
p < 10 Cold cathode ignited Combined Pirani / cold cathode mode
LED High voltage on (LED on)
±30% (in the range
-8
… 100 mbar)
1×10
±5% (in the range
-8
1×10
… 100 mbar)
>9.5 V Pirani sensor de­fective (filament rupture)
Ω, short-circuit proof
(pressure dependent) <10 ms
85 kΩ, referenced to supply common
Low = 0 V
Low = 0 V
High = 15 … 30 VDC
10
Page 11
FRG-700 Inverted Magnetron Pirani Gauge
Supply
DANGER
The gauge may only be connected to power sup­plies, instruments or control devices that conform to the requirements of a grounded protective extra­low voltage (SELV-E according to EN 61010). The connection to the gauge has to be fused
1)
.
Voltage at the gauge 15 … 30 VDC
(ripple
1 V
)
1 V
pp
) 2)
pp
-4
mbar
Power consumption
1)
Fuse
Voltage at the supply unit with maximum cable length
Adjustment
Potentiometer <HV> Potentiometer <ATM>
2 W
1 AT
16 … 30 VDC (ripple
adjustment under 10 adjustment at atmospheric pressure
Electrical connection FCC68 socket, 8 poles Sensor cable 8 poles, shielded Line length
50 m (8×0.14 mm²)
Operating voltage 3.3 kV Operating current
Grounding concept
Vacuum connection –measuring common
500 µA
"Electrical Connection"
connected via 10 kΩ (max. voltage differential with respect to safety ±50 V accuracy ±10 V)
Supply common – signal common conducted separately
1)
VARIAN controllers fulfill this requirement.
2
)
The minimum voltage of the power supply unit must be increased
proportionally to the length of the sensor cable
11
Page 12
FRG-700 Inverted Magnetron Pirani Gauge
Materials on the vacuum side
Vacuum connection Measurement chamber Feedthrough isolation Internal seal Anode Ignition aid Pirani measurement tube Pirani filament
Internal volume Pressure
stainless steel stainless steel ceramic FPM 75 Mo stainless steel Ni, Au W
20 cm³ 10 bar (absolute),
limited to inert gases
Temperatures
Operation Bakeout
Pirani filament Storage
Relative humidity
+5 … +55 °C 150 °C (without electronics and magnetic shielding) 120 °C
-40 … +65 °C
80% at temperatures +31°C decreasing to 50%
at +40°C
Mounting orientation
any
Use indoors only, altitude up to
2000 m
Type of protection IP 40
12
Page 13
FRG-700 Inverted Magnetron Pirani Gauge
Dimensions [mm]
75
60
92
28 55
18
DN 25 ISO-KF
Weight
FRG700KF25 FRG700CF35
ø 63.5
24
DN 40 CF-F
700 g 980 g
13
Page 14
FRG-700 Inverted Magnetron Pirani Gauge
Measuring Signal vs. Pressure
Pressure p
1E+04
1E+02
1E+00
1E–02
underrange
sensor error
1E–04
1E–06
1E–08
0.00.5 1.5 2. 5 3.5 4.5 5.5 6.5 7.5 8.5 9.5 10.5
p = 10
1.667U-d
p = 10
U [V]
Measurement signal U [V]
1.667×U-d
U = c + 0.6log10 p
mbar Pa Torr
d 11.33 9.33 11.46 c 6.8 5.6 6.875
valid in the range 5×10
-9
mbar <p< 1000 mbar
-9
Torr <p< 750 Torr
3.8×10
-7
Pa <p< 1×105 Pa
5×10
overrange
Pa
mbar
torr
sensor error
14
Page 15
FRG-700 Inverted Magnetron Pirani Gauge
Gas Type Dependence
p (mbar)
2
10
8 6
4
2
1
10
8
Indication range
6 4
2
0
10
8 6
4
2
–1
10
8 6
4
2
–2
10
–3
10
Water vapor
2
46
>10
-2
mbar)
–2
10
–1
10
46
2
2
46
10
0
2
H2He Ne
46
Air O2 CO N2
CO
2
Ar
Freon 12 Kr
Xe
1
10
2
10
46
2
p
(mbar)
eff
15
Page 16
FRG-700 Inverted Magnetron Pirani Gauge
p (mbar)
–1
10
8 6
4
2
Indication range
–2
10
8 6
4
2
10
8 6
4
2
10
8 6
4
2
10
8 6
4
2
10
In the range below 10
–3
–4
–5
–6
10
-6
246
-6
-1
... 10
10
mbar
Ar
Air N2 O2 CO
He
-6
10
–4
-3
246
10
-5
10
246
mbar, the pressure indication is linear.
246
10
-2
246
p
eff
-1
10
(mbar)
For gases other than air, the pressure can be determined by means of a simple conversion formula:
p
= K × pressure reading
eff
Where Gas type K (mean values)
Air (O2, CO, N2)
Xe Kr Ar H
2
Ne He
1.0
0.4
0.5
0.8
2.4
4.1
5.9
16
Page 17
FRG-700 Inverted Magnetron Pirani Gauge
3 Installation
3.1 Vacuum Connection
DANGER
DANGER: overpressure in the vacuum system >1 bar
Injury caused by released parts and harm caused by escaping process gases can result if clamps are opened while the vacuum system is pressurized.
Do not open any clamps while the vacuum system is pressurized. Use the type clamps which are suited to overpressure.
DANGER
DANGER: overpressure in the vacuum system >2.5 bar
KF flange connections with elastomer seals (e.g. O-rings) cannot withstand such pressures. Process media can thus leak and possibly damage your health.
Use O-rings provided with an outer centering ring.
17
Page 18
FRG-700 Inverted Magnetron Pirani Gauge
DANGER
DANGER: protective ground Products that are not correctly connected to ground
can be extremely hazardous in the event of a fault. Electrically connect the gauge to the grounded
vacuum chamber. This connection must conform to the requirements of a protective connection ac­cording to EN 61010:
CF flanges fulfill this requirement.
For gauges with a KF flange, use a conductive
metallic clamping ring.
18
Caution: vacuum component Dirt and damages impair the function of the vac-
uum component. When handling vacuum components, take appro-
priate measures to ensure cleanliness and prevent damages.
Caution: dirt sensitive area Touching the product or parts thereof with bare
hands increases the desorption rate. Always wear clean, lint-free gloves and use clean
tools when working in this area.
Caution
Caution
Page 19
FRG-700 Inverted Magnetron Pirani Gauge
WARNING
WARNING: electric arcing Helium may cause electric arcing with detrimental
effects on the electronics of the product. Before performing any tightness tests put the pro-
duct out of operation and remove the electronics unit.
The gauge may be mounted in any orientation. To keep
condensates and particles from getting into the measur­ing chamber preferably choose a horizontal to upright position and possibly use a seal with a centering ring and filter.
If adjustment should be possible after the gauge has
been installed, be sure to install it so that potentiometers <HV> and <ATM> can be accessed with a screwdriver (
"Adjusting the Gauge").
When making a CF flange connection, it can be ad-
vantageous to temporarily remove the electronics and the magnet unit (
Instruction Manual tqna48e1).
19
Page 20
FRG-700 Inverted Magnetron Pirani Gauge
Remove the protective lid and connect the product to the vac­uum system.
Seal with
centering ring
or
Protective lid
20
Clamp
Keep the protective lid.
Seal with centering ring and filter
Page 21
FRG-700 Inverted Magnetron Pirani Gauge
3.2 Electrical Connection
Make sure the vacuum connection is properly made
( "Vacuum Connection").
n If no sensor cable is available, make one according to the
following diagram.
10 k
Signal
Ident
6
10
10
3
5
4
1
2
+
+ –
1
Electrical connection Pin 1 Supply (
Pin 2 Supply common Pin 3 Signal output (measuring signal) Pin 4 Identification Pin 5 Signal common Pin 6 Status Pin 7, 8 n.c.
15 … 30 VDC
8
)
8 poles FCC-68
o Connect the gauge to the controller using the sensor cable.
21
Page 22
FRG-700 Inverted Magnetron Pirani Gauge
4 Operation
When the supply voltage is applied, the measuring signal is available between pins 3 and 5. Over the whole measurement range, the measuring signal is output as a logarithm of the pressure (measuring signal vs. pressure "Technical Data").
Allow for a stabilizing time of 10 minutes. Once the gauge has been switched on, permanently leave it on irrespective of the pressure.
The Pirani measurement circuit is always on.
The cold cathode measurement circuit is controlled by the
Pirani circuit and is activated only at pressures <1×10
4.1 Gas Type Dependence
The measurement value depends on the type of gas being measured. The value displayed is accurate for dry air, O
. It can be mathematically converted for other gases
and N
2
( "Technical Data"). If the gauge is operated in connection with an VARIAN vacuum
gauge controller, a calibration factor can be entered for correction of the reading.
4.2 Ignition Delay
When cold cathode measurement systems are activated upon switching the gauge on, an ignition delay occurs, which is typically:
As long as the cold cathode measurement circuit has not yet ignited, the measurement value of the Pirani is output as measuring signal ("Pirani underrange" is displayed for pressures <5×10
-5
10
mbar 1 second
-7
mbar 20 seconds
10
-9
5×10
mbar 2 minutes
-4
mbar).
-2
mbar.
, CO
2
22
Page 23
FRG-700 Inverted Magnetron Pirani Gauge
4.3 Adjusting the Gauge
The gauge is factory-calibrated. If used under different climatic conditions, through extreme temperatures, aging or contamina­tion, and after exchanging the sensor, the characteristic curve can be offset and readjustment may become necessary.
The cold cathode measurement circuit, which is dominant for low pressures (<1×10 trast, the Pirani measurement circuit can be adjusted. Any ad­justment has a negligible effect on the pressure range between approx. 10
-3
mbar), is factory-calibrated. By way of con-
-2
mbar and 102 mbar.
n If you are using a seal with centering ring and filter, check
that they are clean or replace them if necessary ( "Dein­stallation").
o Activate the gauge.
p Evacuate it to p << 10
-4
mbar and wait at least 10 minutes.
q Turn the nameplate counter-clockwise until the mechanical
stop is reached.
23
Page 24
FRG-700 Inverted Magnetron Pirani Gauge
r
While depressing the tactile switch with a cylindrical pin
3 mm), adjust the <HV> potentiometer by means of a
… to 4.20 V or … to 5×10
1.5 mm screwdriver …
s Vent the gauge with air or nitrogen to atmospheric pres-
sure, and wait at least 10 minutes.
t Turn the nameplate clockwise until the mechanical stop is
reached.
24
-4
mbar
Page 25
FRG-700 Inverted Magnetron Pirani Gauge
u
Using the 1.5 mm screwdriver, adjust the <ATM>
potentiometer …
… to 8.60 V or … to 1×10
3
mbar
v Turn the nameplate back to its original position (it catches).
25
Page 26
FRG-700 Inverted Magnetron Pirani Gauge
5 Deinstallation
DANGER
DANGER: contaminated parts Contaminated parts can be detrimental to health
and environment. Before beginning to work, find out whether any
parts are contaminated. Adhere to the relevant regulations and take the necessary precautions when handling contaminated parts.
n Vent the vacuum system.
26
Caution
Caution: vacuum component Dirt and damages impair the function of the vac-
uum component. When handling vacuum components, take appro-
priate measures to ensure cleanliness and prevent damages.
Caution
Caution: dirt sensitive area Touching the product or parts thereof with bare
hands increases the desorption rate. Always wear clean, lint-free gloves and use clean
tools when working in this area.
Page 27
FRG-700 Inverted Magnetron Pirani Gauge
o Put the gauge out of operation and unplug the sensor
cable.
p Remove the gauge from the vacuum system and place the
protective lid.
When deinstalling a CF flange connection, it can be ad-
vantageous to temporarily remove the electronics and the magnet unit ( Instruction Manual tqna48e1).
6 Maintenance, Troubleshooting
Instruction Manual tqna48e1
If operated at high pressures or under dirty conditions,
the gauge must be regularly cleaned. Gauge failures due to contamination, as well as expend-
able parts (filament), are not covered by the warranty.
27
Page 28
FRG-700 Inverted Magnetron Pirani Gauge
7 Returning the Product
WARNING
WARNING: forwarding contaminated products Contaminated products (e.g. radioactive, toxic,
caustic or microbiological hazard) can be detrimen­tal to health and environment.
Products returned to VARIAN should preferably be free of harmful substances. Adhere to the for­warding regulations of all involved countries and forwarding companies and enclose a duly com­pleted declaration of contamination.
Products that are not clearly declared as "free of harmful sub­stances" are decontaminated at the expense of the customer.
Products not accompanied by a duly completed declaration of contamination are returned to the sender at his own expense.
8 Disposal
DANGER
DANGER: contaminated parts Contaminated parts can be detrimental to health
and environment. Before beginning to work, find out whether any
parts are contaminated. Adhere to the relevant regulations and take the necessary precautions when handling contaminated parts.
28
Page 29
FRG-700 Inverted Magnetron Pirani Gauge
WARNING
N
WARNING: substances detrimental to the environ­ment
Products or parts thereof (mechanical and electric components, operating fluids etc.) can be detrimen­tal to the environment.
Dispose of such substances in accordance with the relevant local regulations.
Separating the components
After disassembling the product, separate its components ac­cording to the following criteria:
Contaminated components Contaminated components (radioactive, toxic, caustic or bio-
logical hazard etc.) must be decontaminated in accordance with the relevant national regulations, separated according to their materials, and disposed of.
Other components Such components must be separated according to their ma-
terials and recycled.
9 Literature
[1] www.varianinc.com
Instruction Manual Inverted Magnetron Pirani Gauge FRG-700 tqna38e1 Varian Vacuum Technologies, MA 02421, USA
29
Page 30
FRG-700 Inverted Magnetron Pirani Gauge
EC Declaration of Conformity
We, VARIAN, hereby declare that the equipment mentioned below complies with the provisions of the Directive relating to electrical equipment designed for use within certain voltage limits 2006/95/EC and the Directive relating to electromagnetic compa­tibility 2004/108/EC.
FRG-700 Inverted Magnetron Pirani Gauge
Part numbers
FRG700KF25 FRG700CF35
Standards
Harmonized and international/national standards and specifi­cations:
EN 61000-6-2
EN 61000-6-3
EN 61010-1
for measurement, control and laboratory use)
Signatures
Varian Vacuum Technologies, USA
9 December 2008
(EMC: generic immunity standard)
(EMC: generic emission standard)
(Safety requirements for electrical equipment
Frederick C. Campbell Operations Manager
30
Page 31
FRG-700 Inverted Magnetron Pirani Gauge
Notes
31
Page 32
Varian Vacuum Technologies 121 Hartwell Avenue Lexington, MA, 02421 USA Tel: (781) 861 7200 Fax: (781) 861 5437
Original: English custserv@varianinc.com
t qma48e1
www.varianinc.com